PRELIMINARY SPECIFICATIONS TO BE
APPROVED BY THE DEPARTMENT OF
SERVICE OF Christie Innomed Inc.
1-866-888-0885
1-866-888-0885
Date: May. 29/2018
CLIENT
BULKLEY VALLEY DISTRICT HOSPITAL -NORTHERN HEALTH AUTHORITY BULKLEY
DISTRICT
SHIMADZU RADSPEED PRO DIGITAL RADIOGRAPHY SUITE HOSPITAL
PROJECT
SUITE
By: null
01 COVER / TITLE
N.T.S.
provided by and under the responsibility of THE CUSTOMER IS INFORMED OF MINIMAL TECHNICAL REQUIRED SPECIFICATIONS
Christie Innomed inc. 02 ROOM LAYOUT DATE
03 STRUCTURAL PLAN
These recommendations must be approved SIGNATURE: DATE: 04 ELECTRICAL PLAN
18/05/08
by Christie Innomed inc. department of
installation and service.
05 ELECTRICAL NOTES PROJECT No.
PRINT NAME: TITLE: 06 SERVICE
07 SHIELDING
BVH2018-02
PAGE
01 07
NOTE
PRELIMINARY SPECIFICATIONS TO BE
APPROVED BY THE DEPARTMENT OF
SERVICE OF Christie Innomed Inc.
SHIMADZU
196.9 102 112.21 838
DOOR
5 CH-200 (5m x 2.6m RAILS)
500 260 285 380
CEILING MOUNTED TUBE SUPPORT 5
SHIMADZU
1707 4
15.75 12.59 15.69 67
6 CH-200 TUBE SUPPORT ELECTRONIC CABINET 6
41 32 40 30
(ON BK-200 EC) WITH R-30H COLLIMATOR
SHIMADZU
24.81 25.6 87.5 353
7 BR-120T
63 65 222 160
119
7'-9"
[2514]
VERTICAL WALL BUCKY
8'-3"
[2359]
8 DR SYSTEM WORKSTATION - - - - 1000 3
RIS / PACS
9 CONSOLE (CUSTOMER SUPPLY) - - - - - 1-866-888-0885
CLIENT
4"
[1516]
4'-113
X03 BULKLEY VALLEY
SITE CONDITIONS FOR RADSPEED PRO
CEILING
MINIMUM CEILING HEIGHT 106'' ( 270 cm) ROOM LAYOUT MARK ROXBOROUGH
RECOMMANDED CEILING HEIGHT 112'' ( 285 cm) SCALE: 1/4'' = 1'-0'' DRAFTMAN
KK, CV
SCALE
1/4''=1'-0''
DATE
18/05/08
PROJECT No.
BVH2018-02
0 1 2 3 4 5 6 7 8 9 10
PAGE
02 07
NOTE
STRUCTURE PRELIMINARY SPECIFICATIONS TO BE
APPROVED BY THE DEPARTMENT OF
SERVICE OF Christie Innomed Inc.
PROVIDE AND INSTALL A CEILING STRUCTURE HAS DEMONSTRATED
ANTI-VIBRATION. THE PLAN SHOULD BE APPROVED BY
THE PHYSICIST FOR THE PHYSICAL
STRUCTURE MUST SUPPORT MOBILE CHARGE OF (455kg) 1000 POUNDS POSITIONING OF THE EQUIPMENT AND
SPECIFICATIONS OF THE SHIELDING.
MAXIMUM STRUCTURE DEFORMATION RECOMMENDED (0.76mm)1/32"
POSITIONING OF THE EQUIPMENT TO BE
MAXIMUM (1.52mm) 1/16" APPROVED BY THE CLINICAL PERSON IN
CHARGE OF THE DEPARTMENT.
STRUCTURE MADE OF TWO BEAMS PARALLEL AND LEVEL
CONSULT A BUILDING STRUCTURE ENGINEER
HARDWARE TO BE SUPPLY
50 NUT 5/16''
PATIENT
LIFT RAIL May 17, 2018
50 WASHER 5/16''
04 COORDINATION
50 LOCK WASHER 5/16'' 03 COORDINATION
125 02 COORDINATION
[1979]
GALVANIZED FINISHES
X-RAY 01 INFORMATION
6'-6"
NO. DESCRIPTION
[2514]
8'-3"
CEILING
[433]
ALL INFORMATION
1'-5"
CONTAINED ON THESE PLANS
DISTANCE BETWEEN FLOOR AND STRUCTURE MAY NOT BE USED FOR
MINIMUM (2700mm)106" WITH LOW CEILING ADAPTER PLANNING, CONSTRUCTION
RECOMMENDED (2845mm) 112" PATIENT OR INSTALLATION PURPOSES
[534]
1'-9"
LIFT RAIL
NOTE WALL REINFORCEMENT
LF02
1- REINFORCEMENT FOR LEAD APRON (SEE PERSON IN CHARGE)
2- REINFORCEMENT FOR WALL BUCKY(50mmx 610mmx 610mm) 2"x24"x24",
CENTER AT (2032mm) 80" OF THE FINISH FLOOR.
UNISTRUT
GENERAL NOTE (TOTAL 7)
THE STRUCTURE MUST BE ACCEPTED BY CHRISTIE INNOMED INC. THE CEILING
SURFACE MUST BE FREE OF (SPRINKLER, DETECTOR, FIXTURE, ETC) 883" 2'-6" 2'-6" 2'-6" 2'-6" 2'-6" 2'-6" 883" 1-866-888-0885
CLIENT
381" 1'-0"
BULKLEY VALLEY
DISTRICT
BEAM HOPSITAL
CEILING
PROJECT
SUITE
2845
112''
[1018]
3'-4"
MARK ROXBOROUGH
DRAFTMAN
KK, CV
[1065]
3'-6"
FLOOR
SCALE
DETAIL "AA"
SCALE: N.T.S. 1/4''=1'-0''
FLOOR
DATE
18/05/08
ELEVATION 1 PROJECT No.
SCALE: 1:50
BVH2018-02
0 1 2 3 4 5 6 7 8 9 10
PAGE
03 07
NOTE
PRELIMINARY SPECIFICATIONS TO BE
APPROVED BY THE DEPARTMENT OF
SERVICE OF Christie Innomed Inc.
W1
4
May 17, 2018
A2 MC2 6
W1 2
04 COORDINATION
03 COORDINATION
CBA1 MC1 1 02 COORDINATION
A1 C2 G D 01 INFORMATION
E-STOP CBA1 3 1 CBA2 NO. DESCRIPTION
[1989]
C3
1"
6'-64
E-STOP C1
ALL INFORMATION
CONTAINED ON THESE PLANS
3
MAY NOT BE USED FOR
PLANNING, CONSTRUCTION
D
B OR INSTALLATION PURPOSES
G
C2,C1
2 ELECTRICAL DIAGRAM
- SCALE: N.T.S.
C3
E CEILING
CLIENT
4"
[1515]
4'-113
BULKLEY VALLEY
C2 DISTRICT
C3
C1 HOPSITAL
[610]
E
2'
[306] PROJECT
1'
FLOOR
1 SUITE
-
MARK ROXBOROUGH
DRAFTMAN
KK, CV
SCALE
1/4''=1'-0''
DATE
18/05/08
PROJECT No.
BVH2018-02
0 1 2 3 4 5 6 7 8 9 10
PAGE
04 07
NOTE
PRELIMINARY SPECIFICATIONS TO BE
APPROVED BY THE DEPARTMENT OF
SERVICE OF Christie Innomed Inc.
Line maximum impedance of 0,16 the run of the Box D: 6"x6"x4" flush mount , bottom at 12" from the finish floor with removable
cover.
May 17, 2018
feeder.
L1, L2, L3, neutral, dedicated and isolated ground from transformer and regular Box E: 8"x8"x6" in the ceiling, with removable cover. 04 COORDINATION
ground.Copper wires. Box G: 6"x6"x4" surface mount, bottom at 36" from the finish floor with removable 03 COORDINATION
cover. 02 COORDINATION
Isolated ground same gage as feeder. Regulard ground per code 01 INFORMATION
W1: Hexagonal boxes and XRAY warning indicators at the door. Contact and power
Wire size per code, maximum impedance and run. (120VAC) from XRAY generator. NO. DESCRIPTION
ALL INFORMATION
Run 0" to 50'-0" , 5 Cupper conductor #3 AWG,
CONTAINED ON THESE PLANS
Run 50'-0" to 100'-0", #1 AWG, 100'-0" to 200'-0", #00 AWG MAY NOT BE USED FOR
PLANNING, CONSTRUCTION
OR INSTALLATION PURPOSES
Conduit #4: 4"x4" duct conduit inside new fur out wall between box E and box C3.
Conduit #5: NA
Conduit #6: conduit and/or conductor for warning indicators X-Ray at the doors
Generator short term 120 kVA , long term 500 VA between W1 and box C2.
PRELIMINARY SPECIFICATIONS TO BE
APPROVED BY THE DEPARTMENT OF
SERVICE OF Christie Innomed Inc.
4' 2'-8"
[1219] [813]
NEW
May 17, 2018
EX 30 x70 EX 40 x70 28 x70
DOOR LEADED DOOR 04 COORDINATION
DOOR 03 COORDINATION
02 COORDINATION
01 INFORMATION
NO. DESCRIPTION
ALL INFORMATION
CONTAINED ON THESE PLANS
MAY NOT BE USED FOR
PLANNING, CONSTRUCTION
EX 28 x70 OR INSTALLATION PURPOSES
DOOR
[5251]
4"
17'-23
1-866-888-0885
CLIENT
BULKLEY VALLEY
DISTRICT
HOPSITAL
PROJECT
MARK ROXBOROUGH
SHIELDING:
DRAFTMAN
FOR RECOMMENDATION OF THE PHYSICIST KK, CV
SCALE
1/4''=1'-0''
DATE
18/05/08
PROJECT No.
BVH2018-02
0 1 2 3 4 5 6 7 8 9 10
PAGE
07 07