UNIQUE CALIBRATION AND CONSULTANT SERVICES

MEASURAND = Dial Pressure Gauge


MAKE = --------
MODEL = --------
LABORATORY I.D. = QA/PG/09
SR. NO. = --------
RANGE OF CALIBRATION = 0 TO 150 bar

LEAST COUNT = 1 bar

DATE = 10/11/2023

SOURCES OF UNCERTAINTY
A: Type A Uncertainty UA1 Uncertainty due to repeatability

B: Type-B Uncertainty UB1 Unertainty due to applied pressure (as per calibration certificate of Dig. P. Gauge)
UB2 Uncertainty due to the resolution of Pressure Gauge (DUC)
UB3 Uncertainty due to Zero Error
UB4 Uncertainty due to hystresis
UB5 Uncertainty due to Maximum Error in Calibration Certificate

UA1 UNCERTAINTY DUE TO REPEATABILITY OF MEASUREMENT: Uncertainty Evaluation


TYPE A TYPE B
Referenc Value (bar) UE = UC * k,
1 2 3 Standard Where k=2
DUC Standard Uncertainty (± bar) UE= ± ( % of
Average UB1 UB2 UB3 UB4 UB5 UC Veffective
Value (bar) UP DN UP UP DN UP UP DN UP Deviation (± bar) UA1 rdg.)
M1 M2 M3 M1 M2 M3 M1 M2 M3
0 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.000000 0.000000 0.000000 0.000000 0.0000 0.00000 0.00000 0.00000 #DIV/0! #DIV/0! #DIV/0!
20 19.91 19.90 19.89 19.88 19.89 19.88 19.89 0.011690 0.004773 0.100000 0.288675 0.0000 0.00289 0.05773503 0.31096266 144178423 0.62193 3.1%
50 49.73 49.72 49.71 49.70 49.71 49.69 49.71 0.014142 0.005774 0.100000 0.288675 0.0000 0.00385 0.08660254 0.31761845 73274928 0.63524 1.3%
100 99.51 99.50 99.48 99.47 99.48 99.46 99.48 0.018619 0.007601 0.100000 0.288675 0.0000 0.00385 0.08660254 0.31765693 24400665 0.63531 0.6%
120 119.27 119.26 119.25 119.24 119.25 119.23 119.25 0.014142 0.005774 0.100000 0.288675 0.0000 0.00385 0.08660254 0.31761845 73274928 0.63524 0.5%
150 149.09 149.08 149.07 149.06 149.07 149.05 149.07 0.014142 0.005774 0.100000 0.288675 0.0000 0.00385 0.08660254 0.31761845 73274928 0.63524 0.4%

Observed By Checked By