Calibration Report of Mitutoyo 211-626K Roundness Measuring Instruments

MPC Control #: 200071011 Serial Number: 200071011

Asset ID: N/A Calibration Date: November 11, 2019

Section 1 - Roundness Test


Measurement Value Uncertainty
Standard Values Average Correction
1 2 3 (±)
2.000 µm 2.008 µm 2.009 µm 2.010 µm 2.009 µm 2.010 µm 0.050 µm

Section 2 - Perpendicularity Test


Standard Measurement Value Uncertainty
Test Point Average Correction
Values 1 2 3 (±)
1.500 µm 1.500 µm 1.490 µm 1.493 µm 1.494 µm 1.492 µm -0.006 µm 0.050 µm

Section 3 - Flatness Test


Standard Measurement Value Uncertainty
Test Point Average Correction
Values 1 2 3 (±)
2.000 µm 2.000 µm 2.010 µm 2.012 µm 2.013 µm 2.012 µm 0.013 µm 0.050 µm

The user shall determine the suitability of the equipment for it's intended use. The calibration status is defined as Report of Value.
The expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2,
which for a normal distribution corresponds to a coverage probability of approximately 95%, unless otherwise stated.

- End of Calibration Report -

Confidential: Controlled Document. Page 1 of 1


Cannot be reproduced without the approval of the Quality Department. TQDG-FM-002 rev04