You are on page 1of 12

Product Information PI

ROTANODE™
E7813X
E7813GX

Rotating Anode X-ray Tube Assembly

 Rotating anode X-ray tube for the purpose of


general diagnostic X-ray procedures.

 Specially processed tungsten faced


molybdenum target of 74 mm diameter.

 These tubes have foci 2.0 and 1.0, and are


available for a maximum tube voltage 150 kV.

 Accommodated with IEC60526 type


high-voltage cable receptacles.

General Data

IEC Classification (IEC60601-1:2005) ................................. Class Ⅰ ME EQUIPMENT

Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage (IEC60613:2010):
Radiographic ...................................................................................................................... 150 kV
Fluoroscopic ....................................................................................................................... 125 kV
Nominal Focal Spot Value (IEC60336:2005):
Large Focus ............................................................................................................................. 2.0
Small Focus ............................................................................................................................. 1.0
Nominal Anode Input Power (at 0.1s):
50 Hz 60 Hz 180 Hz
Large Focus ................................................................................. 42.8 kW 46 kW 80 kW
Small Focus .................................................................................... 19 kW 21.2 kW 35 kW
Nominal Radiographic Anode Input Power (IEC60613:2010)
50 Hz 60 Hz 180 Hz
Large Focus ................................................................................. 42.8 kW 46 kW 80 kW
Small Focus .................................................................................... 19 kW 21 kW 35 kW

★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
E7813X / E7813GX

Motor Ratings:
Stator: XS-AL
Starting Running
Driven Frequency [Hz] 180 60 180 60
Input Power [W] 1100 910 83 83
Voltage 3) 5) [V] 220 130 60 40
Current 4) [A] 5.7 7.8 1.6 2.3
Min. Speed Up 1) 5) [s] 1.2 0.8 - -
Capacitor [μF] 6 44 6 44
Min. Braking 2) 7) [s] 3 / 90V (DC)
Note 1) The speed up time from normal speed to high speed is 2/3 times of the specified speed
up time from 0 to high speed, which is described on motor table.
2) To be applied for high speed rotation.
3) Applied voltage between common and main terminal.
4) Common current.
5) The every applied voltage must be never exceeded 110% of the above specification.
6) No more than two high speed starts per minute are permissible.
7) The speed-up time is allowed up to 110% of the above specification.

Anode Speed:
50 Hz ............................................................................................................ Minimum 2700 min-1
60 Hz ............................................................................................................ Minimum 3200 min-1
180 Hz ............................................................................................................ Minimum 9700 min-1
Stator Resistance:
Common-Main Winding ........................................................................................................ 9.4 Ω
Common-Auxiliary Winding ................................................................................................ 28.3 Ω
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 ℃
Mode of Operation ........................................................................................................... Intermittent

Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 476 mm
Maximum Diameter ....................................................................................................... 152.4 mm
Target:
Anode Angle ............................................................................................................... 16 degrees
Diameter ............................................................................................................................ 74 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Filtration:
Permanent Filtration ................................................................ 0.9 mm Al / 75 kV IEC60522:1999
Available Additional Filter combination (0.4 - 1.5 mm) ................... Maximum 2.4 mm Al / 75 kV
Radiation Protection (In accordance with IEC60601-1-3:2008):
Leakage Technique Factor ................................................................................... 150 kV, 3.4 mA
X-ray Coverage .................................................................................. 354 × 354 mm at SID 750 mm
Weight (Approx.) ........................................................................................................................ 18 kg
High Voltage Receptacle ..................... To meet the requirements of IEC60526 Corrigendum1:2010
Cooling Method .................................................................................................. Natural or forced air
Tube Housing Model Number:
E7813X .......................................................................................................................... XH-106V
E7813GX ....................................................................................................................... XH-180

-2-
E7813X / E7813GX

Absolute Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage (IEC60613:2010):


Radiographic ...................................................................................................................... 150 kV
Fluoroscopic ....................................................................................................................... 125 kV
Between Anode (or Cathode) and Ground ................................................................................ 75 kV
Minimum X-ray Tube Voltage ................................................................................................... 40 kV
Maximum X-ray Tube Current (IEC60613:2010) .................................................... See rating charts
Large Focus .................................................................................................................... 1000 mA
Small Focus ...................................................................................................................... 600 mA
Maximum Filament Current:
Large Focus .......................................................................................................................... 5.2 A
Small Focus .......................................................................................................................... 5.2 A
Filament Voltage:
Large Focus (At maximum filament current 5.2 A) ....................................................... 7.3 ~ 9.9 V
Small Focus (At maximum filament current 5.2 A) ....................................................... 5.6 ~ 7.6 V
Filament Frequency Limits ................................................................................................ 0 ~ 25 kHz
Continuous Anode Input Power (IEC60613:2010) ................................................ 142 W (200 HU/s)
(Fluoroscopic, repeated radiographic or mixed exposure)
Thermal Characteristics:
Anode Heat Content ......................................................................................... 210 kJ (300 kHU)
Maximum Anode Heat Dissipation ................................................................... 475 W (667 HU/s)
X-ray Tube Assembly Heat Content ................................................................ 900 kJ (1250 kHU)
Nominal Continuous Input Power (IEC60613:2010):
Without Air-circulator ................................................................................ 200 W (16 kHU/min)

-3-
E7813X / E7813GX

Environmental Limits

Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 ℃
Humidity ........................................................................................................................ 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -20 ~ 70 ℃
Humidity ........................................................................................................................ 20 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa

-4-
E7813X / E7813GX

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant Potential High-Voltage Generator
Stator Power Frequency 50Hz

Nominal Focal Spot Value: 2.0


70kV 60kV 50kV
700
40kV
80kV

600 90kV
100kV
TUBE CURRENT [mA]

500
110kV
125kV
400
150kV

300

200

100

0
0.01 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10
0.03 0.07 0.3 0.7
EXPOSURE TIME [s]
Refer to IEC60613:2010

Nominal Focal Spot Value: 1.0


50kV 40kV
400

60kV

300 70kV
TUBE CURRENT [mA]

80kV
90kV
100kV
200 110kV
125kV
150kV

100

0
0.01 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10
0.03 0.07 0.3 0.7
EXPOSURE TIME [s]
Refer to IEC60613:2010

-5-
E7813X / E7813GX

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant Potential High-Voltage Generator
Stator Power Frequency 60Hz

Nominal Focal Spot Value: 2.0


80kV 70kV 60kV 50kV
700
40kV
90kV
600
100kV
110kV
TUBE CURRENT [mA]

500
125kV

400 150kV

300

200

100

0
0.01 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10
0.03 0.07 0.3 0.7
EXPOSURE TIME [s]
Refer to IEC60613:2010

Nominal Focal Spot Value: 1.0


50kV 40kV
400 60kV

70kV

300 80kV
TUBE CURRENT [mA]

90kV
100kV
110kV
200 150kV
125kV

100

0
0.01 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10
0.03 0.07 0.3 0.7
EXPOSURE TIME [s]
Refer to IEC60613:2010

-6-
E7813X / E7813GX

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant Potential High-Voltage Generator
Stator Power Frequency 180Hz

Nominal Focal Spot Value: 2.0


90kV
1000 80kV
100kV 70kV
110kV
60kV

800 125kV 50kV


TUBE CURRENT [mA]

40kV
150kV
600

400

200

0
0.01 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10
0.03 0.07 0.3 0.7
EXPOSURE TIME [s]
Refer to IEC60613:2010

Nominal Focal Spot Value: 1.0


600
70kV 60kV

80kV
500 50kV
90kV
TUBE CURRENT [mA]

40kV
400 100kV
110kV
125kV
300
150kV

200

100

0
0.01 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10
0.03 0.07 0.3 0.7
EXPOSURE TIME [s]
Refer to IEC60613:2010

-7-
E7813X / E7813GX

Emission & Filament Characteristics


Constant Potential High-Voltage Generator

Nominal Focal Spot Value: 2.0


1000 16

900 40kV
60kV
800 14

FILAMENT VOLTAGE [V]


80kV
TUBE CURRENT [mA]

700
100kV
600 12
125~150kV
500

400 10

300

200 8

100
Ef
0 6
4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0 5.1 5.2

FILAMENT CURRENT [A]


Note1) For Reference Only
Note2) Refer to IEC60613:2010

Nominal Focal Spot Value: 1.0


600 16

40kV
500 60kV
FILAMENT VOLTAGE [V]

80kV
TUBE CURRENT [mA]

400 100kV 12

125~150kV
300

200 8

100

Ef
0 4
4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0 5.1 5.2

FILAMENT CURRENT [A]

Note1) For Reference Only


Note2) Refer to IEC60613:2010

-8-
E7813X / E7813GX

Thermal Characteristics

X-ray Tube Assembly Heating / Cooling Curve


900
200W
800

700
HEAT STORAGE [kJ]

600

500

400
COOLING

300

200

100
HEATING
0
0 60 120 180 240 300 360

TIME [min]

Anode Heating / Cooling Curve

210
200 475W
180

160
HEAT STORAGE [kJ]

140

120
COOLING
100

80

60
HEATING
40

20

0
0 2 4 6 8 10 12
TIME [min]
The heating curves are showing examples of
average input power to the anode in operation.

-9-
E7813X / E7813GX

Dimensional Outline of E7813X


Unit: mm

φ140

CENTRAL RAY AND REFERENCE AXIS
ANODE
4-M6 8 DEEP

φ152.4

237
11.1

146.5
°
16
125.4

476
90


φ
°
45

134.5
4-M5 10 DEEP

90
11.1

18

239

CATHODE

TERMINAL CONNECTIONS
FOCAL SPOT
53

1 2 3 4 5 6 7 8 9 E
81

NC NC NC NC

C1 M A ET
111
(φ70)

TEMPERATURE RELAY
(NORMALLY CLOSED)

Note) Do not connect terminal No.1 and
      No.5 or No.6 in series circuit.

(71) 2-M63.5 P=1.27

(192)

CENTER OF FOCAL SPOT

EXPLANATION OF SYMBOLS 
CATHODE TERMINAL
C :COMMON B
L :LARGE FOCUS -1.5mm≦A≦1.5mm
S :SMALL FOCUS -1.5mm≦B≦1.5mm

TERMINAL CONNECTIONS
C1:COMMON
M :MAIN WINDING OF THE STATOR
A :AUX. WINDING OF THE STATOR   :CENTRAL X-RAY
NC:NON-CONNECTION ANODE & CATHODE TERMINAL
ET:EARTH TERMINAL   :IEC60526 TYPE

- 10 -
E7813X / E7813GX

Dimensional Outline of E7813GX


Unit: mm

111
TERMINAL CONNECTIONS
2-M63.5 P=1.27 (φ70)

1 2 3 4 5 6 7 8 9 E

NC NC NC NC

(192)
FOCAL SPOT
C1 M A ET
TEMPERATURE RELAY
(NORMALLY CLOSED)

(71)
Note) Do not connect terminal No.1 and

81
      No.5 or No.6 in series circuit.

53
φ140
CENTRAL RAY AND REFERENCE AXIS

ANODE

4-M6 8 DEEP
237

φ152.4
146.5
16
°

125.4
φ

476

90

45
°
134.5

90
4-M5 10 DEEP

18

239

L S

CATHODE

CENTER OF FOCAL SPOT
EXPLANATION OF SYMBOLS 

CATHODE TERMINAL
C :COMMON
L :LARGE FOCUS B
S :SMALL FOCUS -1.5mm≦A≦1.5mm
-1.5mm≦B≦1.5mm
TERMINAL CONNECTIONS
C1:COMMON
M :MAIN WINDING OF THE STATOR
A :AUX. WINDING OF THE STATOR   :CENTRAL X-RAY
NC:NON-CONNECTION ANODE & CATHODE TERMINAL
ET:EARTH TERMINAL   :IEC60526 TYPE

- 11 -
E7813X / E7813GX

OVERSEAS SUBSIDIARIES AND AFFILIATES

EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

For Sales & Technical Services, please contact the following representative:

・TOSHIBA ELECTRONICS EUROPE GMBH


HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

・TOSHIBA AMERICA ELECTRONIC COMPONENTS, INC.


2150 EAST LAKE COOK ROAD, SUITE 310
BUFFALO GROVE, ILLINOIS 60089 USA
PHONE +1 (847) 484-2400 FAX +1 (847) 541-7287

・TOSHIBA ELECTRON DEVICES & MATERIALS (SHANGHAI) CO., LTD. (TEMS)


RM1606, SH-PLAZA,
No.336, XIZANG ROAD (MIDDLE), SHANGHAI, 200001, CHINA
PHONE +86 (21) 6361-0077 FAX +86 (21) 6351-5760

(Product Development Engineering Department)


1385 SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN, 324-8550, JAPAN
PHONE: +81-287-26-6666 FAX: +81-287-26-6060
http://www.toshiba-tetd.co.jp/

・Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, ISO14001.
・Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO9001, ISO13485.

You might also like