Ultrasound Imaging Gets High-Performance Design


NEW HOT
Home News Tech Room Downloads Webinars In Focus Videos EE Awards Asia EE Tech Summits 5G Forum   

 Article By : M. Di Paolo Emilio

 Category : Medical  2019-09-19  (0) Comments

Medical imaging has made great strides. Advances in the field of


microelectronics are fueling instrumentation towards portable and
high-efficiency solutions.

In modern medicine, medical imaging has made great strides. Transmitting acoustic energy in the body and
receiving and processing the reflections of the signal, ultrasound technology can generate images of our internal
organs, map the tissues, and other medical-scientific information. Ultrasound is represented by a high-frequency
sonar system, which measures the tiny echoes of sound waves that pass through the body. The advances in the
field of microelectronics have directed the instrumentation towards portable and high-efficiency solutions, with
performances that are getting closer and closer to the more complex and decidedly larger systems.

Synchronization is a must!

Ultrasound has a wide range of medical applications, including imaging. The main design elements are the
transmission and reception of the signal for a correct interpretation of the images. In a pulsed wave (PW)
ultrasound system, a high voltage pulse stimulates the piezoelectric transducer (a crystal), causing a corresponding
mechanical compression (an inverse piezoelectric effect), and thus creating the ultrasound wave that passes 
through the body (Figure
We use 1). to ensure that we give you the best experience on our website. If you continue to use this site we will assume that you are happy with it.
cookies Ok

Figure 1: Block diagram of an ultrasound system. (Source: Maxim Integrated)

The transmitter sends ultrasound signals, while the receiver collects the echoes that are reflected from the body
structure, generally through the same piezoelectric transducer with the support of digital processing and
amplification circuits.

The wave generated by the transducer through relaxation traverses the body and is focused by phased array
techniques. The reflection signal will have different intensity characteristics depending on the substances it
encounters, such as soft tissues, bones, fat, and blood.

The transmission and reception circuits must be synchronized to achieve the best results. The management of the
synchronization clock is fundamental for the information to be consistent and the image to have minimal
disturbances.

The phase noise (frequency-domain) or jitter (time-domain) is a measure of the phase of a clock signal that
deviates from its ideal position and is an important parameter that can determine the impact on the ultrasound
image. Higher phase noise can cause artifacts, which represents errors in the calculation and visualization of blood
flow.

The power supplies are synchronized with a master clock in the range of 100 kHz – 1 MHz to avoid switching
effects. Higher frequencies require additional measures in order to avoid coupling with the ultrasound signal.
Synchronization at lower frequencies avoids design complications helping to reduce electromagnetic interference
(EMI) too.

Amplification

The transmitted sound pulse propagates through the body as a wave in the frequency range between 1 MHz and
15 MHz. Sound waves are attenuated during the journey to the human body.

STHV1600 is a monolithic, high-voltage, high-speed pulse generator with 16 independent channels and a
beamformer. The IC provides a switch structure that ensures effective isolation during the transmission phase. Both
PW output stages (TX0 and TX1) can provide a peak output up to ± 2 A (Figure 2).

As the transmitted signal travels, part of the wave energy is reflected back to the transducer with very high-
intensity levels. The reflections that occur long after the start of the transmission are mostly weak signals. The
reflected signal is processed and displayed as an image on a screen. With appropriate amplification electronics,
extremely sensitive microelectronic solutions have been developed with a wide range of dynamics. The echoes
received from the focal points close to the surface require minimal or no amplification; while the others, those
received from the deep focal points in the body, must be amplified by a factor of 100 or more.

Figure 2: Block diagram of a complete ultrasound imaging layout with the external implementation of STHV1600
pulser. (Source: ST Microelectronics)

Most receivers must be able to adapt in both conditions, without saturating or distorting the signal, thus
introducing a minimum noise. For ultrasound systems, the amplifier consists of a low-noise, highly linear low-noise
amplifier (LNA) with digitally programmable gain and impedance; and a variable gain amplifier (VGA) before being
digitized by the analog-to-digital (ADC) to provide a better signal-to-noise radio (SNR) and phase noise for the
path (Figure 3).

MAX40077 is a wide-band, low-noise, operational amplifier with 2.7V to 5.5V single power supply. The family of
these devices absorbs about 2.2 mA of quiescent power supply with very low distortion (0, 0002% THD + N), as well
as a low density of noise in input voltage (4.2 nV/√Hz).

Figure 3: Block diagram of the reception system for the ultrasound system. The transmit/receive (T/R) switch isolates
the receiver during high voltage bursts in transmission. (Source: Maxim Integrated)

Since ultrasonic signals are attenuated as they travel through the body, the reflected (received) signal is much
smaller than the transmitted signal. Its detectable peak at the entry of the LNA is limited by the linear input range
of the LNA, typically 500 mVPP with an LNA gain of 18 dB.

During the transmission interval, overload conditions can occur, limiting the processing of the reflection signals and
making it difficult to see small signals in the presence of larger ones or completely losing the reflected signal.
Therefore, it is essential to choose a reception front-end circuit with excellent overload recovery features.

The LNA and VGA stages, as well as the ADC, should be designed in such a way that they all recover quickly from
an overload condition since the receiver could be blind until the end of recovery so as not to preserve the image
quality or information on the Doppler phase shift.

The MAX2082 is a fully integrated ultrasonic transceiver in the industry, optimized for high-performance, high-
channel portable systems. The transmission can generate high voltage pulses up to ± 105 V. The device is
composed of T/R switch, LNA, variable gain amplifier (VGA), antialiasing filter (AAF), analog to digital converter
(ADC) and digital high-pass filter (HPF) with a power dissipation of 131 mW per very-low channel at 50 Mbps
(Figure 4).

Figure 4: Block diagram of the MAX2082 and an application circuit. (Source: Maxim Integrated)

The choice of an ultrasound transceiver with excellent operating characteristics can significantly simplify the
HW/SW design and debugging challenges, speeding up time-to-market and improving image quality. Ultrasound
imaging offers a diagnostic procedure contributing to the general diagnosis of the patient's status.

Subscribe to Newsletter

email address

Subscribe

Share this:

Twitter Facebook LinkedIn More

【EE Tech Summits】 13:30, Oct. 22  Join us with senior industry leaders to
share their insights into future technology trends! Register Free Now!

Related

Diamond Quantum Technology for Medical imaging chip unit volume to ST Introduces 64-Channel High-Voltage
Medical Imaging reach 73m Switch IC
2021-06-21 2016-09-13 2019-11-13
In "Market News" In "Vital Signs" In "MOSFET"

Tags : Medical Tech Watch

Leave a comment

You must Register or Login to post a comment.

Subscribe to
Newsletter

email address

Subscribe

Latest Posts Most Popular

2021-10-21
Asia and
Middle East EV
Sales Up 177%
in Jan-Aug
2021

2021-10-21
SiC Technology
Powers the
Shift to an All-
Electric Future

2021-10-21
GaN Emerges
for Next-Gen
Power
Electronics

2021-10-21
Aurora
Accelerates Its
AV
Deployment
Roadmap

2021-10-20
Keysight
Acquires
Scalable
Network
Technologies

LATEST POSTS

ELECTRIC SIC
VEHICLES 2021-10-21
SiC Technology
2021-10-21
Powers the
Asia and Shift to an All-
Middle East EV Electric Future
Sales Up 177%
in Jan-Aug 2021

MOST POPULAR

CONSUMER BIOTECHNOLOGY
ELECTRONICS 2020-10-21
The Biological
2020-10-30 Nutrient Cycle for
The Pandemic Electronic Industry
Pushed ‘Prime
Day’ Into
October

ASPENCORE NETWORK

PRODUCTS: NEWS & ANALYSIS: DESIGN: TOOLS: GLOBAL NETWORK FOR ADVERTISERS
EE Times Asia Contact Sales
Electronic Products EE Times EDN EEWEB
EE Times China Media Guide Request
Datasheets.com EE Times Europe Electronics-Tutorials PartSim
EE Times India
EEM.com Power Electronic News Planet Analog PCBWeb
EE Times Taiwan
TechOnline EBN Embedded Product Advisor
EE Times Japan
Elektroda.pl Embedded Know How Schematics.com
EDN Asia
Electronics Know How Schematics.io
EDN Taiwan
Embedded Central Europe Engage
ESM China
Embedded News
EDN China
IOT Design Zone
EDN Japan
Motor Control Design

All contents are Copyright © 2021 by AspenCore, Inc. All Rights Reserved. About Us Privacy Policy Terms of Use

You might also like