Instruction manual
Contents
Warranty
Defects occurring within 3 years from delivery date shall be
remedied free of charge at our plant (carriage and insurance Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
paid by sender). Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Sensors, fittings, and accessories: 1 year. Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . .7
Subject to change without notice Overview of Stratos e 2402 Oxy . . . . . . . . . . . . . . . . . . . .9
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Package contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Mounting plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Return of products under warranty Pipe mounting, panel mounting . . . . . . . . . . . . . . . . . . . . . . . .12
Please contact our Service Team before returning a defective device. Ship Installation and connection . . . . . . . . . . . . . . . . . . . . . . .14
the cleaned device to the address you have been given. If the device has Information on installation . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
been in contact with process fluids, it must be decontaminated/disinfected Terminal assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
before shipment. In that case, please attach a corresponding certificate, for Wiring examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
the health and safety of our service personnel. Protective wiring of relay outputs . . . . . . . . . . . . . . . . . . . . . . .18
User interface and display . . . . . . . . . . . . . . . . . . . . . . . .20
Disposal Operation: Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Please observe the applicable local or national regulations Safety features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
concerning the disposal of “waste electrical and electronic Sensocheck, Sensoface sensor monitoring . . . . . . . . . . . . . . . .23
equipment”. GainCheck device self test . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Automatic device self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Hold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
To activate the Hold mode from outside . . . . . . . . . . . . . . . . . .25
Knick Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Elektronische Messgeräte Menu structure of configuration . . . . . . . . . . . . . . . . . . . . . . . .27
GmbH & Co. KG Overview of configuration steps . . . . . . . . . . . . . . . . . . . . . . . .28
P.O. Box 37 04 15 Output 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Output 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
D-14134 Berlin
Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Calibration mode, alarm settings . . . . . . . . . . . . . . . . . . . . . . .46
Phone: +49 (0)30 - 801 91 - 0 Limit function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Fax: +49 (0)30 - 801 91 - 200 Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
Control of rinsing and cleaning systems . . . . . . . . . . . . . . . . . .54
Internet: http://www.knick.de
knick@knick.de
3
Parameter set 1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56 Safety information
Default settings of parameter sets . . . . . . . . . . . . . . . . . . . . . .57
Parameter set, individual settings . . . . . . . . . . . . . . . . . . . . . . .58
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 Be sure to read and observe the following instruc-
Calibration to saturation (SAT) . . . . . . . . . . . . . . . . . . . . . . . . .62 tions!
Calibration to concentration (Conc) . . . . . . . . . . . . . . . . . . . . .64
Zero calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
The device has been designed in accordance with the state of
Product calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
Adjusting temp probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70 the art and complying with the applicable safety regulations.
Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70 When operating the device, certain conditions may neverthe-
Diagnostics functions . . . . . . . . . . . . . . . . . . . . . . . . . . . .71 less lead to danger for the operator or damage to the device.
Controller functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
PID controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74 Caution!
Pulse length / pulse frequency controller . . . . . . . . . . . . . . . . . .76 Commissioning may only be carried out by trained experts.
Connecting a rinsing system . . . . . . . . . . . . . . . . . . . . . . .77 Whenever it is likely that protection has been impaired, the
Operation with automatic cleaning system . . . . . . . . . .77
device shall be made inoperative and secured against un-
Error messages (error codes) . . . . . . . . . . . . . . . . . . . . . .78
Calibration error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . .80 intended operation.
Operating states . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
Sensoface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82 The protection is likely to be impaired if, for example:
Sensocheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83 • the device shows visible damage
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85 • the device fails to perform the intended measurements
Product line and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . .85 • after prolonged storage at temperatures above 70 °C
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86 • after severe transport stresses
Explosion protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
Warnings and notes to ensure safe operation . . . . . . . . . . . . .93
Control Drawing CSA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94 Before recommissioning the device, a professional routine test
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98 in accordance with EN 61010-1 must be performed. This test
should be carried out by the manufacturer.
Trademarks
70084
• Instruction manual
Do not
1 Output 1 9 + Output 1
connect!
Stratos® XXXX X
Safety precautions . . . . . . . . . . . . . . 3
10
Sicherheitshinweise . . . . . . . . . . . . . 5
- Output 1/2
Consignes de sécurité . . . . . . . . . . . 7
Cathode 2
Norme di Sicurezza . . . . . . . . . . . . . . 9
Avisos de seguridad . . . . . . . . . . . . 11
Turvallisuusohjeet . . . . . . . . . . . . . . 13
DO
Säkerhetsanvisningar . . . . . . . . . . . 15
Sikkerhetsinstrukser . . . . . . . . . . . . 17
Veiligheidsvoorschriften. . . . . . . . . 19
Sikkerhedshenvisninger . . . . . . . . . 21
Indicações de segurança . . . . . . . . 23
Sigurnosne upute . . . . . . . . . . . . . . 25
input
Saugumo nurodymai. . . . . . . . . . . . 27
Wskazówki dotyczącebezpieczeństwa 29
Указания по безопасности . . . . . . 31
Bezpečnostní pokyny . . . . . . . . . . . 33
Biztonsági tudnivalók . . . . . . . . . . . 35
Output 2 11 + Output 2
Guard 3
• Safety instructions /
Reference R1
• EC Declarations of Conformity 4 12 Relay 1
electrode
Kurzanleitung
Cal-Info
Kalibrierung
Produktkalibrierung
Verschiebung Nullpkt.
Temp. Abgleich
Cal Info
Calibration
pH
Short instructions
Product calibration
Zero point adjustm.
Temp probe adjustm.
Instructions courtes
Cal-Info
Calibrage
Calibrage du produit
Réglage du point zéro
Réglage sonde temp.
Anode 5 13 Relay 1/2
1100 1100 0000 * Letzter Fehler Last error Dernière erreur
1200 * Konfigurierung Configuration Configuration
2222 * Sensormonitor Sensor monitor Contrôle capteur
5555 * Stromgeber 1 Current source 1 Générateur courant 1
5556 Stromgeber 2 Current source 2 Générateur courant 2
5557 Relaistest Relay test Test des relais
5559 Regler manuell Manual controller Régulateur man.
7654 Parametersatz 1/2 Parameter set 1/2 Jeu de paramètres 1/2
1989 * Paßzahl-Editor * Passcode editor * Editeur code d´accès *
Fehlermeldungen Error messages Messages d‘erreur
Steilheit / Slope / Pente /
R2 14 Relay 2
Asymmetriepotential Asymmetry potential Potentiel asymétrique
RTD E
Ausgangsstrom 1 Output current 1 Courant sortie 1
Err 11: zu klein too low trop bas
Err 12: zu groß too high trop haut
Temp
Err 13: Spanne Span Plage
Ausgangstrom 2 Output current 2 Courant sortie 2
Err 21: zu klein too low trop bas
Err 22: zu groß too high trop haut
Err 23: Spanne Span Plage
Sensocheck Sensocheck Sensocheck
Err 33: Glaselektrode Glass electrode Electrode de verre
Err 34: Bezugselektrode Reference electrode El. de référence
Spülsonde Rinsing probe Sonde rincage
Err 41: Kommunikation Communication Communication
input
Err 42: Cal-Fehler Cal error Erreur cal.
Alarm 15 Alarm
Err 98: Cal / Conf-Daten Cal / Conf data Données Cal / Conf
Meß-Modus / Measuring mode / Mode mesure Err 99: Systemausfall System failure Défaillance système
• Short instructions
RTD D
16 Alarm
shield C
Clean 17 Clean
Devices only:
HOLD 6 Hold 18 Clean
input
HOLD/
CONTROL 7 Power 19 Power
Control
CONTROL 8 input 20 Power
• EC-Type Examination Certificate
144
1
11
15
42 21
10 84 43
2
1 2
3
32 80
9 1 Cable gland (3 pieces)
2 Breakthroughs for cable gland
or conduit 1/2”, dia 21.5 mm
(2 breakthroughs)
72
8 Conduits not included!
3 Breakthroughs for pipe mounting
(4 breakthroughs)
4 Breakthroughs for wall mounting
6,2
3
7 6 5 4 (2 breakthroughs)
165 132
1 1
173
Information on installation
4 3 1
Caution!
• The Stratos may only be installed by trained experts
in accordance with this instruction manual and as
per applicable local and national codes.
• Be sure to observe the technical specifications and input ratings.
• Be sure not to notch the conductor when stripping the
insulation.
• Before connecting the device to the power supply, make sure
that its voltage lies within the range 20.5 to 253 V AC/DC.
• When commissioning, a complete configuration must be
carried out by the system administrator.
The terminals are suitable for single wires and flexible leads up to
2
2.5 mm2 (AWG 14).
Warning!
Additional safety precautions have to be taken for applications in
hazardous locations to CSA (CLI DIV2 GPA,B,C,D T4, Ex nA IIC T4)!
(See Pg 93.) 1 ESD shield covering the sensor inputs
(Screw off for assembly)
Terminal assignments Note: The cable shield must end under the ESD shield.
(Cut lines if required)
2 Terminals for temperature probe
3 Terminals for sensor
4 Connection of power supply
Fig.: Information on installation, rear side of Stratos
Division 2 wiring
The connections to the Stratos are incendive and
must be installed in accordance with the National
Electric Code (ANSI-NFPA 70) Division 2 hazardous
Fig.: Terminal assignments Stratos e 2402 Oxy (classified) location incendive wiring techniques.
transparent
white
red
yellow /
green
red
yellow /
green
black
white
black
green
blue
gray
SE 704 / SE 705 sensor SE 706 / SE 707 sensor
(or InPro6000) (or InPro6800/6900)
Connection Terminal Cable (e.g. ZU 0231) Connection Terminal VP cable (e.g. ZU 0313)
1 not connected 1 not connected
cathode 2 white (coax core) cathode 2 transparent (coax core)
guard 3 not connected guard 3 gray / SE 707 (InPro6900)
ref. el. 4 (jumper 4-5) ref. el. 4 (jumper 4-5), blue
anode 5 red (coax shield) anode 5 red (coax shield)
RTD E black RTD E green
RTD D black RTD D white
shield C yellow / green shield C yellow/green
16 Stratos e 2402 Oxy 17
Protective wiring of switching outputs
Typical AC applications
with inductive load
1
1 Load
1 2 2 RC combination, e.g. RIFA PMR 209
Typical RC combinations
for 230 V AC:
2
3 Capacitor 0.1µF / 630V,
3 Resistor 100 Ohms / 1 W A: DC application with inductive load
3 Contact
B: AC/DC applications with capacitive load
C: Connection of incandescent lamps
A1 Inductive load
A2 Free-wheeling diode, e.g. 1N4007 (Observe polarity)
A3 Contact
B1 Capacitive load
B2 Resistor, e.g. 8 Ohms/1 W at 24 V / 0.3 A
B3 Contact
C1 Incandescent lamp, max 60 W / 230 V, 30 W / 115 V
C3 Contact
Warning!
Make sure that the maximum ratings of the relay
contacts are not exceeded even during switching!
1 2 3 4 5 6 7 8 9 10
20
11
19
18 12
13
17 16 15 14
End End with conf. The measured value and Alarm settings AL.
Hold are displayed alternately, “enter”
flashes. End Hold mode with enter.
Relay / Controller rL. Previous
The display shows the measured value. menu group:
The output current remains frozen for
another 20 sec (HOLD icon on, “hour- Rinsing and
Pb.
glass” flashes). cleaning probes
Menu group Code Display Select menu item o1. Set output current range 4 - 20 mA
Select with key (0 - 20 mA)
Proceed with enter
Output 1 o1.
Select sensor type
Select meas. procedure Current beginning 0000 %
Enter lower end of scale, depending (mg/l,
Select 0-20 / 4-20 mA on the measurement procedure ppm)
selected (Saturation or
Enter current beginning Concentration)
Select with key, edit number with
Enter current end key, proceed with enter
Menu group Code Display Select menu item o1. Time constant of output filter 0 sec
Default setting: 0 sec (inactive). (0 - 120 sec)
To specify a time constant:
Select sensor type
Output 1 o1. Select with key,
edit number with key,
Select meas. procedure proceed with enter.
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end Time constant of output filter
Set output filter To smoothen the current output, a low-pass filter with
adjustable filter time constant can be switched on. When there
22 mA in the case of error is a jump at the input (100 %), the output level is 63 % after
the time constant has been reached.
Hold mode
The time constant can be set from 0 to 120 sec.
If the time constant is set to 0 sec, the current output follows
the input.
Note:
The filter only acts on the current output, not on the display, the
limit values, or the controller!
End:
Press conf, then enter
Menu group Code Display Select menu item o1. 22 mA signal for error message OFF
Select with key (ON)
Proceed with enter
Output 1 o1.
Select sensor type
Select meas. procedure Output signal during HOLD LAST
LAST: (FIX)
Select 0-20 / 4-20 mA During HOLD the last measured value
is maintained at the output
Enter current beginning FIX:
During HOLD a value (to be entered)
Enter current end is maintained at the output
Select with key
Set output filter Proceed with enter
End:
Press conf, then enter 4
HOLD active HOLD active
Menu group Code Display Select menu item o2. Specify temperature unit °C
Select with key (°F)
Proceed with enter
Output 2 o2. Select °C/°F
Select temp probe Select temperature probe 22NTC
Select with key (30NTC)
Select 0-20 / 4-20 mA Proceed with enter
Menu group Code Display Select menu item o2. Time constant of output filter 0 sec
Default setting: 0 s (inactive). (0 - 120 sec)
To specify a time constant:
Output 2 o2.
Select °C/°F Select with key,
edit number with key,
Select temp probe proceed with enter.
End:
Press conf, then enter
Menu group Code Display Select menu item o2. 22 mA signal for error message OFF
Select with key (ON)
Proceed with enter
Output 2 o2.
Select °C/°F
Select temp probe Output signal during HOLD LAST
LAST: (FIX)
During HOLD the last measured
Select 0-20 / 4-20 mA
value is maintained at the output
FIX:
Enter current beginning
During HOLD a value (to be entered)
is maintained at the output
Enter current end Select with key
Proceed with enter
Set output filter
22 mA for temp error Only with FIX selected: 21.0 mA
Enter current which is to flow at the (00.0 to
Hold mode output during HOLD 21.0 mA)
Select with key,
edit number with key,
proceed with enter.
End:
Press conf, then enter 4
HOLD active HOLD active
Menu group Code Display Select menu item Co. Enter polarization voltage 0675 mV
Select with key, edit number with
key, proceed with enter.
Polarization voltage
Correction CO.
End:
Press conf, then enter
Menu group Code Display Select menu item CA. Specify calibration mode SAt
(Calibration to saturation or (Conc)
Calibration mode concentration)
Calibration mode CA. Select with key
Proceed with enter
Cal timer interval
Delay
AL. Select Sensocheck ON / OFF
LED in HOLD mode (continuous monitoring of
sensor)
Select with key
End: Proceed with enter
Press conf, then enter
Alarm contact
Alarm 15 Delay for alarm 0010 s
The alarm contact is closed during normal operation Select with key, edit number with (xxxx s)
16 (N/C). It opens in the case of alarm or power outage. key, proceed with enter.
As a result, a failure message is provided even in the
case of line breakage (fail-safe behavior).
For contact ratings, see Specifications. LED in HOLD mode ON / OFF
Error messages can also be signaled by a 22 mA output Select with key, edit number with
current ( see Pg 37, 43, 78). key, proceed with enter.
Menu group Code Display Select menu item rL. Use of relays: LiMIT
• Limit function (LiMIT) (CtROL)
• Controller (CtROL)
Relay / Controller rL. Use of relays Select with key
Proceed with enter
L1. Contact function
Contact response Note: Selecting CtROL leads to
Controller menu group Ct.
Enter switching point
L1. Limit 1 function , Lo
Enter hysteresis see Fig. on Pg 51. (Hi)
Select with key,
Delay proceed with enter
L2. Relay 2 menu group Limit 1 contact response N/C
N/C: normally closed contact (N/O)
Ct. Controller menu group N/O: normally open contact
Select with key
Proceed with enter
Menu group Code Display Select menu item L2. Select Limit 2, see Fig. below. Hi
Select with key (Lo)
Proceed with enter
Relay / Controller rL. Use of relays
L1. Relay 1 menu group Limit 2 contact response N/C
N/C: normally closed contact (N/O)
L2. Contact function N/O: normally open contact
Select with key
Contact response Proceed with enter
Limit Lo Limit Hi
Signal Signal
Hysteresis +
Switching
Switching point
point
Hysteresis -
1 1
End: Contact Contact
Menu group Code Display Select menu item Ct. Setpoint 0100 %
Select with key, edit number with (xxxx %)
key, proceed with enter.
Relay / Controller rL. Use of relays
L1. Relay 1 menu group Neutral zone (dead band) 0010 %
Select with key, edit number with (xxxx %)
key, proceed with enter.
L2. Relay 2 menu group
Ct. Controller setpoint Controller: P-action component 0100 %
Select with key, edit number with (xxxx %)
Enter neutral zone key, proceed with enter.
o1. Sensor type ____________ ____________ rL. Relay function ____________ ____________
o1. %, mg/l, ppm ____________ ____________
o1. 0/4–20 mA ____________ ____________ L1. Contact function ____________ ____________
o1. Current beginning ____________ ____________ L1. Contact response ____________ ____________
o1. Current end ____________ ____________ L1. Switching point ____________ ____________
o1. Filter time ____________ ____________ L1. Hysteresis ____________ ____________
o1. 22mA signal ____________ ____________ L1. Delay ____________ ____________
o1. Hold behavior ____________ ____________
o1. Fix current ____________ ____________ L2. Contact function ____________ ____________
L2. Contact response ____________ ____________
o2. Unit °C / °F ____________ ____________ L2. Switching point ____________ ____________
o2. Temp probe ____________ ____________ L2. Hysteresis ____________ ____________
o2. 0/4–20mA ____________ ____________ L2. Delay ____________ ____________
o2. Current beginning ____________ ____________
o2. Current end ____________ ____________ Ct. Setpoint ____________ ____________
o2. Filter time ____________ ____________ Ct. Neutral zone ____________ ____________
o2. 22mA signal ____________ ____________ Ct. P action ____________ ____________
o2. Hold behavior ____________ ____________ Ct. I action ____________ ____________
o2. Fix current ____________ ____________ Ct. D action ____________ ____________
Ct. PLC/PFC controller ____________ ____________
Co. Polarization voltage ____________ ____________ Ct. Pulse length ____________ ____________
Co. Pressure unit ____________ ____________ Ct. Pulse frequency ____________ ____________
Co. Pressure ____________ ____________ Ct. Hold behavior ____________ ____________
Co. Salinity ____________ ____________
Pb. Probe selection ____________ ____________
CA. Cal mode ____________ ____________ Pb. Rinsing interval ____________ ____________
CA. Cal interval ____________ ____________ Pb. Rinse duration ____________ ____________
Pb. Contact type ____________ ____________
AL. Sensocheck ____________ ____________
Pb. Cleaning interval ____________ ____________
AL. Alarm delay ____________ ____________
Pb. Lock interval _____________ _____________
AL. LED Hold ____________ ____________
The drift check might take some Information on concentration calibration (Conc)
time. Calibration in air. This calibration method is recommended when the sensor can
be removed for calibration. Air has a stable oxygen content. Therefore the adjust-
ment processes during calibration run more quickly.
• For 2-point calibration, perform zero point calibration first.
Display Action Remark Display of new slope and zero New calibration:
(related to 25°C and 1013 mbars) Press cal key.
Product calibration step 1: The type of product End calibration with enter
Activate calibration calibration (SAT or
(press cal key) Conc) is selected dur-
Enter mode code 1105 ing configuration
(Select with key, (measurement proce-
edit number with key, dure). If an invalid The measured value is shown in After end of calibra-
proceed with enter.) code is entered, the tion, the outputs
the main display alternately with
Stratos returns to remain in Hold mode
“Hold”; “enter” flashes.
measuring mode. for approx. 20 sec.
End with enter.
Measurement
(Error info)
Press conf key while in measuring mode and enter code
0000. The last error message is displayed for
Display Remark approx. 20 sec.
After that the message will be deleted.
In measuring mode the main display shows the configured (immediate return to measurement at pressing enter).
process variable (%, mg/l, or ppm), the secondary display
shows the temperature. During calibration you can return
to measuring mode by pressing the cal key, during configu-
ration by pressing the conf key.
(Waiting time for measured value stabilization is approx.
20 sec).
Controller
[%] O2 %
output Y
KC =100
Setpoint
50 %
KC =50 %
-100
Note:
In Hold mode the controller output acts
Xw 10 / (1.00) 20 / (2.00) 30 / (3.00) 40 / (4.00) 50 / (5.00)
as configured (Y = const. or Y = 0).
Deviation Process variables: % / ( mg/l, ppm)
Red LED
Possible causes
Red LED
Possible causes ing)
ERR 33 Sensocheck x x x
ERR 99 “FAIL” Factory settings x x x x Sensor: Connecting cable defective
flashes EEPROM or RAM defective
This error message only occurs in the • Zero error, Sensoface active, see Pg 82
case of a complete defect. The Stratos
must be repaired and recalibrated at • Slope error, Sensoface active, see Pg 82
the factory.
• Response time exceeded, Sensoface active, s. Pg 82
Problem
Alarm contact
Symbol flashes:
Possible causes Operating
Limit value
Controller
Cleaning
Time out
state
contact
Rel.1/2
Rel.1/2
Out 1
Out 2
Slope out of range
LED
Wrong calibration values specified (relative humidity,
pressure, saturation, concentration)
Product cal 1
Calibration aborted after 12 minutes (cal) 1105
Sensor defective or dirty
Product cal 2
•No electrolyte in the sensor
(cal) 1105
In addition •Sensor cable insufficiently shielded or defective
“CAL Err” •Strong electric fields influence the measurement Configuration 20
is flashing •Temperature fluctuation of calibration solution (conf) 1200
min
Sensor monitor 20
Current source 1 20
Alarm contact
min
(conf) 5555
Operating
Limit value
Controller
Time out
Cleaning
Current source 2 20
state
contact
Rel.1/2
Rel.1/2
min
Out 1
Out 2
(conf) 5556
LED
Relay test 20
Measurement (conf) 5557
min
Manual controller 20
Cal Info 20
(conf) 5559
min
sec
(cal) 0000
Rinsing function
Error Info 20
sec
(conf) 0000
HOLD input
Calibration
(cal) 1100
Explanation: active
Temp adjustment
(cal) 1015 as configured (Last/Fix or Last/Off)
Mounting accessories
Pipe-mount kit ZU 0274
Panel-mount kit ZU 0275
Protective hood ZU 0276
DO input Sensor Type A: SE 704 / 705 / 706 Temperature NTC 22 kOhms / NTC 30 kOhms*
InPro6000–6800 input 2-wire connection, adjustable
Sensor Type B: Inpro6900, SE 707 Measurement range -20.0 to +150.0 °C / -4 to +302 °F
Adjustment range 10 K
Measuring current -2 to 1800 nA, Resolution 0.1 °C / 1 °F
Resolution 0.05 nA Meas. error1,2,3 < 0.5 K (<1 K at >100°C)
(with Vpol = 800 mV and Vref = 200 mV)
Saturation (-10 to 80 °C) 0 to 500 %
Meas. error1,2,3 0.5 % meas.val. + 0.5 % HOLD input Galv. separated (OPTO coupler)
Function Switches Stratos to HOLD mode
Concentration (-10 to 80 °C) 0.00 to 50.00 mg/l Switching voltage Inactive 0 to 2 V (AC/DC)
0.00 to 50.00 ppm Active 10 to 30 V (AC/DC)
Meas. error 1,2,3
0.5 % meas.val. + 0.05 mg/l
or 0.05 ppm CONTROL input Galv. separated (OPTO coupler)
Adm. guard current ≤ 20 µA Function Control input for automatic cleaning system
Polarization voltage * 0 to 1000 mV, Switching voltage Inactive 0 to 2 V (AC/DC)
Process pressure * 0.000 to 9.999 bars Active 10 to 30 V (AC/DC)
(to 999.9 kPa / to 145.0 psi)
Salt correction * 00.00 to 45.00 g/kg Output 1 0/4 to 20 mA, max. 10 V, floating
(galv. connected to output 2)
Sensor standardization Process variable * DO saturation/DO concentration
Operating modes * DO saturation (automatic) Overrange * 22 mA in the case of error messages
DO concentration (automatic) Output filter * Low-pass, filter time constant 0 to 120 s
Product calibration Meas. error 1 < 0.3 % current value + 0.05 mA
Zero point calibration
Start/end of scale Configurable within selected range
Calibration range Zero point ± 2 nA Adm. span 5 to 500 % / 0.5 to 50 mg/l (ppm)
Sensor Type A Slope 25 to 130 nA
(at 25 °C, 1013 mbars) Output 2 0/4 to 20 mA, max. 10 V, floating
Calibration range Zero point ± 2 nA (galv. connected to output 1)
Sensor Type B Slope 200 to 550 nA Process variable Temperature
(at 25 °C, 1013 mbars) Overrange * 22 mA in case of temp error messages
Calibration timer * 0000 to 9999 h Output filter * Low-pass, filter time constant 0 to 120 s
Pressure correction * 0.000 to 9.999 bars / 999.9 kPa / 145.0 psi Meas. error 1 < 0.3 % current value + 0.05 mA
Sensocheck Monitoring for short circuits / open circuits (can be disabled) Start/end of scale* -20 to +150 °C / -4 to +302 °F
Adm. span 20 to 170 K (68 to 338 °F)
Sensoface Provides information on the sensor condition
Evaluation of zero/slope, response,
calibration interval, Sensocheck
Alarm contact Relay contact, floating Cleaning function* Relay contact, floating, for controlling a simple rinsing
Contact ratings AC < 250 V / < 3 A / < 750 VA system or an automatic cleaning system (EasyClean)
DC < 30 V / < 3 A / < 90 W Contact ratings AC < 250 V / < 3 A / < 750 VA
Contact response N/C (fail-safe type) DC < 30 V / < 3 A / < 90 W
Response delay * 0000 to 0600 s Contact response* N/C or N/O
Interval * 000.0 to 999.9 h
Limit values Output via relay contacts R1, R2 (000.0 h = cleaning function switched off)
Contacts R1, R2 floating but inter-connected Rinse duration* 0000 to 1999 s
Contact ratings AC< 250 V / < 3 A / < 750 VA Display LC display, 7-segment with icons
DC< 30 V / < 3 A / < 90 W Main display Character height 17 mm, unit symbols 10 mm
Contact response* N/C or N/O Secondary display Character height 10 mm, unit symbols 7 mm
Response delay * 0000 to 0600 s Sensoface 3 status indicators (friendly, neutral, sad Smiley)
Switching points* Within selected range
Hysteresis* 000.0 to 050.0 % / 00.00 to 05.00 mg/l (ppm) Mode indicators 5 status bars
“meas”, “cal”, “alarm”, “cleaning”, “config”
PID process Output via relay contacts R1, R2 18 further icons for configuration and messages
controller (Relay R1: below setpoint, Alarm indication Red alarm LED in case of alarm or HOLD (user defined)
Relay R2: above setpoint)
Setpoint specification* 0 to 500 % / 0.00 to 50.00 mg/l (ppm) Keypad 5 keys
Neutral zone* 000.0 to 050.0 % / 00.00 to 05.00 mg/l (ppm) Service functions
P-action component* Controller gain Kc: 0010 to 9999 % Current source Current specifiable for output 1 and 2 (00.00 to 22.00 mA)
I-action component* Reset time Tr: 0000 to 9999 s Manual controller Controller output entered directly (start of control process)
(0000 s = no integral action) Device self-test Automatic memory test (RAM, FLASH, EEPROM)
D-action component* Rate time Td: 0000 to 9999 s Display test Display of all segments
(0000 s = no derivative action) Last Error Display of last error occurred
Controller type* Pulse length controller or pulse frequency controller Sensor monitor Display of direct, uncorrected sensor signal
Pulse period* 0001 to 0600 s, min. ON time 0.5 s Relay test Manual control of the four switching contacts
(pulse length controller)
Max. pulse frequency* 0001 to 0180 min-1 Parameter sets Two selectable parameter sets for different
(pulse frequency controller) applications
EMC EN 61326
Emitted interference Class B (residential environment)
Class A for mains supply > 60 V DC
Immunity to
interference Industrial environment
Explosion protection
FM: NI Class I Div 2 Group A, B, C & D, T4 Ta = 55 °C; Type 2
NI Class I Zone 2 Group IIC, T4 Ta = 55°C; Type 2
Configuration
1200 Configuration
Manual controller
5559 Manual specification of controller output