You are on page 1of 9

Int. J. Miner. Process.

76 (2005) 225 – 233


www.elsevier.com/locate/ijminpro

Effect of frothers on bubble size


Rodrigo A. Graua,T, Janusz S. Laskowskib, Kari Heiskanena
a
Helsinki University of Technology, Helsinki, Finland
b
British Columbia University, Vancouver, B.C., Canada
Received 21 May 2004; received in revised form 18 January 2005; accepted 20 January 2005

Abstract

The size of bubbles in flotation cells was measured using the UCT bubble size meter and the HUT bubble size analyser. While
both methods provided similar critical coalescence concentration (CCC) values for the three tested frothers, they also revealed
important differences. Using the HUT bubble size analyser, which does not impose any lower bubble size limit, it was possible to
find differences in bubble size measured for three frothers at concentrations exceeding the corresponding CCC values.
Frothers affect both the break-up phenomena in the impeller/stator zone in flotation cell, and bubble coalescence. While
longer chain polyglycols (e.g. DF-1012) are able to produce larger stable bubbles than the short chain DF-250 and DF-200
polyglycols, the CCC values for DF-1012 are lower than those for DF-250 and DF-200.
D 2005 Elsevier B.V. All rights reserved.

Keywords: flotation; flotation frothers; bubble size; critical coalescence concentration; surface tension

1. Introduction first-order flotation rate constant is linearly related to


S b, the bubble surface area flux, which in turn is
Flotation frothers are used to facilitate air dis- inversely related to the bubble size.
persion into fine bubbles, and to stabilize the froth. The effect of frothers on bubble size has recently
According to Leja–Schulman’s penetration theory been extensively studied (Cho and Laskowski,
(Leja and Schulman, 1954; Leja, 1956/57), frothers 2002a,b; Laskowski et al., 2003; Laskowski, 2003).
accumulate preferentially at the water/gas interface It was found that the effect of frothers on bubble size
and interact with collector molecules adsorbed onto results from their ability to prevent bubble coales-
solid particles in the particle-to-bubble collision and cence. As shown schematically in Fig. 1, with
attachment. The most dramatic effect of the frothers increasing frother concentration, the degree of bubble
on flotation come from their effect on bubble size. The coalescence decreases and at a particular concentra-
tion (critical coalescence concentration, CCC), the
T Corresponding author. Tel.: +358 9 451 3829; fax: +358 9 451 coalescence of the bubbles is completely prevented.
2795. As Fig. 1 explains, the CCC values are obtained by
E-mail address: Rodrigo.Grau@hut.fi (R.A. Grau). finding the intersection of the horizontal asymptote to
0301-7516/$ - see front matter D 2005 Elsevier B.V. All rights reserved.
doi:10.1016/j.minpro.2005.01.004
226 R.A. Grau et al. / Int. J. Miner. Process. 76 (2005) 225–233

Coalescence determines Table 1


bubble size
The breakage process in the List of tested frothers
impeller neighbourhood
Bubble Size

controls the size of the Common name Chemical formula Molecular weight
bubbles
DF-200 CH3(OC3H6)3OH 206.29
DF-250 CH3(OC3H6)4OH 264.37
DF-1012 CH3(OC3H6)6.3OH 397.95
dbccc
Zone I Zone II
Two series of tests were carried out: one at the
CCC Helsinki University of Technology using Outokumpu
Frother concentration flotation cell, and the other one at the University of
Fig. 1. Schematic diagram of the effect of the frother concentration British Columbia using an Open-Top Leeds lab
upon bubble size in a flotation cell. flotation cell. In the former the HUT bubble size
analyser was utilized, while the UCT bubble meter
was used in the latter.
the bubble size–concentration curves at higher con- The HUT bubble size analyser (HUT BSA)
centrations with the sloped line approximating the consists basically of a sampler tube, a viewing
curve at lower concentrations. Further increase of the chamber and image acquisition system as shown in
frother concentration above the CCC value does not Fig. 2. The sampler tube is partially immersed in the
affect the bubble size. liquid phase, below the froth–liquid interface. The
In the experiments discussed in this paper, the sampler tube is used to collect rising air bubbles,
bubble size was determined with the use of the UCT which are drawn into the viewing chamber. The
(University of Cape Town) bubble size meter and the function of the chamber is to expose the bubbles to a
HUT (Helsinki University of Technology) bubble size monochrome CCD camera which is connected to a
analyser. In the UCT technique, bubbles are suctioned personal computer, where the visual information is
into a capillary tube, where the length and velocity of stored and processed. The images captured during
movement of each bubble are measured. The size of sampling are automatically processed using commer-
the capillary tube imposes a lower bubble size limit. In cial image analysis software and in-house visual basic
addition, the suction rate and the inner diameter of the user-interface. The operation of the HUT bubble size
capillary tube seem to impose a maximum bubble size analyser has been described in detail by Grau and
limit. Grau and Heiskanen (2002) observed that Heiskanen (2002, 2003). Bubbles captured by the
bubbles greater than 3 mm in diameter broke when HUT BSA are shown in Fig. 3.
entering a capillary tube of 1 mm in inner diameter.
Since large number of the experimental data have been
generated using the UCT bubble size meter, and since Viewing
Chamber
the critical coalescence concentration for many froth- Progressive
ers has been given based on such measurements, in this Scan Camera

project, the bubble size was also measured using a


visual technique developed at the Helsinki University LED Backlight
of Technology (HUT method, Grau and Heiskanen,
Sampling Analog signal to
2002) to verify how reliable are the CCC values tube frame grabber
obtained with the use of the UCT bubble sizer.

2. Experimental

The tests were carried out with three commercial


Dow Frothers (Table 1). The acronym DF stands for
the trade name Dowfroth. Fig. 2. Schematic diagram of the HUT bubble size analyser.
R.A. Grau et al. / Int. J. Miner. Process. 76 (2005) 225–233 227

(a) (b)

5 5
Fig. 3. Images captured using the HUT BSA method: (a) frother DF-200 9.7  10 M (20 ppm) and (b) DF-1012 2.510 M (10 ppm).

The experimental work was conducted in a 50 dm3 flow rate was set at 77 dm3/min (superficial gas
Outokumpu laboratory flotation cell. Bubble size velocity J g = 1.3 cm/s). On average about 4800
distributions were measured at only one location in bubbles were sized in each run. Each measurement
the cell, 110 mm above the plane of the stator, as was at least duplicated The number of measured
shown in Fig. 4. In a non-coalescing environment, it bubbles, at low frother concentrations (2–4 ppm), was
can be assumed that the bubble size distribution is relatively small in comparison with the number of
entirely determined in the impeller/stator zone and it is bubbles present at high frother concentrations. The
preserved throughout the rest of the cell. Thus, it bubble sizes measured with the use of the HUT
should be possible to collect a representative sample technique are reported at standard temperature (298
irrespective of the sampling location in the cell. Even K) and pressure (101325 Pa) conditions.
though this assumption does not consider a likely The UCT bubble-sizing system comprises a glass
spatial rearrangement of the bubbles, the validity of capillary tube and two pairs of photo-transistor-LED
this hypothesis is backed up by the experiments detectors. Bubbles are drawn up through the capillary
conducted by Jameson and Allum (1984) in a tube under vacuum and collected in a modified
medium-sized flotation cell and by the results burette. The suction rate is controlled by a peristaltic
published by Parthasarathy et al. (1991). pump. The bubbles are transformed into cylinders
The cell was filled with aqueous solutions of the inside of the glass capillary as shown in Fig. 5, and
tested frothers (Table 1), which were prepared using their velocities and length measured using the optical
municipal tap water. The HUT sampler system was detectors. The total volume of gas collected in the
also filled with the same aqueous solution. All the burette is utilized in order to estimate the absolute size
experiments were conducted at room temperature (it
fluctuated between 19 and 22 8C). The impeller speed
was set at 1050 rpm (tip speed: 6.9 m/s) and the air
350 mm
225 mm

Fig. 5. Bubble entering the glass capillary of the UCT bubble-sizing


Fig. 4. Measurement location in the cell. system.
228 R.A. Grau et al. / Int. J. Miner. Process. 76 (2005) 225–233

Concentration (mmol/dm3)
0 0.04 0.08 0.12 0.16 0.2 0.24 0.28
4

Sauter mean diameter (mm)


3.5 UCT HUT

2.5

1.5

0.5

0
0 0.04 0.08 0.12 0.16 0.2 0.24 0.28
3
Concentration (mmol/dm )

Fig. 6. Graphical determination of the CCC values for Dowfroth 200.

of the bubbles. A detailed description of the instru- water. The impeller speed was set at 1000 rpm and the
ment has been published by Randall et al. (1989). air flow rate at 5 dm3/min (superficial gas velocity
The experimental work was carried out in a 1 dm3 J g =0.43 cm/s). Bubble size was determined utilizing
Open-Top Leeds flotation cell. Aqueous solutions of the UCT Bubble size analyser. A glass capillary
the tested frothers were prepared using distillated sampler of 0.8 mm in inner diameter was selected for

3
Concentration (mmol/dm )
0 0.04 0.08 0.12 0.16 0.2
4

3.5 UCT HUT

3
Sauter mean diameter (mm)

2.5

1.5

0.5

0
0 0.04 0.08 0.12 0.16 0.2
3
Concentration (mmol/dm )

Fig. 7. Graphical determination of the CCC values for Dowfroth 250.


R.A. Grau et al. / Int. J. Miner. Process. 76 (2005) 225–233 229

the experimental work. The bubble sampler of the Table 2


UCT bubble size meter was positioned 50 mm above CCC values for the investigated frothers
the stator. All the tests were conducted at room Frother CCC, HUT technique CCC, UCT technique
temperature (21 F1 8C), and approximately 3000 mmol/dm3 ppm mmol/dm3 ppm
bubbles were sampled for each run. DF-200 0.067 13.8 0.084 17.3
DF-250 0.035 9.1 0.033 8.7
2.1. Surface tension DF-1012 0.017 6.6 0.015 6.0

Du Nouy Ring Tensiometer was used to measure obtained using the HUT method (Fig. 6) is not linear
the surface tension of aqueous solutions at varying over the whole concentration range from Co (system
frother concentrations. without frother) to CCC. In this case, two linear
portions could be identified over the Co to CCC
concentration range, and in line with the adopted
3. Results principle, the intersection of that linear portion which
meets the horizontal asymptote is used to determine
The results of the measurements are shown in the CCC. The experimentally determined CCC values for
form of bubble size vs. frother concentration curves in the tested frothers are given in Table 2.
Figs. 6–8. As these figures reveal, each curve is
clearly divided into two concentration ranges, the low
concentration range which can be approximated by a 4. Discussion
steep linear fit and the flat horizontal portion. The
intersection of the horizontal asymptote to the Sauter Perhaps the most noteworthy are patterns clearly
bubble size–concentration curve with the steep linear visible in Figs. 6–8. All the experimental bubble size
portion can easily be determined. This is so-called vs. frother concentration curves have the same shape:
critical coalescence concentration (Cho and Laskow- a very steep part at low frother concentrations and a
ski, 2002a,b). Only the curve for DF-200 frother horizontal flat part at higher frother concentrations.

3
Concentration (mmol/dm )
0 0.02 0.04 0.06 0.08
4

3.5 UCT HUT

3
Sauter mean diameter (mm)

2.5

1.5

0.5

0
0 0.02 0.04 0.06 0.08
3
Concentration (mmol/dm )

Fig. 8. Graphical determination of the CCC values for Dowfroth 1012.


230 R.A. Grau et al. / Int. J. Miner. Process. 76 (2005) 225–233

The curves can easily be fitted with two linear hole sparger. That indicates that the CCC values can
segments with the intersection giving the CCC value. be treated as material constants for frothers.
Despite the differences in the size of bubbles as The primary factors that determine bubble size
measured using the HUT method and the UCT include the following: bubble generating system (its
method, very similar CCC values were obtained in geometry and hydrodynamics) and superficial gas
these two series of the measurements (Table 2). flow rate. The secondary factors, which practically
While the trends are clear, Figs. 6, 7 and 8 also may be more important that the primary ones, first of
show characteristic differences. The bubble sizes all include bubble coalescence that is determined by
obtained from extrapolation of the horizontal asymp- frother concentration (Cho and Laskowski, 2002a,b).
totes for the UCT measurements give very similar In mechanical cells, like the Outokumpu and the
bubble size values for all three tested frothers (see also Open-Top Leeds flotation cells, air is dispersed into
Fig. 9). When the HUT method was utilized, for all fine bubbles in the impeller/stator zone. Experimental
three tested frothers the larger bubbles were measured. evidence indicates that as air is introduced into the
There may be many reasons for these differences. flotation cell, the air collects behind the blades of the
First, it is to be pointed out that these two series of the impeller forming air cavities (Grainer-Allen, 1970).
tests were carried out in different flotation machines. Bruijn et al. (1974) observed that the shape and the
Also, whereas tap water was utilized in the HUT size of the gas cavities change with the air flow rate.
measurements, distilled water was utilized in the UCT Further, the air cavities enlarge as the gas flow rate
measurements. In spite of these differences almost increases and eventually the air cavities can cover
identical CCC values were obtained. This implies that entirely the rear face of the blade. Primary bubbles are
the CCC values for a given frother are independent of generated from the air cavities; according to Rigby et
the type of machine or mechanism used for dispersing al. (1997), gas bubbles break from the cavity due to
the air and independent of the operating conditions instabilities on the upper and lower interfaces of
(air flow rate and impeller speed) in the cell. The same cavity. The primary bubbles are further broken up
conclusion was earlier drawn by Cho and Laskowski under turbulent conditions, mainly in the neighbour-
(2002a), as they compared CCC values measured in a hood of the impeller and dispersed throughout the
lab flotation cell and in the system comprising a three- volume of the cell. The two utilized flotation cells, the

3
Concentration (mmol/dm )
0 0.1 0.2
4
DF-200 HUT
3.5 DF-250 HUT
DF-1012 HUT
Sauter mean diameter (mm)

DF-200 UCT
3
DF-250 UCT
DF-1012 UCT
2.5

1.5

0.5

0
0 0.1 0.2
3
Concentration (mmol/dm )

Fig. 9. Comparison of the bubble size vs. concentration plots for DF-frothers obtained using UCT and HUT methods.
R.A. Grau et al. / Int. J. Miner. Process. 76 (2005) 225–233 231

Outokumpu cell and the Open-Top Leeds cell, have the surface tension data with the CCC values given
different impeller/stator mechanisms. In addition, they in Table 2 indicates that while for the DF-200
were operated under different conditions with a frother the surface tension at the CCC concentration
superficial gas velocity J g = 0.43 cm/s in the Open- for this frother (8.4  10 5 M) is almost that of pure
Top Leeds cell when the UCT method was used to water, the surface tension for DF-250 at its CCC
follow bubble size, and J g = 1.3 cm/s in the Out- value (3.3  10 5 M) is about 66 mN/m and for the
okumpu cell when HUT method was utilized to DF-1012 (CCC =1.5  10 5 M) it is about 61 mN/
follow the bubble size. The influence of the gas flow m. Since surface tension values are interrelated with
rate (superficial gas velocity J g) upon bubble size in adsorption via Gibbs adsorption isotherm, this
mechanical cells is illustrated in Fig. 10 (Grau and means that the adsorption of DF-1012 at its CCC
Heiskanen, 2003). Because of these differences, it is is much larger than the adsorption of DF-250 at the
difficult to identify all the reasons that may be corresponding CCC values for DF-250; in turn, the
responsible for dissimilarities in the sizes of bubbles. adsorption of DF-250 is much higher than the
The following is obvious, in the case of the UCT adsorption of DF-200 at their respective CCC
bubble meter the size of the bubbles at the frother values. Yet, the finest bubbles were measured with
concentration exceeding CCC seems to be determined the DF-200 frother and the coarsest with the DF-
to a large extent by the measuring method. Therefore, 1012 frother. It follows then that the adsorption for
the differences in bubble sizes over the C N CCC larger (and more surface active) molecules of
concentration range is very small for all the three propylene polyglycols must be larger to prevent
tested frothers. These differences are quite substantial bubbles from coalescence. This conclusion may as
for the measurements carried out using the HUT well indicate that the bstaticQ surface tension
bubble-sizing equipment. It is surprising that the least measurements cannot be directly utilized in the
surface active frother (see Figs. 9 and 11) produced analysis of the properties of dynamic systems, as
the finest bubbles, and that the largest bubbles were foams. These foams are very unstable, exist only
measured when the most surface active frother, DF- during bubbling gas and collapse when the foam
1012, was used. formation process is stopped. It is likely that the
Fig. 11 shows surface tension isotherms for DF- stability of such systems is determined by surface
200, DF-250 and DF-1012 frothers. Comparison of elasticity forces. A deformation (compression or

3.5

2.5
Bubble size, mm

1.5

1 Sauter diameter d32, OK Cell, 50 dm3

Average diameter d10, OK Cell, 50 dm3


0.5
Laskowski et al. (2003), 1 dm3 Open-
Top Leeds Cell
0
0 0.5 1 1.5 2
Superficial gas velocity Jg, cm/s (STP)

Fig. 10. Effect of gas velocity upon bubble size. Measurements conducted with DF-250 at concentrations exceeding its CCC value (Grau and
Heiskanen, 2003). OK stands for Outokumpu.
232 R.A. Grau et al. / Int. J. Miner. Process. 76 (2005) 225–233

1E-005 0.0001 0.001 0.01

75 75

70 70

65 65

60 60
Surface Tension (mN/m)

55 55

50 50

45 45
DF-200

DF-250
40 40

DF-1012
35 35

30 30

1E-005 0.0001 0.001 0.01


3
Concentration (mol/dm )

Fig. 11. Surface tension isotherms for DF-200, DF-250 and DF-1012 frothers.

dilation) of the liquid–bubble interface produces 5. Conclusions


variations in surface tension, which provide a
restoring force that tends to counteract the disturb- The following conclusions can be drawn from this
ance (Malysa, 1992; Harris, 1982; Leja, 1982). This project
is known as the Marangoni effect. Under dynamic
conditions the coalescence must then be related not 1. The bubble size measurements carried out using
so much to the frother adsorption as to the rate with the HUT bubble size analyser in a 50 dm3
which it can be adsorbed (Comley et al., 2002). Outokumpu lab flotation cell and the UCT bubble
Furthermore, surface elasticity forces might also size meter in a 1 dm3 Open-Top Leeds flotation cell
have a strong effect on bubble break-up, therefore, the gave practically the same CCC values for three
frother adsorption rate may also be controlling bubble Dow Frothers (DF-200, DF-250 and DF-1012).
generation in the cell. Walter and Blanch (1986) The former tests were carried out in tap water while
concluded that long hydrocarbon chain surfactants the latter were conducted in distilled water.
produce larger stable bubbles than short chain 2. Almost identical CCC values obtained when using
surfactants in solutions with similar surface tensions. different equipment confirms that the CCC values
A similar effect is observed in the measurements can be treated as material constants for frothers.
carried out with the use of the HUT bubble size 3. The experimental bubble size–frother concentra-
analyser (see Fig. 9); shorter chain molecules of tion curves are clearly divided into two concen-
propylene glycols (DF-200) were found to produce tration ranges: the low concentration range which
finer bubbles than the longer chain molecules of DF- can be approximated by a steep linear fit and the
1012 propylene glycol at concentrations exceeding the flat horizontal portion. Since the CCC values are
CCC values. determined graphically as the intersection of the
R.A. Grau et al. / Int. J. Miner. Process. 76 (2005) 225–233 233

horizontal asymptote to the Sauter bubble size– Grainer-Allen, T.J., 1970. Bubble generation in froth flotation
concentration curves with the steep low concen- machines. Trans. Inst. Min. Metall., Sect. C 79, C15 – C22.
Grau, R., Heiskanen, K., 2002. Visual technique for measuring
tration range linear portion the precision of the bubble size in flotation machines. Miner. Eng. 15 (7), 507 – 513.
bubble size measurements apparently are not Grau, R., Heiskanen, K., 2003. Gas dispersion measurements in a
critical for obtaining the CCC values. flotation cell. Miner. Eng. 16 (11), 1081 – 1089.
4. While the obtained experimental bubble size– Harris, P.J., 1982. Frothing phenomena and frothers. In: King,
R.P. (Ed.), Principles of Flotation. South African Institute of
frother concentration curves show the same trends
Mining and Metallurgy Monograph Series, pp. 237 – 263.
irrespective of the bubble-size measuring method Johannesburg.
the bubble sizes measured at frother concentrations Jameson, G., Allum, P., 1984. A survey of bubble sizes in industrial
exceeding the CCC values were much larger when flotation cells, report prepared for AMIRA limited.
the HUT method was utilized for all the three Laskowski, J.S., 2003. Fundamental properties of flotation frothers.
tested frothers. In: Lorenzen, L., Bradshaw, D.J. (Eds.), Proc. 22nd Int. Mineral
Processing Congress, vol. 2. The South African Institute of
Mining and Metallurgy, Cape Town, pp. 788 – 797.
Laskowski, J.S., Tlhone, T., Williams, P., Ding, K., 2003.
Acknowledgments
Fundamental properties of the polyoxypropylene alkyl ether
flotation frothers. Int. J. Miner. Process. 72, 289 – 299.
The authors would like to thank Ms. Maaria Leja, J., 1956/57. Mechanism of collector adsorption and dynamic
Kinnunen for the bubble size measurements at attachment of particles to air bubbles as derived from surface-
Helsinki University of Technology, and Mr. Francisco chemical studies. Trans. IMM 66, 425 – 437.
Leja, J., 1982. Surface Chemistry of Froth Flotation. Plenum
Melo for the surface tension measurements at the
Press.
University of British Columbia. The authors are Leja, J., Schulman, J.H., 1954. Flotation theory: molecular
grateful to Outokumpu Technology for their contri- interactions between frothers and collectors at solid–liquid–air
bution to the research project carried out at Helsinki interfaces. Trans. AIME 199, 221 – 228.
University of Technology. Malysa, K., 1992. Wet foams: formation, properties and mechanism
of stability. Adv. Colloid Interface Sci. 40, 37 – 83.
Parthasarathy, R., Jameson, G.J., Ahmed, N., 1991. Bubble breakup
in stirred vessels—predicting the Sauter mean diameter. Trans.
References
IChemE, Part A 69, 295 – 301.
Randall, E.W., Goodall, C.M., Fairlamb, P.M., Dold, P.L., O’Con-
Bruijn, W., Van’t Riet, K., Smith, J.M., 1974. Power consumption nor, C.T, 1989. A method for measuring the sizes of bubbles in
with aerated Rushton turbines. Trans. Inst. Chem. Eng. 52, two- and three-phase systems. J. Phys., Sect. E: Sci. Instrum. 22,
88 – 104. 827 – 833.
Cho, Y.S., Laskowski, J.S., 2002a. Effect of flotation frothers on
Rigby, G.D., Evans, G.M., Jameson, G.J., 1997. Bubble breakup
bubble size and foam stability. Int. J. Miner. Process. 64, 69 – 80. from ventilated cavities in multiphase reactors. Chem. Eng. Sci.
Cho, Y.S., Laskowski, J.S., 2002b. Bubble coalescence and its effect 52, 3677 – 3684.
on dynamic foam stability. Can. J. Chem. Eng. 80, 299 – 305.
Walter, J.F., Blanch, H.W., 1986. Bubble break-up in gas-liquid
Comley, B.A., Harris, P.J., Bradshaw, D.J., Harris, M.C., 2002.
bioreactors: Break-up in turbulent flows. Chem. Eng. J.
Frother characterization using dynamic surface tension measure- Biochem. Eng. J. 32, B7 – B17.
ments. Int. J. Miner. Process. 64, 81 – 100.

You might also like