Professional Documents
Culture Documents
10
11
12
13
14
15
Note: Third column is to be filled in, when the document is being updated/modified
CONTENTS
9.0 EXAMINATION..................................................................................................... 17
1.0 SCOPE
1.1 This procedure using phased array ultrasonic examination using Olympus NDT
Omniscan MX2 Phased Array System using semi-auto scanner for the NDT personnel
performing examination on materials and weld.
1.2 This procedure shall in accordance with Article 4, Mandatory Appendix VII
“Ultrasonic Examination Requirements for a Workmanship Based Acceptance
Criteria” of ASME Section V and DNVGL-ST-F101_2017.
1.3 This procedure to be used for detecting, locating and evaluating indications within the
welds and heat affected zone using the contact inspection technique for carbon steel
and alloy steel for Pipeline 26’’x23.8 &19.1mm and 18’’x17.5 mm.
2.0 REFERENCE
- ASME Section V, Article 1 and Article 4, 2019 edition
- SNT-TC-1A, 2016 edition;
- DNVGL-ST-F101_2017: Submarine Pipeline Systems
- ISO 9712: Non-destructive testing - Qualification and certification of NDT –
personnel;
- 191364-TPGM-15-SPE-0002_Spec for Pipeline Welding and NDT-SV to
NCS2 Rev.0
- 191364-TPGM-15-SPE-0003_Specification for Pipeline Welding and NDT-
TU to SV_1
- AlphaNDT’s Written Practice No: 8408-30-01 Rev.10E.
Procedure
No. Requirement Non Remark
Essential Paragraph No.
essential
Personnel performing NDT and interpretation of test results shall be certified to Level
2 by a Certification Body or Authorised Qualifying Body in accordance with ISO 9712
or the ASNT Central Certification Program (ACCP). Personnel qualification to an
employer based qualification scheme as SNT-TC-1A may be accepted if the employer’s
written practice is reviewed and found acceptable and the Level 3 is ASNT Level III or
ACCP Professional Level III and certified in the applicable method.
5.0 EQUIPMENT
All the equipment mentioned in the appendix 2 shall be checked by all parties at
site before the work's commencement and sign off.
A) PROBES
- The application requirements will dictate the design of the phased array probe used.
Phased array probes may be used with a removable or integral wedge, delay line.
- Phased array probes used for weld examination shall be of 1D design.
- The number of elements in the phased array probe and the element dimensions and
pitch shall be selected based on the application requirements and the manufacturer’s
recommended limitations.
- The probe selected shall not have more elements than the number of elements
addressable by the pulser-receivers available in the phased array instrument being
used.
- When refracting wedges are used to assist beam steering, the natural incident angle
of the wedge shall be selected such that the angular sweep range of the examination
technique used does not exceed the manufacturer’s recommended limits for the
probe and mode (compression or transverse) used.
- Refracting wedges used on curved surfaces shall require contouring to match the
surface curvature if the curvature causes a gap between the wedge and examination
surface exceeding 0.5 mm (0.020 in.) at any point.
Where:
A = length of search unit footprint during circumferential scanning or the width
when scanning in the axial direction, in. (mm)
D = the component diameter at inspection surface (I.D./O.D.), in. (mm)
The footprint is defined as the physical dimension of the search unit in the curved
direction of the component.
(b) The search unit contoured dimension shall be selected from the tables in (1) and
(2) below, and shall be determined using the same component dimension from
which the examination is being performed (I.D. or O.D.).
(1 ) Maximum contour for examinations performed from O.D.
TomoView
Feature
Analysis
Metric and Imperial Units
Multi-group Combined Display (PA, UT, and TOFD)
Ability to Zoom in/out in Display
Predefined Weld Overlay Display
Selectable Information Groups (Readings)
Offline Data Management and Processing
Indication Table Tools
Built-in Report Generator (Customizable)
Ability to Modify/Create Color Palette
Offline Thickness C-Scan Tool
Offline TOFD Calibration
TOFD Lateral Wave Synchronization Tool
TOFD Lateral Wave Removal Tool
TOFD SAFT Tool
Volumetric Merge Tool (Automatic or Manual)
Software Gain Adjustment
Custom Layouts (Creation and Saving)
Zone Tool for Statistical Measurements
3D Cursor
Ability to Open Multiple Files Simultaneously
Data File Merging
C-Scan Data Merging
Signal to Noise Ratio (SNR) Analysis Tool
Binarizer Processing Tool
Matrix Filters Tool
Backwall Following C-Scan Tool
Hysteresis Correction
Ability to Export Data Group in .txt File
FFT Calculation
Microsoft Excel® Exchanges for Custom Readings
Acoustic Field Simulation (AFiSiMO)
Polar View Display
A-Scan Resynchronization Tool
5.4 COUPLANT
6.1 REFLECTORS
Side drilled holes shall be used to establish primary reference responses of the
equipment according with table D-1-DNVGL-OS-F101_2017.
Calibration block 26”x20mm is machined form pipe 26’’x23.8mm shall cover for pipe
26’’x23.8mm,19.1mm and 18’’17.5mm (See figure 1)
The reference block shall normally be manufactured from the actual material tested.
The finish on the scanning surfaces of the block shall be representative of the scanning
surface finishes on the component to be examined.
7.0 TECHNIQUES
The techniques described in this procedure are intended for applications where probes
are used to produce:
- Angle beam longitudinal waves, where both refracted longitudinal and shear
waves are present in the material under examination. Dual matrix probes with
high refract angle shall be used (Dual Matrix A17 and A27 probes).
- Angle beam shear waves, where incident angles in wedges produce only
refracted shear waves in the material under examination are generally termed
angle beam examinations. (linear A12,A15 and A32 probes shall be used)
Detail of beam angle, mode of wave propagation in material shall be detailed in
plan scan.
8.0 CALIBRATION
A calibration check shall be performed at intervals not to exceed one month or prior to
first use thereafter, by moving the encoder a minimum distance of 20 in. (500 mm). The
display distance shall be within 1% of the actual distance moved.
- Enter the reference amplitude. (default 80%) All focal laws will be adjusted to
this amplitude by addition of a focal law gain offset after sensitivity wizard
completion.
- Enter tolerance of 5%. The tolerance is a visual indicator for amplitude
verification after the sensitivity calibration correction
- Select the last focal law to be calibrated for sensitivity. When there are no
obstructions from corners or adjacent calibration targets the entire sector scan
can be calibrated at one time
- Set the start and width of gate A to ensure that the calibration target can be
detected by all focal laws in the S-scan.
- Set the gate A threshold as low as possible to ensure the last focal law can be
detected. Gate thresholds below approximately 10% may have an amplitude
error beyond the tolerance when the focal law gain offset is calculated to correct
to 80%.
- Gain compensation is a tool that is used to ease calibration, The longer the sound
path to the calibration target, the more gain is added
- Move the probe across the calibration block with consistent pressure and
coupling.
- As the calibration target travels through the gate of each focal law the green
envelope records the maximum amplitude detected.
- The MX2 will calculate the required gain offset in dB for each focal law to
correct to 80%. Focal laws with amplitude over 80% but less than 100% will
receive a negative gain offset.
- Successful sensitivity calibration of the S-scan is visualized by repeating the
probe movement over the calibration block with all focal laws corrected to 80%
amplitude within the tolerance.
- If any focal law is over 100% after calibration the complete process must be
repeated by selecting Restart. Saturated focal laws cannot be corrected.
- Calibrate may be selected repeatedly until all laws are within tolerance.
- Adjust the general gain if needed. Unlike the sensitivity calibration, the TCG
wizard will not allow any point over 80% amplitude to be corrected.
- Adjust the start and range to ensure sufficient range for all required TCG points
to be detected by all focal laws in the S-scan.
- Set the start and width of gate A to ensure that the calibration target can be
detected by all focal laws in the S-scan.
- Set the gate A threshold as low as possible to ensure the last focal law can be
detected. Gate thresholds below approximately 10% may have an amplitude
error beyond the tolerance when the focal law gain offset is calculated to correct
to 80%
- Move the probe across the calibration block with consistent pressure and
coupling.
- As the calibration target travels through the gate of each focal law the green
envelope records the maximum amplitude.
- Successful TCG sensitivity calibration of the S-scan is visualized by repeating
the probe movement over the calibration block with all focal laws corrected to
80% amplitude within the tolerance
- Re-check sensitivity setting on notch of calibration block for all angle beam so
that the response of reflector shall not less than 100% TGC.
Element 1 2 3 4 5 6 7 8 9 10 … … 59 61 62 63 64
Gain
Active
inactive
Table 1: Probe Element Activity Chart
A calibration check on at least one of the reflectors in the basic calibration block or a
check using a simulator shall be performed at the completion of each examination or
series of similar examinations, and when examination personnel (except for automated
equipment) are changed. The distance range and sensitivity values recorded shall satisfy
the requirements 8.7.3.
8.7.3. Confirmation Acceptance Values.
a. Distance Range Points.
If any distance range point has moved on the sweep line by more than 10% of the
distance reading or 5% of full sweep, whichever is greater, correct the distance range
calibration and note the correction in the examination record. All recorded indications
since the last valid calibration or calibration check shall be reexamined and their values
shall be changed on the data sheets or re-recorded.
b. Sensitivity Settings.
If any sensitivity setting has changed by more than 20% or 2 dB of its amplitude, correct
the sensitivity calibration and note the correction in the examination record. If the
sensitivity setting has decreased, all data sheets since the last valid calibration check
shall be marked void and the area covered by the voided data shall be re-examined. If
the sensitivity setting has increased, all recorded indications since the last valid
calibration or calibration check shall be re-examined and their values shall be changed
on the data sheets or re-recorded.
9.0 EXAMINATION
(g) When multi scans are required to cover full length of weld, overlap of scanning shall
be 50 mm or larger.
These typically show multiple facets and edges visible in the A-scan and S-scan. There
is a distinct start and stop on the A-scan, and a significant echo-dynamic travel to the
signal as the probe is moved in and out from the weld (if the crack has significant
vertical extent). The reflector is usually detectable and can be plotted from both sides
of the weld. The reflector should plot to the correct I.D. depth reference or depth
reading, as shown in Picture 1.
LOF (Lack of Fusion) plots correctly on the weld fusion line, either through geometrical
plotting or via weld overlays. There may be a significantly different response from each
side of the weld. LOF is usually detected by several of the angles in an S-scan from the
same position. The A-scan shows a fast rise and fall time with short pulse duration
indicative of a planar flaw. There are no multiple facets or tips. Skewing the probe
slightly does not produce multiple peaks or jagged facets as in a crack. There may be
mode converted multiple signals that rise and fall together and maintain equal
separation. Picture 2 shows an example.
Porosity shows multiple signal responses, varying in amplitude and position. The
signals plot correctly to the weld volume. The signals’ start and stop positions blend
with the background at low amplitude. The A-scan slow rise and fall time with long
pulse duration is indicative of a non-planar flaw. Porosity may or may not be detected
from both sides of the weld, butshould be similar from both sides. Picture 3 shows an
example of porosity.
Toe cracks typically show multiple facets and edges visible in the A-scan and S-scan.
There is significant echo-dynamic travel to the signal as the probe is moved in and out
from the weld. The reflector is usually detectable and can be plotted from at the correct
O.D. depth reference line or depth reading. Normally, toe cracks are best characterized
on S-scans and lower angle E-scan channels. Picture 4 shows an example.
Incomplete Penetration (IP) typically shows high amplitude signals with significant
echo-dynamic travel or travel over the I.D. skip line. IP will typically respond and plot
from both sides of the weld in common weld geometries near centerline reference
indicators. Generally, IP is detected on all channels, with highest amplitude on a high
angle E-scan. The A-scan shows a fast rise and fall time with short pulse duration
indicative of a planar flaw. Picture 4 shows an IP signal.
Note that IP must be addressed relative to the weld bevel. For example, a double V weld
will have IP in the midwall, whereas a single V bevel will be surface-breaking.
However, the rise-fall time of the signal is similar to that for toe cracks and other root
defects. This requires extra care on the part of the operator. Note that incomplete
penetration can look similar to surface lack of sidewall fusion.
Slag typically shows multiple facets and edges visible in the A-scan and S-scan. The
A-scan shows a slow rise and fall time with long pulse duration, indicative of a non-
planar flaw. Typically slag shows lower amplitude than planar flaws, and may be
difficult to distinguish from porosity, or from some smaller planar defects. Slag is
typically detectable from both sides, can be plotted from both sides of the weld and is
often best characterized using an S-scan. A slag reflector will typically plot to the
correct depth area and reference lines that coincide to the weld volume. Picture 6 shows
an example.
A scan plan (documented examination strategy) shall be provided showing search unit
placement and movement that provides a standardized and repeatable methodology for
the examination. All information of scan plan shall be showed in appendix 3
Reference level(TCG) L ≤ t (but max. 8 mm) TCG + 4 dB L ≤ 0.5t (but max. 12.5 mm)
0.5 t < L ≤ t (but max. 25
TCG – 6 dB L > t (but max. 8 mm) TCG – 2 dB
mm)
L > t (but max. 25 mm in
- - TCG – 6 dB
both outer t/3)
L > t (but max. 50 mm in
- - TCG – 6 dB
middle t/3)
Cracks are not permitted.
For welds in duplex stainless steel, CRAs and clad/lined steel: Lack of fusion and lack of penetration are
not permitted.
Transverse indications: Indications shall be considered as transverse if the echo amplitude transversely
exceeds the echo amplitude from the same indication longitudinally with more than 2 dB. Transverse
indications are unacceptable unless proven not to be planar, in which case the acceptance criteria for
longitudinal indications apply.
Total accumulation of discontinuities: The total length of acceptable indications with echo amplitude of
reference level – 6 dB and above shall not exceed 3 t, maximum 100 mm in any weld length of 300 mm nor
more than 12% of total weld length. Any accumulation of defects in any cross section of weld that may
constitute a leak path or reduce the effective thickness of weld more than t/3 is not acceptable.
If only one side of the weld is accessible for testing 6 dB shall be subtracted from the maximum echo
permitted above.
Notes:
1) Reference level is defined as the echo amplitude corresponding to the echo from the reflector in the
reference blocks or equivalent reflector.
2) All indications exceeding 20% of the reference level shall be investigated to the extent that the
operator determines the shape, length and location of the imperfection.
3) Indications that cannot be established with certainty shall whenever possible be tested with
radiography. Indications that are type determined in this way shall meet the RT acceptance criteria.
4) Longitudinal imperfections where the echo height intermittently is below and above the acceptance
level shall if possible be investigated with radiography. Indications that are determined in this way
shall meet the RT acceptance criteria. If radiography cannot be performed, the length shall not exceed
3 t, maximum 100 mm in any weld length of 300 mm.
5) Length and depth shall be determined by an appropriate method.
6) Detectable imperfections are not permitted in any intersection of welds.
7) Systematic imperfections that are distributed at regular distances over the length of the weld are
not permitted even if the size of any single imperfection meets the requirements above.
In case strain level equal or over 0.4%, Acceptance criterial for pipeline girth welds
can be based on an Engineering Critical Assessment (ECA)
15.0 DOCUMENTATION
Results of ultrasonic examination shall be reported on the Phased Array and
examination report sheet. All Phased Array examination data, A-scans, B, C, D, and S
Scans shall be saved as a digital encoded recording.
The Ultrasonic Profile, Ultrasonic Calibration, and Ultrasonic Instrument Linearity
verification (if applicable) shall be considered part of the examination report, assigned
the NDT Report Log Numbers, and page numbered appropriately.
All NDT procedure qualification documentation shall be maintained as quality control
record.
16.0 RECORDS
16.1 Ultrasonic examination results shall be recorded in the attached form by certified
personnel who perform the ultrasonic examination.
16.2 For each ultrasonic examination, the following information shall be recorded:
a. procedure identification and revision;
b. ultrasonic instrument identification (including manufacturer's serial number);
c. search unit(s) identification(including manufacturer's serial number, frequency,
and size);
d. beam angle(s) used;
e. couplant used, brand name or type:
f. search unit cable(s) used, type and length:
g. special equipment when used (search units, wedges, shoes, automatic scanning
equipment, recording equipment, etc.);
h. computerized program identification and revision when used;
i. calibration block identification;
j. simulation block(s) and electronic simulator(s) identification when used;
k. instrument reference level gain and, if used, damp
l. calibration data [including reference reflector(s), indication amplitude(s), and
distance reading(s)];
m. data correlating simulation block(s) and electronic simulator(s), when used, with
initial calibration;
n. identification and location of weld or volume scanned;
o. surface(s) from which examination was conducted, including surface condition;
p. map or record of rejectable indications detected or areas cleared;
q. areas of restricted access or inaccessible welds;
r. examination personnel identity and, when required by referencing Code Section,
qualification level;
s. date and time examinations were performed. Items (b) through (m) may be
included in a separate calibration record provided the calibration record
identification is included in the examination record.
t. search unit element size, number, and pitch and gap dimensions
u. focal law parameters, including, as applicable, angle, element numbers used,
range of elements, element incremental change, angular range, and angle
incremental change
v. instrument settings to include, as a minimum, excitation pulse type, duration
and voltage settings, digitization rate (e.g., nominal rate as affected by
compression and points quantity), rectification, pulse repetition rate, range
start and stop, band pass filters, smoothing, focal type, and length
w. scan plan variables
x. wedge natural refracted angle
y. scanner and adhering and guiding mechanism
z. indication data [i.e., position in weld, length, and characterization (e.g., crack,
lack of fusion, lack of penetration, or inclusion)]
aa. the final display processing levels
bb. supplemental manual technique(s) indication data, if applicable [same
information as (y)]