You are on page 1of 2

Calibration certificate 2920619

Device: FT10-ES
Serial number: P451485
Date of calibration: 14.10.2020
Technician ID: 46
Calibration method: NKM-FT10CS-PK-V1.0

Calibration method
The device to be calibrated is connected to the reference device with a dedicated cable harness. A computer
software with a script is used to control both devices to adjust their readings to match at Pt100 temperatures -30
and 300 °C. The device to be calibrated stores the resulting values in its EEPROM memory. Then the device and
the reference readings are compared at several points. This calibration does not include the sensor nor its
inaccuracy.

Calibration standards used


Description Device
Pt100 calibrator Nokeval Ac2002 A157256

Calibrated ranges
Rref

Pt100-4w input

Adjust point Device reading Reference reading Offset


-30 °C -30.000 -30.000 0.000
300 °C 300.000 300.000 0.000

1/2
Calibration certificate 2920619

Pt100-4w input

Test point Device reading Reference reading Offset


-100 °C -99.948 -100.000 -0.052
-50 °C -49.988 -50.000 -0.012
0 °C 0.015 0.000 -0.015
50 °C 50.024 50.000 -0.024
100 °C 99.997 100.000 0.003
150 °C 150.018 150.000 -0.018
200 °C 200.005 200.000 -0.005
250 °C 249.977 250.000 0.023
300 °C 299.978 300.000 0.022
350 °C 349.957 350.000 0.043
400 °C 400.004 400.000 -0.004
450 °C 449.982 450.000 0.018
500 °C 499.939 500.000 0.061
550 °C 549.969 550.000 0.031
600 °C 599.943 600.000 0.057

Pt1000-4w input

Test point Device reading Reference reading Offset


0 °C -0.18176 0.00000 0.18176

Juha Hämäläinen, Design Engineer