You are on page 1of 2

Artis zee - Study Protocol

Follow-up after stent-assisted


aneurysm coiling
Interventional Neuroradiology

Supported by* Case Description


• syngo InSpace 3D
• syngo DynaCT Patient History Treatment
• syngo DynaCT 360 73-year-old female patient. The quality of the study was very good,
• syngo InSpace 3D/3D Fusion Left middle cerebral artery (MCA) aneu- with no evidence of any other abnormali-
• syngo iPilot enhanced rysm that was treated with stent-assisted ties in the circle of Willis. She tolerated
• syngo iGuide Toolbox coil embolization in 2005. Because of her this imaging study without incident and
• syngo iGuide
history of contrast allergies, we decided was discharged home. Our plan will be to
• syngo Neuro PBV IR
• syngo DynaPBV Body
to complete a syngo DynaCT run with follow up with this patient in 12 months
• syngo Embolization Guidance IV injection of contrast medium in the an- with repeated imaging study.
• syngo iFlow giography suite for the purpose of im-
General Comments
• syngo Advanced Roadmap proved visualization and monitoring the
• syngo iDentify
At our institution we are using
patient post IV injection.
syngo DynaCT runs with IV injection
Courtesy of Diagnosis routinely for patients with the following
Kiffon Keigher, APN, In 3-D rotational angiography we identi- concerns: stent patency, in-stent stenosis,
Demetrius K. Lopes, MD, fied that the left MCA aneurysm was residual filling of aneurysms s/p clipping
Rush University Medical Center, completely occluded, status post stenting and/or coiling, and vasospasm.
Chicago, USA
and coiling. We were able to obtain accu-
System & Software rate visualization of the parent vessel and
Artis zee biplane VC21 verify occlusion of this aneurysm and pa-
syngo X Workplace VB21 tency of the stent. There was no evidence
of in-stent stenosis.

*This list of applications is not


complete. Not all applications
available for all software ver-
sions.

We were able to obtain accurate


visualization of the parent vessel and
verify occlusion of this aneurysm and
patency of the stent.

Answers for life.


Follow-up after stent-assisted aneurysm coiling

Acquisition Protocol 20sDCT Head 109kV Reconstruction Protocol DynaCT Head Nat Fill HU
VOI Size: Full
Injection Protocol Slice Matrix: 512x512
Contrast Media (CM): 370 mg Iodine/ml Kernel Type: HU
Dilution (CM/Saline): No Image Characteristics: Normal
Injection Volume: 80 ml Reconstruction Mode: Nat Fill
Power Injector Used: Yes Viewing Preset: DynaCT Head
Injection Rate: 4 ml/s Secondary Recon In order to improve the visualization
of the stent, a 2nd reconstruction
Duration of Injection: 20 s
with a small size FoV was per-
X-ray Delay: 14 s formed, using the same parameter
Injection/Catheter Position: IV injection antecubital as for the initial reconstruction.
vein (18 G) VOI Size: Small
allergy prep medication
Slice Matrix: 512x512
Kernel Type: HU
Image Characteristics: Normal
Reconstruction Mode: Nat Fill
Viewing Preset: DynaCT Head

VRT (Volume Rendering Technique) of syngo DynaCT run with IV injec- Axial MPR reconstruction outlining full patency of the X stent in the
tion, visualizing the X stent. left middle cerebral artery (with secondary reconstruction).

On account of certain regional limitations of sales rights and service


availability, we cannot guarantee that all products included in this
brochure are available through the Siemens sales organization
worldwide.
All rights reserved.

Global Business Unit


Siemens AG
Medical Solutions
Angiography & Interventional X-Ray Systems
Siemensstr. 1
DE-91301 Forchheim
Germany
Phone: +49 9191 18-0
www.siemens.com/healthcare

Axial MPR reconstruction showing X stent in the left middle cerebral Global Siemens Healthcare Global Siemens Headquarters
artery. Headquarters Legal Manufacturer
Siemens AG Siemens AG
Healthcare Sector Wittelsbacherplatz 2
Henkestrasse 127 DE-80333 Muenchen
91052 Erlangen Germany
Phone: +49 9131 84-0
Germany