You are on page 1of 12

Sensation

CT
Update Instructions CT023/02/P
Title: Optimizing the receiver connection

Reason for update: Safety X Performance

Urgency: Immediate X Within 6 month


Update materials required? X Yes No
Material free of charge? X Yes No
Return of parts? Yes X No
Estimated completion time: 1,0 hrs Number of CSE’s:1
&XVWRPHU,QIRUPDWLRQ" Yes X No

Systems/Products affected/System identifying IVK

Name Part No. Serial No.


Sensation 16 73 93 114 50008, 50012 -50014,
50016 - 50025, 50027 -
50040
Sensation Cardiac 75 43 106 52001 - 52003

Remark:

Components affected/to be modified/IVK

Name Part No. Serial No. Component status


affected not affected
n.a n.a n.a n.a n.a

Remark:
© Sie mens AG 2002
Chg. Rev. No.: 115227 The reproduction, transmission or
use of this document or its contents
Name: J.Kästner is not permitted without express
written authority. Offenders will be
Dept.: CS PS22 liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

Print No.: C2-023.896.05.01.02 Doc. Gen. Date: 08.02


Replaces: n.a
0-2 Revision

Chapter Page Revision


all all 01

Document revision level


The document corresponds to the version/revision level effective at the time of system delivery. Revi-
sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.

Sensation C2-023.896.05 Page 2 of 4 Siemens AG


Rev. 01 08.02 CS PS22 Medical Solutions
Contents 0-3

Page
1 _______General Information ____________________________________________ 1 - 1

Systems / Products affected. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 1


Reason for the Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 1
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Special tools / documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 2
Contents of the update kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Returns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 2
Spare parts handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 2

2 _______Work Steps____________________________________________________ 2 - 1

Modifying the slip ring connections . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1


Tools and auxiliary equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 1
Installing the shrink tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 1
Customer information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 2
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 2
Changes to Previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 3

3 _______Update Completion Form ________________________________________ 3 - 1

Completion Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1

Siemens AG C2-023.896.05 Page 3 of 4 Sensation


Medical Solutions Rev. 01 08.02 CS PS22
0-4 Contents

This page intentionally left blank.

Sensation C2-023.896.05 Page 4 of 4 Siemens AG


Rev. 01 08.02 CS PS22 Medical Solutions
General Information 1 1-1

Systems / Products affected 1

SOMATOM Sensation 16 (serial nos.: 50008, 50012 -50014, 50016 - 50025, 50027 -
50040).
Somatom Sensation Cardio (serial nos.: 52001 - 52003).

Reason for the Update 1

To optimize receiver functionality, the shielding for cable W375-S6 has to be isolated.

Prerequisites 1

n.a.

Special tools / documents 1

Standard service kit

Siemens AG C2-023.896.05 Page 1 of 2 Sensation


Medical Solutions Rev. 01 08.02 CS PS22
1-2 General Information

Ordering information 1

Place an order as follows for each system to be updated:


75 41 175 update kit

List the system serial number on the order.

Note: To limit the number of on-site visits, this update should be performed in com-
bination with a planned visit (e.g. maintenance or other UI’s).

Contents of the update kit 1

Update kit 75 41 175 contains the following parts:

Pos. Quan. Part No. Name


1 1 n.a Shrink tube
2 1 C2-023.896.05.01.02 UI CT023/02/P
3 4 n.a Cable ties
Tab. 1

Based on the above description, the CSE should be able to check the update kit for com-
pleteness prior to beginning the update.

Returns 1

n.a.

Spare parts handling 1

n.a.

Sensation C2-023.896.05 Page 2 of 2 Siemens AG


Rev. 01 08.02 CS PS22 Medical Solutions
Work Steps 2 2-1

Modifying the slip ring connections 2

NOTICE Read the most recent General Safety Notes and Safety Informa-
tion (CT02-023.860.01.xx.xx) before replacing parts inside the gan-
try.

Tools and auxiliary equipment 2

1. Standard tool kit

Preliminary steps 2

œ Switch off the system.


œ 6ZLWFKRIISRZHUXVLQJWKHRQVLWH2Q2IIVZLWFK

Warning The system has to be switched off completely, using the on-site
ON/OFF switch before working inside the gantry. Make sure that
the system is secured against DFFLGHQWDOVZLWFKRQ.

œ Remove the gantry back cover by opening the screws on the back of the gantry.
œ Remove the funnel by removing the 4 screws, and disconnect X312A to 312D.

Installing the shrink tube 2

CAUTION Follow C-Mos guidelines when touching the receiver.


Failure to follow these guidelines when touching the housing will
destroy the extremely sensitive receiver.

3 œ Connect ESD protection to slipring


ground (e.g 1 Fig. 1).
œ Remove plug of cable W375-S6 (3 Fig.
1).
œ Remove cable ties as necessary(3 Fig. 1)
Note: Do not damage the optical fiber
cable while performing the next steps
( 4 Fig. 1)!
4 1
2

)LJ Receiver cable with not isolated shield

Siemens AG C2-023.896.05 Page 1 of 4 Sensation


Medical Solutions Rev. 01 08.02 CS PS22
2-2 Work Steps

œ Slip the shrink tube onto the cable


œ Use a soldering iron to shrink the tube.

)LJ Shrinking the tube

œ Reattach the receiver cable in


accordance with the original
configuration.
Note: Cable shield should now be
completely isolated against the
holder.
œ Secure the cable with cable ties.

)LJ Receiver connection with isolated cable shield.

œ Install the back funnel and reconnect X312A to 312D.


œ Install the gantry back cover.
œ Install the plexi ring.
œ Switch on the system and perform some test scans.

Customer information 2

n.a.

Final work steps 2

œ Fill out the attached "Update Completion Form" and insert it in Register 10 of the system
owner manual.
œ Updates completed prior to publication of this UI must also be reported.
œ Updates are reported as follows:

Sensation C2-023.896.05 Page 2 of 4 Siemens AG


Rev. 01 08.02 CS PS22 Medical Solutions
Work Steps 2-3
- The modification reply cards (Type606) previously distributed with the publication of
updates are no longer applicable.
- The modification reply report has to be prepared by authorized personnel using an
application on the Intranet.

Changes to Previous Version 2

First version

Siemens AG C2-023.896.05 Page 3 of 4 Sensation


Medical Solutions Rev. 01 08.02 CS PS22
2-4 Work Steps
This page intentionally left blank.

Sensation C2-023.896.05 Page 4 of 4 Siemens AG


Rev. 01 08.02 CS PS22 Medical Solutions
Update Completion Form 3 3-1

Completion Protocol 3

The update with the number CT023/02/P has been completed.

Part Number: ................................ Model No: ................................

Serial Number: ................................ SKZ: ................................

Customer: ................................ Site: ................................

Customer No.: ................................

Name ( CSE ): ................................ Telephone: ................................

Country: ................................ Location: ................................

Date: ................................ Signature: ................................

Remark: ....................................................................................................

....................................................................................................

NOTICE After completing the update, make a copy of this page, fill it out
and file it in Register 10 of the System Manual at the correspond-
ing site.

Siemens AG C2-023.896.05 Page 1 of 2 Sensation


Medical Solutions Rev. 01 08.02 CS PS22
3-2 Update Completion Form
This page intentionally left blank.

Sensation C2-023.896.05 Page 2 of 2 Siemens AG


Rev. 01 08.02 CS PS22 Medical Solutions

You might also like