Professional Documents
Culture Documents
C
3
H A P T E R
Electronic Radiation
Sources
Radiation of an unknown type and origin target. Much of the energy of the
was discovered in the year 1895 by electrons’ motion, called kinetic energy, is
Wilhelm Roentgen. Dubbed X-rays by transformed into heat energy as the
their discoverer, they were soon found to stream strikes the target. In fact, except
be a form of electromagnetic radiation for the case of very high energy
with extremely short wavelengths. The generators, almost all of the electrons’
X-radiation is called ionizing because X-ray kinetic energy (more than 97 percent) is
absorption kicks out an electron from an converted into heat; disposal of this
atom of absorption material, creating an thermal energy is an important design
ion. The production of ionizing radiation consideration.
is well covered in the literature.1-5
Technological advances in image
processing, flash radiography, real time
radiography, microfocus and metal Bremsstrahlung
ceramic tubes have sustained interest in A small portion of the energy will also be
the electronic production of ionizing given off as packets of electromagnetic
radiation. The versatility of an electronic energy, called photons. The X-ray photons
source is difficult for an isotopic source to can have energies ranging from a few
match, even for specific applications. The thousand volts to a maximum determined
advent of solid state components has by the original kinetic energy of the
overcome many of the objections related electron and how rapidly the electron is
to size, weight and durability of electronic decelerated by interaction with a target
sources. atom nucleus. This process produces the
This chapter discusses the basic continuous portion of the X-ray spectrum
physics, construction and application of and is known by the German term
electronic sources. The emphasis will be Bremsstrahlung for braking radiation.
on industrial X-ray machines widely used Energies of the electrons (and the
in radiography and radiology. Where X-rays) are frequently given in terms of
special details are significant to particular kiloelectronvolt (keV) or megaelectronvolt
applications, these are discussed in the (MeV). The meaning of this unit becomes
appropriate chapters of this volume. clear if the charged electron is considered.
Under the influence of a voltage
difference (technically called a potential
difference), charged particles will
Conservation of Energy experience a force that causes them to
Electromagnetic theory had long accelerate. A negatively charged particle,
predicted that a charged particle such as an electron, will move from a
undergoing deceleration would emit place of low voltage (–) to a place of
radiation. This theory can be used to higher voltage (+) and increase its kinetic
qualitatively explain the continuous
portion of a typical X-ray spectrum (see
Fig. 1). Although a complete study of the
continuous spectrum requires modern FIGURE 1. Typical X-ray spectrum.
physics, an understanding of its basic
principles may be gained by considering Characteristic peaks
just one of the fundamental laws of
Relative intensity
56 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 57
energy as it does so. Thus the unit different energy levels. If a vacancy in a
kiloelectronvolt (keV) corresponds to the particular level were created by knocking
amount of kinetic energy that an electron an electron from its orbit, an electron
would gain while moving between two from a higher energy level would, after a
points that differ in voltage by 1 kV. time, drop down to fill the void. To do
Similarly, an electron would gain 1 MeV this, it would have to lose (emit) energy.
of kinetic energy while moving between Because the energy levels of an atom are
two points that differ by 1 MV. The points well specified, the exact amount of energy
of differing voltage are called the cathode lost by an electron making such a
(–) and the anode (+) and will be discussed transition would also be discrete.
later in this section. According to the law of conservation of
energy, the energy is not really lost but
given off as electromagnetic radiation in
the form of an X-ray photon. Because
Characteristic X-Rays each atomic element has its own distinct
In addition to the bremsstrahlung there set of energy levels, the line spectrum
are several characteristic peaks in a typical produced in such a manner is
X-ray spectrum. These intensity spikes are characteristic of the particular target
caused by interaction between the material. Interest in these characteristic
impinging stream of high speed electrons X-rays lies in their application to X-ray
and the electrons that are bound tightly diffraction and other analytical
to the atomic nuclei of the target applications. By a combination of
material. If an atom is visualized as a filtration and signal processing,
planetary system, with the nucleus of characteristic X-rays can be used to
protons and neutrons at the center of the produce a nearly monoenergetic beam of
system and with the electrons moving in X-rays.
orbits around the nucleus (see Fig. 2), the
law of energy conservation can again be
applied to describe the origin of
characteristic radiation. Thermionic Emission
Modern physics predicts that these Two final concepts are of interest in this
orbital electrons near the nucleus will brief treatment of the basic physics of
have very well defined energies, with X-ray production. The first process,
electrons in different orbits having thermionic emission, provides the
electrons that are to be accelerated to
high speeds by the voltage difference
between the cathode and anode of an
FIGURE 2. Planetary model of atomic X-ray source.
structure. When metal (in this case the X-ray
tube filament) is heated to incandesence,
a small portion of the free electrons are
E
actually able to escape the surface of the
material. Without an accelerating voltage,
the electrons would form a cloudy space
charge around the filament. Under the
influence of a potential difference,
however, the electrons are quickly
accelerated toward the anode of the tube.
N
P
N P Absorption
N The second concept is X-ray absorption,
P
discussed in greater detail in the chapter
on radiation and particle physics. As
E
X-rays pass into or through a material,
they are absorbed, or attenuated, in a
manner that depends on the energy of the
X-radiation and on the density of the
material. In equation form, the intensity I
E
as a function of thickness has a standard
exponential form:
Legend (1) I = Io e – µ x
E = electron
N = neutron where I is the intensity after passing
P = proton through the material; x is the thickness of
58 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 59
A conventional X-ray generator consists of envelope and have been replaced in most
three main components: (1) X-ray tube, industrial applications with the more
(2) high voltage source and (3) control durable metal ceramic tubes.
equipment. Although each of the major The vacuum envelope of the metal
components are examined in some detail, ceramic tubes consists of a metal cylinder
it should be remembered that all capped on both ends with ceramic disks,
conventional units will have similar usually composed of aluminum oxide.
construction. These ceramic insulators are designed to
allow for more effective use of the
insulation characteristics of both the
ceramic and the high tension grease used
X-Ray Tubes in sealing connections between the high
Early X-ray tubes used gas filled tubes and
a cold cathode from which electrons were
freed by positive ion bombardment.
FIGURE 3. Glass X-ray tube.
Modern tubes used in radiography are of
the high vacuum variety, allowing for Glass envelope
reduction in size, extended tube life and
more stable operation. Target mounting
Electrons are supplied by thermionic Focusing cup structure
emission from the filament. The
accelerating potential and the tube
Target
current can then be varied independently, Filament
with the exception that, at low
accelerating voltages, tube current is
affected by the space charge that
accumulates around the cathode.
Envelope
Envelopes for X-ray tubes are usually glass
or a combination of metal and ceramic
(Figs. 3 and 4). Glass envelope tubes,
although still in common use, are far
more susceptible to thermal and
mechanical shock than the metal ceramic
Filament Target
mounting mounting
structure structure
Filament Target
voltage source and the tube. This design principles). The relationship for geometric
allows for reduction in the size of the tube unsharpness Ug is:
housing, especially important for higher
energy units.
F F⋅t
(2) Ug = =
D D
Cathode
t
The cathode includes the tungsten
filament that provides the thermal where geometric unsharpness Ug is a
electrons for acceleration. The filament is measure of the penumbra of the focal
usually powered by alternating current spot, F is the focal spot size, D is the
(50 to 60 Hz) from a separately controlled distance from the target (focal spot to the
transformer, although in some units the source surface of the object) and t is the
filament current is fixed or automatically thickness of the object plus its distance to
controlled to maintain a constant tube the image plane.
current. Normally, filament currents range In accordance with Eq. 2, geometric
from 1 to 10 A. The tube current, passing unsharpness Ug increases directly as the
between the cathode and anode by means focal spot size increases. Because the
of the high speed electrons, ranges from electron beam originates at the filament,
less than 300 µA for microfocus units to reduction of the filament size might seem
over 20 mA for conventional industrial to solve geometrical unsharpness by
radiographic units. reducing the beam size but this approach
is limited by the durability of the
Beam Focusing filament.
One alternative, called line focusing, is
At times the filament is located in a recess
to project the virtually rectangular beam
in the cathode called a focusing cup. This
produced by the filament onto a target
surrounds the emerging beam of electrons
angled with respect to the beam (about
with an electric field that repels the beam
21 degrees). By masking all of the X-ray
away from the cup wall and into a more
beam except that emitted toward one
localized form. The importance of a well
side, the useful X-ray beam appears to
defined beam of electrons arises from the
issue from a focal spot with about equal
fact that the sharpness or unsharpness of
lateral dimensions (Fig. 6). In practice,
an image depends on the focal spot size
this technique allows production of units
(Fig. 5 and the chapter on radiographic
with effective focal sizes in the range of
1.0 to 3.0 mm (0.04 to 0.12 in.).
By use of a deep focusing cup,
FIGURE 5. Illustration of geometric unsharpness. advantage can also be taken of the screen
effect (Fig. 7). This refers to the removal of
F the lower energy electrons produced
during that portion of the alternating
current cycle where the potential
difference between cathode and anode is
significantly less than maximum. In
practice this improvement is not without
D
Discontinuity
t
Image
Ug Ug
Legend
D = distance from focal spot to target Effective Actual focal size
F = focal spot size focal size
t = thickness of test object
Ug = geometric unsharpness
60 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 61
X-ray beam
FIGURE 7. Graph of screen effect.
Kilovolt peak
Tube current (relative scale)
Electron beam
High screen effect
0 0.5 1.0
than 1 mm (0.04 in.) for short exposure beam intensity also shown. Figure 10b
times, which is of value in medical as well shows a similar representation for a
as some specialized industrial microfocus beam. The beam distribution
applications, such as high speed imaging. in Fig. 10b is said to be gaussian (bell
shaped) because of the shape of the
Target intensity curve. Such a beam profile is
required when it is necessary to define
In radiographic applications, the target is very closely spaced objects, such as
usually tungsten and is bonded to the microcircuitry components.
copper anode; however, analytical units
use several other target materials to take
advantage of the characteristic X-rays Hood
produced. Some of these materials include Addition of a hood to the anode provides
copper, iron and cobalt. the two fold function of (1) eliminating a
The orientation of the target with portion of the X-ray beam outside the
respect to the electron beam strongly central cone of radiation and
influences the size and shape of the focal (2) electrically shielding the insulating
spot. Orientations from 0 degrees to portions of the envelope (glass or ceramic)
30 degrees are used for various from charge buildup due to electrons
applications. For example, zero is the scattered from the tungsten target or
angle used for panoramic units. An angle released by the photoelectric effect (see
of 20 degrees is commonly selected for Fig. 11).
directional units because, in this case, the Removing the unused radiation directly
distribution of X-rays is predominantly in at the anode reduces the amount of
a direction perpendicular to the tube axis. radiation shielding that must be provided
This is shown graphically in Fig. 9. The externally or incorporated into the tube
actual maximum of intensity occurs at housing. The hood, normally constructed
+12 degrees. For radiography of objects of copper, may have materials with high
whose lateral dimensions are less than atomic numbers, such as tungsten,
half the focus-to-film distance (objects incorporated to increase absorption. The
that subtend an angle of less than electrical shielding function of the hood
30 degrees) the variation seldom matters.
Another cause for intensity variation is
the electron beam itself. A cross section of
the electron beam from the filament FIGURE 10. Electron beam distributions: (a) conventional
would resemble Fig. 10a, with relative beam; (b) microfocus beam.
(a)
Electron beam intensity
20 degrees
0 0.5 1.0
(b)
Electron beam intensity
(relative scale)
62 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 63
Anode hood
Cathode
(c)
mA
Beryllium
window
with respect to the cathode, no tube unit time. This system is widely used in
current flows. medical applications but is used less than
One drawback of the self-rectified constant potential units in industrial
system is the possibility of tube backfire. applications.
If the target or anode overheats, reverse Greinacker Circuit. As can be seen from
conduction can occur during the negative the output waveform, the greinacker
half cycle. This type of unit is normally circuit is of the constant potential type.
used for tubes producing X-rays in the Basically a variation of the villard circuit,
range of 50 to 300 kV peak and tube
currents from 2 to 8 mA.
With the addition of capacitors and
FIGURE 14. Graetz grounded anode circuit:
diode rectifiers, the transformer is
(a) circuit; (b) waveform.
normally placed in a tank separate from
the head. The additional elements allow (a)
the current to be rectified by means of
valve tubes or solid state diodes and to be To filament
transformer
filtered and smoothed to provide a more
nearly constant accelerating voltage.
Several popular grounded circuits are
shown in Figs. 13 to 15.
Villard Circuit. An extension of the half
wave system, the villard circuit allows To high
production of accelerating potentials of voltage
control
twice the transformer peak voltage.
Capacitors are charged during one half of
the cycle and discharged when current
passes through the tube, augmenting the
voltage produced by the transformer.
(b)
Graetz Circuit. The full wave circuit, or
graetz circuit allows use of both halves of X-ray tube
the alternating current cycle with a
Voltage
Rectified
FIGURE 13. Villard grounded anode circuit: alterations
(a) circuit; (b) waveform. Time
(a)
To filament transformer
To high
voltage
control
(b) (b)
X-ray tube X-ray tube
Voltage
Voltage
Transformer
Transformer
Time
Time
64 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 65
in which capacitors are charged during power the high tension transformer.
both halves of the cycle, the voltage is not Although this requires additional
only doubled but remains near maximum electronic circuitry or a motor generator,
value throughout the cycle. This gives the core of the high tension transformer
enhanced high energy output and can be reduced in size because of
eliminates the electrical stress placed on increased reactance at higher frequencies.
the tube and insulation. Enhanced tube The smaller transformer can be an
life, as well as about 30 percent reduction advantage in portable or mobile units.
in radiographic exposure times, are the Also, if filtering is to be done, the
results. variation or ripple of the output voltage
A common misconception is that can be reduced even further.
constant potential units provide a beam A variation of this technique is to use
of constant energy X-rays. Although the three phase input power with the high
electron beam is nearly monoenergetic, tension transformer. Commonly used in
the X-rays are produced during random medical X-ray generators, this technique
deceleration processes that give the is used by several industrial manufacturers
statistical energy distribution shown in as well.
Fig. 1. The absence of low energy Another approach is to use an output
electrons reduces the quantity of low waveform other than the standard sine
energy X-rays but does not eliminate wave. Approximate square wave outputs,
them. in conjunction with both phase inversion
circuitry and a high frequency
Cathode Grounded Circuit transformer, can provide accelerating
potentials with extremely low ripple
The main advantage of the cathode characteristics. Such units are available for
grounded configuration is that it allows industrial applications.
the filament transformer to be external to
the tank, because the cathode is at ground
potential and does not require isolation. High Tension Connections
The tube head can be reduced in size and Also important for conventional high
weight and is often gas filled to further voltage sources is the connection of the
decrease weight. high tension transformer to the tube. For
tank units this is not a major
Center Grounded Circuit consideration, because the transformer
can be connected directly to the tube
Both center grounded and anode electrodes. However, for units with
grounded units require isolated filament separate components, insulation and
transformers that must be insulated connection of leads (which may carry
adequately. For the center grounded unit, voltages in excess of 200 kV) are
this is justified by the reduction of the important considerations. The high
high tension transformer insulation. The tension cables themselves are shielded to
transformer needs to supply only one half provide protection against electrical
of the potential difference to each shock. Cables used at lower energies are
electrode, rather than having either the relatively flexible but as the amount of
cathode or anode held at ground potential insulation is increased, the flexibility
and supplying the entire accelerating decreases and sharp bends during
voltage to the other electrode. installation should be avoided.
In the range of 200 to 420 kV peak The cables are inserted into
with beam currents up to 15 mA, center terminations usually made of phenolic
grounded systems can be made smaller (thermosetting plastic) or ceramics and
than comparable end grounded units. are sealed against air by use of insulating
epoxy materials called potting compound.
Anode Grounded Circuit Figure 16 shows the phenolic termination,
used primarily with glass and lower
For the anode or end grounded system, energy metal ceramic tubes.
the advantage lies in the specialized use of These connectors are larger than the
the rod anode tube or metal ceramic tube ceramic terminations used primarily with
for access through small openings. As higher energy metal ceramic tubes. Both
mentioned earlier, cooling of the target is styles use high dielectric grease to seal out
also simplified. This type of unit is air at joining surfaces. Because of the
normally used for tubes producing X-rays tendency for the long male-to-female
in the range of 30 to 200 kV. connections to trap air and substantially
reduce the insulation capabilities of the
Alternative Circuit Designs grease, these joints are normally rated at
A technique for improving tube output about one tenth their theoretical values,
that can be used with any of the above or at about 10 kV per 10 mm (0.4 in.).
circuits including those in tank units, is The linear dimensions of such a
the use of a higher frequency waveform to termination must be correspondingly
Connector to
high tension
generator
(b)
Air space
Connector to
high tension Connector to tube
generator
66 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 67
B
F FIGURE 20. Betatron, 25 MeV.
A G
Legend
A. Electric charge is sprayed on belt.
B. Belt carries charge to terminal.
C. Charge transferred from belt to terminal establishes lower voltage.
D. Compressed nitrogen insulates terminal from accelerator shell.
E. Heated cathode provides electrically charged particles.
F. Charged particles escape via vacuum tube.
G. Potential difference between terminal and lower end accelerates particles.
H. Particle beam bombards targets to make X-rays.
68 Radiographic Testing
3RT03_Layout(055_072) 9/6/02 3:15 PM Page 69
(b)
FIGURE 22. Electron linear accelerator.
Upper Pole
Doughnut
Hr
Steel wafer
Steel wafer
Hr Electron
H orbit
Lower Pole H
H
Radio
Radio frequency source Laser
frequency
system
Pulse transformer
Accelerator
Electron
gun Collimator
70 Radiographic Testing
3RT03_Layout(055_072) 10/2/02 1:51 PM Page 71
References