Siemens Healthcare Diagnostics Products GmbH

P. O. Box 1149
35001 Marburg, Germany
diagnostics.siemens.com

Certificate of Traceability
Assigned Values

This is to certify the traceability of the assigned values of the product listed below to a reference material.

The uncertainty of the assigned value for this product was calculated according to the ISO norm 17511 by using
a coverage factor of two. Expressing the uncertainty by a coefficient of variation can be done by dividing the
given uncertainty by two.

Product Standard Human Plasma

Product Code ORKL

Parameter Fibrinogen (with Dade Thrombin)

Unit g/L

Device for usage of assigned value all

Reference material WHO 09/264

Measuring method used for value assignment Nephelometry, Turbidimetry

Uncertainty of assigned values [%] 2.7

This document was produced by means of an electronic system, therefore without signature

Name of document: D-00064 Effective date: 2013-06-26 Page 1 of 2


Siemens Healthcare Diagnostics Products GmbH
P. O. Box 1149
35001 Marburg, Germany
diagnostics.siemens.com

Traceability Chart
This is to certify the traceability chain of the assigned values of the product listed below to a reference material.

Product Standard Human Plasma

Product Code ORKL

Parameter Fibrinogen (with Dade Thrombin)

Unit g/L

Materials & Procedures Section1)

a
International Standard WHO 09/264
⎯⎯⎯⎯→
Primary Calibration of Standard
Human Plasma Master-Lot with b
reference material
←⎯⎯⎯⎯
Master Calibrator Standard Human
b
Plasma Lot defined
⎯⎯⎯⎯→
Secondary Calibration of new
b
Production Lot vs. Master-Lot
←⎯⎯⎯⎯
Standard Human Plasma-Lot
b
calibrated; value assigned
⎯⎯⎯⎯→
Customer: use of assigned value on
c
instrument

-
1)
a = External material / procedure
b = Company internal
c = Use of assigned value by customer

Name of document: D-00064 Effective date: 2013-06-26 Page 2 of 2