Vascular Access The Introcan Safety® and Vasofix® Safety are equipped with IV Therapy
a self-activating clip technology that automatically protects
you by simply using them. It´s that simple!
e
This Pocket Guid
Safety features
the
does not replace
Self-activating clip technology – no extra action needed
r Use!
Safety system covers needle tip immediately after use
Instructions fo
B. Braun Melsungen AG | 34209 Melsungen | Germany | Tel. +49 5661 71-0 | www.bbraun.com
IV Therapy
Yellow LED: Pre-alarm, Press to reset single values to zero Press to switch on the pump and wait
reminder alarm and switch back to the previous
until self test is finished.
screen/menu level.
Green/Red LED: Infusion running/
Infusomat® Space
device alarm, operating alarm Press to ini- Press to open the door and confirm with
tiate bolus. “Yes” . Insert the infusion line from right
Blue LED:
Currently connected to Press to turn (container) to the left (patient). Insert the two-
SpaceControl pump on/off. hole clip of the IV Set in the line guide fixa-
tion which is only possible in one direction.
Opening Down-
lever stream
pressure
Line Single Two Line Push the line into the up-stream pressure
sensor
outlet clip pin inlet sensor 6 , the down-stream pressure sensor
Air line 7 and the air sensor 8 .
sensor fixation Make sure the line has good contact to the
surface of the sensor.
Yellow LED: Pre-alarm, Press to reset single values to zero Press to switch on the pump. Open the
reminder alarm and switch back to the previous
door. Pull and turn the syringe-holder to
screen/menu level.
Green/Red LED: Infusion running/ the right to open the green axial fixation .
device alarm, operating alarm Press to ini-
tiate bolus.
Blue LED:
Currently connected to Press to turn
SpaceControl pump on/off.
Insert a syringe in an 45° angle of the wings
and turn it up.
Perfusor® Space
Syringe must be fixed with wings upright in
Arrow up and -down: Press to Start/
Scroll through menus, change setting of numbers the slot between the green axial fixation
Stop infusion.
from 0-9, answer Yes/No questions. and the housing of the pump.
Arrow left and -right:
Select data from a scale and switch Press to link the pump to SpaceControl If the display shows the message “Invalid
between digits when numbers syringe” the wings of the syringe are not
are entered. Open a function while Open certain functions and press
pump is running or stopped. to confirm values/settings/alarms. properly fixed by the green axial fixation .
Keep the syringe fixed with your left hand
and close the syringe holder with the right
Green axial hand. The piston brake prevents uninten-
fixation ded boluses.
Optionally a warning hint may appear. Edit the parameter or the sub menu with or .
Confirm this message with or .
It’s also possible to configure a drug With the user can fade in diffe-
specific maximum dosage, that a user rent information e.g. remaining time
can not exceed. until syringe is empty, in the bottom line
of the display.
Navigation
Here many parameters e.g. the sensibility
of the pressure sensor or the bolus rate
are adjustable. Bolus with volume pre-selection
Press . Then press and set bolus
Status Menu dose using the arrow buttons.
In this menu all therapy status parame- Press to confirm and start bolus. An
ters are visible. E.g. the continuous coun- acoustic signal will sound after finishing
ting of the infused volume or amount of the bolus volume.
drug can be seen.
Standby Mode
For volume balancing use to reset Hold the button pressed for 3
the counter. Press “Yes” to reset the seconds to activate the standby mode.
counter or “No” .
Start the PCA therapy with . Depen- The signal tone and the staff call are
ding on the settings in the drug library turned off with . Display and LED stay
the pump might lock the dosage settings in pre-alarm until the operating alarm
and the syringe/drug full automatically. goes off . Pre-alarms don't lead to an
interruption of the infusion.
If the “pain service screen” appears the
patient can choose the external PCA but-
ton, or use at the pump to deliver a Pressure Alarm (downstream to patient)
preconfigured PCA bolus. The volume of The sensibility of the pressure sensor
the confirmation beep can be adjusted can be adjusted in the options menu.
via the option menue or with .
PCA
Above the 12hrs time axis you see the Check the roller clamp or the fluid
administered PCA boluses, below the level of the burette and try to restart
denied PCA boluses. Changes of one or the delivery with .
more of the treatment parameters are If that is not possible please follow
indicated by an ^ at the bottom line of the dialog on the screen.
the screen.
IV Therapy
Port P3
for future Port P2 for Power Supply, SpaceStation, 12V Lead, Combi Lead
options and further accessory leads (staff call, service)
Operation alarms
Port P3
Port for for future Port P2 for Power Supply, SpaceStation, 12V Lead,
dropsensor options Combi Lead and further accessory leads (staff call, service)
IV Therapy
Length Latex PVC Sales unit Code No. B. Braun is a global supplier on the corpo- Efficiency
Infusomat® Space Line
(cm) free free Pcs. (REF) rate healthcare market with a workforce
Today's healthcare systems increasingly
single packed 250 ● - 100 8700036SP of more than 37,000 employees in over 50
demand that the relationship between cost
in 10 unit ward pack 250 ● - 100 8700435SP countries all over the world.
and benefit be a sensible one. It is in every-
single packed with PrimeStop and AirStop 250 ● - 100 8701148SP
"Sharing Expertise" expresses the spirit of body's self-interest to keep the social securi-
single packed, 3 m 300 ● - 100 8270350SP
B. Braun's corporate culture best: It is the ty systems functioning by raising the level
with Eurofix injection port 270 ● - 100 8751668SP
promise to share medical knowledge and of efficiency at corporations, hospitals and
with orange injection port 270 ● - 100 8700087SP
expertise with customers and colleagues alike. in outpatient sector. Almost two centuries of
Neutrapur® 250 ● ● 100 8250731SP
experience enable B. Braun to provide smart
Neutrapur® – with Safeflow needle-free injection port 270 ● ● 100 8700110SP B. Braun is committed to its close partner-
strategies, products, and services for more
Neutrapur® – with inline 2.0 μm Sterifix® Filter 250 ● ● 20 8700095SP ships with physicians, nurses, hospital man-
efficient operational processes.
agers and researchers. The aim is to benefit
Length Latex PVC Sales unit Code No.
Lines for advanced IV Therapy the health of everybody, to use the knowl- Sustainability
(cm) free free Pcs. (REF)
Opaque – light protected, black tubing 250 ● - 100 8700125SP edge gained effectively, to expand it logically,
As a corporate citizen in a global society,
Opaque – light protected, with orange injection port 250 ● - 100 8250430SP consistently and with determination –
B. Braun feels responsible for its employees
Transfusion – with 200 μm blood filter 250 ● - 100 8270066SP always in dialogue with its partners.
and for the geographical regions in which it
Piggyback – with needle-based port + 200 μm filter 250 - - 25 8250715SP As one of the leading system providers, is active. This responsibility obviously encom-
Piggyback – secondary line, needle-based 40 ● - 25 A2581F B. Braun develops services and pipeline passes environmental protection, but also the
Piggyback – with Safeflow injection port 290 ● ● 100 8250710SP products that provide optimum therapeutic commitment to create perspectives for future
Piggyback – secondary line, needle-free 40 ● ● 20 4062877T benefits for patients in both clinical and generations by promoting cultural and sport-
Dosifix® – 150 ml burette, with Safeflow injection port 270 - ● 40 8250245SP outpatient settings. ing events. In these times of constant change,
Length Latex PVC Sales unit Code No. sustainability means offering dependability
Lines for Enteral Nutrition The expertise of B. Braun is based on three
(cm) free free Pcs. (REF) and consistency not only to customers, but
supporting pillars: innovation, efficiency
with 1000 ml Enterofix® bag 230 ● - 25 8250839SP also to world's other less fortunate.
and sustainability. These corporate values
with multi-connector for bottles and bags 320 ● ● 40 8250857SP help B. Braun deliver on its promise of
”Sharing Expertise“.
Innovation
B. Braun has a long tradition of developing
innovative products and services to simplify
doctors’ and nurses' workload and to improve
therapies for patients – a key factor in our
success. The ongoing dialog with our cus-
tomers combined with the creative drive
of our human resources feed the constant
flow of new ideas systematically incorporated
into every product and new advancement.
Disposables