You are on page 1of 17

3RT03_Layout(055_072) 10/2/02 1:51 PM Page 55

C
3
H A P T E R

Electronic Radiation
Sources

George C. Wheeler, Materials and Processes


Consultants, Schenectady, New York
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 56

PART 1. Physical Principles

Radiation of an unknown type and origin target. Much of the energy of the
was discovered in the year 1895 by electrons’ motion, called kinetic energy, is
Wilhelm Roentgen. Dubbed X-rays by transformed into heat energy as the
their discoverer, they were soon found to stream strikes the target. In fact, except
be a form of electromagnetic radiation for the case of very high energy
with extremely short wavelengths. The generators, almost all of the electrons’
X-radiation is called ionizing because X-ray kinetic energy (more than 97 percent) is
absorption kicks out an electron from an converted into heat; disposal of this
atom of absorption material, creating an thermal energy is an important design
ion. The production of ionizing radiation consideration.
is well covered in the literature.1-5
Technological advances in image
processing, flash radiography, real time
radiography, microfocus and metal Bremsstrahlung
ceramic tubes have sustained interest in A small portion of the energy will also be
the electronic production of ionizing given off as packets of electromagnetic
radiation. The versatility of an electronic energy, called photons. The X-ray photons
source is difficult for an isotopic source to can have energies ranging from a few
match, even for specific applications. The thousand volts to a maximum determined
advent of solid state components has by the original kinetic energy of the
overcome many of the objections related electron and how rapidly the electron is
to size, weight and durability of electronic decelerated by interaction with a target
sources. atom nucleus. This process produces the
This chapter discusses the basic continuous portion of the X-ray spectrum
physics, construction and application of and is known by the German term
electronic sources. The emphasis will be Bremsstrahlung for braking radiation.
on industrial X-ray machines widely used Energies of the electrons (and the
in radiography and radiology. Where X-rays) are frequently given in terms of
special details are significant to particular kiloelectronvolt (keV) or megaelectronvolt
applications, these are discussed in the (MeV). The meaning of this unit becomes
appropriate chapters of this volume. clear if the charged electron is considered.
Under the influence of a voltage
difference (technically called a potential
difference), charged particles will
Conservation of Energy experience a force that causes them to
Electromagnetic theory had long accelerate. A negatively charged particle,
predicted that a charged particle such as an electron, will move from a
undergoing deceleration would emit place of low voltage (–) to a place of
radiation. This theory can be used to higher voltage (+) and increase its kinetic
qualitatively explain the continuous
portion of a typical X-ray spectrum (see
Fig. 1). Although a complete study of the
continuous spectrum requires modern FIGURE 1. Typical X-ray spectrum.
physics, an understanding of its basic
principles may be gained by considering Characteristic peaks
just one of the fundamental laws of
Relative intensity

physics, the conservation of energy. This


law states that energy can neither be
created nor destroyed, although it is
possible to change it from one form to
another.
In the case of X-rays, fast moving
charged particles, usually electrons, strike
a target material and are slowed or
stopped. In some specialized applications,
high energy electrons are accelerated 0.1 0.2 0.3 0.4
radially by a strong magnetic field,
producing X-rays without the use of a Wavelength (µm)

56 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 57

energy as it does so. Thus the unit different energy levels. If a vacancy in a
kiloelectronvolt (keV) corresponds to the particular level were created by knocking
amount of kinetic energy that an electron an electron from its orbit, an electron
would gain while moving between two from a higher energy level would, after a
points that differ in voltage by 1 kV. time, drop down to fill the void. To do
Similarly, an electron would gain 1 MeV this, it would have to lose (emit) energy.
of kinetic energy while moving between Because the energy levels of an atom are
two points that differ by 1 MV. The points well specified, the exact amount of energy
of differing voltage are called the cathode lost by an electron making such a
(–) and the anode (+) and will be discussed transition would also be discrete.
later in this section. According to the law of conservation of
energy, the energy is not really lost but
given off as electromagnetic radiation in
the form of an X-ray photon. Because
Characteristic X-Rays each atomic element has its own distinct
In addition to the bremsstrahlung there set of energy levels, the line spectrum
are several characteristic peaks in a typical produced in such a manner is
X-ray spectrum. These intensity spikes are characteristic of the particular target
caused by interaction between the material. Interest in these characteristic
impinging stream of high speed electrons X-rays lies in their application to X-ray
and the electrons that are bound tightly diffraction and other analytical
to the atomic nuclei of the target applications. By a combination of
material. If an atom is visualized as a filtration and signal processing,
planetary system, with the nucleus of characteristic X-rays can be used to
protons and neutrons at the center of the produce a nearly monoenergetic beam of
system and with the electrons moving in X-rays.
orbits around the nucleus (see Fig. 2), the
law of energy conservation can again be
applied to describe the origin of
characteristic radiation. Thermionic Emission
Modern physics predicts that these Two final concepts are of interest in this
orbital electrons near the nucleus will brief treatment of the basic physics of
have very well defined energies, with X-ray production. The first process,
electrons in different orbits having thermionic emission, provides the
electrons that are to be accelerated to
high speeds by the voltage difference
between the cathode and anode of an
FIGURE 2. Planetary model of atomic X-ray source.
structure. When metal (in this case the X-ray
tube filament) is heated to incandesence,
a small portion of the free electrons are
E
actually able to escape the surface of the
material. Without an accelerating voltage,
the electrons would form a cloudy space
charge around the filament. Under the
influence of a potential difference,
however, the electrons are quickly
accelerated toward the anode of the tube.
N
P

N P Absorption
N The second concept is X-ray absorption,
P
discussed in greater detail in the chapter
on radiation and particle physics. As
E
X-rays pass into or through a material,
they are absorbed, or attenuated, in a
manner that depends on the energy of the
X-radiation and on the density of the
material. In equation form, the intensity I
E
as a function of thickness has a standard
exponential form:

Legend (1) I = Io e – µ x
E = electron
N = neutron where I is the intensity after passing
P = proton through the material; x is the thickness of

Electronic Radiation Sources 57


3RT03_Layout(055_072) 10/2/02 1:51 PM Page 58

the absorbing material; Io is the initial


intensity; and µ is the linear absorption
coefficient (characteristic of the material
for a particular X-ray energy range). The
importance of absorption in selection of
materials for tube construction, beam
filtration and shielding are discussed
below.

58 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 59

PART 2. Basic Generator Construction

A conventional X-ray generator consists of envelope and have been replaced in most
three main components: (1) X-ray tube, industrial applications with the more
(2) high voltage source and (3) control durable metal ceramic tubes.
equipment. Although each of the major The vacuum envelope of the metal
components are examined in some detail, ceramic tubes consists of a metal cylinder
it should be remembered that all capped on both ends with ceramic disks,
conventional units will have similar usually composed of aluminum oxide.
construction. These ceramic insulators are designed to
allow for more effective use of the
insulation characteristics of both the
ceramic and the high tension grease used
X-Ray Tubes in sealing connections between the high
Early X-ray tubes used gas filled tubes and
a cold cathode from which electrons were
freed by positive ion bombardment.
FIGURE 3. Glass X-ray tube.
Modern tubes used in radiography are of
the high vacuum variety, allowing for Glass envelope
reduction in size, extended tube life and
more stable operation. Target mounting
Electrons are supplied by thermionic Focusing cup structure
emission from the filament. The
accelerating potential and the tube
Target
current can then be varied independently, Filament
with the exception that, at low
accelerating voltages, tube current is
affected by the space charge that
accumulates around the cathode.

Envelope
Envelopes for X-ray tubes are usually glass
or a combination of metal and ceramic
(Figs. 3 and 4). Glass envelope tubes,
although still in common use, are far
more susceptible to thermal and
mechanical shock than the metal ceramic

FIGURE 4. Metal ceramic X-ray tube.


Metal envelope
Ceramic disk insulator

Filament Target
mounting mounting
structure structure

Filament Target

Electronic Radiation Sources 59


3RT03_Layout(055_072) 10/2/02 1:51 PM Page 60

voltage source and the tube. This design principles). The relationship for geometric
allows for reduction in the size of the tube unsharpness Ug is:
housing, especially important for higher
energy units.
F F⋅t
(2) Ug = =
D D
Cathode
t
The cathode includes the tungsten
filament that provides the thermal where geometric unsharpness Ug is a
electrons for acceleration. The filament is measure of the penumbra of the focal
usually powered by alternating current spot, F is the focal spot size, D is the
(50 to 60 Hz) from a separately controlled distance from the target (focal spot to the
transformer, although in some units the source surface of the object) and t is the
filament current is fixed or automatically thickness of the object plus its distance to
controlled to maintain a constant tube the image plane.
current. Normally, filament currents range In accordance with Eq. 2, geometric
from 1 to 10 A. The tube current, passing unsharpness Ug increases directly as the
between the cathode and anode by means focal spot size increases. Because the
of the high speed electrons, ranges from electron beam originates at the filament,
less than 300 µA for microfocus units to reduction of the filament size might seem
over 20 mA for conventional industrial to solve geometrical unsharpness by
radiographic units. reducing the beam size but this approach
is limited by the durability of the
Beam Focusing filament.
One alternative, called line focusing, is
At times the filament is located in a recess
to project the virtually rectangular beam
in the cathode called a focusing cup. This
produced by the filament onto a target
surrounds the emerging beam of electrons
angled with respect to the beam (about
with an electric field that repels the beam
21 degrees). By masking all of the X-ray
away from the cup wall and into a more
beam except that emitted toward one
localized form. The importance of a well
side, the useful X-ray beam appears to
defined beam of electrons arises from the
issue from a focal spot with about equal
fact that the sharpness or unsharpness of
lateral dimensions (Fig. 6). In practice,
an image depends on the focal spot size
this technique allows production of units
(Fig. 5 and the chapter on radiographic
with effective focal sizes in the range of
1.0 to 3.0 mm (0.04 to 0.12 in.).
By use of a deep focusing cup,
FIGURE 5. Illustration of geometric unsharpness. advantage can also be taken of the screen
effect (Fig. 7). This refers to the removal of
F the lower energy electrons produced
during that portion of the alternating
current cycle where the potential
difference between cathode and anode is
significantly less than maximum. In
practice this improvement is not without
D

FIGURE 6. Diagram of line focusing setup.


Electron
beam
Object size

Discontinuity
t

Image

Ug Ug

Legend
D = distance from focal spot to target Effective Actual focal size
F = focal spot size focal size
t = thickness of test object
Ug = geometric unsharpness

60 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 61

cost to the output of the unit. A loss of Anode


about 25 percent is experienced in units
As mentioned previously, heat is the
with high screen effect. This can be
major form of energy produced as the
compensated for, in part, with higher
electrons strike the target. Uncontrolled,
filament current although this current
this heat would quickly cause the surface
adversely affects the lifetime of the
of the target to erode, which in turn
filament. An alternate technique of
reduces the definition of the focal spot. In
removing low energy components of the
addition, the vaporized target material
electron beam is found in the discussion
reduces the high vacuum of the tube and
on constant potential units.
leads to premature failure due to
If still further focus of the beam is
conduction within the tube. To avoid
desired, as in microfocus radiography and
overheating of the target, the anode to
some analytical applications, additional
which it is attached is composed of a
techniques may be used: (1) conversion of
material with high thermal conductivity,
the conventional diode arrangement of
such as copper. If the cooling demands are
cathode and anode into a triode
relatively low, as for a low energy unit or
arrangement, including a focusing
intermittent use, cooling is often
electrode or grid, and (2) electrostatic or
accomplished by means of a conductor
magnetic deflection systems.
that passes through the tube end for
For the triode arrangement used widely
connection to the high voltage source;
in the microfocus industry, a negative bias
this allows for radiation of heat into an
of up to –150 V is applied to the third
oil or gas reservoir surrounding the tube.
element of the tube to further focus the
Although such a design is not the most
beam and remove lower energy
efficient, the weight of such a tube is
components. This configuration allows a
minimal because pumps and heat
reduction of beam size, producing focal
exchangers are absent.
spots at least as small as 10 µm (4 ×
For higher energy units in continuous
10–4 in.) and a consequent drop in tube
use, it is usually necessary to cool the
current.
anode by injecting coolant directly into it.
In the case of electrostatic deflection,
This is accomplished by hollow
even more elements are included within
construction of the anode conductor.
the envelope whereas a magnetic
Another way of alleviating the problem
deflection system is external to the tube.
of localized heating of the target is with a
These types of deflection systems have an
rotating anode in which the target, a
additional advantage in that the beam
tungsten disk, is driven as shown in
may also be deflected to various areas of
Fig. 8. This allows the tube current to be
the target for added service life. Units that
increased by as much as ten times the
incorporate their own vacuum systems
value for a stationary target. The focal
usually allow for replacement of both
spot on such units can be reduced to less
filament and target components. These
types, though formerly limited to
analytical and research units, are available
off the shelf as standard microfocus and FIGURE 8. Rotating anode.
analytical units.

X-ray beam
FIGURE 7. Graph of screen effect.

Kilovolt peak
Tube current (relative scale)

Electron beam
High screen effect

Low screen effect

0 0.5 1.0

Time (half cycles)

Electronic Radiation Sources 61


3RT03_Layout(055_072) 10/2/02 1:51 PM Page 62

than 1 mm (0.04 in.) for short exposure beam intensity also shown. Figure 10b
times, which is of value in medical as well shows a similar representation for a
as some specialized industrial microfocus beam. The beam distribution
applications, such as high speed imaging. in Fig. 10b is said to be gaussian (bell
shaped) because of the shape of the
Target intensity curve. Such a beam profile is
required when it is necessary to define
In radiographic applications, the target is very closely spaced objects, such as
usually tungsten and is bonded to the microcircuitry components.
copper anode; however, analytical units
use several other target materials to take
advantage of the characteristic X-rays Hood
produced. Some of these materials include Addition of a hood to the anode provides
copper, iron and cobalt. the two fold function of (1) eliminating a
The orientation of the target with portion of the X-ray beam outside the
respect to the electron beam strongly central cone of radiation and
influences the size and shape of the focal (2) electrically shielding the insulating
spot. Orientations from 0 degrees to portions of the envelope (glass or ceramic)
30 degrees are used for various from charge buildup due to electrons
applications. For example, zero is the scattered from the tungsten target or
angle used for panoramic units. An angle released by the photoelectric effect (see
of 20 degrees is commonly selected for Fig. 11).
directional units because, in this case, the Removing the unused radiation directly
distribution of X-rays is predominantly in at the anode reduces the amount of
a direction perpendicular to the tube axis. radiation shielding that must be provided
This is shown graphically in Fig. 9. The externally or incorporated into the tube
actual maximum of intensity occurs at housing. The hood, normally constructed
+12 degrees. For radiography of objects of copper, may have materials with high
whose lateral dimensions are less than atomic numbers, such as tungsten,
half the focus-to-film distance (objects incorporated to increase absorption. The
that subtend an angle of less than electrical shielding function of the hood
30 degrees) the variation seldom matters.
Another cause for intensity variation is
the electron beam itself. A cross section of
the electron beam from the filament FIGURE 10. Electron beam distributions: (a) conventional
would resemble Fig. 10a, with relative beam; (b) microfocus beam.
(a)
Electron beam intensity

FIGURE 9. X-ray distribution graph.


(relative scale)

20 degrees

0 0.5 1.0

Distance across target

(b)
Electron beam intensity
(relative scale)

Angle –20 –13 –10 0 10 13 20


(degrees)
0 0.5 1.0
Intensity 31 70 80 100 104 105 95
(percent) Distance across target

62 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 63

may be improved by the addition of a Tank Head


beryllium window over the X-ray port. A
For a tank head, the housing itself
window 3 or 4 mm (0.1 or 0.2 in.) in
structurally protects the tube, contains the
thickness will stop electrons with
coolant and forms the structural support
negligible effect on the overall X-ray
for the tube insert, electrical connections,
beam.
fittings, pumps, thermal and high voltage
overload sensors and radiation shielding
Rod Anode contained in the head. For a tank unit,
The rod anode (sometimes referred to as the tube head also houses the high
an oxtail) is another adaptation of the voltage and filament transformers. If the
anode. This type of tube arrangement unit has separate components, the tube
requires special circuit considerations that head will also provide for connection to
allow the anode to be grounded. This the high voltage source.
tube, developed for use through small
openings, has been partially replaced by
the metal ceramic tube, which can have a
diameter of less than 50 mm (2 in.) and
High Voltage Sources
tube head diameter as small as 75 mm From line voltages in the range of 100 to
(3 in.). The target of such an end grounded 460 V, the high tension circuitry supplies
tube can be cooled by circulating water in potential differences to the tube from
direct contact with the anode. Beam 5 kV to as much as 420 kV for the larger
focusing is often required for longer tubes. industrial radiographic units. Several
standard circuit designs are used for
various applications. A portable tank unit
Coolant typically uses one of the designs shown in
With the exception of the end grounded Fig. 12.
configurations and units designed for low These circuits are all self-rectified; the
energy output (less than 50 kV), the tube X-ray tube itself limits the flow of
insert is surrounded by an insulating electrons to one direction in the circuit.
coolant and encased in a housing called When the anode is at negative potential
the tube head.
The coolant may be highly dielectric
gas or oil. If oil is used, simple convection
may be sufficient for lower output units. FIGURE 12. Standard high voltage circuit designs for portable
For larger units, an oil circulating pump tank-type units: (a) cathode grounded; (b) center grounded;
may be combined with a heat exchanger (c) anode grounded.
either internal or external to the tube
head. (a)
For units using a fixed amount of oil in
the tube and a circulating pump to
circulate it within the tube head, an oil
resistant bellows is incorporated to allow mA
for expansion and contraction of the oil.
Because of the compressibility of
insulating gases, this is not required for
gas filled heads but a pressure gage is
normally included to monitor possible
(b)
loss of coolant insulation.

FIGURE 11. Hooded anode tube.


mA

Anode hood
Cathode

(c)

mA

Beryllium
window

Electronic Radiation Sources 63


3RT03_Layout(055_072) 10/2/02 1:51 PM Page 64

with respect to the cathode, no tube unit time. This system is widely used in
current flows. medical applications but is used less than
One drawback of the self-rectified constant potential units in industrial
system is the possibility of tube backfire. applications.
If the target or anode overheats, reverse Greinacker Circuit. As can be seen from
conduction can occur during the negative the output waveform, the greinacker
half cycle. This type of unit is normally circuit is of the constant potential type.
used for tubes producing X-rays in the Basically a variation of the villard circuit,
range of 50 to 300 kV peak and tube
currents from 2 to 8 mA.
With the addition of capacitors and
FIGURE 14. Graetz grounded anode circuit:
diode rectifiers, the transformer is
(a) circuit; (b) waveform.
normally placed in a tank separate from
the head. The additional elements allow (a)
the current to be rectified by means of
valve tubes or solid state diodes and to be To filament
transformer
filtered and smoothed to provide a more
nearly constant accelerating voltage.
Several popular grounded circuits are
shown in Figs. 13 to 15.
Villard Circuit. An extension of the half
wave system, the villard circuit allows To high
production of accelerating potentials of voltage
control
twice the transformer peak voltage.
Capacitors are charged during one half of
the cycle and discharged when current
passes through the tube, augmenting the
voltage produced by the transformer.
(b)
Graetz Circuit. The full wave circuit, or
graetz circuit allows use of both halves of X-ray tube
the alternating current cycle with a
Voltage

substantial increase in tube output per

Rectified
FIGURE 13. Villard grounded anode circuit: alterations
(a) circuit; (b) waveform. Time

(a)
To filament transformer

FIGURE 15. Greinacker grounded anode


circuit: (a) circuit; (b) waveform.
(a)
To filament transformer

To high
voltage
control

To high voltage control

(b) (b)
X-ray tube X-ray tube
Voltage

Voltage

Transformer
Transformer
Time
Time

64 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 65

in which capacitors are charged during power the high tension transformer.
both halves of the cycle, the voltage is not Although this requires additional
only doubled but remains near maximum electronic circuitry or a motor generator,
value throughout the cycle. This gives the core of the high tension transformer
enhanced high energy output and can be reduced in size because of
eliminates the electrical stress placed on increased reactance at higher frequencies.
the tube and insulation. Enhanced tube The smaller transformer can be an
life, as well as about 30 percent reduction advantage in portable or mobile units.
in radiographic exposure times, are the Also, if filtering is to be done, the
results. variation or ripple of the output voltage
A common misconception is that can be reduced even further.
constant potential units provide a beam A variation of this technique is to use
of constant energy X-rays. Although the three phase input power with the high
electron beam is nearly monoenergetic, tension transformer. Commonly used in
the X-rays are produced during random medical X-ray generators, this technique
deceleration processes that give the is used by several industrial manufacturers
statistical energy distribution shown in as well.
Fig. 1. The absence of low energy Another approach is to use an output
electrons reduces the quantity of low waveform other than the standard sine
energy X-rays but does not eliminate wave. Approximate square wave outputs,
them. in conjunction with both phase inversion
circuitry and a high frequency
Cathode Grounded Circuit transformer, can provide accelerating
potentials with extremely low ripple
The main advantage of the cathode characteristics. Such units are available for
grounded configuration is that it allows industrial applications.
the filament transformer to be external to
the tank, because the cathode is at ground
potential and does not require isolation. High Tension Connections
The tube head can be reduced in size and Also important for conventional high
weight and is often gas filled to further voltage sources is the connection of the
decrease weight. high tension transformer to the tube. For
tank units this is not a major
Center Grounded Circuit consideration, because the transformer
can be connected directly to the tube
Both center grounded and anode electrodes. However, for units with
grounded units require isolated filament separate components, insulation and
transformers that must be insulated connection of leads (which may carry
adequately. For the center grounded unit, voltages in excess of 200 kV) are
this is justified by the reduction of the important considerations. The high
high tension transformer insulation. The tension cables themselves are shielded to
transformer needs to supply only one half provide protection against electrical
of the potential difference to each shock. Cables used at lower energies are
electrode, rather than having either the relatively flexible but as the amount of
cathode or anode held at ground potential insulation is increased, the flexibility
and supplying the entire accelerating decreases and sharp bends during
voltage to the other electrode. installation should be avoided.
In the range of 200 to 420 kV peak The cables are inserted into
with beam currents up to 15 mA, center terminations usually made of phenolic
grounded systems can be made smaller (thermosetting plastic) or ceramics and
than comparable end grounded units. are sealed against air by use of insulating
epoxy materials called potting compound.
Anode Grounded Circuit Figure 16 shows the phenolic termination,
used primarily with glass and lower
For the anode or end grounded system, energy metal ceramic tubes.
the advantage lies in the specialized use of These connectors are larger than the
the rod anode tube or metal ceramic tube ceramic terminations used primarily with
for access through small openings. As higher energy metal ceramic tubes. Both
mentioned earlier, cooling of the target is styles use high dielectric grease to seal out
also simplified. This type of unit is air at joining surfaces. Because of the
normally used for tubes producing X-rays tendency for the long male-to-female
in the range of 30 to 200 kV. connections to trap air and substantially
reduce the insulation capabilities of the
Alternative Circuit Designs grease, these joints are normally rated at
A technique for improving tube output about one tenth their theoretical values,
that can be used with any of the above or at about 10 kV per 10 mm (0.4 in.).
circuits including those in tank units, is The linear dimensions of such a
the use of a higher frequency waveform to termination must be correspondingly

Electronic Radiation Sources 65


3RT03_Layout(055_072) 10/2/02 1:51 PM Page 66

increased. At the transformer tank, this


size increase is not so important but at the
tube head the increase in size and weight
can make the unit very cumbersome and
bulky. For a 400 kV tube head, 200 kV is
applied to each electrode. This requires a
termination of 180 to 200 mm (7 or 8 in.)
in length at each end of the tube for
proper insulation.
A ceramic insulator can make full use
of the dielectric strength of the insulating
grease by providing rigid, flat mating
surfaces that exclude air from the joint.
This allows for a substantial decrease in
the length of the joint. This design has
been incorporated into tubes used by
several equipment manufacturers for units
up to 420 kV.
As stated above, a transformer is used
to provide the potential difference for
conventional X-ray units. As the
accelerating voltages are increased toward
1 MeV, standard transformer and
insulation technologies become
inadequate. Higher energy X-ray
generators are needed.

FIGURE 16. High voltage transformer terminations:


(a) phenolic connection; (b) ceramic connection.
(a)
Connector to tube
Air space

Connector to
high tension
generator

(b)
Air space

Connector to
high tension Connector to tube
generator

66 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 67

PART 3. Megavolt Radiography6

Resonant transformer X-ray machines possible to extract a 360 degree


were developed in the late 1930s to circumferential beam from the machine.
provide X-ray energies above 500 keV. The The focal spot size is typically as large as
technology of high voltage X-radiography 10 mm (0.4 in.) in diameter in the
is well documented.5-21 transmitted direction. The apparent size
Resonating the high voltage circuit to of the target will depend on the angle
the frequency of the alternating current that is measured from the flat plane of the
power supply obviates an iron transformer target. The radiation output from these
core. In addition, insulating the high machines is about 25 mGy·s–1
voltage sections with (150 rad·min–1) when measured at 1.0 m
dichlorodifluoromethane (refrigerant-12) (39 in.) from the target.
or sulfur hexafluoride (SF6) obviates These machines are very durable, many
transformer oil. These changes permit a having been in service for over 30 years
large reduction in the size and weight of and one having been in daily use for over
the machines, making them much more 60 years.
practical.
The resonant transformer X-ray
machines, similar to Fig. 17, consist of an
operating console, 180 cycles per second Electrostatic Generators
alternating current motor generator set, Van de graaff X-ray machines (Fig. 18)
motor controller, heat exchanger and a operate on the principle illustrated in
large steel tank. An air core transformer is Fig. 19. The van de graaff electrostatic
secured to the base of the steel tank by generator shown in Fig. 18 has undergone
insulating glass rods. A multisection X-ray some changes since its origin in the early
tube is mounted coaxially with the 1930s. However, the basic principles of
transformer stack. A heated filament operation presented in Fig. 19 remain the
provides the free electrons in the tube and same. The machine consists of a control
each section of the multisectional tube is console and a metal tank that houses
connected to an appropriate tap on the power supplies, drive motors, an insulated
transformer. Acceleration of the electrons charging belt, a generator column, high
through the tube is accomplished by voltage terminals and electrical
uniform voltage distribution throughout connections. An accelerator tube, which is
the length of the tube. The electrons highly evacuated, extends through the
strike a water cooled tungsten target at generator column. This metal tank is
the end of the tube, thereby generating pressurized with nitrogen, carbon dioxide
X-rays. and sulfur hexafluoride to about 2.6 MPa
Most X-rays are available in the (375 lbf·in.–2) to prevent high voltage
forward direction but opening ports
around the lead collimator makes it

FIGURE 18. Van de graaff 2.5 MeV


FIGURE 17. Resotron, 2 MeV. electrostatic generator.

Electronic Radiation Sources 67


3RT03_Layout(055_072) 10/2/02 1:51 PM Page 68

arcing. Within the tank, electron sprayers


deposit electrons onto a moving belt that
carries them to a terminal shell. Some of
Betatrons
these electrons are emitted into the The first successful betatron was built in
accelerator tube as beam current through 1941. It produced 2.3 MeV X-rays and had
a direct connection to the filament. The an X-ray output equivalent to 1 g of
filament, when properly heated, emits radium. Later, machines producing X-ray
electrons that are replaced from the energies up to 300 MeV or more were
terminal shell. The electrons flow through built, though the most common
the accelerator tube at an energy radiographic machines operated at
determined by the terminal shell voltage. 24 MeV.
A target is located at the end of the To accelerate electrons to high speed,
accelerator tube. This design results in a the betatron (Fig. 20) uses the magnetic
direct current, constant potential X-ray induction effect used in a transformer.
machine. Constant potential machines The primary winding in a transformer is
made in the 1990s are in the 2.5 MeV connected to an alternating current
range, have a 2.5 mm (0.1 in.) target and voltage source that establishes a varying
produce about 28.3 mGy·s–1 at 1 m flux in an iron core. The secondary
(170 rad·min–1 at 40 in.). winding on this core has induced in it a
voltage equal to the product of (1) the
number of turns in the secondary
winding and (2) the flux time rate of
FIGURE 19. Operation of electrostatic generator. change. The resulting electric current is
made up of the free electrons present in
the wire. The betatron (as shown in
Fig. 21) is essentially such a transformer
D except that, instead of wire, the secondary
is a hollow circular tube. This tube, called
a doughnut, is used to contain the
E electrons for many thousand revolutions.
C The doughnut tube is usually made of
porcelain and is coated on the inside with
a conductive layer of palladium
connected to a ground. The doughnut is
placed between the poles of an
electromagnet that produces a pulsating

B
F FIGURE 20. Betatron, 25 MeV.

A G

Legend
A. Electric charge is sprayed on belt.
B. Belt carries charge to terminal.
C. Charge transferred from belt to terminal establishes lower voltage.
D. Compressed nitrogen insulates terminal from accelerator shell.
E. Heated cathode provides electrically charged particles.
F. Charged particles escape via vacuum tube.
G. Potential difference between terminal and lower end accelerates particles.
H. Particle beam bombards targets to make X-rays.

68 Radiographic Testing
3RT03_Layout(055_072) 9/6/02 3:15 PM Page 69

field. Electrons injected into the tube as


the magnetic field increases will be
accelerated in a circular path. The force
Electron Linear
acting on the particles is proportional to Accelerators
the rate of change of flux and the Electron linear accelerators, similar to the
magnitude of the field. Because the one shown in Fig. 22, are commonly
electrons circle the orbit many times referred to as linacs. Linear accelerators
before striking the target, there is a large accelerate electrons down a guide by
amount of energy gain. For example, in a means of radio frequency voltages. These
24 MeV betatron, the electrons circle the voltages are applied so that the electron
orbit about 350 000 times, traveling a reaches an acceleration point in the field
distance of 418 km (260 mi). The average at precisely the proper time. The
voltage gain per turn at the orbit is about accelerator guide consists of a series of
70 V, which gives about 24 MeV. As the cavities that causes gaps when the radio
electrons reach maximum energy, they are frequency power is applied. The cavities
deflected by an electrical pulse and caused have holes in each end that allow
to spiral outward until they strike the electrons to pass to the next cavity. When
target. Betatrons are equipped with an electron is injected at the proper time
platinum wire targets with dimensions of it gains energy as it is accelerated across
0.13 × 0.25 mm (0.005 × 0.010 in.) and these gaps and out the other end of the
have a radiation output of 25 mGy·s–1 cavity. When the radio frequency power is
(150 rad·min–1) when measured at 1 m phased properly, increased acceleration is
(40 in.) from the target. achieved. Figure 23 shows the general
arrangement of a linear accelerator’s
component parts.
For research purposes, heavier particles
FIGURE 21. Diagram of betratron generator: (a) top view; have been accelerated by linear
(b) cross section. accelerators. An example of this type
linear accelerator is the 3.2 km (2 mi)
(a) long Stanford linear accelerator. There are
Earth ground
two general types of radiographic linear
accelerators: (1) one using the principle of
Equilibrium orbit the traveling wave and (2) the other using
Expanded orbit a standing wave technique of
acceleration.
Target structure Radiographic linear accelerators have
been produced that have X-ray energies of
1 to 16 MeV and with output as high as
4.17 Gy·s–1 (25 krad·min–1) measured at
1 m (40 in.). Although not as durable as
Injector lower voltage machines, some have been
in service for over 30 years. Increased
operating frequencies, up to 9.3 GHz,
permit smaller, lighter X-ray heads.
Palladium Configurations exist that allow for the
coated interior
X-ray beam operation of the accelerator and beam

(b)
FIGURE 22. Electron linear accelerator.

Upper Pole

Doughnut
Hr

Steel wafer

Steel wafer
Hr Electron
H orbit
Lower Pole H
H

Electrons moving toward reader at this point

Electronic Radiation Sources 69


3RT03_Layout(055_072) 10/2/02 1:51 PM Page 70

FIGURE 23. General arrangement of linear accelerator.

Radio
Radio frequency source Laser
frequency
system

Pulse transformer
Accelerator

Electron
gun Collimator

collimator at a distance from the radio


frequency source; the source supplies
power through a flexible wave guide. The
total weight of the X-ray head is greatly
reduced, permitting easy positioning for
field inspection of pipelines, valves and
other test objects of limited accessibility.
One such system is being used for
inservice inspection of nuclear power
plants.

70 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 71

References

1. Beiser, A. Concepts of Modern Physics, 14. Pollitt, C.G. “Radiography with


fifth edition. New York, NY: High-Energy Radiation.” Journal of the
McGraw-Hill Book Company (1963). British Steel Castings Research
2. Halliday, D., R. Resnick and J. Walker. Association. Sheffield, United
Fundamentals of Physics, sixth edition, Kingdom: British Steel Castings
extended. New York, NY: John Wiley Research Association (1962).
and Sons (2001). 15. O’Connor, D.T. and B.G.
3. Industrial Radiography Manual. Cunningham. “Installation of the
Washington, DC: United States Office Flexible 10 MeV Betatron.”
of Education, Division of Vocational Non-Destructive Testing. Vol. 7, No. 4.
and Technical Education; and United Columbus, OH: American Society for
States Atomic Energy Commission, Nondestructive Testing (Spring 1949):
Division of Nuclear Education and p 20-23.
Training (1968). 16. Halmshaw, R. and C.C. Pollitt.
4. Lapp, R.E. and H.L. Andrews. Nuclear “Radiology with High-Energy X-Rays.”
Radiation Physics, fourth edition. Progress in Nondestructive Testing,
Upper Saddle River, NJ: Prentice-Hall Vol. 2. London, United Kingdom:
(1972). Heywood and Company (1959).
5. Nondestructive Testing Handbook, first 16. Cusick, J.H. and J. Haimson. “10-MeV
edition. Columbus, OH: American Rotating Target Linear Accelerator for
Society for Nondestructive Testing Radiography of Large Rocket Motors.”
(1959). Proceedings: Missiles and Rockets
6. Brunty, B.J. “High-Energy Symposium. Concord, CA: United
Radiography.” Nondestructive Testing States Naval Ammunition Depot
Handbook, second edition. Columbus, (1961).
OH: American Society for 18. Haimson, J. “Radiography of Large
Nondestructive Testing (1985): Missiles with the Linear Electron
p 256-298. Accelerator.” Nondestructive Testing.
7. Wilshire, W.J., ed. A Further Handbook Vol. 21, No. 2. Columbus, OH:
of Industrial Radiology. London, United American Society for Nondestructive
Kingdom: Edward Arnold and Testing (March-April 1963): p 102-112.
Company (1957). 19. Bly, J.H. and E.A. Burrill. “High-Energy
8. Nondestructive Testing Handbook, first Radiography in the 6 to 30 MeV
edition. Columbus, OH: American Range.” Special Technical Publication
Society for Nondestructive Testing No. 278. West Conshohocken, PA:
(1959). ASTM International (1959).
9. Hogarth C.A. and J. Blitz. Techniques of 20. Morgan, R.H. and K.E. Corrigan, eds.
Nondestructive Testing. London, United Handbook of Radiology. Chicago, IL:
Kingdom: Butterworth and Company Year Book Publishers (1955).
(1960). 21. Wheeler, G.C. “Radiography of Steel
9. McGonnagle, W.J. Nondestructive with the 25,000 R/min Linac.”
Testing. New York, NY: McGraw-Hill Paper B-41. Conference Proceedings:
Book Company (1961). The Seventh International Conference
10. Halmshaw, R., ed. Physics of Industrial on Nondestructive Testing [Warsaw,
Radiology. New York: American Elsevier Poland, June 1973]. Vol. 1. Warsaw,
Publishing Company (1966). Poland: Instytut Podstawowych
11. Radiography in Modern Industry. Problemów Techniki Polskiej
Rochester, NY: Eastman Kodak Akademii Nauk (1973): p 241-249.
Company.
12. Halmshaw, R. Industrial Radiology
Techniques. London: Wykeham
Publishing Company (1971).
13. High-Energy X-Ray Applications for
Nondestructive Testing. RAD-1936. Palo
Alto, CA: Varian Associates (1982).

Electronic Radiation Sources 71

You might also like