You are on page 1of 44

GE Healthcare

Technical
Publications

Direction 5121592
Revision 1

GEHC 0.2T Profile Magnet


Advanced Shimming Instruction

For The M20092PM, M20082PM & M20552PY


U

Permanent Magnet Installation Only


U

Copyright© 2004 by GE Healthcare


P P U

Operating Documentation
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

DAMAGE IN TRANSPORTATION

All packages should be closely examined at time of delivery. If damage is


apparent, have notation “damage in shipment” written on all copies of the
freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or
concealed, damage MUST be reported to the carrier immediately upon
discovery, or in any event, within 14 days after receipt, and the contents and
containers held for inspection by the carrier. A transportation company will not
pay a claim for damage if an inspection is not requested within this 14-day
period.

If shipment was handled by moving van service uncrate, call Traffic Milwaukee
immediately when any damage is found. At this time be ready to describe type
of damage, type of equipment, serial numbers and if possible the order number.

The above paragraph is in regard to equipment requiring installation only, and


does not apply to supply items

20-September-2004

GE Healthcare
GE Healthcare: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)

GE Healthcare — Europe: Telex 698626


283, rue de la Minere, B.P.34, 78533 Buc Cedex
France
TO GEHC: TECHNICAL PUBLICATION GROUP – MR
FAX NO: 86–510–522–6688
REPORT ON TECHNICAL PUBLICATIONS – MR
(OMISSION, ERRORS, SUGGESTIONS)

PRODUCT PUBLICATION #
U .
PAGE # PAGE REV DATE REPORTEDU .
SERVICE RECORD SUBMITTED □ YES □ NO
PLEASE BE SPECIFIC IN YOUR CORRECTIONS AND SUGGESTIONS

IF ADDITIONAL PAGES ARE INCLUDED, ATTACH THEM TO THIS SHEET.

SERVICE REP
(Please Print)

PHONE
FAX
COUNTRY
CITY

REV 5

DIRECTION NO. 15214


GE HEALTHCARE
TECHNICAL PUBLICATION GROUP – MR

19# Changjiang Road


Wuxi, Jiangsu Province, 214028, P.R.China
FAX No. 86 - 510 - 5226688

(TEAR AT PERF)
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

WARNING
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE
OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY
TO PROVIDE TRANSLATION SERVICES.
DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO
THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM
ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS.

AVERTISSEME

CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN


ANGLAIS.
SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE
AUTRE LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL
INCOMBE DE LE FAIRE TRADUIRE.
NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT
QUE LE MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER
CHEZ LE TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES
BLESSURES DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUES
OU AUTRES.

WARNUNG
DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN
ENGLISCHER SPRACHE.
FALLS EIN FREMDER KUNDENDIENST EINE ANDERE
SPRACHEBENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINE
ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN.
VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR
DIESES KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN
UND ERSTANDEN WURDE.
WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU
VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES
BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE
SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

ATENÇÃO
 
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA
UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL
CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A
QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL
PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS
ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA.

AVVERTENZA

ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA


DISPONÍVEL EM INGLÊS.
SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE
NÃO A GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE
TRADUÇÃO.
NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO
ECOMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A
SEGURANÇA DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO
A‘ CHOQUES ELÉTRICOS,MECÂNICOS OU OUTROS.

AVISO
IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE
SOLTANTO IN INGLESE.
SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC
RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È
TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA
SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED
AVERNE COMPRESO IL CONTENUTO.
NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE
FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI
ALL’ADDETTO ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL
PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI
MECCANICI OD ALTRI RISCHI.
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

警 告

注意
本安装手册仅存英文本。
非 GEHC 公司的安装/维修人员要求非英文本的手册时,客户需自行负责翻
译。
未详细阅读和完全了解本手册之前,不得进行安装/维修。
忽略本文注意事项会对安装人员, 操作维修人员或病人造成触电,机械或
其他伤害。
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

REVISION HISTORY
REV . DATE PRIMARY REASON FOR CHANGE

1 ……… October 20th, 2004


P P …… Released

1
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

CAUTION
DO NOT USE THE FOLLOWING DEVICES NEAR THE MAGNET. THE USE OF
THESE DEVICES NEAR THE MAGNET MIGHT BE DAMAGED BY, OR
CAUSES DAMAGES TO THE MAGNET.

DEVICES NOT TO BE USED NEAR THIS EQUIPMENT

DEVICES WHICH INTRINSICALLY TRANSMIT RADIO WAVES SUCH AS:


CELLULAR PHONE, MOBILE RADIO TRANSMITTER, RADIO TRANSCEIVER,
RADIO CONTROLLED TOY, ECT.

KEEP POWER TO THESE DEVICES TURN OFF WHEN NEAR THE MAGNET.

MEDICAL STAFF IN CHARGE OF THIS EQUIPMENT IS REQUIRED TO


INSTRUCT TECHNICIANS, PATIENTS AND OTHER PEOPLE WHO MAY BE
AROUND THIS EQUIPMENT TO FULLY COMPLY WITH THE ABOVE
REGUALTION.

Warning
Never use the following devices near this equipment. They
may cause erratic function of this equipment.

Cellular phone, transceiver, mobile radio transmitter and


radio-controlled toy, etc

Keep those devices power-off near this equipment.

a
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

INTRODUCTION
This manual introduces the advanced magnet shimming method to the
GEHC 0.2T Profile Permanent Magnet (M20092PM, M20082PM &
M20552PY, in the following part of this manual, all the “magnet” refers
to the GEHC 0.2T Profile Permanent Magnet). The followings are
instructed in this manual:

Shimming Equipments / Devices

Magnetic Field Operation Safety

Shimming Specification

Shimming procedures

Usage of the GEHC Advanced Shimming Software

All preliminary site plans and concepts must be reviewed by GE


Healthcare MR Installation Group prior to installation.

The shimming mission is highly sophisticated, and special engineering


competence is required. GE will use its own specially trained field
engineers. The purchaser of GE equipment shall only utilize qualified
personnel (i.e., GE’s field engineers) to perform shimming servicing on
the equipment.

All the contents of this Magnet Installation Manual are used for the
GEHC 0.2T Profile Permanent Magnet only.

All works must be in compliance with national and local safety codes.

b
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

TABLE OF CONTENT
REVISION HISTORY ……………………………………………………..1
T T T

SECTION 1 MAGNETIC FIELD SAFETY …………………………… 2

SECTION 2 MAGNETIC FIELD SHIMMING PREPARATION ……..5

2-1. Tools & Equipments …..………………………………………..5

2-2. Pre-requirements before starting the shimming…………. ....5

SECTION 3 FIELD SHIMMING SPECIFICATIONS …………………..9

3-1. Magnet field homogeneity specification…….……………..….9

3-2. Magnet field shimming specification quick guides…..…….. .9

SECTION 4 MAGNETIC FIELD SHIMMING PROCEDURES ……...10

4-1. Hardware setup. ……………………………………………….11

4-2. Shimtool working directory setup ……………………………11

4-3. Record the magnetic field point by point ……………………12

4-4. Compare the Pk-Pk PPM value with spec. …………………12

4-5. Mechanical shimming ………………………………………...13

4-6. Passive Shimming …………………………………………….13

4-7. Go to 4-3……..…………………………………………………14

4-8. Make the Data Sheet……………...…………………………..14

SECTION 5 FIELD SHIMMING SOFTWARE INTRODUCTION……15

5-1. Software Installation…………………………………………...15

5-2. Software Execution and Interface introduction…...………...15

5-3. Magnet Shimming Feedback……………………………... 34

1
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

SECTION 1 MAGNETIC FIELD SAFETY

WARNING

THERE IS A MAGNETIC FIELD PRESENT AROUND THE MAGNET AT TIME OF MAGNET DELIVERY,
EVEN WHEN THE MAGNET IS FIXED INSIDE THE SHIPPING CRATE. UP MOST CARES MUST BE
TAKEN WHEN WORKING AROUND THE MAGNET. BE SURE TO KEEP ALL FERROUS OBJECTS
AND EQUIPMENT THAT MAY BE INFLUENCED BY THE MAGNET AWAY FROM THE MAGNET.
DETAILS ARE ILLUSTRATED IN TABLE 1-1.

WARNING

THE MAGNETIC AND ELECTRO-MAGNETIC FIELD CAN CAUSE MALFUNCTION OF ACTIVE


IMPLANTS, SUCH AS CARDIAC PACEMAKER, NEUROSTIMULATORS, AND BIOSTIMULATION
DEVICES. PERSONS WITH CARDIAC PACEMAKERS, NEUROSTIMULATORS AND
BIOSTIMULATION DEVICES MUST REMAIN OUTSIDE THE 5 GAUSS MAGNETIC FRINGE FIELD
EXCLUSION ZONE AS ILLUSTRATED IN TABLE 1-1.

WARNING

THE STRONG FORCE GENERATED BY MAGNETIC FIELD CAN CAUSE DISPLACEMENT OF


INTRACRANIAL ANEURYSM CLIPS IN PATIENTS’ BRAIN, AS WELL AS OTHER METAL
PROSTHESIS. PERSONS WITH THESE KIND DEVICES MUST REMAIN OUTSIDE THE
RECOMMENDED FIVE GAUSS EXCLUSION ZONE AS ILLUSTRATED IN TABLE 1-1.

WARNING

THE STAFF, WHO WILL BE INVOLVED IN THE MAGNET OPERATIONS, MUST BE CHECKED FOR
ANY MEDICAL / PHYSICAL CONTRAINDICATION (INCLUDING BUT NOT LIMIT TO THE ABOVE) TO
THIS OPERATION, AND THIS CONFIRMING SHOULD BE PERFORMED BY LICENSED
CONTRACTORS, AS WELL AS GE STAFF IN CHARGE OF THIS EQUIPMENT. APPROPRIATE
PERSONAL PROTECTION EQUIPMENT SUCH AS SAFETY GLASSES, SAFETY SHOES AND
LEATHER GLOVES MUST BE WORN WHEN PERFORMING MAGNET INSTALLATION TASKS TO
MINIMIZE RISK OF PERSONAL INJURY.

WARNING

THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED. BE CAUTIONS ABOUT
THIS CHANGE OF THE MAGNETIC FRANGE FILED CHANGE.

2
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

CAUTION
DO NOT USE THE FOLLOWING DEVICES NEAR THE MAGNET. USE OF THESE DEVICES
NEARS THE MAGNET MIGHT BE DAMAGED BY, OR CAUSE DAMAGES TO THE MAGNET.

DEVICE WHICH INTRINSICALLY TRANSMIT RADIO WAVES SUCH AS: CELLULAR PHONE,
MOBILE RADIO TRANSMITTER, RADIO TRANSCEIVER, RADIO CONTROLLED---TOY, ETC.
KEEP POWER OF THESE DEVICES TURN OFF WHEN NEAR THE MAGNET.

PLEASE CLEAR FERROUS OBJECTS OUT OF THE DELIVERY ROUTINE AND SCAN ROOM
BEFORE DELIVERING THE MAGNET TO THE SCAN ROOM. THESE FERROUS OBJECTS
COULD CAUSE DAMAGES TO THE MAGNET, OR INJURIES TO PERSON AROUND THE
MAGNET. PREVENT PEOPLE WITH MEDICAL AND PHYSICAL CONTRAINDICATIONS FROM
ENTERING THE MAGNET OPERATION AREA.

CONSIDERATION MUST BE GIVEN TO THE MAGNET DELIVERY ROUTE TO ENSURE THAT


THE FLOOR CAN SUPPORT THE WEIGHT OF THE MAGNET. A STRUCTURAL ANALYSIS
SHOULD BE PERFORMED BY A PROFESSIONAL STRUCTURAL ENGINEER.

SPECIAL RIGGING EQUIPMENT MUST BE CONSIDERED IF THE MAGNET IS TO BE LOWERED


TO A DIFFERENT FLOOR LEVEL THAN THE RECEIVING LOCATION LEVEL.

THE TEMPERATURE FOR THE MAGNET IN THE DELIVERY PROCESS MUST BE BETWEEN –30
℃ AND 45℃ AND HUMIDITY UP TO 80%. IF MAGNET TEMPERATURE IS OUT OF THIS RANGE,
PLEASE CONTACT THE LOCAL SERVICE SUPPORT CENTER.

MAGNETIC SENSITIVE DEVICE PROXIMITY LIMITS


Table 1-1

GAUSS (mT) LIMIT EQUIPMENT


(See Notes 1&2)
0.5 GAUSS(0.05mT) OR LESS Nuclear cameras
1 GAUSS(0.1mT) OR LESS Positron Emission Tomography scanner
Video display (color, B/W, monochrome)
Linear Accelerator CT scanner
Cyclotrons Ultrasound
Accurate Measuring scale Lithotriptor
Image intensifiers Electron microscope
Color TV/Color Monitor( without shielding)
Advantage Windows Workstation
3 GAUSS(0.3mT) OR LESS
5 GAUSS(0.5mT) OR LESS Cardio pacemakers Bio stimulation devices
Neurostimulators Uncontrolled public and staff access
10 GAUSS(1mT) OR LESS Shielded Color Monitor for Stallion.
30 GAUSS(3mT) OR LESS MR equipment (Scan Control, Scan Room Unit, System
Gradient Unit, Step Down Transformer
50 GAUSS(5mT) OR LESS MR equipment (Regulated Step Down Telephones
Transformer or Main Disconnect Device Metal detector for screening
(in U.S.A). )

3
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

100 GAUSS(10mT) OR LESS Service Tool Magnet Power Supply Cabinet Smoke/heat detector
SM Laser Camera (without film processor)
Note
1. Recommended limits given above are based on general MR site planning guidelines. Actual susceptibility of
specific devices may vary significantly depending on electrical design orientation of the device relative to the
magnetic field and the degree of interference considered unacceptable.
2. Verify operating limits with provider of helicopter service.

4
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

SECTION 2 MAGNETIC FIELD SHIMMING


PREPARATION

2-1 Tools & Equipments:


For field shimming, special tools and equipments are needed. The following tools
and devices should be prepared by service engineers who preside the shimming.
Please refer to table 2-1 for tools/equipments details.

PART# DESCRIPTION COMMENT


46-282239G1 Non-magnetic Metric tool kit
46-255839P2 Tesla Meter PT2025 Metro Lab Teslameter
46-255839P102 Single Probe: #3 Range 0.17~0.52 Tesla
2141561 Mapping Fixture Assy
N/A Lap top computer with CD-ROM Reader Windows2000/98
Shim Chips, Shim Spacer, N/S
2370113 PASSIVE SHIM KIT Checker..etc.
Used to connect the Telemeter to
2396997 RS232C Cable the LapTop
GEHC Signa Profile Magnet Advanced
Field Shimming Service CDROM (Class
5122077 C) TAB 497
2359127 or 2219069 Profile Advanced CDROM SV manual TAB 497

TOOLS/EQUIPMENTS
Table 2-1

2-2 Pre-requirements before starting the shimming


a. The temperature of the shimming room should be set within the range from 22℃ to 26℃,
which can be controlled by the air-conditioner and measured by the Temperature Meter.
The room temperature should be kept in stable to control the B0 field drift.

b. Set the magnet TCU at the same temperature as the factory magnet shimming TCU
setting. The factory shimming TCU setting number could be found from the magnet
“ Acceptance Test Report”.
* The “Acceptance Test Report” could be found in the plastic folder placed on the pole face of the bottom Yoke.

c. Take off the Overall Mask and the Ring Protector Set from the magnet.

d. Make sure the insulators are installed properly on the magnet.

e. Install the mapping fixture and the Probe.

5
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

Please find more detail procedures from: Service CD (Part NO. 2359127 or 2219069),
Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo 2. Setup
(Mapping Fixture setup).

f. Copy the GEHC 0.2T Profile Magnet Class C field shimming software to the Laptop
computer. This software is distributed by the TAB497, and the CDROM P/N is 5122077.

g. Teslameter setup and adjustment.

In order to let pc communicate with teslameter correctly, several items should be checked.

TESLA METER IO MODULE


Illustration 2-1
1) With ShimTool class C, the micro-switch (red base with white button) on the IO
module of teslameter (PT2025, from the back) should be set as shown in Table 2-2.
Switch Name Value
1 1
2 0
3 0
4 1
5 0
6 0
7 0
8 0
9 0

* Please confirm the settings before go the next step.

TESLAMETER MICROSWITH ADJUSTMENT


Table 2-2
2) Set the Teslameter to Frequency mode as shown in illustration 2-2.

6
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

Frequency
mode

SET THE TESLA METER TO FREQUENCY MODE


Illustration 2-2
If one wants to enable the functions of B0 drift monitoring and auto data gathering of this
software, he should also do the following:

3) Physically connect the PC to the Tesla meter with an RS232c cable, 9-pin terminal to
pc and 25-pin terminal to the Teslameter (The RS232C cable could be found in the
magnet accessories shipping collector). Please refer to Illustration 2-3.
4) Setup serial port and baud rate correctly in the software. Please refer to section 5-2 ,
2.1~2.3
If the communication is not established successfully, the B0 drift monitoring and auto
data gathering functions of the software could be unavailable for the shimming. One
has to manually record the data.

LAPTOP PC SERIAL PORT (RS232C STANDARD PORT)

RS232C CABLE CONNECT PC LAPTOP TO TESLAMETER


Illustration 2-3

7
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

h. Heat the magnet no less than 24 hours until the temperature of the magnet is stable
enough.

The field drift monitoring can be accomplished by the advanced shimming software
automatically. Please refer to the Section 5-2 2.4 “ B0 Drift Monitoring “ for detail
introduction. The log file with a fixed name "B0DriftRecord.dat" which store the B0 drift
information will be saved to the directory you have set up.(Please refer to section 4
4.2"Shimtool working directory setup")

Field Drift Inspection Specification:


Drift of the central field should be less than 150Hz per 10 minutes and this situation should
last at least two hours. To measure drift, please place the probe in the ISO center of the
magnetic field of Plane 7 (Refer to Section 5 2.5 B0 Drift Monitoring for details about the
location of the probe in Plane 7).

Inspection method: Measure the central field and calculate the drift every 10 minutes.
Suppose the initial central field is B1 and drifts to B2 10 minutes later, then the drift can be
gotten by: |B2-B1| ; (|B2-B1|<=150 Hz).

NOTE: THE FIELD SHIMMING CANNOT START UNTIL ALL THE ABOVE REQUIREMENTS
ARE MET.

8
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

SECTION 3 FIELD SHIMMING SPECIFICATIONS

3-1: Magnet field homogeneity specification.


The field shimming cannot be finished until the following two specifications are met.

a. Magnet Final Central Frequency Range is within 8.50 ~ 8.58 MHz.

b. At the FOV of 40*40*35cm, the inhomogeneity of the field must be less than 45 ppm
(135 pts Pk-Pk measurement).

In the next section there is an introduction about how to shim the magnetic field to meet above
specifications.

3-2: Magnet field shimming specification quick guides.


Parameter Specification Notes
Magnet Drift Rate ≤150Hz per 10 minutes In 2 hours period.
Room Temperature 24.0° C ±2.0° C Keep the temperature stable.
Suggest to be the same setting with
the factory TCU setting. Refer to
Magnet TCU setting 32.5 ° C ± 1.0° C
“Acceptance Test Report” for the
factory temperature setting number *.
Magnet Final Frequency 8.50 ~ 8.58 MHz
Range
40x40x35 cm DSV ≤ 45.0 ppm Compensated
Specification
18 cm DSV ** ≤ 3.3ppm Compensated

* The “Acceptance Test Report” could be found in the plastic folder placed on the pole face of the bottom Yoke.
** Only required to the Profile/I system, or EPI function.

SPECIFICATGION QUICK GUIDE

Table 3-1

9
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

SECTION 4 MAGNETIC FIELD SHIMMING PROCEDURES


This section introduces the magnetic filed shimming procedures of the GEHC 0.2T Profile
Magnet. The basic magnetic field shimming procedures for GEHC 0.2T Profile Magnet are as
shown in flowchart 4-1.
Start

Hardware setup
(Mapping fixture, probe)

Run shimtool C, setup the working


directory for the first time

Shim Data Gathering , Section 5-2, 2.5)

Yes
Homogeneity in Spec?

No

Mechanical shimming Calculation


Section 5-2, 3.1)

Yes
|Z1| < 7 & |Z2| < 7 ppm

No
Execute Mechanical
shimming.*

Scan passive shimming


Section 5-2, 3.5)

Choose a good result & execute


Passive Shimming
Section 5-2, 3.6)

Data Sheet

End
* There is no Transverse shimming in the GEHC 0.2T Profile Magnet mechanical shimming.
MAGNETIC FIELD SHIMMING PROCEDURE FLOWCHART
Flowchart 4-1

10
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

4-1. Hardware setup.


1. Fix the single probe fixture onto the bottom ring of the pole piece.

Please refer to Profile Service CD (Part NO. 2359127 or 2219069), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo 2. Setup (Mapping
Fixture setup) for details instructions.

2. Fix the single probe on the Horizontal bar, and mount it on the single probe fixture.

Please refer to Profile Service CD (Part NO. 2359127 or 2219069), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo 2. Setup ( probe
setup) for detail instruction.

3. Power on the Tesla-meter PT2025. Set / confirm the Teslameter to the Frequency
Mode.

4. Set / confirm the micro-switch is adjusted. Refer to Section 2-2.g.


5. Connect the RS232C cable to the Teslameter and COM port of the Laptop.

Please refer to section 2-2, g.3). “ Pre-Requirements before starting the shimming”
for teslameter setting direction.

6. Power on the Tesla-meter PT2025, and confirm the communication between


Teslameter and the Laptop being established;

Please refer to section 5-2 2.1~2.3 for the serial port and baud rate setup in the
software.

4-2. Shimtool working directory setup


Confirm the Shimming Software is installed successfully in your Laptop. The software
CDROM (GEHC Signa Profile Magnet Advanced Shimming software CDROM, PN
5122077) is distributed by TAB 497.

After running the software, please specify a directory for the software before a new
magnet is to be shimmed. Once the directory is specified, all the files the software
generates in the shimming process will be saved to this folder automatically. Please
follow section 5 1.1 “Set Working Dir…”.

11
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

4-3. Record the magnetic field point by point


Please refer to Service CD (Part NO. 2359127 or 2219069), Profile 3/4 Installation
(For Profile 4b) 6th STEP Sumitomo Section 5 Data Gathering for the
guidance about the data gathering procedures.

Please refer to section 5 2.5 “Data Gathering” for details about how to use the
software to record/ gather the magnetic field data.

NOTE: THE ORIENTATION SYSTEM OF THE GEHC 0.2T PROFILE MAGNET IS


DIFFTERENT FROM THE SUMITOMO / SHINETSU MAGNET AS SHOWN IN
THE ILLUSTRATION 4-1. PLEASE PAY ATTENTION TO MEASURE THE
MAGNETIC FIELD DATA FROM THE GEHC 0.2T MAGNET SHIMMING
COORDINATION 0 DEGREE.

180
210 150
240 120

270 90

300 60
330 30
0
Start Measuring the Data from here
Table Side
GEHC 0.2T Profile Magnet Coordinate System

Illustration 4-1 MAGNET OREINTATION SYSTEM

4-4. Compare the Pk-Pk PPM value with spec.


Magnetic Field inhomogeneity pk-pk ppm value can be read from the field map in the
software main window (refer to illustration 5-1). If the overall inhomogeneity is in spec
(see section 3-2), field shimming could be finished, otherwise go to the Step 4-5;

12
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

4-5. Mechanical shimming


If either Z1 or Z2 is more than 7ppm, please perform the Mechanical Shimming, Axial
only. Please refer to the Section5-2 3.1) to calculate “mechanical shimming”.

Please refer to Profile Service CD (Part NO. 2359127 or 2219069), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo 9. Mechanical Shim (Axial)
for detail procedures about how to implement the mechanical shimming. The GEHC
0.2T Profile does not designed to implement the Mechanical Shimming, transverse
method.

After this mechanical shimming is finished, please go back to 4-3. Only when Z1 or Z2
are all less than 7 ppm, could the passive shimming begin as introduced in section 4-6.

4-6. Passive Shimming.


If the mechanical shimming meets the spec of all Z1& Z2 less than 7ppm, one can
execute “scan passive shimming”, which is described in Section 5-2 3.5, to execute the
passive shimming. After scan and calculation finished, one can go through the results
and select a relative good shimming plan from the shimming calculation results by
evaluating the number of the shimming chips would be used, the location number to be
used and the anticipated homogeneity value to be achieved. Run the passive shimming
with the parameter setting as the selected target ppm and shims number to get the
passive shimming plan. Please refer to Section 5-2 3.6 as instruction to the passive
shimming.

For locate the shim chips, please refer to Please refer to Profile Service CD (Part NO.
2359127 or 2219069), Profile 3/4 Installation (For Profile 4b) 6th
STEP Sumitomo 11. Passive Shimming

NOTE: THE ORIENTATION OF THE SHIM HOLDER SOCKET POSITION is


DIFFERENT WITH THE SUMITOMO OR SHINETSU MAGNET (REFER TO
ILLUSTRATION 4-1). PLEASE FOLLOW THE GEHC 0.2T PROFILE
MAGNET SHIM SOCKET ORIENTATION TO PLACE THE SHIMMING CHIPS.

13
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

4-7. Go to 4-3.
Compare with the shimming specs in section 3. Do the passive shimming or mechanical
shimming again until the specs are met.

4-8. Make the Data Sheet.


After the magnet shimming is accomplished, please fill the magnet installation spreadsheet
which could be found in the Advanced Shimming software Service CDROM (5122077).

After fill this form, please send the spreadsheet to ProfileMagnet.GEMSC@GE.com.


H H You
can ask assistantship from the magnet engineering via this email account if you have any
question in the installation process.

NOTE: THE OVERALL MASK AND THE RING PROTECTOR SET SHOULD BE PROPERLY
PRESERVED FOR THE USAGE IN THE FUTURE. DURING ANY MAGNET RELATED
MAINTENANCE OR MANGET TRANSPORTATION, THE OVERALL MASK AND THE
RING PROTECTOR SET MUST BE USED PROPERLY TO PROTECT THE MAGNET
FROM ANY DAMAGES OR PERSONNEL INJURIES.

NOTE: IN ANY KIND OF MRI SYSTEM MAINTENANCE, PLEASE BE CAUTIONS TO THE


STRONG MAGNETIC FIELD. PLEASE REMOVE ALL THE FERROUS OBJECT FROM
THE MAINTENANCE OPERATION SITE. PLEASE REFER TO THE FORNER SECTION
FOR THE SAFETY MEASURES.

14
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

SECTION 5 FIELD SHIMMING SOFTWARE


INTRODUCTION
There is a GEHC advanced shimming software prepared to the field shimming of GEHC 0.2T
Profile magnet. This shimming software is mounted in the GEHC 0.2T Profile Magnet advanced
field shimming Service CDROM (5122077), which is distributed by Tab497. To implement the
field shimming to the GEHC 0.2T Profile Magnet, the service engineers must install this software
to the Laptop first.

5-1: Software Installation


In the CDROM (5122077) there are one folder and two independent files, which are GEHC
shimming software, GEHC 0.2T Profile Magnet Advanced Shimming Manual (PN:5121592) and
Magnet installation status investigation excel sheet. Please copy all the contents of the CDROM
to your local Laptop hard disk.

5-2: Software Execution and Interface introduction


Go into the shimming software directory, and [double click] the executive file named “ShimToolC”.
The shimming software User Interface will jump out. In the shimming software operation window
there are Manual Bar, Tool Bar, Main Window and output window. Please refer to illustration
5-1 for details. All the calculation results and the shimming direction will be instructed in the
Output Window.

ADVANCED SHIMMING SOFTWARE OPERATION WINDOW


Illustration 5-1

15
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

1. File Menu

FILE MANU
Illustration 5-2

1.1 ) Set Working Dir...

SET WORKING DIR….


Illustration 5-3

After the click the manual the following directory-setting dialog, illustration 5-4 will appear:

Illustration 5-4

16
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

When click on Button, the desired target directory could be selected, established or renamed.
Please refer to the dialog in Illustration 5-5.

Select a folder as a working directory


Illustration 5-5

Before beginning to shim a new magnet, it's recommended to set the Working Directory. With
this setting, all files generated in the shimming process, including the maps, passive shimming
analysis result files, B0 drift monitoring records and output log files, could be saved to this
directory by default.

NOTE: THE BUTTON IN THIS SOFTWARE ALWAYS USED TO SELECT THE


TARGET FILE OR DIRECTORY.
NOTE: TO SPECIFY A DIRECTORY FOR THE MAGNET SHIMMING IS STRONGLY
SUGGESTED BEFORE EXECUTING THE FIELD SHIMMING. PLEASE SPECIFY
A NEW DIRECTORY EACH TIME ONE SHIM A NEW MAGNET.

1.2 ) New
To Setup a new empty document which is waiting for data-gathering (Auto/manually data
gathering).

Illustration 5-6

17
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

1.3) Open...

Illustration 5-7
To open an existing *.map / *.psm file which will automatically search the base map and show
the result with the predicted map. Please refer to Illustration 5-7.

1.4) Save

Illustration 5-8

To save current map file as shown in Illustration 5-8.

1.5) Save As...

Illustration 5-9

18
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

To save current map file with a different name as shown in Illustration 5-9.

1.6) Most Recent Used Files

Illustration 5-10

The most recent used file will be listed for the convenience. Please refer to Illustration 5-10.

2. Data Menu

Illustration 5-11

Sub menu items under this menu are mainly related with data gathering, which includes auto
data gathering through COM port and manually data gathering. Please refer to Illustration 5-11.
For the details about the hardware setup and physically connection of pc laptop and teslameter,
please refer to section 2-2.g.

2.1) Select Serial Port...

Illustration 5-12

19
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

This function allow you to select an available serial port according to the following dialog. Please
refer to the Illustration 5-13. The effective port with the correct baud rate could be identified by
the shimming software automatically also. Please refer to 2.3) for details.

Illustration 5-13

2.2)Set Baud Rate...

Illustration 5-14

The Teslameter can be set from 300bps to 19200bps, and 2400 bps or 9600 bps are the most
possible settings. Please refer to the following illustration 5-15.

Set the Baud Rate


Illustration 5-15.

20
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

2.3) Serial Port Auto Find...

Illustration 5-16

This function helps to find the working COM port connected with the teslameter. The
programming will scan all the available COM ports with different baud rate until correct COM port
with correct baud rate is identified. Please let the Tesla meter lock the field before using this
function. And the whole scanning process will last for about 1 minute or so. When the working
baud rate of the teslameter is found, the edit box in “Found” will show “Yes”. Please refer to
Illustration 5-17.

Illustration 5-17

2.4)B0 Drift Monitoring

Illustration 5-18

21
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

This function enables the automatic B0 drift monitoring instead of manually calculation. To use
this function, the Teslameter must be connected to the PC successfully by RS232C cable and
serial communication parameters correctly setup in software. Please refer to the following B0
drift-monitoring dialog shown in illustration 5-19.

10

B0 drift monitoring
Illustration 5-19

The B0 drift data will be saved under the working directory automatically with file name
"B0DriftRecord.dat". There are 5 columns in each line of the B0 drift data file. Please refer to the
Table 5-3 for the content of each columns.

Time From The time begin to record data at each time interval
B0 now(MHz) The B0 value of each 2nd second
P P

PkPk(Hz) The peak-peak value during 2 record-time interval


The max value recorded during 2 record-time
Max
interval
The min value recorded during 2 record-time
Min
interval

Content of each record of B0 drift


Table 5-1

22
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

The internals between two records could be set in the "Rec Time", which can be found on the
right side and will be effective after click the "Apply" button).

2.5) Data gathering

Illustration 5-20

Data gathering dialog is for users to collect map data automatically by the communication
between the teslameter and the Laptop computer. Please refer to Illustration 5-20.

There are totally 135 points in 13 planes should be measured for one map. In plane 2~ plane 12,
each plane has 12 angles (0 degree to 330 degree, 30 degree for each step); There is only 1
point to be measured for each of the Plane 1 and plane 13. B0 field at the ISO center (plane7)
should be measured twice, firstly in the beginning of taking the map and secondly in the end of
the map data gathering. Please refer to illustration 5-21 for the arrangement of the planes. The
angles are marked on the shimholders of the magnet. Also, please refer to Profile Service CD
(Part NO. 2359127 or 2219069), Profile 3/4 Installation (For Profile 4b) 6th
STEP Sumitomo 11. Passive Shim for more details.

Z -AXIS

Illustration 5-21

23
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

The map data to be collected from each "Plane" and "Angle" are clearly indicated by the data
collection dialog, as shown in illustration 5-22.

Illustration 5-22

NOTE: THE ORIENTATION SYSTEM OF GEHC MAGNET IS DIFFERENT FROM SUMITOMO


OR SHINETSU MAGNET. PLEASE REFER TO ILLUSTRATION 4-1 FOR DETAILS.

One can also change the current position of the space point manually during data gathering by

click on the right side.

There are 2 modes of auto data gathering: "auto-measure" and "auto-accept".

a. In "auto-measure" mode, the button at right-low side shows "Auto Measure".


The program will read data through COM port automatically. After every data read, it will
compare the most recent 5 data acquired, if the peak-peak value of them (in Hz) is less than
what specified in "Max" edit box, the "Accept" button is enabled. And after clicking on the
"Accept" button, the average data(of the recent 5) will be recorded down to the current map file
automatically. And the recording line will go down to the next line.
Hint: Operation Procedures:
"Auto Measure" mode usually needs 2 field engineers' cooperation. One( named A) is moving
the position of the probe, another (B) is operating the computer.

24
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

In each point field value gathering:


A changes the position of the probe and then gives a sign to B --> B waits for read-number valid
and decide (by himself) to accept (save) the data and then gives a sign to A to change the
position of probe.

b. In "auto-accept" mode, the button at right-low side shows "Auto Accept". By click the “Auto-
measure” and “Auto-accept” bottom, you can switch between the two data recording methods.

Illustration 5-23

In this mode, after you click on the "Measure" button, the button itself will be disabled. And then
the program will compare the recent 5 data acquired from the tesla meter. A new data is
accepted and recorded automatically, once the peak-peak difference of these 5 continuous data
is less then what specified in "Max" edit box. After data is recorded to the map file, "Measure"
button is enabled again. The default value of Max is 10Hz.
Hint: Operation Procedures:
"Auto Accept" mode could be accomplished by at least 1 field engineer (named A).
In each point field value gathering:
A changes the position of the probe and click on "Measure" button --> A Waits the computer
accept the data (after valid) and then gives a beep (last for 2 seconds) to A to change the
position of probe.

25
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

2.6) Manual Data gathering

Illustration 5-24

Manual Data gathering dialog is used for recording the field value manually on the specified
points to generate a map. The field value is combined by Prefix and suffix. For example, if the
read value for teslameter is 8.5012345MHz, the prefix 8.5 could be specified as default value,
and 012345 is the only digits to be typed in. After typing, please click "Accept" key or type
“Enter” key on the keyboard to accept the data.

"Overall Warning Bound (PPM +/-)" edit box let you set the overall warning boundary for the
whole map, if field value typed in is outside of the range, a warning will alert you attention that

abnormal data is inputted. And you can click right side to change the current point and
recollect the data if you made a mistake.

"Short Term Warning Bound (PPM +/-)" edit box is similar with the above edit box, which focus
on the difference between the current field value and last, theoretically there should be no big
difference consequent data.

26
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

3) Analysis Menu

Illustration 5-25

3.1) Mechanical Shimming


After run “mechanical shimming” command to an existing map, the program will bring out the
mechanical shimming plan in the output Window similar as the following Illustration 5-26. If the
software gives the result that either |Z1| or |Z2| is larger than 7, one should use the shimming
plan and perform the mechanical shimming physically.

Mechanical Shimming Report


Illustration 5-26
One can watch teslameter’s display value and tune the Top B0 plug and then Bottom B0 plug
until it is as same as what is in the output window, respectively.

3.2) Decompose Field


This function allows you to read the axial harmonic of the magnetic field. Only harmonics from
(0,1)~(0,12) are listed. The most important items are (0,1) and (0,2), or say Z1 and Z2.

3.3) Set Scan Shim Param...

27
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

Illustration 5-27
The dialog shown in Illustration 5-27 is used to set the parameters for the “ Scan Shim”. “ Scan
Shim” function is used to assisting service engineer to find the best Passive Shimming “ target “.
Please refer to 3.5) Passive shimming.

With Only ONE time setting here, one can execute passive shimming analysis and get a series
of simulation results simply and easily (Please refer to Illustration 5-28). The target PPM can be
from 0.01 to 100. “Shims limit” stands for the shimming chips number limitation in each socket,
and it should be less than or equal to 4. The default values are suitable to most cases. You can
also change them before run “scan passive shimming” in 3.4) to see if better result could be
generated (The better result comes from the comprehensive consideration of the
“Anticipated PPM” value and “Shims Number”. Both this two parameter should be as
small as possible).

Target PPM from 12 to 36, with delta of 0.2 and Shims Limit from 4 to 4 with delta of 1 is set as
the default value. When the best analysis results occur at the boundary, the Target PPM value
scanning range could be extended in have more analysis result around the target PPM value.
For Example: If the target is set as from 12ppm to 36ppm, while the best analysis result is just
near 12ppm, you could change the Target PPM range as from 5PPM to 14PPM and run
scanning again to see if there is an even better result could be found.

3.4) Scan passive shimming


“ Scan Shim” function is used to assisting you to find the best Passive Shimming “ target “. To
execute scanning of passive shimming with the specified target ppm and shims number limit in
each pocket, which is described in 3.4), to get the simulated shimming. A dialog, as shown in
Illustration 5-28, will appear after clicking the “scan passive shimming” command. If this is the

28
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

first-time passive shimming (NO shims in the shimholders), keep “2nd or More Iteration shim”
unchecked, and the “Shim File of Last Time” Edit Box is therefore disabled. Please refer to
Illustration 5-28

First Scan Passive Shimming Setting


Illustration 5-28
Otherwise (There are shims already on shim holders), please check the “2nd or More Iteration
P P

shim” and specify the psm file of last time before the current loaded map is taken. Please refer
to Illustration 5-29.

Scan Passive Shimming After the First Iteration


Illustration 5-29

After click on “OK” Button, the brief summary of analysis result for each target is shown in Output
Window. (One can also click on “Scan Para.” Bottom to set parameters for scanning passive
shimming). Please refer to Illustration 5-30.

29
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

Scan Passive Shimming Simulation Results


Illustration 5-30

After scanning is finished, one can look through the brief summary, choose the best anticipated
result and execute 3.5) “the passive shimming” with the corresponding “ settings”. Please
consider the “shims number” and the “Anticipated PPM” comprehensively with main principle the
smaller “shims number” with smaller “Anticipated PPM” the better.

3.5) Passive shimming


A dialog will appear and let you specify the passive shimming parameter. Please refer to
Illustration 5-31.

Passive Shimming Parameter Setting – First Iteration


Illustration 5-31

30
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

Passive Shimming parameter Setting – After the First Iteration


Illustration 5-32
If this is the first-time passive shimming (NO shims on shim holders), then keep “2nd or More
P P

Iteration shim” unchecked, and the “Shim File of Last Time” edit-box is therefore disabled.
Otherwise (There are shims already on shim holders), please check the “2nd or More Iteration
P P

shim” and specify the psm file of last time before the current loaded map is taken. These items
are the same as the introduction in 3.3)

Please specify the “Target PPM” value and “Shim Limit” value you choose from the “Scan
passive shimming” results (section 3.4). Click on “OK” button, and then the program will start to
calculate, generate the result and open the calculation results by the notepad. (For detail
information, please refer to section 4).

The string in the “Output Shim File” edit box is the name of the output shim file, which is
consistent with the name of the base map file as default ( Specified by user or assigned by the
software). The program will automatically add the anticipated inhomogeneity ppm value, target
ppm value to the file name, which has an extension of “.psm”. For example, if you type
“afterMech1” in this edit box as the file name, the target ppm is 13.6, the anticipated ppm is 22.5
after calculation, then the actually output shim file will be saved as the name
“afterMech1_22.5(13.6).psm” automatically. If one accept the result of the psm file and perform
shimming with it, please specify this name for the next iteration of passive shimming if needed.
(Please put this name in the “Shim File of Last Time”, as shown in Illustration 5-32.)

3.6) Stop Scan


To interrupt the current scanning. DO NOT close the shimtool software when scanning passive
shimming is running.

31
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

4) Passive shimming analysis result file.

This kind of file has its extension as "psm". Pease try to find the section of "Display Shim
Direction and Shim Location - Lower Delta ","Display Shim Direction and Shim Location - Upper
Delta", "Display Shim Direction and Shim Location - Lower Total and ","Display Shim Direction
and Shim Location - Upper Total" and execute passive shimming (put small shims into pockets
on shimholder), as shown in the Illustration 5-33.

NOTE: THE SHIMHOLDER ORIENTATION OF THE GEHC 0.2T PROFILE MAGNET IS


DIFFTERENT FROM THE SUMITOMO OR SHINETSU MAGNETS SHIMHOLDER.
PLEASE FOLLOW THE SHIMMING PROCEURES IN THE CDROM, WHILE USING
THE GEHC 0.2T PROFILE MAGNET SHIMHOLDER ORIENTATION TO PLACE THE
SHIM CHIPS. PLEASE REFER TO ILLUSTRATION 4-1 FOR MORE INFORMATION.

32
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

Passive Shimming Analysis Result


Illustration 5-33

33
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592

5-3. Magnet Shimming Feedback.

After the magnet shimming is accomplished, please fill the magnet installation data sheet which
could be found in the Shimming software Service CDROM (5122077).

After fill this form, please send the spreadsheet to ProfileMagnet.GEMSC@GE.COM, or you
can ask engineering for support via this email account if you have any question in the installation
process.