You are on page 1of 63

Chapter 1

INTRODUCTION TO
INSTRUMENTATION
OBJECTIVE
• At the end of this chapter, we will be able to
understand-
1. Static and Dynamic characteristics of an
instrument.

2. Calculate and analyze the measurement error,


accuracy, precision and limiting error.

3. Describe the basic elements of electronic


instrument.
2
INTRODUCTION
• Instrumentation is a technology of measurement which
serves sciences, engineering, medicine and etc.

• Measurement is the process of determining the amount,


degree or capacity by comparison with the accepted
standards of the system units being used.

• Instrument is a device for determining the value or


magnitude of a quantity or variable.

• Electronic instrument is based on electrical or electronic


principles for its measurement functions.

3
FUNCTION AND ADVANTAGES
• The 3 basic functions of instrumentation :-
– Indicating – visualize the process/operation
– Recording – observe and save the measurement reading
– Controlling – to control measurement and process

• Advantages of electronic measurement


– Results high sensitivity rating – the use of amplifier
– Increase the input impedance – thus lower loading effects
– Ability to monitor remote signal

4
Typical Measurement System Architecture
Noise and Interference

Proce
Proce Sensor
Signal
ss or Amp Conditioner
ss Transducer
or
or
Test
Test
ADC
Converter
OUR TOPIC IS HERE
Proces
s
PC
Controller comp
… and control and
over the process or experiment data
storage
5
Examples of Electronic Sensor applications

Uses infrared optical sensor


New Solar Power Faucet
• Electronic sensor automatically turns
water on/off
• Integral temperature control

6
PERFORMANCE CHARACTERISTICS

• Performance Characteristics - characteristics that show the


performance of an instrument.
– Eg: accuracy, precision, resolution, sensitivity.
• Allows users to select the most suitable instrument for a
specific measuring jobs.

• Two basic characteristics :


– Static - measuring a constant process condition.
– Dynamic - measuring a varying process condition.

7
PERFORMANCE CHARACTERISTICS
STATIC

• Accuracy – the degree of exactness (closeness) of


measurement compared to the expected (desired) value.
• Resolution – the smallest change in a measurement variable to
which an instrument will respond.
• Precision – a measure of consistency or repeatability of
measurement, i.e successive reading do not differ.
• Sensitivity – ratio of change in the output (response) of
instrument to a change of input or measured variable.
• Expected value – the design value or the most probable value
that expect to obtain.
• Error – the deviation of the true value from the desired value.

8
ERROR IN MEASUREMENT

• Measurement always introduce error


• Error may be expressed either as absolute or percentage of error

Absolute error, e = Yn  X n
where
Yn value
– expected
X n – measured value

Y  Xn
% error = n 100 = Absolute _ error 100
Yn Expected _ value

9
ERROR IN MEASUREMENT

Relative accuracy, Yn  X n
A 1 
Yn
% Accuracy, a = 100% - % error
= A 100
If measurement is accurate, it must also be precise.

Precision, P = 1  X n  X n
where Xn
-X value of the nth measurement
n
- average set of measurement
Xn

10
The precision of a measurement is a quantitative or numerical
indication of the closeness with which a repeated set of
measurement of the same variable agree with the average set of
measurements.

Example 1.1

Given expected voltage value across a resistor is 80V.


The measurement is 79V. Calculate,
i. The absolute error
ii. The % of error
iii. The relative accuracy
iv. The % of accuracy

11
Solution (Example 1.1)

Given that , expected value = 80V


measurement value = 79V

i. Absolute error, e = Yn  X n = 80V – 79V = 1V


Yn  X n 80  79
ii. % error = Y 100 = 100 = 1.25%
n 80

Yn  X n
iii. Relative accuracy, A 1  = 0.9875
Yn

iv. % accuracy, a = A x 100% = 0.9875 x 100%=98.75%

12
Example 1.2

From the value in table 1.1 calculate Table 1.1


the precision of 6th measurement? No Xn
1 98
2 101
Solution
3 102
4 97
the average of measurement value 5 101
98  101  ....  99 1005 6 100
Xn   100.5
10 10 7 103
8 98
the 6th reading
9 106
Precision = 1  100  100.5 1  0.5 0.995 10 99
100.5 100.5

13
LIMITING ERROR

• The accuracy of measuring instrument is


guaranteed within a certain percentage (%) of full
scale reading
• E.g manufacturer may specify the instrument to be
accurate at 2 % with full scale deflection. This
specification is called limiting error.
• For reading less than full scale, the limiting error
increases

14
LIMITING ERROR (cont)

Example 1.6

Given a 600 V voltmeter with accuracy 2% full scale.


Calculate limiting error when the instrument is used to measure a
voltage of 250V?

Solution

The magnitude of limiting error, 0.02 x 600 = 12V


Therefore, the limiting error for 250V = 12/250 x 100 = 4.8%

15
LIMITING ERROR (cont)

Example 1.7

Given for certain measurement, a limiting error for voltmeter at


70V is 2.143% and a limiting error for ammeter at 80mA is
2.813%. Determine the limiting error of the power.

Solution

The limiting error for the power = 2.143% + 2.813%


= 4.956%

16
Exercise
• A voltmeter is accurate 98% of its full scale
reading.
i. If the voltmeter reads 200V on 500V range, what
is the absolute error?
ii. What is the percentage error of the reading in (i).

17
Significant Figures

• Significant figures convey actual information regarding the


magnitude and precision of quantity
• More significant figure represent greater precision of
measurement

Rules regarding significant figures in calculation

1) The number of decimal places in the result equals the number


of decimal places in the least precise measurement.
2) For addition and subtraction, answer must show the same
number of decimal places as the number in the calculation with
the least number of decimal places.

18
Significant Figures

Example 1.3
Find the precision value of X1 and X2?

===>> 2 s.f
===>> 3 s.f

X n 101
X 1 98
X 2 98.5

19
Solution (Example 1.3)

X n 101
X 1 98 ===>> 2 s.f
X 2 98.5 ===>> 3 s.f

98  101
X1  Precision = 1 0.97
101
98.5  101
X2  Precision1=
 0.===>more
975 precise
101

20
Significant Figures (cont)

Example 1.4
V1 = 6.8 V
+ V2 = 11.934 V

Therefore VT = 18.734 V
 18.7 V

21
Significant Figures (cont)

2) For multiplication and division, retain only as many


significant figures as the least precise quantity contains

Example 1.5
From the value given below, calculate the value for R1, R2 and
power for R1?

I = 0.0148 A ===> 3 s.f


V1 = 6.31 V ===> 2 s.f
V2 = 8.736 V ===> 3 s.f
22
Solution (Example 1.5)

V 6.31V
R1  1  4263.s.f
===> 35 426
I 0.0148 A

V2 8.736V ===> 3 s.f


R2   590.27 590
I 0.0148 A

P1 V1 I  6.31V   0.0148 A


= 0.09339
= 0.0934 ===> 3 s.f

23
Significant Figures (cont)

3) When dropping non-significant figures

0.0148 ==> 0.015 (2 s.f)


==> 0.01 (1 s.f)

24
STATISTICAL ANALYSIS)

Statistical analysis define and calculate arithmetic mean,


deviation from mean, average deviation and standard
deviation.

Arithmetic Mean:- The arithmetic mean of ‘n’


measurements at specific count of the variables ‘x’ is given
by the expression:-
n

x1 x2  x3 .......... xn
 xn
n 1
x  n  n

25
STATISTICAL ANALYSIS)

Deviation from Mean:- This is the departure of a given


reading from the arithmetic mean of the group of readings.
The deviation from mean is:-

d  x  x , d Similarly,
 x  x, d x  x
n n
1 2 2
The deviation may be positive or negative. The algebraic
sum of all the deviations must be zero.

26
STATISTICAL ANALYSIS)

Average Deviation:- This is the indication of the


precision of the instrument used in measurement.
It is defined as the sum if the absolute values of the deviation
divided by the number of readings.
d1  d 2 ............. d n  dn
D  n  n
av
Highly precise instruments yield a low average deviation
between readings.
Where,
average
D  deviation
av
27
STATISTICAL ANALYSIS)

Standard Deviation:- This is a root mean square


deviation, and is the most important factor in the statistical
analysis of measurement of data. Reduction in this quantity
effectively means improvement in measurement.
d12 d 2
2 d 2 ........... d 2
3 n d n
2
  
n n

Where,
Standard
  deviation

28
STATISTICAL ANALYSIS)

Standard Deviation and Probable of Error


Variance: the mean-squared value of the deviations
2 d12  d 22  ...  d n2
 
n
Standard deviation or root mean squared (rms)

d12  d 22  ...  d 2n
SD or σ 
n
For the case of a large number of measurements in which
only random errors are present, it can be shown that the
probable error in any one measurement is 0.6745 times
the standard deviation:
Probable Error 0.6745
29
TYPES OF STATIC ERROR
It is the numerical difference between the true value of a
quantity and its value as obtained by measurement.
Repeated measurement of the same quantity gives
different indications.

• Types of static error

1) Gross error/human error


2) Systematic Error
3) Random Error

30
1. Gross Errors or Human Errors
– Resulting from carelessness, e.g. misreading, incorrectly
recording

Serious measurement errors can occur if an instrument is not read


correctly. The digital instrument is on a 300 mA range, so its
reading is in milliamperes. For the analog meter, the range
selection must be noted, and the pointer position must be read from
the correct scale 31
Absolute Errors and Relative Errors

Percentage accuracy gives the relative error in a measured, or


specified quantity. The absolute error can be determined by
converting the percentage error into an absolute quantity

32
• Accuracy, Precision, Resolution, and Significant
Figures
– Accuracy (A) and Precision
• The measurement accuracy of 1% defines how close the
measurement is to the actual measured quality.
• The precision is not the same as the accuracy of measurement,
but they are related

Measurement precision depends on the smallest change that can be observed


in the measured quantity. A 1mV change will be indicated on the digital
voltmeter display above. For the analog instrument, 50 mV is the smallest
33
change that can be noted
a) If the measured quantity increases or decreases by 1 mV, the reading
becomes 8.936 V or 8.934 V respectively. Therefore, the voltage is
measured with a precision of 1 mV.
b) The pointer position can be read to within one-fourth of the smallest
scale division. Since the smallest scale division represents 0.2 V, one-
fourth of the scale division is 50 mV.
 Resolution
The measurement precision of an instrument defines the smallest
change in measured quantity that can be observed. This smallest
observable change is the resolution of the instrument.
 Significant Figures
The number of significant figures indicate the precision of
measurement.
34
Example 2.1: An analog voltmeter is used to measure voltage of
50V across a resistor. The reading value is 49 V. Find
a) Absolute Error
b) Relative Error
c) Accuracy
d) Percent Accuracy

Solution
a) e  X t  X m 50V  49V 1V
Xt  Xm
b) % Error  100%
Xt
50V  49V
 100% 2%
50V
c) A 1  % Error 1  2% 0.98
d) % Acc 100%  2% 98% 35
Measurement Error Combinations

When a quantity is calculated from measurements made on two


(or more) instruments, it must be assumed that the errors due to
instrument inaccuracy combine is the worst possible way.
Sum of Quantities
Where a quantity is determined as the sum of two
measurements, the total error is the sum of the absolute
errors in each measurement.

E V1 ΔVV1   V2 ΔVV2 


giving E V1  V2   ΔVV1  ΔVV2 

36
Difference of Quantities
The error of the difference of two measurements are again additive

E  V1 ΔVV1    V2 ΔVV2 
 V1  V2   ΔVV1  ΔVV2 

Product of Quantities
When a calculated quantity is the product of two or more quantities,
the percentage error is the sum of the percentage errors in each
quantity
P EI
 E ΔVE  IΔVI 
EI EΔV I IΔV E ΔVEΔVI
since ΔVEΔVI is very small,
P EI  EΔV I IΔV E 

37
Example: An 820Ω resistance with an accuracy of10% carries a current
of 10 mA. The current was measured by an analog ammeter on a 25mA range
with an accuracy of 2% of full scale. Calculate the power dissipated in the
resistor, and determine the accuracy of the result.

Solution
2
P I 2 R 10 mA 820
82 mW
error in R 10%
error in I 2% of 25 mA
0.5 mA
0.5 mA
 100% 5%
10 mA

%error in I 2 2 5%  10%


 
%error in P  %error in I 2   %error in R 
10%  10%  20%
38
Example: The accuracy of five digital voltmeters are checked by using each
of them to measure a standard 1.0000V from a calibration instrument. The
voltmeter readings are as follows: V1 = 1.001 V, V2 = 1.002, V3 = 0.999, V4 =

0.998, and V5 = 1.000. Calculate the average measured voltage and the average
deviation.
Solution
V  V2  V3  V4  V5
Vav  1
5
1.001  1.002  0.999  0.998  1.000
 1.000 V
5
d1 V1  Vav 1.001  1.000 0.001V
d 2 V2  Vav 1.002  1.000 0.002 V
d 3 0.999  1.000  0.001V
d 4 0.998  1.000  0.002 V
d 5 1.000  1.000 0 V
d1  d 2  ...  d 5
D 
5
0.001  0.002  0.001  0.002  0
 0.0012V
5 39
Instrument ‘loading’ effect : Some measuring instruments depend for
their operation on power taken from the circuit in which measurements are
being made. Depending on the ‘loading’ effect of the instrument (i.e. the
current taken to enable it to operate), the prevailing circuit conditions may
change.
The resistance of voltmeters may be calculated since each have a stated
sensitivity (or ‘figure of merit’), often stated in ‘k per volt’ of f.s.d. A
voltmeter should have as high a resistance as possible ( ideally infinite).In
a.c. circuits the impedance of the instrument varies with frequency
and thus the loading effect of the instrument can change.

Example:
Calculate the power dissipated by the voltmeter and by resistor R in
Figure 10.9 when (a) R=250 Ω, (b) R=2 MΩ. Assume that the
voltmeter sensitivity (sometimes called figure of merit) is 10 kΩ/V.

40
41
2. Systematic Error: due to shortcomings of the instrument
(such as defective or worn parts, ageing or effects of the
environment on the instrument)
• In general, systematic errors can be subdivided into static and
dynamic errors.
– Static – caused by limitations of the measuring device or the
physical laws governing its behavior.
– Dynamic – caused by the instrument not responding very fast
enough to follow the changes in a measured variable.
- 3 types of systematic error :-
(i) Instrumental error
(ii)Environmental error
(iii) Observational error
42
Types of static error

(i) Instrumental error


- These are inherent in measuring instrument because of
their mechanical structure (eg: in a D’Arsonval meter, friction in the
bearings of various moving component, irregular spring tension,
stretching of spring, etc)
- Instrumental error can be avoid by:
(a) selecting a suitable instrument for the particular
measurement application
(b) apply correction factor by determining
instrumental error
(c) calibrate the instrument against standard

43
(ii) Environmental error
- due to external condition effecting the
measurement including surrounding area condition
such as change in temperature, humidity,
barometer pressure, etc
- to avoid the error :-
(a) use air conditioner
(b) sealing certain component in the instruments
(c) use magnetic shields

(iii) Observational error


- introduce by the observer
- most common : parallax error and estimation error
(while reading the scale)
- Eg: an observer who tend to hold his head too far to the
left while reading the position of the needle on the scale.
44
3) Random error:-
- due to unknown causes, occur when all systematic
error has accounted
- This is accumulation large number of small effect
and may be of real concern only in measurements
requiring a high degree of accuracy.
- These errors can be avoid by:-
(a) increasing number of reading
(b) use statistical means to obtain best
approximation of true value

45
Systematic Errors and Random errors in brief

Systematic Errors
Instrumental Errors
 Friction
 Zero positioning
Environment Errors
 Temperature
 Humidity
 Pressure
Observational Error
Random Errors 46
Dynamic Characteristics
• Dynamic – measuring a varying process condition.

• Instruments rarely respond instantaneously to changes in the


measured variables due to such things as mass, thermal
capacitance, fluid capacitance or electrical capacitance.

• Pure delay in time is often encountered where the instrument


waits for some reaction to take place.

• Such industrial instruments are nearly always used for


measuring quantities that fluctuate with time.

• Therefore, the dynamic and transient behavior of the


instrument is important.
47
Dynamic Characteristics
• The dynamic behavior of an instrument is determined
by subjecting its primary element (sensing element)
to some unknown and predetermined variations in the
measured quantity.

• The three most common variations in the measured


quantity are:-

– Step change
– Linear change
– Sinusoidal change

48
Dynamic Characteristics
• Step change-in which the primary element is subjected to an
instantaneous and finite change in measured variable.

• Linear change-in which the primary element is following the


measured variable, changing linearly with time.

• Sinusoidal change-in which the primary element follows a


measured variable, the magnitude of which changes in
accordance with a sinusoidal function of constant amplitude.

49
Dynamic Characteristics
• The dynamic performance characteristics of an
instrument are:
– Speed of response - The rapidity with which an instrument
responds changes in measured quantity.
– Dynamic error - The difference between the true and
measured value with no static error.
– Lag – delay in the response of an instrument to changes in
the measured variable.
– Fidelity – the degree to which an instrument indicates the
changes in the measured variable without dynamic error
(faithful reproduction).

50
Dynamic Response of
Zero Order Instruments
• The relation between input and output can be represented
by following differential equation.
d n xo d n  1 xo d xo
an n
 a n 1 n 1
 ......  a 1  a0 xo
dt dt dt
d m xi d m  1 xi d xi
bm  bm  1  ......  b1  b0 xi
dt m dt m  1 dt
• Where,
xo = Output quantity
xi = Input quantity
t = Time
a,b = constants

51
Dynamic Response of
Zero Order Instruments
• In zero order system all the a’s and b’s other than a0 and b0 are
assumed to be zero. Hence, the differential equation degenerates to:-
a0 xo b0 xi
• The static sensitivity of zero order system is defined as
b0
xo  xi  Kxi
a0
• This equation xo = Kxi is an algebraic equation. It shows that output
follows input perfectly with no distortion or time lag. Thus a zero
order instrument represents ideal or perfect dynamic performance.
• A practical example of zero order instrument is the displacement
measuring instrument.

52
Dynamic Response of
First Order Instruments
• In first order system all the a’s and b’s other than a2 , a1, a0 and b0
are assumed to be zero. Hence, the differential equation
degenerates to:-
a
dxo
 a x b x
1 0 o 0 i
dt

• Any instrument follows this equation is called a first order


instrument. By dividing by a0 , we get,
a1 dxo b
 xo  0 xi   .D  1.xo kxi
a0 dt a0

a1 b0
Timeconst . K  StaticSens
• Where,  
a0 a0
xo K

• The operational Transfer Function will be:- xi D  1
• A very common example of first order instrument is a mercury in
53
glass thermometer.
Dynamic Response of
Second Order Instruments
• In second order system all the a’s and b’s other than a 2 , a1 a0 and
b0 are assumed to be zero. Hence, the differential equation
degenerates to:- d 2 xo d xo
a2 2
 ......  a1  a0 xo b0 xi
dt dt
• Any instrument follows this equation is called a second order
instrument. The above equation reduced to :-
 D2 2D 

 2    1
.xo  Kxi
 n n 

• Where,
a0
n  = Undamped natural frequency
a2
a1 b0
2   Damping ratio Static
K  sensititvity

a0 a 2 b1

A practical example of this type is the Spring Balance. 54


Standard
• A standard is a known accurate measure of physical quantity.
• Standards are used to determine the values of other physical
quantities by the comparison method.
• For example:- the fundamental unit of length in the
international system (SI) is the metre, defined as the distance
between to fine lines engraved on gold plugs near the ends of a
platinum iridium alloy at 00C and mechanically supported in a
prescribed manner.
• All standards are preserved at the International Bureau of
Weight and Measures (BIMP), Paris.
• Four categories of standard:
– International Standard
– Primary Standard
– Secondary Standard
– Working Standard
55
Standard
• International Std
– Defined by International Agreement
– Represent the closest possible accuracy attainable by
the current science and technology

• Primary Std
– Maintained at the National Std Lab (different for every
country)
– Function: the calibration and verification of secondary
std
– Each lab has its own secondary std which are
periodically checked and certified by the National Std
Lab.
– Primary standards are not available for use outside the
national laboratory.
56
Standard
• Secondary Standard
– Secondary standards are basic reference standards used by measurement
and calibration laboratories in industries.
– Each industry has its own secondary standard.
– Each laboratory periodically sends its secondary standard to the National
standards laboratory for calibration and comparison against the primary
standard.
– After comparison and calibration, the National Standards Laboratory
returns the secondary standards to particular industrial laboratory with a
certification of measuring accuracy in terms of a primary standard.

• Working Std

– Used to check and calibrate lab instrument for accuracy and performance.
– For example, manufacturers of electronic components such as capacitors,
resistors and many more use a standard called a working standard for
checking the component values being manufactured.
57
ELECTRONIC INSTRUMENT

• Basic elements of an electronics instrument

Signal Indicatin
Transducer Modifie g Device
r
1) Transducer
- convert a non electrical signal into an electrical signal
- e.g: a pressure sensor detect pressure and convert it to electricity for display
at a remote gauge.
2) Signal modifier
- convert input signal into a suitable signal for the indicating
device
3) Indicating device
- indicates the value of quantity being measure
58
INSTRUMENT APPLICATION GUIDE

• Selection, care and use of the instrument :-


 Before using an instrument, students should be thoroughly
familiar with its operation ** read the manual carefully
 Select an instrument to provide the degree of accuracy
required (accuracy + resolution + cost)
 Before used any selected instrument, do the inspection for
any physical problem
 Before connecting the instrument to the circuit, make
sure the ‘function switch’ and the ‘range selector switch’
has been set-up at the proper function or range

59
INSTRUMENT APPLICATION GUIDE

Analog Multimeter
60
INSTRUMENT APPLICATION GUIDE

Digital Multimeter
61
Assignment-1
• Define the terms accuracy, error, precision,
resolution, expected value and sensitivity.
• State three major categories of error.
• A person using an ohmmeter reads the measured
value as 470 ohm when the actual value is 47 ohm.
What kind of error does this represent?
• State the classifications of standards.
• What are primary standards? Where are they
used?
• What is the difference between secondary
standards and working standards?
• State three basic elements of electronic instrument.62
THE END

63

You might also like