Operating Documentation
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
DAMAGE IN TRANSPORTATION
If shipment was handled by moving van service uncrate, call Traffic Milwaukee
immediately when any damage is found. At this time be ready to describe type
of damage, type of equipment, serial numbers and if possible the order number.
20-September-2004
GE Healthcare
GE Healthcare: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)
PRODUCT PUBLICATION #
U .
PAGE # PAGE REV DATE REPORTEDU .
SERVICE RECORD SUBMITTED □ YES □ NO
PLEASE BE SPECIFIC IN YOUR CORRECTIONS AND SUGGESTIONS
SERVICE REP
(Please Print)
PHONE
FAX
COUNTRY
CITY
REV 5
(TEAR AT PERF)
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
WARNING
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE
OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY
TO PROVIDE TRANSLATION SERVICES.
DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO
THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM
ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS.
AVERTISSEME
WARNUNG
DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN
ENGLISCHER SPRACHE.
FALLS EIN FREMDER KUNDENDIENST EINE ANDERE
SPRACHEBENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINE
ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN.
VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR
DIESES KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN
UND ERSTANDEN WURDE.
WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU
VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES
BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE
SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
ATENÇÃO
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA
UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL
CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A
QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL
PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS
ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA.
AVVERTENZA
AVISO
IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE
SOLTANTO IN INGLESE.
SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC
RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È
TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA
SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED
AVERNE COMPRESO IL CONTENUTO.
NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE
FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI
ALL’ADDETTO ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL
PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI
MECCANICI OD ALTRI RISCHI.
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
警 告
注意
本安装手册仅存英文本。
非 GEHC 公司的安装/维修人员要求非英文本的手册时,客户需自行负责翻
译。
未详细阅读和完全了解本手册之前,不得进行安装/维修。
忽略本文注意事项会对安装人员, 操作维修人员或病人造成触电,机械或
其他伤害。
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
REVISION HISTORY
REV . DATE PRIMARY REASON FOR CHANGE
1
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
CAUTION
DO NOT USE THE FOLLOWING DEVICES NEAR THE MAGNET. THE USE OF
THESE DEVICES NEAR THE MAGNET MIGHT BE DAMAGED BY, OR
CAUSES DAMAGES TO THE MAGNET.
KEEP POWER TO THESE DEVICES TURN OFF WHEN NEAR THE MAGNET.
Warning
Never use the following devices near this equipment. They
may cause erratic function of this equipment.
a
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
INTRODUCTION
This manual introduces the advanced magnet shimming method to the
GEHC 0.2T Profile Permanent Magnet (M20092PM, M20082PM &
M20552PY, in the following part of this manual, all the “magnet” refers
to the GEHC 0.2T Profile Permanent Magnet). The followings are
instructed in this manual:
Shimming Specification
Shimming procedures
All the contents of this Magnet Installation Manual are used for the
GEHC 0.2T Profile Permanent Magnet only.
All works must be in compliance with national and local safety codes.
b
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
TABLE OF CONTENT
REVISION HISTORY ……………………………………………………..1
T T T
4-7. Go to 4-3……..…………………………………………………14
1
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
WARNING
THERE IS A MAGNETIC FIELD PRESENT AROUND THE MAGNET AT TIME OF MAGNET DELIVERY,
EVEN WHEN THE MAGNET IS FIXED INSIDE THE SHIPPING CRATE. UP MOST CARES MUST BE
TAKEN WHEN WORKING AROUND THE MAGNET. BE SURE TO KEEP ALL FERROUS OBJECTS
AND EQUIPMENT THAT MAY BE INFLUENCED BY THE MAGNET AWAY FROM THE MAGNET.
DETAILS ARE ILLUSTRATED IN TABLE 1-1.
WARNING
WARNING
WARNING
THE STAFF, WHO WILL BE INVOLVED IN THE MAGNET OPERATIONS, MUST BE CHECKED FOR
ANY MEDICAL / PHYSICAL CONTRAINDICATION (INCLUDING BUT NOT LIMIT TO THE ABOVE) TO
THIS OPERATION, AND THIS CONFIRMING SHOULD BE PERFORMED BY LICENSED
CONTRACTORS, AS WELL AS GE STAFF IN CHARGE OF THIS EQUIPMENT. APPROPRIATE
PERSONAL PROTECTION EQUIPMENT SUCH AS SAFETY GLASSES, SAFETY SHOES AND
LEATHER GLOVES MUST BE WORN WHEN PERFORMING MAGNET INSTALLATION TASKS TO
MINIMIZE RISK OF PERSONAL INJURY.
WARNING
THE 5 GAUSS LINE WIDENS ABOUT 1m WHEN THE CRATE IS OPENED. BE CAUTIONS ABOUT
THIS CHANGE OF THE MAGNETIC FRANGE FILED CHANGE.
2
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
CAUTION
DO NOT USE THE FOLLOWING DEVICES NEAR THE MAGNET. USE OF THESE DEVICES
NEARS THE MAGNET MIGHT BE DAMAGED BY, OR CAUSE DAMAGES TO THE MAGNET.
DEVICE WHICH INTRINSICALLY TRANSMIT RADIO WAVES SUCH AS: CELLULAR PHONE,
MOBILE RADIO TRANSMITTER, RADIO TRANSCEIVER, RADIO CONTROLLED---TOY, ETC.
KEEP POWER OF THESE DEVICES TURN OFF WHEN NEAR THE MAGNET.
PLEASE CLEAR FERROUS OBJECTS OUT OF THE DELIVERY ROUTINE AND SCAN ROOM
BEFORE DELIVERING THE MAGNET TO THE SCAN ROOM. THESE FERROUS OBJECTS
COULD CAUSE DAMAGES TO THE MAGNET, OR INJURIES TO PERSON AROUND THE
MAGNET. PREVENT PEOPLE WITH MEDICAL AND PHYSICAL CONTRAINDICATIONS FROM
ENTERING THE MAGNET OPERATION AREA.
THE TEMPERATURE FOR THE MAGNET IN THE DELIVERY PROCESS MUST BE BETWEEN –30
℃ AND 45℃ AND HUMIDITY UP TO 80%. IF MAGNET TEMPERATURE IS OUT OF THIS RANGE,
PLEASE CONTACT THE LOCAL SERVICE SUPPORT CENTER.
3
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
100 GAUSS(10mT) OR LESS Service Tool Magnet Power Supply Cabinet Smoke/heat detector
SM Laser Camera (without film processor)
Note
1. Recommended limits given above are based on general MR site planning guidelines. Actual susceptibility of
specific devices may vary significantly depending on electrical design orientation of the device relative to the
magnetic field and the degree of interference considered unacceptable.
2. Verify operating limits with provider of helicopter service.
4
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
TOOLS/EQUIPMENTS
Table 2-1
b. Set the magnet TCU at the same temperature as the factory magnet shimming TCU
setting. The factory shimming TCU setting number could be found from the magnet
“ Acceptance Test Report”.
* The “Acceptance Test Report” could be found in the plastic folder placed on the pole face of the bottom Yoke.
c. Take off the Overall Mask and the Ring Protector Set from the magnet.
5
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Please find more detail procedures from: Service CD (Part NO. 2359127 or 2219069),
Profile 3/4 Installation (For Profile 4b) 6th STEP Sumitomo 2. Setup
(Mapping Fixture setup).
f. Copy the GEHC 0.2T Profile Magnet Class C field shimming software to the Laptop
computer. This software is distributed by the TAB497, and the CDROM P/N is 5122077.
In order to let pc communicate with teslameter correctly, several items should be checked.
6
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Frequency
mode
3) Physically connect the PC to the Tesla meter with an RS232c cable, 9-pin terminal to
pc and 25-pin terminal to the Teslameter (The RS232C cable could be found in the
magnet accessories shipping collector). Please refer to Illustration 2-3.
4) Setup serial port and baud rate correctly in the software. Please refer to section 5-2 ,
2.1~2.3
If the communication is not established successfully, the B0 drift monitoring and auto
data gathering functions of the software could be unavailable for the shimming. One
has to manually record the data.
7
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
h. Heat the magnet no less than 24 hours until the temperature of the magnet is stable
enough.
The field drift monitoring can be accomplished by the advanced shimming software
automatically. Please refer to the Section 5-2 2.4 “ B0 Drift Monitoring “ for detail
introduction. The log file with a fixed name "B0DriftRecord.dat" which store the B0 drift
information will be saved to the directory you have set up.(Please refer to section 4
4.2"Shimtool working directory setup")
Inspection method: Measure the central field and calculate the drift every 10 minutes.
Suppose the initial central field is B1 and drifts to B2 10 minutes later, then the drift can be
gotten by: |B2-B1| ; (|B2-B1|<=150 Hz).
NOTE: THE FIELD SHIMMING CANNOT START UNTIL ALL THE ABOVE REQUIREMENTS
ARE MET.
8
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
b. At the FOV of 40*40*35cm, the inhomogeneity of the field must be less than 45 ppm
(135 pts Pk-Pk measurement).
In the next section there is an introduction about how to shim the magnetic field to meet above
specifications.
* The “Acceptance Test Report” could be found in the plastic folder placed on the pole face of the bottom Yoke.
** Only required to the Profile/I system, or EPI function.
Table 3-1
9
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Hardware setup
(Mapping fixture, probe)
Yes
Homogeneity in Spec?
No
Yes
|Z1| < 7 & |Z2| < 7 ppm
No
Execute Mechanical
shimming.*
Data Sheet
End
* There is no Transverse shimming in the GEHC 0.2T Profile Magnet mechanical shimming.
MAGNETIC FIELD SHIMMING PROCEDURE FLOWCHART
Flowchart 4-1
10
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Please refer to Profile Service CD (Part NO. 2359127 or 2219069), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo 2. Setup (Mapping
Fixture setup) for details instructions.
2. Fix the single probe on the Horizontal bar, and mount it on the single probe fixture.
Please refer to Profile Service CD (Part NO. 2359127 or 2219069), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo 2. Setup ( probe
setup) for detail instruction.
3. Power on the Tesla-meter PT2025. Set / confirm the Teslameter to the Frequency
Mode.
Please refer to section 2-2, g.3). “ Pre-Requirements before starting the shimming”
for teslameter setting direction.
Please refer to section 5-2 2.1~2.3 for the serial port and baud rate setup in the
software.
After running the software, please specify a directory for the software before a new
magnet is to be shimmed. Once the directory is specified, all the files the software
generates in the shimming process will be saved to this folder automatically. Please
follow section 5 1.1 “Set Working Dir…”.
11
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Please refer to section 5 2.5 “Data Gathering” for details about how to use the
software to record/ gather the magnetic field data.
180
210 150
240 120
270 90
300 60
330 30
0
Start Measuring the Data from here
Table Side
GEHC 0.2T Profile Magnet Coordinate System
12
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Please refer to Profile Service CD (Part NO. 2359127 or 2219069), Profile 3/4
Installation (For Profile 4b) 6th STEP Sumitomo 9. Mechanical Shim (Axial)
for detail procedures about how to implement the mechanical shimming. The GEHC
0.2T Profile does not designed to implement the Mechanical Shimming, transverse
method.
After this mechanical shimming is finished, please go back to 4-3. Only when Z1 or Z2
are all less than 7 ppm, could the passive shimming begin as introduced in section 4-6.
For locate the shim chips, please refer to Please refer to Profile Service CD (Part NO.
2359127 or 2219069), Profile 3/4 Installation (For Profile 4b) 6th
STEP Sumitomo 11. Passive Shimming
13
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
4-7. Go to 4-3.
Compare with the shimming specs in section 3. Do the passive shimming or mechanical
shimming again until the specs are met.
NOTE: THE OVERALL MASK AND THE RING PROTECTOR SET SHOULD BE PROPERLY
PRESERVED FOR THE USAGE IN THE FUTURE. DURING ANY MAGNET RELATED
MAINTENANCE OR MANGET TRANSPORTATION, THE OVERALL MASK AND THE
RING PROTECTOR SET MUST BE USED PROPERLY TO PROTECT THE MAGNET
FROM ANY DAMAGES OR PERSONNEL INJURIES.
14
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
15
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
1. File Menu
FILE MANU
Illustration 5-2
After the click the manual the following directory-setting dialog, illustration 5-4 will appear:
Illustration 5-4
16
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
When click on Button, the desired target directory could be selected, established or renamed.
Please refer to the dialog in Illustration 5-5.
Before beginning to shim a new magnet, it's recommended to set the Working Directory. With
this setting, all files generated in the shimming process, including the maps, passive shimming
analysis result files, B0 drift monitoring records and output log files, could be saved to this
directory by default.
1.2 ) New
To Setup a new empty document which is waiting for data-gathering (Auto/manually data
gathering).
Illustration 5-6
17
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
1.3) Open...
Illustration 5-7
To open an existing *.map / *.psm file which will automatically search the base map and show
the result with the predicted map. Please refer to Illustration 5-7.
1.4) Save
Illustration 5-8
Illustration 5-9
18
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
To save current map file with a different name as shown in Illustration 5-9.
Illustration 5-10
The most recent used file will be listed for the convenience. Please refer to Illustration 5-10.
2. Data Menu
Illustration 5-11
Sub menu items under this menu are mainly related with data gathering, which includes auto
data gathering through COM port and manually data gathering. Please refer to Illustration 5-11.
For the details about the hardware setup and physically connection of pc laptop and teslameter,
please refer to section 2-2.g.
Illustration 5-12
19
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
This function allow you to select an available serial port according to the following dialog. Please
refer to the Illustration 5-13. The effective port with the correct baud rate could be identified by
the shimming software automatically also. Please refer to 2.3) for details.
Illustration 5-13
Illustration 5-14
The Teslameter can be set from 300bps to 19200bps, and 2400 bps or 9600 bps are the most
possible settings. Please refer to the following illustration 5-15.
20
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Illustration 5-16
This function helps to find the working COM port connected with the teslameter. The
programming will scan all the available COM ports with different baud rate until correct COM port
with correct baud rate is identified. Please let the Tesla meter lock the field before using this
function. And the whole scanning process will last for about 1 minute or so. When the working
baud rate of the teslameter is found, the edit box in “Found” will show “Yes”. Please refer to
Illustration 5-17.
Illustration 5-17
Illustration 5-18
21
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
This function enables the automatic B0 drift monitoring instead of manually calculation. To use
this function, the Teslameter must be connected to the PC successfully by RS232C cable and
serial communication parameters correctly setup in software. Please refer to the following B0
drift-monitoring dialog shown in illustration 5-19.
10
B0 drift monitoring
Illustration 5-19
The B0 drift data will be saved under the working directory automatically with file name
"B0DriftRecord.dat". There are 5 columns in each line of the B0 drift data file. Please refer to the
Table 5-3 for the content of each columns.
Time From The time begin to record data at each time interval
B0 now(MHz) The B0 value of each 2nd second
P P
22
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
The internals between two records could be set in the "Rec Time", which can be found on the
right side and will be effective after click the "Apply" button).
Illustration 5-20
Data gathering dialog is for users to collect map data automatically by the communication
between the teslameter and the Laptop computer. Please refer to Illustration 5-20.
There are totally 135 points in 13 planes should be measured for one map. In plane 2~ plane 12,
each plane has 12 angles (0 degree to 330 degree, 30 degree for each step); There is only 1
point to be measured for each of the Plane 1 and plane 13. B0 field at the ISO center (plane7)
should be measured twice, firstly in the beginning of taking the map and secondly in the end of
the map data gathering. Please refer to illustration 5-21 for the arrangement of the planes. The
angles are marked on the shimholders of the magnet. Also, please refer to Profile Service CD
(Part NO. 2359127 or 2219069), Profile 3/4 Installation (For Profile 4b) 6th
STEP Sumitomo 11. Passive Shim for more details.
Z -AXIS
Illustration 5-21
23
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
The map data to be collected from each "Plane" and "Angle" are clearly indicated by the data
collection dialog, as shown in illustration 5-22.
Illustration 5-22
One can also change the current position of the space point manually during data gathering by
24
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
b. In "auto-accept" mode, the button at right-low side shows "Auto Accept". By click the “Auto-
measure” and “Auto-accept” bottom, you can switch between the two data recording methods.
Illustration 5-23
In this mode, after you click on the "Measure" button, the button itself will be disabled. And then
the program will compare the recent 5 data acquired from the tesla meter. A new data is
accepted and recorded automatically, once the peak-peak difference of these 5 continuous data
is less then what specified in "Max" edit box. After data is recorded to the map file, "Measure"
button is enabled again. The default value of Max is 10Hz.
Hint: Operation Procedures:
"Auto Accept" mode could be accomplished by at least 1 field engineer (named A).
In each point field value gathering:
A changes the position of the probe and click on "Measure" button --> A Waits the computer
accept the data (after valid) and then gives a beep (last for 2 seconds) to A to change the
position of probe.
25
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Illustration 5-24
Manual Data gathering dialog is used for recording the field value manually on the specified
points to generate a map. The field value is combined by Prefix and suffix. For example, if the
read value for teslameter is 8.5012345MHz, the prefix 8.5 could be specified as default value,
and 012345 is the only digits to be typed in. After typing, please click "Accept" key or type
“Enter” key on the keyboard to accept the data.
"Overall Warning Bound (PPM +/-)" edit box let you set the overall warning boundary for the
whole map, if field value typed in is outside of the range, a warning will alert you attention that
abnormal data is inputted. And you can click right side to change the current point and
recollect the data if you made a mistake.
"Short Term Warning Bound (PPM +/-)" edit box is similar with the above edit box, which focus
on the difference between the current field value and last, theoretically there should be no big
difference consequent data.
26
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
3) Analysis Menu
Illustration 5-25
27
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Illustration 5-27
The dialog shown in Illustration 5-27 is used to set the parameters for the “ Scan Shim”. “ Scan
Shim” function is used to assisting service engineer to find the best Passive Shimming “ target “.
Please refer to 3.5) Passive shimming.
With Only ONE time setting here, one can execute passive shimming analysis and get a series
of simulation results simply and easily (Please refer to Illustration 5-28). The target PPM can be
from 0.01 to 100. “Shims limit” stands for the shimming chips number limitation in each socket,
and it should be less than or equal to 4. The default values are suitable to most cases. You can
also change them before run “scan passive shimming” in 3.4) to see if better result could be
generated (The better result comes from the comprehensive consideration of the
“Anticipated PPM” value and “Shims Number”. Both this two parameter should be as
small as possible).
Target PPM from 12 to 36, with delta of 0.2 and Shims Limit from 4 to 4 with delta of 1 is set as
the default value. When the best analysis results occur at the boundary, the Target PPM value
scanning range could be extended in have more analysis result around the target PPM value.
For Example: If the target is set as from 12ppm to 36ppm, while the best analysis result is just
near 12ppm, you could change the Target PPM range as from 5PPM to 14PPM and run
scanning again to see if there is an even better result could be found.
28
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
first-time passive shimming (NO shims in the shimholders), keep “2nd or More Iteration shim”
unchecked, and the “Shim File of Last Time” Edit Box is therefore disabled. Please refer to
Illustration 5-28
shim” and specify the psm file of last time before the current loaded map is taken. Please refer
to Illustration 5-29.
After click on “OK” Button, the brief summary of analysis result for each target is shown in Output
Window. (One can also click on “Scan Para.” Bottom to set parameters for scanning passive
shimming). Please refer to Illustration 5-30.
29
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
After scanning is finished, one can look through the brief summary, choose the best anticipated
result and execute 3.5) “the passive shimming” with the corresponding “ settings”. Please
consider the “shims number” and the “Anticipated PPM” comprehensively with main principle the
smaller “shims number” with smaller “Anticipated PPM” the better.
30
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
Iteration shim” unchecked, and the “Shim File of Last Time” edit-box is therefore disabled.
Otherwise (There are shims already on shim holders), please check the “2nd or More Iteration
P P
shim” and specify the psm file of last time before the current loaded map is taken. These items
are the same as the introduction in 3.3)
Please specify the “Target PPM” value and “Shim Limit” value you choose from the “Scan
passive shimming” results (section 3.4). Click on “OK” button, and then the program will start to
calculate, generate the result and open the calculation results by the notepad. (For detail
information, please refer to section 4).
The string in the “Output Shim File” edit box is the name of the output shim file, which is
consistent with the name of the base map file as default ( Specified by user or assigned by the
software). The program will automatically add the anticipated inhomogeneity ppm value, target
ppm value to the file name, which has an extension of “.psm”. For example, if you type
“afterMech1” in this edit box as the file name, the target ppm is 13.6, the anticipated ppm is 22.5
after calculation, then the actually output shim file will be saved as the name
“afterMech1_22.5(13.6).psm” automatically. If one accept the result of the psm file and perform
shimming with it, please specify this name for the next iteration of passive shimming if needed.
(Please put this name in the “Shim File of Last Time”, as shown in Illustration 5-32.)
31
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
This kind of file has its extension as "psm". Pease try to find the section of "Display Shim
Direction and Shim Location - Lower Delta ","Display Shim Direction and Shim Location - Upper
Delta", "Display Shim Direction and Shim Location - Lower Total and ","Display Shim Direction
and Shim Location - Upper Total" and execute passive shimming (put small shims into pockets
on shimholder), as shown in the Illustration 5-33.
32
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
33
GE HEALTHCARE GEHC 0.2T Profile Magnet Advanced Shimming Instruction
REV 1 DIRECTION 5121592
After the magnet shimming is accomplished, please fill the magnet installation data sheet which
could be found in the Shimming software Service CDROM (5122077).
After fill this form, please send the spreadsheet to ProfileMagnet.GEMSC@GE.COM, or you
can ask engineering for support via this email account if you have any question in the installation
process.