Professional Documents
Culture Documents
User’s Manual
MYTHIC 60 Copyright© Orphée SA. All Rights Reserved. Page 1/118
REVISIONS
CONTACT ADDRESS
MANUFACTURER
Manufactured in Germany for:
ORPHEE SA
19, chemin du champ des filles
CH-1228 Geneva / Plan-les-Ouates
SWITZERLAND
LOCAL AGENT
Indicates that wearing gloves is mandatory before performing the described operation due to
risk of contact with materials that may be infectious.
NOTE: Misuse of electrical equipment may cause electrocution, burns, fire and other hazards.
Connect MYTHIC 60 power supply block to main power supply outlet with an earth connection.
Preserve a good access to the supply outlet to be able to unplug MYTHIC 60 in emergency case.
Do not place the power supply adapter in liquid, nor put it where it could fall into liquid. If the power
Do not use MYTHIC 60 if it is not working properly or if it has suffered any damage (damage to
the supply cord or its plug; damaged caused by dropping the power supply adapter).
Do not let the power supply adapter or its flexible cord meet surfaces which are too hot to touch.
Do not use MYTHIC 60 where aerosol sprays are being used, or where oxygen is being administered.
Always switch off MYTHIC 60 and disconnect the power adaptor before dismantling any part.
NOTE: MYTHIC 60 is an automated hematology analyzer for in vitro diagnostic use in clinical
laboratories by a representative people.
Only human blood or artificial control and calibration blood must be run.
The optimum performances can be only achieved if the cleaning and maintenance procedures are
strictly respected.
Due to the use and the environment of this equipment, all parts and surfaces of MYTHIC 60 are
potentially infective. Wearing gloves and hands washing after work completion are strongly
recommended.
Basic safety precautions should always be taken. If the equipment is not used per the
The treatment of waste and the elimination of a part or the complete instrument must be done in
Any output or input connections (except the printer and the barcode reader) cannot be done without
representative authorization.
Do not open the door located on the right side of the instrument when hydraulic cycle is in progress,
2.1 Introduction
Mythic 60 is a fully automated analyzer performing hematological analysis on whole blood collected on
EDTA tubes.
Sampling volume is 15.6 µl and analysis cycle duration is 60 seconds.
NOTE: Result is displayed and printed before the end of the analysis cycle.
Here below the list of all parameters determined by the analyzer for each analysis:
1. Front cover.
2. Fluidic part.
3. Reagent area.
4. Connection board.
5. External power supply block.
6. Printer (optional).
7. External Barcode reader.
The fluidic part is located on the right side of the instrument directly accessible after opening of
the fluidic door.
The fluidic part consists of the six elements listed below:
1. Sampling module.
2. Syringe module.
3. Syringe valve module.
4. Counting chambers.
5. Counting valve module.
6. Optical Bench.
1
3
4 USB Ports
TP/TCIP Port
In the case of replacement of the main power cord supplied with MYTHIC 60, the
new cord must comply with the local regulation.
MYTHIC 60 have been certified with the power supply block provided with. Use with
another external power supply block is not guaranteed.
2.2.6 Printer
MYTHIC 60 is not delivered with the printer. The instrument is equipped already with PCL3 & PCL6
standard protocol printer drivers which cover compatible printer models. Here below examples of
printers compatible with the instrument. It is not an exhaustive list:
EPSON M100/M105/M200/M205
HP1102w/1106/1108
Lexmark CS310/MS410.
2.3 Configuration
• 1 analyzer
• 1 power cord
• 1 power supply block
• 1 flat screwdriver
• 1 certificate of approval
• 1 diluent pick up tubing
• 1 waste tubing
• Maintenance kit
• 1 USB Barcode scanner
3.1.1 Location
To ensure MYTHIC 60 fulfills its function, it must be installed on strong and stable table which
supports the weight of the unit as well as the one of the printer and reagents (around 30 Kg). Ten
centimeters space at the rear of the instrument and 30 centimeters on each side are required for
maintenance tasks and reagents change. Exposure to direct sunlight shall be avoided as well exposure
to close neon light or any apparatuses which could create magnetic or high frequency disturbances.
Note: If the ambient temperature moves more than 10°C during the working day, MYTHIC 60
must be calibrated more frequently.
Contact your distributor to use the MYTHIC 60 in other conditions than the ones
described above. Please take care of reagents temperatures constraints.
3.2 Unpacking
During device unpacking, person in charge of the installation must control the good presence of all
elements needed for installation comparing to the PACKING LIST.
Installation of the device must be done according to the procedure described hereafter.
If MYTHIC 60 has been stored at a temperature less than 10°C, it must stay at
room temperature during 24 hours to let the time all elements to reach the ambient
temperature before switching it on.
3.3 Installation
This instruction describes the different stages to follow for the physical installation of MYTHIC 60
instrument.
1. Remove the instrument from the cardboard and place it on a stable table.
Whether any element would be missing, contact immediately your ORPHEE distributor.
Before handling the reagents, read carefully their specifications described in section.
3. Connect the diluent pickup tubing to the dedicated hydraulic connector located at the back of
the instrument.
Warranty does not cover any defect, malfunction or damage due to failure to use the
inappropriate reagents or chemicals specified for the product.
Do not modify the type and the length of the diluent and waste tubes.
Collect the waste in a container preferably placed on the floor and treat it in
compliance with your local legislation.
For 20 liters of waste, produced by the MYTHIC 60, add 50 ml of a solution of sodium hydroxide
(NaOH) at 200g/l, mix the container, add 100ml of bleach at 36°C, mix the container, wait one hour
before empty it.
Only use this procedure if there is no specific local legislation for the treatment of
the waste.
MYTHIC 60 has been certified with the power supply block provided with.
Any use of another external power supply block could not be guaranteed.
In the case of change the power cord provided with the instrument, the replacing
cord must be in compliance with the local regulation and the instrument specification
in matter of consumption.
NOTE: The power supply block must be placed at the rear of MYTHIC 60 and if possible, upper
than instrument level to avoid any risk of contact with liquid.
• 100-240 VAC
• 50/60 Hz
• Single phase with ground.
12. Switch on the instrument by pressing the start button and wait for loading until the display of
the login screen.
Regarding login, two different levels of access are available, BIOLOGIST AND USER.
Thereafter, additional access level could be created.
USER level is the more limited in terms of access.
It is possible to create, change or modify ID and/or associated password for Biologist and User
access levels only (see describtion of ADMINUSER screen).
ARBORESCENCE ACCESS
MENU SUB-MENU FUNCTION OPERATOR ADMINISTRATOR TECHNICIAN
STARTUP
SHUTDOWN
CHANGE LOAD
QC LJ GRAPHS
RESULT
DETAIL
CHANGE
STATS
CALIBRATION RESULT Calibrate
LOAD
DEL.UNSELECT
REPEATABILITY
DETAIL
RUN SAMPLE NEXT SAMPLE
DATE
RESULTS
DETAILS
PRIME ALL
DILUENT PRIME
LYSE PRIME
REAGENTS CLEANER PRIME
WASTE RESET
Cycle Counters RESET
Reagents Log
SYSTEM INIT
EVENT LOGS
ERROR LOGS
BACKUP
BACKUP & RESTORE
RESTORE
LAB PREFERENCES
UNITS
CBC TRHESHOLDS AND
FLAGS
LAB PARAMETERS
DIFF TRHESHOLDS AND
FLAGS
REFERENCE RANGES
CALIBRATION FACTORS
SETTINGS DATE AND TIME
AUTOMATIC CYCLES
PRINTER SETTINGS
PRINTER
PRINTERS MANAGEMENT
COMMUNICATION COMMUNICATON SETUP
ADD
USERS MANAGEMENT CHANGE PASSWORD
REMOVE
SOFTWARE UPDATE
FLUIDIC CONTROL
SERVICE
BLEACH CLEANING
DRAIN BATHS
BACKFLUSH APERTURES
BACKFLUSH OPT.
BENCH
NEEDLE DISMANTLING
CHECK ROCKER
PARK MOTORS
RINSE
CLEAN
DRAIN FOR PACK UP
TROUBLESHOOTING
SYRINGE GREASING
DRAIN FLOW CELL
EV1-x
ALL EV ON
EV CHASER
CHECK SET HGB LED OFF
SENSORS/VALVES SET CO ON
SET OPT LED ON
TEST VACUUM
PUMP ON
CHECK SYRINGE
CHECK NEEDLE
The main menu is displayed immediately after validation of the login screen.
From this screen it is possible to access to the following sub-menus:
▪ STATS,
▪ RUN SAMPLE,
▪ RESULTS,
▪ REAGENTS,
▪ SERVICE.
Note : All menus and sub-menus are displayed with the same header as described below.
Make a printout.
▪ Confirmation button.
▪ Cancel button.
▪ Selection in tables.
Uncheck to unselect.
Check to select.
1. Select and by using Barcode Scanner fill in the fields Lot, Expiry, Capacity, SERIAL and
CODE. Information is provided on the label of diluent container in form of two barcodes.
2. Select and by using Barcode Scanner fill in the fields: Lot, Expiry, Capacity, SERIAL and
CODE. Information is provided on the label of lyse container in form of two barcodes.
By leaving the WASTE OPEN DRAIN option unselected , the waste capacity is required. With
this configuration, the waste capacity is managed and is required to be reset when the waste level is
ALMOST FULL OR FULL.
NOTE: It is recommended to use a waste container at least from the same capacity than the
diluent one.
Select on WASTE menu to reset the waste current level, the following window is displayed.
Selecting the WASTE OPEN DRAIN option removes the WASTE CAPACITY from the user interface.
Be sure that all reagents and waste tubes are properly connected before starting.
When the cycle LED turns red, no cycle can be performed before it turns back
green.
The reset of the Cycle Counter is only allowed to the Technician.
3.3.5 STARTUP
Startup cycle is dedicated to daily determine the background (blank) values of measured parameters.
It must be launched everyday before QC and then before patient’s analysis.
Press on .
The front cycle LED turns red , meaning no cycle can be launched
before it turns green.
Instrument will proceed first with counting chambers rinsing then,
1 to 3 blank cycles to control the background values.
If any level from any parameter is higher than expected value, the system warns the user with an alarm
message “STARTUP FAILED”, it is suggested in this case to perform a new start up.
NOTE: If the user chooses to run patient blood after a startup failed, all results will be displayed
and printed with the indication “Startup failed”.
Chapter 4. STATS
4.1 Introduction
To control the stability of the instrument before running patient samples, it is mandatory to run QC
bloods from different levels at the beginning of the working day. In case of result out of range on QC
bloods, it is recommended to perform a calibration, then run QC blood again.
4.2 QC
QC menu is dedicated to Quality Control of the instrument. MYTHIC 60 can store in internal memory,
up to 100 results (results and histograms) per lot for 12 different lots. Results of each QC lot and level
can be viewed in tables and as Levey-Jennings graph.
Press STATS on the main menu, then QC, the following screen is displayed.
This screen is the table of the entire existing QC runs already recorded in the instrument.
4.2.1 Change
This option allows the user: To modify data of an existing QC after selection in QC table.
To create a new item of QC after selection of a free field in QC table.
In case of modification of an existing QC, all recorded results linked with this
modified control lot number will be erased.
1
Enter the QC Lot number.
2
Select the QC Level among the options L (low), N (normal) or H (high).
Option allows importing QC data from usb drive instead of using touch screen keyboard.
4.2.2 Load
How to proceed to import QC from usb drive:
Units format must be specified in order to the system to know which unit is used
Unit;1 stands for US units.
Parameters must be specified in US units ONLY.
4.2.3 Results
This screen is dedicated to:
• See results of a QC after selection in QC table.
• Run a QC analysis, after selection in QC table.
The column on the left shows each parameter with the target values and the limits.
The value under the parameter name is the result of the QC run where the cursor is located.
The key enable displaying following pages giving access to all parameters view.
The keys and enable displaying first or last page directly.
The keys and enable to move the cursor from QC result to another one.
The keys and enable to move the cursor to the first QC result or the last one directly.
4.2.5 Details
This screen allows to see the detailed result of a selected QC in RESULTS table.
Note: To select a QC results in RESULTS table, first press on the QC line at the place you
want to see the details, then select DETAIL.
QC lines selected.
4.3 Calibration
• Lot number.
• Expiry.
• Date.
• Operator name.
• Target and ranges values for calibrated parameters.
• Current coefficients from last calibration.
4.3.1 Change
To enter new data or to modify existing ones, press , the following screen is displayed.
1
Enter the Calibrator Lot number.
2
Enter the Calibrator expiration date.
3
Enter targets and limits values for all parameters.
4
Select to confirm modification, the following window is displayed.
4.3.2 Results
This screen is dedicated to:
• See results of last Calibration.
• Run a new Calibration.
1. Select , following screen is displayed and sampling needle go down in sampling position.
NOTE: At least three consecutive runs are mandatory to validate the Calibration.
Displayed down in the screen, automatic statistical calculations are implemented also
after run.
NOTE: fields are calculated functions of the selected results ( ) in the table.
In the field below: N is the selected number of results, LOT and EXPIRY information from calibrator
used.
N= 3
LOT PUUS1116
EXPIRY 05/01/2017
The calibration blood must be used before its expiry date, be mixed and stored in
accordance with the instructions of use recommended by the manufacturer.
7 Press to validate.
6. Select to exit without saving.
7. Press on to restart the calibration, the following window is displayed.
To load calibrator data from Udrive, press on , the following screen is displayed.
1. Under the directory MYTHIC 60, previously created for QC, create the directory Cali.
2. Under the directory Cali, place the Calibration file.
Units format must be specified in order for the system to know which unit is used
Unit;1 stands for US units.
Parameters must be specified in US units ONLY.
Mythic 60
1. Place the sample tube in sampling position and press the start cycle trigger.
2. The cycle LED flickers alternatively red to green, when it becomes red only the tube can be removed.
3. Repeat the operation as desired (maximum 50 runs).
4. Statistical calculations are automatically carried out with each run.
5. The column # allows to validate or to unselect a result.
6. The button allows deleting the unselected result(s).
5.1 Introduction
The human venous blood sample must be collected in an EDTA K2 or K3 (Ethylene Diamine Tetracetic
Acid, two or tri potassic) tube in sufficient quantity. K3 is recommended. Time between blood collection
and measurement on the Mythic 60 should not exceed 24 hours for optimal results. If the blood was
cooled, please remove it from cooler and keep it at room temperature for minimum 20 minutes in
vertical position.
Sample must be correctly homogenized before analysis. It is recommended to use a rotary agitator,
turning between 20 to 30 turns / minute, for 10 minutes.
The insufficient blood volume for the quantity of anticoagulant or a bad mixing
may involve an erroneous result.
If the room temperature moves more than 10°C during the working day,
MYTHIC 60 must be calibrated again.
2. The needle goes down in sampling position and the following screen is displayed.
Select successively fields day, month, year and ajust with arrows.
Note: It is not possible to enter a date <1900, future or not realistic date.
10. Mode field, to select blood type among DIF WB for DIF whole blood mode, DIF PD for DIF
pre-diluted mode, CBC WB for CBC Whole Blood or CBC PD for CBC pre-diluted mode.
11. Gender field, to select gender among F for female, M for male or U for unknown.
12. To valid the entry and return to the previous screen, press , following screen is displayed.
14. The mode is also accessible from the mode button localized on the RUN SAMPLE screen.
The user can then select among the four available modes.
Once the mode is selected, the button label is updated with the following name: DIF WB, DIF PD,
CBC WB or CBC PD.
Pre-dilution:
If the Pre-diluted mode is selected, the following prompt is displayed and the DIL DISPENSE
button is displayed on the RUN SAMPLE screen.
Clicking on the DIL DISPENSE button allows dispending 300µL to prepare the dilution.
Wear rubber gloves and wash hands with a disinfectant after completion of work.
1. If the needle is not down in sampling position, press on start cycle trigger and needle will go
down in sampling position.
2. Place the tube in order to introduce the needle two-thirds inside the blood volume and press
on start cycle trigger.
two-thirds
of the Blood volume
needle inside
NOTES:
Note: If you press anywhere on System Alerts field, an enlarged window is displayed. In this
window, the messages are displayed in expanded version. For Example, “QC Not Done” will be
displayed as “QC Not Done During The Day”.
Note: If you press anywhere on DIF Plot or histograms fields, a zoom of the one selected is
displayed.
• Over range on linearity high limit: Result is displayed such as “XXXX +”.
• Over range on reportable limit: Result is displayed such as “++++”.
• Over range on extreme limits (user controlled): Result is displayed such as “XXXX L/H”.
• Over range on normal limits (user controlled): Result is displayed such as “XXXX l/h”.
• Rejected result: Result is displayed such as “XXXX *”.
• Suspicious result is displayed such as “XXXX!”.
• Invalid result is displayed such as “----“.
The counting as well as the discrimination of the cellular elements of the blood sample is based on The
Coulter Principle.
This technics is based on the modification of the impedance of a calibrated aperture soaking in an
electrolyte and going through a constant current delivered by two electrodes located on both sides of
the aperture.
A vacuum applied on a side of the aperture allows the cells passage. Cells oppose their physical volume
to the current passage. A voltage impulse is registered at the electrodes terminal. The height of this
impulse is proportional to the cell volume.
The principle of this technology is based on the introduction of the sample solution flow in the flow cell
with low pressure and the introduction of the diluent sheath with a high pressure. Thus, the diluent
sheath drives the sample flow straight across the cuvette through the detection area.
The main advantages are:
Hydraulic connection
Nozzle holder
Hydraulic connection
Absorbance
Eosinophils Neutrophils
Lymphocytes Monocytes
Diffraction
Basophils
The hemoglobin measurement is done directly in the WBC chamber, by spectrophotometry ( = 545
nm).
Hemoglobin is detected by formation of oxyhemoglobin.
HGB blank check is performed at each STARTUP cycle.
HGB blank measurement is done also at the beginning of each analysis cycle.
By comparison of the two values, it is possible to follow potential evolution of the blank value in order
to warn the user if necessary.
The White Blood Cell Total Count is obtained by Impedance metric in the WBC counting chamber; the
other ten parameters are obtained by flow cytometry measurement.
L1 L5
WBC histogram is displayed after pressing on DIF Plot on the result screen.
L1 and L5 thresholds are displayed on curve in full vertical blue color line.
NOTE: First peak on the left side of the histogram represents the lymphocytes cells, the other
one located on the right side represents all the others WBC populations.
WBC 5 differential absolute values and percentages are obtained by optic measurement.
The measured pulses on the two optical channels are displayed on DIF Plot ALL (Y axis) and
FSC (X axis).
Each dot on the DIF Plot represents the height in ALL and FSC of each pulse.
The erythrocyte analysis is done by impedance metric in RBC counting chamber. Eight parameters are
obtained:
R1 R2
Hematocrit (HCT) is measured by integration of the volume of the red blood cells which flow in
the RBC counting chamber aperture.
𝐇𝐂𝐓 𝐱 𝟏𝟎
𝐌𝐂𝐕 =
𝐑𝐁𝐂
The RBC distribution curve analysis allows the measurement of RDW-CV which is an expression
𝐤 𝐱 𝐒𝐃
𝐑𝐃𝐖 =
𝐌𝐂𝐕
Mean Corpuscular Hemoglobin (MCH) calculation is obtained from HGB and RBC by the following
formula:
𝐇𝐆𝐁 𝐱 𝟏𝟎
𝐌𝐂𝐇 =
𝐑𝐁𝐂
Mean Corpuscular Hemoglobin Concentration (MCHC) is calculated from HGB and HCT by the
formula below:
𝐇𝐆𝐁 𝐱 𝟏𝟎𝟎
𝐌𝐂𝐇𝐂 =
𝐇𝐂𝐓
RDW-SD is an actual measure of size. It is derived by finding the width at the 20% height of
the distribution histogram.
RDW-CV is determined by taking the standard deviation of RDW-SD and the mean corpuscular
volume (MCV) number.
Parameters Pathologies
PLT Platelets Thrombopenia PLT<PLT b
Thrombocytosis PLT>PLT h
MPV Mean Platelet Volume Giant platelets MPV> MPV h
Small platelets MPV<MPV l
PCT Thrombocrit
PDW Platelet Distribution Width
PLCR Platelet Large Cell Ratio
CP1 CP2
P CP3
𝐏𝐋𝐓 𝐱 𝐌𝐏𝐕
𝐏𝐂𝐓 =
𝟏𝟎𝟎𝟎𝟎
6.7 Alarms
MYTHIC 60 manages different alarms. These alarms allow user to be alerted if there is a mistake
which can affect the quality of the results. These alarms appear on screen at the right of the result.
NOTE: Whenever one of the following symbol is triggered: *, -, +, ++++, display tagged result
in bold (flag included).
MYTHIC 60 Copyright© Orphée SA. All Rights Reserved. Page 54/118
Whenever one of the following flag is triggered: L, l, h, H, display tagged result normally and
flagged in bold.
Change color of all flagged result to a specific color in accordance to the UI specifications (flag
included):
When result is flagged with: *, +, ++++,!, the parameter printout on table view is the parameter
value followed or replaced (++++) by the appropriate symbol.
When result is flagged with L,l, h or H, the parameter printout on table view is in reverse video
only.
L1 L5
Two editable thresholds L1, L5 enable to discriminate the resistive count flags.
MORPHOLOGICAL ALARMS
On Curve
Alarm Short
Trigger Condition Alarm Flags enlarged
version
Review on Smear:
Determination of the population defined from channel
Small lymphocytes,
CL1 to CL1-2.
Probable Incomplete,
The alarm is raised if this population is higher than L1
Erytrolysis,
an absolute limit AND a percentage of the
Platelet Aggregates,
lymphocytes population.
Erythroblats.
Determination of the population defined by CL5
threshold and curve ending.
Review on Smear:
The alarm is raised if this population is higher than L5
Immature Cells.
an absolute limit AND a percentage of the WBC
population.
MORPHOLOGICAL ALARMS
On DIF Plot
Trigger Condition Alarm Short version Alarm Flags enlarged
Presence of cells in the zone Review on Smear:
regarding the total number of N1 Debris or Small Cells,
leukocytes in the DIF Plot. Platelet Aggregates.
Review on Smear:
Presence of cells in the zone Probable Incomplete,
regarding the number of N2 Erytrolysis,
lymphocytes. Platelet Aggregates,
Erythroblasts.
Review the DIF.
Presence of Immature Cells (from
IMM Review on Smear:
the mono or polynucleated cells line).
Immature Cells.
Review the DIF.
Presence of atypical lymphocytes. ALY Review on Smear:
Atypical Lymphocytes.
Review the DIF.
Presence of atypical lymphocytes or
RL Review on Smear:
basophiles.
Right Lymphocytes.
Presence of basophiles, small
Review the DIF.
neutrophils (without granulations or HL
High Lymphocytes.
few segmented), band cells.
Presence of small neutrophils
(without granulations or few Review the DIF.
NL
segmented), band cells or hyper Low Neutrophils.
basophil monocytes.
Two editable thresholds R1 and R2 enable to discriminate between microcytes, normocytes and
macrocytes. CR1 CR2
100%
68.26% of total
distribution area
RDW
RDW-SD
20%
MCV
MORPHOLOGICAL ALARMS
On The Curve
Trigger Condition Alarm Short version Alarm Flags enlarged
Determination of the population defined from
channel 0 to CR1.
The alarm is raised if this population is higher than R1 Microcytes
an absolute limit OR a percentage of the RBC
population.
Determination of the population defined by CR2
threshold and curve ending.
The alarm is raised if this population is higher than R2 Macrocytes
an absolute limit OR a percentage of the RBC
population.
MORPHOLOGICAL ALARMS
On The Curve
Trigger Condition Alarm Short version Alarm Flags enlarged
Determination of the population defined from channel 0
to CP1.The alarm is raised if this population is higher
than an absolute limit OR a percentage of the PLT P1 Platelet Debris
population.
The fluidic part is located on the right side of the instrument directly accessible after opening the
fluidic door.
The fluidic part consists six elements listed below:
1. Sampling module.
2. Syringe module.
3. Syringe valve module.
4. Counting module.
5. Counting valve module.
6. Optical Bench.
1
3
6.9 Software
MYTHIC 60 software runs using the Linux operating system. The software is embedded on a Q7 CPU
board.
This board is equipped with an Intel ATOM CPU (32 bits), RAM memory and flash where software and
data are stored.
allows to opens a contextual menu for actions linked to the current menu (save, delete, print,
sent…).
An exit button allows closing the tools window and returning to current menu.
allows coming back to the MENU display where ever you are in the arborescence.
• Instrument Model.
• Operation System version.
• Software version.
• FPGA / uBlaze versions.
• Algorithm (dynamic clustering version).
• Cycles modules.
• Fluidic module: N/A.
• Instrument’s Serial Number.
• User - Login level.
• IP address.
6.9.2 Windows
Mythic 60 displays few types of window to communicate with the operator in case of needs.
1. Confirmation message.
2. Information message.
7.1.1 Linearity
Linearity of WBC, RBC, HGB, HCT and PLT will be verified using linearity kits.
HCT % 5 – 70 ± 2 or ± 3% 0-80
MPV fL 5 – 25 ± 1 or ± 10% 0 – 25
7.1.3 Precision
Short term (within run) imprecision will be tested by assaying the same normal whole blood (collected
in K2EDTA) specimen consecutively 20 times.
WBC / DIFF
> 0 < 0.2 >15 <1%
(103/µL)
US
Parameters SI SI MOD JAPANESE
(Standard)
LYM, MONO, NEU, EOS, BASO, ALY, IMM # XXX.X XXX.X XXX.X XXX
103/µL 109/L 109/L 102/µL
LYM, MONO, NEU, EOS, BASO, ALY, IMM % XX.X XX.X XX.X XX.X
% % % %
General:
Ambient temperature: from 18 to 32°C.
Relative Humidity: maximum 80% at 32°C.
Storage temperature: -10 to 50°C.
If MYTHIC 60 has been stored at a temperature less than 10°C, it must stay at
room temperature during 24 hours before switching it on.
If the room temperature moves more than 10°C during the working day, MYTHIC 60
must be calibrated.
- Dimensions: height: 405 mm (approx.) * width: 270 mm (approx.) * depth: 430 mm (approx.).
- Weight: 12kg (approx.).
- Power supply input: 100-240VAC 50-60Hz.
- Power supply output: 24V – 6.75A.
- Electric consumption: 160W
- Screen: LCD Touch-screen 8.4 inch (800*600).
- Memory capacity: 35000 Files (demographics, results and histograms).
QC: 12 lots (100 results per lot).
- Connection: 5 USB ports/ Ethernet-RJ45/ Serial LIS-SUB D9, PC connection/Serial Port.
Dimensions:
• Height: 31 mm (approx.)
• Width: 58,5 mm (approx.)
• Depth: 132 mm (approx.)
• Weight: 0,35 kg (approx.)
• Power supply Input: 100 to 240V AC - 1,5A, 50-
60Hz
Application: The diluent is used to carry out the necessary dilutions for the measurement performed
by the MYTHIC 60 (see section 6).
Storage: At temperature between 18C and 30C until the expiry date printed on the bottle label.
Precautions:
Classification according to Regulation (EC) No. 1272/2008 [CLP]/GHS: Not classified.
To our knowledge, this product does not present any particular risk, provided it is handled in
accordance with good occupational hygiene and safety practice.
This mixture does not contain any substances to be mentioned according to the criteria of section 3.2
of REACH annex II.
Application: OnlyOne Lytic reagent is used as unique leukocyte and hemoglobin reagent to lyse red
blood cells, enable cells subpopulations differentiation and counting and quantitatively
determine hemoglobin content of blood samples on MYTHIC 60 (see section 6).
Storage: At temperature between 4C and 25C, until expiry date printed on the bottle label.
Precautions: Avoid contact with skin or eyes. Do not swallow or inhale. DO NOT REUSE THE
CONTAINER.
To our knowledge, this product does not present any particular risk, provided it is handled in
accordance with good occupational hygiene and safety practice.
Application: The cleaning solution is used to carry out the cleaning of the measurement system and
hydraulic circuit (see section 6).
Storage: At temperature between 18C and 30C, until expiry date printed on the bottle label.
Precautions: Avoid contact with skin or eyes. Do not swallow or inhale. DO NOT REUSE THE
CONTAINER.
To our knowledge, this product does not present any particular risk, provided it is handled in
accordance with good occupational hygiene and safety practice.
Application: Mythic 18-22 FLUSH - CLEANER is hypochlorite-based reagent. Its function is periodic
and routine cleaning of the measuring system of haematology analysers (see section 9.7.2).
Storage: At temperature between 2°C and 25°C, until the expiry date printed on the bottle label if
closed.
Precautions:. Product is classified as causing serious eye damage, irritating to skin and acute toxic to
aquatic life. Contact with acids liberates toxic gas. For detailed precautions please refer to relevant
Material Safety Data Sheet.
MAINTENANCE Recommendations: Please respect the maintenance procedure, calibration and the
quality control procedure. Otherwise, results can be affected.
GENERALITIES: Some abnormal samples may give incorrect results by automated cell counting
methods. The following table shows examples of specific specimens that could cause errors.
Each result for a new patient out of lab linearity limits or with an alarm must be
checked with a conventional method or checked with blood smear.
7.6.1 Interferences
8.1 Introduction
MYTHIC 60 can save in internal memory up to 35 000 patient reports with results and histograms.
Here below explanations how to use results areas to search, print, send, save or erase results.
8.2 Results
All parameters are listed in first column, results of first tube of the day in second column, results of
second tube of the day in third column, etc.
The keys and allow moving from first to last page of results of the day.
The keys and allow moving from a patient file to another one.
The keys and allow moving from beginning to the end of the list of parameters.
The field under the table, describes NAME, PID, SID, Seq number, birth date, gender, mode, date
& time of the selected patient file.
To select a result in the table, select the corresponding column, result will appear in reverse video.
Once in detailed view, use the keys moving in detailed views as the same way than
on results table view.
9.1 Introduction
NOTE: Can be used whether the instrument is switched on with the fluidic door opened and due
to that, did not performed the automatic mechanic initialization.
9.3 Logs
This option allows to backup and/or to restore the patients result (.csv format) the set up and
adjustments.
9.5.1 Backup
In BACKUP field check the options you want backup “Data and Setup” and/or “.csv file”, then select
.
9.5.2 Restore
In restore field, press on , the following window is displayed.
9.6 Settings
This menu is dedicated to set all the instrument with different options as printer, communication, limits,
flags, etc.
Select from menu, the following screen is displayed.
Auto Increment
Check in box if you want auto-increment of SID from start number specified.
QC
Check in the box if you want QC alert displayed and printed on patient files (QC Not Done and
QC Failed).
RUO
Check in the box if you want to display the RUO parameters.
This screen gives the possibility to the user to choose the units among the following list
Select the model of unit you want use, then press to valid, the following window is displayed.
Thresholds modification can affect the quality of the results or can affect the alarm
detection. We recommend to modify these values only after training.
NOTE: In both CBC & DIF thresholds and flags menus, the button allows returning
to the original values.
Once modifications are done, press to validate or to exit without any modification.
The modification of these factors without running a calibration blood could affect
the quality of the result.
If the calibration factors are manually modified, the calibration information in the
calibration screen will be flagged with “MODIFIED” and previous calibration results
are modified
If the calibration factors are manually modified, the following prompt is displayed to
inform the biologist that the calibration results will be removed.
To change the date, select the field among day/month/year and change the value using the
arrows or enter it manually using the displayed numeric keyboard.
To change the date format, press on the arrow and choose the desired format.
D for day
M for month
Y for year
To change the time, select the minute or hour field and change the value using the arrows or
enter it manually using the displayed numeric keyboard.
To change the time format, select the arrow and choose the desired format.
Once modifications are done, press to validate or to exit without any modification.
To enable automatic power up setting, Select Enable and choose the time and date for the automatic
power up.
To modify the autoclean frequency, adjusted by default to be done every 25 analyses, the value can
be set from 10 to 50.
25
If you want set an automatic shutdown, check in the option enable and enter the time chosen.
NOTE: Time is expressed in minutes and shutdown frequency can be set from 30 to 720.
Once modifications are done, press to validate or to exit without any modification.
Report Headers: allows to enter the laboratory header using touch screen keyboard.
Patient Report Options: options selected will be printed on the patient report.
Control Report Options: allows to select LMG format, percentage or absolute value for all
analysis and not only for control.
Once modifications are done, press to validate or to exit without any modification.
MYTHIC 60 Copyright© Orphée SA. All Rights Reserved. Page 90/118
9.6.4.2 Printer Management
This screen is dedicated to the printer installation.
It must be used the first time the printer is connected to the instrument, following the instruction
below.
Connect the printer to an USB port located at the rear of the instrument.
Switch ON the printer.
Select , the following screen is displayed.
No available in this
screen.
NOTE: It needs few seconds for the instrument to detect the printer as in the example below.
NOTE : As in the exemple above, if the field of the printer, connected and detected, is not
present in the list, choose pcl3 or pcl6 drivers depending from printer compatibility.
Generally: pcl3.ppd in dedicated to black & white printers and pcl6.ppd for colors printers.
Exit the menu pressing PREVIOUS, then go back and check that the printer is well installed.
9.6.5 Communication
This menu allows to setup the connection between MYTHIC 60 and the Host Computer of the Lab.
Select from menu, the following screen id displayed.
Communication field allows the user to select the type of communication required.
NOTE: MYTHIC 60 can be connected to the host computer in two different ways:
1. Serial link (type RS232),
2. Ethernet.
3. Select NONE if no communication is required.
allows to configure the communication data accordingly to the host computer data.
This menu allows to add, to modify or to delete an operator and/or a biologist login information.
Select from menu, the following window is displayed.
9.6.6.3 Remove
Select to delete the selected user, the following window is displayed.
NOTE: Software upgrade can take few minutes, do not remove the USB drive or power off the
instrument until the following message.
This menu gives access to different fluidic options as well as mechanical checks.
Select , the following screen is displayed:
option allows performing a complete control cycle of the mechanic and the fluidic
of the instrument in case of needs. A control bar indicates the progress of the cycle.
NOTE: Control cycle allows to control mechanic, hydraulic and electric functions to completely
reinitialize the instrument when it is needed.
2. Instrument begins to drain the counting chambers and the following window is displayed.
3. Open the fluidic door and put 4ml of Flush-Cleaner in each counting chamber. Close the fluidic
door when bleach dispense is done, then press . The system will perform fluidic actions cleaning
the mains counting module parts like apertures, counting chambers and optic flow cell.
NOTE: On each counting bath, an arrow indicates the volume of 4ml for bleach.
The bleach cycle takes 15 minutes to be completed.
It is not possible to use the instrument during that time.
MYTHIC 60 Copyright© Orphée SA. All Rights Reserved. Page 98/118
9.7.3 Drain Baths
NOTE: This option is principally used by Field Service Engineer in case of parts replacement
which does not require a complete instrument draining.
*Backflush onto the apertures: The system push cleaner under pressure into the WBC and RBC
apertures in order to remove a potential clogging. It can be used in case of permanent rejection for
one or few measured parameters.
option allows to perform a backflush onto optic flow cell* in case of needs.
*Backflush onto optic flow cell*: The system push cleaner under pressure into the flow cell in order to
remove a potential clogging. It can be used in case of permanent rejection for optic measurement.
option allows moving automatically the sampling module to give the access to the
needle in case of needs (needle and/or rinse head O-ring replacement).
Hereafter the instruction to replace the needle and/or the rinse head O-ring.
2. Pull on the top of the needle to remove it from the needle carriage.
4. Pull up the needle to remove it from the rinsing head. Disconnect it from tubing and remove
it from the instrument.
Wear rubber gloves and wash hands with a disinfectant after completion of
work.
option allows controlling the functionality of the sampling module transversal axis; it
checks the motor and sensors.
option allows to place the syringe pistons into the maximum high position.
NOTE: This option is mainly used for instrument transportation in order to protect the
mechanisms.
9.7.9 Rinse
option allows the counting chambers rinsing, following tool bar indicates the
progress of the rinse cycle.
At the end of the rinse cycle the following window indicates the user that rinse cycle is completed.
9.7.10 Clean
option allows cleaning the apertures with cleaner. It performs three back flush,
a drain chambers and refill.
NOTE: This option must be always used before a long term shut down.
option allows moving the pistons down in order to perform the greasing.
NOTE: The piston greasing must be performed minimum every 6 months, please proceed as
describe in the procedure below.
2. Wear rubber glove on one hand and place a bit of silicon grease at the tip of the index.
NOTE: With a tool key type T20, turn the two bigger pistons (waste pistons) in order to
spread the grease all around the pistons.
HOMES STATES
SWITCHS
COUNT
Instrument temperature.
VALVES
CHECK DEVICE
option allows controlling the syringe module functionality, this function checks the
syringe motor and sensor.
option allows controlling the functionality of the Y axis of the sampling module, this
function checks the motor and sensors.
The quality of the results and the reliability of MYTHIC 60 are directly linked to the strict respect
of the maintenance hereafter described.
NOTE: This maintenance table is dedicated to the user and to FSE. It is established an average
workload of 50 daily patients. For bigger workload, please increase proportionally the frequency of
maintenance actions.
Needle
Motor
Pistons O-ring
Startup Shutdown Bleach screw
greasing replaceme
greasing
nt
U
Daily
Weekly
Semi
annually
Annually
This Chapter allows knowing what to do when a troubleshooting message appears on the screen.
In any case, if a problem is not solved, contact your distributor.
System detected optical stop cock Open the stop cock if it’s closed.
as closed. Check that the optical bench isn’t fouled up.
System detected that temperature Wait a moment, the heater will reach set
is < to 18°C or > to 36.5°C. points.
System detected a reagent Wait a moment, the heater will reach set
temperature that is < to target – points.
2.5°C.
System communication with I/O Check HGB and FSC/ALL boards connection.
board is out at automaton power on. Reboot the automaton.
Auto clean cycle initiated. Press on and wait the end of the cycle.
Trying to run a cycle while another Press on and wait the end of the cycle in
one is in progress. progress before to run another cycle.
Cycle stopped upon user request Press on and perform a control cycle.
(emergency stop button).
System detects waste container Press on , empty the waste container and
full. reset waste level in reagent menu.
The adjustment of the HGB LED is 1. Verify that WBC bath is filled with
failed. diluent during HGB LED adjustment.
2. Change HGB board.
Note: HGB LED light flux must be 3. Change WBC bath.
included in the range [18000 ;
22000] after gain adjustment.
Cycle prime all is mandatory after Press on and perform a prime all cycle.
clean out.
Rinse cycle is mandatory after drain Press on and perform a rinse cycle.
cycle.
System detected lot number already Redo using another lot number.
used.
System detected default during Correct the default entering good data.
reagent data recording.
System detected diluent reagent is You still can perform ten analysis then
almost empty. system will generate diluent is empty.
System detected failure during Correct the default entering good data.
reagent data recording.
System detected cleaner bottle You still can perform ten analysis then
almost empty. system will generate cleaner is empty.
System detected lyse bottle almost You still can perform ten analysis then
empty. system will generate lyse is empty.
3. The hydraulic circuit of the counting and the optic modules will be rinsed.
4. Counting chambers will be filled with cleaner reagent.
5. At the end of the shutdown cycle, MYTHIC 60 will automatically turn off.
6. Shut Down can be set to be automatically performed (see ).
NOTE: After a shutdown, it is not possible to perform an analytical cycle without performing
startup first.
Every 24 hours MYTHIC 60 must stay at least with cleaning solution, for three
hours.