System
SP
Update Instructions SP008/07/S
Securing the fork and collimator
Urdaneta
SP008/07
Safety Update
Urgent? H Yes Q No
Time period within 3 months
Intranet download available? Q Yes H No
Remote update handling feasible? Q Yes without CSE Q Yes with CSE H No
(Category 2) (Category 1/3)
Materials for logistic update H Yes Q No
required?
Materials free of charge? H Yes Q No
Q No, but credit at return
Return of parts? Q Yes H No
Estimated completion time 1.5 hour(s) Number of CSE’s: 1
Customer application training? Q Yes H No
Criteria Value
System material number 1818363, 1818454, 1818447
System software version n.a.
Component material number n.a.
Affected device serial no. see page 4
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
1 _______ Preparation_____________________________________________________ 4
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Systems/products affected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Reason for the update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Special tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Material information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Contents of the update kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Return of parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 _______ Update_________________________________________________________ 7
Work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Replace the screws from the fork mount. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Replace the screws from the flange mount at the single tank . . . . . . . . . . . . . . . . . . 8
Final check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Customer information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Changes to previous version. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Systems/products affected 0
The mounting bolts that secure the single tank fork assembly and the mounting bolts that
secure the collimator flange could potentially become loose.These specific mounting bolts
will be replaced with this update.
Prerequisites 0
n.a.
Special tools 0
Fig. 1: Kit
Update kit 10188210 (Fig. 1 / p. 5) contains the following parts:
Ordering information 0
The following update kit has to be ordered from CSML (SAP Distribution Channel, plant
2050):
Return of parts 0
no.
In this section, the replacement of the 6 screws (1/Fig. 2 / p. 7) from the fork mount is
shown. It is very important to follow the sequence described.
Work steps
• Switch off the system and disconnect it from the power supply. (In the case of Mobilett
XP Hybrid, operating mode switch = 0)
• Use the screws shown in (1/Fig. 1 / p. 5).
NOTE Don´t mix up the new screws with the old screws during replace-
ment.
• Take out only one screw at a time (1/Fig. 3 / p. 8), replacing each one in turn with a new
screw (1/Fig. 1 / p. 5); tighten each screw to a torque of 7 Nm.
NOTE The sequence for replacing all screws will be crosswise (screws
1-2-3-4-5-6) (Fig. 3 / p. 8); first tighten all screws to a torque of 7
Nm.
Replace the screws from the flange mount at the single tank 0
• Switch off the system and disconnect it from the power supply. (In the case of Mobilett
XP Hybrid, operating mode switch = 0)
• Use a rope or tightening belt to keep the system support arm from lifting up.
(A/Fig. 4 / p. 9).
• Loosen the four Allen screws of the collimator and remove the collimator.
(1/Fig. 7 / p. 10).
• Determine the type of screws that will be required depending on the measurement
result (type A (Fig. 10 / p. 12)or type B (Fig. 11 / p. 12)).
¹ Use the appropriate screws (delivered with this kit) for the distance measured.
If the distance X > 36.5 mm --> Use the type A screws (2/Fig. 1 / p. 5)
If the distance X < 35,0 mm --> use the type B screws (3/Fig. 1 / p. 5)
• Take out and replace only one screw and washer at a time; tighten each one to a torque
of 3.5 Nm.
NOTE Don´t mix up the new screws with the old screws during replace-
ment.
NOTE Only when the deviation between light field and radiation field is
>1.7 % will the collimator be adjusted using the four mounting
screws.
¹ Place the single tank in the exposure position while taking into consideration the
missing weight of the collimator cover.
¹ Correct any deviations > 1.7% via the four mounting screws of the collimator.
• Attach the collimator cover.
• Check the balance of the support arm.
• Dispose of the screws replaced and any that were not used.
NOTE Darken the room if necessary, to allow you to see the edges of the
light field more clearly.
• In the case of the XP Digital, place the lead measuring tape (centering cross) in the cen-
ter of the detector.
• Measure the light field and make note of the dimensions for the XP Digital.
• For XP/ Hybrid/ Eco, mark the light field with 4 washers (Fig. 13 / p. 14).
• Place one washer as a side marker (Fig. 13 / p. 14).
Final check 0
1. Function check of the system according to the modifications that have been made.
Customer information 0
n.a.
1. Fill out the attached "Completion Protocol" and file it in the corresponding binder.
2. The update shall be reported by authorized personnel, either via an intranet application
or by automated data transfer (only for countries connected via update handling inter-
face).
No.
Remarks: .....................................................................................................
.....................................................................................................
NOTE After completing the update, make a copy of this page, fill it out
and file it in the corresponding binder.