How to create a protocol

on GE Rev HD
How to create a protocol on GE Rev HD
1
On the CT console’s left screen, select Protocol Management twice.
If asked, enter the administrator password.
Note: If you don’t know this password, refer to the person responsible
for the protocols on your site or your GE application specialist.

3 Click on the anatomy corresponding to the protocol you’ll use


as base to create your new protocol (e.g. Non-Contrast Head protocol)

GE Reference Your customized


Protocols protocols
How to create a protocol on GE Rev HD

- Select the protocol you’ll use as base to create your new multiphase CTA protocol (e.g. Non-Contrast Head protocol) 4
4 and click on “Copy”, then click on an available line in the protocols list, and click on “Paste”. 5
- Rename this new protocol (e.g. Stroke Multiphase CTA protocol). To do so, click on the protocol you just pasted in the list and type in the new name.

- Click on “Edit” to build this new protocol. 6

6 5
- Make sure the figure is set in the right position
(supine, head first)

- Adjust the scouts’ start and end locations to make


sure the scouts will be long enough to include the
aortic arch up to the vertex, depending on where you
plan on centering the patient (OM or Sternal Node?)

7 (ex: if SN, start: S 250 end: I 200


if OM, start: S 150 end: I 300)

- Adjust other parameters if necessary


7

8
8
Once the parameters have been adjusted, click on
“Next series” to adjust the other series.
How to create a protocol on GE Rev HD
The 1st series is the non-contrast series. If you used the copy-paste method at the beginning, the parameters have been copied from your base protocol.

Adjust acquisition/reconstruction parameters of


9 non-contrast if necessary.

Once the parameters


have been adjusted, click
on “Create New series”
to build the multiphase
10
CTA series.
9
Select “Axial” and
“Create After”
10
11
How to create a protocol on GE Rev HD
The 2nd series is the multiphase CTA series. You will need to adjust all the parameters.

- Add a series description 12


- Click on Scan Type and adjusts parameters as
shown on screenshot.
13

- GSI Mode OFF (if applicable)


13 - HI Res Mode OFF (if applicable)
- Scan Type Helical
12 - Rotation Time 0.6
- Rotation Length Full
- Shuttle Mode OFF (if applicable)
- MAR Mode OFF (if applicable)

à OK

13
How to create a protocol on GE Rev HD
The 2nd series is the multiphase CTA series. You will need to adjust all the parameters.

- Adjust Start and End locations: Type in I50 and D


(à D Duplicates the end from the previous series (e.g. non
contrast) 14
- Number of images will automatically adapt,
no action needed.

- Select Thick Speed and adjust as shown on 15


screenshot.

15
- Detector Coverage 40
- Helical Thickness 0.625
- Pitch & Speed 0.984:1 / 39.37
- Rotation Time 0.6

à OK

14 15 16 17 18 19 - Adjust Interval to 0.625 16


- Gantry tilt should be S0.0 17
- Adjust SFOV to Small Body 18
- Adjust kV to 120 19
How to create a protocol on GE Rev HD
The 2nd series is the multiphase CTA series. You will need to adjust all the parameters.

- Adjust mA 20
- Auto mA AND Smart mA ON (=yellow)

- Adjust mA range : Min – Max - Noise Index


depending on desired IQ-Dose
Example on opposite screenshot
(Don’t pay attention to the Reference Noise Index)

- Adjust SmartPrep with parameters used to scan


and head CTA (example in screenshot
neck
below,
21
others parameters can be selected if desired, such as
21 20 Dynamic Transition, ...)
How to create a protocol on GE Rev HD
The 2nd series is the multiphase CTA series. You will need to adjust all the parameters.

- Adjust reconstruction parameters as wished. 22


- D can be used for DFOV and R/L, A/P center of
recons to duplicate values from previous series,
boxes will be exactly the same.

- Recon Type can be Detail, Standard or Soft.

- The following Recon Option are advised:

Recon Mode Plus


IQ Enhance ON (yellow)
22 Window Level 440 40
ASIR 50%
Flip/Rotate None

- Other recons/DMPR can be activated depending on


needs.
How to create a protocol on GE Rev HD
The 2nd series is the multiphase CTA series. You will need to adjust all the parameters.

- Use Add Group to add 2 or 3 groups depending 23


if
you want to scan 3 or 4 phases.

- If these images are to be reviewed using the


GE FastStroke software, DO NOT activate 24
the
split series function.

- Adjust acquisition time (Prep Group) 25


between
groups to get the 2nd group to start ~10s after the
1st one, and then the 3rd one to start ~8s after the
2nd one. Same time between 3rd and 4th if a 4th
group is needed.
23 (Don’t forget to take the total exposure time into
account = time to actually run the scan)

24
- Adjust noise indexes for each group if you 26
wish
to lower the doses a bit on the different
acquisitions.

26 25 27
Once the parameters
have been adjusted, click
on “Create New series”
to build the CT Perfusion
series (if applicable).

Select “Axial” and


“Create After”
How to create a protocol on GE Rev HD
The 3rd series is the CT Perfusion series. You will need to adjust all the parameters as desired. If you already have a perfusion protocol, you can look at your parameters and use the same ones.

Adjust ALL acquisition/reconstruction parameters as 28


desired.

(Series description, Scan type, Start/End locations,


Thick Speed,..

Opposite screenshot is an example.

The GE CT Brain Perfusion reference protocol you can find in the


GE list of protocols in your CT scanner can also be used.
(see page 2 of this document to access the GE list of protocols in
your CT scanner)

28

Once the parameters have been


adjusted, click on “Accept” to save 29
this new protocol.