You are on page 1of 14

Product Information PI

ROTANODE™
E79016X

Rotating Anode X-ray Tube Assembly

 The rotating anode X-ray tube assembly E79016X is


designed with Hydrodynamic bearing lubricated by
liquid metal and its rotor offers continuous high
speed, extra low noise and high reliability.

 This tube has the grid control function, which is


capable of high-speed pulse-fluoroscopy suitable for
high speed such as cine- fluoroscopy.
 It performs high cooling rate 3.5kW and high patient
throughput with the featuring of 2130kJ anode heat
capacity and water cooling heat exchanger.

General Data

IEC Classification (IEC60601-1:2005+A1:2012) .................... Class Ⅰ ME EQUIPMENT

Electrical:
Circuit:
High Voltage Generator ............................................. Constant Potential High-voltage Generator
Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage:
Radiographic ...................................................................................................................... 125 kV
Fluoroscopic:
Continuous .................................................................................................................... 125 kV
Grid-Control ................................................................................................................... 120 kV
Nominal Focal Spot Value:
Large Focus ............................................................................................................................. 1.0
Middle Focus ............................................................................................................................ 0.6
Small Focus ............................................................................................................................. 0.3
Nominal Anode Input Power (at 0.1s) ..................................................................... See rating charts
Large Focus ...................................................................................................................... 100 kW
Middle Focus ....................................................................................................................... 48 kW
Small Focus ........................................................................................................................ 17 kW
Nominal Radiographic Anode Input Power:
Large Focus ........................................................................................................................ 96 kW
Middle Focus ....................................................................................................................... 46 kW
Small Focus ........................................................................................................................ 16 kW

★The information contained herein is presented only as a guide for the application of our products. No responsibility is assumed by Canon Electron Tubes
& Devices Co., Ltd. (CETD) for any infringements of patents or other rights of the third parties which may result from its use.
No license is granted by implication or otherwise under any patent or patent rights of CETD or others.
★The information contained herein may be changed without prior notice. It is therefore, advisable to contact to CETD before processing with the design of
equipment incorporating this product.
E79016X

Anode Speed ..................................................................................................... Minimum 9000 min-1


Cut-off Grid Voltage (Small Focus) .................................................................................. -2900 VDC
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
(Measured by DC 500 V)
Heat Exchanger Input Voltage / Input Current .............................................................. AC 100V, 2A
Heat Exchanger Source Frequency ..................................................................................... 50/60 Hz
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 °C
Mode of Operation ........................................................................................................... Intermittent
Envelope Current (125kV/100mA) ............................................................................ (Approx) 10 mA
Envelope Voltage (125kV/100mA) .............................................................................................. 0 kV

Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 562 mm
Maximum Diameter .......................................................................................................... 213 mm
Target:
Anode Angle (Effective) .............................................................................................. 11 degrees
Diameter .......................................................................................................................... 140 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Permanent Filtration ................................................................. 1.1 mm Al / 75 kV IEC60522:1999
Radiation Protection (In accordance with IEC60601-1-3:2008):
X-ray Leakage ............................................................................................................... 0.44 mGy/h
Leakage Technique Factor ................................................................................... 125 kV, 18 mA
X-ray Coverage ................................................................................ 389 × 521 mm at SID 1000 mm
Weight (Approx.) ........................................................................................................................ 46 kg
High Voltage Receptacle ............................................................................... CLAYMOUNT MINI-75
Cooling Method ................................................................................. Water-Cooled Heat Exchanger
Inner Cooling Method .................................................................................................. Oil Circulation
Tube Model Number ................................................................................................. DSR-T7345G,U
Tube Housing Model Number ................................................................................................ XH-177

-2-
E79016X

Absolute Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage:


Radiographic ...................................................................................................................... 125 kV
Fluoroscopic:
Continuous .................................................................................................................... 125 kV
Grid-control ................................................................................................................... 120 kV
Between Anode (or Cathode) and Ground ............................................................................. 62.5 kV
Minimum X-ray Tube Voltage .....................................................................................................50 kV
Maximum X-ray Tube Current ................................................................................. See rating charts
Large Focus (1.0) ........................................................................................................ 1100 mA
Middle Focus (0.6) .......................................................................................................... 650 mA
Small Focus (0.3) .......................................................................................................... 180 mA
Maximum Filament Current:
Large Focus (1.0) .............................................................................................................. 5.9 A
Middle Focus (0.6) .............................................................................................................. 5.2 A
Small Focus (0.3) .............................................................................................................. 5.4 A
Filament Voltage: (At maximum filament current)
Large Focus (5.9 A) .............................................................................................. 12.5 ~ 16.7 V
Middle Focus (5.2 A) .............................................................................................. 10.0 ~ 13.4 V
Small Focus (5.4 A) .................................................................................................. 3.7 ~ 5.0 V
Filament Frequency Limits ................................................................................................ 0 ~ 25 kHz
Continuous Anode Input Power ........................................................................ 2200 W (3100 HU/s)
20 minutes ................................................................................................... 3000 W (4230 HU/s)
Thermal Characteristics:
Anode Heat Content ..................................................................................... 2130 kJ (3000 kHU)
Maximum Anode Heat Dissipation ............................................................... 5500 W (7700 HU/s)
X-ray Tube Assembly Heat Content .............................................................. 2050 kJ (2890 kHU)
Nominal Continuous Input Power:
With Water-Cooling Heat Exchanger (HEX-125 or HEX-119) .............. 3500 W (296 kHU/min)

Environmental Limits
Operating Limits:
Temperature ................................................................................................................ 18 ~ 40 ℃
Relative Humidity .......................................................................................................... 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Transport and Storage:
Temperature:
With cooling water empty ...................................................................................... -20 ~ 70 ℃
With cooling water filled ............................................................................................ 2 ~ 60 ℃
Relative Humidity .......................................................................................................... 20 ~ 90 %
(No condensation)
Atmospheric Pressure .................................................................................................50 ~ 106 kPa

-3-
E79016X

Absolute Maximum Rating Charts


(Single Load Rating Charts)

Conditions: Constant Potential High- voltage Generator


Anode Rotation Speed 9000min-1

Nominal Focal Spot Value: 1.0


100kV 90kV 80kV 70kV
1100
110kV
60kV
1000
125kV 50kV
900
TUBE CURRENT [mA]

800

700

600

500

400

300

200

100

0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 20 30 40
0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s]

Do not exceed values in above table at all the time.


Equipment setting and calibration errors must be considered not to exceed the table values.

Nominal Focal Spot Value: 0.6


80kV 70kV
650
60kV
600 90kV

100kV
50kV
500 110kV
TUBE CURRENT [mA]

125kV
400

300

200

100

0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 20 3040
0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s]

Do not exceed values in above table at all the time.


Equipment setting and calibration errors must be considered not to exceed the table values.

-4-
E79016X

Absolute Maximum Rating Charts


(Single Load Rating Charts)

Conditions: Constant Potential High-voltage Generator


Anode Rotation Speed 9000min-1

Nominal Focal Spot Value: 0.3


100kV 90kV 80kV
180
110kV 70kV
160
125kV
140 60kV
TUBE CURRENT [mA]

120 50kV

100

80

60

40

20

0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 20 30 40
0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s]

Do not exceed values in above table at all the time.


Equipment setting and calibration errors must be considered not to exceed the table values.

-5-
E79016X

Cine-fluorographic Rating Charts

Conditions: Constant Potential High-voltage Generator


Anode Rotation Speed 9000min-1

Nominal Focal Spot Value: 1.0


3
×10
100

90

80 SERIES DURATION
PULSE RATING [kV × mA]

70
5S
60
10S
50 15S
20S

40 30S
40S
30

20

10

0
0 100 200 300 400 500
PULSE WIDTH

1ms
50 100 150 200 250
2ms
20 40 60 80 100 120 140 160
3ms
20 40 60 80 100 120
4ms

FRAME RATE [F/s]

Do not exceed values in above table at all the time.


Equipment setting and calibration errors must be considered not to exceed the table values.

-6-
E79016X

Cine-fluorographic Rating Charts

Conditions: Constant Potential High-voltage Generator


Anode Rotation Speed 9000min-1

Nominal Focal Spot Value: 0.6

3
×10
60

50
PULSE RATING [kV × mA]

SERIES DURATION
40
5S
10S
15S
30 20S
30S
40S

20

10
0 100 200 300 400 500
PULSE WIDTH

1ms
50 100 150 200 250
2ms
20 40 60 80 100 120 140 160
3ms
20 40 60 80 100 120
4ms

FRAME RATE [F/s]

Do not exceed values in above table at all the time.


Equipment setting and calibration errors must be considered not to exceed the table values.

-7-
E79016X

Angiographic Rating Charts

Conditions: Constant Potential High-voltage Generator


Anode Rotation Speed 9000min-1
3
×10
90
Nominal Focal Spot Value: 1.0

80 N=5
PULSE RATING [kV × mA]

N=10
70
N=20
N=30
60

50
Nominal Focal Spot Value: 0.6
40 N=10
N=20
N=30
30

N : TOTAL NUMBER OF EXPOSURES


20
0.005 0.01 0.02 0.03 0.05 0.1 0.2
0.007 0.04 0.07
INDIVIDUAL EXPOSURE TIME [s]

Do not exceed values in above table at all the time.


Equipment setting and calibration errors must be considered not to exceed the table values.

3
×10
20
PULSE RATING [kV × mA]

18

16
Nominal Focal Spot Value: 0.3

14 N=10
N=20
N=30

12

N : TOTAL NUMBER OF EXPOSURES


10
0.005 0.01 0.02 0.03 0.05 0.1 0.2
0.007 0.04 0.07
INDIVIDUAL EXPOSURE TIME [s]

Do not exceed values in above table at all the time.


Equipment setting and calibration errors must be considered not to exceed the table values.

-8-
E79016X

Emission & Filament Characteristics


Constant Potential High-voltage Generator

Nominal Focal Spot Value: 1.0


1100 22
80kV
1000 20
90kV
900 100kV 18
TUBE CURRENT [mA]

FILAMENT VOLTAGE [V]


800 16
120KV
700 14
Ef
600 12

500 10
70kV
400 8
60kV
300 6
50kV
200 4

100 2

0 0
4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8 5.9

FILAMENT CURRENT [mA]

For Reference Only

Nominal Focal Spot Value: 0.6


650 17
600 16
70kV
550 15
80kV
500 14
FILAMENT VOLTAGE [V]
TUBE CURRENT [mA]

90kV
450 13
100kV
400 12
120kV
350 11
300 Ef 10
250 9
200 8
60kV
150 7
50kV
100 6
50 5
0 4
3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2

FILAMENT CURRENT [mA]

For Reference Only

-9-
E79016X

Emission & Filament Characteristics


Constant Potential High-voltage Generator

Nominal Focal Spot Value: 0.3


180 4.5

160 4
Ef
140 3.5

FILAMENT VOLTAGE [V]


TUBE CURRENT [mA]

80kV
120 90kV 3
100kV
100 2.5
120kV
80 70kV 2
60kV
60 1.5
50kV
40 1

20 0.5

0 0
4.0 4.2 4.4 4.6 4.8 5.0 5.2 5.4

FILAMENT CURRENT [mA]

For Reference Only

- 10 -
E79016X

Thermal Characteristics

X-ray Tube Assembly Heating / Cooling Curve


(Environmental Temperature : 25℃)
2050
2000

COOLING
1500
HEAT STORAGE [kJ]

2200W

1000 HEATING

1100W

500

0
0 10 20 30 40 50 60 70 80

TIME [min]

Anode Heating / Cooling Curve

2130
2000

1800
COOLING
1600 2200W
HEAT STORAGE [kJ]

1400
1500W
1200

1000 1000W

800

600
500W
400

200
HEATING
0
0 5 10 15 20 25 30
TIME [min]

- 11 -
E79016X

Dimensional Outline (1)


Unit: mm

Pb COLLIMATOR
259.5
130
126
74 4-M8 15DEEP

□22

φ136
φ177
φ165

φ213
111.8

128
61R
φ92

45°

4-M6 8DEEP

127 A CENTRAL OF FOCAL SPOT


211

B
-1.0mm A 1.0mm
-1.0mm B 1.0mm
CENTRAL RAY AND REFERENCE AXIS
562

293 269
11°
CATHODE2

C B
109.5

CATHODE1 12 ANODE
193
182 209

M (0.6) C (COMMON) S (0.3) C (COMMON)

G (GRID) L (1.0) G (GRID) C (COMMON)

CATHODE1 CATHODE2

ANODE & CATHODE TERMINAL: CLAYMOUNT MINI-75

- 12 -
E79016X

Dimensional Outline (2)


Unit: mm

VIEW from C VIEW from B

Focal Spot 133.5 CENTRAL X-RAY


146
60
116

114
73

170
64.5
73

73
106.5
650
114

114
107.5
31 109 109 32

TERMINAL CONNECTIONS 208.5

PROTECTIVE COVER (remove at installation)

LOW VOLTAGE WIRING DIAGRAM PORT TERMINAL CONFIGURATION

CONNECTOR (CNNA)
TERMINAL BLOCK (TB1)

6-M8 10DEEP

CONNECTOR (CNNB)
140
170
110

CONNECTOR (CNNA) TERMINAL BLOCK (TB1) CONNECTOR (CNNB)

R1
.
90 5
120
C1 S1 S2
62.5
107
SURGE ABSORBER 129
151

SYMBOL
LOW VOLTAGE TERMINAL
ET ・・・・・EARTH TERMINAL P ・・・・・PUMP MOTOR (50/60Hz 100V)
U ・・・・・STATOR U C1 ・・・・・TEMPERATURE RELAY COMMON
V ・・・・・STATOR V S1 ・・・・・TEMPERATURE RELAY 80℃ (NORMALLY CLOSE)
W ・・・・・STATOR W S2 ・・・・・TEMPERATURE RELAY 70℃ (NORMALLY CLOSE)
G ・・・・・GETTER FE ・・・・・FUNCTION EARTH TERMINAL
MT・・・・・METAL CENTER TERMINAL

- 13 -
Marketing Engineering Group, Sales Department
1385, Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan
Tel: +81-287-26-6666 Fax: +81-287-26-6060
https://etd.canon

・The head office of Canon Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Environmental Management System ISO14001.
・Canon Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Quality Management Systems ISO9001 and ISO13485.
Product scope is referred to the following URL. https://etd.canon/eng/company/quality.htm.