Professional Documents
Culture Documents
Optima XR200amx *
imagination at work
System Application Generator
The Optima XR200amx is a self-contained battery operated • 300 mA max
mobile X-ray imaging system designed for performing • kVp and mAs controls
radiographic exams at the point of care when it may not be safe
or practical to move the patient to the radiographic room. Ratings Steps
kVp 50 - 125 ±1 kVp increments
Optima XR200amx is suited for routine radiographic procedures
within a hospital including radiology departments, intensive care mAs .2 - 630 36 (25% up - 20% down)
units, cardiac care units, emergency departments, operating • Exposure time: 4.0 msec – 6.0 sec
rooms, orthopedics, pediatrics, hospital rooms and clinics. • Super-resonant inverter with varying frequency
• Less than 2% low frequency ripple
Key Features • Frequency: greater than 100 kHz
• Designed to be used with conventional film or CR cassettes
• Image chain and power management components are built-in
X-ray Source
• Can be upgraded to image on GE’s wireless (where available) • Nominal Tube Voltage (Radiographic) 40 ~ 150 kV
digital detector, FlashPad* • Nominal Focal Spot size (IEC 60336):
• “Stand-by” mode eliminates boot up cycles and allows • Large Focus... 1.2 mm
exposure within 25 seconds • Small Focus... 0.6 mm
• Designed for 24/7 availability. Use with battery or make • Anode Rotation Speed (minimal): 3200 min-1
exposures while the unit is charging • Permanent Filtration: 0.9 mm Al/75 kV IEC60522: 1999
• Automatic charging algorithms allow the system to be • Maximum X-ray Tube Current
re-charged even if batteries are not fully drained
• Large Focus: 500 mA
• 15kW (nominal) generator and available 30kW (nominal) option
• Small Focus: 200 mA
• Maximum Continuous Heat Dissipation:
Productivity
Without Air-circulator: 170 W (238 HU/s)
• Up to 1200W of power available to minimize charge time • Maximum anode heat units 140 kHU
• Average system charge time of three hours
• System can be driven within four seconds of activation Tube Positioning
• Pre-programmed techniques per anatomy and patient size • The column may be rotated up to ± 270° from the park position
• Tube Angle display and SID measuring tool
Networking • Focal Spot range from 62.5 cm (24.6 in) to 200 cm (78.7 in)
• Worklist can be retrieved from DICOM compliant
®
GE Healthcare
3000 N. Grandview Blvd.
Waukesha, WI 53188
USA
General Electric Company reserves the right to make changes in specification and features
shown herein, or discontinue the product described at any time without notice or obligation.
GE, the GE Monogram and imagination at work are trademarks of General Electric Company.