General Data
Electrical:
Circuit:
High Voltage Generator ............................................... Constant potential high-voltage generator
Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage ................................................................................................... 135 kV
Nominal Focal Spot Value (IEC60336:1993):
Large Focus .................................................................................................................... 1.7 × 1.6
Small Focus ............................................................................................................................. 0.9
Nominal Anode Input Power:
Large Focus ........................................................................................................................ 48 kW
Small Focus ........................................................................................................................ 30 kW
★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
E7804X
Motor Ratings:
Stator: XS-AQ
Boost-up
Start-up Stand-by Scanning
from Stand-by
Driven Frequency [Hz] 50 50 50 50
Input Power [W] 1450 380 820 570
Voltage [V] 200 110 152 134
Current [A] 8.5 4.3 6.6 5.5
Min. Speed Up [s] 6 - 2.7 -
Notes 1) To be obtained with 3-phase starter ST-7013 or equivalent.
2) These data are indicating standard values.
3) Anode keeps continuous rotation regardless X-ray output.
Mechanical:
Dimensions ................................................................................................. See dimensional outline.
Overall length ................................................................................................................... 493 mm
Maximum diameter .......................................................................................................... 213 mm
Target:
Anode Angle ................................................................................................................. 7 degrees
Diameter .......................................................................................................................... 140 mm
Construction ..................................................................................................... Rhenium-tungsten
Permanent Filtration ................................................................. 1.1 mm Al / 75 kV IEC60522:1999
Radiation Protection (To meets requirements of IEC60601-1-3):
Leakage Technique Factor ................................................................................. 135 kV, 29.6 mA
X-ray Coverage:
Longitudinal direction of tube axis ....................................................................... Anode side 3.3 °
Cathode side 21 °
Perpendicular direction of tube axis ..................................................................................... ± 27 °
Weight (Approx.):
Tube Housing Unit .................................................................................................. Approx. 41 kg
Heat Exchanger ...................................................................................................... Approx. 26 kg
High Voltage Receptacle .......................... To meet requirements of IEC60526 Corrigendum 1:2010
Cooling Method ........................................................................................................ Heat Exchanger
Position during operation .................................................................... ±30° with respect to tube axis
Position during shipping ................................................................................... Anode facing upward
Housing Model Number ......................................................................................................... XH-168
Centrifugal acceleration during rounding in a gantry ................................... Maximum 6.3 × 9.8 m/s2
-2-
E7804X
Environmental Limits
Operating Limits:
Temperature ................................................................................................................ 18 ~ 45 ℃
Humidity ........................................................................................................................ 30 ~ 80 %
(no condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Altitude ................................................................................................................................ 2000 m
Shipping and Storage Limits:
Temperature .............................................................................................................. -20 ~ 75 ℃
Humidity ........................................................................................................................ 20 ~ 90 %
(no condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa
-3-
E7804X
100
90 10.2kW 8kW
6kW
80 4.75kW
HEAT STRAGE [%]
70
3kW
60
2kW
50
40
COOLING
30
20
10 HEATING
0
0 3 6 9 12 15 18 21 24 27 30
TIME [min]
Notes
1. Heat storage of the anode is calculated based on the anode thermal characteristics.
This determines the X-ray input conditions for subsequent X-ray exposure. For specifying
conditions, contact TETD.
2. To avoid the over-load, the OLP (overload protection) program should be used to restrict the
X-ray input conditions. The X-ray tube assembly should not be used in systems which do not
incorporate the OLP program.
When absolutely impossible to use a system which incorporate the OLP program, contact TETD
for operating procedures.
3. Before you make the OLP program, contact TETD for the detail conditions.
-4-
E7804X
3000
HEAT STORAGE [kJ]
COOLING
2500
4kW
2000
1500
2kW
1000
500 HEATING
0
0 10 20 30 40 50 60
TIME [min]
-5-
E7804X
Emission Characteristics
Large Focus
Ep: Tube Voltage
400
Ep=120kV
300 Ep=135kV
TUBE CURRENT [mA]
200
Ep=100kV
Ep=80kV
100
0
4.0 4.2 4.4 4.6 4.8 5.0 5.2
Small Focus
Ep: Tube Voltage
400
300
Ep=120kV
TUBE CURRENT [mA]
Ep=135kV
200
Ep=100kV
Ep=80kV
100
0
4.0 4.2 4.4 4.6 4.8 5.0 5.2
-6-
E7804X
Filament Characteristics
12
LF
FILAMENT VOLTAGE [V]
11
SF
10
5
3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2
-7-
E7804X
-8-
E7804X
CABLE B
8 9
WHITE BLACK
TERMINAL TBS01
1 2 3 4 5 6 7 8 9
CABLE A 1 2 3 4 5 6 7
GREEN/YELLOW
ORANGE
YELLOW
PURPLE
BLACK
WHITE
WHITE
CONNECTOR (AMP*206708-1)
1 2 3 5 7 8 9
CONNECTOR (AMP*206705-1)
1 2 3 4 5 6 7 8 9
U V W E G G ME
-9-
E7804X
138
FLOW SWITCH WINDOW 131
140
47 39
6.7
AIR FLOW
SWITCH BUTTON
CABLE C
(4-M8)
2000
26
145 90
154
465
(4-M10)
150
GANTRY
ROTATING
DIRECTION
81 9
140
Y
X
Z
10
4
155 90
4 305 4
- 10 -
E7804X
- 11 -
E7804X
EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402
For Sales & Technical Services, please contact the following representative:
・Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, ISO14001.
・Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO9001, ISO13485.