You are on page 1of 12

Schedule

Issue date: 18 November 2015


Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 1 of 12

LABORATORY LOCATION: PYROMETRO SERVICES (M) SDN. BHD.


(PERMANENT LABORATORY) PMT 2927, JALAN MEGAT HARUN
PUSAT PERNIAGAAN MAJU
14000 BUKIT MERTAJAM
PULAU PINANG, MALAYSIA

This laboratory accredited under Skim Akreditasi Makmal Malaysia (SAMM) meets the
requirements of MS ISO/IEC 17025:2005 ‘General requirements for competence of testing and
calibration laboratories’. This Malaysian Standards is identical with ISO/IEC 17025:2005 published
by the International Organization for Standardization (ISO).

* The expanded uncertainties are based on an estimated confidence probability of approximately 95%
and have a coverage factor of k=2 unless stated otherwise.

FIELD OF CALIBRATION: TEMPERATURE

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Comparison with Pt 100


Liquid-in-glass -30 C to 0 C 0.05 C reference in: liquid bath
thermometer 0 C to 200 C 0.04 C and temperature block
(total & partial immersion) 200 C to 400 C 0.35 C calibrator based on
ASTM E77-98

-30 C to 0 C 0.062 C
Temperature sensor 0 C to 250 C 0.057 C
(Platinum resistance 250 C to 400 C 0.36 C
thermometer, 400 C to 600 C 1.0 C Comparison with Pt 100/
thermocouple, thermistor) 600 C to 1000 C 2.1 C Type R thermocouple
reference in: liquid bath
1000 C to 1200 C 2.7 C
and temperature block
calibrator based on JIS
Temperature sensor with C1602-95, JIS C1604-
-30 C to 0 C 0.04 C 97, JIS C1611-95 and
indicator
0 C to 250 C 0.033 C JIS Z8710-93
(Platinum resistance
250 C to 400 C 0.36 C
thermometer,
thermocouple, 400 C to 600 C 0.77 C
thermistor, temperature 600 C to 1000 C 2.2 C
gauge) 1000 C to 1200 C 2.7 C

Comparison with
Thermohygrometer /
Pt 100 based on CSIRO
Temperature & Humidity 10 oC to 50 oC 0.1 oC ‘Temperature and
Indicator or Recorder 20 %RH to 90 %RH 2.1 %RH Humidity Measurement’
Course Book 3, JIS
B7306-1989 and JIS
Z8806-1995

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 2 of 12

FIELD OF CALIBRATION: TEMPERATURE

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Temperature calibrator
I) Simulation
a) Type K -270 C to -100 C 1.0 C
-100 C to 1370 C 0.1 C

b) Type N -260 C to -100 C 1.1 C


-100 C to 1300 C 0.1 C

c) Type J -210 C to 1200 C 0.1 C

d) Type T -270 C to -100 C 0.6 C


-100 C to 400 C 0.1 C By electrical simulation
using multimeter &
reference table ITS 90
e) Type E -270 C to -100 C 0.6 C
based on JIS Z8704-93
-100 C to 1000 C 0.1 C

f) Type R -50 C to 500 C 0.3 C


500 C to 1760 C 0.2 C

g) Type S -50 C to 500 C 0.3 C


500 C to 1760 C 0.2 C

h) Type B 100 C to 500 C 1.1 C


500 C to 1800 C 0.3 C

i) Pt 100 -200 C to 850 C 0.1 C

II) Measurement
a) Type K -270 C to -100 C 1.1 C
-100 C to 1370 C 0.2 C

b) Type N -260 C to -100 C 1.1 C


-100 C to 1300 C 0.2 C

c) Type J -210 C to -100 C 0.2 C By electrical simulation


using calibrator &
d) Type T -270 C to -100 C 0.6 C reference table ITS 90
-100 C to 400 C 0.2 C based on JIS Z 8704-93

e) Type E -270 C to -100 C 0.6 C


-100 C to 1000 C 0.2 C

f) Type R 0 C to 1760 C 2 C

g) Type S 0 C to 1760 C 2 C

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 3 of 12

FIELD OF CALIBRATION: TEMPERATURE

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Temperature calibrator
II) Measurement (cont.)
h) Type B 600 C to 1800 C 2 C By electrical simulation
using calibrator &
i) Pt 100 -200 C to 850 C 0.2 C reference table ITS 90

Temperature measuring
device (by electrical
simulation)
a) Type K -270 C to -100 C 1.1 C
-100 C to 1370 C 0.2 C

b) Type N -200 C to -100 C 1.1 C


-100 C to 1300 C 0.2 C

c) Type J -210 C to 1200 C 0.2 C


By electrical simulation
using calibrator &
d) Type T -270 C to -100 C 0.6 C
reference table ITS 90
-100 C to 400 C 0.2 C
based on JIS C1601-83
and JIS C1603-83
e) Type E -250 C to -100 C 0.6 C
-100 C to 1000 C 0.2 C

f) Type R 0 C to 1760 C 2 C

g) Type S 0 C to 1760 C 2 C

h) Type B 600 C to 1800 C 2 C

i) Pt 100 -200 C to 850 C 0.2 C

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 4 of 12

FIELD OF CALIBRATION: TEMPERATURE

SITE CALIBRATION: CATEGORY I

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Temperature sensor
(Platinum resistance 0 C to 200 C 0.2 C
thermometer, 200 C to 400 C 0.4 C Comparison with Pt 100
thermocouple, thermistor) reference in: liquid bath
and temperature block
calibrator based on JIS
Temperature sensor with
C1602-95, JIS C1604-
indicator
97, JIS C1611-95 and
(Platinum resistance 0 C to 200 C 0.08 C
JIS Z8710-93
thermometer, 200 C to 400 C 0.33 C
thermocouple, thermistor,
temperature gauge)

Temperature measuring
device (by electrical
simulation)
a) Type K -270 C to -100 C 1.1 C
-100 C to 1370 C 0.2 C

b) Type N -200 C to -100 C 1.1 C


-100 C to 1300 C 0.2 C

c) Type J -210 C to 1200 C 0.2 C By electrical simulation


using calibrator &
d) Type T -270 C to -100 C 0.6 C reference table ITS 90
-100 C to 400 C 0.2 C based on JIS C1601-83
and JIS C1603-83
e) Type E -250 C to -100 C 0.6 C
-100 C to 1000 C 0.2 C

f) Type R 0 C to 1760 C 2 C

g) Type S 0 C to 1760 C 2 C

h) Type B 600 C to 1800 C 2 C

i) Pt 100 -200 C to 850 C 0.2 C

-80 C to -30 C 0.7 C Calibrate by using


-30 C to 100 C 0.2 C temperature recorder
Temperature controlled
100 C to 250 C 0.6 C with Pt 100 /
enclosure
250 C to 600 C 1.7 C thermocouple based on
600 C to 1300 C 2.0 C AS 2853-1986

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 5 of 12

FIELD OF CALIBRATION: TEMPERATURE

SITE CALIBRATION: CATEGORY I

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Calibrated by using
Pt 100 /
thermohygrometer
5 C to 60 C 0.2 C based on AS 2853-1986
Humidity enclosure
20 %RH to 95 %RH 2.4 %RH and CSIRO
‘Temperature and
Humidity Measurement’
Course Book 3

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 6 of 12

FIELD OF CALIBRATION: DIMENSIONAL

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*
Calibrated by using
0 mm to 200 mm 0.006 mm caliper checker / gauge
Precision caliper (internal
200 mm to 400 mm 0.007 mm block as standards
& external)
400 mm to 600 mm 0.008 mm according to JIS
B7507:1993

Calibrated by using
gauge block and optical
0 mm to 25 mm 0.002 mm
Micrometer (external) parallel as standards
25 mm to 150 mm 0.003 mm
according JIS
B7502:1994

Calibrated by using
caliper checker & dial
Up to 300 mm 0.007 mm test indicator and
Height gauge
300 mm to 600 mm 0.008 mm standard square as
standard according to
JIS B 7517:1993

Calibrated by using
gauge block as
1 µm
Dial thickness gauge Up to 20 mm standards according to
JIS B 7503:1997 clause
7.1

Calibrated by using dial


gauge tester and
Up to 25 mm 1.0 µm
Dial gauge micrometer head as
25 mm to 50 mm 1.7 µm
standards according to
JIS B 7503:2011

Calibrated by using
Up to 0.28 mm 1.5 µm
gauge block and dial
Dial test indicator 0.28 mm to 0.6 mm 1.8 µm
gauge tester according
0.6 mm to 1.5 mm 5.0 µm
to JIS B 753:1990

Calibrated by using
gauge block as
Digital indicator Up to 50 mm 2.0 µm
standards according to
JIS B 7503:2011

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 7 of 12

FIELD OF CALIBRATION: DIMENSIONAL

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Calibrated by using
Digital Micrometer as
Feeler Gauge 0.01 mm to 1 mm 2.2 µm
standards according to
BS 957: 2008

Calibrated by using 3
External Parallel Screw
Major Diameter: 2.0 µm wire pin set scale and
Thread Plug Gauge
2 mm to 24 mm Simple Pitch Diameter: Micrometer as standard
(Simple Pitch Diameter
3.0 µm according to JIS B
and Major Diameter)
0261:2004

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 8 of 12

FIELD OF CALIBRATION: DIMENSIONAL

SITE CALIBRATION: CATEGORY I

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Calibrated by using
Up to 600 mm X 600 mm
2.1 m beam comparator
Surface Plate Above 600 mm
Up to 2500 mm X 1600 mm 3.6 m micro indicator dial
gauge block

Toolmaker’s Microscope Calibrated by using


(Independent Linear Scale 0 mm to 140 mm 1.5 m glass scale according
Only) to JIS B 7184:1999

Profile projector Calibrated by using


(Independent linear scale 0 mm to 140 mm 1.5 m glass scale according
only) to JIS B7184:1999

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 9 of 12

FIELD OF CALIBRATION: MASS

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Calibrated using
standard weight and
Standard Weight 2 kg to 25 kg 0.2 g comparator as standard
according to OIML
R111-1:2004 (E)

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

FIELD OF CALIBRATION: MASS

SITE CALIBRATION: CATEGORY I

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*
Up to 200 g 0.2 mg
Calibrated using
200 g to 1000 g 1.2 mg
standard weight as
1 kg to 5 kg 8 mg
standard according to
Balance 5 kg to 20 kg 0.4 g
ASTM E 898-88 and
20 kg to 50 kg 0.7 g
LAB 14 EDITION 2006-
50 kg to 100 kg 2.9 g
off centre loading
100 kg to 300 kg 6g

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 10 of 12

FIELD OF CALIBRATION: PRESSURE

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*
Pressure Measuring
Device
Vacuum Up to –0.95 bar 0.006 bar
Calibrated using test
Pneumatic 0 mbar to 70 mbar 0.012 mbar gauge as standard
70 mbar to 700 mbar 0.14 mbar according to AS 1349
0 bar (abs) to 7 bar (abs) 0.0011 bar (abs) (1986) and DKD-R-6-1
0.7 bar to 30 bar 0.0051 bar (edition 01/2003)

Hydraulic 0 bar to 350 bar 0.06 bar


350 bar to 700 bar 0.12 bar

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

FIELD OF CALIBRATION: PRESSURE

SITE CALIBRATION: CATEGORY I

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*
Pressure Measuring
Device
Vacuum Up to -0.95 bar 0.006 bar
Calibrated using test
Pneumatic 0 mbar to 70 mbar 0.012 mbar gauge as standard
70 mbar to 700 mbar 0.14 mbar according to AS 1349
0 bar (abs) to 7 bar (abs) 0.0011 bar (abs) (1986) and DKD-R-6-1
0.7 bar to 30 bar 0.0051 bar (edition 01/2003)

Hydraulic 0 bar to 350 bar 0.06 bar


350 bar to 700 bar 0.12 bar

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 11 of 12

FIELD OF CALIBRATION: FORCE

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Calibrated by using dead


0 N to 4.9 N 0.001 N
Force measuring device weight with reference to
4.9 N to 980 N 0.02 N
ISO 376:2011

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

FIELDS OF CALIBRATION: FORCE

SITE CALIBRATION: CATEGORY I

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Up to 100 kgf 0.006 kgf


Calibrated by using load
Compression / tensile 100 kgf to 1000 kgf 0.8 kgf
cell based on ISO 7500-
testing machine 1000 kgf to 5000 kgf 3 kgf
1:2004
5000 kgf to 10000 kgf 8 kgf

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 18 November 2015
Valid until: 12 April 2018
MS ISO/IEC 17025
NO: SAMM 325
(Issue 3, 18 November 2015 replacement
of SAMM 325 dated 8 June 2015)
Page: 12 of 12

FIELDS OF CALIBRATION: TORQUE

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

0 Nm to 20 Nm 0.29 % of reading
Calibrate by using
Torque Measurement 20 Nm to 150 Nm 0.20 % of reading
torque calibrator based
Device 150 Nm to 400 Nm 0.10 % of reading
on ISO 6789:2003
400 Nm to 1000 Nm 0.14 % of reading

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

FIELD OF CALIBRATION: TIME & FREQUENCY

SITE CALIBRATION: CATEGORY I

SCOPE OF ACCREDITATION:

Calibration and
Instrument Calibrated/ Measurement
Range Remarks
Measurement Parameter Capability Expressed
as an Uncertainty (±)*

Stop watch (max. 1/100 s) 0 s to 300 s 0.19 s


300 s to 3600 s 0.60 s Calibrate by using
3600 s to 10800 s 1.8 s comparison method
according to NIST stop
Timer (max. 1/100 s) 0 s to 300 s 0.10 s watch and timer
300 s to 3600 s 0.58 s calibration
3600 s to 10800 s 1.7 s

RPM Meter
Up to 5000 rpm 1.0 rpm Calibrated using
Contact
60 rpm to 20000 rpm 1.0 rpm tachometer
Non-contact

Signatories:

1. Hasnas binti Hussain


2. ** Teo Hun Wei
** Non-resident signatory

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA

You might also like