Product Information PI

TOSHIBA X-RAY TUBE


D-063 / D-063R / D-063S / D-063SR

Stationary Anode X-ray Tube

 Especially designed for mobile type surgical


X-ray TV apparatus.
 Provided with an insulation cylinder.
(D-063S / D-063SR)
 These tubes have focus 0.6, and are available
for maximum tube voltage 110 kV.

General Data

Application ........................................................ Diagnostic Radiography for Mobile C-arm

Electrical:
Circuit:
High Voltage Circuit .................................................. Constant Potential High-Voltage Generator
Grounding ......................................................................................................... Center-Grounded
Nominal X-ray Tube Voltage ................................................................................................... 110 kV
Nominal Focal Spot Value ............................................................................................................ 0.6
Nominal Anode Input Power (at 1.0s) .................................................................................... 2020 W
Exposure Duty Cycle ..................................................................................................... 1:60 or more
(Exposure Time : interval time)

Mechanical:
Dimensions:
Overall Length ......................................................................................... See dimensional outline
Maximum Diameter ................................................................................. See dimensional outline
Target:
Anode Angle ................................................................................................................. 8 degrees
Material .......................................................................................................................... Tungsten
Inherent Filtration .................................................................................... At least 0.8 mm Al at 50 kV
X-ray Coverage ......................................................................................... φ245 mm at SID 970 mm
Weight:
D-063 ...................................................................................................................... Approx. 320 g
D-063R .................................................................................................................... Approx. 440 g
D-063S .................................................................................................................... Approx. 450 g
D-063SR ................................................................................................................. Approx. 570 g

★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
D-063 / D-063R / D-063S / D-063SR

Cooling Method .............................................. Oil immersed (60℃ Max.) and convection oil cooling.
Tube Holding:
D-063 / D-063R .......................... Holding the glass envelope of the anode end and cathode end,
or the screw of the anode shank.
D-063S / D-063SR ........................................................................ Holding the insulation cylinder.

Absolute Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage ................................................................................................ 110 kV


Between Anode (or Cathode) and Ground ................................................................................ 55 kV
Minimum X-ray Tube Voltage ................................................................................................... 40 kV
Maximum X-ray Tube Current .................................................................................................. 20 mA
Filament Characteristics:
Maximum Filament Current ................................................................................................... 3.7 A
Filament Frequency Limits ...................................................... DC or AC (Sine Wave) 0 ~ 20 kHz
Thermal Characteristics:
Anode Heat Content:
D-063 / D-063S ............................................................................................................... 28 kJ
D-063R / D-063SR .......................................................................................................... 34 kJ
Maximum Anode Heat Dissipation:
D-063 / D-063S .............................................................................................................. 265 W
D-063R / D-063SR ......................................................................................................... 600 W
Maximum Radiographic Exposure Time ..................................................................................... 10 s
Maximum Fluoroscopic Duration:
D-063 / D-063S ................................................................................................. 265 W-20minutes
D-063R / D-063SR ............................................................................................ 600 W-20minutes

-2-
D-063 / D-063R / D-063S / D-063SR

Environmental Limits

Operating Limits (in dielectric oil):


Oil Temperature ........................................................................................................... 10 ~ 60 ℃
Oil Pressure ............................................................................................................. 70 ~ 140 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -40 ~ 70 ℃
Humidity ........................................................................................................................ 10 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa

-3-
D-063 / D-063R / D-063S / D-063SR

Maximum Rating Charts


(Absolute maximum rating charts)

Constant Potential High-Voltage Generator


Nominal Focal Spot Value: 0.6
100kV 90kV
20 80kV-50kV
110kV

18 40kV
TUBE CURRENT [mA]

16

14

12

10
0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10
EXPOSURE TIME [s]

Filament & Emission Characteristics

20 4.5
40kV
80kV
16 4.0
FILAMENT VOLTAGE [V]

110kV
TUBE CURRENT [mA]

12 3.5

Ef

8 3.0

4 2.5

0 2.0
2.9 3.1 3.3 3.5 3.7

FILAMENT CURRENT [A]

Note: This graph indicates typical characteristics.

-4-
D-063 / D-063R / D-063S / D-063SR

Anode Heating / Cooling Curve

D-063 / D-063S

28
265W
25
HEAT STORAGE [kJ]

20

150W
15

10
COOLING

HEATING
0
0 100 200 300 400 500 600 700 800 900
TIME [s]

D-063R / D-063SR

35
600W

30
HEAT STORAGE [kJ]

25

300W
20

15
COOLING

10
100W
5

HEATING
0
0 60 120 180 240 300 360 420 480 540 600

TIME [s]

-5-
D-063 / D-063R / D-063S / D-063SR

Dimensional Outline of D-063


Unit: mm

CENTRAL X-RAY

φ16

M5 10DEEP M3

65±1

120±5

CENTRAL X-RAY

φ52MAX

FOCAL SPOT

TERMINAL SLEEVE
12MAX

C2 F 4 )
2 0.
(φ


CENTRAL X-RAY

SHORT PIN

C1 F
F    :FILAMENT
C1,C2:CATHODE
N    :NOT FOR USE

-6-
D-063 / D-063R / D-063S / D-063SR

Dimensional Outline of D-063R


Unit: mm





±



φ

CENTRAL X-RAY




±


φ
φ40
M5 10DEEP

RADIATOR

21

26

86±1

141±5

CENTRAL X-RAY

φ52MAX

FOCAL SPOT

TERMINAL SLEEVE
12MAX

C2 F )
.4
20
(φ


CENTRAL X-RAY

SHORT PIN

C1 F
F    :FILAMENT
C1,C2:CATHODE
N    :NOT FOR USE

-7-
D-063 / D-063R / D-063S / D-063SR

Dimensional Outline of D-063S


Unit: mm

CENTRAL X-RAY

φ16

M5 10DEEP M3

65±1
60

φ12

130±3
125

CENTRAL X-RAY

FOCAL SPOT
65

4MAX

TERMINAL SLEEVE φ60

C2 F )
.4
φ 20


CENTRAL X-RAY

SHORT PIN

C1 F
F    :FILAMENT
C1,C2:CATHODE
N    :NOT FOR USE

-8-
D-063 / D-063R / D-063S / D-063SR

Dimensional Outline of D-063SR


Unit: mm





±



φ


CENTRAL X-RAY




±


φ

φ40
M5 10DEEP

RADIATOR

21

26

86±1
60

φ12

151±3
125

CENTRAL X-RAY

FOCAL SPOT
65

4MAX

TERMINAL SLEEVE φ60

C2 F 4 )
2 0.
(φ


CENTRAL X-RAY

SHORT PIN

C1 F
F    :FILAMENT
C1,C2:CATHODE
N    :NOT FOR USE

-9-
D-063 / D-063R / D-063S / D-063SR

OVERSEAS SUBSIDIARIES AND AFFILIATES

EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

For Sales & Technical Services, please contact the following representative:

・TOSHIBA ELECTRONICS EUROPE GMBH


HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

・TOSHIBA AMERICA ELECTRONIC COMPONENTS, INC.


2150 EAST LAKE COOK ROAD, SUITE 310
BUFFALO GROVE, ILLINOIS 60089 USA
PHONE +1 (847) 484-2400 FAX +1 (847) 541-7287

・TOSHIBA ELECTRON DEVICES & MATERIALS (SHANGHAI) CO., LTD. (TEMS)


RM1606, SH-PLAZA,
No.336, XIZANG ROAD (MIDDLE), SHANGHAI, 200001, CHINA
PHONE +86 (21) 6361-0077 FAX +86 (21) 6351-5760

(Product Development Engineering Department)


1385 SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN, 324-8550, JAPAN