Device: FT10-ES
Serial number: P451485
Date of calibration: 14.10.2020
Technician ID: 46
Calibration method: NKM-FT10CS-PK-V1.0
Calibration method
The device to be calibrated is connected to the reference device with a dedicated cable harness. A computer
software with a script is used to control both devices to adjust their readings to match at Pt100 temperatures -30
and 300 °C. The device to be calibrated stores the resulting values in its EEPROM memory. Then the device and
the reference readings are compared at several points. This calibration does not include the sensor nor its
inaccuracy.
Calibrated ranges
Rref
Pt100-4w input
1/2
Calibration certificate 2920619
Pt100-4w input
Pt1000-4w input