You are on page 1of 8

Sensation / Emotion

CT
Service Instruction CT087/05/S
Service Pack 1 for VB28B

Reason for update: Safety

Urgency for update: X Immediate Within months


Intranet update available? Yes X No
Remote update feasible? Yes without CSE Yes, with CSE X No
(Category 2) (Category 1)
Materials for logistic update
X Yes No
required?
Materials free of charge? X Yes No
Return of parts? Yes X No
Estimated completion time: 0,5 hour(s) Number of CSE’s: 1
Customer application training? Yes X No

Systems/Products affected/System identifying IVK

Name Material No. Serial No.


see page 3 see page 3 see page 3

Remark: n.a.

Components affected/to be modified/IVK

Name Material No. Serial No. Component status


affected
n.a. n.a. n.a. n.a.

Remark: n.a.

© Siemens AG 2005
Chg. Ref. No.: 166473 The reproduction, transmission or use
of this document or its contents is not
Name: B. Lapunik permitted without express written
authority. Offenders will be liable for
Dept.: CS PS CT damages. All rights, including rights
created by patent grant or registration
of a utility model _or_ design,_are_
reserved.

Print No.: C2-023.897.11.01.02 Doc. Gen. Date: 12.05


Replaces: n.a. Attachment to TDAR 4.14/01 Template Version 4.0; September 2005 Page 1 of 8
2 Service Instruction CT087/05/S

Document Revision Level 0

This document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer 0

The installation and service of equipment described herein is to be performed by qualified


personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

Reason for the Update 0

An error may occur where the LungCare program does not store the marks for measured
nodules. This Service Pack 1 provides a software patch to correct this issue.

Prerequisites 0

Wizard running with VB28B and option LungCare

Special Tools / Documents 0

n.a.

Sensation / Emotion C2-023.897.11. Page 2 of 8 Siemens AG


Rev. 01 12.05 CS PS CT Medical Solutions
Service Instruction CT087/05/S 3

Systems/Products Affected 0

Model Material no. Serial no.

SOMATOM Sensation 16 50180, 50550, 50862, 50882, 50886,


51562, 51563, 51565, 51570, 51571,
7393114 51573, 51574, 51576, 51577, 51579,
51583, 51588, 51600, 51655, 51736,
51864, 51865, 51968
SOMATOM Sensation 10 7543015 53003
SOMATOM Emotion 16 7734713 39011, 39012, 39013, 39015, 39026,
39035, 39049, 39050, 39077, 39078,
39082, 39087

Affected system numbers may be subject to change, the UI- data published in the Intranet
(Udate Handling Tool) always reflect the current status. UI- Paper versions are not
updated when affected serial numbers change.

Ordering Information 0

The following update kit has to be ordered from CSML (SAP Distribution Channel, factory
2050) :
Update-Kit_VB28B_IES_SP1 10179707 Update Kit Material Number)

Contents of the Update Kit 0

Update kit 10179707 contains the following parts:

Pos. Quan. Material No. Name


1 1 10179706 CD_SOMARIS/5_IES_VB28B_SP1
03078631 CD-R 700MB WHITE................=!
03057585 CD-BOX JEWEL BOX MIT TRAY (GRAU)
C2-023.897.11.01.02 CT087/05/S Service Pack 1 for VB28B
Tab. 1

Return of Parts 0

n.a.

Siemens AG C2-023.897.11. Page 3 of 8 Sensation / Emotion


Medical Solutions Rev. 01 12.05 CS PS CT
4 Service Instruction CT087/05/S

Work Steps 0

Installation at the IES (Wizard) only 0

• Switch ON the system


• Insert the Service Pack CD_SOMARIS/5_IES_VB28B_SP1 into the CD drive.
• Select Options -> Service -> Local Service
• In the service software menu, select Utilities.
• Under Source, select Escape to OS and at Command select NT Command Interpreter.
• Under Parameters, enter R:\install.bat
• Activate with Return or GO.

➪ A message window "The ServPack 1 is successfully installed" is displayed.


➪ The System Shut Down and Restart will be done automatically
• Remove the Service Pack CD

Check Installation 0

Once the application software is running,


Select Options -> Version and check if the new Service Pack 1 is listed.

Final Check 0

Perform a scan in customer mode and expose the film to ensure the camera is functioning
properly.

Sensation / Emotion C2-023.897.11. Page 4 of 8 Siemens AG


Rev. 01 12.05 CS PS CT Medical Solutions
Service Instruction CT087/05/S 5
Backup at the Wizard 0

• Start the service software and select the Backup & Restore function in the main menu.
• Insert the backup CD-R for the Wizard
• From Command, select Backup.
• From Drives, select the corresponding drive.
• From Packages, select CT-Settings-IES and click Go.
➪ BackupRestore Running ...

➪ Wait until Ready is displayed.


➪ Close the CD tray.
• From Packages, select SW-Settings02 and click Go.
➪ BackupRestore Running ...

➪ Wait until Ready is displayed.


➪ Close the CD tray.
• From Packages, select Security-Settings-IES and click Go.
➪ BackupRestore Running ..

➪ Wait until Ready is displayed.


➪ Close the CD tray.
• From Packages, select InSpace (optional) and click Go.
➪ BackupRestore Running ..

➪ Wait until Ready is displayed.


➪ Close the CD tray.
• When "Ready" is displayed, remove the backup media.
Click Home and terminate the service software by closing the browser window.

Siemens AG C2-023.897.11. Page 5 of 8 Sensation / Emotion


Medical Solutions Rev. 01 12.05 CS PS CT
6 Service Instruction CT087/05/S

Customer Information 0

n.a.

Final Work Steps 0

• Update the system documentation.


- Fill out, and if needed, make a copy of the attached "Completion Protocol/ Update
Completion Form" and file it in the corresponding System Binder/User Handbook.

• Updates that have already been completed prior to publication of this UI must also be
reported.The modification reply report has to be prepared by authorized personnel using
an application on the intranet.)

Changes to Previous Version 0

No changes; initial publication

Sensation / Emotion C2-023.897.11. Page 6 of 8 Siemens AG


Rev. 01 12.05 CS PS CT Medical Solutions
Service Instruction CT087/05/S 7

Completion Protocol 0

(Not required for "Remote updates without a service technician")

The update with the number CT087/05/S has been completed.

Material number: ................................

Serial number: ................................

Customer: ................................ Functional Location:................................

Customer No.: ................................

Name ( CSE ): ................................ Telephone: ................................

Country: ................................ Location: ................................

Date: ................................ Signature: ................................

Remark: ....................................................................................................

....................................................................................................

NOTE After completing the update, make a copy of this page, fill it out and file
it in the corresponding System Binder/User Handbook.

Siemens AG C2-023.897.11. Page 7 of 8 Sensation / Emotion


Medical Solutions Rev. 01 12.05 CS PS CT
8 Service Instruction CT087/05/S
This page intentionally left blank.

Sensation / Emotion C2-023.897.11. Page 8 of 8 Siemens AG


Rev. 01 12.05 CS PS CT Medical Solutions

You might also like