Professional Documents
Culture Documents
What‘s Next?
Feb.22,2019
1
The Challenge of Dreams
What‘s Next?
Contents
2
The Challenge of Dreams
What‘s Next?
3
Specification of TOPO REF Comparison Table
Products of other Companies Our products
RT-7000 TMS-4N CA800 ANTARES-2
CT-1000 RET-700
(TOMEY) (TOMEY) (TOPCON) (CSO)
Exterior image
-25D~+20D -20D~+30D
Function SPH × × × ×
Ref (VD=12) (VD=0)
Strength
・Wider measurement range ・Possible to store the patient data
φ0.4~φ10.7 Number of Ring 17or19 Approx. 10-thousand images could be stored
and opened. Use LAN, USB-A and USB-B,
store to the external
φ0.4 and open from there.
※ USB-B: Output Data only
φ10.7
6
The Challenge of Dreams
What‘s Next?
2.Measurement mode
There are 4 Measurement Functions.
Tap on the respective icon to activate each function.
1.Topographer
Measuring light : Red
2.REF/KRT
REF Measuring light:
Near infrared
KRT Measuring light:
Near infrared
3.Dry eye
Measuring light:Red
4.Meibomian
Measuring light:
Near infrared 7
What‘s Next?
3. Topographer
・Measurement principle
The multiple ring pattern for measurement
is projected on the cornea and take a
image by using the light that is reflected LED
from cornea with the sensor. The size of (Central ring)
ring pattern is changed depending on the
corneal curvature radius. The map of
corneal curvature radius are designed
by calculating the size of each ring.
Use the visible light (650nm) for
measurement light.
・Central ring
The inside 2 rings (Central ring) of 19 are
projected by the light inside the optical unit,
which make it possible to take a
measurement for peripheral of top of Sensor
cornea. The other 17 rings are projected by (Monitor)
the light from the Placido’s disk. 8
The Challenge of Dreams
What‘s Next?
9
The Challenge of Dreams
What‘s Next?
・Alignment procedure
As for the alignment procedure, align manually, please.
In case of distance between subject’s eye and device is far: red arrow
In case of distance between subject’s eye and device is close : blue arrow
The number of arrow decrease as approaching the distance.
When alignment is OK, the reticle displayed green
・Alignment OK ・Far from subject’s eye ・Too close with subject’s eye
10
The Challenge of Dreams
What‘s Next?
11
The Challenge of Dreams
What‘s Next?
①Type of map
②Type of scale
③Sim-K’s
④Corneal profile
Cornea
・Elevation ・Refractive
BFS(Best fit sphere)
Cornea
The height of cornea in reference to
“best fit sphere” (BFS) or the difference
between actual corneal surface & BFS.
(BFS: φ3/φ5/φ7 mm)
13
What‘s Next?
・②Type of scale
・Normalized ・Numeric
step RmaxRmin
31
(In case of Step is less than 0.25,
it is considered as STEP is 0.25)
・Absolute ・Adjustable
Step=set randomly
0.12~4.75 (0.25※0.12)
Med= set randomly
39.00~47.00 (0.12)
14
The Challenge of Dreams
What‘s Next?
15
The Challenge of Dreams
What‘s Next?
・④Corneal profile
Sim K’s dk
16
The Challenge of Dreams
What‘s Next?
17
The Challenge of Dreams
What‘s Next?
18
The Challenge of Dreams
What‘s Next?
19
The Challenge of Dreams
What‘s Next?
20
The Challenge of Dreams
Fourier,Zernikeはど What‘s Next?
んな目的に使う?
・Analysis screen-Aberration Fourier
It is possible to show Spherical, Astigmatism, Asymmetry, High order by using Fourier
analysis.
The average value of each component in φ3, φ6, φ9 area is displayed in Value.
【Fourier】 【Value】
21
The Challenge of Dreams
What‘s Next?
【Total】 【Each】
22
The Challenge of Dreams
What‘s Next?
Dia.:Diameter
OZ:Optical zone
BC:Base curve
PC:Peripheral curve
23
The Challenge of Dreams
What‘s Next?
24
The Challenge of Dreams
What‘s Next?
25
The Challenge of Dreams
What‘s Next?
26
What‘s Next?
4. Dry eye
・What is the Dry eye?
Dry eye is the disease to damage the health of eye surface because of lacking
the amount of and being poor of quality of tears.
Aqueous(water) Aqueous(water)
layer layer
Mucous Layer Mucous Layer
Corneal Corneal
27
epithelium epithelium
What‘s Next?
28
What‘s Next?
29
What‘s Next?
5 steps for
switching
sensitivity Play/stop button
Adjustment of
graph scale Grapf about how
10~100 disturb the ring is
30
What‘s Next?
Normal
31
What‘s Next?
5. Meibomian
Oil
・What is the Meibomian? layer
A layer of oil is on the surface to make tears difficult to Water
evaporate.The Meibomian produce this oil. layer
Mucous
The disease to be poor the function of Meibomian,
layer
which is called MGD(Meibomian gland dysfunction).
There are some cases which is to be poor Meibomian Corneal
for MGD. Epithelium
【Normal meibomian】 【Meibominan of MGD patient】
The number of
meibomian
Upper eyelid:
30~40
Lower eyelid:
20~30
32
The Challenge of Dreams
What‘s Next?
33
The Challenge of Dreams
What‘s Next?
・Analysis screen-Meibomian
The measured image can be edited to make it easy to see by adjusting 【Contrast】 and
【Brightness】 as necessary.
Also it is possible to enlarge the random part by zoom function.
34
The Challenge of Dreams
What‘s Next?
6. Setup
Regarding REF/KRT1, REF/KRT2, Topographer, Contact lens, Option,
refer to operation manual.
35
The Challenge of Dreams
What‘s Next?
・Memory
Topographer data(Valid for Analysis screen)
Save exam Set to save of Exam data
Save report Set to save of Report
36
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・Printer
37
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・Registration
Enter the registration code to be effective the option function of dry eye.
Registration code is different depending on each product.
By entering demo code, it is possible to use the dry eye function with the number of usage
limit.
For registration code and demo code, please contact to sales department.
38
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
7. External output
・DR-900
1.2 types of cables are required to connect with DR-900
・USB-Serial converter (CHIPI-X10)
・9P-9P cross cable(SP3303)
39
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・DR-900
3.Set Baurate REF of DR-900 to“115200”
4.After REF measurement on Measure screen, tap the Print/Save
button.The data is output to DR-900.
40
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・Report output
It is possible to output the report by pressing Print/Save button on Analysis screen.
The output report is different depending on the analysis screen.
It is selectable either jpg or pdf for report output.
41
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・LAN
It is possible to output to PC and server. Setting item Setting item
Also you can open the output Exam data. ・Terminal ・Shared folder
Shared folder setting and IP address ・Shared folder ・Network
setting are required for LAN connection. ・Network (IP address)
Refer to Maintenance manual for setting (IP address)
procedure.
LAN cable
42
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
Folder setting
1.Add the following folder to the local disk(C drive)
「 C:/RET-700/SharedFolder 」
43
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・LAN
2.After right-click “SharedFolder” and click “Share with”-”Specific people”
44
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・LAN
5.Check that the folder name is “SharedFolder”
6.After checking , press “Done” button.
IP address setting
1.Start up “Control panel”-”Network and Sharing Center”
2.Click “Change adapter settings”
45
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・LAN
3.Right-click “Local Area Connection” and click “Properties”
46
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・LAN
5.Select “Use the following IP address” and enter “IP address” and “Subnet mask”
as below.
6.Click “OK button.
47
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・LAN
Setting of RET-700
48
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・USB-A
It is possible to output data to USB or external HDD.
(Format is only FAT32)
Also it is possible to open the output Exam data.
Setting item
・Terminal
49
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・USB-A
The RET-700 folder is created.
The Exam / Report folder is created in that.
50
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
8. Procedure of software update (SYSTEM ver4.0)
1. Preparation for Update
Necessary item
•PC
•USB-B driver “CP210xVCPInstaller_x86.exe” (in case that the PC is
64bit,_x64.exe)
•Tool for update:SystemUpdate_1005(DealerMode) (provided
separately)
•USB memory for SYSTEM update (provided separately)
•USB-B cable
51
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
2. Install driver
•Install in PC the driver which recognizes the USB-B.
•In case of Windows, please confirm that “Silicon Labs CP210x USB to
UART Bridge(COM..)” is displayed on the port (COM and LPT) by
proceeding as “Control Panel” → “Device Manager”.
If it is not displayed, please retry it from 1
52
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
3. Connect pc by USB A-B cable to RET700
•Copy the software SystemUpdate_1003(DealerMode) in a folder of
PC (example in Documents)
•Connect the USB A-B cable between RET700 and PC
•Switch on the RET700
•By double click start the software SystemUpdate_1005(DealerMode)
•A file named COM.INI will be created automatically in same folder
•Open the COM.INI and be sure that the port is the same of Silicon
Labs CP210 port
•If different, change the number inside COM.INI, than close the
COM.INI file
53
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
54
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
55
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
5
56
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
5
57
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
9. KRT/TOPO calibration
・How to activate maintenance mode
1.Turn on while pressing the measurement switch.
2. Hold down the measurement switch even if Rexxam logo is switched to the
measurement screen. It takes about 1 min. even in normal condition.
58
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
59
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・Calibration
Calibration screen is following contents.
REF 【0D , -10D , +10D , -5D , CYL】
KRT/TOPO 【Alignment/KRT LED , TOPO LED , KRT/TOPO image , KRT/TOPO radius】
60
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・TOPO Calibration
2)KRT/TOPO image 【R8】
1.Set the attached model eye(R:8.0mm).
2.Align by moving the measurement unit. Make sure to be less than 0±2 for X, Y and Z
coordinates at that time.
※NOTE: Align while lowering the measuring unit as much as possible.
If it is not fully lowered, the upper part of ring may be cut due to the chinrest.
3.Take a measurement by tapping [Set].
4.Check that the analysis point of the ring satisfies the following.
18th:more than 55 points for both eyes each , 19th:more than 30 points for both eyes each
5.Check that the ring detected point is not cut due to the headrest by tapping [Map].
61
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・TOPO Calibration
3)KRT/TOPO image 【R5】
1. Set the model eye for adjustment(R:5.0mm).
2. Align by moving the measurement unit. Make sure to be less than 0±2 for X, Y and Z coordinates at that time.
3. Take a measurement by tapping [Test].
※Be careful not to press [Set] by mistake
When you press[Set] by mistake, please calibrate from KRT/TOPO image 【R8】 again.
4.Confirm that the measure values of A1, A2, A3 displayed at the bottom of screen are within the
following standard value.
|A3-A1|≦0.005、|A3-A2|≦0.005
A1:1st ring measurement value / A2:2nd ring measurement value / A3:3rd ring measurement value
5.If It is out of standard value, adjust to be within the standard value by changing the value of [R1S_R]
and [R2S_R].
[R1S_R]:1st ring adjustment value (R5) / [R2S_R]:2nd ring adjustment value (R5)
62
The Challenge of Dreams
What‘s Next?
・TOPO Calibration
4)KRT/TOPO image 【R10】
1. Set the model eye for adjustment.(R:10.0mm).
2. Align by moving the measurement unit. Make sure to be less than 0±2 for X, Y and Z coordinates at that time.
3. Take a measurement by tapping [Test].
※Be careful not to press [Set] by mistake
When you press[Set] by mistake, please calibrate from KRT/TOPO image 【R8】 again.
4. Confirm that the measure values of A1, A2, A3 displayed at the bottom of screen are within
the following standard value.
|A3-A1|≦0.005、|A3-A2|≦0.005
A1:1st ring measurement value / A2:2nd ring measurement value / A3:3rd ring measurement value
5.If It is out of standard value, adjust to be within the standard value by changing the value of [R1L_R],
[R2L_R].
[R1L_R]:1 ring adjustment value (R10) / [R2L_R]:2 ring adjustment value (R10)
63
The Challenge of Dreams
What‘s Next?
・TOPO Calibration
5)KRT/TOPO radius
1.Set the attached model eye(R:8.0mm).
2.Align 5 places in order from 【a】 to 【e】 and tap 【Set】 respectively.
※When align, please pay attention to the sign of X, Y, Z coordinates.
3.After 【e】, finish is displayed on the upper of screen.
4. Finish the calibration by tapping the 【Exit】.
【a】Center 【b】left 【c】right
X:0±2、Y:0±2、Z:0±2 X:○○±2、Y:0±2、Z:0±2 X:-○○±2、Y:0±2、Z:0±2
64
The Challenge of
The Challenge of Dreams
What‘s Next?
Dreams
・TOPO Calibration
【d】up 【e】down
Finish
X:0±2、Y:-○○±2、Z:0±2 X:0±2、Y:○○±2、Z:0±2
65