Overview Seminar workflow
1. „Preparation“
2
2. „Introduction“
I t d ti “
3. B. Braun Presentation
4. Quality assurance
5. General Info’s/risks of CI
6. System Overview
7. Service manual
8. Disassembly
9. Assembly
10. Service program and Kalibration
11. Checking
12. Documentation und Certifying
Disassembly/Assembly
1
1.3 Checklist Special-/Calibration tools Preparation
Cross section
Disposables
• Syringes (per device): 1x 50ml OPS
1x 30ml (Omnifix)
1x 20ml OPS and/or Omnifix
1x 10ml (Omnifix)
1x 5ml (Omnifix)
1x 2ml (Omnifix)
• Competitor syringes (relating to those used in your markets)
• Original Perfusor Line (one per device)
2
1.3 Checklist Special-/Calibration tools Preparation
Calibration/Checking/Others
• PC
• Interface line (PC connection Interface line)
• Service software
• Service plug Space
• Syringe gauge PSP (with motor power test adapter and plate)
• Length gauge PSP
• Diameter gauge kit (32mm, 23,4 mm, 15,7mm, 9,0 mm)
• Test adapter EL_S (leakage current)
• PCA Key
• DC generator or battery (piston brake disassembly)
6. System Overview
6.1 In general
6 2 Th
6.2 The PPerfusor®
f ®S Space
6.3 Handling overview
6.4 Powering on
6.5 Start of Infusion
6.6 Special functions
6.7 Activating functions
6.8 Alarm
6.9 Drive (principle drawing)
3
6.1 In general System Overview
68 mm
m
Weight of:
Perfusor = approx. 1,4kg
Infusomat = approx. 1.4 kg
4
6.1 In general System Overview
Infusomat Space
p
Perfusor Space
Perfusor- and Infusomat Space are offering exactly the same handling structure.
release button
5
6.1 In general System Overview
Rotation
mechanism for
horizontal and
vertical adaptation
of the pumps
Space clamp
Position 1:
Position 2:
6
6.1 In general System Overview
Power Supply
Integrated
handle
12 V
power supply
7
6.1 In general System Overview
Power Supply
• Input:
Voltage 100….240V~ /
50-60 Hz/550 mA
8
6.1 In general System Overview
Euro-Plug UK-Plug
AUS-Plug
US-Plug
9
6.1 In general System Overview
10
6.1 In general System Overview
PCA KIT
1 3
11
6.1 In general System Overview
Space PCA-Kit:
1 1 On demand Button
1.
2. 1 PCA-syringe holder lock
3. 1 Fixation band
4. 1 Fixation clip
Order data:
• Art.-Nr. 8713554
new (green)
Order data:
• Art.-Nr. 8713556 /4 Stck.
12
6.2 The Perfusor® Space System Overview
6.2.1 Back
Cover of battery
compartment
P3
P3 P2
P2
Guide rails
Socket for Socket for 12 V, Interface,
connection with PCA, nurse call,
SpaceControl Service connector
13
6.2.1 Back The Perfusor®Space
C = Controller
A = Area
N = Network
Battery
connector (P1)
Device feet
Battery pack
14
6.2.1 Back The Perfusor®Space
Charge controller
Pre-alarm: 30 minutes
Temperature
sensor
Battery
yppack
15
6.2.1 Back The Perfusor®Space
After battery is fully charged it switches from charging current to “trickle” charging
t1 t2 t3 t
16
6.2.2 Bottom side The Perfusor®Space
Glued Seal
Side wall screw
PCA Lock
Type plate
• The seal stickers are intended to be used for B.Braun - Hospital Care
Devices only
only.
• The Seal number is exclusively for your Personal use (related on your
name) and not transferable
• The Seal number are only allowed to stick on devices which you have been
trained on (normally by a B|BRAUN certified Technical Trainer) and which
intended to be sealed.
• An intact seal upon the device says that after your sealing, there weren't
anyone who's opened it
following page
17
The Handling of the Seal
• After quitting the job as a service engineer, please send us (Technical Training
Dept.) an information about it and destroy
your remaining seal stickers.
• In the moment you are receiving a new seal number all probably existing older
seal numbers on your name are invalid.
Those older numbers are going to be deleted at BBraun
• Changing your employer does not affect the once assigned seal sticker
number (keep in mind that a changed e-mail
e mail address has an impact to your
“Service Portal” access).
• Please order your Seals like you order your Spare Parts (when ordering,
please write your Seal number as clear as possible)
The seals are now available with the following article
code: 3477 4385.
18
6.2.3 The Front The Perfusor®Space
Syringe illumination
ON/OFF key
Start-/Stop key
Graphic display
Communication key
Cursor for navigation and to SpaceControl or
value inputs OK key to confirm barcode data receipt
inputs
19
6.3 Handling Overview (P. Brake) System Overview
20
6.3 Handling Overview System Overview
Pl
Please watch
t h th
the
respective Instruction
for use !
21
6.4 Powering on System Overview
- Battery status
- Last therapy
Press “On/Off” switch on unit and note the automatic self check:
22
6.5 Start of Infusion System Overview
• Inserting
I ti a syringe
i
• Observable things
Start-/Stop key
Communication key
Arrow field
OK key
23
6.6 Special functions System Overview
Customized
Options
24
6.6.1 Occlusion Pressure Special Functions
Occlusion pressure
L
Length
h and
d di
diameter iinfluence
fl the
h pressure
Pressure
[bar]
∆p
Customized
Options
Occlusion
pressure Data Lock
Data Lock Prevention against non-authorized
Date manipulations.
Clock
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
25
6.6.2 Data Lock Special Functions
Level 1:
No changes of values possible, stop, start, turning on/-off allowed
Level 2:
Same as Level 1, additionally no change of the disposable possible.
26
6.6.2 Data Lock Special Functions
Customized
Options
Occlusion
pressure Date & Clock
Data Lock For documentation reasons and the
Date History Protocol.
Clock
Automatic adjustment if connected to
SpaceStation together with SpaceCom.
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
27
6.6.4 Alarm Volume Special Functions
Customized
Options
Occlusion
pressure Alarm Volume
The loudness of the audible alarm signal can be
Data Lock
adjusted to the environmental conditions in
Date the hospital
Clock
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
Customized
Options
Occlusion
Bolus Rate
pressure
The max.bolus rate can be limited for safety reasons
Data Lock If the bolus rate is set to cero the bolus
Date function is disabled.
Clock
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
28
6.6.8 Macro Mode Special Functions
Customized
Options
Occlusion
Macro Mode
pressure
Individual adjustment of the display content.
Data Lock The user can decide if the delivery rate shall be
Date screened in big numbers or smaller.
Clock
Alarm Volume
Patient Data
Bolus Rate
Macro Mode
Help texts on
Overview:
- Bolus mode („Others“)
- Prime („Rates“)
29
6.8 Alarm System Overview
1
1. St i gauge
Strain
30
6.8 Alarm (influences) System Overview
• Disposables
• etc.
seriell
Functional- synchron Control-
Stepper motor- Movement sensor Driveend- processor processor Real time clock
output stage (2 light barriers) Position sensor (M32) (H8S)
FµP KµP
32kHz
Syringe size- 12MHz 10MHz
Sensor
Motor axis
(Potentiometer)
Syringe pump
spur gear -gearing
drive
(stepper motor)
31
7. Service manual
7.1 Overview
7.2 Specialties
7.3 Features
7.4 Updates
32
7.1 Overview (continuing) Service manual
33
7.2 Specialties Service manual
• Revision service
• Index/ directory
• Real pictures
• Exploded drawing
• etc.
34
7.5 Updates (SW Ident.) Service manual
688 C 030001
device
d i = Perfusor
P f SP
group =C
subgroup= 030001
PFA E 00003
Device = Perfusor fm
group =E
subgroup= 00003
Device SW Code
Infusomat Space (S) 686………
Infusomat Space
p P 687………
Perfusor Space 688………
Space Control 689……….
Space Station 690……… No version display
Space Cover (comf.) 692………
Space Com 695……..
HC-XX-DE08 | Service Department
Page CoE AIS | Perfusor SP | Page 73
35
8. Disassembly
81R
8.1 Required
i d ttools
l
36
8.2.1 Demo/exercise phase A Disassembly
• Opening the device before turning off the device let the
drive head move to „end position“!
(SM page 3-1)
Possible failures:
37
Piston Brake (complete unit) Disassembly
Mind of
this edge
stiff
flexible
38
Piston Brake (Cable routing) Disassembly
39
8.2.2 Demo/exercise phase B Disassembly
• Main PCB
40
Drive Disassembly
• Complete
C l t DDrive
i
• Locking
• The Nut
41
Side wall special Disassembly
- seal material is a relatively hard PE foam (similar to that used in the crown corks)
Seal
Side wall
42
Side wall Disassembly
L b i ti
Lubrication only
l att the
th side
id wall
ll side:
id
The lubrication shall be done directly before inserting the drive into the lower part of
housing by means of a brush (do not use too much grease)
The already lubricated side wall will be delivered ex factory from SN 143631 on.
Drive Disassembly
Caution:
Th two
The t screws from
f the
th drive
d i and d th
the th
three S
Screws ffrom
the main PCB could easily be mixed up!
43
Drive (locking) Disassembly
44
Drive (Drive Head cable) Disassembly
Lead- trough
45
Drive (Possible failures) Disassembly
• Foil
F il cable
bl off d
drive
i h head
d (!) = bending
b di and/or
d/ cutting
tti off
ff
-battery
battery
management S2 S3
2 Kabel
Stepper
S23
S22
4 Kabel
motor
Flex lead
16-adrig
Flex lead 0,5mm Drive Head PCB
Piston Brake 5-wired Raster
S26a
S26b
-pressure sensor
Piston Brake 1mm -Piston plate sensor
Raster Main PCB
S25
Claw
motor
S21
Flex lead
30-adrig
Agenda: green = PCB 1mm Raster
Flexkabel
orange= motor Keypad PCB 15-adrig Display-PCB
S20
46
Main PCB (specialties) Disassembly
• Linear Poti Syringe size detection
Gold Cap
Light barriers
Piezzo
47
Main PCB (specialties) Disassembly
Drive head
Disassembly/assembly
of the drive head see
also service manual
page 3-24 and 3-43.
48
8.2.3 Demophase C Demontage
49
Drive head (interior) Disassembly
50
Drive head (interior) Disassembly
9. Assembly
91C
9.1 Caution
ti
51
9.1 Caution Assembly
9.1.4 Housing
52
9.1.1 Cable routing (good!) Caution
53
9.1.2 Piston Brake Caution
1.Set
1 Set the brake into housing
(certain angle!) and be
aware of the Poti nose
When inserting the drive into the lower housing be aware of the following:
3. Shielding
4. Light barriers
5. etc.
54
9.1.4 Housing Caution
When putting upper housing and lower housing together be aware of the following:
55
9.1.4 Housing (Loudspeaker) Caution
10.1 In General
56
10.1 In general Service Prog./Calib.
• O
One opportunity
t it to
t update
d t Perfusor-/Infusomat
P f /I f t Space
S is
i to
t perform
f a single
i l pump
(serial- or via CAN cable)
On Service CD V3.5 located in the tool folder, you will find CAN driver 4.14.
This driver is compatible with WIN 7
The driver installation routine will detect an existing driver on your PC and will stop the
installation
You need to uninstall the old CAN driver first before you install the new one
The old (big) CAN Box is able to work with the latest driver
(see next slide with pictures of old/new CAN Box)
57
10.1 In general - CAN driver Service Prog./Calib.
58
10.2 With Service Plug Service Prog./Calib.
• Syringe
S i ttype „template“
t l t “ has
h no automatic
t ti bolus
b l reduction
d ti
• While calibrating the syringe diameter be sure to turn the syringe holder completely
clock wise (when asked)
axial fixation sensor has to be opened
59
10.4 Opp. inside Service Prog. Service Prog./Calib.
10.4.1.Disposables
10.4.3 Tests
60
10.4.1 Disposables Others in Service Prog.
Data flow:
1.
1 Read
R d ffrom d
device
i
2. Save to Disp. editor
Perfusor 3. Load from file
4. Save to device
Space
Service Disposable
Program editor
PC
valid up to SW „E“
Inside „Disposables“ you have the opportunity to create your own table of disposables,
adapted to your special needs/situation in your hospital!
1. Open „Disposables“
2. Read the existing syringe table „from Device“
3. „Save as“ to file
4. Open the „disposable editor“ inside „tools“
5. Modify the table and save with „Accept“
6. Leave the editor with „close“
7. Now load your table into Service Program with „load from file“
8. Make sure to have the right table by checking the inactive
syringes at the lower end of the table
9. Save your syringe table „to device“
10. Automatically the new table from the device is shown
61
10.4.2 Modification Data Others in Service Prog
Documentation of customizations
Within HiBaSeD 6.1.0 instruction manual a pump default setting list can be seen
and /or printed out for documentation of customized changes.
62
10.4.2 Modification Data Others in Service Prog
Claw type
Silver Green
A test will be started with “Initializing” the device (with device switched on!).
63
10.4.4 History Viewer Service Prog./Calib.
You can choose the History viewer in the star up window of the Service Program
Th Password
The P d to
t gett the
th complete
l t access to
t the
th History
Hi t viewer
i (incl.
(i l „Service
S i Edit
Editor“)
“)
is “sp3”
5. By pressing „Start“ button (full screen mode) the file will be uploaded
64
10.4.5 Clone Data - Drug file download Service Prog./Calib.
65
10.5 Software Update Service Prog./Calib.
Caution:
For the SW Update please always read and carefully
follow the respective „Instruction for SW Update”
(see Software IFU J04)
Hint:
During the update process the user specific settings (like “Option”
Option , limitations of infusion
rates and so on) are going to be reset to the default setting.
Affected settings are Modification data and the disposable articles as well as the drug
library (the following procedure is not valid for the drug library).
66
10.5 Software Update – Before update Service Prog./Calib.
How to reactivate the previous saved user specific settings (after the SW update):
2. Open the matching HiBaSeD service program (usually a new, higher version) and
chose your pump within HiBaSeD
3. Take the previous saved hard copy of the Modification data (and/or the disposable
article data) and modify all respective settings.
5. Now the file can be taken to upload this settings to all other Space pumps which should
get the same settings (see slide 10.4.5 Clone Data - Upload)
67
10.5 Software Update Service Prog./Calib.
Relation between different SW versions and cables to be used.
N
No.: D i
Device S i
Service C bl to be
Cable b used
d for
f Drug li
D list
Software Program HiBaSeD editor
1. “C” 1.5.0 RS 232
2. “D” 2.0.1 RS 232 1.0
3. “E” 3.1.0 RS 232 or CAN 2.0
4. “F” 4… RS 232 or CAN 3.0
5. “G” 5… RS 232 or CAN 3.0
6. “H” 6… RS 232 or CAN 5.0
7. “J” 7…. RS 232 or CAN SpOS*1
8. “L” 8… RS 232 or CAN SpOS*1
1. Update itself:
- Several pumps in parallel via CAN cable with SpSt and SpaceUpdate tool
3. Adaptation of user specific data like “Modification Data”, “Disposable Articles” and
“Drug
Drug Library
Library” when required (use Clone function within HiBaSeD)
Hint: All the above mentioned sw tools respectively programs comes together as part of
the “HiBaSeD Service-CD” (Article number: 8713301).
68
10.5 Software Update – After update Service Prog./Calib.
IMPORTANT: After a successful update the following steps must be carried out !
1. After the pump has been used with the update program, the device must be switched
off and disconnected from mains power supply before it can be used.
2. After a successful update the User settings, Modification and Calibration data of the
device need to be re-adjusted with the HiBaSeD-Tool.
• Sometimes the Disposables and the Drug data has to be uploaded to the device,
even if they are not marketed in red (see complete “Instruction for SW Update”).
• After the updade of the user- and adjust data sometimes (see complete “Instruction
for SW Update”) a CalPatchAkku or a CalPatch.ini has to be uploaded to the pump.
Note: If you do not download the CalPatchAkku an operation of the pump in battery
mode might not be possible
69
10.5 Software Update – After update Service Prog./Calib.
After all steps are carried out, finally the pump should run thru a functional test
t verify
to if the
th correctt operation.
ti
If no error messages or alarm are coming up, the software update has been successful!
Attention:
For calibration after the update from SW G direct to SW J please use for:
Perfusor Space “Modify”
Infusomat Space “New device”
For calibration after the update from SW E/F direct to SW J please use for:
Perfusor Space “New device”
Infusomat Space “New device”
70
10.5 Software Update – News in SW “J” Service Prog./Calib.
Therapy Modes:
- Ramp and Taper
- Multi Does Mode
- Program Mode
- Dose Over Time
W-LAN Battery
It is possible to load (download) an existing drug file from a single pump (new in
HiBaSeD)
71
How to downgrade Device Software
Serial cable SP
PC (No.: 8713234)
Space pump
5. After the pump software has been erased it is important to initiate the connection
between the SpaceUpdate program and the device by pressing the on/off button for
several seconds (the update program will show the identified pump when connected) .
72
11. Checking
73
11.2 Performing the TSC Checking
For the following procedure use the SM
1. Mains connection
2. Primary side
3
3. Secondary side
4. Test adapter (SP)
5. Measuring device
Test Testadapter
adapter (SP)EL_Safety
74
11.2 Performing the TSC Checking
75
11.2 Performing the TSC Checking
• summary,
• open questions,
ti
• certificates for the students,
76