You are on page 1of 1

Introduction

Laser “On” Symbol


The Model UT300A and UT300B Infrared Thermometer
(hereafter, the “Thermometer”) can determine the
UT300 SERIES surface temperature by measuring the amount of
HOLD SCAN or HOLD
SCAN
infrared energy radiated by the target’s surface. They
Temperature values for the
Non-contact Infrared Thermometer have different distance to spot (D:S) figure and different MAX/MIN
MAX and MIN
temperature range, details see the contents.
Battery indication
Operating Manual Figure 1. Symbols and Safety Markings
The Thermometers are non-contact infrared thermom-
eter with low consumption design so that they can be Display backlit
Caution
used for a longer time, which can solve the frequently
changing battery and low battery issues during Symbol (Celsius/Fahrenheit)
To avoid damaging the thermometer or the equipment
measurement. Intelligent design can make measure- How the Thermometer Works
under test protect them from the following:
ment easier and quicker. Infrared thermometers measure the surface tempera-
EMF (electro-magnetic fields) from arc welders, ture of an opaque object. The Thermometer’s optics
induction heaters, etc. sense infrared energy, which is collected and focused
Contacting Uni-Trend Static electricity. onto a detector. The Thermometer’s electronics then
To contact Uni-Trend. call (852) 2950 9168 or visit translate the information into a displayed temperature
Thermal shock (caused by large or abrupt ambient
Uni-Trend web site at www.uni-trend.com temperature changes – all 30 minutes from the reading which appears on the display. The laser is
Thermometer to stabilize before use). used for aiming purposes only.
Do not leave the Thermometer on or near objects of Operating the Thermometer
Safety Information high temperature. The Thermometer turns on when you press the trigger.
Warning The Thermometer turns off when no activity is detected
for 8 seconds.
A warning identifies conditions and actions that pose Features
hazards to the user. To avoid electrical shock or The Thermometer includes: To measure temperature, aim the Thermometer at the
personal injury, follow these guidelines: Single-spot Laser Sighting target, pull and hold the trigger. Release the trigger to
White colour Backlit Display hold a temperature reading.
Do not point laser directly at eye or indirectly off Current Temperature Plus MIN and MAX
reflective surfaces. Temperature Displays Be sure to consider distance-to-spot size ratio and filed
Before using the Thermometer inspect the case. Do Degree Celsius and Fahrenheit Temperature of view. The laser is used for aiming only.
not use the Thermometer if it appears damaged. Look Selectable
Button / Connector Description
for cracks or missing plastic. Battery power monitoring
Replace the battery as soon as the battery indicator Low Battery Display Press to toggle between
appears. Celsius and Fahrenheit.
Do not use the Thermometer if it operates abnormally. Display Press to toggle between
Protection may be impaired. When in doubt, have the MAX and MIN options.
Thermometer serviced. MAX and MIN values are
MAX/MIN
Do not operate the Thermometer around explosive displayed on the second-
gas, vapor, or dust. ary display.
To avoid a burn hazard, remember that highly Press to toggle between
reflective objects will often result in lower than actual turning on the laser and
temperature measurements. display backlight.
Do not use in a manner not specified by this manual
or the protection supplied by the equipment may be The sequence is:
impaired. , , , none.

P/N: 110401101601

Locating a Hot or Cold Spot Emissivity EN61326: 2006


To find a hot or cold spot, aim the Thermometer Emissivity describes the energy-emitting characteris- EN60825-1: 1994+A2: 2001+A1: 2002 Laser Safety Standard
outside the target area. Then, slowly scan across tics of materials. Most organic materials and painted Certification testing was conducted using a frequency range of 80 to 100MHz with instrument in three
the area with an up and down motion until you or oxidized surfaces have an emissivity of about 0.95. orientations.
located the hot or cold spot. See Figure 5.
If possible, to compensate for inaccurate readings that Specifications
may result from measuring shinymetal surfaces, cover Function UT300A UT300B
the surface to be measured with masking tape or flat Auto Power off
black paint (<150 / 302 ) and use the high emissiv- HOLD
ity setting. Allow time for the tape or paint to reach SCAN
the same temperatures as the surface beneath it. MIN
Measure the temperature of the tape or painted
surface. MAX
Laser turn off
Maintenance
Distance and Spot Size Selectable
As the distance (D) from the target being measured Emissivty 0.95 0.95
Changing the Battery
increases, the spot size (S) of the area measured by Temperature Range -18 ~ 280 -18 ~ 380
To install or change the 9V battery, open the battery
the unit becomes larger. The spot size indicates ±2 or ±2%.
compartment the battery. Maximum Measuring Accuracy
90% encircled energy. The maximum D:S is Assumes ambient operating temperature of 23 ±2 .
obtained when the Thermometer is 1000mm (100 in)
Cleaning the Lens Repeatability <±0.5 or <±0.5%
form the target resulting in a spot size of 20mm (2
Blow off loose particles using clean compressed air. Resolution
in). 0.1 0.1
Carefully wipe the surface with a moist cotton swab.
Response Time 500mS
The swab may be moistened with water.
White Display Backlit
Cleaning the Housing
Use soap and water on a damp sponge or soft cloth.
To avoid damage to the Thermometer, do not soak the ** END **
unit into water. This operating manual is subject to change without notice.
c Copyright 2008 Uni-Trend Group Limited.
Caution All rights reserved.
To avoid damaging the Thermometer, do NOT
submerge it in water. Manufacturer:
Uni-Trend Technology (Dongguan) Limited
Field of View Troubleshooting Dong Fang Da Dao
Make sure that the target is larger than the spot Bei Shan Dong Fang Industrial Development District
Symptom Problem Action Hu Men Town, Dongguan City
size. The smaller the target, the closer you should
OL (on display) Target temperature is over range Select target with specifications Guang Dong Province
be to it. Suggested distance less than 75% of the
-OL (on display) Target temperature is under range Select target with specifications China
theory value.
Postal Code: 523 925
Low Battery Replace Battery
Blank Display Possible dead battery Check and / or replace battery Headquarters:
Uni-Trend Group Limited
Laser does not work Low or dead battery Replace battery
Rm901, 9/F, Nanyang Plaza
Ambient temperature above Use in area with lower ambient temperature. 57 Hung To Road
40 (104 ) Kwun Tong
Kowloon, Hong Kong
CE Certification Tel: (852) 2950 9168 Fax: (852) 2950 9303
The Thermometer conforms to the following standards: Email: info@uni-trend.com http://www.uni-trend.com