Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
Page
1 _______General Information ____________________________________________ 1 - 1
2 _______Backup _______________________________________________________ 2 - 1
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Flow Chart MEMOSKOP software update . . . . . . . . . . . . . . . . . . . . . . .3-1
Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Preliminary work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Saving the MEMOSKOP parameters to diskette . . . . . . . . . . . . . . . . . .3-4
Download MEMOSKOP software . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Download MEMOSKOP with empty Flash Proms . . . . . . . . . . . . . . . . . .3-8
6 _______Backup _______________________________________________________ 6 - 1
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
7 _______Shutdown Delay________________________________________________ 7 - 1
Page
8 ______ Drawer Catches and Terminating Resistor __________________________8 - 1
Customer Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Systems/Products Affected 1
Prerequisites 1
n.a.
Ordering Information 1
The software update kit Iso-C 3D Version P3 contains the following parts:
Based on the above description, the CSE should be able to check the update kit for com-
pleteness prior to beginning the update.
Returns 1
n.a.
n.a.
Workflow 1
Perform backup
Perform restore
Perform backup
Backup 2
General Remarks 2
Software 3
The flow chart SW update MEMOSKOP shows only the main working steps. In the case
of problems during the MEMOSKOP software update, please follow the path of the flow
chart.
SW Update
MEMOSKOP
Install
MEMOSKOP
Service SW to the
Service PC
Check Service PC
settings. See ,
prerequisites
Saving the
MEMOSKOP
paramers to
diskette
NO
worked well?
Yes
Read out all
parameter settings
(User and
technical setup)
and note them.
Download
MEMOSKOP
Software
NO
worked well?
Check Service PC
Yes settings. See
Prerequisites.
Download
according to
"Download
Memoskop with Please try several
empty Flash times
Proms
NO
worked well?
Yes
END
Procedures 3
Prerequisites 3
*FIFO settings
steps for Windows in German
• select on the windows footer
• select "start"
• select "Einstellungen"
• select "Systemsteuerung"
• select "System"
• select "Gerätemanager"
• select "Anschlüsse"
• select "COM-Anschluss"
• select "Anschlußeinstellungen"
• select "erweiterte .."
• "FIFO-Puffer verwenden" must marked with "V"
Preliminary work 3
• Remove the back cover from the monitor trolley (important for installing MEMOSKOP
software).
NOTE Before the use of the host service software or MEMOSKOP ser-
vice software, the write protection of the corresponding parame-
ter diskettes have to be switched off
• Insert the MEMOSKOP diskette "User Programs" in the service PC, drive A.
• Start the MEMOSKOP service software.
• Enter the password and press the Return key on the Service-PC.
• Select the required serial interface on the service PC.
Fig. 1
• After the main menu appears, click on the button "Get from Memoskop" and wait for the
transfer of the parameters from the MEMOSKOP to the PC. After the transfer, the status
line shows "....Receiving configuration done"
• Then save the parameters to the diskette (menu "File", submenu "Save").
• For the file name, enter the actual date. The file extension should be "mem".
Example:
The current date is: (DD_MM_YY) 28_01_02
Enter the file name+file extension: 28_01_02.mem
• Wait for the parameters to be saved on diskette.
• If the storage of the MEMOSKOP parameter set to diskette is not possible, all para-
meters (User setup and technical setup) must be read and noted them manually.
• For the MEMOSKOP displayed insert the corresponding diskette supplied in drive A of
the Service PC (MEMOSKOP software).
• Select the menu "Download".
• Select "Select file for Download:".
• The program window "Open File" opens.
• Select drive A.
• Select the file xVC0xxx.mot ) and confirm selection with "OK".
Fig. 2
• After successful completion of the Download, the "Download complete" message will be
displayed in the status line.
NOTE This section describes the downloading procedure for the MEMOSKOP
with empty Flash Proms. If downloading as described in the previ-
ous section could be carried out successfully, this section must
be skipped.
• Switch the unit off and exit the Memoskop service program.
• The cover of the basic unit and the rear cover of the monitor trolley are removed.
• Set the service switch D1.S3.2 in position ON.
• Switch on the unit. Ignore the error messages, displayed on the user console.
• Insert the corresponding diskette supplied in drive A of the Service PC (MEMOSKOP
software ).
• Start the MEMOSKOP service program and enter the password.
• Select the service PC serial interface.
• Ignore the message "No communication with the X-Ray system. Therefore click on the
button "Ignore".
• For the error message "The Program didn’t receive the version message from the
Memoskop", manually select the MEMOSKOP model.
3 MEMOSKOP C-E 100 /C / C-MOD
4 MEMOSKOP CSub/CSub-MOD
Fig. 3
"power plug"
"power plug"
Restore 4
Monitor 4
• Select "Utilities".
• Select "Settings".
• Select "True Color (32-bit)".
• Select "72 Hz" refresh frequency.
• Confirm by clicking on "OK", even if the other settings are correct.
• Select "Start".
• Select "Shutdown".
• Select "OK" for shutdown
• until the "restart" prompt appears.
• Shut down the SIREMOBIL Iso C 3D.
• Wait until the SIREMOBIL Iso C 3D has shut down completely.
• Connect the MEMOSKOP power plug.
Control software 4
General Remarks 5
Manual Settings 5
• Select the layout that corresponds to the keyboard that has been shipped in
"Keyboard layout".
• Click on "OK".
• Click on "OK".
Backup 6
General Remarks 6
! You will find the board (D210) for the shutdown delay in the monitor cart
(see arrow).
• Remove the rear panel from the monitor cart for this purpose.
• The the parallelism of the front of the drawer has to be adjusted, loosen the two screws.
• Align the front of the drawer parallel and tighten the two screws again.
Terminating resistor 8
This terminating resistor is required for all systems equipped with a navigation system,
e.g. from Medvision.
If the navigation system is not connected, then the terminating resistor must be
unplugged.
This information is enclosed as single page with the update kit as supplement to the oper-
ating instructions. Please insert this page in the operating instructions and draw the cus-
tomer’s attention to this information.
Customer Information 9
Completion Protocol 10
Remark: ....................................................................................................
....................................................................................................
NOTICE After completing the update, make a copy of this page, fill it out
and file it in Register 10 of the System Manual at the correspond-
ing site.