You are on page 1of 10

Product Information PI

ROTANODE™
E7100X

Rotating Anode X-ray Tube Assembly

 High speed rotating anode X-ray tube assembly for


high energy radiographic operations.
 For the purpose of general diagnostic X-ray
procedures.
 This tube has foci 1.2 and 0.6, and is available for
a maximum tube voltage 150 kV.
 This tube housing assembly has specially
processed rhenium-tungsten faced molybdenum
target of 100 mm diameter anode disc and is
accommodated with IEC60526 high-voltage cable
receptacles.

General Data

IEC Classification (IEC60601-1:2005+A1:2012) ...................... Class Ⅰ ME EQUIPMENT

Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding ......................................................................................................... Center-Grounded
Nominal X-ray Tube Voltage:
Radiographic ...................................................................................................................... 150 kV
Fluoroscopic ....................................................................................................................... 125 kV
Nominal Focal Spot Value:
Large Focus ............................................................................................................................. 1.2
Small Focus ............................................................................................................................. 0.6
Nominal Anode Input Power (at 0.1s) ..................................................................... See rating charts
180 Hz 60 Hz 50 Hz
Large Focus .................................................................................. 100 kW 59 kW 55.5 kW
Small Focus .................................................................................... 40 kW 24 kW 21.5 kW
Nominal Radiographic Anode Input Power:
180 Hz 60 Hz 50 Hz
Large Focus .................................................................................... 80 kW 48 kW 44 kW
Small Focus .................................................................................... 30 kW 18 kW 17 kW

★The information contained herein is presented only as a guide for the application of our products. No responsibility is assumed by Canon Electron Tubes
& Devices Co., Ltd. (CETD) for any infringements of patents or other rights of the third parties which may result from its use.
No license is granted by implication or otherwise under any patent or patent rights of CETD or others.
★The information contained herein may be changed without prior notice. It is therefore, advisable to contact to CETD before processing with the design of
equipment incorporating this product.
E7100X

Motor Ratings:
Stator: XS-AG
Starting Running
Driven Frequency [Hz] 180 60 180 60
Input Power [W] 2800 800 200 50
Voltage 3) 5) [V] 352 155 80 38
Current 4) [A] 9.6 10.2 3.0 2.2
Min. Speed Up 1) 7) [s] 1.2 0.8 - -
Capacitor [µF] 6 44 6 44
Min. Braking 2) 7) [s] 3.0 (DC 80V)
Note 1) The speed up time from normal speed to high speed is 2/3 times of the specified speed
up time from 0 to high speed, which is described on motor rating table.
2) To be applied for high speed rotation.
3) Applied voltage between common and main terminal.
4) Common current.
5) The every applied voltage must be never exceeded 110% of the above specification.
6) No more than two high speed starts per minute are permissible.
7) The speed-up time is allowed up to 110% of the above specification.

Anode Speed:
180 Hz .......................................................................................................... Minimum 9700 min-1
60 Hz ........................................................................................................... Minimum 3200 min-1
50 Hz ........................................................................................................... Minimum 2700 min-1
Stator Resistance:
Common-Main Winding ........................................................................................................ 9.4 Ω
Common-Auxiliary Winding ................................................................................................ 28.3 Ω
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 °C
Mode of Operation ........................................................................................................... Intermittent

Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 496 mm
Maximum Diameter .......................................................................................................... 195 mm
Target:
Anode Angle ............................................................................................................... 12 degrees
Diameter .......................................................................................................................... 100 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Permanent Filtration ................................................................. 1.1 mm Al / 75 kV IEC60522:1999
Radiation Protection (In accordance with IEC60601-1-3:2008):
Leakage Technique Factor .................................................................................. 150 kV, 3.4 mA
X-ray Coverage ................................................................................ 430 × 430 mm at SID 1000 mm
Weight (Approx.) ........................................................................................................................ 24 kg
High Voltage Receptacle ..................... To meet the requirements of IEC60526 Corrigendum1:2010
Cooling Method .................................................................................................. Natural or forced air
Tube Housing Model Number ............................................................................................. XH-112V

-2-
E7100X

Absolute Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage:


Radiographic ...................................................................................................................... 150 kV
Fluoroscopic ....................................................................................................................... 125 kV
Between Anode (or Cathode) and Ground ................................................................................ 75 kV
Minimum X-ray Tube Voltage ................................................................................................... 40 kV
Maximum X-ray Tube Current:
Large Focus .................................................................................................................... 1000 mA
Small Focus ...................................................................................................................... 500 mA
Maximum Filament Current:
Large Focus .......................................................................................................................... 5.8 A
Small Focus .......................................................................................................................... 5.2 A
Filament Voltage:
Large Focus (At maximum filament current 5.8 A) ................................................... 12.8 ~ 17.2 V
Small Focus (At maximum filament current 5.2 A) ................................................... 10.2 ~ 13.8 V
Filament Frequency Limits ................................................................................................ 0 ~ 25 kHz
Continuous Anode Input Power ............................................................................ 240 W (335 HU/s)
(Fluoroscopic, Radiographic or mixed exposure)
Thermal Characteristics:
Anode Heat Content ......................................................................................... 210 kJ (300 kHU)
Maximum Anode Heat Dissipation ................................................................. 710 W (1000 HU/s)
X-ray Tube Assembly Heat Content .............................................................. 1070 kJ (1508 kHU)
Nominal Continuous Input Power:
Without Air-circulator ................................................................................ 215 W (18 kHU/min)

Environmental Limits

Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 °C
Humidity ........................................................................................................................ 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature ............................................................................................................... -20 ~ 70 °C
Humidity ........................................................................................................................ 20 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa

-3-
E7100X

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant potential high-voltage generator
Stator Power Frequency 180Hz
Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
110kV 100kV 90kV 90kV 80kV 70kV 60kV 50kV
1000 500
80kV
100kV
900 125kV 70kV
110kV 40kV
800 60kV
400
150kV
TUBE CURRENT [mA]

125kV

TUBE CURRENT [mA]


700 50kV

600 150kV
300
40kV
500

400 200

300

200 100

100

0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]

Maximum Rating Charts


(Spot-Film Rating Charts)

Conditions: Tube Voltage


Constant potential high-voltage generator
Stator Power Frequency 180Hz
Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
80kV 70kV 80kV 70kV 60kV 50kV 40kV
900 400
90kV
60kV 90kV
800 100kV 350
100kV
700 110kV 50kV
300 110kV
TUBE CURRENT [mA]
TUBE CURRENT [mA]

125kV 40kV
600 125kV
250
150kV
500 150kV
200
400
150
300
100
200

100 50

0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]

-4-
E7100X

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant potential high-voltage generator
Stator Power Frequency 60Hz
Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
80kV 70kV
800 400
60kV 70kV 60kV 50kV 40kV
90kV 50kV
700 350 80kV
100kV
40kV 90kV
600 110kV 300
TUBE CURRENT [mA]

TUBE CURRENT [mA]


100kV
125kV
500 250 110kV
150kV 125kV
400 200
150kV

300 150

200 100

100 50

0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]

Maximum Rating Charts


(Spot-Film Rating Charts)

Conditions: Tube Voltage


Constant potential high-voltage generator
Stator Power Frequency 60Hz
Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
70kV 60kV 50kV 60kV 50kV 40kV
600 300
80kV 40kV 70kV
90kV
500 250 80kV
100kV
TUBE CURRENT [mA]

90kV
TUBE CURRENT [mA]

110kV
400 200 100kV
125kV 110kV
125kV
300 150kV 150
150kV

200 100

100 50

0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]

-5-
E7100X

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant potential high-voltage generator
Stator Power Frequency 50Hz
Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
70kV 60kV
800 400
60kV 50kV 40kV
80kV
700 50kV 70kV
90kV
40kV 80kV
600 100kV 300

TUBE CURRENT [mA]


TUBE CURRENT [mA]

110kV 90kV
500 125kV 100kV
110kV
400 150kV 200 125kV
150kV
300

200 100

100

0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]

Maximum Rating Charts


(Spot-Film Rating Charts)

Conditions: Tube Voltage


Constant potential high-voltage generator
Stator Power Frequency 50Hz
Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
50kV 40kV
60kV 50kV 300
600 60kV
70kV 40kV
70kV
80kV 250
500
90kV 80kV
TUBE CURRENT [mA]
TUBE CURRENT [mA]

100kV 200 90kV


400
110kV 100kV
125kV 110kV
150 125kV
300
150kV
150kV
100
200

50
100

0
0 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.02 0.05 0.1 0.3 0.7
0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s]
EXPOSURE TIME [s]

-6-
E7100X

Emission & Filament Characteristics

Constant potential high-voltage generator

Nominal Focal Spot Value: 1.2


1000 28

40kV

800 50kV 24

FILAMENT VOLTAGE [V]


60kV
TUBE CURRENT [mA]

80kV
600 100kV 20

125~150kV

400 16

200 12

Ef

0 8
4.5 4.7 4.9 5.1 5.3 5.5 5.7 5.8

FILAMENT CURRENT [A]

For Reference Only

Nominal Focal Spot Value: 0.6

500 16

60kV
400 80kV 14
FILAMENT VOLTAGE [V]
TUBE CURRENT [mA]

100kV

300 125~150kV 12

200 10
Ef

50kV
100 8
40kV

0 6
4.0 4.2 4.4 4.6 4.8 5.0 5.2

FILAMENT CURRENT [A]

For Reference Only

-7-
E7100X

Thermal Characteristics

X-ray Tube Assembly Heating / Cooling Curve

1070
1000 COOLING 215W

900

800
HEAT STORAGE [kJ]

700

600

500

400

300

200

100 HEATING

0
0 60 120 180 240 300

TIME [min]

Anode Heating / Cooling Curve

210
200

180 COOLING
710W
160
HEAT STORAGE [kJ]

140

120
450W
100

80 300W

60
200W
40

20
HEATING
0
0 2 4 6 8 10
TIME [min]

The heating curves are showing examples of


average input power to the anode in operation.

-8-
E7100X

Dimensional Outline
Unit: mm
φ174

ANODE
4-M5 9 DEEP

φ195

244
CENTRAL RAY AND REFERENCE AXIS

164
12°
FOCAL SPOT POSITION
16.5
41.27

149
90

496
φ
92
45°

145
4-M6 8 DEEP

S
16.5

252
CATHODE
C

TERMINAL CONNECTIONS

9 8 7 6 5 1 2 3 4 E

FOCAL SPOT
60±1

NC NC NC NC
C1 M A
100

ET
TEMPERATURE RELAY
(NORMALLY CLOSED)
Note) Do not connect termial No.1 and No.5 or No.6
57

in series circuit.
100
φ70

CENTER OF FOCAL SPOT


A

85
B
210 2-M63.5,P=1.25
-1.5mm≦A≦1.5mm
-1.5mm≦B≦1.5mm

EXPLANATION OF SYMBOLS
CATHODE TERMINAL
C : COMMON
L : LARGE FOCUS
S : SMALL FOCUS

TERMINAL CONNECTIONS
C1 : COMMON : CENTRAL X-RAY
M : MAIN WINDING OF THE STATOR ANODE & CATHODE TERMINAL
A : AUX. WINDING OF THE STATOR : IEC60526 TYPE
NC : NON-CONNECTION
ET : EARTH TERMINAL

-9-
Marketing Engineering Group, Sales Department
1385, Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan
Tel: +81-287-26-6666 Fax: +81-287-26-6060
https://etd.canon

・The head office of Canon Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Environmental Management System ISO14001.
・Canon Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Quality Management Systems ISO9001 and ISO13485.
Product scope is referred to the following URL. https://etd.canon/eng/company/quality.htm.
PE-E7100X 2018-11