You are on page 1of 2

INFANT WARMER NEOSOLUTION

GIGANTE - BRAZIL
Specifications and Technical Characteristics
Classification of the product as per ANVISA
Fitting Class III (high risk)
Classification of the product as per NBR IEC 60601-1 Standard
Type of Protection against Electric Discharge Class I
Level of Protection against Electric Discharge Equipment BF Type
IPX0 – Common Equipment (Equipment
Protection against Noxious Penetration of Water closed protected against leakage of
water)
Safe level of application in the presence of an anesthetic mixture
inflammable with the air, oxygen or nitrous oxide Not adequate
Operation Mode Continuous
Ambient conditions of operation (as per IEC 601601-2 Item 10.2.1)
Range of ambient operation temperature +18°C to +30°C
Range of ambient temperature recommend by Gigante +21°C to +26°C
Range of relative humidity of operation 30% to 75% (non-condensing)
Note: Incubators does not have the possibility to control the relative
humidity
Range of atmospheric pressure of operation 700hPa to 1060 hPa
(525 mmHg to 795 mmHg)
Ambient conditions of transport or storage
Range of ambient temperature of transport or storage 0°C to +55°C
Range of relative humidity to transport or storage 0% to 90% (non-condensing)
Note: Incubator does not have the possibility to control the relative
humidity
Range of atmospheric pressure of transport or storage 500hPa to 1060 hPa
(375 mmHg to 795 mmHg)
Feeding Tension – Automatic reversion 110-130V ~/ 220-230 V~
Frequency 50 / 60 Hz
N°of phases Mono-phase
Fuse (fast – 20 mm) 10A (127V net) / 5A (220V net)
Potency of the warming element 650 Watts
Potency of auxiliary lighting 18 Watts
Nominal potency of entrance (Total) 800 V.A.
Working Range (control modes) Manual (Potency)
Con trolled Newborn (RN)
Working Range (Manual mode) From 0% to 100%
Working Range (Newborn mode) From 35°C to 38°C
Reading Range (indication in the display of temperature) From 0% to 50°C
Accuracy Reading Range of conformity 34°C to 39°C
Resolution of programming and indication of the displays 0,1°C (Newborn controlled mode)
1% (Manual mode)
Maximum error to the indicator of the Newborn temperature +/- 0,3°C
Thermic limiter ( safety thermometer) Electronic
Undesirable or excessive radiation limiter Electronic
Thermostat (working thermometer) Electronic microprocessed
Thermic limiter (safety thermometer) Electronic
Equipment Dimensions (Length x Width x Height) 1,320 x 0,755 x 1,200 mm
Occupied volume 1,200 m3
Mattress Non-toxic and antiallergic
Mattress Dimensions (Length x Width x Height) 0,636 x 0,445 x 0,025 m
Included
Monitor-Tray
Retractable Drawer for X-Ray
IV stand
Drawer Cabinet
Rotate Movement in the Heater Base
Trendelenburg, Horizontal and Reverse Electric Movement
Halogen Phototherapy - Model H100 Coupled