You are on page 1of 2

MIIN Parameters (1/2)

* This parameter should be used for reference only. Please check the skin reaction of patients.

Wavelength Spot Size Fluence Energy Repetition No. of


Indication (㎚) (㎜) (J/㎠) (mJ) rate(㎐) Pass

Pigmented Lesions

Melasma (Toning) 1064S/D 8 0.8 ~ 1.6 400 ~ 800 8 ~ 10 3

PIH (Toning) 1064S 8 1.2 ~ 1.6 600 ~ 800 8 ~ 10 2

4 3.58 ~ 6.36 450 ~ 800 1 ~ 2 2~3


Nevus of Ota 1064S
6 2.65 ~ 3 750 ~ 850 1 ~ 2 2~3

ABNOM 1064S 4 5.17 ~ 5.96 650 ~ 750 1 2

Café-au-lait- spot 532 3 0.42 ~ 30 ~ 1 ~ 2 3

Freckles, Lentigines 1064S/D 4 3.18 ~ 5.17 400 ~ 650 1 ~ 2 1~3

Freckles, Lentigines 532 2 - 3 0.28 ~ 1.9 20 ~ 60 1 ~ 2 1~3

Seborrheic Keratosis 532 3 0.99 ~ 1.69 70 ~ 120 1 ~ 2 2~4

Ochronosis 1064S 4 2.78 ~ 3.58 350 ~ 450 2 ~ 5 1~2

1064S 4 3.58 ~ 4.37 450 ~ 550 2 ~ 4 2


Dark Eye Circles
Quasi 5 12.73 ~ 2500 ~ 7 2
15.27 3000
532 3.5 1.24 ~ 1.97 120 ~ 190 1 1
Dark Lips
1064S 3 7.07 500 1 1
Quasi 4 7.95 1000 3 ~ 5 3
Nail Fungus
1064S 4 5.57 700 3 ~ 5 3

Tattoo Removal

Body 1064S 3 or 4 2.38 ~ 6.7 300 ~ 850 1 ~ 2


Black & 1
Blue
Eyebrow 1064S 3 or 4 0.79 ~ 5.17 100 ~ 650 1 ~ 2

Red, Orange, Purple 532 3 or 4 1.1 ~ 2.38 140 ~ 300 1 ~ 2 1


MIIN Parameters (2/2)
* This parameter should be used for reference only. Please check the skin reaction of patients.

Indication Wavelength Spot Size Fluence Energy Repetition No. of


(㎚) (㎜) (J/㎠) (mJ) rate(㎐) Pass

MIIN Mode

MIIN 4 19.89 2500 1 1


Pigment
one shot only

Step 1 MIIN 9 3.45 2200 5 ~ 10


Full Face
2pass
Step 2 MIIN 7 5.71 2200 5 ~ 10

NANO pigment
Step 3 MIIN 5 10.2~12.7 2000 ~ 2500 ~ 3 area
GENESIS
(Melasma)

If the reaction of the skin in each step becomes very swollen or


reddened, stop the procedure and then cooling down.

MLA Handpiece Please use MLA handpiece with MIIN Mode.

1064S 10 1.01 800 5 1 ~ 3


Pore
MIIN 10 1.9 ~ 2.5 1500 ~ 2000 3 ~ 5 1 ~ 2

Stacking
1064S 10 1.14 900 5 ~ 7
Scar times

MIIN 10 1.9 ~ 2.5 1500 ~ 2000 3 ~ 5 1 ~ 2

● If the skin is severely swollen or reddened depending on the skin reaction,


lower the energy level.
● When the Tattoo removal (1064nm / 532nm), test shots to find the energy that whitening occurs.
● Do not perform the stacking shot when treat tattoo removal at 532 nm.
● When 1064nm and 532nm are used together, 1064nm should be performed first.
● Depending on the patient, it is advisable to inform the patient in advance that blistering or
swelling may occur after the procedure.