Sl. Complaint No. Date Hospital Machine/Model Serial No: Company Contact Details Toll Free No.

Serial No: Company Contact Details Toll Free No. Contract Status
No.
1 18004258025 22.02.20 Karnal Logiq - V-5 1-5731568182 / ID- 83037060144617 GE- Delcom PU Services 9815274001 18001027750/18004257255 Warranty(Block service)
KPGI-KNL-01

2 KPGI-YMNR-01 02.02.20 Yamunanagar UPS-HiPULSE 201506HX0797 TATA EMERSON 9216285777 NILL PER VISIT CHARGE
80kva

3 KPGI-AMB-01 10.02.20 AMBALA SYING PUMP INFUSIA 1440086 FRESENIUS NL WARRANTY


SP7

4 KPGI-AMB-02 15.02.20 AMBALA DEHUMIDIFIER BD10 BIONAIRE 18883249907 NA PER VISIT CHARGE

5 KPGI-YMNR-02 22.02.20 YAMUNANAGAR ECG MACHINE TC102017 PHILIPS 8558833043 NA PER VISIT CHARGES
TC10

6 KPGI-KHTL-01 16.02.20 KHATAULI UPS SX80 201604SX0921 TATA EMERSON 9699922099 NA PER VISIT CHARGE
COMPLAINT LOG SHEET
OUTSOURCE COMPANY DETAIL Last PM done Bio-Medical Engineer Contact Details In House Repair FSE FSE contact No. Complaint Details Diagnosis

HT ULTRASOUND ,DELHI 22.01.2020 Amandeep Singh 9068820087 No Neeraj Kant 9815274001 CW not coming properly FOUNDED PROBE DEMAGE.
+91 98106 65713

APS SERVICES,NOIDA N/L AMANDEEP SINGH 9068820087 NO ADITI 9015884898 UPS BURNT ACCIDENTLY. REPAIRING AND REPLACING PARTS.
9015884898

NA NA AMANDEEP 9068820087 NO NOT GETTING ON MAIN BOARD PROBLEM

LOCAL MARKET NA AMANDEEP 9068820087 YES NA NA MOTOR NOT WORKING MOTOR FAULT

LOCAL MARKET NA AMANDEEP 9068820087 YES NA NA NOT GIVING PRINT DP CONNECTOR BROKEN

JDM POWER SOLUTION NA AMANDEEP 9068820087 NO NIKUNJ 9310459859 NOT GETTING ON COMMUNICATION + SMPS BOARD FAULT
MEERUT
8393887799
Corrective Action Spare Parts Cost Amount Down Time Up Time Total Status
Required.
NEED TO REPLACE PROBE WITH NEW ONE. Yes 260000 112000 11.00 AM
BUYING A REFUBRISHED PROBE WITH Machne partially functional or part ordered for that perticular call
1 YEAR WARRANTY

REPLACING ALL BURNT PARTS WITH NEW PARTS. YES 350000 230000 9AM Machne partially functional or part ordered for that perticular call
WITH 1 YEAR CMC

YES Machine down

REPAIR YES 5PM Machne partially functional or part ordered for that perticular call

REPLACEMENT YES Machne partially functional or part ordered for that perticular call

REPLACED YES 140000 118000 16-Feb 22-Feb 6 DAYS MACHINE WORKING OK.
5PM 1PM