You are on page 1of 2

TECHNICAL BULLETIN

Ref: TB 125

Author: M. Legrand Date: 04/11/2017


Revision: M. Legrand Date: 2970172018

Signature:

For:
IMMEDIATE ACTION
ACTION
UPDATE
INFORMATION 

Section

Model: Mythic 22 and Mythic 22AL 1

Information: New Generation of Optic Bench with New LED Version 2 2

Action: NA

Send by ORPHEE-MEDICAL

Contacts: For further information, contact: mlegrand@orphee-medical.com

Information feedback asked by ORPHEE-MEDICAL


To inform ORPHEE-MEDICAL from any incident relating to this document.

ForBulletinTech Page 1
Version 03
1 MODEL

Mythic 22 from serial number 200517-002137 => New Optical Bench N° 300-0617-4558
Mythic 22AL from serial number 300617-000896 => New Optical Bench N° 300-0617-4569

(for information, the serial numbers of the “old” benches equipped with “old” LED begin with 264-XXX or
below, the new generation of optical benches have the serial number which begin with 300-XXX)

2 INFORMATION

New generation of Optic Bench, equipped with new LED are also equipped with new FSC and ALL
detection boards which will be installed henceforth on Mythic 22 & 22AL.
Designation: M22-M22AL Optical LED V2 Part Number: 1001280080-SHL

This new design of Optic Bench is totally compatible with the current design of Mythic 22 & 22AL
instruments.
It can be installed instead of previous generation respecting the same instruction in matter of
installation and adjustment.

Still the same rules:

1. Only the complete assy or the flowcell and/or LED can be replaced on customer site.
2. Transverse Adj. stays available for FSC using Latex 7µm particles.
3. FSE and ALL gains Adj. stay available for FSC using G-Cal product as before.

Still the same prohibitions:

1. Do not remove or just release preamplifier boards.


2. Do not remove or just release the 3 fixation screws of Illumination part. (glued in red or green)
3. Do not swap parts from a module to another one, except the flowcell.

Plus, new ones:

1. Do not swap any part from new generation of optic bench with previous one.
2. Do not install new LED (V2) on previous generation of optical bench and vice versa.

ML

ForBulletinTech Page 2