You are on page 1of 14

Product Information PI

ROTANODE™
E7290AX

Rotating Anode X-Ray Tube Assembly

 Rotating anode X-ray tube assembly for


magnification mammographic and soft X-ray
radiographic operations.

 The insert tube has molybdenum target of 86 mm


diameter anode disc and beryllium window.

 This tube has foci 0.3 and 0.1, and is available


for a maximum tube voltage 39 kV.

 This tube assembly is accommodated with


ALDEN P500 type high-voltage connector.

General Data

IEC Classification (IEC60601-1:1988+Amd.1:1991+Amd.2:1995)


............................................................................................. Class Ⅰ ME EQUIPMENT

Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding ....................................................................................................... Cathode-grounded
Nominal Tube Voltage:
Radiographic ........................................................................................................................ 39 kV
Nominal Focal Spot Value: 2)
Large Focus 1) .......................................................................................................................... 0.3
Small Focus 1) .......................................................................................................................... 0.1
Note 1) Focal spots are defined on the central ray.
2) The tube is biased to obtain the Focal Spot.

Nominal Anode Input Power (at 0.1s):


180 Hz
Large Focus ....................................................................................................................... 3.1 kW
Small Focus ....................................................................................................................... 1.0 kW

★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
E7290AX

Motor Ratings:
Stator: XS-AL
Starting Running
Driven Frequency [Hz] 180 50/60 180 50/60
Input Power [W] 1100 2150 83 75
3) 5)
Voltage [V] 220 200 60 40
4)
Current [A] 5.7 11.0 1.6 2.1
1) 7)
Min. Speed Up [s] 2.0 0.5 - -
Capacitor [μF] 6 44 6 44
2) 7)
Braking Time [s] 3 s / 70 V (DC)
Note 1) The speed up time from normal speed to high speed is 2/3 times of the specified speed up
time from 0 to high speed, which is described on motor rating table.
2) To be applied for high speed rotation.
3) Applied voltage between common and main terminal.
4) Common Current.
5) The every applied voltage must be never exceeded 110% of the above specification.
6) No more than two high speed starts per minute are permissible.
7) The speed-up time is allowed up to 110% of the above specification.

Anode Speed:
50 Hz ........................................................................................................... Minimum 2700 min-1
60 Hz ........................................................................................................... Minimum 3200 min-1
180 Hz .......................................................................................................... Minimum 9700 min-1
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
Normal operating range of the housing temperature ........................................................ 16 ~ 75 ℃
Mode of Operation ........................................................................................................... Intermittent

Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 309 mm
Maximum Diameter .......................................................................................................... 138 mm
Target:
Anode Angle:
Large Focus ........................................................................................................... 16 degrees
Small Focus ............................................................................................................... 10 degrees
Diameter ............................................................................................................................ 86 mm
Construction ..................................................................................................... Molybdenum Alloy
Permanent Filtration ......................................................................At equivalent 0.0 mm (Be 0.8 mm)
Radiation Protection (To meet the requirements of IEC60601-1-3):
Leakage Technique Factor .................................................................................... 39 kV, 4.9 mA
X-ray Coverage ........................................................................................... See Dimensional Outline
(Drawing of X-ray Coverage)
Weight (Approx.) ........................................................................................................................ 10 kg
High Voltage Receptacle ............................................................................................. ALDEN, P500
Low Voltage Receptacle .......................................................................................... AMP #206705-1
Cooling Method ................................................................................................................. Natural air

-2-
E7290AX

Absolute Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage:


Radiographic ........................................................................................................................ 39 kV
Between Anode and Ground ..................................................................................................... 39 kV
Minimum X-ray Tube Voltage ................................................................................................... 22 kV
Maximum X-ray Tube Current ................................................................................. See rating charts
Large Focus ...................................................................................................................... 110 mA
Small Focus ........................................................................................................................ 35 mA
Maximum Filament Current:
Large Focus .......................................................................................................................... 3.9 A
Small Focus .......................................................................................................................... 3.9 A
Filament Frequency Limits ................................................................................................ 0 ~ 25 kHz
Long Time Maximum Input ..................................................................................... 120 W (170 HU/s)
Thermal Characteristics:
Anode Heat Content .......................................................................................... 210kJ (300 kHU)
Maximum Anode Heat Dissipation ................................................................... 525 W (740 HU/s)
X-ray Tube Assembly Heat Content .................................................................. 610 kJ (860 kHU)
Maximum Continuous Heat Dissipation:
Without Air-circulator .................................................................................... 72 W (6 kHU/min)

Environmental Limits

Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 ℃
Humidity ........................................................................................................................ 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -20 ~ 70 ℃
Humidity ........................................................................................................................ 20 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa

-3-
E7290AX

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant Potential High-Voltage Generator
Stator Power Frequency 180 Hz

Nominal Focal Spot Value: 0.3


120
EMISSION LIMIT

0.1~5 sec
TUBE CURRENT [mA]

100

90
5~10 sec

80

70
22 25 30 35 38 39

TUBE VOLTAGE [kV]

Nominal Focal Spot Value: 0.1


50

40
0.1~5 sec
TBE CURRENT [mA]

EMISSION LIMIT

5~10 sec

30

20
22 25 30 35 38 39
TUBE VOLTAGE [kV]

-4-
E7290AX

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant Potential High-Voltage Generator
Stator Power Frequency 60 Hz

Nominal Focal Spot Value: 0.3

100
EMISSION LIMIT

90
TUBE CURRENT [mA]

0.1~5 sec
80

5~10 sec
70

60

50
22 25 30 35 38 39

TUBE VOLTAGE [kV]

Nominal Focal Spot Value: 0.1

40

30 EMISSION LIMIT
0.1~5 sec
TUBE CURRENT [mA]

5~10 sec

20

10
22 25 30 35 38 39
TUBE VOLTAGE [kV]

-5-
E7290AX

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant Potential High-Voltage Generator
Stator Power Frequency 50 Hz

Nominal Focal Spot Value: 0.3


100
EMISSION LIMIT

90
0.1~5 sec
TBE CURRENT [mA]

80

5~10 sec

70

60

50
22 25 30 35 38 39

TUBE VOLTAGE [kV]

Nominal Focal Spot Value: 0.1

40

30 EMISSION LIMIT
TUBE CURRENT [mA]

0.1~5 sec

5~10 sec

20

10
22 25 30 35 38 39

TUBE VOLTAGE [kV]

-6-
E7290AX

Emission & Filament Characteristics

Constant Potential High-Voltage Generator

Nominal Focal Spot Value: 0.3


110

100 25kV 8

30kV

FILAMENT VOLTAGE [V]


TUBE CURRENT [mA]

80 35kV 7

39kV 20kV
60 6

40 5
Ef

20 4

0 3
3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9
FILAMENT CURRENT [A]
For Reference Only

Nominal Focal Spot Value: 0.1

35 10
25kV
30kV
30 35kV 9

39kV
FILAMENT VOLTAGE [V]
TUBE CURRENT [mA]

25 8

20 7

20kV
15 6

10 5
Ef

5 4

0 3
3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9
FILAMENT CURRENT [A]

For Reference Only

-7-
E7290AX

Thermal Characteristics

X-ray Tube Assembly Heating / Cooling Curve


610
600

COOLING
500

72W
HEAT STORAGE [kJ]

400

300

200

100

HEATING
0
0 50 100 150 200

TIME [min]

Anode Heating / Cooling Curve

210
200
COOLING
180

160
HEAT STORAGE [kJ]

140

120

100

80
120W
60

40

20
HEATING
0
0 2 4 6 8 10
TIME [min]

The heating curves are showing examples of average


input power to the anode in operation.

-8-
E7290AX

Self-Biassing Method (1)


(With Self Bias Resistor)

[1] Circuit Diagram

Large Focus Bias Resistor


Small Focus Bias Resistor
Earth

Protective
Earth

Motor Power Supply Filament


Power Supply

1 2 3 4 5 6 7 8 9
Filament Stator Terminal

Cathode
Auxiliary Winding Coil

Grid
Main Winding Coil

Small Filament
Common
Large Filament
High Voltage
Supply
X-ray Tube
X-ray Tube Assembly
Protective Earth

[2] Bias Resistor


Bias resistor is specified by the inspection sheet of each tube.
(1) Large Focus : 0, 100, 200, 300, 390, 500, 600, 700, 800, 900, 1000Ω
(2) Small Focus : 0, 0.5, 1, 1.5, 2, 2.4, 3, 3.6, 4kΩ

[3] Bias Resistor Connection


(1) The filament common-terminal (Terminal No.8) shall be connected to the grid-terminal
(Terminal No.5) through the bias resistor.
(2) The grid-terminal (terminal No.5) shall be connected to the functional earth-terminal.

-9-
E7290AX

Biassing Method (2)


(With Bias-Voltage Supply)

[1] Circuit Diagram

Large Focus Bias Voltage


Small Focus Bias Voltage
Earth
+ +


Protective
Earth

Motor Power Supply Filament


Power Supply

1 2 3 4 5 6 7 8 9
Filament Stator Terminal

Cathode
Auxiliary Winding Coil

Grid
Main Winding Coil

Small Filament
Common
Large Filament
High Voltage
Supply
X-ray Tube
X-ray Tube Assembly
Protective Earth

- 10 -
E7290AX

[2] Bias Voltage


* ; Bias resistor is specified by the inspection sheet of each tube.

Large Focus
0 -10 -20 -30 -40 -50 -60 -70 -80 -90 -100
Bias Voltage [V]

0 -8 -16 -24 -32 -40 -48 -56 -64 -72 -80


0 -7 -14 -21 -28 -36 -42 -49 -56 -63 -70

(Bias: 1000Ω)
(Bias: 100Ω)

(Bias: 200Ω)

(Bias: 300Ω)

(Bias: 390Ω)

(Bias: 500Ω)

(Bias: 600Ω)

(Bias: 700Ω)

(Bias: 800Ω)

(Bias: 900Ω)
(Bias: 0Ω)

0 0 0 0 0 0 0 0 0 0 0
20 25 30 35 39
Tube Voltage [kV]

0 -15 -30 -45 -60 -75 -90 -105 -120 Small Focus
Bias Voltage [V]

0 -10 -20 -30 -40 -50 -60 -70 -80


(Bias: 0.5kΩ)

(Bias: 1.5kΩ)

(Bias: 2.4kΩ)

(Bias: 3.6kΩ)
(Bias: 0kΩ)

(Bias: 1kΩ)

(Bias: 2kΩ)

(Bias: 3kΩ)

(Bias: 4kΩ)

0 0 0 0 0 0 0 0 0
20 25 30 35 39
Tube Voltage [kV]

[3] Bias Voltage Connection


(1) The filament common-terminal (Terminal No.8) shall be connected to the earth terminal.
(2) The grid-terminal (Terminal No.5) shall be connected to the minus( ) terminal of the
bias-voltage supply.
(3) The plus( ) terminal of the bias-voltage supply shall be connected to the earth terminal.

- 11 -
E7290AX

Dimensional Outline
Unit: mm

φ138

0.5
.1±
92
90

φ
81

45°
FOCAL SPOT

57
LF 16°
SF 10°

90

90
CENTRAL RAY AND REFERENCE AXIS

(54.5)
11.5±0.9 4-M6 8 DEEP

43±1.5 309
252

CENTRAL OF FOCAL SPOT


The Central of focal spot is the point midway
LF between the centers of the two focal spots.
SF A

B
A:±2.0mm,B:±2.0mm
20

PIN CONFIGURAION
AMP#206705-1
6
3 5
2 9
8
1 7
4 TERMINAL CONNECTION
1 2 3 4 5 6 7 8 9 AMP#206705-1

A M S C L
NC

ET GT

TEMPERATURE RELAY
(NORMALY CLOSED)
HIGH VOLTAGE CONNECTOR
(ALDEN, P500)

C : COMMON GT : GRID TERMINAL


(EARTH TERMINAL OF INSERT TUBE) ET : EARTH TERMINAL OF HOUSING ASS
L : LARGE FOCUS NT : NON-CONNECTION
S : SMALL FOCUS : CENTRAL X-RAY
M : MAIN WINDING OF THE STATOR COIL ANODE TERMINAL
A : AUX. WINDING OF THE STATOR COIL : ALDEN, P500 FOT

- 12 -
E7290AX

X-ray Coverage
Unit: mm

Focal Spot Mark

34
45
90

【X-ray Coverage】
SID 600mm
SID 600

・Large Focus A: 370 mm


B: 260 mm

・Small Focus A: 300 mm


B: 190 mm

B
X-ray Coverage
Cathode Side

- 13 -
E7290AX

OVERSEAS SUBSIDIARIES AND AFFILIATES

EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

For Sales & Technical Services, please contact the following representative:

・TOSHIBA ELECTRONICS EUROPE GMBH


HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

・TOSHIBA AMERICA ELECTRONIC COMPONENTS, INC.


2150 EAST LAKE COOK ROAD, SUITE 310
BUFFALO GROVE, ILLINOIS 60089 USA
PHONE +1 (847) 484-2400 FAX +1 (847) 541-7287

・TOSHIBA ELECTRON DEVICES & MATERIALS (SHANGHAI) CO., LTD. (TEMS)


RM1606, SH-PLAZA,
No.336, XIZANG ROAD (MIDDLE), SHANGHAI, 200001, CHINA
PHONE +86 (21) 6361-0077 FAX +86 (21) 6351-5760

(Product Development Engineering Department)


1385 SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN, 324-8550, JAPAN
PHONE: +81-287-26-6666 FAX: +81-287-26-6060
http://www.toshiba-tetd.co.jp/

・Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, ISO14001.
・Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO9001, ISO13485.

You might also like