Portable Optical
Calibration Laboratory
Computerised Slip Gauge Calibration
By Comparison
Refractometer
 Improved eyepiece and
focus adjuster.
 Improved calibration
mechanism with
covered screw.
Micrometer Pitch Calibration Screw Ring Gauge Calibration Micrometer Parallelism
SIP MUL 1000 SIP MU 214B Angle Dekkor and Dowell Prism  Improved textured
rubber sleeve on
barrel.
 Relocated cover
plate hinge for
easier cleaning.

Screw Plug Gauge Calibration Instrument Repair


Floating Carriage

AVAILABLE FROM YOUR DISTRIBUTOR

OXD-336-1010K
OXFORD PRECISION Victoria Works, Leicester, England
© The Kennedy Group Ltd. 03/00
OPERATOR’S MANUAL
OXFORD Portable Optical Refractometer
An optical instrument which uses the variation in refractive index of fluids at different
Application
For determining the concentration of oil in water To protect the machine and work piece from
emulsions and transparent solutions, such as corrosion and to reduce the risk of health hazards
concentrations to measure dilution ratios. machine shop coolant. arising from incorrect concentrations of water
Maintaining the correct mixture in coolant systems soluble oils.
Cover Plate Calibration Eyepiece & is vital to ensure optimum performance from To economise the use of water base cutting fluids
Screw Cover Focus Adjuster cutting tools. and grinding fluids.

Note: Dilution ratio does not necessarily give an performance. For further details regarding pH level
indication of the condition of the emulsion. please see below.
Bacteria and/or pH level can effect coolant

The pH Scale
15
pH is a measurement of the acidity or alkalinity seconds it will change colour. This colour can
1355
14
13
of an aqueous fluid. then be matched to a standard pH colour chart.
Callibrated 12 pH should be measured using pH papers, which Always test a sample from the outlet pipe of the
Refractive 1350 11
Scale
10
9
are available from your Oxford distributor. machine. This avoids picking up tramp oil from
See step 5 1345 8
7 The end of the paper should be dipped the surface of the coolant tank.
6
5 in to the test fluid. Within a matter of
Refraction 1340 4
3
Prism Barrel 2

No 1333
1
0
More Acidic Neutral More Alkaline
20°C BX %
pH
Level 1 2 3 4 5 6 7 8 9 10 11 12 13

Instructions Very little bacteria growth


Readings will vary according to Temperature. It is 7. After cleaning (drying) the prism and inner Aggressive to the skin
recommended that the Refractometer should be face of the cover plate, place a drop from the Very little bacteria growth Good for bacteria growth No
calibrated before every use. sample onto the centre of the prism. corrosion
Calibration 8. Close the lid and take a reading by looking on ferrous
Use a “standard” of known value, on the through the eyepiece (focus as necessary by materials
refractive index. Distilled water being zero, is rotating the adjuster) and record the figure
recommended. where the boundary line passes Typical fresh cutting fluids are in the range pH 9 to 9.5
1. Open cover plate, and ensure both prism, through the scale.
and inner face of cover plate are clean. Use a 9. Clean prism and cover after use. Each division on the scale is a factor of 10 from at pH9 and as they work they excrete acid waste
soft, clean and lint free cloth. The reading taken should be related to the the next, eg. pH 9 is ten times more alkaline which reduces the pH of the emulsion, further
2. Place a drop of distilled water onto the centre refraction index figure given for the particular than pH 8. improving their environment.
of the prism and close the cover. product and dilution by the manufacturer. Alkaline pH’s reduce corrosion of ferrous
3. The sample will spread over the surface of The pH level should never be above 10 as this materials and above pH 12 there is no corrosion
the prism and become a thin film. If this figure is not available the following is an will cause skin irritation. on these materials.
4. With a light source (daylight is preferable) alternative means of establishing one. Cutting fluids are usually formulated to have a
above the prism cover, look through the After following the calibration procedure (Steps 1 Bacteria prefer slightly acidic conditions, ie. fresh emulsion pH in the range 9.0 - 9.5, ie. as
eyepiece. Focus is adjusted by means of to 4,) mix a small sample of the emulsion in a just below pH 7 but start to be encouraged at high as possible to resist bacteria and corrosion
rotating the adjuster. The line of measuring cylinder, using the manufacturers anything below around pH 8. Bacteria can live without causing skin problems.
measurement is the boundary line recommended proportion of oil to water. This
between the blue and white zones. sample may then be measured following steps 7
5. Using the small screwdriver provided adjust to 9 to establish the correct scale reading for that
the calibration screw until the boundary line particular product and dilution. SPECIFICATIONS:
is exactly at zero on the scale. You can now compare, using steps 6 to 9, a
The refractometer is now ready for use. sample from the machine to determine the Range ND: 1.333 -1.356 Bx:0-15% Accuracy: ± 0.0003 ± 0.2%
6. To measure a sample from the coolant strength of the coolant in the system. Eyepiece Magnification: x11 Weight: 0.3kg
system, take a drop from the outlet pipe of The higher the reading, the more concentrated the Lattic Value: 0.0005 0.2% Dimensions: (W x H x L) 38 x 38 x 185mm
the machine. This avoids picking up any mix is. The nearer to zero the reading is, the Telescopic Lens Magnification: x 4
tramp oil from the surface of the coolant tank. weaker the solution.