1 (2020), 17–20
Article History: Received 24 February 2020, Revised 04 March 2020, Accepted 04 March 2020, Available Online 06 March 2020
Abstract
In this paper, we will discuss the making of the simple photometer instrument to measure the absorbance of natural pigment crude
extract. To make simple photometer, cuvette holder was printed using 3D Printing Machine, the development of simple photometer
instrument also uses some components such as LED light (633 nm) as the light source, Photodiode-TSL250 as a detector and a
variable resistor to set the initial intensity of the light source. In this development, Arduino Uno was also used as USB data
acquisition device to capture the signal from the instrument. The results of sample measurements between simple photometer
instrument and UV-Vis 1800 spectrophotometer showed 98% regression coefficient of determination and simple photometer has a
LOD and LOQ absorbance at 0.0267 and 0.0344. Simple photometer instrument was able to show a similar response as a
commercial instrument. Through the development of simple photometer, the student can easily understand the working principles
of a spectrophotometer in natural pigment analysis that difficult to studied in detailed before. The process of making a simple
photometer will be very useful for the learning process of students.
Vled Vsample
C T (%) A
(mV) (mV)
0% 1002.35 949.64 94.74 0.02
40% 1002.35 816.91 81,50 0.08
50% 1002.35 788.60 78.68 0.10
60% 1002.35 730.04 72.83 0.13
70% 1002.35 695.88 69.43 0.15 Figure 3. Regression of simple photometer and
commercial instrument for different concentration
80% 1002.35 651.96 65.04 0.18
90% 1002.35 520.20 51.90 0.28 After measuring the absorbance of the sample, the next
100% 1002.35 476.28 47.52 0.32 step is the validation of the photometer which is
conducted by measuring the sensitivity of the instrument.
LOD and LOQ are obtained by calculating the average
However, if the regression coefficient of determination absorbance of the blank and its standard deviation. Table
between the absorbance results of the two instruments is 2 present LOD and LOQ calculation.
calculated, then the obtained R2 is 98%. This shows that
Table 2. Simple Photometer LOD and LOQ Calculation
the absorbance measurement using a photometer has a
linear relationship to a commercial spectrophotometer.
The average absorbance of the
Regression between simple photometer and commercial
blank 0.0234
instrument present in Figure 2.
The standard deviation of the
Figure 3 presents a regression curve for measuring absorbance of the blank 1.095×10-3
simple and commercial photometers. The R2 value LOD (Absorbance) 0.0267
generated from the linear regression equation using a
LOQ (Absorbance) 0.0344
simple photometer is 0.9269 while using a commercial
photometer is 0.9745. The low value of R2 simple
photometer shows that the data obtained is less
In table 2, it can be seen that the detection limit or the
accurate. Some factors which are thought to cause
lowest sample concentration that can be detected by a
inaccuracies in measurement are errors in sample
photometer instrument is 2.67%. This means that if the
dilution, inaccurate adjustment of the initial intensity of Spirulina sp crude extract sample is less than 2.67%, the
LEDs which are regulated using a variable resistor, and photometer instrument detector is not able to detect its
different light sources between a simple photometer and
presence. For the LOQ, the simple photometer
a commercial photometer.
instrument has a minimum quantification or concentration
limit that can be detected by the instrument and is
statistically trusted at 3.44%.
Adhiwibawa, et al. (2020)
Content from this work may be used under the terms of the Creative Commons Attribution 4.0 license
19
Indones. J. Nat. Pigm., Vol. 02, No. 1 (2020), 17–20
CONCLUSION [10] Li, Y. C., Melenbrink, E. L., Cordonier, G. J., Boggs, C., Khan, A.,
Isaac, M. K., ... & Kreuter, R. A. (2018). An easily fabricated low-
The results of sample measurements between simple cost potentiostat coupled with user-friendly software for
photometer instrument and UV-Vis 1800 introducing students to electrochemical reactions and
spectrophotometer showed 98% coefficient of electroanalytical techniques.
determination and simple photometer has a LOD and [11] Baird, W. H., Hack, W. N., Tran, K., Vira, Z., & Pickett, M. (2011).
The Light-Emitting Diode as a Light Detector. The Physics
LOQ absorbance at 0.0267 and 0.0344. Simple Teacher, 49(3), 171-174.
photometer instrument was able to show a similar [12] Cao, T. (2016). Development of Inexpensive and Portable Devices
response as a commercial instrument. Through the for Chemical Measurements and Environmental Analysis (Doctoral
development of simple photometer, the students could dissertation).
[13] Grasse, E. K., Torcasio, M. H., & Smith, A. W. (2016). Teaching
easily understand the working principles of a UV–Vis spectroscopy with a 3D-printable smartphone
spectrophotometer in natural pigment analysis that spectrophotometer. Journal of Chemical Education, 93(1), 146-
difficult to studied in detailed before. 151.
[14] Albert, D. R., Todt, M. A., & Davis, H. F. (2012). A low-cost
quantitative absorption spectrophotometer. Journal of Chemical
REFERENCES Education, 89(11), 1432-1435.
[1] Kvittingen, EV, Lise, Birte J., and Richard V. Simple and [15] Dominguez, R. B., Orozco, M. A., Chávez, G., & Márquez-Lucero,
Inexpensive UV-Vis photometer Both Using LEDs as Light Source A. (2017). The evaluation of a low-cost colorimeter for glucose
and Detector. J. Chem. Educ, 2014. detection in salivary samples. Sensors, 17(11), 2495.
[2] McClain, Robert L. Construction of a photometer as Instructional [16] Magro, A., Silva, M., Sousa, G., Cardoso, F., Guimarães, G.,
Tool for Electronics and Instrumentation. J. Chem. Educ, 2014. Muniz, S., & Simões, R. Development of a low-cost colorimeter-
[3] Nandiyanto, A. B. D., Zaen, R., Oktiani, R., Abdullah, A. G., & like for undergraduate classes using microcontroller board and
Riza, L. S. (2018). A simple, rapid analysis, portable, low-cost, and RGB LED. Educación Química, 31(1), 36-48.
Arduino-based spectrophotometer with white LED as a light [17] Malhotra, A., Sandhu, A. S., & Singh, M. (2019, November).
source for analyzing solution concentration. Telkomnika, 16(2), Development of RGB LED Based Compact Multipurpose
580-585. Colorimeter for Hemoglobin Estimation of Blood. In 2019 1st
[4] Yang, B., Patsavas, M. C., Byrne, R. H., & Ma, J. (2014). International Conference on Electrical, Control and
Seawater pH measurements in the field: a DIY photometer with Instrumentation Engineering (ICECIE) (pp. 1-8). IEEE.
0.01 unit pH accuracy. Marine chemistry, 160, 75-81. [18] Pokrzywnicka, M., Koncki, R., & Tymecki, L. (2009). A very simple
[5] Tan, A. F. Y., Talik, N. A., Yap, B. K., & Krishnan, P. S. (2017, photometer based on paired-emitter-detector diodes. Chemia
December). Prospects of low-cost photometers for colorimetric Analityczna, 54(3), 427.
serum iron concentration determination. In International [19] Fonseca, A., & Raimundo Jr, I. M. (2007). A simple method for
Conference for Innovation in Biomedical Engineering and Life water discrimination based on an light emitting diode (LED)
Sciences (pp. 165-169). Springer, Singapore. photometer. Analytica chimica acta, 596(1), 66-72.
[6] Wilson, M. V., & Wilson, E. (2017). Authentic performance in the [20] Porter, LA, Benjamin MW, Mazin HH, and Richard FD Colorimeter
instrumental analysis laboratory: Building a visible 3D Printed User Friendly Models for Student Exploration of
spectrophotometer prototype. Journal of Chemical Education, Instrument Design and Performance. 2011.
94(1), 44-51. [21] Grinias, J. P., Whitfield, J. T., Guetschow, E. D., & Kennedy, R. T.
[7] Kutschera, A., & Lamb, J. J. (2018). Light Meter for Measuring (2016). An inexpensive, open-source USB Arduino data
Photosynthetically Active Radiation. American Journal of Plant acquisition device for chemical instrumentation.
Sciences, 9(12), 2420. [22] Mabbott, G. A. (2014). Teaching electronics and laboratory
[8] Paiva, E. M., Rohwedder, J. J. R., Pasquini, C., & Pereira, C. F. automation using microcontroller boards. Journal of Chemical
(2019). Method for building a portable near infrared photometer Education, 91(9), 1458-1463.
based on LEDs and interference filters chosen by a spectral [23] Riyanto. 2014. Validation & Verification Test Method: According to
variable selection algorithm. Microchemical Journal, 146, 842-849. ISO / IEC 17025 Testing and Calibration Laboratories.
[9] Walkowiak, M., & Nehring, A. (2016). Using chemduino, excel, and Yogyakarta: Deepublish.
Powerpoint as tools for real-time measurement representation in
class.
Abstrak
Pada jurnal ini akan dibahas mengenai pembuatan instrumen fotometer sederhana untuk mengukur absorbansi molekul, mulai dari pembuatan
tempat kuvet menggunakan teknologi 3D Printing, menyusun instrumen fotometer menggunakan lampu LED (633 nm) sebagai sumber
cahaya, Photodiode-TSL250 sebagai detektor dan variabel resistor untuk mengatur intensitas awal sumber cahaya. Selain itu akan dibahas
pula mengenai penggunaan alat Arduino Uno sebagai USB Data Acquisition Devices untuk menangkap sinyal dalam bentuk tegangan dari
instrumen yang dibuat, dimana akan dihubungkan juga dengan penggunaan software Arduino sebagai media untuk menampilkan hasil.
Sebagai uji coba instrumen yang telah dibuat, dilakukan pengukuran absorbansi pada ekstrak pigmen kasar Spirulina dengan menggunakan
persen konsentrasi yang berbeda-beda. Hasil pengukuran absorbansi yang didapat dibandingkan dengan instrumen Spektrofotometer
komersial Shimadzu UV-1800 sehingga dapat dilakukan validasi metode terhadap instrumen fotometer buatan sendiri.
Kata kunci: kata kunci 1, kata kunci 2 (sampai kata kunci kelima)