Professional Documents
Culture Documents
9-500-0169
(
M-IV
Mammography System
Operator's Manual
9-500-0169
Revision 12
CHAPTER
Table of Co,.tents
c Setup Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . • . • . . • . • . • . . • . . . . . . ..
EXPOSURE TECHNIQUE DEFAULTS .••.. •••••••••••••••• • • .•• • 66
65
OVERVIEW: I Chapter 1 provides a brief introduction to the M-IV, and familiarizes the
user with the following :
c OBJECTIVES: I After reading this chapter, the user will be able to:
Chap-ter 1:
Introduction & Specifications
(
(
Introduction & Speci[fi<,atio"15
c Manual Objectiues
This manual serves a two-fold purpose:
•
•
R.l8 eempatible
1. Display Screen
Figure I-I:
2. Input Control Panel The M-IV Operator Console
3. AutoFilm ID
4. Emergency OFF Switch
5. Radiation Shield
6. Floppy Disk Drive
7. Cassette Storage (both sides)
8. Input Power I Data Cable Connector Panel
9. Bar Code Reader (option)
The Gantry
The Gantry houses the high voltage electronics which
produce x-rays at the energy levels needed for mammog-
raphy. The mechanical assemblies which position the
C-arrn (height and rotation), the input transfonner, the
fuse panel, and most of the system electronics are also in
the Gantry. LED readouts on the left and right side of
the Gantry show the C-arm angle.
Legend - Gantry
Use the legend below as a guide for component identifi-
cation.
• Full Pass Linear Sucley Device (24cm x 3Ocm) • 15cm Localization Paddle (open)
c •
•
•
lOcm Compression Paddle (contact & mag)
Face Shield
Radiation Shield
• StereoLoc IT Breast Biopsy Systsm
(
M-N Operator's Manual 7
Introduction & Specifications
OAHTAY
MAONIFICATION
TABLE CONSOLE
BUCKVS
~~
c -
- --D
----IiI
----IiI
---D
C""~"J
LABELl
~LABEL7
LABEL 3: UlCompilance
[®' ®' 1
,"'-
IMPORTANT I When replacing a certified component that Is concealed by a surface (i.e., cover, panel, etc).
the component's duplicate /abel on the exterior of the unit MUST BE replaced at file time of service.
IMPORTANT 1 When replacing a surface (I.e., cover, panel, etc.) that has a certified component duplicate /sbe/(s),
a tabel(s) that mpresents the certified component /abe/(s) MUST BE placed on the new surface
at the time of service.
F;gure 1-3:
Legend - lAbel Locations
• InputLlne
• Standby Power Consumption Surge, fast transient/burst
0.75 kVA nominal
Unit Measurements
• Gantry with C-ann
Height - 75" (190.5cm)
Width - 25.25" (M.2em) - C-arm at 0° position
Depth - 44.5" (I I3ern)
Weight - 866 Ibs (389.7 kg) - crated
• Remote Console
Height - 36" (91.5cm) at front; 42"(106.7cm) at rear
Width - 31" (78.74cm) at radiation shield
Depth - 17" (43. 1Scrn) including radiation shield b
Weight - 218 Ibs (98.1 kg) - crated i!2 g
~
• Radiation Shield
Height - 75" (190.5cm) maximum height installed
Width - 31" (78.74cm)
Ph equivalent - O.5mm Ph equivalent (minimum)
Figure 1-4:
M-IV Dimensions
• Vertical Travel
26" to 55" (66cm to 140cm) from Bucky surface to • Compression Controls
UplDown controls on both sides of C-arm and on 2-
floor for C-ann positions at both 0° and 90° (ACR I
position footswitch (Motorized). Handwheel on
CDC). Movement controls on both sides of C-arm
both sides of Compression Carriage (Manual).
and on footswitch.
• Magnification
1.8x for objects 22.5mm above the Magnification
mode raises Compression Device upon ex.posure
termination available via Setup Mode data field.
Table's breast support surface (breast support surface
provides approximately 1.7x. magnification). • Compression Force Display
LED Display on Compression Carriage shows the (
• Image Receptor Support Device
Houses image receptor accessory detectors and the
compression force through the range of 10lbs to
65Ibs (44.5N to 289N) in lIb (IN) increments.
AEC sensors. Stereotactic configurations: remov-
able via mechanical locking apparatus. Electrical
interlocks prevent C-arm motion and x-ray exposure
• Compression Force Display Accuracy
±31bs from 10lbs to 35 Ibs
(±13.3SN from 44.5N to 155.75N)
when device is removed.
±SIbs above 351bs
(±22.25 N above 155.75 N)
Compression Specifications
• Manual Compression Force • Compression Thickness Display
LED Display on Compression Carriage shows the
6S lbs (289.25N), ±7lbs (±31.15 N) distance from the bottom of the Compression Paddle
(under compression) to the surface of the image
• Motorized Pre-Compression Force
151bs to 30lbs (66.75N to 133.5N)
receptor accessory (calibration compensates for
varying accessory thicknesses) in O.lcrn increments.
c • Thermal Characteristics
Anode Heat Storage Capacity - 210 kJ (300 kHU)
Max. Anode Heat Dissipation - 525 W (740 HU/s)
•
•
Inherent 1\J.be Filtration
O.Omm AI equivalent
Added Filtration
• Anode Rotation 2-position filter changer mechanism that houses a 30
180 Hz (Hi-speed rotation)
micron (0.03mm) Molybdenum foil filter. and a 25
micron (O.025mm) Rhodium filter. Filter is user
• Anode Angle selectable in all exposure modes except Auto-Filter
Bi-angular - Large focal spot at 16", Small focal spot
Mode. where it is selected automatically.
at 10°. X-ray tube angle at 6° to provide 22" (Large)
& 16° (Small) anode to film plane angle. Reference
axis at 6° from chestwall for both focal spots. • Beam Quality
With the Compression Paddle in the x-ray beam. the
HVL for MoIMo operation is equal to or greater than
• Anode Material tV/lOO +0.03 (units of mm of aluminum). but less
Molybdenum
than kVpllOO +0.11 (in units of mID of aluminum).
• X-ray Window
Beryllium - O.8mm thickness (maximum) • Radiation Output
Equal to or greater than 800 mRlsecond for at least 3
seconds (output measured through the Compression
Paddle, 4.5cm above the breast support, 4.0cm from
the chest wall edge. using exposure techniques of
MalMo target/filter. large focal spot. 28kV).
(
M -N Operator's Manual 13
Introduction & Specifications
c Image Receptors
• Standard Types
18cm x 24cm cassette holder
18cm x 24cm Bucky with HTC Grid
• Exposure Termination
The AEC system terminates the exposure within 50
milliseconds or within 5mAs if the calculated
exposure time exceeds the "backup" time of 5
seconds.
24cm x 30cm Bucley with Linear Grid
• Density Range
• Optional Types 11 density adjustment steps provide approximately
18cm x 24cm Budey with Linear Grid 12% difference in mAs from each adjacent step.
24cm x 30cm Bucley with HTC Grid
StereoLoc II
DSM System
• Stanup sequence
• Descriptions and use of standard accessories
• Profiles of optional accessories
(
Chap-ter 2:
Accessory Setup & Installation
(
(
Accessory Set Up & lru.ta.ll~,tUm
The Console to
Gantry Connections
When the M-IV is initially installed, all of the necessary
Operator Console-le-Gantry cabling is completed by an
authorized service representative. Therefore. operator DETAIL CircUit Breaker
action is not required regarding cable connections.
c Connecting to Power
The M-IV Gantry receives AC line power from an
appropriately fused "disconnect" box. The Gantry then
provides power for all unit subsystems, the Operator
Console, and all accessories. The following checks are
recommended prior to startup.
Figure 2 ~1:
Q Check that all cable connections between the Power Configuration
Gantry and the Console are secure.
(
M-N Operator's Manual 17
Accessory Set Up & Installation
Pre-Power Up Checks If, during the diagnostic checks, the system detects a
Before applying power to the system, perform the pre- fault or an error condition, a message will display in the
startup checks below. Correct ALL discrepancies before Message Area. To temporarily stop the startup sequence
applying power. so that the error message remains on the screen, simulta-
neously press the <Shift> & <View> keys. To resume
o Check the unit (Console & Gantry) for open the startup sequence, press the <Shift> & <View> keys
panels, loose or missing hardware. again.
o Check for damaged or disconnected cables, and When the diagnostics are complete. the Display will
for other visible problems. show the system operational mode screen (Run Mode
screen).
o Verify the integrity of the radiation shield (chips,
cracks, breaks. security).
Post-Power Up Checks
Perform the following checks after powering up the
(
System Power Up M·IV Mammography System:
After completing the pre-power up checks, perform the
steps below to power up the system. o Verify that all the C-arm Controls function
properly.
o Switch ON the Input Power Circuit Breaker on the
rear of the Gantry. o Check that the Tubehead Controls are operating.
Q Press the Power ON switch on the control panel. o Verify operation of the Control Panel keys and
switches.
After pressing the power ON switch, a tone will sound
and the control panel display will illuminate, indicating o Install a floppy disk in the floppy disk drive (for
that the system is turned on and is powering up. post-exposure data storage),
The Operator Console display will briefly show the M" NonCE ••• Jhe system will post a ....no,...
system aisle" error messoge if the floppy
message:
aisK is not removea from the arive at .tart
''The M·IV is performing start up diagnostics"
up_
18
CHAPTER
THElMAGE
RECEPTOR SVPPORT DEVICE STANDARD
IRSD
The Image Receptor Support Device (IRSD). on the
M·IV C-arm. accepts standard film image receptors and
the LORAD Digital Spot Mammography (DSM) Image
Receptor.
c
On the back of the IRSD is a rocker-type switch that
permits C-arm Rotation. Note that this switch functions
C·ARM
identically to the C-arm Rotation Switches on either side ROTATION
of the tubehead. SWITCH
The IRSD has internal detection circuitry that senses the Figure 2-2:
installed image receptor holder and automatically: The Image Receptor Support Device
Figure 2-3: (
Mounting & Loading a Bucky Device
c o
o
Stand in front of the unit facing the C-ann.
o Slide the Magnification Table onto the IRSD until Loading the Magngication Table
it stops. To load a cassette into the Magnification Table. follow
these steps:
To remove a Magnification Table:
a Pull the Magnification Table straight off the a Push the cassette finnly into the opening until it
IRSD.
snaps into place.
When installed. the detection circuitry automatically To remove a cassette from the Magnification Table:
recognizes the Magnification Table. The system then
selects the small focal spot. and sets the collimator to the Q Push the cassette until it slips out of the slot in the
18cm x 24cm field size (Manual Collimation). or to the Magnification Table.
size of the installed paddle (Automatic Collimation).
a Pull the cassette straight out of the device.
orr
~
G
Image Receptor: lacm x 24cm
Automatic Collimation: lacm x 24cm
~ Image Receptor: 1acm x 24cm
o
Image Receptor: 1acm x 24cm
Automatic Collimation: 1Scm x 15cm Automatic Collimation: lacm x 1acm
Manual Collimation: lacm x 24cm Manual Collimation: 1acm x 24cm
RADIATION SHlETDING
The unit provides shielding devices to protect the patient
and operator from exposure to x-radiation. ALWAYS
ensure that the shielding devices are attached, and
remember to check their condition prior to use each day.
o Slide the open end of the face shield into the slots
Figure 2-6:
on the tubehead mount.
Installing the Face Shield
o Push the face shield completely on the mount -
until the rear tabs snap in place.
DUAL FUNCTION
FOOTSWITCHES
The M-IV incorporates two footswitches that control
C-arm UplDown movement and Compression UplDown
movement. These footswitches free the operator's hands
for patient and breast positioning. Each dual pedal
footswitch uses a rocker mechanism which permits one
pedal to control both up and down movement.
*
Figu.re 2-7:
The Dual Fu.nction Footswitch
I ~I COMPRESSION UPIDOWN
The footswitches provide the same movement control as
the pushbutton controls on the C-arm and enables the
c
The footswitches plug into receptacles inside the Gantry same M-IV functions :
(near the bottom), and are connected by a service
technician at installation. These receptacles are not user • The light field lamp illuminates upon Compres-
accessible. sion Down.
Place each footswitch on the floor so that one is acces- • LOCKOUT mode disables the motorized Com-
sible from the left side of the C-arm, and the other is pression Release pushbuttons.
accessible from the right side.
• All compression modes are available (Pre-, Full
Always position the footswitches so as to eliminate Range. or Dual Compression).
accidental activation by patient or operators.
FILM LABELING
The system provides identification and technique factor
labeling on mammograms by means of the integrated
AutoFilm ID unit. or the optionallabeJ printer. The
AutoFilm ID "flashes" a photographic image of patient
demographics and the exposure factors onto the film
before it is developed. while the printer provides a hard
copy label for placement onto the film after it is C.~n.
Slot
~§~
-----t::""'~.
developed.
The AutoFilm ID
The Auto Film ID is a photographic labeling unit that
eliminates the need to manually label the mammogram.
This unit is integrated into the Operator Console and
requires the use afDIN-style cassettes. Cassette
The LORAD HTC Grid System The Rotation Collar is installed at the factory (on sales
order). or in the field by a LORAD service representa-
The LORAD HTC Grid System is a Bucky device which
tive. The StagelBiopsy Assembly and the Control
contains aJligh:transmission.£ellular grid in place of a
Module are installed onto the C-arm as necessary for the
linear grid. The LORAn HTC Grid System only accepts
application.
DIN-style cassettes with the sliding window for pholo-
graphic placement of patient and technique information
The StereoLoc II operates with standard film/screen
directly onto the film using the integrated AutoFilm ID.
PREFACE: I The design and function of the M·IV controls. readouts, and indicators
expedites equipment setup, patient positioning, exam parameter setup, and
the monitoring of exam proceedings. System controls are strategically
located on the Operator Console. the Gantry, and the C-ann. Readouts and
indicators are conveniently located on the Gantry and on the Compression
Carriage.
OVERVIEW: I Chapter 3 describes the purpose and function of the controls, readouts. and
indicators on the Operator Console, the Gantry, and the C-arm.
OBJECTIVES: I After reading this chapter. the user will understand use of:
c •
•
•
Operator Console Controls
Gantry, C-arm, and Tubehead Controls
Gantry and Compression Carriage Readouts
and Indicators
Chap-ter 3:
System Controls
(
c
CHAPTER
System Co;ntr'Ols
OPERATOR
CONSOLE CONTROLS
The Control Panel. on the Operator Console, is used to In addition to the controls listed, a Motion Emergency
enter exam infonnation via the Alphanumeric keys. the OFF Switch is positioned on the left side of the Operator
Function keys, and the View keys, and to set the exposure Console frame which, when pressed, immediately
techniques for the cltarrunation using the Exposure removes power from the system.
Technique keys.
A Bar Code Reader (option), located to the right of the
The Control Panel contains the following: Control Panel, provides an alternate means of entering
patient data.
• X-ray Release keys (2)
• Function Keypad
• Alphanumeric Keypad
c •
•
View Keypad
Exposure Technique Keypad
b
COnIP,.UIOn{
R.....
0
~ ~
oI oI
X-ray X·,.y
Sw"," Switch
~~I~
Fu~n
KeyptldLh, ~_ .. _
I
Alph.lnurr.rlc KeylHld
. __________ _ _ .. _ ..... _ _
-
......
I
Figure 3·1:
Th e Control Panel
• Enter patient and technologist identification data. • Switch between the Run Mode and the Setup
Mode (Run I Setup)
• Manually input a CPT Code.
• Switch among the data field choices (Change)
• Type comments pertaining to the patient or exam.
• Erase data field entries (Clear)
In the Additional Setup Mode, use these keys to:
• Select clinical view labels (View keys)
• Make time and date changes.
• Acknowledge error messages (Reset)
• Enter film I screen types for calibration.
The following paragraphs describes the functions of each
• Type site the name and address information. View Keypad component in more detail. Figure 3-2
Figure 3-2:
The View Keypad
System Co'rltTols
Q
Automatically moves the highlight bar in the
Patient ID data field.
contains multiple choices, pressing <Change> toggles Note that the Tech ID data line is only cleared when the
through the available options for that field. Access the Additional Setup data field ''Tech ID cleared with Clear
Run Mode screen and perform the following example Key" is set to YES.
THE EXPOSURE
TECHNIQUE KEYPAD /77 I \ ""'''''''
Use the Exposure Technique keypad, above the alphanu·
merle keys. to set the exposure parameters for the'
examination(see Figure 3·3).
,
The descriptions below detail the functions of each QWEflTYU 0
Fig",e 3-3:
The Expos",e Techniq"e Keypad
The Mode Select Key. • . . .•.. ~ @if":: ~
The Mode Select Key permits the user to select
the exposure mode for the examination. The system runs The Filter Select Key ••••..• ~
in one of three automatic modes, or a manual mode is
available for user selection of all factors. Use <Filter Select> to choose the desired filter
for the exposure.
The four exposure modes are:
c •
•
Manual
AutoTIme
The available filter choices are:
• Mo (Molybdenum)
• Rh (Rhodium)
• Auto-kV
• Auto-Filter To select a filter fOT the exposure perfonn the following:
To select an exposure mode, perform the (ollowing: o Enter the Run Mode and select the desired
exposure mode.
Q Enter the Run Mode.
o Press <Filter Select> to switch between the two
Q Press <Mode Select> several times to cycle choices: Mo or Rh.
through the three available exposure modes.
o Release the key when the preferred filter choice is
Q Release the key when the preferred exposure shown in the data field
mode is shown in the data field.
The Density Set Keys ..•.... ~ The Film Select Key • •..••.• ~ _ ~
~8
The ten keys on the left side of the Control Panel
comprise the Function Keypad (Figure 3-4). -. 1
!
•
The CPT Code Key
~~
• The Stereo Key
• The C-arm Position Key
• The Cursor Keys (arrows)
rff1
m~
• The unlabeled keys (reserved for future use) {r
\
The CPT Code Key .•.. •....
m~
Cv.RI?€"('J'f f ROCf'<>'-!RAL "rf(:Y..J"h.• E.:
:FI
Use this key to scroll through the scheduled exam
billing codes for that patient. The available choices will
Ctrt
~~
(
appear in the CPT Code data field in the Run Mode.
Below are the steps necessary to schedule CPT Codes for Figure 3-4:
a patient. The Function Keys
Q While in the Run Mode, highlight the CPT Code To switch between the scheduled CPT Codes for that
data field. patient, press <CPT Code> several times to scroll
through the scheduled billing codes for that patient.
Q Press <Change> to scroll through the pre-set list
of billing codes. Upon selection of a CPT Code, the associated ell.am
views are automatically placed in the Exam View data
o When the desired billing code appears, press field. However, manually entering a CPT Code will
<Enter> to schedule the patient for the exam. require that the exam. view also be manually entered
using the mammographic view keys.
When scrolling through the CPT Code list using
<Change>, the billing codes appear dim (half-bright).
When a CPT Code is scheduled by pressing <Enter>, the
billing codes appear bright.
The Stereo Key. . • . . • • • • • •• 8 labels, to move the highlight bar through the
selectable on screen data fields. Each arrow key (Cursor)
moves the highlight bar in the direction indicated by the
This key is only functional when the M-JV has label on the key cap. Below details this movement:
been configured for stereotactic use, and when the
LORAD StereoLoc II device has been mounted onto the • UpArrow:
C-ann. Use this key to set the M-JV so that the C-arm highlight moves from bottom to top
only rotates to the stereotactic projection angles (C-ann
shifts ±lSD while the StereoLoc II remains stationary): • Down Arrow:
highlight moves from top to bottom
CI Install the StereoLoc II device on the M-JV
C-arm. The system automatically is set for • Left Arrow:
stereotactic rotation. highlight moves from right to left
(
M-N Operator's Manual 37
System Controls
o
o
Lower the Compression Device.
C-ARM CONTROLS
There are identical pushbutton switches on either side of
the C-ann that control the following operations on the
mammography system:
• C-arm rotation
There are also handwheels on either side of the Compres- Figure 3·5:
sion Carriage for manual movement ofWs device, Use C-arm Pushbunon Controls & Switch Set
of each pushbutton, switch, and the handwheels, is
described next. Note that the C-arm will stop automatically upon
( reaching the upper or lower C-arm travel limits. For
tactile discernment, the C-arm Up switch is convex, and
the C-arm Down switch is concave.
C-arm
l' ~'
Up & Down Controls ° og NOTICE • • _ The Caarm Up and C-arm Down
lunctions are also available using the dual
C-arm Up and C-ann Down switches are on both C-arm lunction Footswitch.
switch panels (one on either side of the C-arm). This set
of switches moves the C-ann up or down for height
positioning of the image receptor support device.
M NOTICE ••• The Caarm Rotation fundions
are also available using tubehead controls.
(
M-N Operator's Manual 39
SW5t",,,, Controls
Compression Compression
t ,j,
. . . ....... .
Up & Down Switch. ... ... 00 Release Switch.
Compression Up and Down switches are on both A Compression Release switch on both C·arm switch
C·ann switch panels (one on either side of the C-arm). pads provides quick release of the patient from compres-
Use these switches to raise the compression device sion, provided the Compression Release Mode is set to
(increase thickness). or to lower the compression device MANUAL. The function and use of this switch is the
(decrease thickness). same as the Control Panel Compression Release Switch.
The Compression Release switch has a raised bump on
The Compression Up switch is convex, and the Compres- its surface for tactile discernment.
sion Down switch is concave for tactile discernment. To
move the Compression device, perform the following: The M·IV has an AUTOMATIC Compression Release
feature which. when pre-selected in the Setup Mode.
a Press and hold <Compression Up> to raise the releases compression automatically at the end of each
Compression Device. exposure. However, if a Localization Compression
Paddle is used, the system senses the paddle and "locks
a Release the switch to stop upward movement out" the Automatic Compression Release function.
C-arm Manual
Rotation Switch " Compression Handwheels
Use this rockeNype switch on the rear of the IRSD to Manual compression Handwheels are provided to adjust
control C-arm rotation. One side of the switch is labeled the Compression Device manually. Use these controls in
for clockwise rotation. the other side for counterclock- addition to the motorized compression methods. The
wise rotation. handwheels work as follows:
To rotate the C-arm to the angle necessary for the clinical Q Rotate the handwheels in one direction to de-
view, perfonn the following: crease object thickness (apply compression).
CI Press and hold the clockwise label side of the Q Rotate the handwheels in the opposite direction to
switch to move the C-arm to a more positive increase object thickness (release compression).
angle.
Moving the Compression Device manually permits the
a Release the switch to stop clockwise rotation. user to apply up to 65lbs of force (maximum) onto the
object. This method of compression is typically used
a Press and hold the counter-clockwise label side of after applying motorized compression.
the switch to move the C-arm to a more negative
c o
angle.
C-ARM
READOUTS & INDICATORS
• Compression thickness
• Compression force
• AEC Sensor Position & Indicator
Compression
Thickness Display . . . . .. . . .
An LED display, on the Compression Carriage cover,
shows the distance between the bottom of the Compres- -,
sion Paddle and the top surface of the image receptor
holder. The readout displays the compression thickness - 5 y- o (
in O. tern increments between Oem and 15cm. To see
how the display fu nctions, perform the fonowing : -3
System Contr·oz..
Compression -,
Force Display. ............ "'9'" Automatic Exposure Control
Sensor Position Indicator • •.
-,
An LED display, on the top Compression Carriage cover,
shows the approximate force being applied to the
compressed object. The readout displays the compres-
A sensor in the Image Receptor Support Device
measures the x-radiation that penetrates the breast
-,
sion force in either pounds (5Ib to 65tb) or Newtons during an exposure. The AEC sensor may be
(20N to 290N). whichever is pre-set as the default in placed in one of seven detent positions. Typically, the
Setup. sensor is placed below the glandular tissue in the anterior
region of the breast. To check the function of the
To see how the force readout works, perform the indicator. perfonn the following:
following:
a Install a large Compression Paddle.
Q Place an object (phantom, breast equivalent
material. etc.) on the image receptor support a Grasp the AEC Sensor handle beneath the IRSD
device, then install a Compression Paddle. and move it back and forth to all seven positions.
a Press a Compression Down control to compress a Observe that the corresponding LED on the
the object between the Paddle and the Image position display changes.
c Q
Receptor, and observe the increase of the force
readout.
TUBEHEAD CONTROLS
A control pad on either side of the Tubehead contains
identical switches for:
• C~arm Rotation
• Collimator Override
• Light Field illumination
C-orm
Rotation SWitch . ... (0. Figure 3-8:
Tubehead Button
Use this rockeNYpe switch, on either side of the
tube head. to control C-arm rotation. One side of the
switch is labeled for clockwise rotation. the other side for
counterclockwise rotation.
Light Field Switch . • • • • . . .. @ (
This switch turns on the collimator lamp. which
This switch fu nctions the same as the Rotation switch on provides a visual representation of the x-ray field. Use
the rear of the IRSD for both standard mammography. this feature as necessary to accurately position the patient
and for stereotactic procedures (if applicable). in the x-ray field. The collimator lamp illuminates for 30
seconds whenever this switch is pressed, and it extin-
guishes automatically.
Corm Rotation Memory
To help with patient positioning. the M·IV employs a
C-arm Rotation Memory feature. After making an
oblique exposure, the C-ann angle is stored in memory.
When the technologist rotates the C-arm past 0° for the
next exposure, the rotation circuitry automatically SlOpS
the C-arm at the opposite oblique angle. The technolo-
gist may then rotate the C-arm to an angle other than the
memorized oblique angle if necessary.
Collimator
Ovenide Switch
The internal beam limiting device uses either an auto· In either mode, use the Collimator Override switch to
matie mode or a manual mode to collimate the x-ray field adjust the collimator blades for field sizes other than the
appropriately for the installed Compression Paddle and pre-set field size:
Image Receptor. Choice of the collimator mode is made
in Setup. The Collimator Override is used to change the a Install the appropriate Image Receptor and
automatically selected x-ray field size to a size other than Compression Padl,ile.
that of the installed Compression Paddle (Automatic
Mode) or the installed Image Receptor (Manual Mode). Q Press the Collimator Override switch to illuminate
the light field lamp and to adjust the collimator
AUTOMATIC MODE: When Automatic collimation is blades to one of the available field sizes, depend-
set, the collimator adjusts to the field size of the installed ing on the selected mode and installed
Compression Paddle, or to the installed Image Receptor accessories.
when no Paddle is attached. The Collimator Override
switch allows changing the field size up to the size of the
installed Image Receptor. Note that collimation override
will not allow field sizes any larger than the size of the
GANTRY
CONTROLS & SWITCHES
.The Gantry contains the foUowing user controls and
switches:
The Motion ~
Emergency OFF Switch. . . . •. ~
The Gantry holds two of the system's three Motion
Emergency OFF switches (one on either side door).
Pressing this switch immediately removes ALL system
power. These switches work identically to the switch on
the Control Console.
(
The Input
Power Circuit Breaker
The Input Power Circuit Breaker is located o n the back
afthe Gantry. Typically. the circuit breaker is left ON.
However. always check the circuit breaker status prior to
applying power each day.
Figun3~9:
The Motion Emergency OFF Switch
GANTRY READOVT
The M-IV Gantry has a separate readout for the C-arm
angle
PREFACE: I The M· IV user interface involves several modes of operation. For clarity
and ease of understanding, each of these modes is presented in a separate
chapter. The Run Mode is the operational mode - where x-ray exposures
are made.
OVERVIEW: I When the M-IY powers up. it goes through a start up routine before
allowing user interaction through. the main operating screen - the Run
Mode screen. This chapter briefly explains the Stanup screen, and details
the components of the Run Mode screen wlUch permits:
c
OBJECITVES: I After reading this chapter. the reader will be able to:
Chap-ter 4:
User Interface - The Run Mode
(
(
User Interface - The Run M(KU
The RUn Mode Screen The left side of thc Message Window displays instruc-
tions for thc action taking place (Le., highlighting the
When the M·IV completes the start up sequence, the
Birth Datc data fields brings instructions on entcring thc
system automatically entcrs the Run Mode. which
patient's datc of birth). During fault conditions, such as
indicates that the system is ready to receive patient and
an empty cassette holder, the Message Area requests a
exposure information. In the Run Mode, the user
user response for corrective action. Errors that occur
performs the following tasks:
during operation also appear in the Message Area. In
cases where user action can remedy the error, appropriate
• Inputs patient information and exposure technique
prompts appear; in cases where the error condition is
factors beyond the user's ability to "clear", the message reads
• Views accessory configuration "Call Service".
• Reads system operational status messages (errors
The Exposure Technique WlDdow displays all the
& help prompts)
parameters by which the exposure will be conducted (Le.,
• Permits user access to the Setup Mode kV, mAs, mode, etc.). Note that the techniques available
for change will differ depending on the exposure mode
The Run Mode screen is divided into three sections: selected (described later).
The paragraphs that follow describe each data field in Q Type in the last name of the patient, followed by a
this column, and the information contained therein. comma (,).
a Type the slash (1), then cnter the 2-digit day (01 a Press the Appropriate letter designation (M or 0).
through 31 are the only valid entries).
a Press <Enter> to move the highlight to the next
a Type the slash (I). then enter the two digit year data field.
(Le., 56 for 1956).
NOTICE • •• Pressing any 'e"or ot"or than
o Press <Enter> to register the birth date entry and tile valid '.".rs lor th. gender designation (
move the highlight to the next data field. '0
causes the sys'em soltwore automa,1-
cally display the d.ICJult gonder (F).
DDIMMIYY Format (DaJIMonthIY~ar)
Q Highlight the Birth Date Data Field. The "Comment" Data Field
Use the Comment data field to display simple remarks
Q Enter the 2-digit day (01 through 31 are the only
regarding the examination. This data field holds up to 25
valid entries).
characters. including numbers. letters, and punctuation
marks (comma, period. hyphen. space, etc.).
a Type the slash (I), then entcr the 2-digit month (01
through 12 are the only valid entries).
To enter a comment, use the following procedure:
o Type the slash (I) , then cnter the two digit year
a Highlight the Comments data field.
(Le. , 56 for 1956).
The "Tech ID" Data Field The "CPT Code" Data Field
The Tech ID data field contains the technologist's The 5-character CPT Code Data Field has two sections:
personal identification code. This field holds up to 3
characters, including numbers, letters. and punctuation • CPT Code Field
marks (comma. period, hyphen, space, etc.),
• Exam Description Field
Entering a Tech ID that has been previously defined in
The CPT Code Field displays the billing code(s) for the
the Technologist Setup Mode causes the following to
scheduled exam. Use the pre-defined codes (CPT Code
occur:
key), or manually enter a billing code (alphanumeric
keys).
• The preferred Compression mode is activated.
• The first view in the pre-defined screening exam The Exam Description field displays the exam name that
view sequence is displayed. is associated with the chosen billing code in the CPT
Code field. For example. if the billing code represents a
• The view sequence is available without manually
entering a CPT code. normal breast screening technique. the Exam Description
field will read "screening". This area is not user acces-
sible.
To enter the technologist's ID code, perform the
c
following:
The user can schedule CPT codes for each patient by
performing the following:
o Highlight the Tech ID data field.
a Press <Enter> to register the entry and move the o Press <Change> to scroll through the available
Highlight to the next data field. CPT codes. (When scrolling through the codes.
the data field appears dim (CPT Code is not yet
scheduled).
The "Exam View" Data Field The Exam View Count Box tracks the number of expo-
sures taken for the displayed view. For example. if the
The Exam View data field shows the associated exam
views L MLO and R w..O are associated with the
views for the selected CPT Code. When the user enters a
selected billing code. the View Count box will change
CPT Code, the first view associated with the entered CPT
similarly to the example below:
code is displayed. Press <View> to scroll through the
remaining views available for the entered CPT Code.
a Select the first exam view (L w..O) by pressing
<View>. then take the exposure. Observe that the
The exam views appear in the data field as follows:
View Count box reads: I.
• Bright when "Selected" in the CPT Setup
a Select the second exam view (R LMO) by
• Dim when "Optional" in the CPT Setup pressing <View>. then take the exposure and
observe that the View Count box reads: 1.
Note that pressing <View> permits selection of only the
exam views that were previollsly marked "Selected" or IT it is necessary to retake the L MLO exposure:
"Optional" in CPT Setup. Exam views defined as "Not
Selected" are not available using <View>. • the Count box will read: 2
(when L MLO is the displayed exam view)
To schedule an exam view that is not pre-defined (Not
• the Count box will read: I
Selected) for the CPT Code, perform the following:
(when R MLO is the displayed exam view) (
o Highlight the Ellam View data field
M di.ploy.d
NOTICE ••• "'S.I.ct.d'" views 'or tit.
CPT Cod. app.ar bright; "'Op-
tional'" and "'Not S.I.ct.d'" views oppeor
dim.
The "Compression" Data Field To change the compression force mode, perform the
following:
Motorized compression is applied using one of the three
force modes below:
a Highlight the Compression data field.
58 M ·W Operator's Manual
User Interface· The Run M(la.
c date display, or to change the date format. The "Compression" Status Field
This field displays the Compression Release mode
selected in the default settings. There are two Compres-
The "Time" Status Field
sion Release modes: Manual and Automatic. Refer to
This field displays the current time in the format of the Chapter 5 (The User Interface - Setup Mode) to change
selected language. Whenever the user interface language the compression release mode. Note however, that when
selection changes, the system automatically displays the a Localization PaddJe is mounted on the Compression
time in the appropriate format. Refer to Chapter 6 (The Caniage, this data line will read: Locked Out.
User Interface - Additional Set Up Mode) to correct the
time display, or to change the time format.
• LORAD HTe Grid System (l8cm x 24<:m) The "View" Status Field
• LORAD Linear Bucky (24cm x 3Ocm) The system incorporates a feature that requires entering a
mammographic view label in the Exam View data field
• LORAD Magnification Table
before enabling an exposure. When this "View Re-
Also detected are the following optional receptors: minder" is ON, the View data field will flash until a
mammographic view is entered for the exam. The status
display will be either Reminder ON or Reminder OFF.
• LORAD HTe Grid System (24cm x 3Ocm)
Refer to Chapter 5 (User Interface - The Setup Mode) to
• LORAD Linear Bucky (18cm x 24cm) change the operational status of the View Reminder.
• LORAD DSM Digital Image Receptor
• LORAD Full Field Digitallmage Receptor
PREFACE: I The Setup Mode permits the user to tailor the M-IV System for specific
operations and personal preferences. In Setup. the user can set commonly
used exposure techniques. choose a preferred compression mode , and
control the operational status of installed peripherals. The user also has
access to a data field which allows selection of other Setup screens.
OVERVIEW: I Chapter 5 describes the Setup Mode screen and guides the user while
setting machine parameter defaults, setting tech.nique factor defaults for
each exposure mode. and for selecting operational status of peripheral
devices and (earures.
OBJECTIVES: I After reading this chapter, the user will be familiar with:
c •
•
Setting desired defaults for each machlne parameter.
Selecting defaults for the techniques available for each exposure
mode.
(
Chap-ter 5:
User Interface - The Setup Mode
(
(
CHAPTER
User Interface - The Setup M,Id.
INTRODUCTION
The Setup Mode allows the user to set the exam and
exposure defaults most frequently used. and is accessed
from the Run Mode by pressing <Run I Setup>. The
choices made in Setup appear in the Run Mode each time
system power is applied. These defaults, however, may
be changed as necessary in the Run Mode.
c light bar will always appear in the first data field of the
Serop Window. Pressing one of the arrow keys <Cursor>
on the Function keypad moves the highlight bar through
the available fields.
Selecting Fields & Entering Data
The following paragraphs describe movement through
the setup screens, changing data field selections, and
eotering system labels.
The Message Window displays instructions for setting DtJItJ Field Sehction: Press the arrow keys <Cursor>, or
andlor changing system operations or preferences. When press <Enter>, to move the highlight through or across
a field is highlighted, instructions appear to guide the the columns on the screen. A data field is selected when
user through the correct sequence of steps for entering the highlight appears in the data field.
data for that particular field.
Clumging Dam Field Selections: After selecting a data
The right side of the Message Window (the Status field, press <Change> to switch between the available
Window), will always read: SETIJP. choices for that machine function. H the data field
requires manual entry, key in the information using the
The Exposure Technique Window permits the user to set alphanumeric keyboard.
or change the default technique factors for each exposure
mode. Selection of these defaults an: made in the same CluJllging Exposure Technique Defaults: The exposure
manner as the Run Mode exposure technique selections. techniques are Changed by pressing the corresponding
Technique Select key. These keys are on the Control
Panel above the alphanumeric keyboard.
PRE-FORCE FULL-FORCE
89.0
50% 22.0 97.9 28.5 126.8
60% 23.5 30.5 135.7
c
Mode. The choices for this data field are:
Locked-Out mode activates upon installing a localization
paddle, and automatically disables the compression
• MANUAL
release function.
• AUTO-TIME
• AUTO-KV
Collimator:
The collimator device within the M-IV tubehead operates
in one of two modes: View Reminder :
The system may be set so that exposures are not possible
• AUTOMATIC unless an exam view is displayed in the Run Mode.
Typically. the View Reminder is ON for clinical examina-
• MANUAL tions. However, the user or service technician may
switch it OFF to perfOITIl verification exposures (using a
Automatic collimation mode has the detection circuitry
phantom), or during unit calibration. The only options
sense the size of the installed Image Receptor and
for the View Reminder are:
Compression Paddle. and adjusts the field size appropri-
ately for the application.
• ON
Manual collimation mode has the detection circuitry
• OFF
sense the size of the installed Image Receptor only. and
limits the field size appropriately. All subsequent field
size adjustments for the application are made using the
Collimator Override Switch on the tubehead.
65
User Interface - The Setup Mode
EXPOSURE
Auto-11m, ExpQsu« Mod,
TECHNIQUE DEFAVLTS
Filter Molybdenum (Mo) or Rhodium (Rh)
Use the area along the bottom of the screen to define the
exposure parameters that will appear as defaults in the kV 20kV through 39kV
Run Mode 's Exposure Technique Window. A unique set mAs not se lectable
of parameters may be chosen for each exposure mode.
including magnification modes. Below are the seven Density -5 to 0 to +5
modes available for setting defaults: Spot Large or Small
• Auto-Filter Density
Spot
-5toOto+5
Large o r Small
(
Choose the exposure mode requiring setup (default
selection). then use the tables that follow to determine FUm one of the three film/screen
the exposure techniques available for each exposure combinations calibrated
mode. Auto-Filtf!IT E:c,pQsua Mode
NOTICE • •• P,....ing tit_ Morle Key perm;'. Filter not selectable (default - Mo)
'0
the us.r choo•• among seven exposure
kV 25kV
mod•• (4 contad mod• • , 3 mas mod• • ).
When the "'mas'"' lab.1 appear. in the mAs not selectable
Mod. da'a fleld, the default• ••, lor that Density -5 to 0 to +5
mod. will b. the •• H;ns_ the M .JV recall.
when the magnifica,ion 'oW. i_ ;mlall. d . Spot Large or Small
Film one of the three filmlscn:en
Manual Exposure Motkj combinations calibrated
Filter
kV
Molybdenum (Mo) or Rhodium (Rh)
20kV - 35kV (Mo)
.&f NOnCE - • • A ulo-Filter expo_ure made i.
nat uvai'abl. lar mugnilieulian . Iudle••
28kV - 39kV (Rh)
mAs 3mAs through 500mAs
Density not selectable
Spot Large or Small
Film one of the three film/screen
combinations calibrated
PREFACE: I The Additional Setup Mode is a second default setup screen that pennits
the setting of the remaining system parameters. and the machine labels
(Le .• film type. date, time).
OVERVIEW: I Chapter 6 describes the Additional Setup Mode screen and guides the user
through default settings for:
• RIS connection
• Film and Flash Duration
This chapter also explains how to change the date and time and how to
enter label data (film types. institution name and address).
c OBJECTIVES: I After reading this chapter, the user will be familiar with:
Chap-ter 6:
User Interface - Additional Setup
(
c
CHAPTER
User Interface - Additional Selmll
c •
•
RIS
Film DEFAULTS -
ADDITIONAL SETUP
• Flash Time
• Date
The following paragraphs detail each of the Additional
Setup data fields, and the choices available for each field,
• Time and instructions for changing or modifying them.
• Institution Name
The Exposure Technique Window is always blank in the To change the compression force units of measure,
Additional Setup mode and is not used. follow these steps:
To change the current date. follow these steps: To enter or change the name of the institution, clinic, or
hospital, follow these steps:
o Highlight the Date data field, then use the
o Highlight the Institution data field.
Backspace key to delete the displayed date.
c separate the month. day, and year digits. Note that the
two digits representing the year automatically change to
the 4-digit year (Le., 01101198 changes to 0110111998).
i5 re5erved lor the unit number. There/ore~
alwaY51imit tlte lengtlt 01 the 'n5titution'5
Name to 31 character5.
To change the current time, follow these steps: o Highlight the Institution address field.
Q Highlight the Time data field, then press Change o Type in tth appropriate name
to select the desired time fonnat. (up to 32 characters).
Q Use the Backspace key to delete the displayed o Press Enter to record the entry.
time, then type in the new time in one of the
above fonnals.
To set the Clear Key mode. perform these steps: CJ Use the Change key to step through tthe offset
numbers.
a Highlight the data field.
CJ Press Enter to record the change.
o Press Change to switch between YES or NO.
To change the Auto ID contrast. perform these steps: To set the Daylight Savings Time option, perform the
following:
o Highlight the Auto ID Contrast data field.
CJ Highlight the data field.
o Use the Change key to step through the contrast
values (1-19). CJ Press Change to switch between YES or NO.
o Press Enter to record the change. CI Press Enter to record the change.
I
The system permits entry of patient demographics prior to the patient's
PREFACE: arrival. The identification information entered is stored on the hard drive
and is recalled at the time it is nceded.
OVERVIEW: I This chapter explains the mechanics of the Pre-schedule feature, describes
the means of entering, recalling and modifying the data, and discusses the
data storage feature.
OBJECTIVES: I After reading this chapter. the user will be familiar with:
Chap-ter 7:
Patient Pre-scheduling & Data Storage
(
(
Patient Pre-Scheduling & Data Stor,~a.
To manually enter Pre-schedule information, perform the Cl Use the Change key to step through the pre-
following:
scheduled data.
Q Thm the unit ON, and then enter the Run Mode.
The system recalls the previously pre-scheduled data and
CI Type in the patient identification information (lD automatically fills in the appropriate data fields. At this
and Name). point. any data field (except the patient ID data field) can
be modified as necessary.
Q Enter any other known demographics (Sex, Birth
AUTOMATED PRE-SCHEDVLlNG When the Reader registers the code, the system will
sound a tone, and the patient information for that data
The following details the automated methods of patient field will appear. The hightUght will move to the next
pre-scheduling. data field. where the process is repeated. If a non-valid
entry is made using lhe Bar Code Reader. the M·IV will
sound a tone. The hightlight does not advance until a
The Bar Code Reader valid entry is made.
An automated means of pre-scheduling data is through
the optional Bar Code Reader. The Bar Code Reader In some cases, the Bar Code Reader may be used to pre-
transfers the data that it reads from a coded strip (on a schedule exams in advance of patient arrival. A group of
patient ID card or admissions slip) to the appropriate data codes can be scanned using the Reader at the start of the
fields on the Run Mode screen. To recall data using the day. As patients arrive in the exam room, the user simply
Bar Code Reader, perform the following: types the exact patient ID into the Patient ID data field to
recall the previously scanned data.
o Thm the unit on and access the Run Mode screen.
I
The steps involved in an examination are usually numerous. For this
PREFACE: reason, it is recommended that the user develop a routine for
perfonning these steps, and adhere to that routine each time the exam is
performed.
• . Contact exposures
• Magnification studies
• Spot exposures
c OBJECTIVES: I After reading this chapter, the user will be familiar with:
Chap-ter 8:
Clinical Procedures
(
c
Clinical Proc,e d,ur•.,
2. Check that all interlocks are set (Le., IRSD lock. 5. Confirm that the "Receptor" data field indicates
Gantry panels closed, etc.). the presence of the mounted Bucky.
c
1. Select the appropriate Compression Paddle for the
4. Check the area near the Gantry for obstructions procedure.
that may hinder C-ann movement.
2. Install the Compression Paddle into the mounting
5. Remove any disk from the floppy disk drive. holes in the Compression Carriage.
~: NOTICE ••• When pollen' demog,..plrla 2. Adjust the C-arm height to adapt to the patient.
0'• • n,....eI u"n. tit_ ",....dr.du/. Ie.
'u~ AU patient Information doto 'Wd. RII 3. Position the patient. Place the patient's arm or
automatKa/Iy..,..... ......... ",. poII- hand on the patient handle.
_r_rwmIMr,
WAaNING, Mo,orlzerl."u'pmon'a U••
care wilen od/u.tlne lor patlen' u...
O •••rve equipment and patlen' at all
tim •• durin••• r up. " a chair I. nec...
MIry lor CI procedure, wit.,.. po_••'.,
".I.h,".1.",
v •• 0 YGrfob'.
minimum
cItaIr ..t .ky. ".
t . . . .,., with rII • •• , up
proc••••
Clinical Proc••d,"......
M-lV Power Up
1. Switch ON the Gantry power Circuit Breaker.
2. Install the Compression Paddle into the mounting 8. Enter any comments (up to 25 characters)
holes in the Compression Carriage. regarding the examination.
3. Check that the automatic collimator adjusts to the ~' NOTICE •.• When potient demographic5
Compression Paddle format. are entered u5ing the pre-schedule 'eature,
ALL patient in'ormation data 'ield. 'ill
4. Raise the Compression Device to provide ample automatically upon entering the patient
room for positioning the patient's breast. identi'ication number.
8. Select the film I screen combination being used M the ma.ximum .iI. 01 the
NOTICE ••• n.. automatic collimator limi,.
~-nEY field to the
C
for the exposure.
m. 01 th.'mag. receptor (I•••, 'BCdlX
24an, 24an" 30an).
Positioning the C-arm & Patient
I. Rotate the C-ann to the necessary angle for the
selected clinical view.
POsitioning the AEC Sensor
1. If using an automatic exposure mode, move the
2. Adjust the C-arm height to adapt to the patient. AEC Sensor handle forward or back to position
the sensors.
3. Position the patient. Place the patient's ann or
hand on the patient handle. 2. Check that the sensor position is beneath the
portion of the breast where the tissue is most
WAaNING' Motor'zerlequlpment. u•• dense.
co,.. wlt.n ad/v.,'ng lor patl.nt u••.
Ob••rve equIpment an" patIent at all
tIme. "urlng .et up. " a cllalr I. nece..
• ary 'or a proce"ur., wII.... poulltle,
u.e a varialtl. lIeiglit clloir .et oltoYe it.
minimum lI.igll, to o ••i.t with the •• , up
proc••••
2. Make any final technique changes and demo- 2. Insert the cassette (DIN-style only) into the
graphic corrections. integrated AutoFilm ID. After the audible tone
sounds. remove the cassette.
3. Move completely behind the radiation shield.
3. Develop the film.
4. Instruct the patient to be as still as possible and to
hold her breath during the exposure. ~' NOTICE ..• The aptianal printer provides a
5. Press and hold both X-RAY bunons for the entire
0'
hard copy record patient demagraphics
and technique in'ormation when the
exposure duration - until the tone stops. printer is connected and selected (in 'he
~' NOTICE •.• The Compr••• ion Carriage will 3. Position the patient (i.e .• compression. AEC
automatically,ois" approximately 6cm sensor, etc.). Place the patient's arm or hand on
alter ,h.
exposure whon the compreuion the patient handle.
r.'ea•• method is sat to Automat;c. 110
Loca/i%ot;on Paelel/. i. us.d lor the exam, 4. Check all patient demographics and make the
the system aisoble. AlL the Compression necessary changes.
Release switches. Use a Compression UP
5. Evaluate the exposure technique factors and make
button, the 'ootswitch, or the handwheel
the necessary changes.
ta release the patient.
6. Verify that the M-IV status reads READY and
make the exposure.
Post-exposure Checks
1. Evaluate the post-exposure mAs value (automatic
modes only) for subsequent exposure technique
& WARNING I Observe ALL sa'ety precau-
tions when making x-ray expasures.
adjustments.
1. Enter a patient identification number (up to 15 6. Switch the Spot data field to indicate the focal
characters), then press Enter. spot size required for the exposure.
2. Enter the patient's name (up to 24 characters). 7. Select the film I screen combination being used
then press Enter. for the exposure.
5. Record the technologist identification number (up 2. Adjust the C-arm height to adapt to the patient.
to 3 characters), then press Enter.
3. Position the patient. Place the patient's arm or
6. If applicable. enter the five character CPT code, hand on the patient handle.
c 7.
then press Enter.
2. Press the light field button to illuminate the x-ray 2. Make any final technique changes and demo-
field an additional 30-seconds (if necessary). graphic corrections.
3. Press the Collimator Override button to adjust the 3. Move completely behind the radiation shield.
x-ray field for the exposure.
4. Instruct the patient to be as still as possible and to
4. Apply adequate manual compression to obtain a hold her breath during the exposure.
homogeneous imaging surface.
5. Press and hold both X-RAY buttons for the entire
NOTICE • •• Th. automatic collimator IIml,. exposure duration - until the tone stops.
fII. maximum me 01 tit. x-n.y field to the
';ze 01 ria. Image rec.ptor (i•••, 'Scm x .&. WARNING I Observe ALL sa/ety precau-
2""", 2"an" 30an) tions when making x-ray exposures.
Post-exposure Checks
1. Evaluate the post-exposure mAs value (automatic
modes only) for subsequent exposure technique
adjustments.
Clinical Prc,.,edu;res
c
(ML) or Lateromedial (LM) views, C-arm rotation 1:Z/12/1934 63 P 226mA. 0 3 f..Oem
memorization is preset. C-arm rotation automatically 3118/1997 :z.43pm 18x2f. BUCKY 123
••
Patient Age 3. Press the OFF Switch on the Control Panel.
Patient Gender
•• Institution Name
C-arm angle
•• Institution Address
Clinical View
•• Exposure Mode
kV
•• mAs
c ••
Density
Filter
AEC Sensor Position
••
Compression Force
Comment
•• Compression Thickness
Tech ID
• Unit Number
Figure 8-2:
AutoFilm ID Lobel
WARNING I Before cleaning anti tlisinfec- The following cleaning solutions have been found to be
tion, remove power from the equipment. compatible with polycarbonate compression paddles and
accessories:
ALL pans of the unit which come in contact with the • 70% (by volume) isopropyl solution
patient must be cleaned and disinfected prior to each
examination. • 90% maximum concentration of Hydrogen
Peroxide solution
The standard hospital regimen for disinfection should be
followed, with the following qualifications: NOTICE • •• A'ter cleaning with the above
solutions, the paddles MUST be washed
• DO NOT clean with solvent-based cleaners
(containing benzene, or stain removers) surfaces
on pans which contact the patient (i.e .• face
shield. cassette holders, Bucky devices, compres-
with soap and warm water.
c
sion trays).
PREFACE: I The CPT Setup.Setup screen permits the user to select clinical views for
standard and custom billing codes. The Tech 10 Setup screen permits
custom setup of technologist preferences. such as the view ordcr and the
compression mode.
OVERVIEW: I This first part of this chapter describes the CPT Setup Mode screen and
guides the user through the steps for selecting:
The second part of this Chapter describes the Tech ID Setup Mode screen
and guides the user (Mough the steps to:
c
OBJECTIVES: I After reading this chapter. the user will be familiar with:
• The basic layout and functions of the CPT Setup & the Tech ID
Setup mode screens.
• Defining clinical views for standard and custom billing codes.
• Setting Technologist ID data and defining preferences.
(
Chap-ter 9:
CPT Code & Tech ID Setup
(
(
CHAPTER
To view the billing codes, perform the following: The View Selection Fields
The center portion of the CPT Setup screen shows a
o Move the highlight to the CPT Code data field, or
listing of the 12 most common exam view labels and
press the Clear key to remove existing CPT data.
their descriptions. Selection boxes to the left of the
descriptions are used to select or deselect the appropriate
o Press the CPT Code key to scroll through the
exam views for the left and right breast for each dis-
available choices.
played CPT Code.
To clear the CPT Code and Exam Description data fields
The choices available for each selection box are:
for entry of a custom billing code, perform the following
steps:
• S - selected
o Press the Enter key until the highlight is in the • 0 - optional
CPT Code data field.
• X - not selected
o Press the Clear key to remove the existing data
Perform the following steps to add a custom CPT Code
from the two data fields.
to the CPT Code list.
o Press the CPT Code key until the desired billing o Use th'e arrow keys to move the highlight through
code appears. the list of exam views.
o Press the Shift and Clear keys together to delete o Label each exam view for the CPT Code by
the billing code and description. pressing the Change key to switch among the
three settings.
TECHNOLOGIST m
SETUP SCREEN
The user can assign Tech ID codes. and define individual
preferences for the Compression mode and the screening
exam's view order by accessing the Technologist ID
Setup screen. After an ID code is assigned and defined,
entering that code in the Run Mode's Tech ID 6eld
brings that users preferences to the appropriate fields in
the Run Mode.
c a
appears as highlighted choice.
Press Entcr to access the Tech ID Setup screen. • Technologist ID: this is a unique alphanumeric
ID that contains up to 3 characters
The Technologist Setup screen is divided into two main
sections: • Compression Mode: this data field permits
selection of one of three compression modes
• T he Setup Window (Full. Pre. Dual).
• The Message Window • First View: permits selection of the prefered first
view of a screening exam.
The Setup Window contains the data fields used to define
the preferences and view order for the selected technolo- • Second View: permits selection of the prefered
gist. second view of a screening exam.
The Message Window displays instructions for entering • Third View: permits selection of the prefered
the ID data and for selecting the view order for screening third view of a screening exam.
examinations. An example of the Technologist ID Setup
screen is shown in Figure 9-2. • Fourth View: permits selection of the prefered
fourth view of a screening exam.
r-.
'0 '
. '----~
c
92 M-W Operator's Manual