Professional Documents
Culture Documents
■ EXPERIMENTAL SECTION
Materials. TMB was purchased from Sigma-Aldrich Korea (Seoul,
Korea). HBsAg ELISA kits were purchased from Perfemed Group, Inc.
(San Francisco, CA, USA). The photoresist (AZ-GXR601) was
purchased from Merck Co. (Darmstadt, Germany). Polystyrene
microplates were purchased from SPL Co. (Seoul, Korea). Ag/AgCl
reference electrodes were purchased from Warner Instruments LLC
(Hamden, CT, USA). Pieces of 2-mm-diameter Pt wire were used as
the counter electrodes.
IDE Fabrication. The IDEs were fabricated on a glass substrate
using an ultraviolet (UV) photolithographic process.17,25,26 A 3-μm-
thick layer of a positive photoresist (AZ-GXR-601) was spin-coated
onto a 4 in. glass wafer.27,28 After soft-baking the photoresist layer at
105 °C for 1 min, UV-photolithography was performed using a
photomask of the IDEs. The IDEs were designed to have 100 pairs of
finger electrodes, each with a width of 5 μm (de) and a 5 μm (dgap)
space between each. After the IDE pattern was developed, a 5-nm-
thick Ti layer was deposited as an adhesive layer; then, a 50-nm-thick
Au layer was deposited using sputter deposition. The IDE pattern was
obtained after a lift-off process.
Redox Cycling. The amperometric measurements were carried out
using a commercial potentiostat from Ivium Technologies (Eind-
hoven, Netherlands). Cyclic voltammetry was performed at potentials
between −200 and +600 mV versus Ag/AgCl at a scan rate of 50 mV/
s. Conventional cyclic voltammetry for the oxidation of TMB
molecules was carried out, and the CV shown in Figure 1a was
analyzed. The signal for the TMB solution (reduction current) was
calculated as the difference in the reduction current for the TMB stock
solution (OD = 0 without any oxidized molecules).6−10 To analyze
TMB using redox cycling, two distinct potentials were sequentially
Figure 1. Redox cycling of TMB. (a) Conventional cyclic voltammo- applied to the collector: an oxidative potential (+600 mV versus Ag/
gram (CV) obtained using one set of IDEs as a working electrode. (b) AgCl) and a reductive potential (−200 mV versus Ag/AgCl). For each
Two chronoamperomograms from redox cycling with IDEs at TMB sample, two CVs were obtained at the generator electrode, one
oxidative and reductive potential. (c) Structure of IDE used for for the oxidative potential and one for the reductive potential of the
redox cycling. collector. The signal for the TMB solution during redox cycling was
calculated as the difference between the redox cycling signal of the
TMB stock solution (OD = 0, without oxidized molecules) and that
sensitivity when the redox cycling was applied to medical for the TMB sample at the same reduction potential (−200 mV versus
diagnosis. In this work, the redox cycling of TMB was carried Ag/AgCl). That is, the signal of the TMB solution during redox
out without cyclic voltammetry at the generator using cycling = TMB0 current − TMB current.11
chronoamperometry.12−15 The chronoamperometric signal In this work, the chronoamperometric assay was carried out by
was obtained at the same potential conditions of cyclic application of two kinds of electric pulses to the generator at the fixed
voltammetry. The chronoamperometric signal was obtained potential of the collector. For the first chronoamperometry measure-
sequentially under two potential conditions: (1) with the ment, the potential of the generator was changed from a reductive
potential (−200 mV against a Ag/AgCl reference electrode) for 10 s to
generator at the oxidative potential of TMB and the collector at an oxidative potential (+600 mV against Ag/AgCl reference electrode)
the reductive potential of TMB, and (2) with the generator at with the same reductive potential at the collector (−200 mV against a
the reductive potential of TMB and the collector at the Ag/AgCl reference electrode). For the second chronoamperometry
oxidative potential of TMB as shown in Figure 1b. From such a measurement, the potential of the generator was changed from
chronoamperometric method the redox cycling was possible reductive (−200 mV against Ag/AgCl reference electrode) for 10 s to
Figure 2. Chronoamperometry based redox cycling of TMB. (a) Redox reactions of TMB according to the potential of generator and collector. The
potential conditions could result in the iterative redox reaction (redox cycling) for a single TMB molecule. (redox cycling at reduction and oxidation
with collector). (b) Simulation of redox cycling at different potentials of generator and collector.
oxidative (+600 mV against Ag/AgCl reference electrode) with the (Sunnyvale, CA, USA). Redox cycling was performed as described in
same oxidative potential at the collector (+600 mV against Ag/AgCl redox cycling section by dipping an IDE into each well of the
reference electrode). The currents at the generator and collector were microplate before the sulfuric acid quenching process.17,27
■
separately measured, and the signal was determined from the
difference between the current at the end of the first amperometric RESULTS AND DISCUSSION
profile and that at the beginning of the second amperometric profile,
as shown in Figure 1b. Redox Cycling and Modified Chronoamperometry. As
ELISA. Commercial ELISA tests were carried out according to the previously mentioned, TMB undergoes a two-step oxidation.
manufacturer’s instructions. The HBsAg (HBV) ELISA test was From the CV analysis, the oxidation of TMB begins at +100
performed using a 96-well microplate, which was coated with anti- mV (versus an Ag/AgCl reference electrode) from the oxidative
HBV antibodies. The cutoff value was estimated using the positive and wave and the reduction of TMB begins at +400 mV (versus an
negative samples included in the commercial ELISA kit. The standard Ag/AgCl reference electrode) from the reductive wave. For the
carcinoembryonic antigen (CEA) samples were prepared by diluting
redox cycling of TMB, the second working electrode
the positive CEA sample to 1−120 ng/mL. Each CEA sample was
incubated for 1 h at room temperature. After repeating the washing
(collector) is used at a fixed potential to reduce (or oxidize)
steps three times, HRP-conjugated secondary antibodies were the analyte from the first working electrode (generator). As
incubated for 1 h at room temperature. The amount of CEA bound previously reported, the redox cycling of TMB could be carried
to the 96 wells was quantified by a reaction with TMB and a hydrogen out using the collector electrode at the reductive potential of
peroxide solution for 20 min. After incubation, the OD was measured −200 mV and at the oxidative potential of +600 mV against an
at 650 nm using an ELISA reader (VersaMax) from Molecular Devices Ag/AgCl reference electrode.4−6,11
108 DOI: 10.1021/acssensors.7b00681
ACS Sens. 2018, 3, 106−112
ACS Sensors Article
Figure 3. Optimization of chronoamperometry based redox cycling of TMB. Influence of potential at collector and pulse time.
In this work, chronoamperometry was performed by applying also set to the oxidative potential of TMB. In this case, the
two kinds of electric pulses to the generator at the fixed oxidized product of TMB at the generator could not be
potential of the collector. As shown in Figure 1b, the redox recycled at the collector because the potential of the collector is
reaction of TMB could be controlled according to the set to the oxidative potential.
potentials of the generator and collector. The first chro- Such redox cycling processes between the generator and the
noamperometry is carried out at the potential range in Figure collector, according to the oxidative and reductive potentials,
1b-1 where the potential of the generator is set to the reductive were simulated using a commercial analysis software called
potential of TMB and that of the collector is also set to the COMSOL Multiphysics (COMSOL).11,17,29−31 For the simu-
reductive potential of TMB. In this case, the reduced product of lation, two pairs of generator and collector at the same size and
TMB at the generator could not be recycled at the collector electrode distance as IDE in this work. When the potential of
because the potential of the collector is also set to the reductive the generator is set to the reductive potential of TMB and that
potential. At the potential range in Figure 1b-2, the potential of of the collector is also set to the reductive potential of TMB, as
the generator is set to the oxidative potential of TMB and that shown in Figure 2b-1, the simulated reduced product of TMB
of the collector is set to the reductive potential of TMB, as at the generator is not recycled at the collector, as in the case of
shown in Figure 2a-chronoamperometry 1. In this case, the Figure 1b-1. The same results are obtained when the potential
oxidized product of TMB at the generator could be recycled at of the generator is set to the oxidative potential of TMB and
the collector because the potential of the collector is set to the that of the collector is also set to the oxidative potential of
reductive potential. The second chronoamperometry is carried
TMB, as shown in Figure 1b-4. When the potential of the
out at the potential range in Figure 1b-3 where the potential of
generator is set to the oxidative potential of TMB and that of
the generator is set to the reductive potential of TMB and that
the collector is set to the reductive potential of TMB, as shown
of the collector is set to the oxidative potential of TMB, as
in Figure 1b-2, the simulated oxidized product of TMB at the
shown in Figure 2a-chronoamperometry 2. In this case, the
reduced product of TMB at the generator could be recycled at generator is recycled at the collector. When the potential of the
the collector because the potential of the collector is set to the generator is set to the reductive potential of TMB and that of
oxidative potential. The currents at the generator and collector the collector is set to the oxidative potential of TMB, as shown
were separately measured, and the signal was determined from in Figure 1b-3, the simulated reduced product of TMB at the
the current at the end of the first profile and that at the generator is recycled at the collector. These results show that
beginning of the second amperometric profile. Therefore, redox the redox cycling to reduce (or oxidize) the oxidized (or
cycling to reduce (or oxidize) the oxidized (or reduced) TMB reduced) TMB molecule at the generator could be performed
molecule at the generator could be carried out iteratively using iteratively using the collector.
the collector, and the amplified signal from the redox cycling In this work, the redox cycling of TMB was carried out
process is as much as 10-fold higher than that from without cyclic voltammetry at the generator using chronoam-
conventional amperometry with a single working electrode. In perometry. As shown in Figure 2a, the potential of the collector
this work, the difference in the current from the generator is changes from a reductive potential (−200 mV against Ag/AgCl
taken as the redox cycling signal. From the comparison reference electrode) during the first chronoamperometry
between the signals at the generator and collector, the redox measurement (Figure 1b-chronoamperometry 1) to an
cycling efficiency is estimated to be 94% (n = 10). At the oxidative potential during the second chronoamperometry
potential range in Figure 1b-4, the potential of the generator is measurement (+600 mV against Ag/AgCl reference electrode)
set to the oxidative potential of TMB and that of the collector is (Figure 1b-chronoamperometry 1).
109 DOI: 10.1021/acssensors.7b00681
ACS Sens. 2018, 3, 106−112
ACS Sensors Article
Figure 4. Comparison of chronoamperometry based redox cycling with the conventional chronoamperometry with a single electrode for the analysis
of TMB. (a) Current profiles for standard TMB solutions at OD = 0 and OD = 2.0. (b) Comparison of sensitivities for TMB analysis.
Figure 5. Application to immunoassay. (a) Procedure of ELISA for the medical diagnosis of human hepatitis B (hHBsAg). (b) Comparison of
sensitivities between chronoamperometry based redox cycling and the conventional chronoamperometry with a single electrode for the medical
diagnosis of human hepatitis B (hHBsAg).
As shown in Figure 3, the initial current of TMB analysis concentration range of hHBsAg in serum.32,33 From such
increases as the potential of the collector changes from reasons, the high sensitivity of the immunoassay was required
reductive to oxidative because of the oxidation of TMB on for the hHBsAg test. In this work, the oxidized TMB was
the surface of the collector. In addition, the chronoampero- quantified by inserting an IDE into each well of the microplate
metric signal was optimized, which was determined from the for the chronoamperometry-based redox cycling. As shown in
difference between the current at the end of the first Figure 5b, the chronoamperometry-based redox cycling
amperometric profile (Figure 3-2) and that at the beginning provided a sensitivity of 5.51 μA/OD for the signal at the
of the second amperometric profile (Figure 3-3). Because the generator. The sensitivity is estimated to be 0.30 μA/OD using
second oxidation step of TMB occurred at a potential of +400 conventional chronoamperometry with a single working
mV against a Ag/AgCl reference electrode, as shown in Figure electrode. These results show that the chronoamperometry-
1a, the initial oxidation current of TMB for chronoamperom- based redox cycling could have more than a 10-fold higher
etry is similar above the oxidation potential of the collector sensitivity than conventional chronoamperometry using a single
(+600 mV against a Ag/AgCl reference electrode). From these working electrode when applied to a commercial ELISA kit for
results, the optimal potentials of the collector for the first and the medical diagnosis of human hepatitis B virus surface antigen
the second chronoamperometry of TMB were determined to (hHBsAg).
be a reductive potential (−200 mV against an Ag/AgCl
reference electrode) and an oxidative potential (+600 mV
against an Ag/AgCl reference electrode), respectively.
■ CONCLUSIONS
Redox cycling usually reduces (or oxidizes) the oxidized (or
Application to Immunoassay. The immunoassay OD reduced) product iteratively between two kinds of working
value of the TMB solution was measured to quantify the electrodes, and an amplified signal for the single analyte
oxidized TMB molecule. The standard TMB solutions at molecule can be obtained from the redox cycling process. In
different OD values were prepared, and chronoamperometry this work, the redox cycling of TMB was performed without
was performed to quantify oxidized TMB molecules in the cyclic voltammetry at the generator using chronoamperometry.
standard solutions. The signal from chronoamperometry was obtained at the same
As a first step, the current signal from the chronoamperom- potential conditions of cyclic voltammetry. The signal was
etry-based redox cycling method was compared to the obtained from two sequentially performed chronoamperomet-
conventional chronoamperometry using a single working ric profiles: (1) with the generator at the oxidative potential of
electrode for TMB solutions with different OD values. As TMB and the collector at the reductive potential of TMB, and
shown in Figure 4a, conventional chronoamperometry using a (2) with the generator at the reductive potential of TMB and
single working electrode was performed at +600 mV against an the collector at the oxidation potential of TMB. The
Ag/AgCl reference electrode. For the chronoamperometry- chronoamperometry-based redox cycling showed a signal
based redox cycling (dual mode), the potential conditions of sensitivity of 1.49 μA/OD at the generator. The signal at the
Figure 4a were applied, and the signal was measured for the collector was estimated to be 1.40 μA/OD. From the
same standard TMB solutions. comparison between the signals at the generator and collector,
The sensitivities of several kinds of amperometric methods the redox cycling efficiency was estimated as 94% (n = 10). In
were compared to quantify the oxidized TMB. As shown in this work, the difference in the currents from the generator was
Figure 4b, the chronoamperometry-based redox cycling (dual used to determine the redox-cycling signal. The sensitivity of
mode) shows a sensitivity of 1.49 μA/OD for the signal at the the conventional redox cycling with the same IDE was
generator. The signal at the collector is estimated to be 1.40 estimated to be 0.67 μA/OD, and the chronoamperometry
μA/OD. As previously mentioned, such a signal difference for using a single working electrode was estimated to be 0.18 μA/
the generator and collector could be explained from the redox OD. These results showed that the chronoamperometry-based
cycling efficiency which was estimated to be 94%. The redox cycling could be more effectively used to quantify
sensitivity of the conventional redox cycling with the same oxidized TMB than other kinds of amperometric methods. The
IDE is estimated to be 0.67 μA/OD, and the chronoamper- chronoamperometry-based redox cycling was applied to an
ometry using a single working electrode is estimated to be 0.18 immunoassay by using a commercial ELISA kit for the medical
μA/OD. These results show that the chronoamperometry- diagnosis of hHBsAg. Finally, the chronoamperometry-based
based redox cycling could be more effectively used to quantify redox cycling showed a sensitivity of 5.51 μA/OD for the signal
oxidized TMB than other kinds of amperometric methods. at the generator. The sensitivity of conventional chronoamper-
The chronoamperometry-based redox cycling was applied to ometry using a single working electrode was estimated to be
the immunoassay using a commercial ELISA kit for the medical 0.30 μA/OD. These results showed that chronoamperometry-
diagnosis of hHBsAg.17,27 The immunoassay was performed by based redox cycling could have more than a 10-fold higher
treating a known concentration of hHBsAg in serum on the sensitivity than conventional chronoamperometry using a single
microplate coated with the anti-hHBsAg antibodies as shown in working electrode when applied to a commercial ELISA kit for
Figure 5a. To quantify the amount of bound hHBsAg on the the medical diagnosis of hHBsAg.
■
microplate, the anti-hHBsAg antibodies labeled with HRP were
treated and then reacted with the TMB solution. The cutoff AUTHOR INFORMATION
value for the positive and negative determination of each Corresponding Author
sample (OD = 0.105) was calculated according to the *E-mail: jcpyun@yonsei.ac.kr.
manufacturer’s instructions. Usually, the test sample with
higher OD than the cutoff value of ELISA test was determined ORCID
to be positive. For the hHBsAg ELISA test the cutoff value is Ga-Yeon Lee: 0000-0001-5273-6085
known to be quite low in comparison with other infectious Jun-Hee Park: 0000-0002-2871-0093
diseases because the acute patients of hHBsAg have a wide Jae-Chul Pyun: 0000-0001-9954-6860
111 DOI: 10.1021/acssensors.7b00681
ACS Sens. 2018, 3, 106−112
ACS Sensors Article
■
Pyun, J. C. Capacitive biosensor based on an interdigitated electrode
with nanoislands. Anal. Chim. Acta 2014, 844, 27−34.
ACKNOWLEDGMENTS (18) Niwa, O.; Morita, M.; Tabei, H. Electrochemical Behavior of
This work was supported by the Nano-Convergence Reversible Redox Species at Interdigitated Array Electrodes with
Foundation (R201602210) funded by the Ministry of Science, Different Geometries: Consideration of Redox Cycling and Collection
ICT and Future Planning (MSIP, Korea) and the Ministry of Efficiency. Anal. Chem. 1990, 62, 447−452.
(19) Aggarwal, A.; Hu, M.; Fritsch, I. Detection of dopamine in the
Trade, Industry and Energy (MOTIE, Korea), the Industry
presence of excess ascorbic acid at physiological concentrations
Technology Development Program (10063335) funded by the through redox cycling at an unmodified microelectrode array. Anal.
Ministry of Trade, Industry and Energy (MOTIE, Korea), and Bioanal. Chem. 2013, 405, 3859−3869.
the National Research Foundation of Korea (NRF- (20) Goluch, E.; Wolfrum, B.; Singh, P. S.; Zevenbergen, M. S.;
2017R1A2B4004077, NRF-2017R1A2B2004398, NRF- Lemay, G. Redox cycling in nanofluidic channels using interdigitated
2017R1A6A3A11034081, NRF-2017R1D1A1A09000712), and electrodes. Anal. Bioanal. Chem. 2009, 394, 447−456.
Yonsei University Research Fund (Yonsei Frontier Lab. Young (21) Zhou, X.; Liu, L.; Hu, M.; Wang, L.; Hu, J. Detection of
Researcher Supporting Program) of 2017. hepatitis B virus by piezoelectric biosensor. J. Pharm. Biomed. Anal.
■
2002, 27, 341−345.
REFERENCES (22) Ivanov, I.; Gnedenko, O. V.; Nikolaeva, L. I.; Konev, A. V.;
Kovalev, O. B.; Govorun, V. M.; Pokrovskii, V. I.; Archakov, A. I.
(1) Pyun, J. C.; Kim, S. D.; Chung, J. W. New immobilization Detection of hepatitis B visur surface antigen with optical biosensor.
method for immunoaffinity biosensor by using thiolated proteins. Zhurnal Microbiology Epidemiology Immunology 2003, 58−62.
Anal. Biochem. 2005, 347, 227−233. (23) Chung, J. W.; Kim, S. D.; Pyun, J. C. Application of SPR
(2) Park, M.; Bong, J. H.; Yoo, G.; Jose, J.; Kang, M. J.; Pyun, J. C. biosensor for medical diagnostics of HBV-antibodies. Sens. Actuators, B
Flow cytometric immunoassay by using E.coli with autodisplayed Z- 2005, 111−112, 416.
domains. Enzyme Microb. Technol. 2013, 53, 181−188. (24) Choi, Y. H.; Lee, G. Y.; Ko, H.; Kang, M. J.; Pyun, J. C. SPR
(3) Pyun, J. C.; Park, M. Development of a wash-free immunoassay biosensor for the detection of human hepatitis virus surface antigen
using E.coli cells with autodisplayed Z-domains. Analyst 2017, 142, (hHBsAg) using plasma-treated parylene-N film. Biosens. Bioelectron.
1720−1728. 2014, 56, 286−294.
(4) Pyun, J. C.; Kim, S. D.; Chung, J. W. Gold film array electrode for (25) Smirnov, W.; Yang, N.; Hoffmann, R.; Hees, J.; Obloh, H.;
an ELISA analyser based on an amperometric detector. Sens. Actuators, Vuller-Sebert, W. M.; Nebel, C. E. Integrated all-diamond ultra-
B 2005, 111−112, 416−422. microelectrode arrays: optimization of faradaic and capacitive currents.
(5) Lee, J. K.; Yoo, G.; Park, M.; Jose, J.; Kang, M. J.; Pyun, J. C. Anal. Chem. 2011, 83, 7438−7443.
Electrochemical ELISA based on E.coli with autodisplayed Z-domains. (26) Duay, J.; Goran, J. M.; Stevenson, K. J. Facile fabrication of
Sens. Actuators, B 2012, 175, 46−52. carbon ultramicro- to nanoelectrode arrays with tunable voltammetric
(6) Lee, G. Y.; Choi, Y. H.; Chung, H. W.; Cho, S.; Pyun, J. C. response. Anal. Chem. 2014, 86, 11528−11532.
Capacitive immunoaffinity biosensor based on vertically paired ring- (27) An, B. G.; Chang, Y. W.; Kim, H. R.; Kang, M. J.; Park, J. K.;
electrodes. Biosens. Bioelectron. 2013, 40, 227−232. Pyun, J. C. In-situ synthesized CdS nanowires for the application to
(7) Kätelhön, E.; Hofmann, B.; Lemay, S. G.; Zevenbergen, M. A. G.; photosensor. Sens. Actuators, B 2015, 221, 884−890.
Offenhäusser, A.; Wolfrum, B. Nanocavity Redox Cycling Sensors for (28) Noh, Y. Y.; Kim, K. B.; Choi, M. K.; Song, O. S. Properties of
the Detection of Dopamine Fluctuations in Microfluidic Gradients. blocking layer with Ag nano powder in a dye sensitized solar cell.
Anal. Chem. 2010, 82, 8502−8509. Han'guk Seramik Hakhoechi 2016, 53, 105−109.
(8) Wang, Y.; Lin, X.; Su, B. Redox cycling with ITO electrodes (29) Guo, J.; Lindner, E. Cyclic voltammetry at shallow recessed
separated by an ultrathin silica nanochannel membrane. Electrochem. microdisc electrode:Theoretical and experimental study. J. Electroanal.
Commun. 2016, 72, 1−4. Chem. 2009, 629, 180−184.
(9) Ma, C.; Contento, N. M.; Gibson, L. R.; Bohn, P. W. Redox (30) Yagati, A. K.; Pyun, J. C.; Min, J. H.; Cho, S. Reduced Graphene
Cycling in Nanoscale-Recessed Ring-Disk Electrode Arrays for Oxide-Nanoparticle Modified Microarrays for the Direct Impedimetric
Enhanced Electrochemical Sensitivity. ACS Nano 2013, 7, 5483−5490. Detection of C-Reactive Protein. Bioelectrochemistry 2016, 107, 37−46.
(10) Ma, C.; Contento, N. M.; Gibson, L. R. P.; Bohn, W. Redox (31) Odijk, M.; Olthuis, W.; Dam, V. A. T.; van den Berg, A.
Cycling on Recessed Ring-Disk Nanoelectrode Arrays in the Absence Simulation of redox-cycling phenomena at interdigitated array (IDA)
of Supporting Electrolyte. J. Am. Chem. Soc. 2014, 136, 7225−7228. electrodes:amplification and selectivity. Electroanalysis 2008, 20, 463−
(11) Lee, G. Y.; Park, J. H.; Chang, Y. W.; Cho, S.; Pyun, J. C. Redox 468.
cycling-based immunoassay for detection of carcinogenic embryonic (32) Chan, H. L. Y.; Wong, V. W. S.; Tse, A. M. L.; Tse, C. H.; Chim,
antigen. Anal. Chim. Acta 2017, 971, 33−39. A. M. L.; Wong, G. L. H.; Sung, J. Y. Serum Hepatitis B Surface
(12) Bard, A. J.; Crayston, J. A.; Kittlesen, G. P.; Shea, T. V.; Antigen Quantitation Can Reflect Hepatitis B Virus in the Liver and
Wrighton, M. S. Digital Simulation of the Measured Electrochemical Predict Treatment Response. Clin. Gastroenterol. Hepatol. 2007, 5,
Response of Reversible Redox Couples at Microelectrode Arrays: 1462−1468.
Consequences Arising from Closely Spaced Ultramicroelectrodes. (33) Westh, H.; Hoffmann, S.; Christiansen, E.; Worm, A. M.
Anal. Chem. 1986, 58, 2321−2331. Hepatitis B core antibody screening in a high prevalence group:
(13) Brett, C. M. A.; Brett, A. M. O. Electroanalysis; Oxford university comparison of three enzyme immunoassays using receiver operating
press: Oxford, 1998; pp 9−36. characteristic analysis. J. Virol. Methods 1996, 56, 13−18.
(14) Sanderson, D. G.; Anderson, L. B. Filar Electrodes: Steady-State
Currents and Spectroelectrochemistry at Twin Interdigitated Electro-
des. Anal. Chem. 1985, 57, 2388−2393.
(15) Paeschke, M.; Wollenberger, U.; Lisec, T.; Schnakenberg, U.;
Hintsche, R. Highly sensitive electrochemical microsensensors using
submicrometer electrode arrays. Sens. Actuators, B 1995, 26−27, 394−
397.