Simple Photometer Development For Educational Purposes in Natural


Pigment Analysis

Jovine Marcella Kurniawana, and Marcelinus A.S. Adhiwibawaa*


a
Ma Chung Research Center for Photosynthetic Pigments, Universitas Ma Chung, Malang 65151, Jawa Timur,
Indonesia
*Corresponding Authors: marcelinus.setya@machung.ac.id

Article History: Received 24 February 2020, Revised 04 March 2020, Accepted 04 March 2020, Available Online 06 March 2020
Abstract
In this paper, we will discuss the making of the simple photometer instrument to measure the absorbance of natural pigment crude
extract. To make simple photometer, cuvette holder was printed using 3D Printing Machine, the development of simple photometer
instrument also uses some components such as LED light (633 nm) as the light source, Photodiode-TSL250 as a detector and a
variable resistor to set the initial intensity of the light source. In this development, Arduino Uno was also used as USB data
acquisition device to capture the signal from the instrument. The results of sample measurements between simple photometer
instrument and UV-Vis 1800 spectrophotometer showed 98% regression coefficient of determination and simple photometer has a
LOD and LOQ absorbance at 0.0267 and 0.0344. Simple photometer instrument was able to show a similar response as a
commercial instrument. Through the development of simple photometer, the student can easily understand the working principles
of a spectrophotometer in natural pigment analysis that difficult to studied in detailed before. The process of making a simple
photometer will be very useful for the learning process of students.

Simple photometer development: A simple photometer prototype


consisting of a light source(LSR), cuvette holder(C), light
sensor(LSN) and microcontroller(M) can be used as a medium for
learning instrumentation in detail. Students can build their own
instruments with low costs in a short time. It is expected that by
interacting in the development of a simple photometer students can
understand the work process of a spectrophotometer in natural
pigment analysis.

Prototype of simple photometer instrument


© 2020 MRCPP Publishing. All rights reserved.
http://doi.org/10.33479/ijnp.2020.02.1.17
Keywords: Photometer, Natural Pigment, Arduino Uno

INTRODUCTION that experiences maximum absorption, while


Spectroscopy is an analytical method that used to test monochromatic light can make data accurate because it
material based on the light emitted, absorbed or reflected has a narrow wavelength range [12,13,14,15].
by the material [1,2]. The instrument used in The importance of spectrophotometers in the
spectroscopy is a spectrophotometer [4,5]. One type of analysis is often not followed by knowledge of how the
spectrophotometer is a visible light spectrophotometer spectrophotometer works. This development was
[6,7,8]. The tool can be used to determine the conducted with a principle that resembles visible light
characteristics of a material in solution, in the form of spectrophotometers, using simple tools and materials
maximum absorption wavelength, absorbance value and such as LEDs, cuvette holders, light sensors, Arduino
concentration of material in solution, by giving light Uno and computers [16,17,18,19]. Furthermore, in this
radiation to the solution and then measuring the intensity study students can take measurements of natural
of the light absorbed by the solution[9,10,11]. Radiation pigments using the Lambert-Beer principle on a simple
can be polychromatic and monochromatic light; photometer instrument.
Polychromatic light is used to select the spectrum of light
Adhiwibawa, et al. (2020)
Content from this work may be used under the terms of the Creative Commons Attribution 4.0 license
17
Indones. J. Nat. Pigm., Vol. 02, No. 1 (2020), 17–20

EXPERIMENTAL Absorbance measurements and Method Validation


Cuvette Holder Design Absorbance measurement was conducted by
Cuvette holder design was created with the 123D determining the initial intensity of LED light. After that, the
software. Before designing the cuvette holder, solvent absorbance also measured to make sure the
measurements of length, width, and height cuvette to be absorbance of the sample is not affected by the solvent
used was made. To place the light source and the light that used. Then, the absorbance of the sample was
detector precisely it is also necessary to measure the measured. Measurement of the absorbance can be done
hole position in a straight parallel line. After getting the by using the equation:
appropriate design, then the design was printed into
three-dimensional cuvette holder with a three-
dimensional printer. Cuvette holder was printed using a
black colored material in order to avoid light reflection Vsample is a sample light signal detected by Arduino Uno,
that may affect the absorbance values of the Vled is a an initial light source signal and T is light
compound[20]. transmittance. Samples were also measured with a
commercial spectrophotometer Shimadzu UV-1800 as
LED lamp as the light source instrument performance comparison.
LED was used as a light source in the simple LOD (limit of detection) and LOQ (limit of Quantification)
photometer, In this experiment the wavelength of LED is was also conducted to know the sensitivity of the simple
633nm. This light source is used as the sample to be photometer. The limit of detection (LOD) is the lowest
measured is crude extracts of Spirulina sp which have concentration of the sample can be measured and limits
been known to contain chlorophyll pigment. The green of quantitative measurement (LOQ) is the lowest
color of the pigment can absorb red light at wavelength concentration of the sample can be measured
region around 600-700 nm so that light with a wavelength absorbance by means of acceptable precision and
of 633 nm will be suitable for use. accuracy in statistics[23]. Here is the equation used to
determine the sensitivity:
The Light Detector IC-TSL250

The detector is an important part of the photometer


development. The detectors will capture the light signal
and will transform the light signal into an electrical Sensitivity : LOD or LOQ
current. The higher the intensity of light is detected, the k : 3 to 10 for the LOD and LOQ
greater the current generated. In the experiment, IC- SD Blank :The standard deviation of the
TSL250 was used as the detector. absorbance of the blank
μ : Average absorbance of the blank
Arduino Uno Microcontroller
Arduino Uno is a microcontroller that can be used to
convert analog current signals into digital signals. Analog
RESULTS AND DISCUSSION
pins from Arduino Uno connected to IC-TSL250. Then
Arduino Uno ADC converted analog current signal that Design of Simple Photometers
displayed on Arduino IDE by Using PC[21,22]. The design of the photometer using the 123D application
has resulted in a functional cuvette holder model.
Instrument Schematic of Simple Photometer Educational photometer consists of LED light sources,
Once all the equipment is ready, the photometer cuvette holders, light detectors and microcontrollers.
instrument coupled with a scheme as shown below: Educational photometer measurements consist of solvent
and sample measurements. Making a photometer in
accordance with a single beam system. This simple
spectrophotometry system contained IC TSL-250 which
was installed on the cuvette holder. Educational
photometer has dimensions of 5.5 × 1 × 5 cm.
Educational photometers operate at 5 Volts with a
current of 30 mA. The LED light source uses a
Natural Pigment Sample Preparation wavelength of 663 nm adjusted to the chlorophyll light
A total of 0.1 grams of Spirulina sp powder was weighed, absorption as the target compound. Black cuvette holder
CaCO3 and Sodium Ascorbate was added as a to reduce interference from outside light. The light signal
neutralizing agent and antioxidants. Samples were from the TSL-250 IC sensor is processed by Arduino
extracted with 10 ml of 100% methanol solvent, UNO and displayed at the computer layer using the
homogenized with a vortex for 2 minutes, sonicated two Arduino IDE software interface. The picture below is a
times for 2 minutes, and centrifuged for 2 minutes at a simple photometer instrument that used to measure the
speed of 10,000 rpm. The extraction is done in duplicate, absorbance of Spirulina sp crude extract sample with
and the supernatant obtained is stored in a vial and various concentration.
diluted at least 25 times as the stock solution for
absorbance measurement using the simple photometer.

Adhiwibawa, et al. (2020)


Content from this work may be used under the terms of the Creative Commons Attribution 4.0 license
18
Indones. J. Nat. Pigm., Vol. 02, No. 1 (2020), 17–20

Figure 1. Simple Photometer Prototype

Quantitative measurement of Simple


Photometers
Table 1 presents observation data obtained by using
simple photometer, where absorbance shows a positive
trend along with an increase in sample concentration.
The intensity value of the photometer instrument has Figure 2. Regression between simple photometer and
been converted from the digital number to voltage. The commercial instrument
absorbance results obtained by the photometer show a
difference with the absorbance results of commercial
spectrophotometers. This is influenced by the differences
in the light source and the detector used by the device,
the difference in the wavelength of the light source, and
the inaccuracy in diluting stock solutions.

Table 1. Simple Photometer Observation Data

Vled Vsample
C T (%) A
(mV) (mV)
0% 1002.35 949.64 94.74 0.02
40% 1002.35 816.91 81,50 0.08
50% 1002.35 788.60 78.68 0.10
60% 1002.35 730.04 72.83 0.13
70% 1002.35 695.88 69.43 0.15 Figure 3. Regression of simple photometer and
commercial instrument for different concentration
80% 1002.35 651.96 65.04 0.18
90% 1002.35 520.20 51.90 0.28 After measuring the absorbance of the sample, the next
100% 1002.35 476.28 47.52 0.32 step is the validation of the photometer which is
conducted by measuring the sensitivity of the instrument.
LOD and LOQ are obtained by calculating the average
However, if the regression coefficient of determination absorbance of the blank and its standard deviation. Table
between the absorbance results of the two instruments is 2 present LOD and LOQ calculation.
calculated, then the obtained R2 is 98%. This shows that
Table 2. Simple Photometer LOD and LOQ Calculation
the absorbance measurement using a photometer has a
linear relationship to a commercial spectrophotometer.
The average absorbance of the
Regression between simple photometer and commercial
blank 0.0234
instrument present in Figure 2.
The standard deviation of the
Figure 3 presents a regression curve for measuring absorbance of the blank 1.095×10-3
simple and commercial photometers. The R2 value LOD (Absorbance) 0.0267
generated from the linear regression equation using a
LOQ (Absorbance) 0.0344
simple photometer is 0.9269 while using a commercial
photometer is 0.9745. The low value of R2 simple
photometer shows that the data obtained is less
In table 2, it can be seen that the detection limit or the
accurate. Some factors which are thought to cause
lowest sample concentration that can be detected by a
inaccuracies in measurement are errors in sample
photometer instrument is 2.67%. This means that if the
dilution, inaccurate adjustment of the initial intensity of Spirulina sp crude extract sample is less than 2.67%, the
LEDs which are regulated using a variable resistor, and photometer instrument detector is not able to detect its
different light sources between a simple photometer and
presence. For the LOQ, the simple photometer
a commercial photometer.
instrument has a minimum quantification or concentration
limit that can be detected by the instrument and is
statistically trusted at 3.44%.
Adhiwibawa, et al. (2020)
Content from this work may be used under the terms of the Creative Commons Attribution 4.0 license
19
Indones. J. Nat. Pigm., Vol. 02, No. 1 (2020), 17–20

CONCLUSION [10] Li, Y. C., Melenbrink, E. L., Cordonier, G. J., Boggs, C., Khan, A.,
Isaac, M. K., ... & Kreuter, R. A. (2018). An easily fabricated low-
The results of sample measurements between simple cost potentiostat coupled with user-friendly software for
photometer instrument and UV-Vis 1800 introducing students to electrochemical reactions and
spectrophotometer showed 98% coefficient of electroanalytical techniques.
determination and simple photometer has a LOD and [11] Baird, W. H., Hack, W. N., Tran, K., Vira, Z., & Pickett, M. (2011).
The Light-Emitting Diode as a Light Detector. The Physics
LOQ absorbance at 0.0267 and 0.0344. Simple Teacher, 49(3), 171-174.
photometer instrument was able to show a similar [12] Cao, T. (2016). Development of Inexpensive and Portable Devices
response as a commercial instrument. Through the for Chemical Measurements and Environmental Analysis (Doctoral
development of simple photometer, the students could dissertation).
[13] Grasse, E. K., Torcasio, M. H., & Smith, A. W. (2016). Teaching
easily understand the working principles of a UV–Vis spectroscopy with a 3D-printable smartphone
spectrophotometer in natural pigment analysis that spectrophotometer. Journal of Chemical Education, 93(1), 146-
difficult to studied in detailed before. 151.
[14] Albert, D. R., Todt, M. A., & Davis, H. F. (2012). A low-cost
quantitative absorption spectrophotometer. Journal of Chemical
REFERENCES Education, 89(11), 1432-1435.
[1] Kvittingen, EV, Lise, Birte J., and Richard V. Simple and [15] Dominguez, R. B., Orozco, M. A., Chávez, G., & Márquez-Lucero,
Inexpensive UV-Vis photometer Both Using LEDs as Light Source A. (2017). The evaluation of a low-cost colorimeter for glucose
and Detector. J. Chem. Educ, 2014. detection in salivary samples. Sensors, 17(11), 2495.
[2] McClain, Robert L. Construction of a photometer as Instructional [16] Magro, A., Silva, M., Sousa, G., Cardoso, F., Guimarães, G.,
Tool for Electronics and Instrumentation. J. Chem. Educ, 2014. Muniz, S., & Simões, R. Development of a low-cost colorimeter-
[3] Nandiyanto, A. B. D., Zaen, R., Oktiani, R., Abdullah, A. G., & like for undergraduate classes using microcontroller board and
Riza, L. S. (2018). A simple, rapid analysis, portable, low-cost, and RGB LED. Educación Química, 31(1), 36-48.
Arduino-based spectrophotometer with white LED as a light [17] Malhotra, A., Sandhu, A. S., & Singh, M. (2019, November).
source for analyzing solution concentration. Telkomnika, 16(2), Development of RGB LED Based Compact Multipurpose
580-585. Colorimeter for Hemoglobin Estimation of Blood. In 2019 1st
[4] Yang, B., Patsavas, M. C., Byrne, R. H., & Ma, J. (2014). International Conference on Electrical, Control and
Seawater pH measurements in the field: a DIY photometer with Instrumentation Engineering (ICECIE) (pp. 1-8). IEEE.
0.01 unit pH accuracy. Marine chemistry, 160, 75-81. [18] Pokrzywnicka, M., Koncki, R., & Tymecki, L. (2009). A very simple
[5] Tan, A. F. Y., Talik, N. A., Yap, B. K., & Krishnan, P. S. (2017, photometer based on paired-emitter-detector diodes. Chemia
December). Prospects of low-cost photometers for colorimetric Analityczna, 54(3), 427.
serum iron concentration determination. In International [19] Fonseca, A., & Raimundo Jr, I. M. (2007). A simple method for
Conference for Innovation in Biomedical Engineering and Life water discrimination based on an light emitting diode (LED)
Sciences (pp. 165-169). Springer, Singapore. photometer. Analytica chimica acta, 596(1), 66-72.
[6] Wilson, M. V., & Wilson, E. (2017). Authentic performance in the [20] Porter, LA, Benjamin MW, Mazin HH, and Richard FD Colorimeter
instrumental analysis laboratory: Building a visible 3D Printed User Friendly Models for Student Exploration of
spectrophotometer prototype. Journal of Chemical Education, Instrument Design and Performance. 2011.
94(1), 44-51. [21] Grinias, J. P., Whitfield, J. T., Guetschow, E. D., & Kennedy, R. T.
[7] Kutschera, A., & Lamb, J. J. (2018). Light Meter for Measuring (2016). An inexpensive, open-source USB Arduino data
Photosynthetically Active Radiation. American Journal of Plant acquisition device for chemical instrumentation.
Sciences, 9(12), 2420. [22] Mabbott, G. A. (2014). Teaching electronics and laboratory
[8] Paiva, E. M., Rohwedder, J. J. R., Pasquini, C., & Pereira, C. F. automation using microcontroller boards. Journal of Chemical
(2019). Method for building a portable near infrared photometer Education, 91(9), 1458-1463.
based on LEDs and interference filters chosen by a spectral [23] Riyanto. 2014. Validation & Verification Test Method: According to
variable selection algorithm. Microchemical Journal, 146, 842-849. ISO / IEC 17025 Testing and Calibration Laboratories.
[9] Walkowiak, M., & Nehring, A. (2016). Using chemduino, excel, and Yogyakarta: Deepublish.
Powerpoint as tools for real-time measurement representation in
class.

Abstrak
Pada jurnal ini akan dibahas mengenai pembuatan instrumen fotometer sederhana untuk mengukur absorbansi molekul, mulai dari pembuatan
tempat kuvet menggunakan teknologi 3D Printing, menyusun instrumen fotometer menggunakan lampu LED (633 nm) sebagai sumber
cahaya, Photodiode-TSL250 sebagai detektor dan variabel resistor untuk mengatur intensitas awal sumber cahaya. Selain itu akan dibahas
pula mengenai penggunaan alat Arduino Uno sebagai USB Data Acquisition Devices untuk menangkap sinyal dalam bentuk tegangan dari
instrumen yang dibuat, dimana akan dihubungkan juga dengan penggunaan software Arduino sebagai media untuk menampilkan hasil.
Sebagai uji coba instrumen yang telah dibuat, dilakukan pengukuran absorbansi pada ekstrak pigmen kasar Spirulina dengan menggunakan
persen konsentrasi yang berbeda-beda. Hasil pengukuran absorbansi yang didapat dibandingkan dengan instrumen Spektrofotometer
komersial Shimadzu UV-1800 sehingga dapat dilakukan validasi metode terhadap instrumen fotometer buatan sendiri.

Kata kunci: fotometer, pigmen alami, arduino uno