Professional Documents
Culture Documents
C H A P T E R
KEY CONCEPTS
47
Continued—
10 Because the ventilator’s spill valve is closed ventilator–fresh gas flow coupling, and leaks
during inspiration, fresh gas flow from the in the anesthesia machine, the breathing
machine’s common gas outlet normally circuit, or the patient’s airway.
contributes to the tidal volume delivered to 13 Waste-gas scavengers dispose of gases that
the patient. have been vented from the breathing circuit
11 Use of the oxygen flush valve during the by the APL valve and ventilator spill valve.
inspiratory cycle of a ventilator must be Pollution of the operating room environment
avoided because the ventilator spill valve with anesthetic gases may pose a health
will be closed and the adjustable pressure- hazard to surgical personnel.
limiting (APL) valve is excluded; the surge of 14 A routine inspection of anesthesia equipment
oxygen (600–1200 mL/s) and circuit pressure before each use increases operator familiarity
will be transferred to the patient’s lungs. and confirms proper functioning. The U.S.
12 Large discrepancies between the set and Food and Drug Administration has made
actual tidal volume are often observed in the available a generic checkout procedure for
operating room during volume-controlled anesthesia gas machines and breathing
ventilation. Causes include breathing systems.
circuit compliance, gas compression,
No piece of equipment is more intimately associated Much progress has been made in reducing the
with the practice of anesthesiology than the anesthe- number of adverse outcomes arising from anes-
sia machine (Figure 4–1). On the most basic level,
1 thetic gas delivery. Equipment-related
the anesthesiologist uses the anesthesia machine to adverse outcomes are rarely due to device
control the patient’s ventilation, ensure oxygen deliv- malfunction or failure; rather, misuse of anesthesia
ery, and administer inhalation anesthetics. Proper gas delivery systems is three times more prevalent
functioning of the machine is crucial for patient among closed claims. Equipment misuse includes
safety. Modern anesthesia machines have become errors in preparation, maintenance, or deployment
very sophisticated, incorporating many built-in of a device. Preventable anesthetic mishaps are fre-
safety features and devices, monitors, and multiple quently traced to an operator’s lack of familiarity
microprocessors that can integrate and monitor all with the equipment, an operator’s failure to verify
components. Moreover, modular machine designs machine function prior to use, or both. Such mis-
allow a variety of configurations and features within haps accounted for about 1% of cases in the
the same product line. The term anesthesia worksta- American Society of Anesthesiologists’ (ASA)
tion is therefore often used for modern anesthesia Closed Claims Project database from 1990 to 2011.
machines. While two manufacturers of anesthe- Severe injury was found to be related to provider
sia machines in the United States, GE Healthcare errors involving, in particular, improvised oxygen
(Datex-Ohmeda) and Dräger Medical, have the larg- delivery systems and breathing circuit failures, sup-
est market share, other manufacturers (eg, Mindray, plemental oxygen supply problems outside of the
Maquet, Spacelabs) also produce anesthesia deliv- operating room, and problems with an anesthesia
ery systems. Anesthesia providers should be famil- ventilator. In 35% of claims an appropriate preanes-
iar with the operations manuals of all varieties of thetic machine check (see the ASA’s 2008 Recom-
machines present in their clinical practice. mendations for Pre-Anesthesia Checkout) would
Display
Vaporizers
Suction regulator
Bellows assembly
Flow controls
System switch
Primary gas
Carbon dioxide
supply gauges
(CO2) absorber
(usually pipeline)
Oxygen (O2)
flush button
Brake
B Electrical
outlets with
circuit breakers
Main circuit breaker
Oxygen power
Pipeline inlets outlet
Back cover
Cylinders
Scavenging
connector
Noninterchangeable gas-specific connections to pipeline Prevent incorrect pipeline attachments; detect failure, depletion,
inlets (DISS)1 with pressure gauges, filter, and check valve or fluctuation
Pin index safety system for cylinders with pressure Prevent incorrect cylinder attachments; provide backup gas
gauges, and at least one oxygen cylinder supply; detect depletion
Low oxygen pressure alarm Detect oxygen supply failure at the common gas inlet
Minimum oxygen/nitrous oxide ratio controller device Prevent delivery of less than 21% oxygen
(hypoxic guard)
Oxygen failure safety device (shut-off or proportioning Prevent administration of nitrous oxide or other gases when the
device) oxygen supply fails
Oxygen must enter the common manifold downstream Prevent hypoxia in event of proximal gas leak
to other gases
Oxygen concentration monitor and alarm Prevent administration of hypoxic gas mixtures in event of a low-
pressure system leak; precisely regulate oxygen concentration
Automatically enabled essential alarms and monitors Prevent use of the machine without essential monitors
(eg, oxygen concentration)
Vaporizer interlock device Prevent simultaneous administration of more than one volatile agent
Capnography and anesthetic gas measurement Guide ventilation; prevent anesthetic overdose; help reduce
awareness
Oxygen flush mechanism that does not pass through Rapidly refill or flush the breathing circuit
vaporizers
Breathing circuit pressure monitor and alarm Prevent pulmonary barotrauma and detect sustained positive,
high peak, and negative airway pressures
Pulse oximetry, blood pressure, and ECG monitoring Provide minimal standard monitoring
Mechanical ventilator Control alveolar ventilation more accurately and during muscle
paralysis for prolonged periods
Backup battery Provide temporary electrical power (>30 min) to monitors and
alarms in event of power failure
likely have prevented any adverse event. Fortu- The American National Standards Institute and
nately, patient injuries secondary to anesthesia subsequently the ASTM International (formerly
equipment have decreased both in number and in the American Society for Testing and Materials,
severity over the past two decades. However, claims F1850–00) published standard specifications for anes-
for awareness during general anesthesia have thesia machines and their components. Table 4–1 lists
increased. essential features of a modern anesthesia workstation.
Nitrous
oxide
Oxygen
FIGURE 4–3. The anesthesia machine reduces the pressure from the gas supply, vaporizes anesthetic agents, and
delivers the gas mixture to the common gas outlet. The oxygen flush line bypasses the vaporizers and directs oxygen
directly to the common gas outlet. (Reproduced with permission from Rose G, McLarney JT, eds. Anesthesia Equipment Simplified. New York,
NY: McGraw-Hill Education, Inc; 2014.)
North America, the following color-coding scheme optimal use of cylinder gases, machines utilize a
is used: oxygen = green, nitrous oxide = blue, car- pressure regulator to reduce the cylinder gas pres-
bon dioxide = gray, air = yellow, helium = brown, sure to 45 to 47 psig.1 This pressure, which is slightly
nitrogen = black. In the United Kingdom, white is lower than the pipeline supply, allows preferential
used for oxygen and black and white for air. The use of the pipeline supply if a cylinder is left open
E-cylinders attached to the anesthesia machine are a (unless pipeline pressure drops below 45 psig). After
high-pressure source of medical gases and are gener- passing through pressure gauges and check valves,
ally used only as a backup supply in case of pipeline the pipeline gases share a common pathway with
failure. Pressure of gas supplied from the cylinder to the cylinder gases. A high-pressure relief valve pro-
the anesthesia machine is 45 psig. Some machines vided for each gas is set to open when the supply
have two oxygen cylinders so that one cylinder can pressure exceeds the machine’s maximum safety
be used while the other is changed. At 20°C, a full limit (95–110 psig), as might happen with a regula-
E-cylinder contains 600 L of oxygen at a pressure of tor failure on a cylinder. Some machines also use a
1900 psig, and 1590 L of nitrous oxide at 745 psig. second regulator to drop both pipeline and cylinder
pressure further (two-stage pressure regulation). A
second-stage pressure reduction may also be needed
FLOW CONTROL CIRCUITS for an auxiliary oxygen flowmeter, the oxygen flush
Pressure Regulators mechanism, or the drive gas to power a pneumatic
ventilator.
Unlike the relatively constant pressure of the pipe-
line gas supply, the high and variable gas pressure 1
Pressure unit conversions: 1 kiloPascal (kP) = kg/m · s2 =
in cylinders makes flow control difficult and poten- 1000 N/m2 = 0.01 bar = 0.1013 atmospheres = 0.145 psig =
tially dangerous. To enhance safety and ensure 10.2 cm H2O = 7.5 mm Hg.
A B
1000 10
2
900 9
800 8
Millimeters/minute
700 7 1
Liters/minute
Liters/minute
600 6
500 5
400 4 0.8
300 3
200 2
0.6
100 1
0.4
Fine Coarse
flowtube flowtube
FIGURE 4–4 Constant-pressure variable-orifice flowmeters (Thorpe type). A: Two tube design. B: Dual taper design.
are in series and are still controlled by one valve. A can be delivered to the patient (Figure 4–5). To
dual taper design can allow a single flowmeter to reduce this risk, oxygen flowmeters are always posi-
read both high and low flows (Figure 4–4B). Causes tioned downstream to all other flowmeters (nearest
of flowmeter malfunction include debris in the to the vaporizer).
flow tube, vertical tube misalignment, and stick- Some anesthesia machines have electronic
ing or concealment of a float at the top of a tube. flow control and measurement. In such instances,
Should a leak develop within or downstream a backup conventional (Thorpe) auxiliary oxy-
from an oxygen flowmeter, a hypoxic gas mixture gen flowmeter is provided. Other models have
A B C
Oxygen Air Nitrous oxide Air Nitrous oxide Oxygen Nitrous oxide Air Oxygen
Incorrect sequence Datex-Ohmeda sequence Dräger sequence
FIGURE 4–5 Sequence of flowmeters in a three-gas machine. A: An unsafe sequence. B: Typical Datex-Ohmeda
sequence. C: Typical Dräger sequence. Note that regardless of sequence a leak in the oxygen tube or further downstream
can result in delivery of a hypoxic mixture.
1600
e
1400
n
ra
lu
sf
De
1200
e
Vapor pressure mm Hg
an
r
lu
of
1000 Is
ne
a
th
alo
800 H
e
ne ra
600 flu
lura vo
Enf Se
400
200
0
0 5 10 15 20 25 30 35 40 45 50 55 60 65
Temperature °C
with vapor. One milliliter of liquid anesthetic yields flow, and anesthetic concentration exists. However,
approximately 200 mL of anesthetic vapor. Because if total gas flow decreases without an adjustment in
the vapor pressure of volatile anesthetics is greater copper kettle flow (eg, exhaustion of a nitrous oxide
than the partial pressure required for anesthesia, cylinder), the delivered volatile anesthetic concen-
the saturated gas leaving a copper kettle has to be tration rises rapidly to potentially dangerous levels.
diluted before it reaches the patient.
For example, the vapor pressure of halothane is C. Modern Conventional Vaporizers
243 mm Hg at 20°C, so the concentration of halo- 5 All modern vaporizers are agent specific and
thane exiting a copper kettle at 1 atmosphere would temperature corrected, capable of delivering a
be 243/760, or 32%. If 100 mL of oxygen enters
the kettle, roughly 150 mL of gas exits (the initial
100 mL of oxygen plus 50 mL of saturated halo- Oxygen
and 4850 mL
thane vapor), one-third of which would be saturated nitrous
halothane vapor. To deliver a 1% concentration of oxide
5000 mL
halothane (MAC 0.75%), the 50 mL of halothane 1%
vapor and 100 mL of carrier gas that left the copper halothane
to patient
kettle have to be diluted within a total of 5000 mL of
150
fresh gas flow. Thus, every 100 mL of oxygen passing
through a halothane vaporizer translates into a 1%
increase in concentration if total gas flow into the
Halothane
breathing circuit is 5 L/min. Therefore, when total Oxygen 100 mL
flow is fixed, flow through the vaporizer determines
FIGURE 4–7 Schematic of a copper kettle vaporizer.
the ultimate concentration of anesthetic. Isoflurane Note that 50 mL/min of halothane vapor is added for
has an almost identical vapor pressure, so the same each 100 mL/min oxygen flow that passes through the
relationship between copper kettle flow, total gas vaporizer.
constant concentration of agent regardless of tem- bypass area and lead to dangerously high anesthetic
perature changes or flow through the vaporizer. concentrations. In the event of tilting and spillage,
Turning a single calibrated control knob counter- high flow of oxygen with the vaporizer turned off
clockwise to the desired percentage diverts an appro- should be used to vaporize and flush the liquid anes-
priate small fraction of the total gas flow into the thetic from the bypass area. Fluctuations in pressure
carrier gas, which flows over the liquid anesthetic in from positive-pressure ventilation in older anesthe-
a vaporizing chamber, leaving the balance to exit the sia machines may cause a transient reversal of flow
vaporizer unchanged (Figure 4–8). Because some through the vaporizer, unpredictably changing agent
of the entering gas is never exposed to anesthetic delivery. This “pumping effect” is more pronounced
liquid, this type of agent-specific vaporizer is also with low gas flows. A one-way check valve between
known as a variable-bypass vaporizer. the vaporizers and the oxygen flush valve (Datex-
Temperature compensation is achieved by Ohmeda) together with some design modifications
a strip composed of two different metals welded in newer units limit the occurrence of some of these
together. The metal strips expand and contract dif- problems. Variable-bypass vaporizers compen-
ferently in response to temperature changes. When sate for changes in ambient pressures (ie, altitude
the temperature decreases, differential contraction changes maintaining relative anesthetic gas partial
causes the strip to bend, allowing more gas to pass pressure). It is the partial pressure of the anesthetic
through the vaporizer. Such bimetallic strips are also agent that determines its concentration-dependent
used in home thermostats. As the temperature rises physiological effects. Thus, there is no need to
differential expansion causes the strip to bend the increase the selected anesthetic concentration when
other way restricting gas flow into the vaporizer. using a variable-bypass vaporizer at altitude because
Altering total fresh gas flow rates within a wide range the partial pressure of the anesthetic agent will be
does not significantly affect anesthetic concentration largely unchanged. Although at lower ambient pres-
because the same proportion of gas is exposed to the sures gas passing through the vaporizer is exposed to
liquid. However, the real output of an agent would increased vaporizer output, because of Dalton’s law
be lower than the dial setting at extremely high flow of partial pressure the partial pressure of the anes-
(>15 L/min); the converse is true when the flow rate thetic vapor will remain largely unaffected compared
is less than 250 mL/min. Changing the gas composi- with partial pressures obtained at sea level.
tion from 100% oxygen to 70% nitrous oxide may
transiently decrease volatile anesthetic concentra- D. Electronic Vaporizers
tion due to the greater solubility of nitrous oxide in Electronically controlled vaporizers must be utilized
volatile agents. for desflurane and are used for all volatile anesthet-
Given that these vaporizers are agent specific, ics in some sophisticated anesthesia machines.
filling them with the incorrect anesthetic must be 1. Desflurane vaporizer—Desflurane’s vapor pres-
avoided. For example, unintentionally filling a sevo- sure is so high that at sea level it almost boils at
flurane-specific vaporizer with halothane could lead room temperature (Figure 4–6). This high volatil-
to an anesthetic overdose. First, halothane’s higher ity, coupled with a potency only one-fifth that of
vapor pressure (243 mm Hg versus 157 mm Hg) will other volatile agents, presents unique delivery
cause a 40% greater amount of anesthetic vapor to problems. First, the vaporization required for gen-
be released. Second, halothane is more than twice as eral anesthesia produces a cooling effect that would
potent as sevoflurane (MAC 0.75 versus 2.0). Con- overwhelm the ability of conventional vaporizers to
versely, filling a halothane vaporizer with sevoflu- maintain a constant temperature. Second, because it
rane will cause an anesthetic underdosage. Modern vaporizes so extensively, a tremendously high fresh
vaporizers offer agent-specific, keyed, filling ports to gas flow would be necessary to dilute the carrier gas
prevent filling with an incorrect agent. to clinically relevant concentrations. These problems
Excessive tilting of older vaporizers (Tec 4, Tec 5, have been addressed by the development of spe-
and Vapor 19.n) during transport may flood the cial desflurane vaporizers. A reservoir containing
A
Concentration Internal
dial on/off switch
Inlet port
Pressure Temperature-
compensator compensating
bypass
Concentrating
cone
Wick
Vaporizing
chamber
Anesthetic
agent
B Concentration
dial
Bypass path
Mixed gas to
Vaporizer common gas
manifold manifold
Carrier gas
Cooler temperatures
Bimetallic strip
Wick
Sump
FIGURE 4–8 Schematic of agent-specific variable-bypass vaporizers. A: Dräger Vapor 19.n. B: Datex-Ohmeda Tec 7.
desflurane (desflurane sump) is electrically heated supplies gas to the breathing circuit. The term fresh
to 39°C (significantly higher than its boiling point) gas outlet is also often used because of its critical
creating a vapor pressure of 2 atmospheres. Unlike a role in adding new gas of fixed and known composi-
variable-bypass vaporizer, no fresh gas flows through tion to the circle system. Unlike older models, some
the desflurane sump. Rather, pure desflurane vapor newer anesthesia machines measure and report
joins the fresh gas mixture before exiting the vapor- common outlet gas flows. An antidisconnect retain-
izer. The amount of desflurane vapor released from ing device is used to prevent accidental detachment
the sump depends on the concentration selected by of the gas outlet hose that connects the machine to
turning the control dial and the fresh gas flow rate. the breathing circuit.
Although the Tec 6 Plus maintains a constant des- The oxygen flush valve provides a high flow
flurane concentration over a wide range of fresh (35–75 L/min) of oxygen directly to the common gas
gas flow rates, it cannot automatically compensate outlet, bypassing the flowmeters and vaporizers. It is
for changes in elevation as do the variable-bypass used to rapidly refill or flush the breathing circuit,
vaporizers. Decreased ambient pressure (eg, high but because the oxygen may be supplied at a line
elevation) does not affect the concentration of agent pressure of 45 to 55 psig, there is a real potential for
delivered, but decreases the partial pressure of the lung barotrauma to occur. For this reason, the flush
agent. Thus, at high elevations one must manually valve must be used cautiously whenever a patient
increase the desflurane concentration control. is connected to the breathing circuit. Moreover,
inappropriate use of the flush valve (or a situation
2. Aladin (GE) cassette vaporizer—Gas flow from of stuck valve) may result in backflow of gases into
the flow control is divided into bypass flow and liq- the low-pressure circuit, causing dilution of inhaled
uid chamber flow. The latter is conducted into an anesthetic concentration. Some machines use a sec-
agent-specific, color-coded, cassette (Aladin cas- ond-stage regulator to drop the oxygen flush pres-
sette) in which the volatile anesthetic is vaporized. sure to a lower level. A protective rim around the
The machine accepts only one cassette at a time and flush button limits the possibility of unintentional
recognizes the cassette through magnetic labeling. activation
The cassette does not contain any bypass flow chan-
nels; therefore, unlike traditional vaporizers, liquid
anesthetic cannot escape during handling and the THE BREATHING CIRCUIT
cassette can be carried in any position. After leav- In adults, the breathing system most commonly
ing the cassette, the now anesthetic-saturated liquid
used with anesthesia machines is the circle system
chamber flow reunites with the bypass flow before (Figure 4–9); a Bain circuit is occasionally used. The
exiting the fresh gas outlet. A flow restrictor valve components and use of the circle system were previ-
near the bypass flow helps to adjust the amount of ously discussed (see Chapter 3). It is important to
fresh gas that flows to the cassette. Adjusting the note that gas composition at the common gas outlet
ratio between the bypass flow and liquid chamber can be controlled precisely and rapidly by adjust-
flow changes the concentration of volatile anesthetic ments in flowmeters and vaporizers. In contrast,
agent delivered to the patient. Sensors in the cas- gas composition, especially volatile anesthetic con-
sette measure pressure and temperature, thus deter- centration, in the breathing circuit is significantly
mining agent concentration in the gas leaving the affected by other factors, including anesthetic uptake
cassette. Correct liquid chamber flow is calculated in the patient’s lungs, minute ventilation, total fresh
based on desired fresh gas concentration and deter- gas flow, volume of the breathing circuit, and the
mined cassette gas concentration. presence of gas leaks. Use of high gas flow rates dur-
ing induction and emergence decreases the effects
Common (Fresh) Gas Outlet of such variables and can diminish the magnitude
In contrast to the multiple gas inlets, the anesthe- of discrepancies between fresh gas outlet and circle
sia machine has only one common gas outlet that system anesthetic concentrations. Measurement of
A
Patient
Bag
Fresh gas
Inhalation
B C
Patient Patient
To ventilation
bellow
Bag Bag
FIGURE 4–9 Diagram of a typical breathing circuit (Dräger Narkomed). Note gas flow during A: spontaneous
inspiration, B: manual inspiration (“bagging”), and C: exhalation (spontaneous or bag ventilation).
inspired and expired anesthetic gas concentration the anesthesia machine. The sensor should be placed
also greatly facilitates anesthetic management. into the inspiratory or expiratory limb of the circle
In most machines, the common gas outlet is system’s breathing circuit—but not into the fresh gas
attached to the breathing circuit just past the exhala- line. As a result of the patient’s oxygen consumption,
tion valve to prevent artificially high exhaled tidal the expiratory limb has a slightly lower oxygen par-
volume measurements. When spirometry measure- tial pressure than the inspiratory limb, particularly
ments are made at the Y-connector, fresh gas flow can at low fresh gas flows. The increased humidity of
enter the circuit on the patient side of the inspiratory expired gas does not significantly affect most mod-
valve. The latter enhances CO2 elimination and may ern sensors.
help reduce desiccation of the CO2 absorbent.
Newer anesthesia machines have integrated inter- Spirometers
nalized breathing circuit components (Figure 4–10). Spirometers, also called respirometers, are used to
The advantages of these designs include reduced measure exhaled tidal volume in the breathing cir-
probability of breathing circuit misconnects, dis- cuit on all anesthesia machines, typically near the
connects, kinks, and leaks. The smaller volume of exhalation valve. Some anesthesia machines also
compact machines can also help conserve gas flow measure the inspiratory tidal volume just past the
and volatile anesthetics and allow faster changes in inspiratory valve or the actual delivered and exhaled
breathing circuit gas concentration. Internal heating tidal volumes at the Y-connector that attaches to the
of manifolds can reduce precipitation of moisture. patient’s airway.
A common method employs a rotating vane of
Oxygen Analyzers low mass in the expiratory limb in front of the expi-
General anesthesia must not be administered with- ratory valve of the circle system (vane anemometer
out an oxygen analyzer in the breathing circuit. or Wright respirometer, Figure 4–11A).
Three types of oxygen analyzers are available: The flow of gas across vanes within the respi-
polarographic (Clark electrode), galvanic (fuel rometer causes their rotation, which is measured
cell), and paramagnetic. The first two techniques electronically, photoelectrically, or mechanically.
utilize electrochemical sensors that contain cathode In another variation using this turbine principle,
and anode electrodes embedded in an electrolyte the volumeter or displacement meter is designed to
gel separated from the sample gas by an oxygen- measure the movement of discrete quantities of gas
permeable membrane (usually Teflon). As oxygen over time (Figure 4–11B).
reacts with the electrodes, a current is generated that During positive-pressure ventilation, changes
is proportional to the oxygen partial pressure in the in exhaled tidal volumes usually represent changes
sample gas. The galvanic and polarographic sensors in ventilator settings, but can also be due to circuit
differ in the composition of their electrodes and leaks, disconnections, or ventilator malfunction.
electrolyte gels. The components of the galvanic cell These spirometers are prone to errors caused by iner-
are capable of providing enough chemical energy so tia, friction, and water condensation. For example,
that the reaction does not require an external power Wright respirometers under-read at low flow rates
source. and over-read at high flow rates. Furthermore, the
Although the initial cost of paramagnetic sen- measurement of exhaled tidal volumes at this loca-
sors is greater than that of electrochemical sensors, tion in the expiratory limb includes gas that had been
paramagnetic devices are self-calibrating and have lost to the circuit (and not delivered to the patient;
no consumable parts. In addition, their response discussed below). The difference between the vol-
time is fast enough to differentiate between inspired ume of gas delivered to the circuit and the volume of
and expired oxygen concentrations. gas actually reaching the patient becomes very sig-
All oxygen analyzers should have a low-level nificant with long, compliant breathing tubes; rapid
alarm that is automatically activated by turning on respiratory rates; and increased airway pressures.
Y-piece
Absorber pole with
CO2 monitor 19-mm scavender
airway adapter hose terminal
19-mm scavenger
hose
B
PEEP/MAX valve Selection knob for
Expiratory connection port MAN and SPONT on
valve pressure-limiting valve
Expiration
port
Connector for
breathing bag
Carbon dioxide
absorber
FIGURE 4–10 Breathing circuit design. A: Conventional external components. B: Compact design that reduces
external connections and circuit volume (Dräger Fabius GS).
These problems are at least partially overcome by information about airway and lung mechanics.
measuring the tidal volume at the Y-connector to Modifications have been required to overcome inac-
the patient’s airway. curacies due to water condensation and temperature
A hot-wire anemometer utilizes a fine platinum changes. One modification employs two pressure-
wire, electrically heated at a constant temperature, sensing lines in a Pitot tube at the Y-connection
inside the gas flow. The cooling effect of increasing (Figure 4–11D). Gas flowing through the Pitot
gas flow on the wire electrode causes a change in elec- tube (flow sensor tube) creates a pressure difference
trical resistance. In a constant-resistance anemome- between the flow sensor lines. This pressure differ-
ter, gas flow is determined from the current needed ential is used to measure flow, flow direction, and
to maintain a constant wire temperature (and resis- airway pressure. Respiratory gases are continuously
tance). Disadvantages include an inability to detect sampled to correct the flow reading for changes in
reverse flow, less accuracy at higher flow rates, and density and viscosity.
the possibility that the heated wire may be a potential
ignition source for fire in the breathing manifold. Circuit Pressure
Ultrasonic flow sensors rely on discontinui- A pressure gauge or electronic sensor is always used
ties in gas flow generated by turbulent eddies in the to measure breathing-circuit pressure somewhere
flow stream. Upstream and downstream ultrasonic between the expiratory and inspiratory unidirec-
beams, generated from piezoelectric crystals, are tional valves; the exact location depends on the
transmitted at an angle to the gas stream. The Dop- model of anesthesia machine. Breathing-circuit
pler frequency shift in the beams is proportional to pressure usually reflects airway pressure if it is mea-
the flow velocities in the breathing circuit. Major sured as close to the patient’s airway as possible. The
advantages include the absence of moving parts and most accurate measurements of both inspiratory and
greater accuracy due to the device’s independence expiratory pressures can be obtained from the
from gas density.
6 Y-connection. A rise in airway pressure may
Machines with variable-orifice flowmeters usu- signal worsening pulmonary compliance, an
ally employ two sensors (Figure 4–11C). One mea- increase in tidal volume, or an obstruction in the
sures flow at the inspiratory port of the breathing breathing circuit, tracheal tube, or the patient’s air-
system, and the other measures flow at the expiratory way. A drop in pressure may indicate an improve-
port. These sensors use a change in internal diam- ment in compliance, a decrease in tidal volume, or a
eter to generate a pressure drop that is proportional leak in the circuit. If circuit pressure is being mea-
to the flow through the sensor. The changes in gas sured at the CO2 absorber, however, it will not always
flows during the inspiratory and expiratory phases mirror the pressure in the patient’s airway. For exam-
help the ventilator to adjust and provide a constant ple, clamping the expiratory limb of the breathing
tidal volume. However, due to excessive condensa- tubes during exhalation will prevent the patient’s
tion, sensors can fail when used with heated humidi- breath from exiting the lungs. Despite this buildup in
fied circuits. airway pressure, a pressure gauge at the absorber will
A pneumotachograph is a fixed-orifice flow- read zero because of the intervening one-way valve.
meter that can function as a spirometer. A parallel Some machines have incorporated auditory
bundle of small-diameter tubes in chamber (Fleisch feedback for pressure changes during ventilator use.
pneumotachograph) or mesh screen provides a
slight resistance to airflow. The pressure drop across Adjustable Pressure-Limiting Valve
this resistance is sensed by a differential pressure The adjustable pressure-limiting (APL) valve, some-
transducer and is proportional to the flow rate. times referred to as the pressure relief or pop-off
Integration of flow rate over time yields tidal vol- valve, is usually fully open during spontaneous ven-
ume. Moreover, analysis of pressure, volume, and tilation but must be partially closed during manual
time relationships can yield potentially valuable or assisted bag ventilation. The APL valve often
A
To absorber
exhalation port
From
breathing Forward flow
circuit
Sensor clip
Light
emitting
diode A Light
emitting
Light diode B
detector B Light
detector A
Interface panel
Hall effect
Volume transistor (B)
sensor
Exhaled patient gas
Expiratory valve
Spiromed sensor
Absorber
FIGURE 4–11 Spirometer designs. A: Vane anemometer (Datex-Ohmeda). B: Volumeter (Dräger). C: Variable-orifice
flowmeter (Datex-Ohmeda). D: Fixed-orifice flowmeter (Pitot tube).
C
Lower flow
Differential Transducer
pressure diaphragm
transducer
Higher flow
Flow sensor
connector
Reverse flow
Flow sensor
lines
Flow sensor
tube
Gas flow
Flow sensor
flap
requires fine adjustments. If it is not closed suffi- (eg, pneumothorax) or hemodynamic compromise,
ciently excessive loss of circuit volume due to leaks or both. As an added safety feature, the APL valves
prevents manual ventilation. At the same time, if it is on modern machines act as true pressure-limiting
closed too much or is fully closed, a progressive rise devices that can never be completely closed; the
in pressure could result in pulmonary barotrauma upper limit is usually 70 to 80 cm H2O.
HME unit
Patient
circuit
Gas
sample line Patient circuit
HME unit
(heat and moisture exchange)
FIGURE 4–12 Heat and moisture exchanger (HME) functions as an “artificial nose” that attaches between the tracheal
tube and the right-angle connector of the breathing circuit.
A B C
Airway pressure
Gas flow rate
Lung volume
FIGURE 4–13 Pressure, volume, and flow profiles of different types of ventilators. A: Constant pressure. B: Constant
flow. C: Nonconstant generator.
inspiratory pressure but would not alter the flow rate markedly decrease tidal volume, because cycling will
generated by this type of ventilator (Figure 4–14). be delayed until the pressure limit is met. Volume-
cycled ventilators vary inspiratory duration and
B. Transition Phase from pressure to deliver a preset volume. In reality, mod-
Inspiration to Expiration ern ventilators overcome the many shortcomings of
Termination of the inspiratory phase can be trig- classic ventilator designs by incorporating secondary
gered by a preset limit of time (fixed duration), a cycling parameters or other limiting mechanisms.
set inspiratory pressure that must be reached, or a For example, time-cycled and volume-cycled venti-
predetermined tidal volume that must be deliv- lators usually incorporate a pressure-limiting feature
ered. Time-cycled ventilators allow tidal volume that terminates inspiration when a preset, adjustable
and peak inspiratory pressure to vary depending safety pressure limit is reached. Similarly, a volume-
on lung compliance. Tidal volume is adjusted by preset control that limits the excursion of the bellows
setting inspiratory duration and inspiratory flow allows a time-cycled ventilator to function some-
rate. Pressure-cycled ventilators will not cycle from what like a volume-cycled ventilator, depending on
the inspiratory phase to the expiratory phase until the selected ventilator rate and inspiratory flow rate.
a preset pressure is reached. If a large circuit leak
decreases peak pressures significantly, a pressure- C. Expiratory Phase
cycled ventilator may remain in the inspiratory phase The expiratory phase of ventilators normally reduces
indefinitely. On the other hand, a small leak may not airway pressure to atmospheric levels or some preset
To breathing
circuit
Expiratory phase
Drive mechanism
Drive gas
To breathing
circuit
Inspiratory phase
value of positive end-expiratory pressure (PEEP). achieve multiple ventilatory modes, PEEP, accurate
Exhalation is therefore passive. Flow out of the lungs tidal volumes, and enhanced safety features. Some
is determined primarily by airway resistance and anesthesia machines have ventilators that use a
lung compliance. Expired gases fill up the bellows; single-circuit piston design.
they are then relieved to the scavenging system.
A. Double-Circuit System Ventilators
D. Transition Phase from In a double-circuit system design, tidal volume
Expiration to Inspiration is delivered from a bellows assembly that con-
Transition into the next inspiratory phase may be sists of a bellows in a clear rigid plastic enclosure
based on a preset time interval or a change in pres- (Figure 4–16). A standing (ascending) bellows is
sure. The behavior of the ventilator during this phase preferred as it readily draws attention to a circuit
together with the type of cycling from inspiration to disconnection by collapsing. Hanging (descending)
expiration determines ventilator mode. During con- bellows are rarely used and must not be weighted;
trolled ventilation, the most basic mode of all ven- older ventilators with weighted hanging bellows
tilators, the next breath always occurs after a preset continue to fill by gravity despite a disconnection in
time interval. Thus tidal volume and rate are fixed the breathing circuit.
in volume-controlled ventilation, whereas peak The bellows in a double-circuit design ven-
inspiratory pressure and rate are fixed in pressure- tilator takes the place of the breathing bag in the
controlled ventilation (Figure 4–15). anesthesia circuit. Pressurized oxygen or air from
the ventilator power outlet (45–50 psig) is routed
Ventilator Circuit Design to the space between the inside wall of the plastic
7 Traditionally ventilators on anesthesia enclosure and the outside wall of the bellows. Pres-
machines have a double-circuit system design surization of the plastic enclosure compresses the
and are pneumatically powered and electronically pleated bellows inside, forcing the gas inside into
controlled (Figure 4–16). Newer machines also the breathing circuit and patient. In contrast, during
incorporate microprocessor controls and sophis- exhalation, the bellows ascend as pressure inside the
ticated and precise pressure and flow sensors to plastic enclosure drops and the bellows fill up with
A A Mechanical ventilation
VE/VT
Vol Alarms On
VE L/min O2 Pmax cmH2O Volume Mode Circle
5.0 OFF
OFF 25OFF 30
18
40
30
20
VTE ml ƒ /min Pmean 10
500 OFF 12 0
OFF 10 3 6
Sec
9 12 15
Vent On
VT Rate I:E Plimit PEEP
500 mL 10 / min 1:1 40 cmH2O OFFcmH2O
B C
AC power VT
VE/VT
Vol Alarms On
VE L/min O2 Pmax cmH2O Pressure Mode Circle
5.0 OFF
1.4 100OFF 30 18
40
30
20
VTE ml ƒ /min Pmean 10
475 OFF 9 0
OFF 10 3 6
Sec
9 12 15
Vent On
Pinspired Rate I:E Plimit PEEP
20 cmH2O 10 / min 1:2 30 cmH2O OFFcmH2O
D
Pinspired
FIGURE 4–15 Ventilator controls (Datex-Ohmeda). A: Volume control mode. B: Pressure control mode.
the exhaled gas. A ventilator flow control valve regu- L/min and a ventilator is delivering 6 L/min to the
lates drive gas flow into the pressurizing chamber. circuit, a total of at least 8 L/min of oxygen is being
This valve is controlled by ventilator settings in the consumed. This should be kept in mind if the hospi-
control box (Figure 4–16). Ventilators with micro- tal’s medical gas system fails and cylinder oxygen is
processors also utilize feedback from flow and pres- required. Some anesthesia machines reduce oxygen
sure sensors. If oxygen is used for pneumatic power consumption by incorporating a Venturi device that
it will be consumed at a rate at least equal to min- draws in room air to provide air/oxygen pneumatic
ute ventilation. Thus, if oxygen fresh gas flow is 2 power. Newer machines may offer the option of
A
Inspiration Ventilator Expiration Ventilator
driving gas driving gas
Membrane Membrane
and metal lid and metal lid
To EVAC To EVAC
B
Inhalation Drive gas Exhalation Drive gas
To From
patient patient
circuit circuit
To
scavenger To
scavenger
FIGURE 4–16 Double-circuit pneumatic ventilator design. A: Datex-Ohmeda. B: Dräger.
using compressed air for pneumatic power. A leak collapse if the patient generates negative pressure
in the ventilator bellows can transmit high gas pres- by taking spontaneous breaths during mechanical
sure to the patient’s airway, potentially resulting in ventilation.
pulmonary barotrauma. This may be indicated by a
higher than expected rise in inspired oxygen con- B. Piston Ventilators
centration (if oxygen is the sole pressurizing gas). In a piston design, the ventilator substitutes an elec-
Some machine ventilators have a built-in drive gas trically driven piston for the bellows, and the venti-
regulator that reduces the drive pressure (eg, to 25 lator requires either minimal or no pneumatic
psig) for added safety.
8 (oxygen) power. The major advantage of a pis-
Double-circuit design ventilators also incor- ton ventilator is its ability to deliver accurate
porate a free breathing valve that allows outside air tidal volumes to patients with very poor lung com-
to enter the rigid drive chamber and the bellows to pliance and to very small patients.
A 40
Inspiration Expiration
Peak inspiratory
30 pressure
10
0
1 2 3 4 5 6
Time (secs)
B 40
30
Paw (cm H2O)
20
10
0
1 2 3 4 5 6
Time (secs)
C 40
30
Paw (cm H2O)
20
10
0
1 2 3 4 5 6
Time (secs)
FIGURE 4–17 Airway pressures (Paw) can be diagrammatically presented as a function of time. A: In normal persons,
the peak inspiratory pressure is equal to or slightly greater than the plateau pressure. B: An increase in peak inspiratory
pressure and plateau pressure (the difference between the two remains almost constant) can be due to an increase in tidal
volume or a decrease in pulmonary compliance. C: An increase in peak inspiratory pressure with little change in plateau
pressure signals an increase in inspiratory flow rate or an increase in airway resistance.
first few breaths and adjusts subsequent metered drive Safety and Health (NIOSH) recommends limiting
gas flow volumes to compensate for tidal volume the room concentration of nitrous oxide to 25 ppm
losses (feedback adjustment). Another design contin- and halogenated agents to 2 ppm (0.5 ppm if nitrous
ually measures fresh gas and vaporizer flow and sub- oxide is also being used) in time-integrated samples.
tracts this amount from the metered drive gas flow Reduction to these trace levels is possible only with
(preemptive adjustment). Alternately, machines that properly functioning waste-gas scavenging systems.
use electronic control of gas flow can decouple fresh To avoid the buildup of pressure, excess gas vol-
gas flow from the tidal volume by delivery of fresh gas ume is vented through the APL valve in the breath-
flow only during exhalation. Lastly, the inspiratory ing circuit and the ventilator spill valve. Both valves
phase of the ventilator–fresh gas flow may be diverted should be connected to hoses (transfer tubing) lead-
through a decoupling valve into the breathing bag, ing to the scavenging interface, which may be inside
which is excluded from the circle system during venti- the machine or an external attachment. The pressure
lation. During exhalation the decoupling valve opens, immediately downstream to the interface should be
allowing the fresh gas that was temporarily stored in kept between 0.5 and +3.5 cm H2O during normal
the bag to enter the breathing circuit. operating conditions. The scavenging interface may
be described as either open or closed.
An open interface is open to the outside atmo-
WASTE-GAS SCAVENGERS sphere and usually requires no pressure relief valves
13 Waste-gas scavengers dispose of gases that (Figure 4–18). In contrast, a closed interface is closed
have been vented from the breathing circuit to the outside atmosphere and requires negative- and
by the APL valve and ventilator spill valve. Pollution positive-pressure relief valves that protect the patient
of the operating room environment with anesthetic from the negative pressure of the vacuum system
gases may pose a health hazard to surgical person- and positive pressure from an obstruction in the
nel. Although it is difficult to define safe levels of disposal tubing, respectively. The outlet of the scav-
exposure, the National Institute for Occupational enging system may be a direct line to the outside via
Keep bobbin
in optimized
position by
adjusting
suction knob
FIGURE 4–18 Open interface scavenger system. (Reproduced with permission from Rose G, McLarney JT, eds. Anesthesia Equipment
Simplified. New York, NY: McGraw-Hill Education, Inc; 2014.)
(continued)
a ventilation duct beyond any point of recirculation conscientious use of a checkout list is mandatory
(passive scavenging) or a connection to the hospital’s before each anesthetic procedure. A mandatory
vacuum system (active scavenging). A chamber or check-off procedure increases the likelihood of
reservoir bag accepts waste-gas overflow when the detecting anesthesia machine faults. Some anesthe-
capacity of the vacuum is exceeded. The vacuum sia machines provide an automated system check
control valve on an active system should be adjusted that requires a variable amount of human interven-
to allow the evacuation of 10 to 15 L of waste gas tion. These system checks may include nitrous oxide
per minute. This rate is adequate for periods of high delivery (hypoxic mixture prevention), agent deliv-
fresh gas flow (ie, induction and emergence) yet ery, mechanical and manual ventilation, pipeline
minimizes the risk of transmitting negative pressure pressures, scavenging, breathing circuit compliance,
to the breathing circuit during lower flow condi- and gas leakage.
tions (maintenance). Unless used correctly the risk
of occupational exposure for health care providers CASE DISCUSSION
is higher with an open interface. Some machines
may come with both active and passive scavenger
systems. Detection of a Leak
After induction of general anesthesia and intu-
bation of a 70-kg man for elective surgery, a
standing bellows ventilator is set to deliver a
ANESTHESIA MACHINE tidal volume of 500 mL at a rate of 10 breaths/
CHECKOUT LIST min. Within a few minutes, the anesthesiolo-
Misuse or malfunction of anesthesia gas delivery gist notices that the bellows fails to rise to the
equipment can cause major morbidity or mortality. top of its clear plastic enclosure during expira-
tion. Shortly thereafter, the disconnect alarm is
14 A routine inspection of anesthesia equipment triggered.
before each use increases operator familiarity
and confirms proper functioning. The U.S. Food and Why has the ventilator bellows fallen and the
Drug Administration (FDA) has made available a disconnect alarm sounded?
generic checkout procedure for anesthesia gas
Fresh gas flow into the breathing circuit is inad-
machines and breathing systems (Table 4–3). This
equate to maintain the circuit volume required
procedure should be modified as necessary, depend-
for positive-pressure ventilation. In a situation
ing on the specific equipment being used and the
in which there is no fresh gas flow, the volume
manufacturer’s recommendations. Note that
in the breathing circuit will slowly fall because
although the entire checkout does not need to be
of the constant uptake of oxygen by the patient
repeated between cases on the same day, the
(metabolic oxygen consumption) and absorption fresh gas inlet (ie, within the breathing circuit).
of expired CO2. An absence of fresh gas flow could Large leaks within the anesthesia machine are less
be due to exhaustion of the hospital’s oxygen sup- common and can be ruled out by a simple test.
ply (remember the function of the fail-safe valve) Pinching the tubing that connects the machine’s
or failure to turn on the anesthesia machine’s flow fresh gas outlet to the circuit’s fresh gas inlet cre-
control valves. These possibilities can be ruled out ates a back pressure that obstructs the forward
by examining the oxygen Bourdon pressure gauge flow of fresh gas from the anesthesia machine. This
and the flowmeters. A more likely explanation is is indicated by a drop in the height of the flowme-
a gas leak that exceeds the rate of fresh gas flow. ter floats. When the fresh gas tubing is released,
Leaks are particularly important in closed-circuit the floats should briskly rebound and settle at their
anesthesia. original height. If there is a substantial leak within
How can the size of the leak be estimated? the machine, obstructing the fresh gas tubing will
not result in any back pressure, and the floats will
When the rate of fresh gas inflow equals the not drop. A more sensitive test for detecting small
rate of gas outflow, the circuit’s volume will be leaks that occur before the fresh gas outlet involves
maintained. Therefore, the size of the leak can attaching a suction bulb at the outlet as described
be estimated by increasing fresh gas flows until in step 5 of Table 4–3. Correcting a leak within the
there is no change in the height of the bellows machine usually requires removing it from service.
from one expiration to the next. If the bellows col- Leaks within a breathing circuit not connected
lapse despite a high rate of fresh gas inflow, a com- to a patient are readily detected by closing the APL
plete circuit disconnection should be considered. valve, occluding the Y-piece, and activating the
The site of the disconnection must be determined oxygen flush until the circuit reaches a pressure of
immediately and repaired to prevent hypoxia and 20 to 30 cm H2O. A gradual decline in circuit pres-
hypercapnia. A resuscitation bag must be imme- sure indicates a leak within the breathing circuit
diately available and can be used to ventilate the (Table 4–3, step 11).
patient if there is a delay in correcting the situation.
Where are the most likely locations of a How are leaks in the breathing circuit located?
breathing-circuit disconnection or leak? Any connection within the breathing circuit is a
Frank disconnections occur most frequently potential site of a gas leak. A quick survey of the cir-
between the right-angle connector and the tra- cuit may reveal a loosely attached breathing tube
cheal tube, whereas leaks are most commonly or a cracked oxygen analyzer adaptor. Less obvious
traced to the base plate of the CO2 absorber. In causes include detachment of the tubing used by
the intubated patient, leaks often occur in the tra- the disconnect alarm to monitor circuit pressures,
chea around an uncuffed tracheal tube or an inad- an open APL valve, or an improperly adjusted scav-
equately filled cuff. There are numerous potential enging unit. Leaks can usually be identified audibly
sites of disconnection or leak within the anesthesia or by applying a soap solution to suspect connec-
machine and the breathing circuit, however. Every tions and looking for bubble formation.
addition to the breathing circuit, such as a humidi- Leaks within the anesthesia machine and
fier, provides another potential location for a leak. breathing circuit are usually detectable if the
machine and circuit have undergone an estab-
How can these leaks be detected? lished checkout procedure. For example, steps 5
Leaks usually occur before the fresh gas out- and 11 of the FDA recommendations (Table 4–3)
let (ie, within the anesthesia machine) or after the will reveal most significant leaks.