Professional Documents
Culture Documents
AUTOsoft
AUTOsOFT
Control- and evaluation software
for Autobio LUMO and PHOMO
Software Version 2.6.9, Build 160608
Document ID M S01 100 E
Document Version 2.8
Publishing Date 2016-09-20
Contact
If you have any questions related to AUTOsoft, please contact us by email with a short
description of your question to the address below. Any suggestions for improvements to
our products and services are welcome.
Autobio Labtec Instruments Co., Ltd.
No.199 15th Ave, National Eco & Tech Zone, Zhengzhou, Henan, China 450016
e-mail: info@autobio-diagnostics.com
phone: +86 371 6798 5313
fax : +86 371 6798 5804
homepage: www.autobio.com.cn
Operating Manual
1. Introduction AUTOsOFT
Manual Warranty Disclaimer
Except as provided in writing signed by an officer to The Manufacturer, this system and
any related documentation are provided "as is" without warranty of any kind, expressed
or implied, including that the system is "error free". This information is presented in good
faith, but The Manufacturer does not warrant, guarantee, or make any representations
regarding the use or the results of the use of this system and related documentation in
terms of correctness, accuracy, reliability, currentness, omissions, or otherwise. The entire
risk as to the use, results and performance of this system and related documentation is
assumed by the user.
Except as expressly provided herein, The Manufacturer makes no other warranty, whether
oral or written, expressed or implied, as to any matter whatsoever, including but not
limited to those concerning merchantability and fitness for a particular purpose, nor is
freedom from any patent owned by The Manufacturer or by others to be inferred.
Limitation of liability
The Manufacturer shall not be liable, to any extent whatsoever, for any damages
resulting from or arising out of the use or performance of this system and related
documentation or the procedures specified in this manual, regardless of foreseeability or
the form of action, whether in contract, tort (including negligence), breach of warranty,
strict liability or otherwise, and including but not limited to damages resulting from loss
of data, loss of anticipated profits, or any special, indirect, incidental or consequential
damages.
In no event shall The Manufacturers liability to the user exceed the amount paid by the
user to The Manufacturer hereunder. The user assumes full responsibility for the results
obtained from the use of this system and related documentation and for application of
such results.
Trademarks
Microsoft, Windows, and Windows NT are either registered trademarks or trademarks of
Microsoft Corporation in the United States and/or other countries. IBM is a trademark of
International Business Machines Corporation in the United States, other countries, or
both. Pentium is a registered trademark of Intel Corporation. Adobe, Acrobat, and
Postscript are either registered trademarks or trademarks of Adobe Systems Incorporated
in the United States and/or other countries. All other trademarks appearing in this
manual are owned by their respective companies.
Copyright
Copyright by Autobio Labtec Instruments. All rights reserved. No part of this publication
may be reproduced, distributed, or transmitted in any form or by any means, electronic,
mechanical, photocopying, recording, or otherwise, or stored in a database or retrieval
system, without the prior written permission of Autobio Labtec Instruments.
1 Introduction .......................................................................................1
2 Installation, Launch, Initial Setup .......................................................3
2.1 Installing AUTOsoft .........................................................................................................................3
2.1.1 System Requirements ...............................................................................................................3
2.1.2 Special requirements for Widows Vista ....................................................................................3
2.1.3 Running the Setup Program .....................................................................................................4
2.2 Installation of USB-driver (optional) ................................................................................................5
2.3 Running the Autobio Software .......................................................................................................7
2.4 Error logging ..................................................................................................................................7
2.5 Initial setup of AUTOsoft.................................................................................................................7
2.6 Configuring Options .....................................................................................................................10
2.6.1 Data Export ............................................................................................................................10
2.6.2 Item Classification ..................................................................................................................11
2.6.3 Others ....................................................................................................................................12
2.6.4 Sample Info ...........................................................................................................................14
2.7 Configuring the Instrument ..........................................................................................................14
2.8 Directory .......................................................................................................................................17
7 Tools ................................................................................................ 85
7.1 Internal QC ...................................................................................................................................85
7.1.1 Define QC Sera (Data | QC Serum) .........................................................................................87
7.1.2 Define QC Project (Data | View Project) .................................................................................88
7.1.3 View and add QC Data (Data | QC Data) ...............................................................................90
7.1.4 Signing (Setting | Signing) ....................................................................................................93
7.1.5 Print (Setting | Print) .............................................................................................................93
7.1.6 Initialize QC Data (Setting | Initialize QC Data) ......................................................................95
7.2 Export Data ...................................................................................................................................95
7.3 Backup/Restore Database ..............................................................................................................97
7.4 Repair Database ............................................................................................................................98
7.5 Compress Database ......................................................................................................................98
7.6 Initialize Database.........................................................................................................................98
7.7 Log Viewer....................................................................................................................................99
Instrument control
Designed as a common platform, all microplate based measurement instruments
manufactured and marketed by Autobio are seamlessly integrated into AUTOsoft.
We have made all efforts to make routine operation and assay setup with AUTOsoft as
simple and intuitive as possible, well aware that with the variety of equipment our
customers are using every day, only the best products can provide best support to the job.
User administration
AUTOsoft provides access control by user name and password for 2 different
authorization levels.
Evaluation
Quantitative evaluation: Samples of unknown concentrations can be quantified using
various curve fits (linear regression, point to point, cubic spline and 4 parameter fit) with
linear and logarithmic axis scaling and optional extrapolation. Standard curves can be
saved and reused later on.
Qualitative analysis: based on up to 5 different cut off values can be used to assign
samples to up to different groups.
Semi-quantitative evaluation based on mathematical transformations.
Replication Formula combines the values of two replicates in a flexible formula to
provide a value different from the mean value to be used for further evaluation.
Titer Determination to determine titer classification across a series of dilutions of the
sample
5. When the software installation is complete, choose [Close] to exit the setup
program.
6. Restart your computer and AUTOsoft is ready to use.
Table 2-3: Installing AUTOsoft – steps for setup
Please Note!
The use of a USB cable different from the one supplied with the instrument may cause
communication problems.
Click on the [Advanced] button and a new window is displayed titled 'Advanced
Settings for COMX' (X is the corresponding port number) see Fig. 2-3
Click the [Defaults] button to reset all previous settings to their default value.
Perform the following changes to the advanced settings as shown in Fig. 2-3
Change the value for 'Latency Timer (msec)' to 1
Change the two timeouts 'Minimum Read Timeout (msec)' and 'Minimum Write
Timeout (msec)' both to 1000.
! Please Note:
The name of the user will appears in the bottom of the main window during running
AUTOsoft and on the protocol printouts for tests.
Change Password appears (see Fig. 2-7).
! Please Note
Only the user 'admin' authorized to create new, or remove obsolete users and assign/reset
passwords for other users!
Note:
The initial/default password for Common Level users is 123.
The initial/default password for Super Level user is 000.
The initial password for the Administrator user is admin.
Only the admin-user has the right to create or delete users and reset their passwords to
default.
2.6.3 Others
This tab summarizes two system related settings for the language and the printer colour
configuration.
Language
Depending on the individual version AUTOsoft offers the possibility to select between a
number of operating languages.
From the Setup menu, select Language. Chose the desired language and click [OK]. You
will be asked to restart the application to run it in the selected language.
Printer Colour
Depending on the type of printer used the printouts can either use color or black and
white. Especially when using a black and white printer (e.g. laser printer) legibility is
improved when using the corresponding setting (black).
Unit
If LUMO is selected in Instrument Type, you can choose the light units based on
requirements. If RLU is selected, actual measurement value is displayed as the test result;
If ALU is selected, the actual measurement value is divided by 1000 and then displayed as
the test result. The default light unit is ALU.
Laboratory Name
The optional text entered in this field is displayed in the right top corner of all report
printouts. For examples see Appendix 1.
In order to verify the correct connection setting return back to Main Screen and click on
[Reading] icon. The second entry in the status-line at the bottom of the screen indicates
the status of the connection to the instrument. When the Reading function is opened the
first time after starting AUTOsoft the connection is established which may take a short
while.
Common Buttons
All tabs under Directory have the following common buttons:
Change Modify the selected item
New Create a new item
Delete Delete the selected item
Save Save a newly defined or modified item
Exit Back to Main screen.
! Unsaved changes will be discarded.
Section Office
… can be used to store the names of the various sections, departments, within the
organization supplying the samples. The list created here is used to assign the related
'Section Office' to a sample.
The number (No.) can be freely assigned and can also be used to refer to existing
numbering systems within the organization.
Region
… can be used to store the names of the various regions the samples processed in the
lab are originating from. The list created here is used to assign the related 'Region' to a
sample.
The number (No.) can be freely assigned and can also be used to refer to existing
numbering systems within the organization.
Deliver Doctor
… can be used to store the names of the doctors delivering patient samples to the
laboratory. The list created here is used to assign the label 'Deliver Dr.' to a sample.
The number (No.) can be freely assigned and can also be used to refer to existing
numbering systems within the organization.
‘Section Office’ is used to assign a Deliver Doctor to a certain Section Office. After
selecting an entry of Section Office in the Sample window, all doctors assigned to this
entry will appear in the drop-down list of Deliver Doctors.
Sample Type
… can be used to store a list of sample types used for assays processed on AUTOsoft in
your laboratory. The list created here is used to assign the label 'Sample Type' to a sample.
The number (No.) can be freely assigned and can also be used to refer to existing
numbering systems within the organization.
Unit
… can be used to store a list of units for quantitative test results. The list created here is
used to assign the label 'Units' in the 'Quantitative'-tab of the 'Test Definition'.
The number (No.) can be freely assigned and defines the order of the entries in the list.
Remark
The number (No.) can be freely assigned and defines the order of the entries in the list.
The defined remarks can be selected from a picklist when defining a test in the
'Reference'-tab (see 4.1.6).
Reagent
… can be used to store a list of Reagents used for assays processed on AUTOsoft in your
laboratory. The list created here is used to assign the Reagent and it's specific data used
for a test when the result are stored by clicking [Save] in the Start Test window after the
Test is processed.
The "Reagent-data" is show when saving a plate, in the table of Saved Plates and on the
related printouts.
No. The number (No.) can be freely assigned and defines the order of the
entries in the list. In case of the Reagent the "No." is added in front of
the Name when it's shown in the list of Reagents for selection.
Name Name of the Reagent/kit
Manufacturer Manufacturer of the Reagent/kit
Batch Batch of the Reagent/kit
Expiry Date Expiry date of the Reagent/kit
External Tests
This tab can be used to save data on commonly used tests which are not performed on
the connected instrument and evaluated in AUTOsoft, but which can be saved in the
'Results'-tab of a sample to complete the sample related data (see 6.1).
No. The number (No.) can be freely assigned and defines the order of the
entries in the list. In case of the Reagent the "No." is added in front of
the Name when it's shown in the list of Reagents for selection.
E-Name Alternative Test Name (optional)
Test Name Name of the Test
Unit The unit of the result in case the test provides a quantitative result
Qualitative In case of a qualitative Test the possible results for the individual test
can be saved in this field (optional)
Reference 1-6 Additional information to be added to the result (to be shown in the
sample printout)
The functions can be accessed through the items shown in the menu bar and the most
common ones can be reached by clicking on the corresponding icons.
3.1.3 Toolbar
By clicking on the corresponding icon some of the main functions accessible trough the
Menu Bar are directly started.
Reading | Quick Test (see 8.1)
Class
Class (Left) Select the class the programmed test should be assigned to
(see 2.6.2)
Sub-class (right) Select the Sub-class (child) the programmed test should be assigned
to
[…] opens the window 'Item Classification' accessible through
Tools|Options|Item Classification (see 2.6.2)
Test Name
Test No. The number of the test as shown under 'No.' in the table of tests.
You can use Test Numbers to structure your saved Tests according to
your requirements by using number groups like e.g. 1xx, 2xx, 3xx for
specific groups of test (xx will then be the number of the individual
Test in the Test group).
Pr. Index The order of individual test results if several of them are printed for
a specific Sample ID.
E-Name English name. Can be used if a Test has an English name beside the
Units
The unit of the concentration determined in the quantitative
evaluation. Pick the desired value from the list or enter it manually.
The list displayed for selection of the Unit can be created/modified
under Directory > Unit (see 2.8)
Standards
Please Note: Tests using a saved standard curve
If a test run should be performed using a saved reference curve the
same Test can be used without modification. AUTOsoft automatically
determines if the layout used for a test run incorporates all standards
defined in the 'Quantitative'-tab of the test. If the layout includes no
standards or only one or two standards the user will be prompted to
select a previously stored stand curve.
Depending on the number of standards present in the layout the
selected standard curve will either be used without adjustment (no
standards) or corrected based on one or two standards present in the
layout.
[Qualitative] Select this option if the test should also perform a qualitative
evaluation of the concentrations determined in quantitative
evaluation.
Concentration
Enter the corresponding concentration values for the standards.
Groups
AUTOsoft supports up to 5 cutoff values and accordingly the assignment of samples to
up to 6 groups. Cutoff 1 has the lowest and Cutoff 5 highest value. Accordingly the
assignment is performed as follows: Group 1 < Cutoff 1 < Group 2 < Cutoff 2 < Group
3 < Cutoff 3 < Group 4 < Cutoff 4 < Group 5 < Cutoff 5 < Group 6
Replicate Rejection
Replicate Rejection is used to eliminate values within replicated standards, controls and
samples from further calculation based on specific criteria defined by formula.
Formula The Formula defines the criteria which has to be met for each individual
replicate. If a replicate doesn't meet the criteria it is rejected and not used
for further calculation.
The formula may contain:
- Controls (NC1 to NC6 and PC1 to PC6)
- Standards (S1 to S8)
- 0-Standard (B0)
- Numerical constants
- Variables V1 to V6 (see 4.1.7.4)
- min(a;b), max(a;b) returns the minimum or maximum of the two values
- pwr(a;b) calculates 'a' by the power of 'b' ab
- Mathematical operators (, ), +, -, * ,/ ,<=, >=, =, >, <
- Logical operators AND and OR are used to combine different criteria
Please Note: 'AND' and 'OR' have to be separated from the adjacent parts
of the formula by a space before and after the logical operator.
Example:
PC1*0.9<PC1X<PC1*1.1 (all replicates of PC1 have to range within +-
10% from the mean value of PC1)
PC1X>100000 (all replicates of PC1 have to be greater than 100000)
Please Note: Make sure Controls and Standards used in the formula are
also available in the layout used to run the test!
Base 'Meas. Value' (depending on the measurement method) or
Calculated concentration – as determined for the individual well
Shaking
Enable Shaking Tick the 'Enable Shaking'-box to perform shaking prior to the
measurement with the following settings.
Intensity Low, Medium, High
Time 1 to 99 seconds
Table 4-5: Test Definition Parameters - Shaking
'Scan Mode' is used for agglutination data reduction with titer determination based on
an external calculation module called AUTOaggl (see 4.1.7.2)
'Variables' are used to replace fixed numerical values in formulas used in the test
definitions. The values can be checked and modified when the test is started. This feature
Please Note! If 'Titer Test' is activated in the 'Advanced'-tab 'Cutoff 1' is used to
separate negative samples (in Group 1) from positive samples (in Group 2). Additional
groups above Group 2 can be defined but will be irrelevant for Titer determination.
Layout
The Layout required for a Titer Test has to use the number of replicates per sample ID
corresponding to the number of required dilutions of the samples as shown in Fig. 4-7
(example with 8 replicates).
Please Note! Only samples positioned in the layout will be processed to determine the
corresponding Titer. If case special positions (PC, NC and K) are positioned in the layout
the Titer will not be calculated for them, but the values will be available for definition of
the Cutoff and for Validation.
Fig. 4-7: Layout definition with 8 replicates for Titer Evaluation (example)
Advanced_Titer_Layout.png
Factors
Test programming
General information on the various Test-tabs are provided in chapter 4.1.
General–tab: Select the desired wavelength to be used for the measurement. Make
sure the 'Qualitative is selected as 'Mode'.
You may enter a 'Transformation' formula and use the Transformation results as basis for
the Qualitative classification of each individual sample dilution. This may be particularly
useful, to correctly assign Titers in case the reaction provides higher measurement values
for higher dilutions. In this case a formula like X'=1/X will reverse the order and after
applying the same calculation to the Cutoff (see below) the titers for the samples will be
correctly assigned.
Select the previously defined layout as described above as 'Default Layout'.
Qualitative-tab: Define the value for Cutoff 1 to distinguish negative and non
negative (positive or equivocal sample dilutions) with negative samples assigned to
Group1 and positive or equivocal samples to Group 2 and higher.
In order to correctly perform a Titer Test, values below Cutoff 1 have to be considered
negative, values above Cutoff 1 as positive. For the Titer Determination equivocal samples
are considers positive. The names assigned to the groups can be freely defined.
If the assay shows a reversed ratio between dilutions and measurement values the
following message is displayed:
'TEST INVALID: Lower dilutions have to show higher meas. values and higher dilutions
lower meas. values. In this test the order is reversed so the results are invalid!'
In case you have defined a Transformation formula select the 'Transformation' as basis for
the further evaluation. In case special positions are defined in the layout these will not be
processed to provide Titers, but the measurement value (or in case of replicates the mean
measurement value) can be used in a formula to define the required cutoff.
The following table shows an example for better understanding of the calculation
performed. In this case position A6 with a value of 0.223OD is the first non negative
sample (assigned to Group 2) and therefore used to determine the sample specific Titer,
in this case 1/1280.
! Please Note
In order to use the External Calculation feature Microsoft Excel® has to be installed on
the computer AUTOsoft is used on!
By default all related Excel files are located in the directory '\Advanced' below the
AUTOsoft installation directory.
Evaluation workbook
The evaluation workbook to be used to perform the external evaluation is selected in the
'Advanced'-tab in the Single Test definition as 'Calculation File'.
The first worksheet in the evaluation workbook (independent of its name) contains the
qualitative- and quantitative results to be transferred to AUTOsoft and an area defining 6
variables to be optionally defined, modified at the start of the test and used in the
individual calculation similar to the functionality described in 4.1.7.4.)
Content Location Description
Qualitative Results C16:N23 An 8x12 matrix including the qualitative results for
Special considerations
Quantitative results:
Concentrations are checked against the valid range defined in the Quantitative Test tab.
Concentrations exceeding the limits of the standard curve defined in AUTOsoft are
marked as <X or >Y (X=concentration of lowest and Y=concentration of highest
standard) unless Extrapolation (Down and/or Up) is selected.
When using external evaluation to calculate concentrations AUTOsoft will show no
standard curve in the Result-tab or the Matrix printout.
Sample concentrations below the value set in 'Min. Sensitivity'
… are displayed as <X (X=value set for Min. Sensitivity)
Example
The AUTOsoft installation will automatically copy the following files into the installation
directory to be used as an example for external evaluation. When running the test make
sure to select 'Virtual' (in the bottom left corner of the 'Start Test' window in order to use
the virtual measurement data file provided (see below).
External_Example.xls External evaluation file to demonstrate the
external evaluation function.
External_Example.txt Virtual measurement data file (with OD values) to
be used with External_Example.xls. Select this file
as 'Simulation Data' under 'Instrument'.
TestSave_External_Example.tst Test file for import ('Single' [Import]) to be used
with External_Example.xls. The test performs a
photometric measurement on PHOmo and an
external evaluation based on External_Example.xls
Please Note: After import of the Test and Layout
open the newly imported test and check the
assignment of the default layout in the 'General'-
tab and the path of the Calculation File in the
'Advanced'-tab.
LayoutSave_External_Example.plt Layout file for import ('Single' [Import]) to be
used with External_Example.xls
Table 4-10: Files to be used as example for external evaluation
Variables V1 to V6
Value Defines the value assigned to the related Variable. The value can either
be defined by a numerical value or by a formula using:
- Controls (NC1 to NC6 and PC1 to PC6)
- Standards (S1 to S8)
- 0-Standard (B0)
- Numerical constants
- min(a;b), max(a;b) returns the minimum or maximum of the two
values
- pwr(a;b) calculates 'a' by the power of 'b' ab
- Mathematical operators (, ), +, -, * ,/
Please Note: Variable formulas must not use X, V, F, R
Description An optional descriptive text can be entered to describe each individual
variable. The Description is displayed together with the assigned value
under 'Test Results' 'Parameters' and listed in the Table Report.
Check If 'Check Variables' is activated the list of Variables is displayed after
Variables starting the related Test, the values can be modified and the test run can
be continued by clicking [OK]. In this case the programmed test is
updated with the modified values as soon as the results are saved. The
next time the Test is started the updated values are displayed as default.
If the Variable values are modified in a previously measured and saved
Plate which has been opened and recalculated the variable values are
only changed for the opened plate and the Variable settings in the
original Test remains unchanged.
Type
Select the type of sample to be placed in the template
P Patient sample
S Standard (S1 to S8)
NC Negative Control (NC1 to NC6)
PC Positive Control (PC1 to PC6)
BL Blank (will automatically be deducted from all measurement
values prior to any other calculation)
B0 0-Standard
K Control (K1 to K6)
Current Type Click up or down to manually increment or decrement the number of
the sample type selected (see above for numbers per type).
Filling Direction
Vertical or Defines the filling direction when using the [Fill] function (see below).
Horizontal Vertical = 8-way, Horizontal = 12-way
Fill Fills all empty - not previously assigned wells - with samples
according to the direction and replication parameters defined above.
Clear Removes all samples assigned to the template.
Clear All Empties the template by removing all samples and special positions.
View
Layout Displays the plate layout
Factor Displays the dilution factors for each sample location. By default the
factor is set to 1. (e.g. A dilution of 1:10 would have to be defined as
10)
The Factor is automatically used to adjust Calculated Concentrations
and the Results of the Transformation Formula by multiplying the
original result of the calculation by the Factor.
Test – tab
Select [New] and enter the name of the Multitest to be created under 'Name:' Click OK.
In 'Filter', select: 'EMPTY' for a Luminescence Multitest or
The desired filter (from the list displayed) for Photometric Multitest
In 'No. of Tests', select the number of Tests to be used within the Multitest (up to 12).
In T 01, T 02, …, select pre-programmed tests. The tests displayed for selection either
use the same 'Filter' or are Luminescence Tests (see Fig. 4-13).
Click [Save] to save the current settings. Only after clicking [Save] the selected Tests
are made available in the following step Layout-definition.
Date Range
From / To As Multitests have to be defined together with a specific Layout to
accommodate the number of samples to be tested the Date Range
for stored Multitests can be defined. Only Multitests within the
specified Date Range are displayed in the List of Tests.
Layout-tab to define the layout. In 'Current Test', select the first Test and define its
layout corresponding to the procedure also used for Layout definition for a Single Test
Date Range
From / To see above
[Refresh] see above
Test
Current Test Is used to select the Test for which samples should be positioned in
the Multitest Layout. Only Tests are listed which have been selected
in Test-tab of the 'Multitest Definition'.
No. of Samples Defines the number of samples which should be positioned in the
layout for the 'Current Test'
Type
Is used to select the type of sample to be positioned (see 4.1.7)
Filling Direction
Defines the filling direction and provides the possibility to Fill or Clear
samples automatically (see 4.1.7).
! Please Note
After starting AUTOsoft and preparing the first test run as mentioned above by clicking
the related icon or selecting the menu option AUTOsoft automatically establishes
connection to the instrument connected and shows the message 'Instrument connected!'
in the status line at the bottom of the screen. Please make sure your instrument is turned
on and connected to the PC before clicking getting ready for a measurement.
The 'Start Test' window is opened which holds three tabs used to define the necessary
data for the Test and view the results.
Test Parameter – tab (see 5.1)
Sample IDs – tab (see 5.2)
Results – tab (see 5.3)
At the bottom of the Start Test window a number of buttons common to all tabs is
displayed.
[Start] Starts the Test after all settings have been performed. After the Test
run is finished the related result data is displayed.
If communication cannot be established to the instrument the
following message box is displayed:
Click [Yes] to start the new test. Click [No] to cancel the test and wait
to save the last test results.
Samples_Results_Kindling.png
Results – tab
RLU or OD Measurement results (see Fig. 5-6)
Change values or manually reject a value
If necessary the measurement value can either be altered by double-
clicking on the corresponding well and entering the desired value.
Alternatively a right mouse click will open a box showing two options
to be applied to the selected well:
'Modify Value': corresponds to a double-click on a well position. The
original value is displayed and can be altered
'Reject Well': the selected well is excluded from all further calculations.
Please Note!
In order to highlight the manual change the measurement-value (OD or
RLU) and all results calculated based on the changed value are
displayed and printed in italic letters! Rejected values are displayed in
strike through format. Changes and manual rejections applied to
measurement values are logged together with user, date and time. The
logged changes can be viewed under Tools | Log Viewer (see 7.7)
The high values which are displayed in scientific notation to fit into the
matrix can be viewed in total by moving the cursor into the
Name
Shows the list of all available, previously saved Reference Curves.
To select a curve left-click on it's list entry and the related curve
data will be displayed.
Please Note: The list of saved reference curves is automatically
filtered to only show saved curves which have been generated
with the same test.
Standards
X-Value Shows the concentration (X-Value) as defined in the test used to
Y-Value generate the curve and Y-Values - the measurement values (or
transformation value if a formula is defined).as determined during
the test run.
Curve + Corresponds to the entries in the 'Quantitative-tab during test
Axis (X/Y) definition (see 4.1.2)
Please Note: If the user logged on with AUTOsoft is 'admin' or a
'Super User' the values can be changed manually and the changes
can be saved. Changes applied to the values are only visualized in
the curve graph once the [Save]-button as been pressed.
[Save] Saves changes made to the Reference Curve data (see above) and
recalculates/updates the curve graph.
Graphics
Displays the standard curve graph of the selected Reference Curve.
Common Buttons
[New Curve] This button is used to create a new 'Reference Curve' with manually
In case the layout of a test includes Controls (K1 to K6), a Cutoff in Qualitative evaluation
and QC-Project(s) have been defined under 'Internal QC' the right section of the 'Save'-
Window shows picklists to select the desired QC-Project the QC Data should be saved in.
If the Controls of the measured plate should not be saved select the empty position in
the picklist.
The results of a Multitest are displayed corresponding to the colour code used in the
layout.
6.1 Samples
The Samples-tab is used to:
Add, modify and view detailed information to individual sample IDs
Create new sample IDs with the related patient information
Import sample IDs from a text file
View Results for a selected sample ID
Print result reports for selected Sample IDs
The Samples-tab offers all relevant information in one window separated into different
areas, which are described in the following chapter.
Date Range
From Select start date for the time interval to display Sample IDs.
To Select end date for the time interval to display Sample IDs.
[Refresh- Updates the Sample List according the 'Date Range' defined.
button]
Samples
Sample List 1st Column of the Sample List shows a symbol indicating if Test Results
are assigned to the Sample ID or not.
… no results assigned. Sample ID can be modified
… with results assigned. Sample ID cannot be modified
… results printed. Sample ID cannot be modified
2nd Column displays the Sample ID
3rd Column the Name of the Patient (last name, first name)
By clicking on the column header the displayed entries are sorted
according to the selected column. By sorting the first column
previously entered samples can easily be grouped together to select
them for positioning onto a plate.
Please Note: If the plate(s) containing the related result(s) of a sample
is/are deleted the related results are not available anymore.
Accordingly the results are not anymore displayed in the result list and
sample status indicator will turn red to show that no results are
available for the particular sample.
Please Note: The settings for the prefix for batch generation of sample
IDs are not used to create the new sample ID but only the standard
structure!
Please Note: The new Sample ID is only created in the database if the
[Save]-button is clicked afterwards.
[Save] Will save all additions and alterations to the Sample Information.
You can alternatively press F2 to save the added or modified data.
Please Note: This [Add]-button will only save Sample ID data but not
changes to related results!
[Import] After clicking the [Import]-button the following window is displayed.
From File: is used to import a list of Sample IDs and related data from
an external text file.
Select the 'From File'-option and click on the […]-button to open a file
selection box. Select the file to import. Click [Start] to perform the
import of the Sample IDs which are then displayed in the 'Samples-
table.
If the file format is not correct a warning is displayed and the import is
cancelled. In this case please check the import file and correct it
- The following list showing all previously saved 'External Tests' (see
2.8) is displayed. Select the Tests that should be assigned to all
selected Samples and click [Save].
- The selected external test entries are added to the samples and the
results of all selected sample IDs are displayed in the Results list. This
step does not enter an actual result value but just create a result
entry with the information previously saved with the 'External Tests'.
- Select the previously added result entries one by one in the Results
List and fill in the specific values/results for the External Tests and
click [Save]. Proceed with the next added result entry.
[Delete] … will delete the selected Result.
[Save] Saves a newly added or modified Result.
For Results you can alternatively press F3 to save the modified data.
Please Note!
If you don't click [Save] before [EXIT] or selection of another Sample ID
your previously made changes will be discarded.
[Print] Prints the Result report on the selected samples. If multiple samples
are selected a separate Report will be printed for each Sample ID
selected including all Test Results related to the selected sample. For
an example see Fig. 10-1
[Exit] Exit 'Samples'. Return to Main Screen.
Plates - tab
The 'Plates'-tab is divided into two sections:
Upper section
… lists all plates meeting the selected search criteria. All 3 criteria entered have to be met
in order to show a plate in the list.
Test Type Single: Only Single Test are listed
Multi: Only Multitest are listed
The content and view of this tab is identical with the 'Test definition' window with it's
seven tabs (for details see 4.1).
Changes made to the Test Definition are only saved for the individual plate and do not
change the settings for the Test originally used to run the Test.
7.1 Internal QC
Internal QC is a function of AUTOsoft designed for quality control purposes in the
laboratory. It is used to monitor values obtained for controls which are defined in the
layout of a test and measured and evaluated in each test run.
Various criteria can be select to check the status of the controls (QC-Sera) based on a
series of previous measurements of the same QC-Sera under the same test conditions.
Additionally values for QC-sera can also be manually added or deleted.
A report function provides a printed protocol including all relevant data on a selected
QC-project.
Basic procedure
Define a layout for a test (or a group of similar tests) including at least one control (K1
to K6) (see 4.1.7).
Define a test using the layout created above. The test has to include qualitative
evaluation (also combined with quantitative evaluation) and the related Cutoff
definitions. Select the desired Cutoff Value definition to be used to calculate M/CO
(see 4.1.3).
Define one or more 'QC Sera' representing the Controls in the QC calculations (see 0).
Create one or more 'QC Projects' which define the method the results of the 'QC Sera'
is checked with, based on the results of a previous series of measurements of the
same 'QC Sera' (see 7.1.2).
Run the previously defined test.
When saving the results assign the corresponding, previously defined 'QC Project' to
the control(s) on the plate (K1 to K6). The data of the controls is automatically stored
as 'QC Data' to the selected QC Project and used to calculate the related limits for the
Rules based on mean value and standard deviation of M/CO (see above).
QC Data can also be added manually (see 7.1.3)
By using the 'Print' Option (Settings | Print) the necessary QC protocols for each
project can be created (see 7.1.5).
The screen dump below shows an example for 'Internal QC'-main window with some data
of QC-sera previously assigned to the QC-Project 'Demo_K1'.
No. * Can be freely assigned and is used to sort the table of QC-Sera
displayed in the related table.
Name Assign a name to a specific QC-Serum. This name is used to
identify/select the specific QC-Serum as 'QC Item' when defining the
QC-Project (see 7.1.2).
Batch Defines the Batch of the QC-Serum used (optional)
Source Defines the Source (e.g. Supplier/kit) the QC-Serum is sourced from
or linked to (optional)
Concentration Defines the Concentration of the QC-Serum if applicable (optional)
If the QC-Method is 'Real-time' and 'Class' is set to 'Quantitative' this
value is the target value for the concentration of the QC-Sera.
Unit Indicates the unit of 'Concentration' if a concentration is defined
(optional)
StdDev For Real-time method StdDev defines the acceptable standard
deviation for the concentration determined for the QC-Sera.
CV StdDev/Concentration is automatically calculated based on the
corresponding values entered above.
Upper Limit … for concentrations of QC-Sera to be acceptable (Real-time
Method)
= Concentration + StdDev (based on the values defined above)
Lower Limit … for concentrations of QC-Sera to be acceptable (Real-time
Method)
= Concentration - StdDev (based on the values defined above)
Out of Rules
One or more of the following five QC criteria can be selected to check the value of the
QC-Serum selected under 'QC Item'. The criteria are either based on the mean value of the
20 previously measured controls (which proved to be within range) and the standard
deviation of this group of values or the tendency of the values (rising or falling).
Rule Description
one point is out of +-2s One point shows a value exceeding x (mean) ± 2s
four consecutive QC points fall Four consecutive QC point show values which all
on the same side outside of 1s exceed either x (mean) +1s or –1s (they are all
located on the same side of the mean value)
seven consecutive QC points are Seven consecutive QC points show a steady rising or
continuously rising or falling falling tendency.
two consecutive QC points The difference between 2 consecutive QC-points
differ by more than 4s exceeds ± 4s
Project Name of the 'Project' selected (in the picklist) from the previously
defined projects.
Output Mode:
Output Mode Description
Basic Data the top 20 correct data (see Fig. 7-9)
Monitoring Data the other data (see Fig. 7-10)
QC Graphics QC diagram (see Fig. 7-11)
After starting the 'Export Data' function the above screen is displayed.
Date Range From: Start date of selected date interval
To: End date of selected date interval
[Refresh] After selecting the 'Date Range' you have to click the [Refresh]-button
to display the plates/samples meeting the 'Date Range'-criteria.
Saved Plates Displays the list of saved plates according to the selected date interval.
Select the IDs of the plate(s) you would like to have the sample/result
data exported for.
Path By default the path used to save the export data file is the installation
directory of AUTOsoft (e.g. C:\Program Files\Autobio\AUTOsoft\)
The default file name is automatically created in the format:
YYMMDDhhmm.txt (YY=year, MM=month, DD=day, hh=hour,
mm=minute corresponding to the time the Export Data function has
been ) started.
Data Table Shows the sample data according to the plate(s) selected under 'Saved
Plates'
[Export] Click the [Export]-button to export the selected data to the defined file
[Print] Prints a report of the selected plates/samples showing:
Sample ID, Name, Test, Result, RLU, M/CO
[Exit Click [Exit]to return to the main screen.
File format
First line: 'Autobio Co., LTDYYYY-MM-DD,hh:mm:ss'
YYYY=year, MM=month, DD=day, hh=hour, mm=minute, ss=second
Second line: Plate_ID (as assigned when the plate was saved)
! Please Note
Restoring a database backup will replace the existing database of the AUTOsoft
installation. Thus all data stored in the database prior to restoring it will be lost!
Wells
Shows data on manual changes applied to measurement results in the RLU-view.
Operator … logged on during the change was performed
Plate ID … the Sample ID is located on, for which the RLU value was
changed.
Sample ID … for which the RLU value was changed.
Old Value OD/RLU value prior to the manual change
New Value OD/RLU value after the manual change or
'M-reject' to indicate that the value has been manually rejected.
Date Date and time to change has been performed
Please Note:
If the ‘Sample ID’ shows empty, it is shown that the saved plate is deleted, and the ‘Test
Name’ is the saved plate.
The individual features of the Quick Test function are described in the table below:
Instrument Type The type of Instrument on which the test will be delivered
Luminescence Depending on the type of instrument selected the corresponding
or measurement function is activated and the non valid option is
Photometric disabled.
In [Class of Tests], select the Class, In [Saved Tests], select the test. In [Saved Layouts],
select the layout (see Fig. 9-7).
From Start Test, choose NO. Sample IDs, enter the number of sample (for example,8).
Then click Generate (see Fig. 9-8).
Choose Save.
The ‘Save’ Window is
displayed
B. Printing Assay
From Saved Plate, select Samples. Samples appears (see Fig. 9-14).
PHOMO Test
Fitler: 405nm
03.08.2015 12:41:48
OD Values
0,002 0,000 4,500 0,154 0,27 0,385 1,361 2,639 3,903 0,001 4,500 0,002
0,003 0,000 4,500 0,154 0,271 0,384 1,356 2,63 3,727 4,500 0,001 0,004
0,003 0,000 4,500 0,152 0,269 0,383 1,355 2,627 3,692 0,001 4,500 0,004
0,002 0,000 4,500 0,154 0,271 0,386 1,359 2,627 3,727 4,500 0,001 0,003
0,002 0,000 4,500 0,153 0,269 0,384 1,354 2,636 3,852 0,001 4,500 0,002
0,002 0,000 4,500 0,157 0,274 0,390 1,362 2,645 3,806 4,500 0,001 0,002
0,002 0,000 4,500 0,158 0,274 0,390 1,362 2,633 3,806 0,002 4,500 0,002
0,002 0,000 4,500 0,154 0,268 0,384 1,356 2,636 3,765 4,500 0,001 0,003
Scan Data
Cell 1,1 -0,000039 -0,00002 0 0,000098 0,000938 0,010493 0,067237 0,284698 0,619126 0,862946 0,955331 0,984798
0,993239 0,995779 0,996444 0,996287 0,994412 0,967876 0,787678 0,41546 0,149813 0,050726 0,016394 0,004729
0,001466 0,000469 0,000156 -0,00002 -0,000098
Cell 1,2 0,999629 0,999629 0,999668 0,99959 0,999629 0,999629 0,999629 0,99959 0,99955 0,999629 0,999648 0,99959
0,999629 0,99959 0,999629 0,99959 0,99959 0,999511 0,99955 0,99955 0,999511 0,999511 0,99955 0,999453
0,99955 0,999472 0,999472 0,999472 0,999492
Cell 1,3 0 -0,000059 0,000039 0,000117 0,00002 -0,000059 -0,00002 0,00002 -0,00002 0,000039 0,00002 -0,00002
-0,000117 -0,00002 0 0,000059 -0,00002 -0,00002 -0,000078 0,00002 0 0,000039 -0,000059 -0,00002
-0,00002 0,000059 -0,000039 -0,00002 0,00002
...
...
...
Cell 8,10 0,000019 -0,000058 -0,000214 -0,000253 -0,000214 -0,000292 -0,000253 -0,000214 -0,000312 -0,000214 -0,000253 -0,000253
-0,000253 -0,000214 -0,000273 -0,000234 -0,000234 -0,000195 -0,000312 -0,000214 -0,000292 -0,000312 -0,000273 -0,000292
-0,000292 -0,000195 -0,000019 -0,000058 -0,000019
Cell 8,11 0,010198 0,086653 0,327347 0,697709 0,899542 0,966501 0,986078 0,992668 0,994852 0,995944 0,996276 0,996529
0,99651 0,996646 0,996568 0,996646 0,996588 0,996568 0,996451 0,996256 0,995788 0,995262 0,99454 0,993468
0,984791 0,860427 0,52803 0,210471 0,069669
Cell 8,12 0,00002 0,000117 0,000078 0,000176 0,000585 0,004057 0,023894 0,138663 0,426582 0,754993 0,922642 0,973414
0,989604 0,994324 0,995767 0,995884 0,995338 0,989994 0,896524 0,590661 0,256924 0,090934 0,02879 0,009109
0,002789 0,000819 0,000254 0,000117 -0,000078
OD:
-------
0,049 1,466 0,709 1,375 1,069 0,931 0,165 0,874 0,953 1,474 0,954 1,117
0,046 1,483 0,702 1,394 1,085 0,946 0,177 0,879 0,973 1,481 0,960 1,121
0,374 0,418 0,834 0,758 0,206 1,439 1,360 1,263 0,666 1,027 1,448 0,374
0,359 0,912 0,860 0,764 0,208 1,462 1,354 1,295 0,677 1,057 1,482 0,382
0,582 0,731 1,233 0,653 1,024 1,176 0,386 1,205 1,474 1,017 1,321 1,275
0,592 0,748 1,275 0,653 1,018 1,181 0,408 1,225 1,500 1,047 1,315 1,308
0,933 0,229 0,346 0,730 1,465 0,417 0,232 0,476 0,857 0,665 1,364 1,257
0,915 0,225 0,356 0,771 1,505 0,439 0,240 0,479 0,881 0,667 1,394 1,264
Status:
-------
OK OK OK OK OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK OK OK OK OK
OK OK OK OK OK OK OK OK OK OK OK OK
Graphics
Fig. 2-1: Select USB Serial Port in the Device Manager .............................................................................. 5
Fig. 2-2: Restore Default Settings for USB Serial Port ................................................................................ 6
Fig. 2-3: Modify Advanced Settings for USB Serial Port ............................................................................. 6
Fig. 2-4: AUTOsoft desktop icon ............................................................................................................... 7
Fig. 2-5: License (example) ....................................................................................................................... 7
Fig. 2-6: Login admin ............................................................................................................................... 8
Fig. 2-7: Change Password ....................................................................................................................... 8
Fig. 2-8: Add New User ............................................................................................................................ 9
Fig. 2-9: Setup | Options | Data Export .................................................................................................. 11
Fig. 2-10: Setup | Options | Item Classification ........................................................................................ 12
Fig. 2-11: Setup | Options | Others .......................................................................................................... 13
Fig. 2-12: Setup | Options | Sample Info.................................................................................................. 14
Fig. 2-13: Directory | Section Office ......................................................................................................... 17
Fig. 2-14: Directory | Region .................................................................................................................... 18
Fig. 2-15: Directory | Deliver Doctor ......................................................................................................... 18
Fig. 2-16: Directory | Sample Type ........................................................................................................... 19
Fig. 2-17: Directory | Unit ........................................................................................................................ 19
Fig. 2-18: Directory | Remark ................................................................................................................... 19
Fig. 2-19: Directory | Reagent .................................................................................................................. 20
Fig. 2-20: Directory | External Tests .......................................................................................................... 21
Fig. 3-1: Main Window .......................................................................................................................... 22
Fig. 4-1: Test Definition General ............................................................................................................. 27
No.199 15th Ave, National Eco & Tech Zone, Zhengzhou, Henan, China 450016