You are on page 1of 36

Copyright

Copyright © 2006 Huqiu Imaging Technologies (Suzhou) Co., Ltd. All rights reserved.
No copying, distribution, publication, modification, or incorporation of this document, in whole or part, is permitted
without the express written permission of Huqiu. In the event of any permitted copying, distribution, publication,
modification, or incorporation of this document, no changes in or deletion of author attribution, trademark legend, or
copying notice shall be made.
No part of this document may be reproduced, stored in a retrieval system, published, used for commercial exploitation,
or transmitted, in any form by any means, electronic, mechanical, photocopying, recording, or otherwise, without the
express written permission of Huqiu Imaging Technologies (Suzhou) Co., Ltd.
This document is also distributed from your dealer in Adobe Systems Incorporated’s PDF (Portable Document Format).
You may reproduce the document from the PDF file for internal use. Copies produced from the PDF file must be
reproduced in whole.

Trademarks
The hu.q wordmark, hu.q logo, and the names of the hu.q products and services referred to in this document are
trademarks of Huqiu Imaging Technologies (Suzhou) Co., Ltd.
Other brand or product names are the trademarks or registered trademarks of their respective owners.

Limitation of Liability
The product, software or services are being provided on an “as is” and “as available” basis. Except as may be stated
specifically in your contract, Huqiu Imaging Technologies (Suzhou) Co., Ltd, expressly disclaims all warranties of any
kind, whether express or implied, including, but not limited to, any implied warranties of merchantability, fitness for a
particular purpose and non-infringement.
You understand and agree that Huqiu Imgaing Technologies (Suzhou) Co., Ltd. shall not be liable for any direct,
indirect, incidental, special, consequential or exemplary damages, including but not limited to, damages for loss of
profits, goodwill, use, data or other intangible losses (even if Huqiu has been advised of the possibility of such
damages), resulting from: (i) the use or the inability to use the product or software; (ii) the cost of procurement of
substitute goods and services resulting from any products, goods, data, software, information or services purchased;
(iii) unauthorized access to or alteration of your products, software or data; (iv) statements or conduct of any third party;
(v) any other matter relating to the product, software, or services.
The text and drawings herein are for illustration and reference only. The specifications on which they are based are
subject to change. Huqiu may, at any time and without notice, make changes to this document. Huqiu Imaging
Technologies (Suzhou) Co., Ltd. for itself and on behalf of its subsidiaries, assumes no liability for technical or editorial
errors or omissions made herein, and shall not be liable for incidental, consequential, indirect, or special damages,
including, but not limited to, loss of use, loss or alteration of data, delays, or lost profits or savings arising from the use
of this document.

Document No.: X-ray series UserGuide (A3) Hu.q Page 1


Introduction
Congratulations and thank you for buying the HQ series Processor, the well-designed equipment. This
is an advanced, yet easy to use, product with a wide range of features, designed to make processing
of medical, print and photo paper much easier.

This series processor features Microprocessor Control and Automatic Process, Efficient and Quick
Drying, yet it is easy for maintenance: all racks can be released and easily cleaned by water. This
product also features Energy Saving with less chemicals and water consumed. It fits for the medical
organizations that process more than 10 pieces of films for each day.

You can achieve best performance with good equipment and correct operation. This Operation Manual
detailedly describes installation, calibrations, user operation and maintenance for this HQ series
Processor. The processor may efficiently avoid malfunctions if you are familiar with operation
procedures and machine’s functions.

Document No.: X-ray series UserGuide (A3) Hu.q Page 2


Contents
Product Features and Technical Data 5
Product Features…………………………………………………………………………………………………5
IEC 60601-1 Classification…………………………………………………………………………….………..5
Technical Data……………………………………………………………………………………………………6

Overview of Structures and Parts 7


HQ-350XT Structure and Parts………………………………………….……..………………….……………7
HQ-450XT Structure and Parts………………..………………………….…………………….………………9
HQ-450XM Structure and Parts……………………………………………………………………...………..12

Installation 16
Checklist…………………………………………………………...……………………………………...……..16
Before Installing…………………………..……………………………………………………………..……...16
Machine Placement…………………………………………………………………………………………………..…16
-
How to Install……………………………………………………………………………………………….……………17
-

Operation Settings 19
Touch Panel and Keys for Operation……………………………………………………..………………….19
Machine Adjustment……………………………………………………………………………………..……..22
Processing Tesst………………………………………………………………………………………………..23

Safety 25
Safety warning……………………………………………………………………………………..……………25
Caution items……………………………………………………………………………………..……………..25
Marking……………………………………………………………………………………………..……………26
EMC…………………………………………………………………………………………….………………..28

Maintenance 29

Troubleshooting 31

Limited Warranty 33
What hu.q will do…………………………………………………………………………………..……………33

Terms and Definitions 34

Document No.: X-ray series UserGuide (A3) Hu.q Page 3


HQ Film Series Processor
(HQ-350XT HQ-450XT HQ-450XM)
Operation Manual
Fifth edition (August 2006)
This manual is published by Huqiu Imaging Technologies (Suzhou) Co., Ltd, without any warranty.
Improvements and changes to this manual necessitated by typographical errors, inaccuracies of
current information, or improvements to programs and/or equipment, may be made by Huqiu Imaging
Technologies (Suzhou) Co., Ltd at any time and without notice. Such changes will, however, be
incorporated into new editions of this manual.
All rights reserved.
©Huqiu Imaging Technologies (Suzhou) Co., Ltd, 2006

European Union standard


The CE symbol on our products signifies that they have been certified according to the Medical
Devices Directive 93/42/EEC.

Document No.: X-ray series UserGuide (A3) Hu.q Page 4


Product Features and Technical Data

Product Features
HQ-350XT/450XT/450XM (in this document below it is defined as “Product” or “machine”) is new
model of film processor. Developed by our experience of more than 20 years in the imaging industry, it
is designed to process X-ray, CT, MRI or other medical film with high performance.
To guarantee the best performance and reliability, the machine is made by ABS (450XM is made by
PVC) plastic. Internal parts of the machine are all made by anti-corrosion materials. This Product
utilizes micro chip for reliable control and operation. Four processing procedures of Develop, Fix,
Wash, Dry can be realized in one operation cycle. Furthermore, this Product features some automatic
functions:

Using machine’s touch panel, you can set parameters for a specific function or multiple functions when
you use different chemicals and films. To simplify operation, one heating system which controls
developer and fixer’s temperature, is used in 350XT.Two ways for replenishing chemicals, machine
automatically replenish or you manually replenish. During normal operation, processor will replenish all
chemicals automatically after processing every 3pcs of film. When chemical solution level is low or
chemical concentration is less than normal, you can start replenishment manually.

Thanks to frequency modulation control technology, the main motor runs silently and performs
accurate operation. In some abnormal cases, i.e. chemical level low, high solution level in washing
tank, 350XT will alarm to remind you to check the errors.

Once a film is fed into the entrance of machine, the system of transport, wash and dry will start to
operate. 350XT has an energy saving system, which enables itself (i.e. transport system, washing
system, drying system) automatically stop operating when no film fed in.

The transport system will not start automatically unless chemical temperature reaches pre-set value,

even when you feed a film. In this case, you can start transport when you press “》” button on the touch

panel. However, unless the film comes out of the processor, you can not stop processor’s transport

system by pressing “》” button again.

IEC 60601-1 Classification


Type of protection against electric shock: Class I equipment
Degree of protection against electric shock: No applied parts
Possible interference with other equipment:
IEC 60601-1-2 Group 1 Class A
Mode of operation: Continuous mode

Document No.: X-ray series UserGuide (A3) Hu.q Page 5


Technical Data
HQ-350XT/450XT/450XM Technical Data
Technical HQ-350XT HQ-450XT HQ-450XM
Items
Max. film width 365mm (14 inch) 435mm (17 inch)
Mini. film width 76mm (3 inch)
Mini. film length 152mm (6 inch)
Dev. Time Continuous adjusting: 25 second~45 second Continuous adjusting:

17 second~80 second

Processing 556 mm/min (when 849 mm/min (when 1375 mm/min (when
speed Dev. Time=25S) Dev. Time=25S) Dev.Time=20S)
Adjustable 20℃~40℃ 10℃~40℃
Dev./Fix temp.
range
Adjustable dryer 40℃~65℃ 40℃~65℃
temp. range
Adjustable 20ml~200ml∕film,or continuous replenish during 50ml-300ml∕0.5m2,or

range of repl. processing film. continuous replenish

during processing film.

Chemicals Automatic, manual, or continuous replenish during processing.


replenish mode
For automatic mode:
HQ-350XT/450XT, replenish one time when each X-ray film is processed
HQ-450XM, replenish one time when 0.5m2 area of films is processed
Tank volume (L) Dev.: 5.5L, Fix: 5.5L Dev.: 10.5L, Fix:
10.5L
Dimension 865mm × 585mm × 970mm × 585mm 1200mm × 660mm
520mm × 510mm × 1150mm
(L × W × H)
Net Weight 57Kg±1Kg (added basic frame ) 89Kg±1Kg (added
basic-frame)
Power Rated Voltage: a.c. Rated Voltage: a.c. Rated Voltage: a.c.
conditions 220-240V, 50/60Hz; 220-240V, 220-240V, 50/60Hz;
50/60Hz;
Rated Current: 12A; Rated Current: 15A;
Rated Power: Rated Current: Rated Power:
2.64KW 12A; 3.3KW
Rated Power:
2.64KW
Conditions for Ambient temperature: -40℃~+70℃
transport and

Document No.: X-ray series UserGuide (A3) Hu.q Page 6


storage
Relative humidity: 10% 100%

Ambient air pressure: 500hPa 1060hPa


 All specifications are subject to change without prior notice.

Document No.: X-ray series UserGuide (A3) Hu.q Page 7


Overview of Structure and Parts

HQ-350XT Structure and Parts


The figures below are for machine’s general view. The main spare parts indicated by numbers are
listed as critical parts that may be necessary for replacement after a long period of usage.

Overview for front-right side of HQ-350XT

Figure 01: HQ-350XT Front Overview


This picture corresponds to the table listed as follows:

No. Part Number Description


1 J02102 Rack Dev.
2 J02103 Rack Fix
3 A03105 Interior cover
4 C03015 Repl. elbow/Dev.
5 J02101 Rack dryer
6 C03015 Repl. elbow/fix
7 A03104 Cover/shaft
8 C03105 Shaft/transport
9 C03106 Holder/shaft
10 C03109 Latch for rack
11 C03033 Worm
12 J02104 Heater/dryer

Document No.: X-ray series UserGuide (A3) Hu.q Page 8


13 D18010 Blower/dryer
14 D19001 Cir. Pump/dev
15 D19001 Cir. Pump/fix
16 C03116 (Optional)
17 D20012 Main motor
18 C03116 (optional)
19 A03111 Keypad
20 A03102 Machine body
21 A03103 Cover/feeding

Overview for rear-left side of HQ-350XT

Figure 02: HQ-350XT Rear Overview


This picture corresponds to the table listed as follows:

Document No.: X-ray series UserGuide (A3) Hu.q Page 9


No. Part Number Description
22 D27005 Solenoid/cooling water
23 E11010 Drainage valve/water
24 E11010 Drainage valve/Fix
25 E11010 Drainage valve/Dev
26 D27005 Solenoid/water
27 A03101 Main cover
28 A03110 Holder\switch
29 D19004 Repl. Pump
30 A03106 Frame/stand
31 A03107 F. plate/stand
32 F07013 Overflow/dev.
33 C03104 Heat exchanger
34 F07015 Overflow/fix
35 C03113 Standing foot
36 A03108 Reinforcement
37 A03109 R. plate/stand
38 C03070 Jack bolt
39 F08002 Overflow/water

HQ-450XT Structure and Parts


The figures below are for machine’s general view. The main spare parts indicated by numbers are
listed as critical parts that may be necessary for replacement after a long period of usage.

Document No.: X-ray series UserGuide (A3) Hu.q Page 10


Overview for front-right side of HQ-450XT

Figure 03: HQ-450XT Front Overview


This picture corresponds to the table listed as follows:

No. Part Number Description


1 D29026 Keypad
2 J02122 Rack Dev.
3 J02123 Rack Fix
4 J02124 Rack Wash
5 C03015 Repl. elbow/Dev.
6 C03151 Shaft/transport
7 C03015 Repl. elbow/Fix
8 C03033 Worm
9 J02121 Rack Dryer
10 C03133 Exhaust/upper
11 C03131 Sync strap wheel (big)
12 A04111 Right side cover
13 A04102 Stand
14 E09021 Sync. Strap
15 C03132 Sync strap wheel (small)
16 D20012 Main motor
17 F07010 Inlet tube/Fix
18 F07011 Inlet tube/water
19 F07008 Inlet tube/Dev.

Document No.: X-ray series UserGuide (A3) Hu.q Page 11


20 C03135 Retainer ring
21 C03104 Heat exchanger
22 D30041 PCB Box
23 A04115 PCB cover
24 A04105 Feed tray

Overview for rear-left side of HQ-450XT

Figure 04: HQ-450XT Rear Overview


This picture corresponds to the table listed as follows:

No. Part Number Description

Document No.: X-ray series UserGuide (A3) Hu.q Page 12


25 E11010 Drainage valve/wash
26 E11010 Drainage valve/fix
27 D18015 Blower/dryer
28 D27005 Solenoid/cooling water
29 E11010 Drainage, valve/Dev.
30 D27005 Solenoid/wash water
31 A04103 Maincover
32 A04104 Cover/feeding
33 D16003 Switch
34 C03113 Standing foot
35 A04112 Left side cover
36 A04101 Case for tanks
37 F08002 Overflow/wash
38 D19001 Cir. Pump/Dev.
39 D19001 Cir. Pump/Fix
40 A04113 Bottom cover
41 F07013 Overflow/Dev.
42 C03116 Container/Dev.
43 D19004 Repl. pump
44 F07015 Overflow/Fix
45 C03116 Container/Fix
46 C03134 Exhaust/lower
47 A04114 Back cover

HQ-450XM Structure and Parts


The figures below are for machine’s general view. The main spare parts indicated by numbers are
listed as critical parts that may be necessary for replacement after a long period of usage.

Document No.: X-ray series UserGuide (A3) Hu.q Page 13


Overview for front-right side of HQ-450XM

Figure 05: HQ-450XM Front Overview


This picture corresponds to the table listed as follows:

No. Part Number Description


1 D29026 Keypad
2 J02122 Rack Dev.
3 J02123 Rack Fix
4 C03015 Repl. elbow/Dev.
5 A04107 Interior cover
6 C03015 Repl. elbow/Fix
7 C03109 Latch bracket
8 C03108 Latch for rack
9 C03151 Shaft/transport
10 J02121 Rack Dryer
11 J02124 Heater/Dryer
12 D18015 Blower/Dryer
13 C03033 Worm
14 C03020 Holder/shaft
15 D19001 Cir. Pump/Fix
16 Washer
17 C03022 Retainer ring
18 D19001 Cir. Pump/Dev.
19 C03152 Synch wheel
20 D20015 Main motor

Document No.: X-ray series UserGuide (A3) Hu.q Page 14


21 Main PCB cover

Overview for rear-left side of HQ-450XM

Figure 06: HQ-450XM Rear Overview


This picture corresponds to the table listed as follows:

Document No.: X-ray series UserGuide (A3) Hu.q Page 15


No. Part Number Description
22 D27005 Solenoid/cooling water
23 E11010 Drainage valve/water
24 E11010 Drainage valve/Fix
25 E11010 Drainage valve/Dev
26 D27005 Solenoid/wash water
27 A04103 Main cover
28 A04104 Cover/feeding
29 D16003 Switch
30 A04015 Feed tray
31 A04106 PCB Cover
32 A04102 Stand
33 D19004 Repl. pump
34 F07013 Overflow/Dev.
35 C03104 Heat exchange
36 F07015 Overflow/Fix
37 C03116 Container/Dev.
38 C03116 Container/Fix
39 A04101 Case for tanks
40 F08002 Overflow/wash water

Document No.: X-ray series UserGuide (A3) Hu.q Page 16


Installation

Checklist
Be sure that all the following equipment and accessories have been included with your machine. If
anything is missing, contact your dealer.
 Main HQ-350XT/450XT/450XM body.
 3 rack assemblies.
 2 waste containers.
 1 water inlet hose.
 1 water outlet hose.
 1 bag of spare parts.
 Machine stand with fixings (excluding HQ-450XM)
 Documents (Operation Manual and Warranty Card).

Before Installing
This Product is a fully automatic processing equipment that integrates mechanism and electric. To
enjoy its optimum performance, the machine should be installed with electrical technician and service
engineer being onsite. Also, ambient conditions should meet followings:
 Power supply voltage should only range from AC 220V to AC 240V. Maximum output of current
should be no less than 15A. Knife switch for power supply, safe switch of socket, receptacle and
safe grounding, which are specially used for this machine, should be supplied.

 The working room must be clean and well-ventilated, with ambient temperature between 15 ~

25℃ and relative humidity less than 85% (25℃).


 At least one water tap should be specially supplied as washing water, also including relevant
water pressure regulator, water pressure meter and filter. Water pressure should be kept between

1kg~2kg.

 A good draining system is required for draining the water’s overflow from the washing tank. Water
in washing tank runs in a cycle with water supply and water overflow happening simultaneously.
Water overflows much more than the water that is supplied
 The floor for machine’s placement should be flat and reliable in case of machine shake, chemicals
contamination, and thus affecting processing quality.
 A sink of at least 450mm x 700mm acreage should be available in the working room or near the
machine, used as washing or soaking rack assemblies periodically.

Note: Disposal of the waste chemicals must comply with your local regulations.

Document No.: X-ray series UserGuide (A3) Hu.q Page 17


Machine Placement
The machine can be placed in two modes.
 Mode 1: Film collecting in a light room, see figure 07. This placing mode divides film collecting
side from film processing side. You can take any processed film away before the processing of
other films is finished. Meanwhile, this mode enables hot air coming from dryer to exhaust through
light room’s window, and keeps air in working room fresh.
 Mode 2: The whole machine is installed in a dark room. Locating the machine in this mode is
more convenient but you can not operate film processing and film collecting respectively. Hot air
of machine can not exhaust in time. To prevent this and to keep air inside the room fresh, exhaust
fan should be installed in the working room

Figure 07: Light room placement mode

How to Install
 After unpacking the machine’s crate, remove machine body and accessories. Assemble the stand
of machine according to Overview of Structure and Parts on page 7. Place the stand to the site
where the whole machine will be placed. Level the stand so that it is stable.
 Remove the top cover of machine, carefully check:
 Any crack of machine chemical tanks due to transportation of machine.
 Loosing or missing contact of pumps’ hose connector.

After solving the problems that may occur, place machine body on the stand. Level the machine
by adjusting the four leveling feet that are located at the bottom side of the stand.
 Fit the plumbing (water inlet hose, drain hose, chemicals waste container, replenish hoses and
chemicals replenish container) properly according to the instruction labels that are marked on
them.

Document No.: X-ray series UserGuide (A3) Hu.q Page 18


Note: It is recommended that a water pressure regulator is installed on the hose of water inlet.

Water pressure should be controlled within the range of 1kg ~2kg. Pay attention that do NOT put

the wrong hose to the replenishing container. Check that all the drainage valves are closed. Close
them if not, and put the solution overflow hoses in the corresponding waste solution containers.
Connect the solenoid valve for washing to the city water tap by the hose and adaptor supplied. (if
for some reason you will not use the hose and adaptor we supplied, please make sure you’re
using reliable material and safety way.

 Check again all the drain valves are closed. Fill all 3 tanks with clean water until water overflow
from the overflow outlet. Check the leveling of the processor again by the water level. If any water
splash to the section contains electrical components, dry it with a hair dryer.
 Connect the power supply, turn on the power switch, then the circulation pumps will work, and the
water will be circulated. If everything is normal, you may access to the test now.

Document No.: X-ray series UserGuide (A3) Hu.q Page 19


Settings

Touch Panel and Keys for Operation


Your HQ-350XT/450XT/450XM has an operator panel which enables you easily control or know the
status of Chemical/Dryer Temperature, Replenish Rate, and Error Messages when machine has a
problem.
Touch Panel of HQ-350XT Overview

Figure 08: HQ-350XT Touch Panel

Touch Panel of HQ-450XT/450XM Overview

Figure 09: HQ-450XT/450XM Touch Panel

Document No.: X-ray series UserGuide (A3) Hu.q Page 20


Character Meanings for HQ-350XT/450XT
CD-stands for develop temperature
DY-stands for dryer temperature
SP-stands for duration of dev. process
N1-stands for replenish rate of chemicals
GOOD-means that data has been finished in settings and that you may switch on the machine
Self-diagnosis Information:
EE-1 means that saving for the set data malfunctions
EB- means that chemical temperature sensor fails or the sensor is not well connected.
EY- means that dryer temperature sensor fails or the sensor is not well connected.
E7-1 stands for the alarm of chemical overheated
E8-1.2 stands for solution level LOW of Develop or Fix tank.
E8-3 stands for water level HIGH of Wash tank.
Note:
1) N is an ordinal number.
2) Overheat means the temperature is higher than the preset point by 2℃.

Character Meanings for HQ-450XM


CD-stands for the temperature of Develop or Fix
DY-stands for Dryer temperature
SP-stands for duration of Dev. process
N1-stands for replenish rate of Dev. chemicals
N2-stands for replenish rate of Fix chemicals
PB-stands for modifications of Dev. and Fix temperature
PY-stands for modifications of Dryer temperature
PG-stands for programm code
GOOD-means that data has been finished in settings and that you may switch on the machine
GOODP-means that data has been finished in modifications, or that a channel has been selected.
Self-diagnosis Information:
E5-1 means that Dev. temperature sensor fails
E5-2 means that Dryer temperature sensor fails
E7-1 stands for overheated alarm
E7-2 means that CPU (central process unit) fails
E8-1 stands for solution level LOW of Develop tank
E8-2 stands for solution level LOW of Fix tank
E8-3 stands for solution level HIGH of Wash tank
Note:
1) N is an ordinal number.
2) Overheat means the temperature is higher than the preset point by 2℃.

Keys and their functions


DRIVEDual Function key: in standby condition, press the key to turn on / off the drive motor; in setting
>> condition, press the key to move the digital.

Document No.: X-ray series UserGuide (A3) Hu.q Page 21


REPL Dual Function key: in the standby condition, press the key to replenish manually; in setting
+1 condition, press the key to modify value.

SET Dual Function key: in operational condition, press the key to display the temperature one by
one; press both this key and YES key to change running status.
YES Dual Function key: in operational condition, press the key to change brightness; press both the
key and SET key to change running status; in setting condition, press the key to confirm.

Key’s Operation
A. Enter working parameter setting status
Example:
No. Description Key Operation Displayed
information
1 Turn on the processor cd28.0

2 Change operational status to “Good” Press both SET and YES good
status one time

3 Enter setting status and then the preset Press SET one time db20.0
datta is displayed

4 To modify the figure to “3” in decade Press db30.0


REPL
digit one
+1
time

5 ToDRIVE
modify the figure to “2” in unit digit Press db30.0
>> one
time

REPL Press db32.0


+1 two
times

6 ToDRIVE
modify the figure to “5” in decimal Press db32.0
digit>>

REPL Press db32.5


+1 five times

Document No.: X-ray series UserGuide (A3) Hu.q Page 22


7 Save the modified develop Press SET repeat 4-7 steps dy20.0
temperature value , then set the
dryer temperature

8 By repeating step 4 to step7, you can Press SET one time good
set dryer temperature, speed and
replenishment

9 Complete setting and then enter in Press YES one time


operation mode

DRIVE
B. Cancel self-diagnosis display
>>
Once the error display of self-diagnosis is appeared, press both key and YES key, then the error
display is disappeared, but the trouble is not obviated.

If the temperature doesn’t reach the setting point, the drive system will not start automatically. If you
want to feed film at this condition, you must press DRIVE key.
>>

Machine Adjustment
Turn on the processor
Turn on the processor, select correct parameters for the using chemistry, like developing time,
temperature and etc.

Check according to LED indicator


Pay attention to the LED on the Circulation indicator and chemical heating indicator, if the LEDs
are on, check inside of the tank to see if the water is really being circulated and if the
temperature of the water is getting up.

Note: Circulation is very important to the heating of chemicals, if the chemical is not well circulated,
then the temperature can not be raised rapidly and evenly, and this may effect the result of the film
processing as well as damage the chemical heater. If the circulation system is not good, first check if
the LED and the circulation indicator is on, then check if the voltage to the circulation pump is
normal(220V). If the pump is working properly, then there might be some air in the pump which block
the water flow. In this case please switch off the processor and drain some water from the tank. Fill up
the tank and switch on the processor again. Do this two or three times if necessary until circulation is
in good condition.

Feeding the film

Document No.: X-ray series UserGuide (A3) Hu.q Page 23


If circulation is good, close the machine cover, waiting for the chemicals temperature reaching
their setting point. A beeper will remind you this and you may feed with a film now. Choose
several pieces of used film, and feed them into the processor one by one. When you feeding the
film, after the leading edge of the film was caught by the rollers at the entrance of the developing
rack, you may leave it and the film will entering the tank slowing. At the time, the dryer and the
drying fan will start working. You may open the machine cover to check the film driving
condition, the washing water condition as well as the replenishing.

Balance the water


If the water tap for washing water is opened, you can find the city water goes into the washer
tank then overflow from the overflow outlet. Please adjust the water tap or water pressure
regulator, until they reach dynamic balance. If the incoming water flow is too large, then water
can not drain from the overflow outlet on time, and may overflow to the fixer tank or even
developer tank. If the incoming water flow is too small, then the film can not be cleaned enough,
and may result to a bad processing result. Make a mark on the tap when you find the ideal water
flow.

Checking the speed


Checking the developing time with a stopwatch. You should start when the leading edge of the
film enters the developer chemical and stop when it enters the fixer chemical. Check all the
three speeds and too see if they’re accurate.

Checking the temperature


If the chemical temperature reaches the setting point, measure it with a thermometer and check
if it is in corresponding to the figure selected.

Clean the processor


If all above steps are completed without any problem, you may clean the processor and fill with
chemicals. Drain all water in the tanks, take out the rack assemblies from each tank, clean the
tank and the rack with a dry and clean towel. If the tank or rack assemblies are not clean
enough, rinsing with clean water again.

Ready for processing


Close the draining valves re-load the rack assemblies and lock them. You’re ready for fill the
processor with chemicals and start formal processing now.

Processing Test
Please follow below steps when you do processing test:
1) prepare all chemicals(both working and replenishing solutions) according to the instruction of
the manufacturer. Then fill the fixer tank first. Do not spill over any fixer chemicals to other tank
or other parts of the processor. Wipe them off immediately if any.

2) then fill up the developer tank until to the overflow outlet.

Document No.: X-ray series UserGuide (A3) Hu.q Page 24


3) fill the washing tank with clean water, then open the water tape for washing water. (us the
mark you made on the tap as a reference.)

4) Checking all the setting on the control panel to make sure if they meet the requirement of the
using chemistry. Then switch on the processor.

5) Check the circulation of the chemical. If it is in order, close the machine cover, wait until the
chemical temperature reaches the setting point.

6) When the chemical temperature reaches the setting point, a buzzer will raise and you may
feed film into the processor now. It was recommended to have a glove when you handling the
film.

7) After the film come out from the dryer, check the density, contrast, and the drying condition of
the film immediately, adjust the parameters like the temperature and time of developer,
temperature of the dryer as well as the amount of replenishment until the result satisfy you.

Document No.: X-ray series UserGuide (A3) Hu.q Page 25


Safety

Safety warning
1. The machine must be earthed properly because of high humidity in the darkroom.
2. Service on the electrical/electronic components of the processor must be done by a qualified
technician.
3. After machine’s daily operation, be sure to switch off the power supply for your machine.
4. After machine’s daily operation, be sure to close the water supply in case of high water pressure
which will damage machine’s plastic hose and cause water to flow on the ground.
5. Avoid chemical solutions to spatter your eyes or skins. Using fresh water to wash your eyes or
skins if spattering happens.
Caution: We do not bear the responsibility of accident due to machine operator’s absence and
machine’s running with power and water is being supplied.

Caution items
1. After switching on your machine every day, open the cover and check the solution circulation. Do
not proceed unless the circulation pumps are working normally.
2. Please pay usually attention to the water flow of washing water. To the areas that water pressure
are not stable, a flow meter is recommended. Please do not use still water when processing and
drain out the water in the washing tank after your daily operation. (some microbe in still water may
cause alga which will cause problem for proper film transportation).
3. Do not leave the processor un-attended for long time while it is running. If there is any abnormal
noise or smell, stop the machine and have a check.
4. Do not leave the processor along with solutions in for a long period. If you would not use the
processor for more than 7 days, drain out the solutions and wash the solution tanks as well as the
rack assemblies.
5. After switching off the processor when you finish your daily work, please slide the machine cover
a little bit to release the damp inside the processor.
6. It is a better way to switch off your machine during its standby mode, that is dryer, transport roller
and blowers all stop working. This may help the machine to extend its life.
7. To assure high performance of the processor, please maintenance the processor as per following
schedule.
8. Disposal of the waste chemicals and other disposal of waste material must comply with your local
regulations. The waste of equipment must not be disposed as unsorted municipal waste and must
be collected separately.
9. There are one 10A fuse and one 15A fuse inside the machine’s main circuit.

Model R314010 R314015


Specifications 220V/10A 220V/15A

Note: When you replace the fuse, you may lift the dark box 21 up according to Overview of
Structure and Parts, remove the stainless steel entrance plate, open the shelter cover for main

Document No.: X-ray series UserGuide (A3) Hu.q Page 26


board and take the socket for fuse out for fuse placement.

Marking

Electric shock: This mark is sticked on the Electric Box, and it indicates dangerous voltage that is
harmful to human body. Only qualified electric technician may access this area.

N
Neutral wire

L
Line wire

Protective grounding: The part with this mark is grounded.

This mark is sticked on the board cover of transport shaft, and indicates the part has transport
function. To avoid personal injure, you should keep fingers or your body away from the part.

Document No.: X-ray series UserGuide (A3) Hu.q Page 27


This mark is sticked on the outer cover of dryer section, and indicates that:
1. There is 220 Vac voltage inside. Only qualified technician may disassemble the part.
2. There is hot part inside. To avoid heat, do NOT touch it while machine is running.

This mark is sticked on the dark box, and indicates that you should open the cover of dark box when
your machine does not operate.

This mark is sticked on the side surface of rack, and it indicates that you should keep your hands away
from the gears while machine is running.

This mark is sticked on the film entrance of dark box to warn the cutting injure.

Document No.: X-ray series UserGuide (A3) Hu.q Page 28


Caution! Please refer to your operation manual, and safely operate the machine according to the
operation manual.

EMC
EMC test report based on IEC 60601-1-2 2001 (TUV report NO: 15013333 001) Group 1, Class A. Pay
attention to interference between this equipment and other device.

Document No.: X-ray series UserGuide (A3) Hu.q Page 29


Maintenance
Since HQ-350XT/450XT/450XM film processor integrates mechanic and electric, and its operating
environment is not machine-friendly, the machine should be well maintained and be cared by special
person.

Maintenance Schedule and Maintenance Method


Maintenanc Maintenance Schedule Items Maintenance Method
e

Frequency
Everyday 1. Switch off the machine, open 1. Use wash bottle with full level water
machine’s top cover, use wash to spray water to the roller gears
bottle to rinse rollers of Dev. and Fix beside the chemical solution, so
rack with water, rinse the gears that the chemicals left on the gears
beside chemical solutions to avoid can be wiped away in case of
generating crystals on the roller generating crystals.
gears.
2. Open the drain valve for the Wash
2. Drain the water in the Wash tank tank to drain the water out in case
after machine’s daily operation. of generating animalcule in the
3. Use damp towel to wipe machine’s Wash tank.
outer covers. 3. Clean the film entrance plate in the
4. Check that the container for dark box, machine outer covers
replenishing chemicals is empty or including the top cover one time.
not. 4. If no replenish chemicals in the
tank, add chemicals to full level.
Every 7 1. Take out Dev., Fix, and Wash rack 1. Release the racks of Dev. Fix and
days assemblies, rinsing and clean them Wash from the respective tank. Use
in the sink. fresh water to rinse them and use
brush to wipe off the crystals on the
2. Check that the gears, gear clips and roller gears. After cleaning, use a
screws on the rack assembly are dry towel to wipe them.
intact.
3. Clean the stand, replenish container 2. Check that the gears, gear clips
and waste tank one time. and screws are missing or not, and
4. Clean the main driving shaft one check that they are intact.
time. 3. Use damp cloth to wipe dust off
from the stand, replenish container
and waste tank.
4. Use cloth to clean the main driving
shaft to avoid generating crystals

Document No.: X-ray series UserGuide (A3) Hu.q Page 30


and rust..
Every 15 1. Rinse each chemical tank one time 1. Open the respective valve to drain
days respectively. the chemicals and water from the
solution tanks. Fill fresh water to
2. Oil the bearing and chains on the every tank and use brush to clean
two ends of drive shaft. them. After cleaning, close each
valve.

2. Add 2 or 3 drops of oil to the


bearing chains that is on the two
ends of main transport shaft.
Every 30 1. Check the connection of chains, 1. Remove the right and left side
days circulation pump, replenish pump cover of machine to clean the
and solenoid valve. chains one time and you may oil
the chains. Check that circulation
2. Clean the circulation pump and one- pump, replenish pump and solenoid
way valve one time. valve are well connected and check
that there’s no leaking among the
parts.

2. Take out the inlet and outlet one-


way valves from the replenish
pump. Rinse them in the fresh
water and make sure that the
rubber material of them are not
damaged or malfunction. If
everything is ok, return them back
the place. Take care that do NOT
assemble with a wrong direction.
Every 90 1. Check the one-way valve or replace 1. Take out the one-way valve from
days it if necessary. the replenish pump. After rinsing it
and if you find it aged, replace it
with a new one.

Document No.: X-ray series UserGuide (A3) Hu.q Page 31


Troubleshooting

This chapter consists of a troubleshooting list of a selection of suggestions for solving possible
problems with your machine. Please note that this troubleshooting list is for reference of qualified
service engineer. The hu.q or dealer authorized by hu.q does not bear the responsibility for the
following accidents, including but not limited to machine malfunctions, personal injuries, fires, etc., if
unauthorized person repair your machine.

Troubleshooting List for HQ-350XT/450XT/450XM


TROUBLES PROBABLE CAUSE COUNTERMEASURE
Chemical 1. No LED indication on the main PCB

can not be 1-1 the temperature displayed is higher 1-1 check the data setup for temperature
heated than setting point 1-2 repair or replace the temperature
1-2 the temperature probe defective probe
1-3 the circuit for temperature control 1-3 check and repair the circuit
defective
2. There is heating indication on the main
PCB 2-1 check and replace the fuse
2-1 the fuse for heating blown out 2-2 replace the heater
2-2 the heater damaged 2-3 replace the relay
2-3 the relay for heating defective
Temperatur 1. Solution circulation is not in order 1. Check the circulation and solution filter
e not stable
2. Temperature probe poor connection 2. Check probe and its wiring
3. The circuit for temperature control 3. Repair the circuit
defective
Solution 1. The motor for circulation pump is
does not running
1-1 remove the foreign matter in the pump
circulate 1-1 the impeller of the pump blocked 1.2 clean or replace the filter
1-2 the solution filter blocked
2. The circulation pump is not running 2-1 replace the fuse
2-1 the fuse for the pump fails 2-2 replace the pump
2-2 the pump damaged
Solution can 1. There’s LED indication for replenishing
on control panel while the pump works
not be

Document No.: X-ray series UserGuide (A3) Hu.q Page 32


replenished 1-1 the one-way valve in the repl. pump 1.1 clean the valve
blocked
1-2 the pump do not seal 1.2 fasten any connection
1-3 the bellow leaks 1.3 replace the bellow
2. There’s LED indication for replenishing
on control panel while the pump does not
work
2-1 the relay for the repl. pump damaged 2-1 replace the relay
2-2 the fuse for pump blown out 2-2 replace the fuse
2-3 the switch on pump blown out 2-3 switch on or replace the motor
2-4 the motor for pump block or damaged 2-4 repair or replace the motor
3. No LED indication on control panel
3-1 the relay for pump defective 3-1 replace the relay
Driving 1. The key lamp for driving is ON

motor does 1-1 the fuse for driving motor blown out 1.1 replace the fuse
not work 1-2 the motor or gear box damaged 1.2 replace the motor or gear box
1-3 the circuit for motor control is NOT on
2. The key lamp for driving is NOT on 1.3 check and repair the circuit
2-1 DRIVE key was not pressed
2-2 no response when DRIVE key was 2-1 press the DRIVE key
pressed 2-2 re-start the machine
2-3 check the circuit
Drying 1. There’s LED indication on main PCB
temperature
1-1 the fuse for dryer blown out 1.1 replace fuse
can not 1-2 the heating element aged or damaged 1.2 replace the heating element
reach the 1-3 the SSR (relay) for dryer damaged 1.3 replace the relay
setting point 1-4 the blower for dryer damaged 1.4 replace the blower
2. There’s not LED indication on main
PCB
2-1 the processor is at stand-by status 2-1 feed with paper, start
2-2 the setup temperature for drying too 2-2 check and re-setup the data
low
Drying 1. The temperature probe short-circuited 1. replace the probe
temperature 2. check the circuit
2. The circuit for drying temperature
can not be control
controlled defective
Blower for 1. the fuse for blower blown out 1. replace the fuse
2. the blower damaged 2. replace the blower
dryer does

Document No.: X-ray series UserGuide (A3) Hu.q Page 33


not work
Print jams in 1. some retaining rings for gear come 1. check and put on the retaining rings
off, so the relative rollers do not turn

Rack 2. film curves too much 2. flat the film


assembly
Print lost in 1. the film too small 1. use size of film required
2. microbe gets in washing tank and the 2. clean the rack assembly, use fresh
tank rollers getting too smooth water

Document No.: X-ray series UserGuide (A3) Hu.q Page 34


Limited Warranty
Huqiu Imaging Technologies (Suzhou) Co., Ltd (hu.q), provides this Limited Warranty for your machine
(HQ-350XT/450XT/450XM) and original accessory delivered with your machine (hereinafter referred to
as “Product”). Should your Product need warranty service, please ask your dealer from whom it was
purchased, to repair it or to replace it if necessary, or visit www.huq.cn to obtain further information.

What hu.q will do


If, during the warranty period, this Product fails to operate under normal use and service, due to
defects in design, materials or workmanship, hu.q authorized distributors or service partners, in the
country or the region where you purchased the Product, will, at their option, either repair or replace the
Product in accordance with the terms and conditions stipulated herein.
Hu.q and its service partners reserve the right to charge a handling fee if a returned Product is found
not to be under warranty according to the conditions below.

Conditions
1. The warranty is valid only if the original proof of purchase issued to the original purchaser by a
dealer authorized by hu.q for this Product, specifying the date of purchase and machine’s serial
number, is presented with the Product to be repaired or replaced. The hu.q reserves the right to
refuse warranty service if this information has been removed or changed after the original
purchase of the Product from the dealer.
2. This warranty does not cover any failure of the Product due to normal wear and tear, or due to
misuse, including but not limited to use in other than the normal and customary manner, in
accordance with the hu.q instructions for use and maintenance of the Product. Nor does this
warranty cover any failure of the Product due to accident, software or hardware modification or
adjustment, acts of God or damage resulting from liquid.
3. This warranty does not cover Product failures caused by installations, modifications, or repair or
opening of the Product performed by a non-hu.q authorized dealer or person.
4. This warranty does not cover Product failures which have been caused by use of accessories or
other peripheral devices which are not hu.q branded original accessories intended for use with the
Product.
5. Tampering with any of the seals on the Product will void the warranty.
6. There are no express warranties, whether written or oral, other than this printed limited warranty.
All implied warranties, including without limitation the implied warranties of merchantability or
fitness for a particular purpose, are limited to the duration of this limited warranty. In no event shall
hu.q or its authorized dealer be liable for incidental or consequential damages of any nature to the
full extent those damage can be disclaimed by law.
Some countries or regions do not allow the exclusion or limitation of incidental or consequential
damages or limitation of the duration of implied warranties, so the preceding limitations or exclusions
may not apply to you.
The warranty provided does not affect the consumer’s statutory rights under applicable legislation in
force, not the consumer’s rights against the dealer arising from their sales/purchase contract.

Document No.: X-ray series UserGuide (A3) Hu.q Page 35


Terms and Definitions
Develop/Dev. refers to develop process or develop tank which uses developer to process X-ray film.
Fix/Fixer refers to fix process or fix tank which uses fixer to process X-ray film.
Wash/Washing refers to wash process or wash tank which uses water to clean medicals left on X-ray
film.
Dry/Dryer refers to dry process or dryer section which blows hot air for drying X-ray film.
Replenish/Repl. refers to a process for chemicals (developer or fixer) and water’s replenishment.
Temperature/Temp. refers to the temperature of Dev., Fix or Dryer, or ambient temperature for
machine’s operating/transport/storage.
Circulation Pump refers to pumps which allow chemicals or water in the processing tank to circulate
and so that temperature in solutions is even and water keeps fresh.

Huqiu Imaging Technologies (Suzhou) Co,. Ltd

Address: 388 Fengjiang Rd, Mudu Industry Park


Suzhou 215101, China
Tel: (00)86-512-6636-8881
Fax: (00)86-512-6636-7771

Document No.: X-ray series UserGuide (A3) Hu.q Page 36