Professional Documents
Culture Documents
Manuel Service Selectra Prom
Manuel Service Selectra Prom
SELECTRA PRO
M
6003-400-450-02
Copyright © 2011 Vital Scientific B.V.
All rights reserved. No part of this document may be reproduced, stored in a retrieval system,
or transmitted, in any form or by any means (electronic, mechanical, or otherwise) without the
prior written permission of Vital Scientific B.V.
Vital Scientific B.V. reserves the right to modify components of the described instrument at
any time, without prior notifications to customers. The contents of this documentation may
also be changed without notification.
This document describes the Selectra ProM with all optional components. Vital Scientific B.V.
cannot be held responsible for any damage resulting from changes made to the Selectra
ProM after it was supplied. Vital Scientific B.V. cannot be held responsible for any damage
resulting from not complying to the specifications supplied with the Selectra ProM.
Selectra and ProM are trademarks of Vital Scientific B.V. All other trademarks are trademarks
or registered trademarks of their respective owners.
Vital Scientific B.V. is an ELITech Group company.
SELECTRA PRO
6003-400-450-02
M
Preface
Declaration of conformity
This instrument (excluding the computer) conforms to the provisions of the
EU Directive on In Vitro Diagnostic Medical Devices (98/79/EC) of the
European Parliament and the Council of 27 October 1998. The declaration
of conformity is supplied with each device in a separate file.
! CAUTION
TAKE CARE
This manual was written and produced with the utmost care. However,
errors can never be fully excluded. Vital Scientific B.V. does not take any
responsibilities and accepts no liabilities for incidents of any kind that may
occur because of errors in the manual. When you have doubts about
information given in this manual, please contact your supplier.
! CAUTION
TAKE CARE
Read this manual carefully before you use the analyzer. Observe the safety
guidelines listed in the "Safety" chapter. Observe safety procedures that
may be defined in your laboratory. When you are in doubt about any
information in this manual, consult your superior or contact your supplier.
Info
i This manual describes the Selectra ProM with software version 4.3.x.
Info
i This manual describes the Selectra ProM with all optional components. If
components are not included in your analyzer, this does not affect the
behavior of other components.
Manufacturer: Distributor:
Selectra ProM Reagents, consumables
Vital Scientific B.V. Seppim S.A.S.
PO box 100 Zone Industrielle
NL 6950 AC Dieren 61500 Sees
The Netherlands France
Tel.: +31 313 430 500 Tel.: +33 2 33 81 21 00
Fax: +31 313 427 807 Fax: +33 2 33 48 77 51
info.vital@elitechgroup.com info@elitechgroup.com
www.elitechgroup.com www.elitechgroup.com
SELECTRA PRO
M 6003-400-450-02
SELECTRA PRO
6003-400-450-02
M
Table of Contents
1 Introduction
1.1 About this manual ................................................................................1-2
1.1.1 Scope of this manual ........................................................................1-2
1.1.2 Symbols used in this manual............................................................1-3
1.1.3 Typographic conventions..................................................................1-4
1.2 About the instrument............................................................................1-5
1.2.1 The Selectra ProM............................................................................1-5
1.2.2 Intended use.....................................................................................1-6
1.2.3 Technical support .............................................................................1-7
1.2.4 Identification .....................................................................................1-8
1.2.5 Liability..............................................................................................1-9
2 Safety
2.1 General safety instructions ..................................................................2-2
2.1.1 General safety guidelines .................................................................2-2
2.1.2 User qualifications ............................................................................2-4
2.2 Safety symbols ....................................................................................2-5
2.3 Sources of danger ...............................................................................2-7
2.4 Waste disposal ....................................................................................2-8
3 System description
3.1 Overview..............................................................................................3-2
3.1.1 Subsystems identification .................................................................3-2
3.2 Rotors section......................................................................................3-3
3.2.1 Reagent rotor....................................................................................3-3
3.2.2 Cooling unit.......................................................................................3-4
3.2.3 Sample rotor .....................................................................................3-5
3.2.4 Rotor positioning...............................................................................3-6
3.3 Pipetting section ..................................................................................3-7
3.3.1 Pipettor .............................................................................................3-7
3.3.2 Pipetting arm ..................................................................................3-10
3.3.3 Stirrer unit .......................................................................................3-11
3.3.4 ISE needle ......................................................................................3-12
3.4 Measuring section..............................................................................3-13
3.4.1 Optical system ................................................................................3-13
3.4.2 Lamp unit........................................................................................3-14
3.4.3 Filter wheel .....................................................................................3-15
3.4.4 Cuvette rotor...................................................................................3-16
3.4.5 Washing unit...................................................................................3-17
3.4.6 Vertical positioning of the wash arm...............................................3-18
3.4.7 ISE unit ...........................................................................................3-19
4 Installation
4.1 Preparing for installation ......................................................................4-2
4.1.1 Transport and storage ......................................................................4-2
4.1.2 Preparing the installation site ...........................................................4-3
4.1.3 Unpacking.........................................................................................4-4
4.2 Installation............................................................................................4-7
4.2.1 Mounting the panel PC .....................................................................4-7
4.2.2 Installing water and waste containers.............................................4-12
4.2.3 Installing other parts .......................................................................4-14
4.2.4 Installing the application software...................................................4-15
4.2.5 Finishing the installation .................................................................4-16
4.2.6 Parameters for the performance test..............................................4-17
4.3 Taking the analyzer out of use...........................................................4-18
4.3.1 Switching the analyzer off for a long time.......................................4-18
4.3.2 Preparing for transport or storage ..................................................4-19
SELECTRA PRO
ii 6003-400-450-02
M
4.4 Upgrading and restoring ....................................................................4-20
4.4.1 Preparations ...................................................................................4-20
4.4.2 Restoring the operating system......................................................4-21
4.4.3 Restoring the application software .................................................4-22
5 Software
5.1 Overview..............................................................................................5-2
5.1.1 Service menus..................................................................................5-2
5.2 Status and history screens ..................................................................5-4
5.2.1 Needle Rinse History screen ............................................................5-4
5.2.2 Error History screen..........................................................................5-5
5.2.3 Changes Log screen ........................................................................5-6
5.2.4 System Configuration screen ...........................................................5-7
5.3 Functional checks and adjustments.....................................................5-8
5.3.1 Adjust Lamp screen..........................................................................5-8
5.3.2 Filter Check screen...........................................................................5-9
5.3.3 Reagent Arm and Sample Arm screens .........................................5-10
5.3.4 Reagent Disk screen ......................................................................5-12
5.3.5 Sample Disk / Barcode screen .......................................................5-13
5.3.6 Measurement Disc/Filter screen.....................................................5-14
5.3.7 Wash Arm screen ...........................................................................5-15
5.3.8 ISE Arm/ISE Unit screen ................................................................5-16
5.3.9 Pipettor screen ...............................................................................5-17
5.3.10 Vacuum System screen..................................................................5-18
5.3.11 Water System screen .....................................................................5-19
5.3.12 Optical electronics screen ..............................................................5-20
5.3.13 Electronics screens ........................................................................5-21
6 Maintenance
6.1 Introduction ..........................................................................................6-2
6.1.1 Preventive maintenance ...................................................................6-2
6.1.2 Maintenance equipment ...................................................................6-3
6.1.3 Replacement parts ...........................................................................6-4
6.2 General procedures .............................................................................6-5
6.2.1 Cleaning the system .........................................................................6-5
6.2.2 Running the dichromate end test......................................................6-7
6.2.3 Filling a cooling unit ..........................................................................6-9
6.2.4 Removing a measuring unit............................................................6-10
6.2.5 Removing a water/vacuum pumps section.....................................6-12
6.2.6 Removing a bellows pump .............................................................6-13
6.3 Rotors section....................................................................................6-14
6.3.1 Adjusting a reagent or sample rotor ...............................................6-14
6.3.2 Adjusting an encoder opto..............................................................6-16
6.3.3 Checking reagent cooling ...............................................................6-17
6.3.4 Checking a PSID scanner ..............................................................6-18
6.3.5 Adjusting a PSID scanner...............................................................6-20
6.4 Pipetting section ................................................................................6-23
6.4.1 Replacing a syringe ........................................................................6-23
6.4.2 Replacing a syringe plunger ...........................................................6-25
SELECTRA PRO
M 6003-400-450-02 iii
6.4.3 Replacing a pipettor valve ..............................................................6-26
6.4.4 Removing a pipettor frontplate .......................................................6-27
6.4.5 Servicing a pipettor spindle ............................................................6-28
6.4.6 Adjusting a pipetting arm (horizontal) .............................................6-30
6.4.7 Adjusting a pipetting arm (vertical) .................................................6-31
6.4.8 Checking / adjusting a needle switch .............................................6-33
6.4.9 Replacing a stirrer belt....................................................................6-34
6.4.10 Replacing a stirrer bearing .............................................................6-35
6.4.11 Adjusting a stirrer............................................................................6-36
6.5 Measuring section..............................................................................6-37
6.5.1 Adjusting a lamp unit ......................................................................6-37
6.5.2 Replacing a lamp............................................................................6-39
6.5.3 Adjusting a cuvette photo sensor ...................................................6-41
6.5.4 Adjusting a cuvette rotor.................................................................6-42
6.5.5 Adjusting a wash arm .....................................................................6-44
6.5.6 Replacing a drying block ................................................................6-46
6.5.7 Checking cuvette detection ............................................................6-47
6.5.8 Replacing Peltier elements.............................................................6-48
6.6 Fluid and vacuum system ..................................................................6-55
6.6.1 Replacing a water filter ...................................................................6-55
6.6.2 Adjusting a vacuum switch .............................................................6-56
6.6.3 Overhauling a vacuum pump..........................................................6-58
6.6.4 Overhauling a cooling or water pump.............................................6-59
6.6.5 Adjusting a bellows pump...............................................................6-61
6.7 Control system...................................................................................6-62
6.7.1 Checking the power supply ............................................................6-62
6.7.2 Checking the photometer board .....................................................6-63
6.7.3 Checking / adjusting the temperature control board.......................6-65
6.7.4 Temperature control board states and parameters ........................6-67
6.7.5 Adjusting the sample level detection ..............................................6-69
6.7.6 Adjusting the reagent level detection..............................................6-70
7 Troubleshooting
7.1 Error messages ...................................................................................7-2
7.1.1 Error reporting ..................................................................................7-2
7.1.2 System board and general errors .....................................................7-3
7.1.3 Pipettor errors...................................................................................7-6
7.1.4 Measuring unit errors........................................................................7-8
7.1.5 Sample and reagent arm errors......................................................7-10
7.1.6 Sample and reagent rotor errors ....................................................7-12
7.1.7 Washing system errors...................................................................7-14
7.1.8 ISE unit errors.................................................................................7-16
7.1.9 Temperature control board errors...................................................7-18
7.1.10 Cooling unit board errors ................................................................7-21
7.2 Troubleshooting .................................................................................7-24
7.2.1 Personal safety...............................................................................7-24
7.2.2 System initialization sequence .......................................................7-25
8 Parts
8.1 Overview..............................................................................................8-2
SELECTRA PRO
iv 6003-400-450-02
M
8.2 Rotors section......................................................................................8-4
8.2.1 Reagent and sample rotors ..............................................................8-4
8.2.2 Rotor drive ........................................................................................8-6
8.2.3 Cooling unit.......................................................................................8-8
8.2.4 Stirrer unit .......................................................................................8-10
8.2.5 Wash arm .......................................................................................8-12
8.2.6 Wash arm drive ..............................................................................8-14
8.3 Measuring unit ...................................................................................8-16
8.3.1 Measuring unit - lamp section.........................................................8-16
8.3.2 Measuring unit - filter wheel section ...............................................8-18
8.3.3 Measuring unit - drive section.........................................................8-20
8.3.4 Measuring unit - rotor section .........................................................8-22
8.4 Pipettor ..............................................................................................8-24
8.4.1 Pipettor (front).................................................................................8-24
8.4.2 Pipettor (rear) .................................................................................8-26
8.4.3 Pipettor (valve drive).......................................................................8-28
8.5 Reagent and sample arms.................................................................8-30
8.5.1 Pipetting arms (main) .....................................................................8-30
8.5.2 Vertical drive...................................................................................8-32
8.5.3 Horizontal drive...............................................................................8-34
8.5.4 Reagent head .................................................................................8-36
8.5.5 Sample head ..................................................................................8-38
8.6 Water and waste system ...................................................................8-40
8.6.1 Pumps and valves (main) ...............................................................8-40
8.6.2 Pumps and valves (details) ............................................................8-42
8.6.3 Bellows pump (mechanical parts)...................................................8-44
8.6.4 Bellows pump (valves and tubing)..................................................8-46
8.6.5 Water cask......................................................................................8-48
8.6.6 Collecting cask ...............................................................................8-50
8.6.7 Collecting cask (ECO-kit version)...................................................8-52
8.6.8 Water and diluted waste containers ...............................................8-54
8.7 Other parts.........................................................................................8-56
8.7.1 Panels.............................................................................................8-56
8.7.2 Power and electronics section........................................................8-58
8.7.3 ISE needle ......................................................................................8-60
8.7.4 ISE unit ...........................................................................................8-62
8.7.5 PSID scanner .................................................................................8-64
9 Electronics
9.1 Introduction ..........................................................................................9-2
9.1.1 Servicing of electronic boards ..........................................................9-2
9.1.2 Location of main control boards .......................................................9-3
9.1.3 Locations of other boards .................................................................9-4
9.2 Wiring diagrams...................................................................................9-7
9.3 Board layouts and diagrams ..............................................................9-14
9.3.1 6002-773: System board ................................................................9-14
9.3.2 6003-375: Motion control board......................................................9-21
9.3.3 6003-068: Photometer board..........................................................9-26
9.3.4 6003-220: Temperature control board............................................9-30
SELECTRA PRO
M 6003-400-450-02 v
9.3.5 6003-007: Cooling unit board .........................................................9-36
9.3.6 6001-769: Dispenser driver board ..................................................9-42
9.3.7 3367-103: Ambient board ...............................................................9-44
9.3.8 6001-763: LLD board......................................................................9-45
9.3.9 6001-768: Valve board ...................................................................9-47
9.3.10 6001-773: Stirrer driver / ISE needle board....................................9-49
A Technical Specifications
A.1 Performance and technical data ......................................................... A-2
A.1.1 Selectra ProM.................................................................................. A-2
A.1.2 User interface .................................................................................. A-4
A.1.3 Physical data ................................................................................... A-6
A.1.4 Cleaning solutions ........................................................................... A-7
A.1.5 Accuracy and precision ................................................................... A-9
A.2 Fluid and vacuum system ................................................................. A-10
A.2.1 Fluid diagram................................................................................. A-10
SELECTRA PRO
vi 6003-400-450-02
M
1
Introduction
This chapter contains general information about the documentation. It also contains
information about available technical assistance and service training.
SELECTRA PRO
M 6003-400-450-02 1-1
Introduction
This service manual is the prime source of information for service engineers for the Selectra
ProM analyzer. The manual contains information required for preventive maintenance,
periodic replacement of spare parts and servicing of the analyzer.
This service manual does NOT include all information required to operate the Selectra ProM.
That information can be found in the user manual, which should always be available in the
immediate vicinity of the instrument.
Info
i Service engineers should have received proper training on the Selectra
ProM from Vital Scientific or its distributors.
Info
i If you need information that is not given in this manual, or if you are in doubt
about the information given in this manual, contact the system support
department of Vital Scientific. See the address information on the Preface
page.
SELECTRA PRO
1-2 6003-400-450-02
M
Introduction
The safety notes and warnings in this manual are defined according to the ANSI Z535.6
standard for hazard statements. This standard identifies four categories of hazard statements,
with defined color schemes and captions. The icon shown left of the text and the heading text
vary with the type of hazard statement.
NOTICE
PLEASE NOTE
i This paragraph format indicates situations that may lead to damage to the
equipment or the environment if precautions are not taken.
The symbol and header may indicate a particular type of hazard.
! CAUTION
TAKE CARE
This paragraph format indicates situations that may lead to minor or
moderate injuries if no precautions are taken.
The symbol and header may indicate a particular type of hazard.
! WARNING
POTENTIAL HAZARD
This paragraph format indicates situations that may lead to serious injuries
if precautions are not taken.
The symbol and header may indicate a particular type of hazard.
! DANGER
HIGH RISK
This paragraph format indicates situations that will lead to serious injuries or
death if precautions are not taken.
The symbol and header may indicate a particular type of hazard.
Notes that do not identify hazards but merely point out useful information are styled with an
info symbol and no colored heading:
Info
i This paragraph format indicates information that may be useful. They may
concern efficiency or point to information to be found elsewhere, either in this
manual or in other documents.
SELECTRA PRO
M 6003-400-450-02 1-3
Introduction
When names of items on the screen are mentioned in this documentation, they are shown as
Control Name (in bold italic text).
Info
i The analyzer software offers a choice of languages for the user interface.
Some screen items are not translated and will always appear in English.
Info
i The documentation uses the same names that appear on the screens, in the
language in which the documentation is written or translated. Because of
changes to the user interface software after this documentation was
produced, differences may occur.
SELECTRA PRO
1-4 6003-400-450-02
M
Introduction
SELECTRA PRO
M 6003-400-450-02 1-5
Introduction
Disclaimers
Depending on the specific characteristics of the used reagent kit, the results obtained from
a clinical chemistry system may vary. The test parameters for each test and each reagent
supplier need to be developed and validated by appropriate methods (e.g. using ECCLS
or CLSI guidelines) before the system is used for actual measurements of samples.
Info
i ECCLS = European Committee on Clinical Laboratory Standards.
CLSI = Clinical and Laboratory Standards Institute.
Vital Scientific recommends the use of ELITech reagents, calibrators and controls for the
analyzer.
SELECTRA PRO
1-6 6003-400-450-02
M
Introduction
Sources of information
Apart from this service manual, the following sources of information are or should be
available to you:
• User Manual. This should always be available near the analyzer.
• Service Training. See below for details.
• Service Notes. These are issued by Vital Scientific on a regular basis. If you want to
check on the availability of Service Notes, contact the support department of Vital
Scientific or your supplier.
• Vital Scientific website. See the contact data in the Preface.
• System support department of Vital Scientific.
Service training
Vital Scientific organizes service training on a regular basis. These training courses are
either organized at the Vital Scientific headquarters, on customer sites or in the offices of a
supplier of the analyzers. Check for service training opportunities on the Vital Scientific
website or contact the system support department of Vital Scientific for details.
Participants are expected to have knowledge of:
• analog / digital electronics
• microprocessor techniques
• basic photometric techniques
• the English language
Info
i The type and serial number are shown on the identification plate in the back
of the analyzer. The software version is shown in the Release screen. See
par. 1.2.4.
SELECTRA PRO
M 6003-400-450-02 1-7
Introduction
1.2.4 Identification
Analyzer hardware
An identification plate is attached to the back of the analyzer. The identification plate
shows the analyzer type and serial number (SN).
PROM SERIES
SN:
IVD
Analyzer software
The analyzer software version is shown in the Release screen. This screen is available in
the Installation menu.
Info
i The detailed information about the versions of the PROM software in various
control boards of the analyzer is shown after touching the right side of the
screen.
SELECTRA PRO
1-8 6003-400-450-02
M
Introduction
1.2.5 Liability
Conditions of change
Changes to the analyzer, other than described in this manual or in other documentation
provided by Vital Scientific, may only be conducted by qualified personnel after explicit
written permission by Vital Scientific.
Repair and/or maintenance must be performed using parts that are validated and/or
recommended by Vital Scientific.
Vital Scientific cannot be held responsible for any consequential damage, loss or costs
that result from non-conformance to the above rules.
SELECTRA PRO
M 6003-400-450-02 1-9
Introduction
SELECTRA PRO
1-10 6003-400-450-02
M
2
Safety
This chapter contains important safety information. You must read this chapter before
starting any work on the analyzer.
SELECTRA PRO
M 6003-400-450-02 2-1
Safety
The Selectra ProM is designed and manufactured according to state-of-the-art standards and
conforms to safety regulations. All possible risks that were known at the time of
manufacturing were taken into account and either eliminated or reduced. Nevertheless, some
sources of danger cannot be fully eliminated. Please note and follow the guidelines listed in
this chapter carefully. Safety regulations also include requirements on the part of the users of
the instrument: they need to make themselves acquainted with safety procedures and
regulations before starting to work with the instrument. It is important that all personnel who
will work with the instrument reads this chapter and complies to the safety regulations.
! CAUTION
TAKE CARE
The user should be aware that, if the analyzer is used in a manner not
specified by Vital Scientific, the protection provided by the safety provisions
may be impaired.
! CAUTION
TAKE CARE
The guidelines listed below do not replace any safety policy that may be
defined in your company, nor do they replace national or international
regulations regarding safety. Any such safety guidelines and rules remain in
force. In case of conflicting guidelines or doubts about the guidelines given
here, consult your superior.
SELECTRA PRO
2-2 6003-400-450-02
M
Safety
• Employees must avoid wearing rings, wrist watches, jewelry, loose or torn clothing
when operating the instrument. These objects can get caught on moving parts of the
instrument and cause dangerous situations.
• Employees must immediately report to their superior any malfunctions of the
instrument or faults in the tools used for the instrument.
• Employees are not allowed to remove safety, alarm or control elements from the
instrument, without having permission from their superior.
• Employees are not allowed to act on their own initiative, when this may cause
dangerous situations for themselves or others.
• Employees must point out dangerous situations or behaviour to others.
SELECTRA PRO
M 6003-400-450-02 2-3
Safety
Operators
The analyzer should only be used by qualified and trained personnel, who have received a
special operator training for the instrument. For clinical tests, the analyzer should be used
under supervision of a doctor or clinical inspector.
Service engineers
Installation, maintenance and repair should only be done by service engineers, who have
received a special service training for the instrument from the manufacturer or their
representative. A service engineer must also be familiar with the normal operating
procedures as described in the user manual, as well as with the procedures described in
the service manual.
SELECTRA PRO
2-4 6003-400-450-02
M
Safety
The following safety symbols may be used on the analyzer and/or in the manual.
NOTICE
NOTICE
RECYCLE
This notice indicates that the analyzer should not be discarded with normal
household waste. A label with this symbol, in accordance with European
Directive 2002/96/EC, is attached to the back of the analyzer. At the end of
its lifetime, the analyzer must be collected separately and disposed of
according to locally valid regulations.
! CAUTION
TAKE CARE
This caution gives important information that helps preventing personal
injuries and/or serious damage to the analyzer and/or the environment.
! CAUTION
WEAR GLOVES
This caution indicates that gloves must be worn when performing the
described work.
! CAUTION
! WARNING
BIOHAZARD
This caution indicates potentially infectious materials. Safety measures must
be taken according to locally valid regulations when performing the
described work.
SELECTRA PRO
M 6003-400-450-02 2-5
Safety
! WARNING
HOT SURFACE
This warning indicates a hot surface. To prevent fingers or other body parts
from getting burned, the precautions listed in the warning must be taken
when performing the described work.
! WARNING
PINCH POINT
This warning indicates a pinch point. To prevent fingers or other body parts
from getting crushed, the precautions listed in the warning must be taken
when performing the described work.
! WARNING
SHARP EDGES
This warning indicates a sharp edge or point. To prevent fingers or other
body parts from being cut, the precautions listed in the warning must be
taken when performing the described work.
! DANGER
HIGH VOLTAGE
This danger notice indicates a high voltage terminal. The precautions listed
in this danger notice must be taken.
SELECTRA PRO
2-6 6003-400-450-02
M
Safety
This section indicates general sources of danger in the analyzer. The warnings listed here
should always be kept in mind when servicing the instrument.
Electrical hazards
! DANGER
HIGH VOLTAGE
Servicing the power supply, the computer or the main fuses should never be
done while the power cord is plugged in.
! CAUTION
TAKE CARE
Servicing electronic parts may require the power to be switched on. Make
sure you are not causing short circuits. These may lead to irrepairable
damage to the electronic components.
! CAUTION
TAKE CARE
Take the necessary precautions to avoid static electricity discharges. These
may lead to irrepairable damage to the electronic components.
Mechanical hazards
! WARNING
PINCH POINT
Servicing mechanical parts may require activating drives while covers are
open or removed. Do not touch components while their drives are active.
! WARNING
HOT SURFACE
Several components (e.g. lamp, cooling unit components, heating elements)
may be hot while the instrument is switched on. Take care when servicing
these parts or parts in the immediate vicinity of hot surfaces.
Biological hazards
! WARNING
BIOHAZARD
Samples are potentially infectious. Take all required precautions when
handling any sample or waste materials.
SELECTRA PRO
M 6003-400-450-02 2-7
Safety
This section lists notices and cautions concerning the correct disposal of waste materials.
These should always be kept in mind when installing, maintaining, servicing and recycling the
instrument.
Installation
NOTICE
PLEASE NOTE
i Various countries have regulations on the disposal of packaging waste.
Consult local sources for additional information on correct disposal of waste
materials.
Consumables
! WARNING
BIOHAZARD
Fluid and solid consumables are potentially infectious. Take all required
precautions when discarding these materials.
End of life
NOTICE
RECYCLE
No parts of the analyzer should be discarded as normal household waste. At
the end of its lifetime, the analyzer must be collected separately and
disposed of according to locally valid regulations. Contact the System
Support department of Vital Scientific for reuse and treatment information.
SELECTRA PRO
2-8 6003-400-450-02
M
3
System description
This chapter contains functional descriptions of the main components in the analyzer.
SELECTRA PRO
M 6003-400-450-02 3-1
System description
3.1 Overview
Nr Description See:
SELECTRA PRO
3-2 6003-400-450-02
M
System description
Info
i Multiple reagent rotors can be used on the same analyzer. The rotor can be
lifted off the axle after screwing off the center screw [7]. When a rotor is
exchanged, the rotor configuration must also be changed in the software.
SELECTRA PRO
M 6003-400-450-02 3-3
System description
Primary Primary
cooler cooler
Level tube
Cooling pump
Seconday cooler
SELECTRA PRO
3-4 6003-400-450-02
M
System description
8
9
3
4 10
5 11
1 Sample rotor cover 5 Stepper motor 9 Rotor positioning pin
2 Sample rotor 6 Screw 10 Vane
3 Home sensor 7 Sample rotor (PSID) 11 Encoder disc
4 Encoder opto board 8 Driving axle
Info
i If the Selectra ProM is equipped with the optional barcode scanner, the PSID
sample rotor [7] replaces the standard sample rotor [2].
SELECTRA PRO
M 6003-400-450-02 3-5
System description
Info
i The rotation of the pipetting needle is also controlled in this way. The vertical
displacement of the pipetting needle is done in a similar way, except for the
home sensor. See par. 3.3.2.
Fig.3-7 Two encoder signals monitor the exact encoder disc position
SELECTRA PRO
3-6 6003-400-450-02
M
System description
3.3.1 Pipettor
SELECTRA PRO
M 6003-400-450-02 3-7
System description
Pipetting water
Info
i In reality, the tubing on the water and needle side will never be empty. The
diagram is simplified to explain the function of the pipettor.
SELECTRA PRO
3-8 6003-400-450-02
M
System description
SELECTRA PRO
M 6003-400-450-02 3-9
System description
Info
i Analyzers may have separate pipetting arms for samples and reagents or
one pipetting arm that combines both functions. The number of arms on an
instrument depends on the analyzer type. This functional description is valid
for all types.
The pipetting arm aspirates a defined volume of liquid from the reagent bottle or sample tube
and dispenses it into a cuvette. Two stepper motors [6+7] move the shaft [8] of the pipetting
arm via timing belts.
The horizontal stepper motor [7] rotates the arm to defined positions over the reagent or
sample rotor and over the cuvette rotor. The encoder disc [11] and encoder opto board [12]
supervise exact positioning of the arm. The reset position is monitored by a home sensor [10]
that detects the horizontal reset vane [5] on the pulley that rotates the shaft [8]. This is
described in detail in par. 3.2.4.
The vertical stepper motor [6] lifts and lowers the arm to defined positions for aspiration and
dispensing of liquids. The vertical position for aspiration is determined by the remaining
volume in the reagent bottle or sample tube, as explained in par. 3.6.1. The reset position is
monitored by a reset opto board [4] that detects the vertical reset vane [9] on the shaft [8].
SELECTRA PRO
3-10 6003-400-450-02
M
System description
SELECTRA PRO
M 6003-400-450-02 3-11
System description
SELECTRA PRO
3-12 6003-400-450-02
M
System description
Info
i The signal from the detector is not the only factor in calculating the light
absorbance in the cuvette. Blank measurements are also done, to determine
the absorbance by the cuvettes. Optionally, sample blank measurements
are performed to determine the absorbance of sample without reagent.
SELECTRA PRO
M 6003-400-450-02 3-13
System description
SELECTRA PRO
3-14 6003-400-450-02
M
System description
SELECTRA PRO
M 6003-400-450-02 3-15
System description
SELECTRA PRO
3-16 6003-400-450-02
M
System description
Info
i If the concentrated waste option is installed on the instrument, the aspiration
needle in the first washing position is connected to the concentrated waste
container. All other aspiration needles are connected to the diluted waste
container.
SELECTRA PRO
M 6003-400-450-02 3-17
System description
SELECTRA PRO
3-18 6003-400-450-02
M
System description
Info
i The ISE unit is an optional module in the Selectra ProM.
Info
i A bypass electrode is available for systems that do not require all ISE
electrodes.
SELECTRA PRO
M 6003-400-450-02 3-19
System description
3.5.1 Overview
Info
i If the Selectra ProM is equipped with an ECO-kit, a separate container is
fitted for concentrated waste. This container is used for waste from the first
washing position and from the optional ISE unit. With the ECO-kit option, the
collecting cask has two separate chambers, for concentrated and diluted
waste. There is no functional difference between analyzers with and without
the ECO-kit.
SELECTRA PRO
3-20 6003-400-450-02
M
System description
SELECTRA PRO
M 6003-400-450-02 3-21
System description
SELECTRA PRO
3-22 6003-400-450-02
M
System description
Info
i The bellows pump contains six bellows. Four are used for the wash arm.
One is used to fill the rinsing place for the sample needle. The last bellows
is not used. It can be used as a spare when required.
SELECTRA PRO
M 6003-400-450-02 3-23
System description
Info
i If the Selectra ProM is equipped with an ECO-kit, the collecting cask
contains two chambers, as shown. If the ECO-kit is not installed, a simpler
collecting cask is fitted, where the smaller chamber and one of the valves are
not present. There is no functional difference.
SELECTRA PRO
3-24 6003-400-450-02
M
System description
Info
i If the Selectra ProM is equipped with an ECO-kit, a separate container is
fitted for concentrated waste. This container is used for waste from the first
washing position and from the optional ISE unit. The concentrated waste
container also has a float switch that monitors the fill level of the container.
SELECTRA PRO
M 6003-400-450-02 3-25
System description
At the wash position, 4 cuvettes are filled with 500 μl water by the bellows pump. The sample
needle rinse place is filled with the same amount of water. The water is drawn from the water
cask, which is automatically filled from the water container by the water pump.
Info
i The reagent needle rinse place is filled by the reagent needle itself. The
sample needle does not have sufficient volume. Therefore the sample
needle rinse place is filled by the bellows pump.
The filling sequence of the cuvettes and sample needle rinse place is as follows:
1) Valves V6 to V10 are switched off (positions 1 and 3 connected).
2) The bellows pump driving bracket moves down, drawing water into the bellows.
3) Valves V6 to V10 are switched on (positions 2 and 3 connected).
4) The bellows pump driving bracket moves up, pushing the water out of the bellows.
SELECTRA PRO
3-26 6003-400-450-02
M
System description
Emptying the cuvettes and needle rinse places is done by vacuum. The fluids are drawn into
the collecting cask. The vacuum in the collecting cask is created by the vacuum pump (see
par. 3.5.9).
Info
i The Selectra ProM may be fitted with the ECO-kit option. This adds a
concentrated waste circuit. The fluid diagram with the ECO-kit option is
described in par. 3.5.10.
The emptying sequence of the cuvettes and needle rinse places is as follows:
1) The vacuum pump creates a vacuum in the collecting cask.
2) Valve V2 switches on (positions 2 and 3 connected).
3) The contents of the cuvettes is sucked into the collecting cask.
4) Valve V2 switches off (positions 1 and 3 connected).
5) Valves V4 and V5 switch on (positions 2 and 3 connected).
6) Valve V2 switches on (positions 2 and 3 connected).
7) The contents of the reagent needle rinse place is sucked into the collecting cask.
8) Valve V5 switches off (positions 1 and 3 connected).
9) The contents of the sample needle rinse place is sucked into the collecting cask.
10 ) Valves V4 and V2 switch off (positions 1 and 3 connected).
SELECTRA PRO
M 6003-400-450-02 3-27
System description
Emptying the collecting cask is done by pressurized air. The fluids are forced into the diluted
waste container by creating pressure in the collecting cask. The required air pressure is
created by the vacuum pump.
Info
i The Selectra ProM may be fitted with the ECO-kit option. This adds a
concentrated waste circuit. The fluid diagram with the ECO-kit option is
described in par. 3.5.11.
SELECTRA PRO
3-28 6003-400-450-02
M
System description
Emptying the cuvettes and needle rinse places is done by vacuum. The fluids are drawn into
the collecting casks. The vacuum in the collecting casks is created by the vacuum pump (see
par. 3.5.11).
The emptying sequence of the cuvettes and needle rinse places is as follows:
1) The vacuum pump creates a a vacuum in the collecting casks.
2) Valves V2 and V14 switch on (positions 2 and 3 connected).
3) The contents of the cuvettes is sucked into the collecting casks.
4) Valves V2 and V14 switch off (positions 1 and 3 connected).
5) Valves V4 and V5 switch on (positions 2 and 3 connected).
6) Valve V2 switches on (positions 2 and 3 connected).
7) The contents of the reagent needle rinse place is sucked into the collecting cask.
8) Valve V5 switches off (positions 1 and 3 connected).
9) The contents of the sample needle rinse place is sucked into the collecting cask.
10 ) Valves V4 and V2 switch off (positions 1 and 3 connected).
SELECTRA PRO
M 6003-400-450-02 3-29
System description
Emptying the collecting casks is done by pressurized air. The fluids are forced into the waste
containers by creating pressure in the collecting casks. The required air pressure is created
by the vacuum pump.
The emptying sequence of the collecting casks is as follows:
1) Valves V2 and V14 switch off (positions 1 and 3 connected).
2) Valves V3 and V13 switch on.
3) Valves V1 and V12 switch on (positions 2 and 3 connected).
4) The air pressure on the collecting casks pushes the fluid into the waste containers.
5) Valves V1 and V12 switch off (positions 1 and 3 connected).
6) Valves V3 and V13 switch off.
SELECTRA PRO
3-30 6003-400-450-02
M
System description
Info
i The liquid detection is not primarily done to detect an empty bottle / tube. Its
first objective is to control the stepper motor that moves the needle into the
fluid. When the needle reaches the fluid, it is moved down by a defined
number of steps, before aspiration of the fluid starts. Keeping the needle as
close to the surface as possible minimizes the needle surface that comes
into contact with the fluid and thereby reduces pollution and contamination.
It also prevents the formation of a drop of fluid on the outside of the needle,
which would be transferred to the cuvette and change the defined volume
that is being pipetted.
SELECTRA PRO
M 6003-400-450-02 3-31
System description
Info
i The needle switch also works when the needle is blocked in its horizontal
movement, e.g. when a incorrectly placed tube or bottle blocks it.
SELECTRA PRO
3-32 6003-400-450-02
M
System description
SELECTRA PRO
M 6003-400-450-02 3-33
System description
Info
i Resetting the wash arm is done automatically at various occasions, e.g.
when the analyzer is switched on.
SELECTRA PRO
3-34 6003-400-450-02
M
System description
SELECTRA PRO
M 6003-400-450-02 3-35
System description
The Selectra ProM can run tests in mono or dual mode. In mono mode, only one reagent is
used for the test. In dual mode, two or three reagents can be used. Switching between the
mono and dual modes is possible, but as the modes use different timing cycles, this cannot
be done while measurements are being performed. When switching between the modes, the
analyzer will be reset.
422
690
12
-10
-50
45
0
R S ISE
675
519
280
24
-3
-135
54
0
175
283
R1 S ISE R2 R2/R3
SELECTRA PRO
3-36 6003-400-450-02
M
System description
RB K1-K22 E
422
690
12
-10
-50
45
R 0
S ISE
SELECTRA PRO
M 6003-400-450-02 3-37
System description
675
519
280
24
-3
-135
54
0
175
283
R1 S ISE R2 R2/R3
• When three reagents are used, R2 is added earlier and R3 takes the place of R2.
SELECTRA PRO
3-38 6003-400-450-02
M
System description
The cuvette rotor moves in a fixed cycle, making one full revolution plus one step in each
cycle. After one cycle, each cuvette has shifted one position forward compared to the fixed
positions of the functional units in the measurement section (see par. 3.7.5).
Measurements are always performed at fixed points in the cuvette timing cycle. This means
that each action takes place in the next cuvette compared to the previous cycle.
Measurements are always performed, even if they are not required for the sample. The
software in the analyzer determines which measurements are used to calculate the test
results.
K22 K11 K1
0 5 10 15 20
Info
i In dual mode, the cuvette cycle is 27 seconds and the kinetic measurements
are divided over two phases within that cycle. The basic principle remains
the same. Details about the mono and dual mode cycles are given below.
The cuvette rotor can be compared to a production line in a factory, where each cuvette
passes the various stages in the production process at fixed locations and in a fixed timing. If
the analyzer would not use a rotor but a linear sample transport chain, this would require
many measuring units along that production line.
A B
0 5 10 15 20
Fig.3-37 Cuvette timing in mono mode
The cuvette peforms one rotation in 20 seconds. In the first part of its rotation cycle [A], a
number of actions take place to prepare for new measurements. This includes washing the
cuvette, dispensing the reagent and adding the sample. In the second part of the rotation
cycle [B], measurements take place. 22 kinetic measurements are done on subsequent
cuvettes in the rotor. Also, endpoint and reagent blank measurements are performed.
SELECTRA PRO
M 6003-400-450-02 3-39
System description
A B C D
0 5 10 15 20 25 27
Fig.3-38 Cuvette timing in dual mode
The cuvette performs one rotation in 27 seconds. In the first part of its rotation cycle [A], a
number of actions take place to prepare for new measurements. This includes washing the
cuvette, dispensing the first reagent and adding the sample. Also, dichromatic endpoint
measurements are done in this cycle. In the second part of the rotation cycle [B], half of
the kinetic measurements take place. The other half take place during the last part of the
cycle [D]. Also, dichromatic reagent blanks are performed in this part of the cycle. In the
third part of the cuvette rotation cycle [C], addition of second and third reagents, sample
blanks and predilution of the sample take place. Which actions are required depends on
the tests that were requested for each sample.
SELECTRA PRO
3-40 6003-400-450-02
M
System description
SELECTRA PRO
M 6003-400-450-02 3-41
System description
3.8.1 Overview
The Selectra ProM is controlled by a set of electronics boards in the back of the instrument. A
panel PC with touch screen supplies the user interface. The main control board in the
instrument is the System Board (SB), which holds the core microprocessor that drives all the
functional units on other boards. The Motion Control Boards (MCB) each drive two stepper
motors. The Temperature Control Board (TCB) monitors and regulates the temperature of
several components. The Photometer Board (PB) processes the signals from the light
detectors.
Info
i PCB layouts and electrical diagrams are listed in the "Electronics" chapter.
That chapter also contains the wiring diagrams that specify how the boards
are interconnected.
SELECTRA PRO
3-42 6003-400-450-02
M
System description
Master CPU
SIO 1
Serial communication with slave controllers
KIO SIO
CPU
SIO 2
PIO
CTC
CTC
Drivers for
pumps and valves Photometer signals
SELECTRA PRO
M 6003-400-450-02 3-43
System description
Pipettor A2a
Vertical movement of syringes (bottom)
Info
i When the stepper motor runs normally, it may seem that the LEDs are on
continuously, because the switching time is much faster than the human eye
can detect.
SELECTRA PRO
3-44 6003-400-450-02
M
System description
Photo
Diode Photometer Measuring signals
Circuit (to system board)
Photo
Diode
(Ref.)
Photo
Diode
IC 11 Vx
b1
G0
b2
G1
Gain
b4 Control
G2
b8
G3
b(ase) = 1,5
SELECTRA PRO
M 6003-400-450-02 3-45
System description
Info
i The theoretical maximum of the photometer range is 0.436 + 3.82 = 4.256
Abs. At this level, the noise and drift effects are very high. Therefore, the
maximum gain step that is used in the analyzer software is 12. This allows a
maximum measurable absorbance of 3.06 + 0.436 = 3.496.
0 1 0 0 0 1 0
1 1 0 0 1 1.5 0.255
2 1 0 1 0 2.25 0.51
3 1 0 1 1 3.38 0.765
4 1 1 0 0 5.06 1.02
5 1 1 0 1 7.6 1.275
6 1 1 1 0 11.4 1.53
7 1 1 1 1 17.1 1.785
8 0 0 0 0 25.6 2.04
9 0 0 0 1 38.4 2.295
10 0 0 1 0 57.7 2.55
11 0 0 1 1 86.5 2.805
12 0 1 0 0 129.7 3.06
13 0 1 0 1 194.6 3.315
14 0 1 1 0 292 3.57
15 0 1 1 1 438 3.82
• Signal G3 is active high, so its effect is reversed when compared to the other
signals.
Lin/log converter
Photometer board System board
start/stop
CLK
(from KIO)
99 IC5
10V
C4 f IC7a
1 f/2
CTC
100mV reference & 16-bits
Vx IC13 IC6b counter
Input
Amplifier
IC2
meas.
/ref.
Conversion of the linear signal from the amplifier to the required logarithmic scale is done
by using the logarithmic discharge curve of capacitor C4. The capacitor is first charged to
10V. The voltage Vx is compared to the voltage over the capacitor. The time from the start
SELECTRA PRO
3-46 6003-400-450-02
M
System description
of the discharge to the point where the voltages are equal corresponds to the log of the
measured signal. This time is measured by starting a counter at the start of the discharge
and stopping it when the two voltages are equal. The counter is placed on the system
board.
10V
Vcapacitor
reported count
Vmeasured
0V
counter
0
To compensate for possible temperature and component variations, the conversion is also
done for a defined reference signal. Conversions of actual measurements and reference
signals are constantly alternated.
The reference signal voltage is fixed at 100 mV, which corresponds to 20 000 counts and
2.0 Abs. When the reference count differs from 20 000, the count for actual measurements
is changed proportionally. When the reference count is more than 1 000 counts off the
target of 20 000, an error is reported.
SELECTRA PRO
M 6003-400-450-02 3-47
System description
Info
i Using the operating panel on the temperature control board is described in
the Electronics chapter. See par. 6.7.3.
The system board can monitor the temperature within ± 0.5 ºC via the READY signal. If
the target temperature is not reached within 5 minutes the microcontroller generates a
cuvette temperature error.
Info
i If the cuvette temperature error is acknowledged the error will not re-appear
until the instrument is reset. The heating circuit is switched off.
SELECTRA PRO
3-48 6003-400-450-02
M
System description
Lamp voltage
SELECTRA PRO
M 6003-400-450-02 3-49
System description
The cooling unit board drives two Peltier elements and a heat exchanger. The Peltier
elements are provided with controlled power via the cooling unit board. The cooling
temperature is set to 10º C. If the temperature sensor (cold side) measures a higher
temperature of the primary cooler, the measured difference is used to determine the power
required for cooling.
The Peltier elements use water cooling to release the heat. The heated water is transported
to a heat exchanger with forced cooling. The cooling unit board controls a pump and fan that
are always on. For protection of the system, two extra temperature sensors (hot side) are
fitted. When they detect overheating (60º C), these switches interrupt the power to the Peltier
elements and send an alarm to the system board. When the temperature drops below 40º C,
the Peltier elements are switched again on and cooling is resumed.
SELECTRA PRO
3-50 6003-400-450-02
M
System description
Needle
Sample Sample
MCB
Arm Liquid
Sample Arm
Board Detection
Oscillator
Needle
Oscillator
Info
i The LLD board is often referred to as PLL board. LLD refers to the function
(Liquid Level Detection), whereas PLL refers to the electronic mechanism
that is used. The name LLD is preferred.
The liquid detection circuits for the pipetting needles are located on the LLD board. A 10V 10
kHz oscillator signal is generated on the board and connected to the rotor. The metal base
plate of the rotor and the metal needle serve as a capacitor, with the liquid in the reagent
bottle or sample tube serving as the dielectric. When the needle touches the liquid, the
oscillating signal is picked up and fed back to the LLD board.
When the return signal is detected by the LLD board, a signal is sent to the motion control
board for the pipetting arm. The sensitivity of the liquid detection circuit can be adjusted by
potentiometers on the LLD board.
SELECTRA PRO
M 6003-400-450-02 3-51
System description
The DC power for the analyzer is supplied by a switched power supply, which accepts input
voltages from 95 to 264 V AC with frequencies between 47 and 63 Hz. The specifications of
the power supply and the voltages used by the various components in the analyzer are listed
in the table below.
Fuses
The analyzer is fitted with fuses to protect the main components. The specifications and
locations of all fuses are listed below.
! CAUTION
TAKE CARE
When fuses are replaced, the new fuses must match these specifications
exactly. If other fuses are fitted, all warranty will be void.
SELECTRA PRO
3-52 6003-400-450-02
M
4
Installation
This chapter contains instructions for transport, unpacking and installation of the analyzer.
Instructions on re-packaging the analyzer for transport after it has been in use are also
included.
SELECTRA PRO
M 6003-400-450-02 4-1
Installation
NOTICE
PLEASE NOTE
i The boxed instrument should be stored in a dry environment at temperatures
between -10 and +45 ºC (14 to 113 ºF).
SELECTRA PRO
4-2 6003-400-450-02
M
Installation
To prepare the installation site for the analyzer, make sure the following requirements can be
met:
• Only instruments that meet the relevant safety requirments may be connected to the
analyzer.
• Only use UL-listed power supply cables and power distribution blocks.
• The analyzer should not be placed near to sources of strong electromagnetic
radiation (e.g. unshielded intentional radio transmitters).
• The analyzer must be placed on a level surface. Make sure ergonomic
requirements are met.
• The mounting surface must be strong enough to carry the analyzer plus possible
peripherals without bending.
• The mounting surface must be free of vibrations.
• A reliable power connection must be available. See the technical specifications.
• The analyzer and peripherals must be well grounded.
• It is recommended to use an Uninterruptable Power Supply (UPS).
• The analyzer must be free of the wall on all sides.
• The power cords should be accessible without having to move the analyzer.
• Adequate ventilation in the room and around the instrument must be available to
prevent overheating and condensation.
• The analyzer must not be exposed to direct sunlight.
• Distilled or purified water (at least ASTM type II) must be available.
• The analyzer should not be exposed to extreme temperatures or humidity. See the
technical specifications in par. A.1.3.
! CAUTION
TAKE CARE
The analyzer and optional additional devices, parts and accessories must
only be unpacked and installed by a qualified service engineer from Vital
Scientific or a designated representative. Vital Scientific does not accept
liability for any consequential damage, loss or costs as a result of non-
conformance to this rule.
SELECTRA PRO
M 6003-400-450-02 4-3
Installation
4.1.3 Unpacking
Info
i Two boxes contain the panel PC and accessories. The other boxes are
empty. These are used to keep the contents from moving around during
transport.
SELECTRA PRO
4-4 6003-400-450-02
M
Installation
! WARNING
HEAVY LOAD
Use care when lifting, removing or installing the analyzer. The weight of the
analyzer is 93 kg. Make sure there are enough people to safely lift the
analyzer. Consult local sources for additional information on the locally valid
workplace safety regulations for lifting a load.
NOTICE
SELECTRA PRO
M 6003-400-450-02 4-5
Installation
NOTICE
NOTICE
PLEASE NOTE
i In various countries there are regulations on the disposal of packaging
waste. If you choose to discard the packing materials, consult local sources
for additional information on correct disposal of waste.
SELECTRA PRO
4-6 6003-400-450-02
M
Installation
4.2 Installation
The panel PC is mounted on an arm that allows moving the panel PC to a convenient working
position. The panel PC and the mounting arm come in separate boxes and must be
assembled and mounted on the analyzer.
NOTICE
PLEASE NOTE
i Do not use the mounting arm documentation to adjust the tension of the arm.
Adjusting the tension should be done as described below.
NOTICE
NOTICE
PLEASE NOTE
i In various countries there are regulations on the disposal of packaging
waste. If you choose to discard the packing materials, consult local sources
for additional information on correct disposal of waste.
SELECTRA PRO
M 6003-400-450-02 4-7
Installation
Fig.4-7 Attaching the mounting arm and keyboard support to the panel PC
Use the enclosed long screws and a Philips head size 1 screwdriver. Tighten all four
screws.
3 Attach the arm to the panel PC (without the optional keyboard support)
SELECTRA PRO
4-8 6003-400-450-02
M
Installation
SELECTRA PRO
M 6003-400-450-02 4-9
Installation
SELECTRA PRO
4-10 6003-400-450-02
M
Installation
SELECTRA PRO
M 6003-400-450-02 4-11
Installation
SELECTRA PRO
4-12 6003-400-450-02
M
Installation
If the analyzer is ordered with the ECO-kit, a 2 liter concentrated waste container is available
in the accessories kit. In that case, perform the following step.
4 Install the concentrated waste container
a) Attach the CONCENTRATED WASTE sticker to the 2 liter container.
b) Attach the BIOHAZARD sticker to the container.
c) Screw the concentrated waste cap on the concentrated waste container.
d) Place the concentrated waste container next to the analyzer. Make sure it is accessible
for the operator.
e) Lead the tubes and sensor level cable through the opening in the right panel.
f) Connect the tubes to the connection plate behind the container. Make sure the numbers
on the tubes and on the connections match.
g) Connect the level sensor cable to the connector marked CONCENTRATED WASTE.
SELECTRA PRO
M 6003-400-450-02 4-13
Installation
SELECTRA PRO
4-14 6003-400-450-02
M
Installation
Info
i The File Based Write Filter is a safety feature of Windows. It is used to
prevent unauthorized or accidental overwriting of important system
information. It must be deactivated to allow the installation of the analyzer
software.
4 Log in as Maintenance
5 Install the application software
a) Insert the USB flash drive with the application software in a free USB port.
b) Open Windows Explorer and navigate to the USB flash drive.
c) Double-click the Setup application on the USB flash drive.
d) Reboot the panel PC when the installation is finished.
6 Log in as Maintenance
7 Activate the File Based Write Filter
Touch the button Enable write filter. Then touch the OK button in the confirmation dialog
that appears. The panel PC reboots automatically with the File Based Write Filter switched
on.
8 Switch on the analyzer
9 Log in as Analyzer
10 Start the analyzer software
Double-click (or double-tap) the Elitech Clinical Systems logo on the desktop.
SELECTRA PRO
M 6003-400-450-02 4-15
Installation
NOTICE
PLEASE NOTE
i Only touch the cuvette rotor on the handle in the center. Never touch the
cuvettes or the sides of the cuvette rotor.
SELECTRA PRO
4-16 6003-400-450-02
M
Installation
SELECTRA PRO
M 6003-400-450-02 4-17
Installation
If the analyzer is going to be switched off for a relatively long time (e.g. several weeks), some
precautions must be taken to prevent damage to some of the components.
SELECTRA PRO
4-18 6003-400-450-02
M
Installation
NOTICE
PLEASE NOTE
i The analyzer should be stored and/or transported in a dry condition. All
system solution and cooling liquid must be removed.
NOTICE
PLEASE NOTE
i Preparing the analyzer for transport should only be done by a qualified
service engineer from Vital Scientific or its designated representative.
a) Disconnect the upper tube from the mounting block of the sight glass [1].
b) Place the end of the upper tube in a container to catch the cooling liquid. The container
should be placed below the level of the analyzer.
c) Remove the cap [2] from the sight glass.
d) Switch the analyzer on.
e) Wait until all cooling liquid is pumped into the container.
f) Swtich the analyzer off.
g) Place the cap [2] on the sight glass.
h) Connect the tube [1] to the mounting block of the sight glass.
NOTICE
PLEASE NOTE
i Do not keep the analyzer switched on after the cooling liquid is drained. This
may lead to overheating of the cooling pump.
SELECTRA PRO
M 6003-400-450-02 4-19
Installation
4.4.1 Preparations
If the analyzer software and/or operating system must be upgraded or restored, all customer-
specific data must first be backed up, so that they can be restored after the new operating
system and/or analyzer software is installed. This can be done by creating a restore point,
which allows restoring the analyzer to its previous state after upgrading the software. See the
User Manual for instructions on creating and restoring a restore point.
SELECTRA PRO
4-20 6003-400-450-02
M
Installation
SELECTRA PRO
M 6003-400-450-02 4-21
Installation
Info
i If no backup data is available, you can use the installation procedure that is
performed for new analyzers. See par. 4.2.4.
1 Log in as Maintenance
2 Deactivate the File Based Write Filter
Touch the button Disable write filter. Then touch the Yes button in the confirmation dialog
that appears. The panel PC reboots automatically with the File Based Write Filter switched
off.
Info
i The File Based Write Filter is a safety feature of Windows. It is used to
prevent unauthorized or accidental overwriting of important system
information. It must be deactivated to allow the installation of the analyzer
software.
3 Log in as Maintenance
4 Install the application software
a) Insert the USB flash drive with the application software in a free USB port.
b) Open Windows Explorer and navigate to the USB flash drive.
c) Double-click the Setup application on the USB flash drive.
5 Activate the File Based Write Filter
Touch the button Enable write filter. Then touch the Yes button in the confirmation dialog
that appears. The panel PC reboots automatically with the File Based Write Filter switched
on.
6 Switch on the analyzer
7 Log in as Analyzer
8 Start the analyzer software
Double-click (or double-tap) the Elitech Clinical Systems logo on the desktop.
SELECTRA PRO
4-22 6003-400-450-02
M
5
Software
This chapter contains descriptions of the service software available for the instrument.
SELECTRA PRO
M 6003-400-450-02 5-1
Software
5.1 Overview
The service menu gives access to five screens. The first one is a submenu that offers a
choice of service functions. The other screens give direct access to supporting functions. The
main menu and submenu are described together in this paragraph.
SELECTRA PRO
5-2 6003-400-450-02
M
Software
Submenu choices
The state of various subsystems is shown on most of the service screens. The signals
indicate the status of the following subsystems:
• PIPE: Pipettor drive
• CUVR: Cuvette rotor drive
• SARM: Sample arm drive
• RARM: Reagent arm drive
• SDSC: Sample disc drive
• RDSC: Reagent disc drive
• WARM: Wash arm drive
SELECTRA PRO
M 6003-400-450-02 5-3
Software
Info
i The top of the screen shows the date and time when the last needle rinse
was performed. The lower part of the screen gives instructions on
performing a needle rinse procedure.
Function buttons
Wash Rotor : Start a complete washing cycle of the cuvette rotor.
Wash/Fill Rotor : Start a complete washing cycle of the cuvette rotor. After the washing
cycle, the cuvettes are filled with water.
Needle Rinse : Start a cleaning cycle of the sample and reagent needles.
Return : Return to the Service screen.
SELECTRA PRO
5-4 6003-400-450-02
M
Software
Info
i The left section of the screen shows the list of dates/times when errors were
stored. Select one of the entries in the list to see all error messages of that
date/time in the list to the right.
Parameters
Error text : Error that was generated and shown in an error dialog window. For a full list of
possible error messages, see par. 7.1.
User action : Action that was performed by the user. The options are determined by the
available buttons in the error dialog window.
End time : Time when the error signal was lowered (when the error was resolved).
Function buttons
Print One : Print a full error report for the currently selected error.
Print All : Print an error report for all errors shown on the screen.
Service : Return to the Service screen.
Main Menu : Return to the Main screen.
SELECTRA PRO
M 6003-400-450-02 5-5
Software
The Changes Log screen shows all logged user actions. Each action is shown with the time
and date plus the action type and some details.
Parameters
Date : Date when the action was performed.
Time : Time when the action was performed.
Access level : Login level of the user that performed the action.
Action : Action that was performed by the user.
Details : Short description of the item on which an action was performed.
Function buttons
Save And Clear : Open a dialog to save the current list to a file. Then clear the list.
Service : Return to the Service Menu screen.
Main Menu : Return to the Main screen.
SELECTRA PRO
5-6 6003-400-450-02
M
Software
Parameters
Wavelengths : If filters in the filter wheel are replaced by filters with other wavelengths, the
wavelengths must be changed accordingly in this section.
Run mode : Choose between mono or dual mode. In mono mode, only one reagent can
be used per assay; in dual mode, two or three reagents can be used. For a detailed
description of mono and dual modes, see par. 3.7.1.
Info
i Depending on the type of system, you may or may not be able to change this
setting. On closed systems, assays that require two or three reagents cannot
be changed. The mono mode option is not available on such systems.
Use system barcode : Enable or disable the built-in PSID barcode scanner.
ISE installed : Enable or disable the built-in ISE system.
Non-clinical use : These fields allow renaming various standard texts that appear in many
screens in the analyzer software. Redefining the terms may be helpful when the
analyzer is used in non-clinical settings, such as veterinarian practices.
Sample Types : Defining more, or different, sample types makes these types appear in the
various screens of the analyzer software.
Info
i Adding sample types increases the lists for some parameters, such as
reference values. Make sure to also set the reference values for the added
sample types.
Function buttons
Service : Return to the Service screen.
Main Menu : Return to the Main screen.
SELECTRA PRO
M 6003-400-450-02 5-7
Software
Parameters
Lamp absorbance : Absorbance measured by the reference detector.
Cuvette absorbance : Absorbance measured by the detector in the cuvette rotor.
Info
i The analyzer keeps measuring both absorbances while this screen is open.
The countdown at the top only counts down from the moment the screen was
opened.
Function buttons
Filter Check : Open the Filter Check screen.
Return : Return to the Service screen.
SELECTRA PRO
5-8 6003-400-450-02
M
Software
Info
i This screen can only be opened from the Adjust Lamp screen.
Parameters
340 - 700 : Absorbance measured by detector in the cuvette rotor for each of the filter
wavelengths.
Info
i The analyzer keeps checking the filter absorbance while this screen is open.
The blue arrows indicating the measured absorbance are refreshed in a
continuous cycle from left to right.
Function buttons
Lamp Adjustment : Open the Adjust Lamp screen.
Return : Return to the Service screen.
SELECTRA PRO
M 6003-400-450-02 5-9
Software
Info
i The Sample Arm screen is almost identical to the Reagent Arm screen.
The only difference is that the sample arm moves in the opposite direction
and has fewer positions over the rotor.
Info
i The state graph in the bottom section of the screen is the same in most
service screens. It is described in a subsection of par. 5.1.1.
Arm positions
The graph in the top section of the screen shows the current position of the arm: this is
indicated by the red square. The reagent arm can move to 7 distinct positions horizontally:
5 positions over the reagent rotor, one position over the needle rinse place and one over
the cuvette rotor. The sample arm has 3 positions over the sample rotor, so there are only
5 positions in total.
In all positions except the cuvette position, the arms can be in one of two vertical positions:
up (reset position) or down. In the cuvette position, an intermediate vertical position is
possible.
Info
i While the drives are enabled, the arms can be moved to all available
positions using the cursor keys.
Status signals
The status of various signals on the reagent and sample arm driver boards are shown
below the position graph.
Horizontal reset opto : The vane on the horizontal drive is detected by an optical switch.
When the vane is positioned inside the optical switch, the state of this signal is 0.
Encoder opto A/B : The horizontal drive is fitted with an encoder disc. The edges of the
disc are detected by two optical switches on the encoder board. When the disc moves
over a small distance, one of the optical switches changes its state.
Vertical reset opto : The vertical drive of the arm is fitted with a vane that moves through
a set of three optical switches on the arm opto board. When the vane is positioned in the
upper optical switch, the state of this signal is 0.
SELECTRA PRO
5-10 6003-400-450-02
M
Software
Liquid sense : The liquid sensing system is described in par. 3.6.1. When liquid is
detected, the state of this signal is “Yes”.
Needle switch : If the needle is blocked in its downward movement, this is detected by the
needle switch inside the reagent or sample head. If the needle switch is activated, the
state of this signal is 0.
Max. down bit : If the vane on the vertical drive of the arm is positioned in the lower optical
switch, the state of this signal is 0.
Function buttons
Reset Arm : Move the reagent or sample arm to its home position (both horizontally and
vertically).
Disable Hor. Drive : Switch off the horizontal drive control. When switched off, the arm can
be rotated manually.
Disable Vert. Drive : Switch off the vertical drive control. When switched off, the arm can
be moved up and down manually.
Start Stirring : Start the stirrer drives. This is used to adjust the stirrer position. See par.
6.4.11.
Stop Stirring : Switch off the stirrer drives.
Main Menu : Return to the Main screen.
Info
i Enabling the horizontal and vertical drives is done by resetting the arm.
SELECTRA PRO
M 6003-400-450-02 5-11
Software
Info
i The state graph in the bottom section of the screen is the same in most
service screens. It is described in a subsection of par. 5.1.1.
Status signals
Disk position : The position of the rotor is monitored by the motion control board. The
board keeps track of every displacement (signalled by the encoder optos). When the
rotor is in the home position (the reset opto signal is 0), the Disk position is 1.
Reset opto : The vane on the rotor drive is detected by an optical switch. When the vane
is positioned inside the optical switch, the state of this signal is 0.
Encoder opto A/B : The drive is fitted with an encoder disc. The edges of the disc are
detected by two optical switches on the encoder board. When the disc moves over a
small distance, one of the optical switches changes its state.
Function buttons
Reset Disk : Move the rotor to its home position.
Edit Position : Open a dialog in which a position can be entered. Confirming the edit
position makes the rotor move to the specified position.
Disable Drive : Switch off the drive control. When switched off, the rotor can be rotated
manually.
Main Menu : Return to the Main screen.
Info
i Enabling the rotor drive is done by resetting the disk.
SELECTRA PRO
5-12 6003-400-450-02
M
Software
Info
i The state graph in the bottom section of the screen is the same in most
service screens. It is described in a subsection of par. 5.1.1.
Status signals
Disk position : The position of the rotor is monitored by the motion control board. The
board keeps track of every displacement (signalled by the encoder optos). When the
rotor is in the home position (the reset opto signal is 0), the Disk position is 1.
Barcode position : Rotor position that is facing the internal barcode scanner.
Barcode information : Barcode that is read by the internal barcode scanner (only when
activated). If no valid barcode is read, the string “no read” is shown here.
Reset opto : The vane on the rotor drive is detected by an optical switch. When the vane
is positioned inside the optical switch, the state of this signal is 0.
Encoder opto A/B : The drive is fitted with an encoder disc. The edges of the disc are
detected by two optical switches on the encoder board. When the disc moves over a
small distance, one of the optical switches changes its state.
Function buttons
Reset Disk : Move the rotor to its home position.
Edit Position : Open a dialog in which a position can be entered. Confirming the edit
position makes the rotor move to the specified position.
Edit Barcode Position : Open a dialog in which a position can be entered. Confirming the
edit barcode position makes the rotor move until the specified position is facing the
internal barcode scanner.
Scan Barcode : Activate the internal barcode scanner. The barcode is shown on the
screen.
Continuous Scanning On : Switch the barcode scanner to continuous mode. This is used
when adjusting the position of the internal barcode scanner.
Disable Drive : Switch off the drive control. When switched off, the rotor can be rotated
manually.
Return : Return to the Main screen.
Info
i Enabling the rotor drive is done by resetting the disk.
SELECTRA PRO
M 6003-400-450-02 5-13
Software
Info
i The state graph in the bottom section of the screen is the same in most
service screens. It is described in a subsection of par. 5.1.1.
Status signals
Disk position : The position of the cuvette rotor is monitored by the motion control board.
The board keeps track of every displacement (signalled by the encoder optos). When
the cuvette rotor is in the home position, the Disk position is 1.
Filter position : The position of the filter wheel is monitored by the motion control board.
The board keeps track of every displacement of the stepper motor. When the filter wheel
is in the home position, the filter wheel position is 3.
Disk reset opto : A slit in the cuvette rotor base is detected by an optical switch. When the
slit is positioned inside the optical switch, the state of this signal is 1.
Disk encoder opto A/B : The cuvette rotor drive is fitted with an encoder disc. The edges
of the disc are detected by two optical switches on the encoder board. When the disc
moves over a small distance, one of the optical switches changes its state.
Filter reset opto : A hole in the rim of the filter wheel is detected by an optical switch. When
the hole is positioned inside the optical switch, the state of this signal is 1.
Function buttons
Reset Cuvette Rotor : Move the cuvette rotor to its home position.
Reset Filter : Move the filter wheel to its home position.
Disable Cuvette Drive : Switch off the cuvette drive control. When switched off, the
cuvette rotor can be rotated manually.
Disable Filter : Switch off the filter wheel drive control. When switched off, the filter wheel
can be rotated manually.
Edit Cuvette Position : Open a dialog to enter a rotor position. Confirming makes the
cuvette rotor move to the specified position. Positions 1 to 48 can be chosen.
Edit Filter Position : Open a dialog to enter a filter wheel position. Confirming makes the
filter wheel move to the specified position. Positions 1 to 8 can be chosen.
Lamp On : Switch the lamp on. This is done when adjusting the lamp and cuvette rotor
positions. Touch the button again to switch the lamp off.
Main Menu : Return to the Main screen.
Info
i Enabling the cuvette drive is done by resetting the cuvette rotor. Enabling
the filter wheel drive is done by resetting the filter wheel.
SELECTRA PRO
5-14 6003-400-450-02
M
Software
Info
i The state graph in the bottom section of the screen is the same in most
service screens. It is described in a subsection of par. 5.1.1.
Arm positions
The graph in the top section of the screen shows the current position of the arm: this is
indicated by the red square. The wash arm can move to 3 distinct positions vertically.
Info
i While the drive is enabled, the arm can be moved to all available positions
using the cursor keys.
Status signals
The status of various signals on the wash arm driver board are shown below the position
graph.
Arm opto 0-2 : The drive of the arm is fitted with a vane that moves through a set of three
optical switches on the arm opto board. When the vane is positioned in an optical switch,
the state of its signal is 0. Opto 0 is located at the top.
Overflow detection : The cuvette overflow detection system is described in par. 3.6.3.
When a too high liquid level in the cuvettes is detected, the state of this signal is “Yes”.
Cuvettes present : The cuvette rotor detection system is described in par. 3.6.4. When a
cuvette rotor is detected, the state of this signal is “Yes”.
Function buttons
Reset Wash Arm : Move the wash arm move to its home position.
Disable Drive : Switch off the wash arm drive control. When switched off, the arm can be
moved up and down manually.
Main Menu : Return to the Main screen.
Info
i Enabling the wash arm drive is done by resetting the arm.
SELECTRA PRO
M 6003-400-450-02 5-15
Software
Arm positions
The graph in the top section of the screen shows the current position of the arm: this is
indicated by the red square. The ISE arm can move to 2 distinct positions vertically.
Info
i The arm can be moved down using the arrow down key. The arm moves up
automatically after 5 seconds.
Status signals
The status of signals on the ISE/stirrer board are shown below the position graph.
Arm opto : The arm is fitted with a vane that moves into an optical switch on the ISE/stirrer
board. When the vane is detected, the ISE arm is in the “down” position.
Arm control : Shows the status of the ISE arm control circuit. If this signal shows “down”,
the ISE arm should normally move down.
Function buttons
Info
i The information that is available via the function keys does normally not fit
on one screen. Touching the same key again allows jumping from one page
of data to the next. After the last available page is shown, touching the key
again returns to the first available page.
Meas. Data : Shows data from the measurement buffer (13 pages).
EEPROM Data : Shows parameter settings in the EEPROM (6 pages).
Calib. Data : Shows data about calibrations performed (11 pages).
Flow Data : Shows parameter settings of the bubble detector (9 pages).
Pump Test : Activates the peristaltic pump twice (2 pages).
ADC Test : Performs A/D converter self-tests (4 pages).
Valve Test : Opens and closes both valves several times (1 page).
Sipper Test : Checks the ISE sipper control, without moving the arm (2 pages).
Return : Return to the Main screen.
Info
i Access to this screen is usually done in close collaboration with the System
Support Department at Vital Scientific. Therefore, the data is not described
in detail in this documentation.
SELECTRA PRO
5-16 6003-400-450-02
M
Software
Info
i The state graph in the bottom section of the screen is the same in most
service screens. It is described in a subsection of par. 5.1.1.
Status signals
Reagent / Sample valve : The state of the valve: “Water” or “Needle” position.
Reagent / Sample Syringe : The current position of the plunger (which determines the
volume drawn into the syringe), shown in 1000 steps. The steps represent 1 μl for the
reagent syringe and 0.1 μl for the sample syringe. The 0 position is at the top (the
plunger is pushed all the way into the syringe).
Reagent / Sample valve opto 0/1 : The state of the optical switches on the valves opto
board. Opto 0 detects the “Water” position, opto 1 detects the “Needle” position. When
the positioning pin is in an optical switch, its signal is 0.
Reagent / Sample syr res. / down opto : The vane on the syringe carriage is detected by
two optical switches on the syringe opto board. The upper opto detects the reset
position. The lower opto detects the maximum downward position. When the vane is in
an optical switch, its signal is 0.
Function buttons
Reset Pipettor : Move the syringes to their home (upper) position and the valves to their
“Water” position.
Reag. / Sample Water Position : Move the reagent or sample valve to the “Water”
position.
Reag. / Sample Needle Position : Move the reagent or sample valve to the “Neelde”
position.
Reagent / Sample Syringe : Open a dialog to enter a syringe position. Confirming makes
the syringe move to the specfied position. Positions between 0 (top) and 1000 (bottom)
can be chosen.
Reagent / Sample Full Stroke : Make the syringe do a full stroke (reset, move all the way
down and all the way up again).
Return : Return to the Main screen.
SELECTRA PRO
M 6003-400-450-02 5-17
Software
Info
i The state graph in the bottom section of the screen is the same in most
service screens. It is described in a subsection of par. 5.1.1.
Status signals
Valve # : Current state of the valve. States can be “On” (opened) or “Off” (closed).
Vacuum pump : Current state of the vacuum pump. States can be “On” or “Off”.
Vacuum sensor : Current state of the vacuum switch. States can be “Vacuum” or “No
vacuum”.
Function buttons
Valve # : Toggle the valve between its open and closed positions.
Vacuum Pump : Switch the vacuum pump on. Touch the button again to switch the pump
off.
Return : Return to the Main screen.
SELECTRA PRO
5-18 6003-400-450-02
M
Software
Info
i The state graph in the bottom section of the screen is the same in most
service screens. It is described in a subsection of par. 5.1.1.
Status signals
Valve # : Current state of the valve. States can be “On” (opened) or “Off” (closed).
Bellows pump : Current position of the bellows pump. States can be “Up” or “Down”. The
current position may also be shown. Positions can be between 0 (up) and 850 (down).
Water pump : Current state of the water pump. States can be “On” or “Off”.
Pump opto up / down : Current state of the optical switches in the bellows pump unit. The
switches detect a slit in the flange on the bellows pump driving axle. When the slit is
inside one of the optos, its signal is 1.
Waste sensor : Current state of the waste sensor. States can be “Full” or “Not Full”.
Cover : Current state of the analyzer cover. States can be “Open” or “Closed”
Water running out / Insuff. water : Current states of the liquid sensors inside the water
cask. The water running out sensor is fitted lower and causes a warning only. The
insufficient water sensor causes the analyzer to stop. States can be “Yes” or “No”.
Function buttons
Valve # : Toggle the valve between its open and closed positions.
Reset Bellows Pump : Move the bellows pump to its home (upper) position.
B.Pump Down : Move the bellows pump to its lowest position.
B.Pump Posit. : Open a dialog to enter a bellows pump position. Confirming makes the
bellows pump move to the specfied position. Positions between 0 (top) and 850 (bottom)
can be chosen.
Water Pump : Switch the water pump on. Touch the button again to switch the pump off.
Return : Return to the Main screen.
SELECTRA PRO
M 6003-400-450-02 5-19
Software
Info
i For a detailed description of the optical electronics system for which the data
is shown on this screen, see par. 3.8.4.
Status signals
Wavelength : Filter for which the optical electronics data is shown. The function buttons
allow stepping through the available filters.
Gain : Currently selected gain. Values can be 0 to 15.
Abs. : Measured absorbance.
meas. / ref. : Counter values for the measurement and reference signals.
err_ga : Error code. The shown number is composed of two hexadecimal digits: an error
and a gain digit. The gain digit has values 0 to F (15). The error digit represents a bitmap
and has the following meanings:
0 0000 No errors
1, 3, 5, 7, 9, B, D, F xxx1 Underrange error (signal)
2, 3, 6, 7, B, C, E, F xx1x Overrange error (signal)
4, 5, 6, 7, C, D, E, F x1xx Underrange reference counter (CTC)
8, 9, A, B, C, D, E, F 1xxx Overrange reference counter (CTC)
Function buttons
Next / Previous Filter : Select the next or previous filter to view its data.
Blank : Perform a blank measurement for the currently selected filter.
Inc. / Dec. L./C. gain : Increase or decrease the gain for the lamp signal by one step.
Inc. / Dec. C. gain : Increase or decrease the gain for the cuvette signal by one step.
Return : Return to the Main screen.
SELECTRA PRO
5-20 6003-400-450-02
M
Software
The Electronics screen shows a procedure for checking the power supply voltages. See par.
6.7.1.
! CAUTION
TAKE CARE
The listed procedures should only be performed by qualified and trained
electrical engineers. Contact the System Support department of Vital
Scientific when you have doubts.
Function buttons
Return : Return to the Main screen.
SELECTRA PRO
M 6003-400-450-02 5-21
Software
SELECTRA PRO
5-22 6003-400-450-02
M
6
Maintenance
This chapter contains the maintenance schedule for the instrument. It also contains
information about required supplies and equipment.
SELECTRA PRO
M 6003-400-450-02 6-1
Maintenance
6.1 Introduction
Info
i The list does not include preventive maintenance that must be performed by
the operator. These tasks are included in the User Manual.
! WARNING
BIOHAZARD
Materials in the analyzer may be infectious. It is recommended to start the
yearly maintenance with the clean system procedure. If this is not feasible,
make sure to take protective measures, such as wearing a face mask and
gloves when performing the maintenance tasks.
SELECTRA PRO
6-2 6003-400-450-02
M
Maintenance
Info
i Only one of the dichromate solutions is used. Both come in 25 ml bottles.
The dichromate 80 Abs solution must be diluted at 1:10 before being used.
SELECTRA PRO
M 6003-400-450-02 6-3
Maintenance
The special yearly preventive maintenace kit for the Selectra ProM contains consumables
that will normally be required and sufficient to perform the yearly maintenance. The two-yearly
maintenance kit contains items that are required for additional two-yearly maintenance tasks.
Info
i Maintenance requirements depend on the workload of the analyzer as well
as on environmental conditions such as ambient temperatures and humidity.
Individual analyzers may require maintenance more or less often than stated
in this documentation.
a) Sufficient for quarterly user maintenance plus yearly maintenance for 1 year.
SELECTRA PRO
6-4 6003-400-450-02
M
Maintenance
! WARNING
BIOHAZARD
Fluids in the waste container are potentially infectious. These fluids must be
handled with great care. Clean up spills immediately. Use applicable
procedures to discard the fluids from the waste container.
! CAUTION
WEAR GLOVES
This caution indicates that gloves must be worn when performing the
described work.
! CAUTION
NOTICE
PLEASE NOTE
i Always use at least ASTM type II grade water.
Info
i If the position xx is shown on the screen, you must first define the rotor
position for the system cleaning solution. See the Reagent Positions
screen.
Info
i Your assistance is required during the entire procedure. The analyzer waits
for confirmation after each step. The procedure takes about 45 minutes.
SELECTRA PRO
M 6003-400-450-02 6-5
Maintenance
Info
i This step can be done while the system is performing its first cycle of the
cleaning procedure. The cleaned and refilled water container will be required
in later steps of the procedure.
If the results from the control tests are not acceptable, perform the following steps:
6 Empty the system
a) Remove the water filter from the water container.
b) Open the Rotor / System menu.
c) Open the Fill / Empty System screen.
d) Touch the Empty System button.
7 Fill the system
a) Make sure the water container is filled with system liquid (a mixture of 1 part system
solution on 400 parts distilled water).
b) Place the water filter in the water container.
c) Open the Rotor / System menu.
d) Open the Fill / Empty screen.
e) Touch the Fill System button.
8 Repeat the controls for all tests
SELECTRA PRO
6-6 6003-400-450-02
M
Maintenance
Info
i The dichromate end test uses known solutions and defined test parameters.
The tests are designed to check the performance of the pipetting system and
measuring unit. Also, the functionality of the stirrer units is tested.
NOTICE
PLEASE NOTE
i The Dichromate 8 Abs solution should not be exposed to direct sunlight.
Place a cap and store in a dark place when not in use.
Test R1 Sample R2
The water bottles must be full. The Dichromate 8 Abs bottle must contain 5 ml.
3 Request control tests
• 20 x the Dichromate 8 Abs control for the SR3 (Check-S) test.
• 20 x the Water control for the SR75 (Check-R) test.
4 Load the control samples
• 1 tube containing 1 ml Dichromate 8 Abs solution.
• 1 tube containing 1 ml distilled water.
5 Run the control tests
6 Check the results
a) Open the Quality Control screen.
b) Check for drop-outs on the Levy-Jennings plot. These indicate air bubbles.
c) Check for high CV on the SR3 test. This indicates a problem with the sample syringe or
the stirrer on the sample side.
d) Check for high CV on the SR75 test. This indicates a problem with the reagent syrings
or the stirrer on the reagent side.
SELECTRA PRO
M 6003-400-450-02 6-7
Maintenance
• Replace the stirrer belts and/or stirrer bearings to resolve stirring problems.
8 Repeat the dichromate end test
SELECTRA PRO
6-8 6003-400-450-02
M
Maintenance
Info
i Due to air bubbles in the tubing it may be necessary to repeat this procedure
a couple of times after installation of the analyzer.
SELECTRA PRO
M 6003-400-450-02 6-9
Maintenance
Info
i The reagent neelde rinse place has one tube connected to it; the sample
needle rinse place has two tubes connected to it.
Remove the four screws that fasten the measuring unit to the base plate. One screw is
located behind the support beam.
SELECTRA PRO
6-10 6003-400-450-02
M
Maintenance
! CAUTION
TAKE CARE
Avoid hitting the wash arm, stirrer unit or any other components while lifting
out the measuring unit.
SELECTRA PRO
M 6003-400-450-02 6-11
Maintenance
SELECTRA PRO
6-12 6003-400-450-02
M
Maintenance
Info
i It may be easier to do this when the bellows pump unit is lifted and partly
pulled out of the analyzer.
SELECTRA PRO
M 6003-400-450-02 6-13
Maintenance
Info
i The figure shows the adjustment of the sample rotor. Adjusting the reagent
rotor works in the same way, with a reagent bottle instead of a sample tube.
Preparations:
• Remove the work table to get access to sample and reagent rotor.
• Remove parts of the cooling unit to get access to the reagent rotor.
Info
i The reagent bottle has some radial play in the rotor. Make sure the reagent
bottle is centered when checking the reagent needle position.
SELECTRA PRO
6-14 6003-400-450-02
M
Maintenance
SELECTRA PRO
M 6003-400-450-02 6-15
Maintenance
Info
i The encoder optos are glued on the encoder opto board. Adjustment of the
entire encoder opto board only needs to be done when the bracket screws
were loosened.
Info
i The screen update takes some time. After each step, wait one second to
check the changes in the encoder signals.
SELECTRA PRO
6-16 6003-400-450-02
M
Maintenance
Info
i The temperature should be 10ºC (+/- 1) or at least 15ºC below the ambient
temperature (if the ambient temperature is above 25ºC).
SELECTRA PRO
M 6003-400-450-02 6-17
Maintenance
SELECTRA PRO
6-18 6003-400-450-02
M
Maintenance
Info
i The location of the mark depends on the viewing direction. Make sure you
are marking the position in a line that goes through the center of the rotor
and of the sample position.
SELECTRA PRO
M 6003-400-450-02 6-19
Maintenance
Info
i For clarity, the above picture does not show the sample rotor drive and
various other components. It is not required to remove them.
Preparations:
• Remove the sample rotor.
• Remove the working table.
• Check the sample rotor positioning. Adjust if required. See par. 6.3.1.
SELECTRA PRO
6-20 6003-400-450-02
M
Maintenance
Info
i Checking the light beam must be done from the rear of the analyzer, as the
light travels from the rear to the front. The pictures in the following steps
show the adjustment jig as seen from the rear.
SELECTRA PRO
M 6003-400-450-02 6-21
Maintenance
Info
i Each adjustment influences the other setitngs as well. Repeat previous
steps as required.
SELECTRA PRO
6-22 6003-400-450-02
M
Maintenance
NOTICE
SELECTRA PRO
M 6003-400-450-02 6-23
Maintenance
NOTICE
a) Remove the valve block from the pipettor unit and place it upside down on a clean
surface.
b) Carefully remove the old gasket [1].
c) Make sure the bush [2] is clean.
d) Place a new gasket into the bush.
e) Push the gasket down with your finger.
f) Place the valve block on the pipettor unit.
6 Place the syringes
a) Push the plunger into the syringe a little.
b) Push the plunger handle [4] of the syringe over the drive pin [6].
c) Fasten the screw [5].
d) Pull the glass barrel [3] up.
e) Screw the end cap [2] into the valve block [1].
7 Refill the system
a) Press the Reset System button on the screen.
b) Open the Fill / Empty System screen.
c) Press the Fill System button.
d) Check for air bubbles in the tubing.
If air bubbles are visible, repeat the Fill System command.
SELECTRA PRO
6-24 6003-400-450-02
M
Maintenance
NOTICE
Info
i The plunger of the sample syringe has two screws fitted on either side of the
plunger handle.
Info
i The plunger of the sample syringe has two screws fitted on either side of the
plunger handle.
SELECTRA PRO
M 6003-400-450-02 6-25
Maintenance
Info
i Replace one valve block at a time. This reduces the change of errors when
reconnecting the tubes.
SELECTRA PRO
6-26 6003-400-450-02
M
Maintenance
SELECTRA PRO
M 6003-400-450-02 6-27
Maintenance
SELECTRA PRO
6-28 6003-400-450-02
M
Maintenance
NOTICE
PLEASE NOTE
i If the drive nuts do not move smoothly, the measurement results may be
incorrect due to incorrect volumes being pipetted. The same is true for
driving nuts that have too much play on the spindle. These should be
replaced.
SELECTRA PRO
M 6003-400-450-02 6-29
Maintenance
Info
i This procedure is for small horizontal adjustments. If this is not sufficient, the
whole arm must be readjusted. See par. 6.4.7.
SELECTRA PRO
6-30 6003-400-450-02
M
Maintenance
Info
i This adjustment is only required when repairs were done to the arm
mechanism.
SELECTRA PRO
M 6003-400-450-02 6-31
Maintenance
The clamp bracket is fitted with a set screw that pushes the two bearings away from each
other. Turn the set screw clockwise to make the vertical arm movement smoother. Turn
the set screw counterclockwise if the bearings show play on the guide shaft. Make small
adjustments and check until the movement is solid and smooth.
3 Move the arm to the wash position (only horizontally)
Use the buttons in the service menu to move the arm to the wash position horizontally.
4 Move the clamp ring down
a) Loosen the screw in the clamp ring [4].
b) Lower the clamp ring over a short distance.
c) Fasten the screw.
5 Adjust the vertical arm position
a) Remove the clamp screw [8].
b) Apply some locktite 243 to the clamp screw.
c) Mount the clamp screw, but do not tighten it.
d) Move the arm shaft [2] so that the distance between the bottom of the arm [1] and the
top of the mechanical assembly is exactly 97 mm.
e) Make sure that the needle is centered on the wash position.
f) Make sure that the block [3], washers [9] and clamp bracket [7] rest against each other.
g) Make sure the distance between the clamp bracket [7] and the swing unit bracket [6] is
kept at 0.5 mm.
h) Tighten the clamp screw [8].
Info
i To ease the vertical arm adjustment, a mechanical adjustment tool is
available from Vital Scientific (part number 4999-001).
SELECTRA PRO
6-32 6003-400-450-02
M
Maintenance
SELECTRA PRO
M 6003-400-450-02 6-33
Maintenance
Info
i Pull the belt off the pulley in the back first. If it slips out of the tweezers, it will
not be catapulted into the analyzer.
Info
i Place the belt around the eccentric pulley in the front first. If it slips out of the
tweezers, it will not be catapulted into the analyzer.
SELECTRA PRO
6-34 6003-400-450-02
M
Maintenance
SELECTRA PRO
M 6003-400-450-02 6-35
Maintenance
SELECTRA PRO
6-36 6003-400-450-02
M
Maintenance
To optimize the amount of light through the cuvette, the lamp must be correctly positioned.
The lamp is placed on a bracket, which can be adjusted vertically and horizontally via
eccentric adjustment screws. A fastening screw holds the lamp unit in position.
SELECTRA PRO
M 6003-400-450-02 6-37
Maintenance
! CAUTION
TAKE CARE
Looking into the lamp may hurt your eyes. Avoid looking into the lamp while
it is switched on.
SELECTRA PRO
6-38 6003-400-450-02
M
Maintenance
! WARNING
HOT SURFACE
The lamp gets very hot when switched on. Make sure the lamp is cooled
down before removing it from the analyzer.
! CAUTION
WEAR GLOVES
Quartz-iodine lamps are very sensitive. Touching the lamp with your hands
will dramatically reduce its lifetime. Wear gloves when placing a new lamp.
SELECTRA PRO
M 6003-400-450-02 6-39
Maintenance
Info
i To get easier access to the lamp and the lamp holder, remove the mounting
screw [3]. The mounting bracket can be pulled out of the instrument. Take
care to not unplug any cables when doing so.
SELECTRA PRO
6-40 6003-400-450-02
M
Maintenance
For correct measurements, the photo sensor in the cuvette stator must be perfectly aligned
with the light beam. The cuvette stator that holds the photo sensor can be rotated over a very
short distance.
SELECTRA PRO
M 6003-400-450-02 6-41
Maintenance
Info
i Fully transparent tape is not useful for this procedure, as you will not be able
to see the light beam on it. Semi-transparent tape is required to see both the
light beam and the hole behind the tape, as shown in the inset.
Info
i Green light is used because it offers the highest visibility. If visibility of
another color is better for you, you can use another filter wheel position.
SELECTRA PRO
6-42 6003-400-450-02
M
Maintenance
Info
i If the MCB must be replaced, the change to the hexaswitch setting must also
be done on the new MCB.
Finishing work:
• Remove the tape from the cuvette rotor base.
• Place the cuvette rotor.
SELECTRA PRO
M 6003-400-450-02 6-43
Maintenance
Info
i The vertical stop position of the wash arm is determined by the optical
switches on the wash arm board. These detect the position vane that moves
up and down with the wash arm. If the vertical stop position of the wash arm
is incorrect, the wash arm board with the optical switches must be moved
vertically to change the detection moments.
SELECTRA PRO
6-44 6003-400-450-02
M
Maintenance
The difference between the two measured distances must be 0.5 +/- 0.1 mm. If this is not the
case, perform the following step.
5 Adjust the wash arm board position
a) Remove the working table.
b) Loosen the locknut [4]. To do this, insert a 5.5 mm spanner via the slit of the upper plate
of the mechanical assembly.
c) Turn the set screw [3] to adjust the position. Turn the screw clockwise if the measured
difference is too small. When turning the screw counterclockwise, make sure to press
down on the screw to prevent it from moving up, instead of moving the wash arm board.
One full turn moves the wash arm board by 0.5 mm.
d) Tighten the locknut [4].
When the wash arm board position was adjusted, perform the entire checking procedure again.
Finishing work:
• Mount all removed parts in reversed order.
SELECTRA PRO
M 6003-400-450-02 6-45
Maintenance
SELECTRA PRO
6-46 6003-400-450-02
M
Maintenance
Info
i When the wash arm is reset the wash arm moves down as far as it goes. If
a cuvette rotor is present, this will stop the downward movement when the
position vane [5] is still visible for the middle opto switch [4] on the wash arm
board. If no cuvette rotor is present, the wash arm moves down further until
the middle opto switch does not detect the position vane anymore. When this
happens, a “cuvette rotor not present” error is generated.
Preparations:
• Adjust the wash arm. See par. 6.5.5.
SELECTRA PRO
M 6003-400-450-02 6-47
Maintenance
Supplies:
• 6003-127: Peltier replacement kit
• 2201-122: Thermal joint compound
a) Disconnect the electrical cables from the bottom of the cover heating connectors [8].
b) Remove the four screws [1].
c) Lift the top section [7] off the measuring unit.
3 Remove the lamp bracket
a) Loosen the screw [6].
b) Remove the lamp bracket [5].
4 Remove the filter wheel
a) Remove the small screw [4].
b) Remove the filter wheel axle [3].
c) Lift the filter wheel [2] out of its housing.
SELECTRA PRO
6-48 6003-400-450-02
M
Maintenance
SELECTRA PRO
M 6003-400-450-02 6-49
Maintenance
SELECTRA PRO
6-50 6003-400-450-02
M
Maintenance
! CAUTION
TAKE CARE
The Peltier elements unit should not be placed upside down. Make sure the
colors of the wires [1] are as shown. The black wire should be closest to the
side where the filter wheel will be.
SELECTRA PRO
M 6003-400-450-02 6-51
Maintenance
! CAUTION
TAKE CARE
Use the new washers and spring washers from the Peltier replacement kit.
Do not over-tighten the screws: this will damage the Peltier elements.
SELECTRA PRO
6-52 6003-400-450-02
M
Maintenance
SELECTRA PRO
M 6003-400-450-02 6-53
Maintenance
SELECTRA PRO
6-54 6003-400-450-02
M
Maintenance
SELECTRA PRO
M 6003-400-450-02 6-55
Maintenance
SELECTRA PRO
6-56 6003-400-450-02
M
Maintenance
SELECTRA PRO
M 6003-400-450-02 6-57
Maintenance
SELECTRA PRO
6-58 6003-400-450-02
M
Maintenance
Info
i The cooling pump has one extra part that is not shown in the above picture.
This is a foam disc that is placed inside the cover [9]. The instructions below
are valid for both the water and cooling pump.
SELECTRA PRO
M 6003-400-450-02 6-59
Maintenance
Info
i When overhauling the cooling pump, do not forget to place the foam disc
inside the cover [9].
SELECTRA PRO
6-60 6003-400-450-02
M
Maintenance
If the needles are too deep in the water, perform the following step:
2 Change the pump stroke
a) Loosen the two screws [3].
b) Move the driving beam [2] over a short distance.
c) Fasten the two screws [3].
Info
i To increase the stroke, move the driving beam to the right.
To decrease the stroke, move the driving beam to the left.
SELECTRA PRO
M 6003-400-450-02 6-61
Maintenance
Info
i The power supply does not have adjustable or servicable parts. If the power
supply is the cause of the incorrect voltages, it should be replaced.
SELECTRA PRO
6-62 6003-400-450-02
M
Maintenance
Info
i If the difference in the two measurements is not within the specified range,
the photometer board must be replaced.
SELECTRA PRO
M 6003-400-450-02 6-63
Maintenance
Info
i If the difference in the two measurements is not within the specified range,
the photometer board must be replaced.
SELECTRA PRO
6-64 6003-400-450-02
M
Maintenance
Info
i When the power is switched on, the display flashes all segments and then
shows the version of the control software stored in the EPROM. During
normal operation, if no errors have occurred, only three dots are lit (as in the
above picture).
Info
i Possibly, more than one error condition holds. Press the UP button to let the
temperature control board show all current error codes. Press the DOWN
button to move down the list of current errors one by one.
SELECTRA PRO
M 6003-400-450-02 6-65
Maintenance
Info
i The list of states and their possible ranges is included in the description of
the temperature control board in the Electronics chapter. See par. 6.7.4.
Info
i The list of parameters and their valid ranges is included in the description of
the temperature control board in the Electronics chapter. See par. 6.7.4.
SELECTRA PRO
6-66 6003-400-450-02
M
Maintenance
Status codes
The three-digit display on the temperature control board is used to signal error conditions,
show the current state of various signals and set parameter values. The error codes are
listed in par. 7.1.9 The status codes are shown in the table below. The next subsection lists
the available parameters.
Info
i The cuvette temperature setpoint is corrected for the current ambient
temperature. This implies that a different current cuvette temperature does
not always imply an error condition.
SELECTRA PRO
M 6003-400-450-02 6-67
Maintenance
Parameters list
The three-digit display on the temperature control board is used to signal error conditions,
show the current state of various signals and set parameter values. The error codes are
listed in par. 7.1.9 The status codes are shown in the previous subsection. The table below
lists the available parameters.
* Instrument type selection is required when the temperature control board is switched on
for the first time. The required codes are shown in the following list:
1) Selectra ProXS, Selectra ProS, Flexor EL80, Flexor EL150
2) E series, Selectra ProM, Flexor EL200
3) XL series
Select the required code using the UP and DOWN buttons. Press the MODE button when
the correct value is shown. The display shows ACC when the setting is accepted.
** When the cuvette heater is disabled (P07 is set to 01), the heater is not switched on and
no error is sent to the system board when the temperature is too far off the setpoint.
SELECTRA PRO
6-68 6003-400-450-02
M
Maintenance
Info
i If the sample needle stops too early (above the water level), the sensitivity
is too high. If the sample needle stops too late (too deep in the water), the
sensitivity is too low.
SELECTRA PRO
M 6003-400-450-02 6-69
Maintenance
Info
i If the reagent needle stops too early (above the water level), the sensitivity
is too high. If the reagent needle stops too late (too deep in the water), the
sensitivity is too low.
SELECTRA PRO
6-70 6003-400-450-02
M
7
Troubleshooting
This chapter lists all error codes that may appear in the analyzer software, with possible
causes and remedies. A section on troubleshooting is also included.
SELECTRA PRO
M 6003-400-450-02 7-1
Troubleshooting
When a test-related error condition is found, the measurement result is flagged. These flags
are also included on printed test results. A list of flags and their meanings is included in the
User Manual.
When a hardware error occurs, it is signalled to the operator in a dialog window. All errors
have a specific code and are logged. The error history lists all logged error messages.
Info
i A hardware error often leads to subsequent errors. All errors are listed in the
error history. Start checking the cause for the first error that occured at a
specific time. Solving that cause may automatically resolve all subsequent
errors.
SELECTRA PRO
7-2 6003-400-450-02
M
Troubleshooting
E10 No Vacuum
The vacuum is below the defined minimum too long.
Possible causes:
• Defective or worn components in the vacuum system.
SELECTRA PRO
M 6003-400-450-02 7-3
Troubleshooting
SELECTRA PRO
7-4 6003-400-450-02
M
Troubleshooting
• Ambient temperature outside the specified working range. See par. A.1.3.
Actions:
• Check the error code on the TCB. See par. 7.1.9.
SELECTRA PRO
M 6003-400-450-02 7-5
Troubleshooting
SELECTRA PRO
7-6 6003-400-450-02
M
Troubleshooting
SELECTRA PRO
M 6003-400-450-02 7-7
Troubleshooting
SELECTRA PRO
7-8 6003-400-450-02
M
Troubleshooting
SELECTRA PRO
M 6003-400-450-02 7-9
Troubleshooting
SELECTRA PRO
7-10 6003-400-450-02
M
Troubleshooting
SELECTRA PRO
M 6003-400-450-02 7-11
Troubleshooting
SELECTRA PRO
7-12 6003-400-450-02
M
Troubleshooting
SELECTRA PRO
M 6003-400-450-02 7-13
Troubleshooting
E125 No Cuvettes
The wash arm was not stopped by a cuvette rotor when moving down.
Possible causes:
• No cuvette rotor present.
• Incorrect positioning of the wash arm.
• Malfunctioning optical switches on the wash arm board.
Actions:
• Place a cuvette rotor.
• Check the wash arm position. See par. 6.5.5.
SELECTRA PRO
7-14 6003-400-450-02
M
Troubleshooting
• Check the optical switches. Replace the wash arm board if required.
SELECTRA PRO
M 6003-400-450-02 7-15
Troubleshooting
SELECTRA PRO
7-16 6003-400-450-02
M
Troubleshooting
Possible causes:
• Missing ISE reference solution.
Actions:
• Make sure there is sufficient ISE reference solution in the bottle.
SELECTRA PRO
M 6003-400-450-02 7-17
Troubleshooting
SELECTRA PRO
7-18 6003-400-450-02
M
Troubleshooting
SELECTRA PRO
M 6003-400-450-02 7-19
Troubleshooting
• Disconnect PL2 and measure the resistance over pins 3 and 4 of the wire
connector. If the resistance is not between 2 and 10 Ohm, check the Peltier
elements and wiring. Replace if required.
• Disconnect PL5 and measure the resistance between pins 1 and 3 on the wire
connector. If the resistance is not around 40 kOhms, replace the temperature
sensor.
• Disconnect PL5 and measure the resistance between pins 1 and 4, 2 and 4, 3 and 4
of the wire connector. If these resistances are not all very high, check the wiring and
temperature sensor for short circuits. Replace wiring and/or temperature sensor if
required.
• If all of the above measurements were within the specified ranges, replace the TCB.
SELECTRA PRO
7-20 6003-400-450-02
M
Troubleshooting
E05 short-short-long-short-long
The CU board receives no samples.
Possible causes:
• Malfunctioning A/D converter.
Actions:
• Replace the cooling unit board.
E06 short-short-long-long-short
The 12V input supply line is too low.
Possible causes:
• Bad connection to the power supply.
• Malfunctioning power supply.
Actions:
• Check the power cables.
• Check the power supply.
E09 short-long-short-short-long
Short circuit in the cooling pump or fan.
Possible causes:
• Short circuit in the cooling pump or wiring.
• Short circuit in the fan or wiring.
Actions:
• Check the wiring and replace if required.
• Check the cooling pump.
• Check the cooling fan.
E10 short-long-short-long-short
The cooling pump and/or fan are not detected during startup.
Possible causes:
• Malfunctioning cooling pump.
• Malfunctioning fan.
• Bad connections to cooling pump and/or fan.
Actions:
• Check all power connections.
• Check cooling pump and fan.
E11 short-long-short-long-long
The unit is not cooling.
Possible causes:
• Bad connections to the Peltier elements.
• Peltier elements are connected the wrong way around.
• Malfunctioning Peltier elements.
Actions:
• Check the connection to the Peltier elements.
• Correct the orientation of the Peltier elements.
SELECTRA PRO
M 6003-400-450-02 7-21
Troubleshooting
• Check the resistance of the Peltier elements. If the resistance is high, replace the
Peltier element.
E13 short-long-long-short-long
No feedback received from the Peltier elements.
Possible causes:
• Malfunctioning voltage regulator circuit.
• Malfunctioning Peltier elements.
• Bad connection to Peltier elements.
Actions:
• Check connections to Peltier elements.
• Check if Peltier elements with an external power supply.
• If the Peltier elements and wiring are working, replace the CU board.
E14 short-long-long-long-short
Temperature sensor does not give a signal.
Possible causes:
• Malfunctioning temperature sensor.
Actions:
• Replace the cooling unit board.
E21 long-short-long-short-long
The main loop in the program is taking too long.
Possible causes:
• Malfunctioning components on the cooling unit board.
Actions:
• Replace the cooling unit board.
E22 long-short-long-long-short
The program has encountered a stack overflow.
Possible causes:
• Malfunctioning components on the cooling unit board.
Actions:
• Replace the cooling unit board.
E23 long-short-long-long-long
The watchdog function did not receive a signal for too long.
Possible causes:
• The program is stuck.
Actions:
• Reset the CU board by interrupting the power.
• If the problem reappears, replace the CU board.
E25 long-long-short-short-long
The microcontroller does not receive a correct clock signal.
Possible causes:
• Malfunctioning crystal.
SELECTRA PRO
7-22 6003-400-450-02
M
Troubleshooting
Actions:
• Replace the cooling unit board.
E26 long-long-short-long-short
The microcontroller cannot read or write the flash memory.
Possible causes:
• Malfunctioning flash memory.
Actions:
• Replace the cooling unit board.
E27 long-long-short-long-long
The CU board has interrupted its program.
Possible causes:
• An external reset signal was received.
Actions:
• Check the reason for the external reset signal.
• Reset the CU board by interrupting the power.
SELECTRA PRO
M 6003-400-450-02 7-23
Troubleshooting
7.2 Troubleshooting
! WARNING
BIOHAZARD
Samples, controls, calibrators and liquid waste are potentially infectious.
Handle all materials according to national and international laboratory safety
regulations.
! CAUTION
WEAR GLOVES
Gloves are required when working on tubing, syringes, rotors and/or waste
containers.
SELECTRA PRO
7-24 6003-400-450-02
M
Troubleshooting
The following sequence indicates the order in which the control system is initialized at startup.
This may be helpful in finding causes of startup problems.
Info
i If errors occur during the system initialization phase, subsequent errors may
occur. Use the error history to find the first error that occurred during the
initialization. Solve that problem and restart the analyzer.
SELECTRA PRO
M 6003-400-450-02 7-25
Troubleshooting
SELECTRA PRO
7-26 6003-400-450-02
M
8
Parts
This chapter contains exploded views and parts lists for the components of the instrument.
It also contains standard lists of spare parts.
SELECTRA PRO
M 6003-400-450-02 8-1
Parts
8.1 Overview
SELECTRA PRO
8-2 6003-400-450-02
M
Parts
a) 6003-904 - Water container assy (including water filter, tubing and cap)
b) 6003-282 - Diluted waste container assy (including float switch, tubing and cap)
c) 6003-395 - Concentrated waste container assy (including float switch, tubing and
cap). This part is only applicable if the ECO-kit option is installed.
SELECTRA PRO
M 6003-400-450-02 8-3
Parts
4 4
2
3
5
10
6 8
7 9
1
1
SELECTRA PRO
8-4 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-5
Parts
26
27
27
15
31
29
30
28
25
13
22
24
20
21
23
13
13
17
16
18
15
13
3
19
2
5
6
13
1
12
8
9
14
11
10
15
SELECTRA PRO
8-6 6003-400-450-02
M
Parts
1 4102-324 SHAFT
2 4802-245 PLATE
3 4102-355 SHAFT
4 3130-077 SPROCKET
5 3053-035 BEARING, BALL 30 X 55 X 13 MM
6 3053-001 BEARING, BALL 6 X 19 X 6 MM
7 4105-187 BUSH
8 3350-023 CONN. 2PM FASTON TAB 6,3 MM
9 4105-186 BUSH
10 3022-025 WASHER 6.4 ST.ST. DIN 125A
11 3002-388 SCREW M6 X 25 ST.ST. DIN 912
12 3023-006 WASHER,LOCK J 3.2 DIN 6798
13 3002-016 SCREW M3 X 6 ST.ST. DIN 7985
14 3002-015 SCREW M3 X 4 ST.ST. DIN 7985
15 3022-015 WASHER 4.3 ST.ST. DIN 125A
16 3002-030 SCREW M4 X 6 ST.ST. DIN 7985
17 3022-020 WASHER M5 ST.ST. DIN 125A
18 3002-336 SCREW M4 X 16 ST.ST. DIN 912
19 4802-250 VANE
20 3064-053 TIMING BELT (KEVLAR) 4T2.5/330
21 6001-930 OPTICAL SWITCH 1.5V/7.5MA
22 4803-472 BRACKET
23 4803-383 BRACKET
24 3381-053 OPTICAL SWITCH 1.5V/7.5MA
25 6002-572 ARM MOTOR UNIT HOR./VERT.
26 3019-049 NUT M4 X 27.5 MM
27 3002-333 SCREW M4 X 12 ST.ST. DIN 912
28 3008-040 SCREW M4 X 5 ST.ST. DIN 913
29 3130-076 SPROCKET T2,5/15
30 3004-204 SCREW M2 X 6 ST.ST DIN 965
31 4700-543 DISC, ENCODER
SELECTRA PRO
M 6003-400-450-02 8-7
Parts
2 16
1 6
4 17
3 18
9
8
7
24
12 26
25
13 27
10 28
19
14 29
15 22 28
20
30
31
32
11 23
21 33
22
34
SELECTRA PRO
8-8 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-9
Parts
24
24
5
25
1
24
3
18
19
6
20
7
14
13
12
2
23
16
4
15
17
22
21
8
9
10
11
27
SELECTRA PRO
8-10 6003-400-450-02
M
Parts
1 4501-512 BLOCK
2 4802-330 PLATE
3 4803-450 BRACKET
4 4105-238 BUSH
5 4509-162 STRIP
6 6001-773 STIRRER DRIVER BOARD
7 4509-167 PLATE
8 3363-625 MOTOR 18V DC BALLBEARING
9 4129-019 PULLEY
10 3053-038 BEARING, BALL 7 X 14 X 5 MM
11 4129-021 PULLEY
12 3053-020 BEARING, BALL 8 X 16 X 5 MM
13 4129-018 PULLEY
14 3064-041 DRIVINGBELT MIXER
15 3019-004 NUT M3 BRASS (HEIGHT 25 MM)
16 3009-048 SCREW, KNURLED TUMB LOW
17 3008-038 SCREW M3 X 16 ST.ST. DIN 913
18 3053-005 BALL 5 MM BALLBEARINGSTEEL
19 3750-043 SPRING
20 3008-125 SCREW M6 X 5 ST.ST. DIN 551
21 3002-122 SCREW M1.6 X 3 ST.ST. DIN 84
22 3002-002 SCREW M2 X 5 ST.ST. DIN 7985
23 3004-204 SCREW M2 X 6 ST.ST DIN 965
24 3002-016 SCREW M3 X 6 ST.ST. DIN 7985
25 3023-006 WASHER,LOCK J 3.2 DIN 6798
26 3019-013 NUT M3 INSERT ST. (1.8)
27 3008-020 SCREW M3 X 3 ST.ST. DIN 913
SELECTRA PRO
M 6003-400-450-02 8-11
Parts
5
2
4
3
14
13
12
7
10 15
11
9 16
17
21
24
23
19
6 20 25 18
12
14
22
SELECTRA PRO
8-12 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-13
Parts
13
12
1 17
16
10
15
14
11
18
9
8
7
3
6
3
1
5
2
1
SELECTRA PRO
8-14 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-15
Parts
26
18
12
23
11
7
17
29
22
16
25
28
15
24
27
14
6
21
3
13
5
20
11
19
12
10
3
4
1
SELECTRA PRO
8-16 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-17
Parts
4
3
5
2
29
9
28
26
30 27
6 10
7 11
8
7
12 23
22
20
25
24
1 21
3
18
19
15 17
14 16
13 10
SELECTRA PRO
8-18 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-19
Parts
7 12
11
14
15
13
9
25 15
26
17 10
2
16
8
24
25
7
6
3 20
22
18
4 21
19
22
5 23
15
SELECTRA PRO
8-20 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-21
Parts
20
29
20
30
2
26
32
27
31
24
10
28
5
25
22
17
21
19
23
20
18
11
10
9
8
3
1
12
14
7
15
16
13
6
5
4
SELECTRA PRO
8-22 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-23
Parts
8.4 Pipettor
25
24
26
25
23
20
17
19
18
18
22
19
14
16
13
21
20
15
1
20
27
5
2
8
3
10, 11
6
12
9
7
SELECTRA PRO
8-24 6003-400-450-02
M
Parts
1 4509-169 PLATE
2 4800-331 FRONTPLATE DISPENSER
3 3002-018 SCREW M3 X 8 ST.ST. DIN 7985
4 3002-032 SCREW M4 X 35 ST.ST. DIN 7985
5 3002-016 SCREW M3 X 6 ST.ST. DIN 7985
6 3066-072 VALVE BLOCK
7 4107-079 SCREW
8 4105-215 BUSH
9 4107-065 KNOB FOR VALVE BLOCK
10 3066-155 SYRINGE LONG LIFE 100 μL a)
11 3066-156 SYRINGE LONG LIFE 1 ML a)
12 4107-066 KNOB FOR SYRINGE PLUNGER
13 4102-306 PIN, DRIVE
14 4501-424 CARRIAGE
15 3002-336 SCREW M4 X 16 ST.ST. DIN 912
16 3022-015 WASHER 4.3 ST.ST. DIN 125A
17 4100-411 DRIVE NUT
18 3053-001 BEARING, BALL 6 X 19 X 6 MM
19 3130-070 SPROCKET T2.5/25
20 3008-040 SCREW M4 X 5 ST.ST. DIN 913
21 3020-017 CIRCLIP 4 DIN 6799
22 4102-305 SHAFT, GUIDE
23 4802-227 VANE ON CARRIAGE
24 4102-304 SHAFT, GUIDE
25 4100-410 BUSH FOR GUIDE SHAFT
26 3750-033 SPRING
SELECTRA PRO
M 6003-400-450-02 8-25
Parts
15
3
16
1
17
7
6
4
5
11
7
13
12
8
13
14
10
9
SELECTRA PRO
8-26 6003-400-450-02
M
Parts
1 4509-169 PLATE
2 6001-459 DISPENSER OPTO BOARD
3 3004-211 SCREW M3 X 8 ST.ST. DIN 965
4 3019-013 NUT M3 INSERT ST. (1.8)
5 6001-769 DISPENSER DRIVER BOARD
6 3381-053 OPTICAL SWITCH 1.5V/7.5MA 6001-769
7 3002-018 SCREW M3 X 8 ST.ST. DIN 7985
8 6001-813 MOTOR, STEPPER 15 DEG. 5V/2.78A
9 3002-026 SCREW M4 X 8 ST ST. DIN 7985
10 3022-015 WASHER 4.3 ST.ST. DIN 125A
11 3064-016 TIMINGBELT 80T/T2.5/200
12 3130-069 SPROCKET T2,5/24
13 3008-040 SCREW M4 X 5 ST.ST. DIN 913
14 6001-780 VALVE DRIVE UNIT par. 8.4.3
15 3002-029 SCREW M4 X 16 ST.ST. DIN 7985
16 3023-010 WASHER,LOCK,ST.ST.M4 DIN 6798
17 3019-048 NUT M3 X 45
SELECTRA PRO
M 6003-400-450-02 8-27
Parts
21
34
9
29
27
28
20
26
31
25
33
32
23
24
30
3
18
22
16
17
1
16
8
19
3
14
2
13
4
15
5
4
6
8
9
7
10
12
11
SELECTRA PRO
8-28 6003-400-450-02
M
Parts
1 4501-519 BLOCK
2 4102-303 SHAFT
3 3020-017 CIRCLIP 4 DIN 6799
4 3053-019 BEARING, BALL 5 X 11 X 4 MM
5 4100-409 BUSH
6 3022-018 SHIM RING ST.ST. PS5 X 7 X 0.1 MM
7 4501-422 BLOCK
8 3008-040 SCREW M4 X 5 ST.ST. DIN 913
9 3022-010 WASHER 3.2 ST.ST. DIN 125A
10 4803-297 STRIP
11 3002-019 SCREW M3 X 10 ST.ST. DIN 7985
12 4102-269 SHAFT
13 3130-062 GEAR
14 4106-203 DISC
15 3004-205 SCREW M2 X 8 ST.ST. DIN 965
16 3053-001 BEARING, BALL 6 X 19 X 6 MM
17 4100-408 BUSH
18 4102-302 SHAFT
19 3022-023 SHIM RING ST.ST. PS6 X 8 X 0,15 MM
20 4501-518 BLOCK (MOTOR UNIT)
21 3002-023 SCREW M3 X 20 ST.ST. DIN 7985
22 3008-041 SCREW M4 X 5 ST.ST. DIN 914
23 3002-016 SCREW M3 X 6 ST.ST. DIN 7985
24 3011-011 NUT M3 ST.ST. DIN 934
25 3002-001 SCREW M2 X 4 ST.ST. DIN 7985
26 3022-005 WASHER ø 2.2 ST.ST. DIN 988
27 3053-002 BEARING, BALL 3 X 10 X 4 MM
28 4102-243 WORM (MOTOR ASSY'S)
29 3020-011 CIRCLIP 2.3 DIN 6799
30 3002-003 SCREW M2 X 6 ST.ST. DIN 7985
31 1513-050 TUBE, PVC 2 X 4 MM CLEAR a)
32 4802-329 PLATE
33 3363-620 MOTOR, 12VDC 20MA 7800 RPM
34 3002-307 SCREW M3 X 8 ST.ST. DIN 912
SELECTRA PRO
M 6003-400-450-02 8-29
Parts
4
5
3
6
1
6
2
7
5
4
SELECTRA PRO
8-30 6003-400-450-02
M
Parts
1 4102-329 SHAFT
2 6002-004 REAGENT HEAD par. 8.5.4
3 6002-003 SAMPLE HEAD par. 8.5.5
4 n.a. Vertical drive par. 8.5.2
5 n.a. Horizontal drive (swing) par. 8.5.3
6 n.a. Home sensor for horizontal drive par. 8.5.3
7 n.a. Encoder board for horizontal drive par. 8.5.3
SELECTRA PRO
M 6003-400-450-02 8-31
Parts
27
28
7
22
26
24
23
21
25
19
20
18
16
15
18
17
7
6
12
9
11
12
5
14
12
13
10
1
3
SELECTRA PRO
8-32 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-33
Parts
10
25
24
10
26
4
23
4
7
22
10
20
17
18
14
15
16
14
21
19
4
5
11
12
2
9
6
1
13
10
7
8
3
SELECTRA PRO
8-34 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-35
Parts
4
31
17
8 15
16
7 9
19 14
6 13
3 12
20 18
27
10
5
30
29 2
25 21
26 24
28 11
22
23
30
SELECTRA PRO
8-36 6003-400-450-02
M
Parts
1 3070-579 CAP
2 4509-183 BRACKET
3 4100-459 SCREW
4 3004-204 SCREW M2 X 6 ST.ST DIN 965
5 4106-192 RING
6 4100-517 PROP
7 3004-229 SCREW M4 X 16 ST.ST. DIN 965
8 3002-019 SCREW M3 X 10 ST.ST. DIN 7985
9 3022-010 WASHER 3.2 ST.ST. DIN 125A
10 6002-939 NEEDLE UNIT
11 3053-037 BEARING, BALL 2.5 X 6 X 2.6 MM
12 3750-056 SPRING COMPRESSION
13 6003-081 SOLDERTAG
14 4107-092 SCREW
15 4106-278 RING
16 3008-021 SCREW M3 X 4 ST.ST. DIN 916
17 3374-244 LLD WIRE ASSY REAGENT ARM
18 3352-015 CONN. 6PF HOUSING
19 3355-004 PLUG,G2033 RETENTION
20 6001-995 ARM BOARD WITH OPTO
21 4105-302 BUSH
22 3002-002 SCREW M2 X 5 ST.ST. DIN 7985
23 3002-003 SCREW M2 X 6 ST.ST. DIN 7985
24 4509-181 BLOCK
25 4509-179 BLOCK
26 3750-058 SPRING COMPRESSION
27 4802-358 PLATE
28 3030-009 DOWEL PIN 2M6 X 14 DIN 7 AISI
29 3022-005 WASHER ø 2.2 ST.ST. DIN 988
30 3002-001 SCREW M2 X 4 ST.ST. DIN 7985
31 4802-344 SILICON SEALING RING
SELECTRA PRO
M 6003-400-450-02 8-37
Parts
30
16
15
8
14
9 7
18
13
3
6
17 12
19
26 5
20 10
29 2
28
24
25
23
27
21 11
22 29
SELECTRA PRO
8-38 6003-400-450-02
M
Parts
1 3070-579 CAP
2 4509-182 BRACKET
3 4100-459 SCREW
4 3004-204 SCREW M2 X 6 ST.ST DIN 965
5 4106-192 RING
6 4100-517 PROP
7 3004-229 SCREW M4 X 16 ST.ST. DIN 965
8 3002-019 SCREW M3 X 10 ST.ST. DIN 7985
9 3022-010 WASHER 3.2 ST.ST. DIN 125A
10 6002-007 SAMPLE NEEDLE UNIT
11 3053-037 BEARING, BALL 2.5 X 6 X 2.6 MM
12 3750-040 SPRING COMPRESSION
13 6003-080 SOLDERTAG
14 4105-199 SCREW
15 4107-073 RING
16 3374-208 LLD WIRE ASSY SAMPLE ARM
17 3352-015 CONN. 6PF HOUSING
18 3355-004 PLUG,G2033 RETENTION
19 6001-995 ARM BOARD WITH OPTO
20 4105-248 BUSH
21 3002-002 SCREW M2 X 5 ST.ST. DIN 7985
22 3002-003 SCREW M2 X 6 ST.ST. DIN 7985
23 4509-181 BLOCK
24 4509-180 BLOCK
25 3750-058 SPRING COMPRESSION
26 4802-358 PLATE
27 3030-009 DOWEL PIN 2M6 X 14 DIN 7 AISI
28 3022-005 WASHER ø 2.2 ST.ST. DIN 988
29 3002-001 SCREW M2 X 4 ST.ST. DIN 7985
30 4802-344 SILICON SEALING RING
SELECTRA PRO
M 6003-400-450-02 8-39
Parts
22
1
7
24
16 23
8
17
19
18
15
14
3
21
20
10
7 12
11
9
5 4
6
13
SELECTRA PRO
8-40 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-41
Parts
14
13
33
13
28
27
29
31
30
32
17
18
23
21
22
24
24
25
20
16
16
26
19
4
3
7
11
14
13
8
10
15
9
12
6
SELECTRA PRO
8-42 6003-400-450-02
M
Parts
1 4803-653 BRACKET
2 3366-920 VALVE 3-WAY 12VDC/4W
3 3366-920 VALVE 3-WAY 12VDC/4W
4 3366-920 VALVE 3-WAY 12VDC/4W
5 1513-052 TUBE, PVC 1.5 X 3.5 MM CLEAR
6 3066-091 PLUG (PP)
7 3002-309 SCREW M3 X 10 ST.ST. DIN 912
8 3370-053 FEED THROUGH
9 4105-130 BUSH
10 3344-004 SWITCH, VACUUM
11 3002-320 SCREW M3 X 30 ST.ST. DIN 912
12 4803-562 BRACKET
13 1513-051 TUBE, PVC 4 X 6 MM CLEAR
14 3066-087 SOCKET NYLON WIT
15 3374-197 VACUUM SWITCH CABLE
16 3002-019 SCREW M3 X 10 ST.ST. DIN 7985
17 3002-020 SCREW M3 X 12 ST.ST. DIN 7985
18 3105-032 BUSH 3.1X5X5MM
19 3019-013 NUT M3 INSERT ST. (1.8)
20 4803-398 BRACKET
21 4100-478 PROP
22 4100-479 PROP
23 4104-091 PIPE
24 3029-010 O-RING ID=32.92 X d=3.53 MM BUNA
N
25 3370-403 CLIP.CABLE.H2P
26 3066-052 TUBE SCREW CONN. 3/32"/1.8MM
27 3070-018 BUFFER
28 3002-029 SCREW M4 X 16 ST.ST. DIN 7985
29 4802-431 PLATE
30 3029-079 O-RING ID=40MM d=7MM
31 3365-014 PUMP, VACUUM 12VDC 12L/MIN.
(EPDM)
32 3350-138 CONN. 2PF HOUSING MOD 1
33 3066-092 TUBE ELBOW 5.9MM
SELECTRA PRO
M 6003-400-450-02 8-43
Parts
29
30
31
17
28
27
26
33
18
32
10
11
8
24
23
34
22
25
19
20
21
6
17
18
12
7
16
17
14 13
15
3
13 14
4
12
2
12
10
9
5
SELECTRA PRO
8-44 6003-400-450-02
M
Parts
1 4802-253 PLATE
2 4509-122 CAP
3 4501-565 PLATE
4 4802-251 PLATE
5 3002-026 SCREW M4 X 8 ST ST. DIN 7985
6 3106-016 BEARING BUSH
7 3002-020 SCREW M3 X 12 ST.ST. DIN 7985
8 4102-246 PLUG
9 4502-291 BEAM
10 3002-019 SCREW M3 X 10 ST.ST. DIN 7985
11 3022-030 WASHER 3.2 ST.ST. DIN 9021A
12 3022-010 WASHER 3.2 ST.ST. DIN 125A
13 3002-001 SCREW M2 X 4 ST.ST. DIN 7985
14 3022-005 WASHER ø 2.2 ST.ST. DIN 988
15 4802-252 PLATE
16 4100-532 BUSH
17 3008-040 SCREW M4 X 5 ST.ST. DIN 913
18 3053-025 BEARING, BALL 6 X 15 X 5 MM
19 4501-464 BLOCK
20 3381-053 OPTICAL SWITCH 1.5V/7.5MA
21 3002-018 SCREW M3 X 8 ST.ST. DIN 7985
22 4105-192 BUSH
23 3020-020 CIRCLIP 5 DIN 6799
24 4102-352 SHAFT
25 6001-831 MOTOR, BELLOWS
26 4802-254 PLATE
27 3004-212 SCREW M3 X 10 ST.ST. DIN 965
28 3064-018 TIMINGBELT 64T/T2,5/160
29 3130-081 SPROCKET
30 3002-028 SCREW M4 X 12 ST.ST. DIN 7985
31 3022-015 WASHER 4.3 ST.ST. DIN 125A
32 4100-347 BELLOWS
33 3029-039 O-RING ID=24 X d=2 MM BUNA N
34 3019-014 NUT M4 INSERT ST. (1.8)
SELECTRA PRO
M 6003-400-450-02 8-45
Parts
21
13 19
17
14
20
18
11
16
12
15 11
11
11
8
7
10
3
9 2
4 6
SELECTRA PRO
8-46 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-47
Parts
12 13
14
11
10
8
6 5
SELECTRA PRO
8-48 6003-400-450-02
M
Parts
1 4100-527 BUSH
2 4100-558 PROP
3 4100-492 PROP
4 3029-020 O-RING ID=60.05 X d=1.7 MM BUNA
N
5 3002-013 SCREW M2,5 X 10 ST.ST. DIN 7985
6 3004-226 SCREW M4 X 8 ST.ST. DIN 965
7 3066-052 TUBE SCREW CONN. 3/32"/1.8MM
8 3344-015 SWITCH, LIQUID LEVEL N.O. a)
9 3344-015 SWITCH, LIQUID LEVEL N.O. a)
10 3020-027 CIRCLIP 7 DIN 6799 RVS
11 4100-531 BUSH
12 3344-015 SWITCH, LIQUID LEVEL N.O. a)
13 3350-107 CONN. 1PM CRIMP CONTACT
14 3352-019 CONN. 6PF HOUSING
SELECTRA PRO
M 6003-400-450-02 8-49
Parts
16
17
5
10
12
18 15
11 4
13
13
14
SELECTRA PRO
8-50 6003-400-450-02
M
Parts
1 4501-566 BLOCK
2 4501-565 PLATE
3 3366-918 VALVE 2/2 WAY 12VDC/5W
4 4802-454 PLATE
5 3004-019 SCREW M3 X 70 ST.ZINC PLATED
DIN 96
6 3019-013 NUT M3 INSERT ST. (1.8)
7 3029-058 O-RING ID=16 X d=2 MM VITON
8 3066-052 TUBE SCREW CONN. 3/32"/1.8MM
9 3029-060 O-RING ID=7 X d=1 MM VITON
10 6001-733 CONNECTOR ASSY
11 3366-918 VALVE 2/2 WAY 12VDC/5W
12 3350-138 CONN. 2PF HOUSING MOD 1 a)
13 1513-051 TUBE, PVC 4 X 6 MM CLEAR b)
14 3066-092 TUBE ELBOW 5.9MM
15 3002-018 SCREW M3 X 8 ST.ST. DIN 7985
16 3066-139 CONN. 10-32 UNF (PP)
17 4501-465 PLATE
18 3066-090 TUBE SCREW CONN. 10/32"/5.9M
SELECTRA PRO
M 6003-400-450-02 8-51
Parts
16
7
9 5
4
15
10
17
13
11
14
13
12
SELECTRA PRO
8-52 6003-400-450-02
M
Parts
1 4501-553 BLOCK
2 4501-554 PLATE
3 3366-918 VALVE 2/2 WAY 12VDC/5W
4 4802-454 PLATE
5 3004-019 SCREW M3 X 70 ST.ZINC PLATED
DIN 96
6 3019-013 NUT M3 INSERT ST. (1.8)
7 3029-058 O-RING ID=16 X d=2 MM VITON
8 3066-052 TUBE SCREW CONN. 3/32"/1.8MM
9 3029-060 O-RING ID=7 X d=1 MM VITON
10 6001-733 CONNECTOR ASSY
11 3366-918 VALVE 2/2 WAY 12VDC/5W
12 3350-138 CONN. 2PF HOUSING MOD 1 a)
13 1513-051 TUBE, PVC 4 X 6 MM CLEAR b)
14 3066-092 TUBE ELBOW 5.9MM
15 3002-018 SCREW M3 X 8 ST.ST. DIN 7985
16 3066-139 CONN. 10-32 UNF (PP)
17 3066-090 TUBE SCREW CONN. 10/32"/5.9M
SELECTRA PRO
M 6003-400-450-02 8-53
Parts
14
15
16
25
2
4 26
17
3
5 21
6 22
5 18
7
23
8 24
19 20
6
12
11
10
9
27
1 28
13
SELECTRA PRO
8-54 6003-400-450-02
M
Parts
1 3062-041 JERRYCAN 5 L
2 4100-548 CAP
3 4100-703 PROP
4 3820-262 STICKER WATER
5 4104-080 PIPE
6 1513-001 TUBE, PVC 2 X 4 MM CLEAR a)
7 3066-093 TUBE CONN. MALE/3.6MM
8 3066-107 NUT NYLON BLUE
9 3066-052 TUBE SCREW CONN. 3/32"/1.8MM
10 4105-228 BUSH
11 4105-228 BUSH
12 3066-083 WATER FILTER
13 3820-266 STICKER "WATER"
14 3820-261 STICKER DILUTED WASTE
15 3004-211 SCREW M3 X 8 ST.ST. DIN 965
16 4100-669 CAP
17 4100-800 PROP
18 3344-016 SWITCH, LIQUID LEVEL N.C.
19 3344-016 SWITCH, LIQUID LEVEL N.C.
20 3020-027 CIRCLIP 7 DIN 6799 RVS
21 4104-104 PIPE
22 1513-051 TUBE, PVC 4 X 6 MM CLEAR a)
23 3066-094 TUBE CONN. MALE/6.4MM
24 3066-105 NUT NYLON RED
25 3351-028 CONN. 3PF CABLE
26 3351-028 CONN. 3PF CABLE
27 3820-364 STICKER "BIOHAZARD"
28 3820-265 STICKER DILUTED WASTE
SELECTRA PRO
M 6003-400-450-02 8-55
Parts
8.7.1 Panels
7
14
6
13
5
12
4
11
3
10
2
9
1
SELECTRA PRO
8-56 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-57
Parts
12
20
14 11
13
15 17
1
18
19
16
11
10
4
5
8
6 9
SELECTRA PRO
8-58 6003-400-450-02
M
Parts
a) Different electronic boards are fitted on the brackets. Two boards can be fitted on
each bracket. See par. 9.1.2.
SELECTRA PRO
M 6003-400-450-02 8-59
Parts
10
26 11
4 3
6 5
19
25
24
20 2
7
14 16
21
17
18
22 12
15
8
13
23
15
SELECTRA PRO
8-60 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-61
Parts
30
11
30
12
32
30
13
28
9
15
14
29
31
27
7
16
33
5
17
1
6
26
8
25
2
10
18
34
21
21
20
4
19
3
4
22
23
24
SELECTRA PRO
8-62 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-63
Parts
6
6
15
20
17
21
19
18
12
15
16
1
15
11
14
2
9
7
4
10
13
5
3
6
8
7
6
SELECTRA PRO
8-64 6003-400-450-02
M
Parts
SELECTRA PRO
M 6003-400-450-02 8-65
Parts
SELECTRA PRO
8-66 6003-400-450-02
M
9
Electronics
This chapter contains PCB layouts and electrical diagrams of the main electronic boards in
the instrument. For electronic boards that are servicable, parts lists are included.
SELECTRA PRO
M 6003-400-450-02 9-1
Electronics
9.1 Introduction
This chapter includes PCB layouts and electronic diagrams of the control boards in the
analyzer. In some cases, an electronic parts list is also included. The information is mainly
given to help you find the cause of a malfunction.
NOTICE
In principle, the information in this chapter allows servicing control boards on the component
level. However, repairing PCB’s is not a simple task and may lead to unsuspected damages
to sensitive components. It is recommended to exchange defective boards rather than try to
repair them.
Info
i Many boards in the Selectra ProM contain SMD components. These boards
cannot be serviced without specialist equipment, if this can be done at all.
Therefore, parts lists for these boards are not included in this manual and
components for these boards cannot be ordered from Vital Scientific.
NOTICE
PLEASE NOTE
i Trying to repair SMD components on PCB’s voids any warranty on these
boards.
NOTICE
SELECTRA PRO
9-2 6003-400-450-02
M
Electronics
SELECTRA PRO
M 6003-400-450-02 9-3
Electronics
Ambient board
The ambient board is mounted inside the left corner compartment. This is the
compartment that is used for the optional ISE module.
LLD board
The LLD board is mounted inside the back compartment, against the back wall of the
measurement compartment.
Valve boards
The valve boards are mounted inside the back compartment, against the back wall of the
pumps and valves unit.
SELECTRA PRO
9-4 6003-400-450-02
M
Electronics
The dispenser driver board is mounted inside the right compartment, in the back of the
pipettor unit.
ISE/stirrer board
The ISE/stirrer board is mounted in the main compartment, on the side of the stirrer unit.
SELECTRA PRO
M 6003-400-450-02 9-5
Electronics
The cooling unit board is mounted in the main compartment, below the cooling body of the
reagent rotor.
Info
i The cooling unit board cannot be accessed from the top. The image is only
intended to indicate the position of the board and does not show the base of
the reagent rotor that covers the cooling unit board.
The PSID interface board is plugged into a connector on the MCB that drives the sample
rotor.
The PSID scanner board is located on a mounting block behind the sample rotor.
Info
i The PSID scanner board has two identical connectors. When replacing the
scanner unit, make sure to connect the cables in the correct positions. The
top connector [PL1] has 3 wires; the bottom connector [PL2] has 2 wires.
SELECTRA PRO
9-6 6003-400-450-02
M
9.2
DISPENSER OPTO BOARD
ZK
1
2 2 X1 X5
' ZK\H
DOWN 12 12
REAGENS ( 4 UG 4
SELECTRA
' EQ 10
10 X6
RESET X2
REAGENS ( X9
PL8
' 8 YW 8
PRO
DOWN
M
SAMPLE ( X3 X7
' J\
6 6
RESET (
SAMPLE EN
9 9 X8
3/ X4
X3 X3
J\
UG
X10 X11
Wiring diagrams
X15
PL2
REAGENT SYRINGE
MOTOR \O X13
X12
EN
X1 EN
J\ ZK
X16
UG PL5 EX X14
A1 /PL2
EQ see page 3/7
YW
SAMPLE SYRINGE PL4
MOTOR X2
\O
6003-400-450-02
EN
REAGENS JQEQ
VALVE PL2
JQZK
( X1
REAGENS '
UG
VALVE NEEDLE ZK
( J\
REAGENS PL7
EN
'
VALVE WATER
ZKUG
X3
EX
SAMPLE (
YW
VALVE NEEDLE EN
' PL6
RJ
UGZK
(
SAMPLE X4 X3
VALVE WATER '
9-7
Electronics
9-8
VACUUM
SWITCH 9
3374-197 3374-143
PUMP
Electronics
8001-831
J\
EN
EN
EN
UP DOWN WATERPUMP
3 4 1 2
X7
'
'
3374-143 PL3 PL2 X5
EN
(
(
9 PL10
EN
X5
8001-790
3374-143 EN
3374-184
9 3/
EN
;
UG
YW
EQ
ZK
UG
J\
EN
EX
JQZK
JQ
ZK
EN
UG
UGZK
3374-143 VACUUM VALVE BOARD
EN
8 1 2 5 7 3 4 6 2 3 1 5 4 6
3374-199 X3 X7 9 3/
bk PL12 PL3
2 EN
V6 PL10 ;
bk 1
X5 3374-143 EN
9
3374-199 3/
bk 2 EN
V7 PL9 ;
bk 1 8001-768
X5 EN
9 3/
3374-199 bk BELLOWS VALVES BOARD
2 EN
V8 PL8 ;
bk 1
6003-400-450-02
X5 EN
3374-199 9 3/
bk 2 EN
V9 PL7 ; 3/
bk 1 1 2 3 4 5 6 7 1 0 11 1 2
X5 ;
ZKYW
ZKRJ
ZKJQ
J\
ZK\H
ENUG
EN
ZKUG
ZKJ\
EN
3374-199 bk 3374-313
2
V10 PL6
bk
6003-400: Selectra ProM wiring diagram (sheet 2 of 7)
1
X5 A1 PL12
see page 3/7
PL11 PL4
1 2 3 4 5 6 12
X3 X3
ZKUG
ZK\H
ZKJ\
ZKRJ
ZKJQ
ZKYW
EN
UGZK
JQEQ
JQZK
ZK
EN
UGEN
YW
UG
UG
EX
EQ
J\
3374-304 3374-315
SELECTRA
A3A PL2/PL6 A1 PL13
See page 5/7 See page 3/7
PRO
M
3374-315 A5A MCB
ISE Sipperarm ZKUG A2B MCB
ZK\H see page 6/7 see page 5/7
driver board 3/
see page 5/6 ZKJ\
; ZKRJ bellows valves
3/
SELECTRA
ZKJQ see page 2/7
ZKYW
0 FAN EN
12V ;
ZKJQ
ZK\H
UG
EN
EN
RJ
3374-308
ZKJQ
EN
UG
RJ
ZK\H
3374-307
UG
EX
3374-313
PRO
ZKYW
ZK\H 1 3 4 8 10 12 1 3 4 6 5
M
.
ZKJ\ 3374-305 X3 X2
PL4 PL2
ISE 3374-326 ZKRJ J\
INTERFACE ZKJQ J\
ISE UNIT BOARD UG 3/ vacuum valves A5- TEMP. EN PLL BOARD
ZKUG BOARD PL5
ZKYW YW
J\ see page 2/7 see page 6/7 YW 8001-763
ZK
3/
3/
ENUG X2
EX PL1
. EN TAB2 PL3 TAB1
3/ EN
EN 275618394 275618394
;
; ; X15 X15
3367-095
3374-211 3374-314 8003-287
8003-377 ZKYW 8003-287
ZK\H
3374-309 ZKJ\ pump valves 8001-995
ZKRJ see page 4/7 ARM BOARD
OPTION A4-PL7 3/ ZKEN
3/ SAMPLE NEEDLE
See page 6/7 EN PL2 PL1
; J\
EQ J\
YW
YW
PL WS YW
12V EN UG
UG see page 4/7 2&
; ; 8001-995 ZK\H
FAN 0 . EN
EX
ARM BOARD ZKJQ ZKJQ
RJ 7
;
EQ 3374-306 EQ PL2 PL1 X3 X3
COOLING UNIT J\ UG
12V 3374-257 3374-324 ZK
EN J\
UG
A1-SYSTEM BOARD
BOARD 3/
JQ ZK YW EQ YW \H YW
COOLING YW
0 . EN \H
EN . EN ZK UG
PUMP 3/
EQ 2& 3374-332
; ; ; ZK\H
; EN
UG ZKJQ REAGENT NEEDLE
RJ ZKJQ
. 3374-310 7
EN
6003-400-450-02
ZK X3 X3
Temp_sensor J\
UG
. EX
EN YW
8003-007 EQ 8003-068
EN ZK
OPTION 3/ ZK\H Peltier heat exchanger
3/
ZKJ\ 3374-328 3374-329
ZKRJ Part of cooling unit OPTION
ZKJQ
ZKYW
A2B-PL5 (p 5/7) 3374-318 ZKUG
A3A-PL5 (p 5/7) 3/ \HUG 8001-786
A5B-PL5 (p 4/7) \HEN Water Cask
6003-400: Selectra ProM wiring diagram (sheet 3 of 7)
; EN
; ;
A6-PHOTOMETER BOARD
A2A-PL5 (p 1/7) 3374-317
A3B-PL5 (p 6/7) 3/ 3/ 3/
A5A-PL5 (p 6/7)
;
3374-223 3374-224
3/
2 3 4 5 6 7 8
;
3374-095
EQ
J\
YW
ZK
EX
EN
ZKJ\
3 2 6 5 4 8 7
; MEASURING REFERENCE
PHOTO CELL PHOTO CELL
6001-756
Panel PC
9-9
Electronics
9-10
Electronics
12V EN
3374-291 3374-323 VACUUM
PUMP 0 PL2
UG
X5 8001-791
JQEQ
JQEQ
3374-288
JQZK JQZK EN
WATER
PUMP 0 PL9
ZK
SAMPLE DISC PL2 X5
MOTOR PUMP VALVE
3374-143 BOARD
UGZK UGZK EN
9 PL7
UGEN UGEN
X1 EN
X10 X5
JQEQ JQEQ 3374-143
8003-375 EN
JQZK JQZK 9 PL8
EN
ZK\H
ZKRJ
ZKJ\
ZKYW
6003-400-450-02
X2
( EN EN
RESET
DIR B J\ J\
6003-400: Selectra ProM wiring diagram (sheet 4 of 7)
' EQ EQ
SAMPLE
X3
UG UG PL1 SPI to RS232 board
PL10
( EN EN see page 5/7
DIR A
REAGENT ZK ZK Option
' EX EX
PL7
UG UG
( YW YW
DIR B
REAGENT J\
J\
' EQ EQ
X3
X9
SELECTRA
PRO
M
UGEN
Bellows Valve board PL11 UGZK
JQEQ JQEQ
see page 2/7 JQEQ
PL2
JQZK JQZK JQZK
X1
UG
SELECTRA
UG
EX
SAMPLE ARM PL1
Bellows Valve board PL11 HOR. MOTOR
EQ
see page 2/7 J\ PL6
YW UGZK UGZK
8003-375
ZK
EN UGEN UGEN
PRO
X1
X3 X10
M
EN JQEQ JQEQ
ZK
JQZK JQZK
EX 3374-318
X9 PL5 A1-PL4
EQ see page 3/7
EN
YW SAMPLE ARM
JQEQ JQEQ PL2 ZK
X2 VERT. MOTOR
PL5 EX A1-PL4
JQZK JQZK see page 3/7
EQ
UGZK UGZK
YW
WASH ARM PL1 UGEN UGEN X2
MOTOR X10 X1
UG UG
( EN EN
UGZK
UGZK
DIR A
ZK ZK
A2B-MOTION CONTROL BOARD
' EX EX
UGEN UGEN PL7
UG UG
X1
X10 ( YW YW
DIR B
J\ J\
J\ J\ ' EQ EQ
' EX EX UG UG X3
HIGH ( EN EN
RESET
( UG UG
HOR. ZK ZK
6003-400-450-02
DOWN
(
; X3
EN EN EN
EN
ZK\H
RJ
ZKJQ
UG
PL2
J\
ZK ZK
X1 PLL Board
WASH ARM BOARD X3 X3 see page 3/7
CUVETTE
\H VS EN EN PL4
ISE ARM
8001-773 \H \H
SOLENOID \H
6003-400: Selectra ProM wiring diagram (sheet 5 of 7)
VS \H \H
X2
EN EN
ZKRJ PL1 RJ RJ
VS
RA. STIRRER M EN VS ZKYW ZKYW
ZKRJ ZKRJ A1-SYSTEM BOARD
ZKJQ ZKJQ
YW YW
PL1
ZKJQ VS EN EN
SA. STIRRER
M EN VS UG
UG 8002-773
ISE SIPPER/STIRRER
DRIVER BOARD
X3 X2
UG UG SPI to RS232
PL2 EN EN
BOARD
BARCODE JQ JQ
YW YW PL2 For programming
PL3
PL1 EN EN PL1 purposes
X2
3365-169
3367-109 X8
See additional info in section
PL10 A5B MCB
“Location of other boards” page 4 Option
9-11
Electronics
9-12
8003-355 3374-321
8003-375 3374-322
UG UG 3
( EN EN
FILTERWHEEL 9 JQEQ JQEQ 8003-375
RESET ZK ZK
7 JQZK JQZK 3374-317
' EQ EX
PL8
Electronics
UG UG 4 EN
(
MEASURING DISC YW YW 5 ZK
REAGENT ARM
RESET J\ HOR. MOTOR PL1 EX
J\ 1 PL5 A1-PL2
' EX EX
8 EQ see page 3/7
UG UG UGEN UGEN
2 X3 YW
( EN EN
MEASURING DISC 8 UGZK UGZK
X2
DIR A ZK ZK 7 X10 X1
' EX EX
JQEQ JQEQ
UG UG
MEASURING DISC ( YW YW
PL7 JQZK JQZK
DIR B
J\ J\
5 3374-317
' EQ EQ 4 REAGENT ARM
ZKJQ UGEN X3 EN VERT. MOTOR PL2
1
ZK
JQ ZKUG 2 UGZK UGZK
EX
PL5 A1-PL2
EQ see page 3/7 UGEN UGEN
X10
FILTERWHEEL YW UG UG
X1
MOTOR ( EN EN
PL2
X2 DIR. A ZK ZK
ZKUG JQEQ ' EX EX PL7
UG UG
ZKJQ ( YW YW
UG 4 DIR. B
4 RES. HOR.
ZK ZK
' EX EX
PL8
UG UG
MEASURING DISC PL1 ( YW YW
MOTOR RES.VERT. J\ J\
' EQ EQ
UGZK UGZK 2 X3
EN X9
UG UGEN 1 X1
3374-308
UG EQ UG
6003-400-450-02
ZKJQ
ZK ZK
8003-220 3374-309 PLL. Board ZK\H
EN PL6
see page 3/7
ZK\H RJ
EN JQ EN
+ ZKJ\
JQ\H EN
PL5 ZKRJ X3
PL7
8003-295 YW RJ ZKJQ A1-PL11
see page 3/7
ZKYW
HEATING ELEMENT J\ EN EN
X2
COVER CUVETTEN
ZK RJ X2
12V
20W ZK ZK
- PL2 3374-305
6003-400: Selectra ProM wiring diagram (sheet 6 of 7)
UG YW
J\
EN J\ J\
X6 EN
PL6 PLL board
X9 YW see page 3/7
YW
PL4
2 1
;
TEMPERATURE
SELECTRA
COMPENSATION
ASSY
2 UG
1 EN
Temp_sensor
PRO
M
SELECTRA
PRO
M
VACUUM VALVE BOARD PUMP VALVE BOARD
ELECTRONIC ELECTRONIC ELECTRONIC
BOARD BOARD BOARD
PL1 F1 PL1 F1
+
+14V
14V +14V
+14V
3.15 A 3.15 A
FAN FAN FAN
8001-790 8001-791
+5.15V
+12V
POWER SUPPLY
1
LINE IN POWER OUT
+14V
3359-048
+24V
DISTRIBUTION BLOCK K1
COOLING UNIT
6003-400-450-02
F2 PL3 PL3 PL3 PL3 PL3
ISE INTERFACE BOARD F1 PL3 PL1
1A
1A
F1 F1 F1 F1 F1 F1
+5V F1 F1 F1 F1
+24V F2 3.15A
F2 F2 F2 F2 F2
3.15A 3.15A
F2 F2 1.A15A
3
PL3 3.15A 3.15A 3.15A 3.15A 3.15A
3.15A F3 F3 F3 3.15A F3 F3
3.15A 3.15A 3.15A F3 F3
3.15A 3.15A 3.15A +14V OPTION
3.15A
+30V 1A 1A 1A 1A 1A VCC
+30V +30V +30V 1A 250MA +30V +30V
VCC +30V +30V
+14V +14V +14V +14V +14V 8003-068
+14V +14V
VCC VCC VCC VCC VCC
8002-773 VCC VCC
8003-375 8003-375 8003-375 8003-375 8003-375
8003-375 8003-220
PL8
8001-768
SYSTEM BOARD
ISE UNIT MCB MCB MCB MCB MCB PHOTOMETER BOARD
MCB TEMP.CONTROL BOARD
6003-400: Selectra ProM wiring diagram (sheet 7 of 7)
8003-377
OPTION
9-13
Electronics
Electronics
SELECTRA PRO
9-14 6003-400-450-02
M
Electronics
SELECTRA PRO
M 6003-400-450-02 9-15
Electronics
SELECTRA PRO
9-16 6003-400-450-02
M
PL13
PL2 KB0 V6 1
D-GND 1 KB1 V7 2
M-TXD 2 KB2 V8 3
VCC VCC M-RXD KB3 V9
3 4
M-RTS 4 KB4 V10 5
M-CTS 5 KB5 V11 6
PL9 D-GND P2B0
6 7
1 P-TXD 7 P2B1 8
VCC VCC P-RXD P2B2
2 H-TXD 8 9
3 MD_COM P-RTS 9 P2B3 10
4 H-RXD P-CTS 10 C45 C49 C44 C48 11
D-GND
SELECTRA
5 H-RTS 11 VCC 12
6 PIP_COM R-TXD 12 1n 1n 1n 1n
7 H-CTS R-RXD 13 BELLOWS VALVES
8 MASTERSIOS R-RTS 14 1
H-DTR R-CTS
9 15
H-AUX RA_COM RN2.A
10 16
4k7
11 BUS BUS
SLAVE COMM. 1 PL12
12 2
MASTERCPU SA_COM KB6 V1
PRO
C47 C40 C46 C41 C36 1
HOST COMM. PL4 KB7 V2 2
D-GND KC0 V3
M
1n 1n 1n 1n 1n 1 3
S-TXD 2 KC1 V4 4
WA_COM S-RXD 3 KC2 V5 5
S-RTS 4 P2A4 V13 6
I-TXD S-CTS KA0 VACUUM SWITCH
5 7
PL8 I-RXD DS_COM D-GND 6 8
W-TXD 7 9
1 D-GND D-GND
W-RXD 8 P2A5 V14 10
2 773-2-1.SCH 773-3-1.SCH W-RTS 9 C43 11
3 W-CTS 10 12
4 D-GND 11 R11 1n
5 VCC D-TXD VCC VACUUM VALVES
12
6 D-RXD 13
D-RTS 14 R9 R10
ISE-UNIT COMM.
D-CTS
4k7
4k7
opt.
15
16 PL10
VCC P2A2 1
SLAVE COMM. 2
P2A3 2
D2 1N4148 3
KC3 V12 4
PL1 VCC P1A5 WATER PUMP 5
1 VCC ISE-DOWN P1B6 VCC FAN 6
2 ISE-OPTO P1B5 KA1 WASTE SWITCH 7
ISE-OUT P1B4 1 P2A0 CONC. WASTE SWITCH
3 8
4 S_STIRR P2A6 RN2.C P2A1 9
R_STIRR P2A7 KC5 VACUUM PUMP
4k7
5 10
6 MASTERPIOS C37 C38 C42 11
C50 C16 PL6 4 12
ISE SIPPER 1n 1n 1n
R7 R8 1 KA6
STIRR CONN. 1n 1n PUMP UNIT
BUS
33k
33k
2
IC10.A
6002-773: Electric diagram - main
3 P1A4
4
1 2
VCC
SPARE PL3
74HCT04 MEAS/REF 1
1 1 P1A7 MEAS/REF 2
P1A6 STOP/START 3
6003-400-450-02
RN2.B RN2.D
PH-STRT 4
4k7
4k7
D-GND REF-STRT 5
PL7 3 5 6
773-4-1.SCH
1 WATER ALARM 1 KA2 P1A0 P-G0 7
2 P1A1 P-G1 8
3 WATER ALARM 2 P1B7 P1A2 P-G2 9
4 P1A3 P-G3 10
P1B0 R-G0 11
WATER CASK P1B1 R-G1 12
P1B2 R-G2 13
C39 C35 P1B3 R-G3 14
15
1n 1n
16
PHOTOM. BOARD
PL5
F2 VCC
1
2 3.15 A
3
4 F1
PL11 5 +30V
NEEDLE THERM. CONTROL KC7 6 3.15 A
1
2 LAMP CONTROL KC6
POWER SUPPLY C34 C30
3 NEEDLE TEMP OK KA5
4 CUVET TEMP OK KA4
15U 47U
5 CUVET THERM. CONTROL KA3
6 R5
5k1
4k7
4k7
C32 C31 100
RN3.A
RN3.B
3.6V
1 1 1n 1n
TAB7
TAB6 TAB4
GND-BEAM
VCC +30V
B1
1 2 3 4
G1 G2 G3 G4
í3.2
í3.2
í3.2 í3.2
STICKER
9-17
Electronics
TAB2
9-18
VCC BAUDR BUS
A4 O4 A4 D4 WR A4 D4 PA4
A5 7 19 D5 A5 5 17 D5 RD 21 35 A5 D5 69 26 KA5 R2 VCC
A5 O5 A5 D5 RD A5 D5 PA5
10k
A6 6 20 D6 A6 4 18 D6 36 A6 D6 70 27 KA6 5 1
A6 O6 A6 D6 A6 D6 PA6 RST PBRST
A7 5 21 D7 A7 3 19 D7 28 37 A7 D7 71 28 KA7 SW2
A7 O7 A7 D7 REFSH A7 D7 PA7
A8 27 A8 25 38 A8 RES 6 2
A8 A8 A8 RST TD
A9 26 A9 24 18 39 A9 A0 1 45
A9 A9 HALT A9 A0 ARDY
A10 23 A10 21 40 A10 A1 84 46 KA7 7 3
25 A10 23 A10 24 A10 1 83 A1 ASTB ST TOL
A11 A11 VCC A11 A11 WAIT WAIT A11 A11 A2 A2
A12 4 A12 2 2 A12 A3 82 42 KB0 GND
28 A12 26 A12 16 A12 3 A3 PB0 41
A13 A13 A13 A13 INT INT A13 A13 PB1 KB1 MAX1232
A14 29 1 P2B7 1 NMI 17 4 A14 KIO 81 40 KB2 4
A14 VPP A14 NMI A14 CS PB2
A15 3 5 A15 39 KB3
A15 A15 PB3 TAB3 TAB5
A16 2 31 MW 27 RES 26 M1 80 38 KB4
A16 PGM RW RESET M1 PB4
MR 22 TAB1 14 D0 RD 79 37 KB5
OE D0 RD PB5
RAM 20 BRQ 25 15 D1 IORQ' 77 36 KB6
ROM 22 CE BUSRQ D1 IORQ PB6
CE 23 12 D2 INT 54 35 KB7
BUSAK D2 INT PB7
MR 24 8 D3
OE 55257 D3
CLK CLK 6 7 D4 CLK 57 43
CLK D4 9 CLK BRDY 44 IC19
D5 D5 BSTB
27C1001A 10 D6 RES 76 C22 1 2
D6 13 RESET 11 C1+ V+
D7 D7 PC0(WT/RDYB) KC0 C25
IE0 51 12 KC1 15U 3
IEI PC1(SYNCB)__ C1- 15U
IE1 52 13 KC2
Z80-CPU IEO PC2(DTRB)___
14 KC3 C27 4
PC3(RTSB)___ C2+ C29
72 29
VCC CLK/TRG0 PC4(RTSA)___ 15U
50 30 KC5 15U 5 6
ZC/TO0 PC5(DTRA)___ C2- V-
IC4 IC2.A 31 KC6
PC6(SYNCA)__
VCC D0 25 1 REF-STRT CLK 73 32 KC7 11 14
D0 CLK/TRG1 PC7(WT/RDYA) IN1 OUT1 H-TXD
D1 26 23 3 BAUDR 49
27 D1 CLK/TRG0 ZC/TO1
D2 7 2 18 12 13
28 D2 ZC/TO0 74 RXDA 17 OUT3 IN3 H-RXD
D3 D3 CLK/TRG2 RXCA
C19 C18 C17 C13 D4 1 48 15 10 7
D4 74HCT08 VCC ZC/TO2 TXDA IN2 OUT2 H-RTS
D5 2 16
100N 100N 100N 100N D5 TXCA
D6 3 22 75 9 9 8
D6 CLK/TRG1 CLK/TRG3 CTSA OUT4 IN4 H-CTS
D7 4 8 4 47 8
D7 ZC/TO1 IC1.A ZC/TO3 DCDA
IC2.B BAUDR MAX232
CTC 16 4 5 2 C14 2
CE Q PR D RXDB
A0 18 21 6 33p 60 3
CS0 CLK/TRG2 XTAL0 RXCB
A1 19 9 5 3 CLK 5
VCC CS1 ZC/TO2 CLK X1 7.3728Mhz TXDB
M1 14 59 4
10 M1 6 XTAL1 TXCB 6
IORQ' IORQ 74HCT08 Q C15 CTSB IC18
RD 6 20 CL 33p 56 7
C12 C10 C11 C9 RD CLK/TRG3 CLKOUT DCDB C21 1 2
IE4 13 C1+ V+
IEI 1 55
74HCT74 C24
100N 100N 100N 100N OSC 15U 3
IE5 11 C1- 15U
IEO
INT 12 17 CLK
INT RESET VCC C26 4
CLK 15 Z80-KIO C2+ C28
CLK VCC
5 6 15U
PH-STRT 15U
6002-773: Electric diagram - CPU core
Z80-CTC C2- V-
10
6003-400-450-02
VCC RES IC1.B 11 14
IN1 OUT1 I-TXD
VCC 9 PR 12
Q D 12 13
OUT3 IN3 I-RXD
C7 C4 C5 C8 IC3 11 P2B6
74HCT08 CLK 10 7
A14 1 15 74HCT08 1 1 1 IN2 OUT2 H-DTR
100N 100N 100N 100N A Y0
A15 2 14 13 R4 8
B Y1 RN1.A Q P2B5 9 8
3 13 ROM CL R1 C6 OUT4 IN4 H-AUX
10k
C Y2 11 10
12
4k7
4k7
4k7
1ook
VCC
RN1.C
RN1.B
Y3 12 8 74HCT74
D-GND 11 IC2.D 9 INT 4 3 2 13 1N MAX232
Y4
6 10 IC2.C NMI
G1 Y5
4 9 RAM BRQ
G2A Y6 VCC
5 7 IE0
A16
G2B Y7
IC12.A
RD 1 74HCT138
3 MR IC11
MREQ 2
A4 1 15 KIO
A Y0
A5 2 14 SIO1
74HCT32 B Y1
A6 3 13 SIO2
C Y2
VCC 12 SIO3
Y3
11 CTC
Y4
IC12.B 6 10 PIO1
G1 Y5
WR 4 4 9 PIO2
G2A Y6
6 MW 5 7
5 G2B Y7
MREQ VCC
74HCT138
74HCT32 VCC
C2 C1 C3 C23 C33
M1
1 1u 100N 100N 100N 100N
RN1.D
IC10.B IC8
4k7
5 3 14 IC9.A IC10.D
A QA
CLK 3 4 4 13 1 IC9.B
5 B QB 12 3 9 8 4
6 C QC 11 2 6
SELECTRA
D QD IORQ'
74HCT04 15 5
RCO
7 74HCT04
ENP 74HCT32
IC10.C 10 74HCT32
ENT
2 IC9.D
CLK
M1 5 6 9 12
LOAD
1 11 WAIT
CLR
13
74HCT04
PRO
74HCT161
IORQ 74HCT32
M
Electronics
BAUDR
IC17A
D0 1 13
D0 RXDA
D1 2 14
D1 RXCA
D2 44 16
D2 TXDA
D3 3 15
43 D3 TXCA 12 S-TXD
D4 D4 SYNCA
D5 4 11 S-RXD
42 D5 W/RDYA
D6 D6
5 19 S-RTS
D7 D7 RTSA
20 S-CTS
CTSA
SIO2 40 18
CE DTRA
RES 23 21 C55 C56 SA_COM
RESET DCDA
M1 9
M1
IORQ' 41 33 1n 1n
IORQ RXDB
RD 37 32
RD RXCB
A0 38 30
39 CD TXDB 31
A1 BA TXCB 34 R-TXD
6 SYNCB 35
INT INT W/RDYB R-RXD
IE2 7
IEI R-RTS
IE3 8 26
IEO RTSB
25 R-CTS
CTSB
CLK 22 27
CLK DTRB
24 C57 C58 RA_COM
DCDB
Z80-SIO/4
1n 1n
BAUDR
BUS
IC21A
D0 1 13
D0 RXDA
D1 2 14
44 D1 RXCA 16
D2 D2 TXDA
D3 3 15
43 D3 TXCA 12 D-TXD
D4 D4 SYNCA
D5 4 11 D-RXD
D5 W/RDYA
D6 42
D6 D-RTS
D7 5 19
D7 RTSA
20 D-CTS
CTSA
SIO3 40 18
CE DTRA
RES 23 21 C59 C60 DS_COM
RESET DCDA
M1 9
M1
IORQ' 41 33 1n 1n
37 IORQ RXDB
RD RD 32
A0 38 RXCB 30
39 CD TXDB 31
A1 BA TXCB
34 W-TXD
SYNCB
INT 6 35 W-RXD
INT W/RDYB
IE3 7
IEI W-RTS
IE4 8 26
IEO RTSB
25 W-CTS
CTSB
CLK 22 27
CLK DTRB
24 C61 C62 WA_COM
DCDB
Z80-SIO/4
1n 1n
BAUDR
SELECTRA PRO
M 6003-400-450-02 9-19
Electronics
Z80-PIO
BUS
IC15
D0 19 15 P2A0
D0 PA0
D1 20 14 P2A1
D1 PA1
D2 1 13 P2A2
D2 PA2
D3 40 12 P2A3
D3 PA3
D4 39 10 P2A4
D4 PA4
D5 38 9 P2A5
D5 PA5
D6 3 8 P2A6
D6 PA6
D7 2 7 P2A7
D7 PA7 18
ARDY
INT 23 16 D-GND
INT ASTB
IE6 24
IEI
22 27 P2B0
IEO PB0
28 P2B1
PB1
PIO2 4 29 P2B2
CE PB2
RD 35 30 P2B3
RD PB3
IORQ' 36 31 P2B4
IORQ PB4
M1 37 32 P2B5
M1 PB5
A0 5 33 P2B6
CD PB6
A1 6 34 P2B7
BA PB7
21
BRDY
CLK 25 17 D-GND
CLK BSTB
Z80-PIO
SELECTRA PRO
9-20 6003-400-450-02
M
Electronics
SELECTRA PRO
M 6003-400-450-02 9-21
X1 X2 X3 sheet 4
TP TP TP
PL4
1
1
1
+30V +30V
9-22
1 VCC
2 +14V
3 +30V PA4
INA1 A1
4
5 1 X4 PA5 PL1
TP INA2 A2
6 1
PC2 2
Header6 X5 TAB2 ENA
3
TP TP 4
F3
1
Electronics
1 2
step-driver Header-4
1
VCC
PA6
1
Fuse holder (LT5) INB1 B1
+ C9 PA7
X6 47u/50V TAB8 INB2 B2
TP TP PC3
2
PL3 +14V +30V ENB
F2
1
1 1 2 PB7
1
+14V PWM-Imax PGND
2
1
1
1
3 Fuse holder (LT5) PB6
+ C10 I70
4 R21 R22 GND
5 X7 47u/50V TAB1
TP 10k0 10k0
6 TP
2
F1
1
2
2
Header6 1 2
1
+30V
1
1
1
TAB6 1 Fuse holder (LT5) sheet 3
TP + C11
470uF/50V R23 R24
1k00 1k00 +30V +30V
2
2
2
PA0
INA1 A1
sheet 2 PA1 PL2
PL5 INA2 A2
1
1 PC0 2
ENA
2 TxD 3
3 RxD ADC0 4
4
1
1
5
step-driver Header-4
C12 C13 ADC1 PA2
6 INB1 B1
1n 1n
2
2
Header6 PA3
INB2 B2
CTS PC1
DTR ENB
1
1
PB4
C14 C15 X112 X113 PWM-Imax PGND
1n 1n TP TP PB5
2
2
VCC I70
PL9 PA[0..7] PA[0..7]
1
1
PA[0..7] R25 GND
2 1 TCK
4 3 PB[4..7] 1 2 PB[4..7]
TDO PB[4..7] VCC
JTAG 6 5 TMS
8 7 PC[0..3] PC[0..3]
PC[0..3] 270E
10 9 TDI
6003-375: Electrical diagram - main
X8
R26
header5x2 TP
1 2 PL8 PL7
1
nRES VCC
1 VCC 1
PL10 2 2
6003-400-450-02
controller PE4 100E PD2
1 MISO VCC 3 3
SPI 2 4 PD3 4
1
MOSI VCC
3 5 R29 5
1
1
1
1
2
R27 R28 100n 100n
6 8 270E 8
2
2
1
+14V
11 11
SW10 C19 12 PD6 12
1 8 PG0 100n PD7
2
2 7 PG1 Header-6x2 Header-6x2
1
1
3 6 PG2 PE6
4 5 PG3 C20 C21
1
PG[0..3] 1n 1n
2
2
PG[0..3] C22
SW-DIP-4
100n X9
2
TP
PL6
1
PG4
SW3 1
1
VCC
Hex switch 2
6 PC7 C23 PD0 3
3 PC6 1n PD1 4
8
2
A
4 PC5 R32 5
1
1
1
1
C
1 PC4 PE7 1 2 6
2 4 6
E
PC[4..7] C25 C26
0
5 7
1
C C 1 2 4 8
1k00 1k00
C24 9
1
1
1
1
1
1n 10
2
2
2
+14V
R59 R34 R35 R36 R37 11
PD4 12
1k00 1k00 1k00 1k00 1k00
PD5
2 Header-6x2
2
2
2
2
1
1
C28 C29
1 HOLE1 1n 1n
2
2
3mm2 / 7mm5
PE[4..7] PE[4..7]
SELECTRA
PE[4..7]
1 HOLE2
3mm2 / 7mm5
PRO
N1
M
pcb_MCB2_R1B
SELECTRA
X10 X12
TP TP
R33
SW1
1
1
1 2 1 2
4 3 D3A D3B
1
100E 1 & 5 &
SW-NO C27 3 4
1n 2 6
PRO
TAB7 4093 4093
M
nRES TP
1
X11
TP
X13
1
PD[0..7] TP
C
PD[0..7] R56
1
1 2 B V8
1
VCC
BC847C
G1 X125 X23 X34
X17 X15 X18 X19 X20 X16 X21 X22 TP TP 10k0 TP E
2 8.00 MHz D3C R57 D3D
D2 TP TP TP TP TP TP TP TP 8 12
1
1
X27
& & 1
10 1 2 11
1
1
1
1
1
1
1
1
TP 1 25 PD0 9 13
3
1
nRES PEN PD0/SCL/INT0 PD1
20 26
1
RESET PD1/SDA/INT1 PD2 270k C40
27 4093 4093
PD2/RXD1/INT2 PD3 100n
24 28
2
XTAL1 PD3/TXD1/INT3 VCC
X29 X30 X24 X25 X31 X32 X26 X33 23 29 PD4
2
TP TP TP TP TP TP TP TP XTAL2 PD4/IC1 PD5
PD5/XCK1 30
31 PD6
1
1
1
1
1
1
1
1
PA[0..7] PD6/T1 R62
PA0 51 32 PD7
PA0/AD0 PD7/T2 10k0
PA1 50
PA2 PA1/AD1
49 PL11
1
PA3 PA2/AD2 PDI PDO
48 PA3/AD3 PE0/RXD0/PDI 2 RxD 1 2 VCC
PA4 47 3 PDO SCK 3 4 PDI
PA4/AD4 PE1/TXD0/PDO TxD
PA5 46 4 nRES 5 6
PA5/AD5 PE2/XCK0/AIN0 CTS
PA6 45 5
PA6/AD6 PE3/OC3A/AIN1 DTR
PA7 44 6 PE4 header3x2
X35 X36 X37 X38 X39 X40 X41 X42 PA7/AD7 PE4/OC3B/INT4 PE5
PE5/OC3C/INT5 7
TP TP TP TP TP TP TP TP 8 PE6 SPI
1
1
1
1
1
1
1
1
PE6/T3/INT6 PE7
SSn 10 PB0/SS PE7/IC3/INT7 9
SCK 11 X43 X44
1
1
1
1
1
1
1
1
SCK PB1/SCK PE[4..7]
12 TP TP
1
1
MOSI PB2/MOSI
PB[4..7] MISO 13 PB3/MISO PF0/ADC0 61 ADC0
PB4 14 60
PB4/OC0 PF1/ADC1 ADC1 X45 X46 X47 X48 X49 X50 X51 X52
PB5 15 59 1 X126TP
PB6 PB5/OC1A PF2/ADC2 X127TP TP TP TP TP TP TP TP TP
16 PB6/OC1B PF3/ADC3 58 1
PB7 17 57
PB7/OC2/OC1C PF4/ADC4/TCK TCK
56
6003-400-450-02
PF5/ADC5/TMS TMS
PC[0..3] PF6/ADC6/TDO 55 TDO
PC0 35 54
PC0/A8 PF7/ADC7/TDI TDI
PC1 36
PC2 PC1/A9 PG[0..3]
37 PC2/A10 PG[0..3]
PC3 38 33 PG0
PC[4..7] PC3/A11 PG0/WR
PC4 39 34 PG1
PC5 PC4/A12 PG1/RD PG2
40 PC5/A13 PG2/ALE 43
PC6 41 18 PG3
PC7 PC6/A14 PG3/TOSC2 PG4
42 PC7/A15 PG4/TOSC1 19 PG4
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
X128
6003-375: Electrical diagram - controller section
TP
62
1
Aref
22 64
1
X54 X55 X56 X57 X58 X59 X60 X61 GND AVcc X62 X63 X64 X65 X66 X67 X68 X69 X114
53 GND
TP TP TP TP TP TP TP TP 63 52 C42 TP TP TP TP TP TP TP TP TP
GND Vcc 100n
21
Vcc 2
ATmega64L
1 X129
TP
1
2
C41
2
100n R60
100E
X14
TP L1
1
1 2
1
VCC
1
300E@100MHz
C30
100n
2
9-23
Electronics
+30V
A1A
9-24
VS 4
X117
TP
10 15
1
INA1 IN1 OUT1 A1
1
logic
11 5 X118
INA2 IN2 OUT2 TP
R38
8
1
Electronics
ENA EN 6k80
X72
prot 12 TP
1
X70 SENSE
1
TP one shot TP
2
1
13 9 X71 C2
1
RC VREF 100n B7 B5
1
1
1
1
red red
1
C43
R2 47n R17 R18
VCC C1
2
82k0 L6207 0E47 5% 0E47 5%
1n
2
1
2
X74
2
2
2
C
R3 A2
TP
1 2 B V1
1
PWM-Imax BC847C
1
E
100k C3 X77
100n 1 TP
2
R5
1 2
10k0
1 X115
TP
R6
I70 1 2
1
2k70
A1B
R7
1k20
VS 33
2
X119
TP
26 22
1
INB1 IN1 OUT1 B1
1
logic
27 32 X120
INB2 IN2 OUT2 TP
R39
29
1
ENB EN 6k80
X84
prot 25 TP
2
SENSE 1
X82
1
TP one shot
2
1
24 28 C5
1
RC VREF X83 100n B8 B6
2
1
1
1
1
TP red red
6003-400-450-02
C44
R9 47n R19 R20
C4
2
82k0 L6207 0E47 5% 0E47 5%
1n
2
1
2
2
2
2
B2
6003-375: Electrical diagram - stepper driver 1
PGND
X87 V3 X88
TP 1 2 1 1 TP
A1C
1N4148SMD
30
1
VBOOT charge
pump GND
C7 7
220n/16V VCP
2
2 NC GND 1
3 NC GND 18
SELECTRA
V4 6 NC
1 2 GND 19
14 NC GND 36
1N4148SMD 16
X89 NC
17 NC GND (SLUG) 37
R15 TP
C8
100E X90 L6207
1
1 2 1 2 1 TP
PRO
10n
M
+30V
A3A
VS 4
X121
TP
10 15
1
INA1 IN1 OUT1 A1
1
logic
11 5 X122
INA2 IN2 OUT2 TP
R41
8
1
ENA EN 6k80
X93
prot TP
SELECTRA
12
1
X91 SENSE
1
TP one shot TP
2
1
13 9 X92 C32
1
RC VREF 100n B20 B18
1
1
1
1
red red
1
C45
R42 47n R43 R55
VCC C31
2
82k0 L6207 0E47 5% 0E47 5%
1n
PRO
2
1
2
X95
2
2
2
C
R44 A2
TP
M
1 2 B V5
1
PWM-Imax BC847C
1
E
100k C33 X98
100n 1 TP
2
R46
1 2
10k0
1 X116
TP
R48
I70 1 2
1
2k70
A3B
R47
1k20
VS 33
2
X123
TP
26 22
1
INB1 IN1 OUT1 B1
1
logic
27 32 X124
INB2 IN2 OUT2 TP
R50
29
1
ENB EN 6k80
X105
prot 25 TP
2
SENSE 1
X103
1
TP one shot
2
1
24 28 C35
1
RC VREF X104 100n B19 B17
2
1
1
1
1
TP red red
6003-400-450-02
C46
R51 47n R52 R53
C34
2
82k0 L6207 0E47 5% 0E47 5%
1n
2
1
2
2
2
2
B2
6003-375: Electrical diagram - stepper driver 2
PGND
X108 V6 X109
TP 1 2 1 1 TP
A3C
1N4148SMD
30
1
VBOOT charge
pump GND
C36 7
220n/16V VCP
2
2 NC GND 1
3 NC GND 18
V7 6 NC
1 2 GND 19
14 NC GND 36
1N4148SMD 16
X110 NC
17 NC GND (SLUG) 37
R54 TP
C37
100E X111 L6207
1
1 2 1 2 1 TP
10n
9-25
Electronics
Electronics
SELECTRA PRO
9-26 6003-400-450-02
M
VP15 R1 1M00
A10 1 2 VP5V0 VP15
2 6 Vref_10V
Vin Vout
1 2
5
1
1
1
1
GND Trim R2 10k00 0.1%
C1 ADR01ARZ C2 C3
4 R3
100n/50V 99:1 100n/50V 100n/50V
2
2
2
100E0 0.1%
12
13
A5A
2
2 VL Vdd 3
D1 S1
SELECTRA
1 Logic
VP15 VP5V0
1 pos
"PHOTOM. CELL" TAB10 GND Vss C4
TP 1 2
5
4
1
1
PL4 GUARD PHOTOM. CELL DG445DY
1
1p-coax R12 100E C5 C6
1
100n/50V 10n/50V 100n/50V
1 1 2
2
2
IN OUT R4
1
1
470E
5
4
3
2
PRO
P-G[0..3] R5
C30 C7 A8
2
8
5
6
G[0..3] 1u/50V 100n/50V 1 2
2
2
M
2 V+B BS
1
+
VP15 7 PH-STRT
VN15 100E C8
VP15 VP15 3 -
A5B 10n/50V V- G
2
15 14 LT1011AI
4
1
1
VP5V0 VP5V0 D2 S2
1
C9 GUARD 16 Logic
GND 100n/50V 1 pos C10
2
10n/50V
2
1
VN15 VN15 DG445DY
C11
7
8
A5D
R6 1K60 A7 100n/50V
2
Sensor Amplifier V+ OPA124U
7 6 1 2 3 +
D4 S4 6
1
8 Logic 2
1
-
1 pos C12 V- VP5V0 VP15 VN15
R7 68n/25V
4
1
5
DG445DY 33K0 2%
2
1
1
C13 C14
1
100n/50V 100n/50V
2
2
NOT-MEAS/REF
MEAS/NOT-REF
C15
12
13
100n/50V A6A
2
2 VL Vdd 3
D1 S1
1 Logic
VP15 VP5V0
VN15 1 pos
GND Vss C16
1 2
5
4
1
1
DG445DY
1
C17 C18
STP/STR 100n/50V 10n/50V 100n/50V
2
2
R8
1
470E
R9
C19 A9
6003-068: Electrical diagram - main
8
5
6
TAB9 100n/50V 1 2
2
TP 2 V+B BS
1
+
"LAMP REF. CELL" 7 REF-STRT
LAMP REF CELL VN15 100E C20
PL3 GUARD 3
1
-
1p-coax R22 100E A6B 10n/50V V- G
2
6003-400-450-02
1 1 2 15 14 LT1011AI
4
1
IN OUT D2 S2
1
1
16 Logic
5
4
3
2
R-G[0..3] C38 1 pos C21
G[0..3] 1u/50V 10n/50V
2
2
1
DG445DY
C22
100n/50V
2
VP15 VP15
VP5V0 R30 1M00
VP5V0
1 2
VN15
GND
VN15 VN15
Sensor Amplifier
PCB N1
pcb_VPB2_R1B
Mounting holes
Powersupply A6D
VP5V0 TAB1 TAB2 VP15 7 6
PL2 TP A11 H1 H2 H3 D4 S4
F1
1
1
TP
1 1 2 1 11 3mm2 / 7mm5 3mm2 / 7mm5 3mm2 / 7mm5 8 Logic
+Vi +Vo "System BRD." 1 pos
2 24 14
1
+Vi +Vo DG445DY
Fuse holder (LT5) PL1
+ C24 NOT-MEAS/REF MEAS/NOT-REF
3 1 2
1
1
1
1
1
1
Header2 F2 Go 15u/25V STP/STR PH-STRT
Go 10 3 4
+ C25 + REF-STRT
15 5 6
2
V1 15u/25V Go P-G0 P-G1
1A T Go 22 7 8
C23 ES1C P-G2 9 10 P-G3
2
2
15u/25V R-G0 R-G1 H4 H5
12 2 11 12
2
1
Gi -Vo R-G2 R-G3 3mm2 / 7mm5 3mm2 / 7mm5
13 Gi -Vo 23 13 14
+ C26 15 16
1
TAB11 EC05-1502DM-S 15u/25V
TP Y1 Header-8x2
2
1
1
keying plug
R-G[0..3]
1
TAB3 H6 H7
TP VN15 P-G[0..3] 3mm2 / 7mm5 3mm2 / 7mm5
1
1
9-27
Electronics
VP5V0 VP5V0
9-28
1
R33
1K50
E
R34
A6C
2
10 11 1 2 B V5
D3 S3 BC857A
9 Logic V4 PMLL4448
Electronics
C
1 pos 1K50
2 1
DG445DY
K2
9402-05-00
1 2
GUARD
5 6
1 2
3 4
C35 6n8/50V
1 2 2 1
C36
1 2
22p/50V
VP15
1
C37 GUARD
100n/50V
2
7
8
GUARD A2
3 V+
+
6 OUT
IN 2 -
V-
OPA124U
4
1
5
1
R23
1
1K50
C39
2
R24 R25
100n/50V
2
1 2 2 1
1
1
6003-400-450-02
1
R26
3K00 R27
G[0..3] 16K0 R28
2
G[0..3] 47K0
2
2
G0
G1
G2
6003-068: Electrical diagram - photometer section
G3
9
10
A4C
D3
VP15
1 pos
Logic
S3
VP5V0
VP15
DG445DY
16
15
11
A4B
1
2
VP5V0 A4A
D2
VP15
D1
5 12
VL
C40
1
VP5V0 100n/50V
4 13
GND Vss
Vdd
1
1
2
1 pos
Logic
S2
GND 1 pos
Logic
S1
2
2
1
14
DG445DY
D4
R29
VN15 15K0
2
VN15
SELECTRA
1 pos
Logic
S4
DG445DY
6
PRO
M
VP5V0 VP5V0
1
R31
1K50
E
R32
A5C
2
10 11 1 2 B V3
D3 S3 BC857A
9 Logic V2 PMLL4448
C
1 pos 1K50
2 1
DG445DY
SELECTRA
K1
9402-05-00
1 2
GUARD
5 6
1 2
3 4
PRO
C27 6n8/50V
M
1 2 2 1
C28
1 2
22p/50V
VP15
1
C29 GUARD
100n/50V
2
7
8
GUARD A1
3 V+
+
6 OUT
IN 2 -
V-
OPA124U
4
1
5
1
R13
1
1K50
C31
2
R14 R15
100n/50V
2
1 2 2 1
1
1
6003-400-450-02
1
R16
3K00 R17
G[0..3] 16K0 R18
2
G[0..3] 47K0
2
2
G0
G1
6003-068: Electrical diagram - reference section
G2
G3
9
10
A3C
D3
VP15
1 pos
Logic
S3
VP5V0
VP15
DG445DY
16
15
11
A3B
1
2
VP5V0 A3A
D2
VP15
D1
5 12
VL
C32
1
VP5V0 100n/50V
4 13
GND Vss
Vdd
1
1
2
1 pos
Logic
S2
GND 1 pos
Logic
S1
2
2
1
14
DG445DY
D4
R19
VN15 15K0
2
VN15
1 pos
Logic
S4
DG445DY
6
9-29
Electronics
Electronics
SELECTRA PRO
9-30 6003-400-450-02
M
+5V +14V +30V
TP13 TP14 TP15 TP39 TP40 TP41 TP42
250 mA F3 +30V
PL1:A 1
PL1:B 2 3.15 A F2 1 PL3:A
PL1:C 3 2 PL3:B
PL1:D 4 3.15 A F1
PL1:E 5
PL1:F 6
C22 C8
SELECTRA
470uF 50V 470uF 50V
PRO
M
LAMP_ON
LEDG1
LEDG2
LEDG3
HEAT_PWR_EN
CV_H1
CV_H2
CUVET_TEMP_OK
LAMP_CONTROL
CUVET_THERM_CONTROL
NEEDLE_TEMP_OK
NEEDLE_THERM_CONTROL
+14V +30V
SCLK
R70 R68
MOSI
30.9k 30.9k
TP89 A_CS
TP94TP93 MISO
R23 4.75k
REF_2.5V C_CS
MEASURE_LAMP N_CS
MEASURE_30V
MEASURE_14V TP60 TP62 TP66 R42 10k
AMB_TEMP +5V
R44 10k
R47 10k
R71 R69
C58 4.75k 2.49k C57 TP79 TP84
0.1uF 0.1uF
SW1
R60 10k
SW2 +5V
R61 10k
SW3 +5V
WHEAT_EN
CV_L2
CV_L1
R84
6003-220: Electrical diagram - main
6003-400-450-02
R83
TP127 4.75k TP121 U10:C C50 LEDC
0.1uF TP78 TP81 TP85 LEDB
5 6
PL9:G 7 CTS_B LEDA
74AC14 C42
R85 0.0022uF
TP125 4.75k TP119 U10:A CTS
TXD
1 2
PL9:C 3 RXD_B RXD
C44 Y1
74AC14 22pF 7.3728 MHz
R82 R86
4.75k C63 4.75k C64
0.001uF 0.001uF +5V +5V C45
22pF U10:D
R92 R64 9 8
4.75k 4.75k TP72
TP140 TP86
PL9:A 1 REF_4V +5V 74AC14
+5V
PL9:D 4 U10:E
PL9:E 5 PL10:A 1 BKGD BKGD TP55 TP88 C40 11 10
PL9:F 6 PL10:B 2 0.1uF
PL9:H 8 PL10:C 3 NC
PL9:I 9 PL10:D 4 RESET RESET 74AC14
BDM
PL9:J 10 PL10:E 5 NC +5V U10:F
PL9:K 11 PL10:F 6 C51 13 12
+5V U11 0.001uF C52 C24 C53 +5V
0.01uF 0.01uF 0.01uF
VIN RST +5V
74AC14
GND
MC34064 R63
300
C32 C41 C33
TP87 0.1uF 0.1uF 0.1uF
REF_4V
U14 C49
LM4040BIM3-4.1 0.1uF
9-31
Electronics
9-32
+30V +30V
TP67 R51 TP47 R27
100K TP73 100K TP52
TP71 TP50
TP68 R50 Q12 TP48 R26 Q4
4.75k TP74 MMBT2907 4.75k TP53 MMBT2907
Q11 Q2
CV_H1 NTR4501NT1G CV_H2 NTR4501NT1G
Electronics
R54 R32
4.75k 4.75k
TP83 TP57
+30V C61
TP1280.01uF
TP96
TP122 L2
C66 C65 C62
TP136 TP139 68uH
4.7uF/50V 4.7uF/50V 4.7uF/50V D8
R88 C55 C56
MBRS540T3G
25.5k 33uF/25V 33uF/25V
R90 TP135
C67 9.31k Q13 Q7
1000pF
FDB031N08
FDB031N08
R91
+5V 2.49k D4 R62 R37 D3
TP145 MMBZ5232 4.75k C46 C26 4.75k MMBZ5232
1000pF 1000pF
TP11 TP12
C_CS TP141
SCLK CUV_CTRL 3 PL2:C
4 PL2:D
MOSI +5V C69
100pF
C68 R53 R31
0.1uF TP70 499 TP76 Q14 Q8 TP51 499 TP49
CV_L1 FDB031N08
FDB031N08 CV_L2
R52 R28
10k 10k
TP45 TP25
+5V L1 +5VA +5VA REF_2.5V
1 2
10uH
C9 C10 C18 D1 C11
47uF/25V 47uF/25V 0.1uF MMBZ5232 1uF
6003-400-450-02
REF_2.5V REF_2.5V
C1
0.1uF
+5VA +5V R2
6003-220: Electrical diagram - cuvette driver
C6
0.1uF
REF_2.5V
R6
6.98k 0.1% C4
0.1uF
SELECTRA
W_THERM
R7
TP20 10k 0.1% C3
0.1uF
PRO
M
+14V R72 TP102 Q15 TP82
0.012 1/4W FDB031N08
2 PL2:B
SELECTRA
+5V +14V
+14V
TP114 D5 C27
+30V U16 0.1uF R41
4.75k
VIN VOUT R56
ADJ BAT54C C47 1k +5V R40 TP59 R35
PRO
C60 R78 C59 0.1uF 10k 30.9k
0.1uF LM317 1uF/35V 5
M
499 4 R59
TP115 U7 - 87.6k 0.1%
V+
1 TP56
TP75
V-
R79 3
+ MEASURE_LAMP
7.87k 2
R58 R36
U9 10k 0.1% C43 C25 4.75k
TP69 R46 TP63 C29 NL 0.1uF
TP80 TP77 LM4041BIM3-1.2 0.1uF
R57 30.9k
10k
LAMP_ON
C31
0.01 uF
R55
10k
74AC14
TP2 TP26 U3:B TP36 REF_2.5V
3 4
PL7:B 2 LAMP_CONTROL
10k R1
6003-400-450-02
R17 R15 R14 74AC14 10k
10k 10k C13 C15 TP10
0.001uF 0.001uF
6003-220: Electrical diagram - lamp driver
+5V +5V
AMB_TEMP 1 PL4:A
R20 R19
10k 10k 1k 2 PL4:B
C2 R3 3 PL4:C
0.001uF 4 PL4:D
TP3 R13
100 TP16 U3:C TP30
6 5
PL7:C 3 NEEDLE_TEMP_OK
74AC14
TP4 R12
100 TP17 U3:D TP31
8 9
PL7:D 4 CUVET_TEMP_OK
74AC14
R11
TP5 10k TP18 U3:E TP27
11 10
PL7:E 5 CUVET_THERM_CONTROL
PL7:F 6 C7 74AC14
R4
10k 0.001uF
+5V
C12
0.1uF
U3:F
13 12
74AC14
9-33
Electronics
+5V
9-34
+5V R67
10k
R38
10k
TP6
TP61
Q10 N_THERM 1 PL6:A
NTR4501NT1G 2 PL6:B
Electronics
8
2 U5:A R66
- 2.49k
1
3 +5V
N_THERM +
4
TP90
TP38 +5V
N_CS TP91
C21
0.1uF SCLK
C54
MOSI 100pF
+5V
+5V
C48
0.1uF
8
2 U1:A
-
1
3
W_THERM +
4
+5V R5
100k
R18 C14
C16 10k 0.1uF
1uF
TP23
W_THERM 1 PL8:A
+30V TP21 2 PL8:B
W_PWR_CTRL
N_PWR_CTRL
3
FQB34P10TM
PL8:C
Q1
+30V_SW
4 PL8:D
5 PL8:E
6 PL8:F
6003-400-450-02
R24 D2 C17
TP95 10k MMBZ5232 Not Loaded TP32
REF_2.5V TP35
+5V +5V
6003-220: Electrical diagram - needle heater
+ TP44 TP54
8
2 U8:A
Q9 -
MMBT2222 1
TP64
Q5 3
NTR4501NT1G +
4
R29
0.33 R43 R48
6 U1:B 10k 10k +5V
W_THERM -
7 C36
0.1uF
5
+
+5V TP24
R30
4.75k
6 U8:B
SELECTRA
-
TP46 Q6 7
NTR4501NT1G
HEAT_PWR_EN
5
+
PRO
M
+5V
R45
4.75k
TP131
TP113 +5V
SW1
LEDA S3
SELECTRA
BSS84 KSC321G
Q20 C30
0.001uF
TP112
TP108 +5V
R77
LEDB
300
BSS84
PRO
Q18 +5V
M
TP107
TP123 +5V R39
R75 4.75k
LEDC
300
BSS84 TP111
Q22
SW2
TP118
TP117 +5V S2
R81 KSC321G
LEDD C28
300
BSS84 0.001uF
Q21
TP116
TP97 +5V
R80
LEDE +5V
300
BSS84
Q16 R34
4.75k
TP99
TP106 +5V
R73 TP98
LEDF
300
BSS84 SW3
Q17 S1
KSC321G
TP103
TP110 +5V C23
R74 TP124 TP134 TP132 TP104 TP100 TP105 TP101 0.001uF
LEDG
300
BSS84
Q19
TP109
R76
300
6003-400-450-02
TP130 +5V
6003-220: Electrical diagram - adjustment panel
LEDDP
BSS84
Q23
TP129
R87
300
TP133
TP144
TP142
TP137
9-35
Electronics
Electronics
SELECTRA PRO
9-36 6003-400-450-02
M
Electronics
SELECTRA PRO
M 6003-400-450-02 9-37
Electronics
T2 MOSFET SI4884DY
T3 MOSFET SI4836DY
T4 MOSFET BTS3134N
T5 TRANS. BC847
X1 CRYSTAL 6MHz
SELECTRA PRO
9-38 6003-400-450-02
M
U11 U8
SELECTRA
L M 1117M 5V L M 3940 100u/6V 3/ESR-0.5 Peltier
F4
3 2 1 3 Peltier.sch K3
12V V in V out V IN V OUT 3V 3
4
multifuse 200mA V out TEC_PW M TEC_K 1 1
GN D
TEC_K 2 2
1
GN D
GN D
R31 C32 DS_PW R TEC 1
10k 100n C26 C28 C29 C27 DS_PROG
K 10
1
100n 10u/16V 100n
PRO
2
4
12V
I_M EA S 1
2
V _M EA S 2
M
K1 GND
TEC_OFF TEC 2
GND
3 TEC_FA IL
2 R16 CPU
1 2k2 CPU.sch
power D6
TV S_SM PW R_M EA S TEC_PW M
PGND 3V 3
DS_PW R
DS_PROG D2 R22
GND GND BA V 23S 10k K4
I_M EA S R24 R23
3V 3
V _M EA S 1
10k 10k SW ITCH
TEC_OFF 2
R27 TEC_FA IL C20
K2 10k 10n/100v GND
R35
3
330E GND 12V
2 ENA BL E
1
8
1 A L A RM COV ER_OPEN
Host_Comm U10A K5
C23 OPTO_M OCD213 PUM P_FA N_OFF C21
10n/100v R30 PUM P_FA N_L 10n/100v 2
PUM P
PUM P_FA N_CHK 1
2
7
10k
GND
L ED2 R15 T4
5V
L ED3 3k3
TSENS_PW M protect
R36 FET_N_HITFET_SOT223
330E TX D GND R25 K6
RX D 10k C22
10n/100v 2
6003-007: Electrical diagram - part 1
FA N
1
6
C24 U10B GND
6003-400-450-02
10n/100v OPTO_M OCD213
U4A
16 1 5V 5V
5
OUT IN
8 9
GND COM
DRV R_UL N2003_DIS
GND 5V R33 R34
330E 330E
1
R29
U7 470E
U9 SM T160-30 D4 D5
R42 1 2 ROOD GROEN
Output V cc
2 1 5K 6
V+ C1+ U4F U4E
GN D
6
2
3V 3 IN OUT IN OUT
100n 16 3 100n 10K 100n
V CC C1- DRV R_UL N2003_DIS DRV R_UL N2003_DIS
3
6 4
V- C2+
C12 C11 C9 GND
100n 10 0n 15 5 100n GND
GND GND C2-
K7
14 11
3 T1OUT T1IN
2
7 10
1 T2OUT T2IN
RS232- 13 12
R1IN R1OUT
GND
8 9
R2IN R2OUT
M A X 3232CW E
9-39
Electronics
9-40
U2 U3
Electronics
DS1821 DS1821
3V 3 3V 3
GN D
DQ
V CC
GN D
DQ
V CC
DS_PW R
R38 R39
4k7 4k7
1
2
3
1
2
3
3V 3
12V D7 2
3 U4B C3 U4C C4
F1 L1 PGND R5 100n 100n
1 15 2 14 3
4k7 OUT IN OUT IN
1.2uH 6CW Q03FN
multifuse 6A
DRV R_UL N2003_DIS DRV R_UL N2003_DIS
C14 C34
5V
D1 C2
R1 GND
PGND PGND
10E 94SV P566X 0016E7 DS_PROG
BA T54 100n
U4D C1 94SV P566X 0016E7
4 13 2.2uF T1
TEC_OFF IN OUT IRF7467
U1
3
2
1
7
PGND SI4836DY U6
6003-007: Electrical diagram - part 2
6003-400-450-02
A M P_A D820
C5 R13 V- 5V A M P_L M V 358
C15 C35 100n 0.005E R10
4
1
R14
3k6
GND GND C7
150 100n R18
7k5
GND
94SV P566X 0016E7
PGND 94SV P566X 0016E7 C36
4n7
GND
5V
R11 R32
5V 3k6 680
R12
8
15k
U5A
3 R7
1 R19 1k2
R20 2
TEC_PW M 7k5
120k
A M P_L M V 358 GND
4
5V
SELECTRA
PRO
M
SELECTRA
PRO
M
3V 3
R46
100E
3V 3 3V 3
DS_PW R
R26 R40 T5
10K 10K R44 NPN_BC847
10K
U12 R45
21 10K
L ED2 P1.0/TA CL K /A DC10CL K
P1_1 22
L ED3 P1.1/TA 0
TEC_PW M 23
P1.2/TA 1
TSENS_PW M 24
JTA G_TCK P1.3/TA 2 GND
25
JTA G_TM S P1.4/SM CL K /TCK
26
JTA G_TDI P1.5/TA 0/TM S
27
JTA G_TDO P1.6/TA 1/TDI
28
P1.7/TA 2/TDO/TDI
8
PW R_M EA S P2.0/A CL K /A 0
I_M EA S 9
P2_2 P2.1/INCL K /A 1
V _M EA S 10
P2.2/TA 0/A 2
DS_PROG 19
P2.3/TA 1/A 3/V ref-/V eref-
20
P2.4/TA 2/A 4/V ref+/V eref+
3
A L A RM P2.5/Rosc K8
JTA G_TDO
1
P1_1
2
11 JTA G_TDI
COV ER_OPEN P3.0/STE0/A 5 3
6003-007: Electrical diagram - part 3
12 P2_2
ENA BL E P3.1/SIM O0 4
13 JTA G_TM S
TEC_FA IL P3.2/SOM I0 5
14 X OUT
PUM P_FA N_CHK P3.3/UCL K 0 6
15 JTA G_TCK 7
TX D P3.4/UTX D0
6003-400-450-02
16 JTA G_TEST
RX D P3.5/URX D0 8
17 9
TEC_OFF P3.6/A 6 GND
18 3V 3 RST
PUM P_FA N_OFF P3.7/A 7 R43 10
3V 3
2 33K ISP
V cc
JTA G_TEST 1
R37 TEST
R47 33K C13 GND
120k RST 7 100n
RST/NM I
4
V ss
GND 6
X IN
GND
X1
X OUT 5
X OUT
6M Hz CPU_M SP430X 12X 2
C30 C31
16pF 16pF
GND GND
9-41
Electronics
Electronics
SELECTRA PRO
9-42 6003-400-450-02
M
Electronics
R6
560
PL4 OC4
1 30
30 H21B1
2
3 30
4 30
VCC
OC3
H21B1
PL3
1 STM1
2 STM2
3 STM3
4 STM4
VCC
5 40
VCC
6
7
8
9
560
10
11 R1
12 R.OPTO-R
13 R.OPTO-S
14 L.OPTO-R
L.OPTO-S OC2
15
16 +14V
VALVE_A H21B1
17
18 VALVE_B
19 STM5
STM6 VCC
20
21 STM7
STM8 OC1
22
H21B1
MF1 PL1
30 1
2
0.65 A R3 3
+14V
4.7
PL2 4
1 30
C4
2 30 C2
22n
3 30 IC1 9
22n
4 30 12 IN1 VS B1 11
16 IN2 O1 10
2 E MF2
R4 R5 VCC
10k
10k
O2 8 30
18 VR B2 17 0.65 A R2
4.7
S
C5 L6202
1
220n 22n
C3 C1
22n
SELECTRA PRO
M 6003-400-450-02 9-43
Electronics
SELECTRA PRO
9-44 6003-400-450-02
M
Electronics
SELECTRA PRO
M 6003-400-450-02 9-45
Electronics
SELECTRA PRO
9-46 6003-400-450-02
M
Electronics
SELECTRA PRO
M 6003-400-450-02 9-47
Electronics
SELECTRA PRO
9-48 6003-400-450-02
M
Electronics
SELECTRA PRO
M 6003-400-450-02 9-49
9-50
VCC
C11 C9
C10 C8 15μ 1n IC2 9
12 IN1 VS B1 11 C14
15μ 1n SA STIRRER
Electronics
22n MF1
C15
16 IN2 O1 10
1n SP1
C18 VCC
2 E 0.2 A R3
1n 8
O2 150
18 VR B2 17
C12 S C1
1
L6202 C13 1n
220n
22n SP2
PL1
1 SIPPER-ARM
2
3 SA_STIRRER
MF2 RA STIRRER
4
5 RA_STIRRER SP3
6 0.2 A R5
7 OPTO
150
3 4
8
9 VCC
RN1.B RN1.C
4k7
4k7
10 +14V C17
6001-773: Electrical diagram
11 +30V
+30V 1 1 1n
12
SP4
C4 C3
6003-400-450-02
1n IC1 9
47μ C7
12 IN1 VS B1 11 ISE-SIPPERARM
22n MF3
C16
16 IN2 O1 10
2 1n SP5
VCC
R1 2 E 0.65 A R4
RN1.A
4k7
1k
33
O2 8
C6
1 18 17 22n
VR B2
C5 S C2
1
OC1 220n L6202 4n7
SP6
H22B1
SELECTRA
R2
4k7
PRO
M
appendix A
Technical Specifications
SELECTRA PRO
M 6003-400-450-02 A-1
Technical Specifications
Performance
Maximum throughput Mono mode: 180 tests/hour (360 with Dry Electrode ISE unit)
Dual mode: 133 tests/hour (333 with Dry Electrode ISE unit)
Sample system
Syringe 100 μl
• All positions can be used for patient samples (normal, ASAP and STAT), controls
and calibrators. Normally, controls and calibrators are placed on the inner ring and
patient samples on the two outer rings (readable by the internal barcode reader).
Reagent system
SELECTRA PRO
A-2 6003-400-450-02
M
Technical Specifications
Syringe 1000 μl
Measurement station
Wavelengths (2 options) Option 1: 340, 405, 505, 546, 578, 620, 660 and 700 nm
Option 2: custom
Wavelength uncertainty ± 2 nm
Temperature 37 °C ± 0.2 °C
Approvals
CE
CB
UL
IEC 61326-2-6 The analyzer complies with the emission and immunity
requirements described in the IEC 61326-2-6.
CISPR 11 Class A This equipment has been designed and tested to CISPR 11
Class A. In a domestic environment it may cause radio
interference, in which case, you may need to take measures to
mitigate the interference.
Info
i The approvals listed here refer only to the instrument and operator console,
not to additional devices. For the approvals for these devices, see the
corresponding manuals.
SELECTRA PRO
M 6003-400-450-02 A-3
Technical Specifications
Serial ports 2 x RS232 (one for analyzer, one for host connection or printer)
Technology CCD
Barcodes Code 39
Code 93
Code 128
Codabar
Code 2 of 5 family
Info
i Other barcodes can be configured when required. A total of 5 barcodes can
be programmed for the internal barcode reader. Contact the system support
department of Vital Scientific if you need more information.
Amount of characters Max. 11 (Code 39, Code 93, Code 128, Codabar)
Info
i For Codabar barcodes the start and stop characters must be the same.
SELECTRA PRO
A-4 6003-400-450-02
M
Technical Specifications
Info
i Instructions on correctly attaching barcode labels to sample tubes are
included in the user manual.
Technology CCD
Barcodes Code 39
Code 128
Codabar
Code 2 of 5 family
Info
i Other barcodes can be configured when required. See the configuration
manual that came with the hand-held barcode reader. Contact the system
support department of Vital Scientific if you need more information.
SELECTRA PRO
M 6003-400-450-02 A-5
Technical Specifications
Width 125 cm
Depth 62 cm
Power requirements
Environmental requirements
Degree of protection IP 20
! CAUTION
TAKE CARE
Do not use the analyzer in close proximity to sources of strong
electromagnetic radiation (e.g. unshielded intentional RF sources), as these
may interfere with proper functioning of the analyzer.
Info
i Vital Scientific advises the user to evaluate the electromagnetic environment
prior to operation of the Selectra ProM.
SELECTRA PRO
A-6 6003-400-450-02
M
Technical Specifications
Introduction
All cleaning solutions are EC marked according to directive 98/79/EC. They are for
professional use only and intended for in vitro diagnostic devices.
Undiluted cleaning solutions are stable between 15 and 25 ºC (59 - 77 ºF) until the expiry
date stated on the package label. Diluted solutions are stable at least 2 weeks on board of
the analyzer.
NOTICE
PLEASE NOTE
i Cloudy solutions must be discarded.
Respect local regulations concerning waste disposal.
System solution
Reference SLSY-5900
Manufacturer SEPPIM
Intended use Washing of cuvettes and needles; also used for the pipetting
system (wetting effect for fluidified system).
Preparation Dilute with distilled water at 1:400
(12.5 ml system solution on 5 liters of distilled water)
Composition Aqueous solution with a detergent containing sodium azide
(< 0.1%).
Precautions None.
Reference SLNA-5900
Manufacturer SEPPIM
Intended use Cleaning of pipetting needles; also used in the system
cleaning procedure (see par. 6.2.1).
Preparation Dilute with distilled water at 1:10 for needle rinsing;
Use undiluted for the system cleaning procedure.
Composition Aqueous solution of Sodium hypochlorite (< 2% of active
chlorite) and Sodium hydroxide (< 5.4%).
Precautions Corrosive (C)
R 35 Causes severe burns
S 26 In case of contact with eyes, rinse immediately with plenty of
water and seek medical advice.
S 28 In case of contact with skin, wash immediately with plenty of
water.
S 36 / 37 / 39 Wear suitable protective clothing, gloves and eye / face
protection.
S 45 In case of accident or if you feel unwell, seek medical advice
immediately (if possible show the package label).
SELECTRA PRO
M 6003-400-450-02 A-7
Technical Specifications
Acid solution
Reference SLHC-5900
Manufacturer SEPPIM
Intended use Cleaning of the reagent needle; also used in the incompatibility
procedure of tests.
Preparation Solution is ready to use.
Composition Aqueous solution of hydrochloric acid (< 0.5%).
Precautions Irritant (Xi)
R 36 / 37 / 38 Irritating to eyes, respiratory system and skin
S 26 In case of contact with eyes, rinse immediately with plenty of
water and seek medical advice.
SELECTRA PRO
A-8 6003-400-450-02
M
Technical Specifications
The chemical performance of clinical chemistry analyzers, in terms of accuracy and precision,
depends on the characteristics of the instrument, the measurement techniques and the
materials used. Therefore, the chemical performance characteristics of a clinical chemistry
analyzer can only be established and postulated in terms of: the analyte; the specific reagent
kit and calibrator(s) used; the type and constitution of the specimens involved; etc.
The analyzers manufactured by Vital Scientific may be supplied with open channels. ‘Open’
implies that most clinical chemistry tests and techniques that require photometric
measurement, can be adapted on the system. Only the test parameters for a specific test
need to be adjusted. The user needs to establish the required test parameter settings to
achieve satisfactory results, utilizing appropriate methods. The methods are preferably based
on international guidance documents, for example ECCLS or CLSI guidelines. Vital Scientific
recommends the use of ELITech reagents, calibrators and/or controls on their analyzers.
Obtain information on the performance characteristics from the selected reagent distributor
and/or manufacturer. Various reagent manufacturers have performed performance studies on
the analyzers from Vital Scientific in combination with their reagent kits. Therefore, they have
application sheets available for various analytes. The required information usually can be
obtained from the reagent package inserts. Please contact your local representative and/or
reagent manufacturer for further information on the chemical performance of their reagents on
analyzers from Vital Scientific.
Disclaimer
Vital Scientific assumes no responsibility for erroneous test results caused by reagent kits
and/or test parameters that are not explicitly provided or recommended by Vital Scientific.
SELECTRA PRO
M 6003-400-450-02 A-9
Technical Specifications
SELECTRA PRO
A-10 6003-400-450-02
M
8003-400-154-00 TUBING TABLE ProM
from to tube
tube # qt part location / assy part location / assy codeno length marking
1 1 4104-103 Pipe, cuvette wash place 1 suction 8001-961 Needle plate unit 3066-089 Thread connector, farrest from outlet 8003-388 Connection block 1513-010 580 1, 1
1 1 4104-103 Pipe, cuvette wash place 1 suction 8001-961 Needle plate unit 3066-088 Straight through reduction connector 8003-325 CW connection block assy 1513-010 580 1, 1
2 1 4104-103 Pipe, cuvette wash place 2 suction 8001-961 Needle plate unit 3066-089 Thread connector, next to previous 8003-388 Connection block 1513-010 580 2, 2
SELECTRA
2 1 4104-103 Pipe, cuvette wash place 2 suction 8001-961 Needle plate unit 3066-089 Thread connector, farrest from outlet 8003-326 CW connection block 1513-010 580 2, 2
3 1 4104-103 Pipe, cuvette wash place 3 suction 8001-961 Needle plate unit 3066-089 Thread connector, next to previous 8003-388 Connection block 1513-010 580 3, 3
3 1 4104-103 Pipe, cuvette wash place 3 suction 8001-961 Needle plate unit 3066-089 Thread connector, next to previous 8003-326 CW connection block 1513-010 580 3, 3
4 1 4104-103 Pipe, cuvette wash place 4 suction 8001-961 Needle plate unit 3066-089 Thread connector, next to previous 8003-388 Connection block 1513-010 580 4, 4
4 1 4104-103 Pipe, cuvette wash place 4 suction 8001-961 Needle plate unit 3066-089 Thread connector, next to previous 8003-326 CW connection block 1513-010 580 4, 4
5 1 4104-103 Pipe, cuvette wash place 5 suction 8001-961 Needle plate unit 3066-089 Thread connector, next to previous 8003-388 Connection block 1513-010 580 5, 5
PRO
5 1 4104-103 Pipe, cuvette wash place 5 suction 8001-961 Needle plate unit 3066-089 Thread connector, next to previous 8003-326 CW connection block 1513-010 580 5, 5
M
6 1 4104-102 Pipe, cuvette wash place 6 dryblock 8001-961 Needle plate unit 3066-089 Thread connector, next to previous 8003-388 Connection block 1513-010 580 6, 6
6 1 4104-102 Pipe, cuvette wash place 6 dryblock 8001-961 Needle plate unit 3066-089 Thread connector, next to previous 8003-326 CW connection block 1513-010 580 6, 6
7 1 4104-110 Pipe 8001-963 Filter block assy 4105-243 Bush 8001-961 Needle plate unit 1513-010 33
8 1 3066-095 Female Luer connector, pumps-valves unit port 4 8003-391 NOCW pumps-valves unit Valve V4 port 1 8003-390 NOCW valves assy 1513-001 150 red
9 1 3066-052 Thread connector, outlet 8003-388 Connection block 3066-093 Male Luer connector, pumps-valves unit port 4 8003-391 NOCW pumps-valves unit 1513-001 570 4, red
9 1 3066-052 Thread connector, outlet 8003-326 CW connection block 3066-093 Male Luer connector, pumps-valves unit port 4 8003-342 CW pumps-valves unit 1513-001 570 4, red
10 1 Valve V5 port 3 8003-390 NOCW valves assy Valve V4 port 2 8003-390 NOCW valves assy 1513-001 65
11 1 3066-118 Straight through connector 8003-299 Bellows pump unit 4104-084 Pipe, cuvette wash place 1 water 8001-961 Needle plate unit 1513-010 700 11, 11
12 1 3066-118 Straight through connector 8003-299 Bellows pump unit 4104-084 Pipe, cuvette wash place 2 water 8001-961 Needle plate unit 1513-010 700 12, 12
13 1 3066-118 Straight through connector 8003-299 Bellows pump unit 4104-084 Pipe, cuvette wash place 3 water 8001-961 Needle plate unit 1513-010 700 13, 13
14 1 3066-118 Straight through connector 8003-299 Bellows pump unit 4104-084 Pipe, cuvette wash place 4 water 8001-961 Needle plate unit 1513-010 700 14, 14
15 1 Valve V4 port 3 8003-390 NOCW valves assy Valve V2 port 3 8003-390 NOCW valves assy 1513-001 65
16 1 3066-091 Press-in plug 8003-390 NOCW valves assy Valve V2 port 1 8003-390 NOCW valves assy 1513-009 25
17 1 Valve V2 port 2 8003-390 NOCW valves assy 3066-095 Female Luer connector, pumps-valves unit port 3 8003-391 NOCW pumps-valves unit 1513-001 140 red
18 1 3066-093 Male Luer connector, pumps-valves unit port 3 8003-391 NOCW pumps-valves unit 3066-052 Thread connector, collecting cask port B 8003-385 Collecting cask unit 1513-001 335 3, red
19 1 3066-088 Straight through reduction connector 8003-325 CW connection block assy 3066-093 Male Luer connector, pumps-valves unit port 8 8003-342 CW pumps-valves unit 1513-001 505 8, black
20 1 3066-095 Female Luer connector, pumps-valves unit port 8 8003-342 CW pumps-valves unit Valve V14 port 3 8003-341 CW valves assy 1513-001 140 black
21 1 3066-091 Press-in plug 8003-341 CW valves assy Valve V14 port 1 8003-341 CW valves assy 1513-009 25
22 1 Valve V14 port 2 8003-341 CW valves assy 3066-095 Female Luer connector, pumps-valves unit port 7 8003-342 CW pumps-valves unit 1513-001 150 black
23 1 3066-093 Male Luer connector, pumps-valves unit port 7 8003-342 CW pumps-valves unit 3066-052 Thread connector, collecting cask port C 8003-321 CW collecting cask unit 1513-001 445 7, black
24 1 3066-139 Thread connector, collecting cask port A 8003-385 Collecting cask unit 3066-094 Male Luer connector, pumps-valves unit port 2 8003-391 NOCW pumps-valves unit 1513-024 300 2, green
25 1 3066-153 Female Luer connector, pumps-valves unit port 2 8003-391 NOCW pumps-valves unit Valve V1 port 3 8003-390 NOCW valves assy 1513-009 145 green
26 1 Valve V1 port 1 8003-390 NOCW valves assy 3066-050 Tee connector 8003-339 Vacuum switch assy 1513-001 315
27 1 3344-004 Vacuum switch 8003-339 Vacuum switch assy 3066-087 Straight through reduction connector 8003-339 Vacuum switch assy 1513-002 30
28 1 3066-087 Straight through reduction connector 8003-339 Vacuum switch assy 3066-050 Tee connector 8003-339 Vacuum switch assy 1513-001 255
29 1 3066-050 Tee connector 8003-339 Vacuum switch assy 3066-052 Thread connector 8002-754 Vacuum tube assy 1513-001 65
30 1 3066-052 Thread connector 8002-754 Vacuum tube assy Valve V11 port 1 8003-390 NOCW valves assy 1513-001 360
31 1 Valve V11 port 3 8003-390 NOCW valves assy 3066-087 Straight through reduction connector 8003-340 Vacuum pump assy 1513-001 290
6003-400-450-02
32 1 3066-087 Straight through reduction connector 8003-340 Vacuum pump assy 3066-092 Elbow connector 8003-340 Vacuum pump assy 1513-002 28
33 1 3066-092 Elbow connector 8003-340 Vacuum pump assy 3365-014 Vacuum pump, suction inlet 8003-340 Vacuum pump assy 1513-002 25
34 1 3365-014 Vacuum pump, pressure outlet 8003-340 Vacuum pump assy 3066-092 Elbow connector 8003-340 Vacuum pump assy 1513-002 25
35 1 3066-092 Elbow connector 8003-340 Vacuum pump assy 3066-146 Straight through reduction connector 8003-340 Vacuum pump assy 1513-002 28
36 1 3066-146 Straight through reduction connector 8003-340 Vacuum pump assy Valve V12 port 3 8003-390 NOCW valves assy 1513-009 255
37 1 Valve V12 port 2 8003-390 NOCW valves assy Valve V1 port 2 8003-390 NOCW valves assy 1513-009 65
38 1 Valve V12 port 1 8003-390 NOCW valves assy 3066-153 Female Luer connector, pumps-valves unit port 1 8003-391 NOCW pumps-valves unit 1513-009 140 white
39 1 3066-094 Male Luer connector, pumps-valves unit port 1 8003-391 NOCW pumps-valves unit 3066-096 Female Luer connector, connection plate port 3 8003-389 NOCW connection plate 1513-002 730 white, 1, 3, red
40 1 3066-090 Thread connector, Valve V3 port 2 8003-385 Collecting cask unit 3066-092 Elbow connector 8003-385 Collecting cask unit 1513-002 40
41 1 3066-092 Elbow connector 8003-385 Collecting cask unit 3066-096 Female Luer connector, connection plate port 4 8003-389 NOCW connection plate 1513-002 370 4, red
42 1 3066-090 Thread connector, Valve V13 port 2 8003-321 CW collecting cask unit 3066-092 Elbow connector 8003-321 CW collecting cask unit 1513-002 30
43 1 3066-092 Elbow connector 8003-321 CW collecting cask unit 3066-096 Female Luer connector, connection plate port 6 8003-322 CW connection plate 1513-002 380 6, red
44 1 3066-052 Thread connector, top right 8001-786 Watercask 3066-095 Female Luer connector, connection plate port 2 8003-389 NOCW connection plate 1513-001 570 2, blue
45 1 3066-095 Female Luer connector, connection plate port 1 8003-389 NOCW connection plate 3066-093 Female Luer connector, pumps-valves unit port 5 8003-391 NOCW pumps-valves unit 1513-001 565 blue, 1, 5, blue
46 1 3066-095 Female Luer connector, pumps-valves unit port 5 8003-391 NOCW pumps-valves unit 3066-087 Straight through reduction connector 8003-338 Water pump assy 1513-001 115 blue
47 1 3066-087 Straight through reduction connector 8003-338 Water pump assy 3365-021 Water pump, inlet 8003-338 Water pump assy 1513-002 26
48 1 3365-021 Water pump, outlet 8003-338 Water pump assy 3066-087 Straight through reduction connector 8003-338 Water pump assy 1513-002 26
49 1 3066-087 Straight through reduction connector 8003-338 Water pump assy 3066-095 Female Luer connector, pumps-valves unit port 6 8003-391 NOCW pumps-valves unit 1513-001 95 blue
50 1 3066-093 Female Luer connector, pumps-valves unit port 6 8003-391 NOCW pumps-valves unit 3066-052 Thread connector, top front 8001-786 Watercask 1513-001 145 blue, 6
51 1 3066-052 Thread connector, bottom 8001-786 Watercask 3066-050 Tee connector 8003-320 W-OUT tube assy 1513-001 840
52 1 3066-050 Tee connector 8003-320 W-OUT tube assy 3066-095 Female Luer connector 8003-297 Bellows valve assy 1513-001 580
53 1 3066-128 Male Luer connector 8003-297 Bellows valve assy 3066-117 Tee connector 8003-297 Bellows valve assy 1513-009 70 blue
54 3 3066-117 Tee connector 8003-297 Bellows valve assy 3066-117 Tee connector 8003-297 Bellows valve assy 1513-009 27
55 1 3066-117 Tee connector 8003-297 Bellows valve assy Valve V6 port 1 8003-297 Bellows valve assy 1513-009 46
56 4 3066-117 Tee connector 8003-297 Bellows valve assy Valves V7..V10 port 1 8003-297 Bellows valve assy 1513-009 20
57 5 Valves V6..V10 port 3 8003-297 Bellows valve assy 4104-085 Pipe, bellows port 8001-839 Bellows pump 1513-009 95
58 1 Valve V6 port 2 8003-297 Bellows valve assy 3066-118 Straight through connector 8003-297 Bellows valve assy 1513-009 190 S
59 1 Valve V7 port 2 8003-297 Bellows valve assy 3066-118 Straight through connector 8003-297 Bellows valve assy 1513-009 160 11
60 1 Valve V8 port 2 8003-297 Bellows valve assy 3066-118 Straight through connector 8003-297 Bellows valve assy 1513-009 130 12
61 1 Valve V9 port 2 8003-297 Bellows valve assy 3066-118 Straight through connector 8003-297 Bellows valve assy 1513-009 100 13
62 1 Valve V10 port 2 8003-297 Bellows valve assy 3066-118 Straight through connector 8003-297 Bellows valve assy 1513-009 70 14
63 1 3066-050 Tee connector 8003-320 W-OUT tube assy 3066-050 Tee connector 8003-324 Dispenser unit 1513-001 100
A-11
Technical Specifications
A-12
64 1 3066-050 Tee connector 8003-324 Dispenser unit 4105-215 Bush in 3066-072 Valve block reagent, left port 8003-324 Dispenser unit 1513-001 150
65 1 3066-050 Tee connector 8003-324 Dispenser unit 4105-215 Bush in 3066-072 Valve block sample, left port 8003-324 Dispenser unit 1513-001 55
66 1 4100-683 Bush 8003-336 Reagent tube assy 3370-053 Feed through 8003-336 Reagent tube assy 1513-001 1210 R
67 1 4105-214 Bush in 3066-072 Valve block reagent, right port 8003-324 Dispenser unit 4100-456 Needle 8002-939 Reagent needle unit 1503-008 1500
68 1 4105-305 Bush 8003-347 Bracket assy reagent 4100-456 Needle 8002-939 Reagent needle unit 1513-002 245
69 1 4100-683 Bush 8003-337 Sample tube assy 3370-053 Feed through 8003-337 Sample tube assy 1513-001 930 S
70 1 4105-214 Bush in 3066-072 Valve block sample, right port 8003-324 Dispenser unit 4100-456 Needle 8002-007 Sample needle unit 1503-007 1215
71 1 4105-305 Bush 8003-359 Bracket assy sample 4100-456 Needle 8002-007 Sample needle unit 1513-002 245
72 1 4104-085 Pipe, rinsing place reagent needle outlet, low 8003-355 Measuring unit 3066-088 Straight through reduction connector 8003-355 Measuring unit 1513-009 50
73 1 3066-088 Straight through reduction connector 8003-355 Measuring unit 3066-093 Male Luer connector, pumps-valves unit port 10 8003-391 NOCW pumps-valves unit 1513-001 870 R, 10, green
74 1 3066-095 Female Luer connector, pumps-valves unit port 10 8003-391 NOCW pumps-valves unit Valve V5 port 2 8003-390 NOCW valves assy 1513-001 155 green
75 1 3066-118 Straight through connector 8003-299 Bellows pump unit 4104-089 Pipe, rinsing place sample needle inlet, side 8003-355 Measuring unit 1513-009 545 S, S
Technical Specifications
76 1 4104-085 Pipe, rinsing place sampe needle outlet, low 8003-355 Measuring unit 3066-088 Straight through reduction connector 8003-355 Measuring unit 1513-009 50
77 1 3066-088 Straight through reduction connector 8003-355 Measuring unit 3066-093 Male Luer connector, pumps-valves unit port 9 8003-391 NOCW pumps-valves unit 1513-001 670 S, 9, white
78 1 3066-095 Female Luer connector, pumps-valves unit port 9 8003-391 NOCW pumps-valves unit Valve V5 port 1 8003-390 NOCW valves assy 1513-001 155 white
79 1 3066-164 Panel mount reduction connector 8003-378 Dry ISE module 3066-096 Female Luer connector, connection plate port 5 8003-389 NOCW connection plate 1513-002 1220 5, red
80 1 3066-052 Thread connector 8001-620 Water filter 4104-080 Pipe 8002-904 5L water container 1513-001 300
81 1 4104-080 Pipe, outlet 8002-904 5L water container 3066-093 Male Luer connector 8002-904 5L water container 1513-001 300 1, blue
82 1 3066-093 Male Luer connector 8002-904 5L water container 4104-080 Pipe, inlet 8002-904 5L water container 1513-001 300 blue, 2
83 1 3066-094 Male Luer connector 8003-282 Diluted waste container 4104-104 Pipe 8003-282 Diluted waste container 1513-002 300 red, 3
84 1 3066-094 Male Luer connector 8003-282 Diluted waste container 4104-104 Pipe 8003-282 Diluted waste container 1513-002 300 red, 4
85 1 3066-094 Male Luer connector 8003-282 Diluted waste container 4104-104 Pipe 8003-282 Diluted waste container 1513-002 300 red, 5
86 1 3066-094 Male Luer connector 8003-395 Concentrated waste container 4104-104 Pipe 8003-395 Concentrated waste container 1513-002 1100 red, 5
87 1 3066-094 Male Luer connector 8003-395 Concentrated waste container 4104-104 Pipe 8003-395 Concentrated waste container 1513-002 1100 red, 6
88 1 4104-083 Pipe 8003-379 Stirrer-ISE sipper assy 8003-396 Sample pipe assy 8003-378 Dry ISE module 1513-100 725
89 1 3070-058 Tube clamp 8003-284 Basic mechanical Assy 3070-058 Tube clamp 8003-274 Basic-frame 1513-002 480
100 1 3066-147 Male Luer connector, level tube assy outlet, low 8002-933 Level tube assy 5000-259 Pipe assy, inlet 8002-941 Sec. heat exchanger 1513-002 345
101 3 5000-259 Pipe assy, outlet 8002-941 Sec. heat exchanger 5000-259 Pipe assy, inlet 8002-941 Sec. heat exchanger 1513-002 125
102 1 5000-259 Pipe assy, outlet 8002-941 Sec. heat exchanger 3066-147 Male Luer connector, cooling pump inlet 8003-370 Sec. heat exch. assy 1513-002 260
103 1 3066-144 Female Luer connector 8003-367 Cooling pump assy 3365-020 Cooling pump, inlet 8003-367 Cooling pump assy 1513-002 140
104 1 3365-020 Cooling pump, outlet 8003-367 Cooling pump assy 3066-147 Male Luer connector 8003-367 Cooling pump assy 1513-002 140
105 1 3066-144 Female Luer connector, cooling pump outlet 8003-361 Stator assy 8002-934 Prim. heat exch. assy inlet, low 8003-361 Stator assy 1513-002 785
106 1 8002-934 Prim. heat exch. assy outlet, high 8003-361 Stator assy 8002-934 Prim. heat exch. assy inlet, high 8003-361 Stator assy 1513-002 170
107 1 8002-934 Prim. heat exch. assy outlet, low 8003-361 Stator assy 3066-092 Elbow connector 8003-361 Stator assy 1513-002 22
6003-400-450-02
108 1 3066-092 Elbow connector 8003-361 Stator assy 3066-147 Male Luer connector, level tube assy inlet, high 8002-933 Level tube assy 1513-002 900
120 1 8003-396 Sample pipe assy 8003-378 Dry ISE module Y connector 8003-452 Y-connector tube assy - 65
121 1 Y connector 8003-452 Y-connector tube assy 3918-005 Potassium (K) electrode 8003-378 Dry ISE module - 230
122 1 Straw 8003-453 Cap assy Y connector 8003-452 Y-connector tube assy - 230
123 1 Straw 8003-453 Cap assy 4104-135 Restriction pipe 8003-376 Restriction pipe assy - 280
124 1 3918-005 Potassium (K) electrode 8003-378 Dry ISE module 3918-004 Sodium (Na) electrode 8003-378 Dry ISE module - -
125 1 3918-004 Sodium (Na) electrode 8003-378 Dry ISE module 3918-006 Chloride (Cl) electrode 8003-378 Dry ISE module - -
126 1 3918-006 Chloride (Cl) electrode 8003-378 Dry ISE module 3918-007 Bypass electrode 8003-378 Dry ISE module - -
127 1 3918-007 Bypass electrode 8003-378 Dry ISE module 3900-049 Ground electrode 8003-378 Dry ISE module - -
128 1 3918-002 Reference electrode 8003-378 Dry ISE module 3900-049 Ground electrode 8003-378 Dry ISE module - -
129 1 4104-135 Restriction pipe 8003-376 Restriction pipe assy 3918-002 Reference electrode 8003-378 Dry ISE module - 45
130 1 3900-049 Ground electrode 8003-378 Dry ISE module 3066-164 Panel mount reduction connector 8003-378 Dry ISE module - - black, white
tube
codeno description total length (CW option & Dry-ISE option)
1503-007 PTFE, ID1.0 x OD1.6 1215
1503-008 PTFE, ID0.7 x OD1.6 1500
1513-001 PVC, ID2.0 x OD4.0 11955
1513-002 PVC, ID4.0 x OD6.0 9915
1513-009 PVC, ID1.5 x OD3.5 2208
1513-010 PVC, ID1.5 x OD2.5 6313
1513-024 PVC, ID4.0 x OD8.0 300
1513-100 Tygon®, ID0.79 x OD2.38 (ID1/32"xOD3/32") 725
SELECTRA
PRO
M