Professional Documents
Culture Documents
CT
Update Instructions CT023/02/P
Title: Optimizing the receiver connection
Remark:
Remark:
© Sie mens AG 2002
Chg. Rev. No.: 115227 The reproduction, transmission or
use of this document or its contents
Name: J.Kästner is not permitted without express
written authority. Offenders will be
Dept.: CS PS22 liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Page
1 _______General Information ____________________________________________ 1 - 1
2 _______Work Steps____________________________________________________ 2 - 1
Completion Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
SOMATOM Sensation 16 (serial nos.: 50008, 50012 -50014, 50016 - 50025, 50027 -
50040).
Somatom Sensation Cardio (serial nos.: 52001 - 52003).
To optimize receiver functionality, the shielding for cable W375-S6 has to be isolated.
Prerequisites 1
n.a.
Ordering information 1
Note: To limit the number of on-site visits, this update should be performed in com-
bination with a planned visit (e.g. maintenance or other UI’s).
Based on the above description, the CSE should be able to check the update kit for com-
pleteness prior to beginning the update.
Returns 1
n.a.
n.a.
NOTICE Read the most recent General Safety Notes and Safety Informa-
tion (CT02-023.860.01.xx.xx) before replacing parts inside the gan-
try.
Preliminary steps 2
Warning The system has to be switched off completely, using the on-site
ON/OFF switch before working inside the gantry. Make sure that
the system is secured against DFFLGHQWDOVZLWFKRQ.
Remove the gantry back cover by opening the screws on the back of the gantry.
Remove the funnel by removing the 4 screws, and disconnect X312A to 312D.
Customer information 2
n.a.
Fill out the attached "Update Completion Form" and insert it in Register 10 of the system
owner manual.
Updates completed prior to publication of this UI must also be reported.
Updates are reported as follows:
First version
Completion Protocol 3
Remark: ....................................................................................................
....................................................................................................
NOTICE After completing the update, make a copy of this page, fill it out
and file it in Register 10 of the System Manual at the correspond-
ing site.