You are on page 1of 14

Sensation

CT
Update Instructions CT003/06/P
Facelift Sensation

Reason for update: Performance

Urgency for update:: Immediate X Within 6 months


Intranet update available? Yes X No
Remote update feasible? Yes, without CSE Yes, with CSE X No
(Category 2) (Category 1)
Materials for logistic update
X Yes No
required?
Materials free of charge? X Yes No
Return of parts? Yes X No
Estimated completion time: 7 hour(s) Number of CSE’s: 2
Customer application training? Yes X No

Systems/Products affected/System identifying IVK

Name Material No. Serial No.


Sensation see page 3 see page 3

Remark:
Order the proper Kit for the corresponding System. See Ordering Information

Components affected/to be modified

Name Material No. Serial No. Component status


affected
n.a. n.a. n.a. n.a.

© Siemens AG 2006
Chg. Ref. No.: 166142 The reproduction, transmission or use
of this document or its contents is not
Name: H.Smyk permitted without express written
authority. Offenders will be liable for
Dept.: CSPSCT damages. All rights, including rights
created by patent grant or registration
of a utility model _or_ design,_are_
reserved.

Print No.: C2-023.897.19.01.02 Doc. Gen. Date: 01.06


Replaces: n.a. Attachment to TDAR 4.14/01 Template Version 4.0; September 2005 Page 1 of 14
2 Update Instructions CT003/06/P

Document Revision Level


This document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

Sensation C2-023.897.19 Page 2 of 14 Siemens AG


Rev. 01 01.06 CSPSCT Medical Solutions
Update Instructions CT003/06/P 3

Systems/Products Affected 0

Attention: 0

➪ Order the proper Kit for the corresponding System. See Ordering Information

System Material number Serial number


Sensation 40 88 74 427 57001
Sensation 40 88 74 427 57004
Sensation 64 83 77 520 54012
Sensation 64 83 77 520 54013
Sensation 64 83 77 520 54016
Sensation 64 83 77 520 54021
Sensation 64 83 77 520 54025
Sensation 64 83 77 520 54034
Sensation 64 83 77 520 54040
Sensation 64 83 77 520 54053
Sensation 64 83 77 520 54058
Sensation 64 83 77 520 54071
Sensation 64 83 77 520 54175
Sensation 64 83 77 520 54321
Sensation 64 83 77 520 54323
Sensation 64 83 77 520 54411
Sensation Open 88 72 017 49225

Reason for the Update 0

This update provides instructions for exchanging the gantry covers. The new covers
reflect the new corporate design.

Prerequisites 0

n.a.

Special Tools / Documents 0

Standard tools

Siemens AG C2-023.897.19 Page 3 of 14 Sensation


Medical Solutions Rev. 01 01.06 CSPSCT
4 Update Instructions CT003/06/P

Ordering Information 0

The following update kit has to be ordered from CSML (SAP Distribution Channel, factory
2050) :
Update Kit, Cover Sensation Open Update Kit: 88 73 692
Update Kit, Cover Sensation 40/64 Update Kit: 88 74 815

List the system Serial Number on the order !

Contents of the Update Kit 0

n.a.

Return of Parts 0

Parts must be scrapped locally.

Sensation C2-023.897.19 Page 4 of 14 Siemens AG


Rev. 01 01.06 CSPSCT Medical Solutions
Update Instructions CT003/06/P 5

Work Steps 0

Sensation 40/64 0

8874559 Adapter 0

See number: 5
Remove adapter. Install new adapter. Each of the 2x3 nuts M6 has to be tighten with a
torque of 10 Nm

8874062 Front door compl. green 0

See number: 6
See CB-DOC Service - Gantry - Replacements of parts - Gantry front side - Gantry front
door.
Please, take the display panel (see number: 12) from the old front door and install it in the
new front door .See CB-DOC Service - Gantry - Replacements of parts - Gantry front side
- Gantry front door - Display panel.

8874088 Rear funnel compl. green 0

See CB-DOC Service / Gantry / Replacements of parts / Gantry rear side / Rear funnel
➪ NOTE: SELF-LOCKING NUTS AND WASHERS MUST BE REPLACED

8874112 Rear control panel 0

See number: 71/701


See CB-DOC Service / Gantry / Replacements of parts / Gantry rear side / Back control
panels
➪ NOTE: MIND THE JUMPER SETTING! The left control panel is jumpered differ-
ent from the right control panel. [compare with old control panel]

8874120 Shield Somatom Sensation 0

See number: 611


Attach label " Somatom Sensation " at the upper right side of the gantry front cover.

8874146 Corporate label Siemens 0

See number: 610


Attach label " SIEMENS " at the upper left side of the gantry front cover.

Siemens AG C2-023.897.19 Page 5 of 14 Sensation


Medical Solutions Rev. 01 01.06 CSPSCT
6 Update Instructions CT003/06/P

Final steps: 0

Check the function of the light markers.

Switch on light markers 0

• The position of static light markers must be checked / respectively must be adjusted if
necessary as described in
CB-Doc / Gantry / Replacements of parts / Gantry front side / Static light markers Sensa-
tion 10/16/40/64 / Cardiac/Cardiac64

The Light marker distance has to be adjusted. 0

• Adjust the table to a height 145


• Gantry tilt angle 0
• Select Service -> Tuneup -> Expert Mode -> LightM
• Mark the postion of the outer light marker
• Move table top till mark is at the inner light marker
• Insert the distance of Light marker
• Click Go

Test of function: 0

• Position mark at outer light marker


• Press button ->||<- at control panel
➪ The mark has to stop at inner light marker
• Check the intercommunication system.

Sensation C2-023.897.19 Page 6 of 14 Siemens AG


Rev. 01 01.06 CSPSCT Medical Solutions
Update Instructions CT003/06/P 7
Sensation 0

Siemens AG C2-023.897.19 Page 7 of 14 Sensation


Medical Solutions Rev. 01 01.06 CSPSCT
8 Update Instructions CT003/06/P

Sensation OPEN 0

8874559 Adapter 0

See number: 5
Remove Adapter. Install in new Adapter. Each of the 4 nuts M6 has to be tightened with a
torque of 10 Nm

8873668 Front Front door compl. green 0

See number: 6
See CB-DOC Service / Gantry / Replacements of parts / Gantry front side / Gantry front
door
Take the display panel (see number: 12) from the old front door and install it in the new
front door .See CB DOC Service / Gantry / Replacements of parts / Gantry front side -
Gantry front door / Display panel.

8873676 Rear funnel compl. green 0

See CB DOC Service / Gantry / Replacements of parts / Gantry rear side / Rear funnel

8874112 Back control panel 0

See number: 71/701


See CB-DOC Service - Gantry - Replacements of parts -- Gantry rear side - Back control
panels
NOTE: MIND THE JUMPER SETTING ! The left control panel is jumpered different from
the right control panel. [compare with old control panel]

8874120 Shield Somatom Sensation Open 0

See number: 611


Attach label " Somatom Sensation Open" at the upper right side of the gantry front cover.

8874146 Corporate labelSiemens 0

See number: 610


Attach label " SIEMENS " at the upper left side of the gantry front cover.

Final steps: 0

Check the function of the light markers 0

Swicht on the light markers


• The position of static light markers must be checked / and adjusted if necessary as
described in: CB Doc / Gantry / Replacements of parts / Gantry front side / Static light
markers Sensation OPEN

The Light marker distance has to be adjusted. 0

• Adjust the table to a height 145


• Gantry tilt angle 0

Sensation C2-023.897.19 Page 8 of 14 Siemens AG


Rev. 01 01.06 CSPSCT Medical Solutions
Update Instructions CT003/06/P 9
• Select Service -> Tuneup -> Expert Mode -> LightM
• Mark the postion of the outer light marker
• Move table top till mark is at the inner light marker
• Put in the distance of light marker
• Click Go

Test of function: 0

• Position mark at outer light marker


• Press button ->||<- at control panel
The mark has to stop at inner light marker

Siemens AG C2-023.897.19 Page 9 of 14 Sensation


Medical Solutions Rev. 01 01.06 CSPSCT
10 Update Instructions CT003/06/P

Sensation C2-023.897.19 Page 10 of 14 Siemens AG


Rev. 01 01.06 CSPSCT Medical Solutions
Update Instructions CT003/06/P 11

PHS (Sensation 40/ 64/ OPEN) 0

Note: Install the short bumper first. Affix the long bumper, starting on the short bumper
side for approx. 25 cm, then fix the rest of the bumper (pull or press it).

7034296 Rubber bumper, short 0

See number:14

Remove elastic bumper on the right an left side. Affix the new green elastic bumper, it is
magnetically affixed. 0

7034304 Rubber bumper, long 0

See number:13
Remove elastic bumper on the right and left side. Clean surface e.g. with alcohol. Affix the
new green elastic bumper.(it is equipped with double sided adhesive tape).

Siemens AG C2-023.897.19 Page 11 of 14 Sensation


Medical Solutions Rev. 01 01.06 CSPSCT
12 Update Instructions CT003/06/P

Final Check 0

• Check the function of the horizontal and sagittal lightmarkers. If needed, please adjust
them ( see Replacement of parts)
• Perform a function check of the system according to the modifications that have been
made.

Customer Information 0

n.a.

Sensation C2-023.897.19 Page 12 of 14 Siemens AG


Rev. 01 01.06 CSPSCT Medical Solutions
Update Instructions CT003/06/P 13

Final Work Steps 0

• Update the system documentation.


Update the revision level, the operating instructions and the technical documentation.
Fill out, and if needed, make a copy of the attached "Completion Protocol/ Update
Completion Form" and file it in the corresponding System Binder/User Handbook.
• The modification reply report has to be prepared by authorized personnel using
an application on the intranet.)

Changes to Previous Version 0

First version

Siemens AG C2-023.897.19 Page 13 of 14 Sensation


Medical Solutions Rev. 01 01.06 CSPSCT
14 Update Instructions CT003/06/P

Completion Protocol 0

(Not required for "Remote updates without a service technician")

The update with the number CT003/06/P has been completed.

Material number: ................................

Serial number: ................................

Customer: ................................ Functional Location:................................

Customer No.: ................................

Name ( CSE ): ................................ Telephone: ................................

Country: ................................ Location: ................................

Date: ................................ Signature: ................................

Remark: ....................................................................................................

....................................................................................................

NOTE After completing the update, make a copy of this page, fill it out and file
it in the corresponding System Binder/User Handbook.

Sensation C2-023.897.19 Page 14 of 14 Siemens AG


Rev. 01 01.06 CSPSCT Medical Solutions