Professional Documents
Culture Documents
Recommended
Moisture Methods
Precise
Practical
Fast
Food Industry
Moisture Methods
2 METTLER TOLEDO Method Collection for Food Industry
Content
There are several types of methods available for analyzing the moisture content of foodstuffs, one of them being
thermogravimetric measurements, such as oven drying, halogen/IR drying and microwave drying.
Detailed information
on how to prepare
your sample to ensure
meaningful analysis
The measurement performance of the HMA is illustrated with a diagram for each substance, showing the
corresponding drying curve. The drying curve shows the progress of the moisture content determination over time,
providing useful insight into the drying dynamics of a given substance.
Drying curve over time, slope
shows the speed of LOD and
Exemplary Drying Curve
defines the switch-off criterion
Automatic calculation and
display of the chosen unit
The mean and the repeatability (SD, n=6) of the results using both techniques are displayed in an easy-to-read manner.
Moisture e-Learning
Moisture Guide
} www.mt.com/moisture-elearning
} www.mt.com/moisture-guide
Description of sample
Powder
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 125 °C 103 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 2g 5g
Sample preparation Stir sample, use spatula to put Dry weighing container with cover
and procedure the sample on the pan then in oven (103 °C, 1 h), allow to cool
distribute it evenly by gently in desiccator, weigh. Stir sample,
shaking the pan add to container, weigh.
Dry in oven (5 h), allow to cool down to
room temperature in desiccator, weigh
3.50
3.00
2.50
2.00
%MC
1.50
1.00
0.50
0.00
0 60 120 180
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 3.34 %MC 3.29 %MC
Standard deviation (SD) 0.07 0.09
Measurement duration (mean) 3 min 300 min
Description of sample
Milk-biscuit dough
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 135 °C 103 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 7g 5g
Sample preparation Press sample in batter spreader Dry weighing container with cover in oven
and procedure into a circular, thin sheet. Place (103 °C, 1 h), allow to cool in desiccator,
as one on sample pan weigh. Knead sample to homogenous
mixture. Add to container, weigh.
Dry in oven (4 h), allow to cool down to
room temperature in desiccator, weigh.
Dry sample one more time in oven (1 h,),
cool in desiccator, weigh
25.00
20.00
15.00
%MC
10.00
5.00
0.00
0 180 360 540 720 900 1080 1260 1440 1620 1800
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 19.81 %MC 19.73 %MC
Standard deviation (SD) 0.09 0.24
Measurement duration (mean) 26 min 240 min
Description of sample
Dry biscuit
Reference method
Oven
Method parameters:
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 120 °C 103 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 3g 5g
Sample preparation Grind sample in mortar, use Dry weighing container with cover in oven
and procedure spatula to put the sample on the (103 °C, 1 h), allow to cool in desiccator,
pan then distribute it evenly by weigh. Grind sample in mortar, stir well,
gently shaking the pan add to container, weigh.
Dry in oven (4 h), allow to cool down to
room temperature in desiccator, weigh.
Dry sample again in oven (1 h), cool in
desiccator, weigh
2.00
1.50
%MC
1.00
0.50
0.00
0 60 120 180 240 300
Time [s]
Measurement results:
Halogen Moisture Analyzer Drying Oven
Result (mean) 2.32 %MC 2.30 %MC
Standard deviation (SD) 0.04 0.32
Measurement duration (mean) 4 min 270 min
Description of sample
Butter
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 110 °C 102 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 3g 2g
Sample preparation Allow sample to come to room Dry metal pan (approx. 5 cm) with
and procedure temperature, distribute evenly on cover in the oven (103 °C, 1 h), allow
the sample pan to cool in desiccator, weigh. Allow
sample to come to room temperature,
stir well, add to pan, weigh.
Dry in oven (4 h), allow to cool down to
room temperature in desiccator, weigh
16.00
14.00
12.00
10.00
%MC
8.00
6.00
4.00
2.00
0.00
0 60 120 180 240 300 360 420 480 540 600
Time [s]
Description of sample
Potato chips, salted and flavored with paprika
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 122 °C 102 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 5g 5g
Sample preparation Grind sample in mortar, use Dry weighing container with cover in
and procedure spatula to distribute evenly on oven (102 °C, 1 h), allow to cool in des-
the sample pan iccator, weigh. Grind sample in mortar,
stir well, add to container, weigh.
Dry in oven (4 h), allow to cool down to
room temperature in desiccator, weigh
1.60
1.40
1.20
1.00
0.80
0.60
0.40
0.20
0.00
0 60 120 180 240 300 360 420 480 540 600
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 1.38 %MC 1.42 %MC
Standard deviation (SD) 0.04 0.01
Measurement duration (mean) 9 min 240 min
Description of sample
Powder
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 100 °C 102 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 2g 5g
Sample preparation Stir sample, use spatula to put Dry weighing container with cover in
and procedure the sample on the pan then oven (102 °C, 1 h), allow to cool in
distribute it evenly by gently desiccator, weigh. Stir sample, add to
shaking the pan container, weigh.
Dry in oven (4 h), allow to cool down to
room temperature in desiccator, weigh
2.50
2.00
1.50
%MC
1.00
0.50
0.00
0 60 120 180 240 300
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 3.05 %MC 3.02 %MC
Standard deviation (SD) 0.05 <0.01
Measurement duration (mean) 4 min 240 min
Description of sample
Powder of ground coffee beans
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 120 °C 103 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 3g 5g
Sample preparation Stir sample, use spatula to dis- Dry weighing container with cover in
and procedure tribute evenly on the sample pan oven (103 °C, 1 h), allow to cool in
desiccator, weigh. Stir sample, add to
container, weigh.
Dry in oven (5 h), allow to cool down to
room temperature in desiccator, weigh.
Dry sample again in oven (1 h), cool in
desiccator, weigh
5.00
4.00
3.00
%MC
2.00
1.00
0.00
0 60 120 180 240 300
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 4.42 %MC 4.46 %MC
Standard deviation (SD) 0.03 0.01
Measurement duration (mean) 5 min 360 min
Description of sample
Viscous liquid, non-sweetened
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 140 °C 102 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 2g 6g
Sample preparation Stir sample, apply evenly in Dry metal pan with 20 g of sand, glass
and procedure between two glass fiber filters stirrer and cover in the oven (102 °C,
using a pipette 1 h), allow to cool in desiccator, weigh.
Stir sample, add to pan, weigh, allow to
evaporate in oven for 30 min at 90 °C,
add 5 ml of water, mix.
Dry in oven (2 h), allow to cool down to
room temperature in desiccator, weigh.
Dry sample two more times in oven (1 h,
each time), cool in desiccator, weigh
80.00
60.00
%MC
40.00
20.00
0.00
0 60 120 180 240 300 360 420 480
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 74.60 %MC 74.56 %MC
Standard deviation (SD) 0.17 0.08
Measurement duration (mean) 8 min 240 min
Description of sample
Granulate
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 160 °C 130 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 5g 5.5 g
Sample preparation Stir sample, use spatula to put Dry weighing container with cover in
and procedure the sample on the pan then dis- oven (130 °C, 1 h), allow to cool in
tribute it evenly by gently shaking desiccator, weigh. Stir sample well, add
the pan to container, weigh.
Dry in oven (1.5 h), allow to cool down to
room temperature in desiccator, weigh
16.00
14.00
12.00
10.00
%MC
8.00
6.00
4.00
2.00
0.00
0 120 240 360 480 600 720 840
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 14.54 %MC 14.56 %MC
Standard deviation (SD) 0.03 <0.01
Measurement duration (mean) 13 min 90 min
Description of sample
Hard cheese with >35% fat
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 110 °C 102 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 3g 2g
Sample preparation Grate sample, stir, use spatula Dry weighing container with 10 g of sand,
and procedure to distribute evenly on a glass glass stirrer and cover in the oven (102 °C,
fiber filter 1 h), weigh. Grate sample, stir well, add to
container, weigh, grind with sand.
Dry in oven (4 h), allow to cool down to
room temperature in desiccator, weigh
35.00
30.00
25.00
20.00
%MC
15.00
10.00
5.00
0.00
0 240 480 720 960 1200 1440 1680
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 31.48 %MC 31.99 %MC
Standard deviation (SD) 0.18 0.03
Measurement duration (mean) 26 min 240 min
Description of sample
Loose tea
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 125 °C 103 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 2.5 g 5g
Sample preparation Stir sample, use spatula to dis- Dry weighing container with cover in
and procedure tribute evenly on the sample pan oven (102 °C, 1 h), allow to cool in
desiccator, weigh. Stir sample, add to
container, weigh. Dry in oven (16 h),
allow to cool down to room temperature
in desiccator, weigh
8.00
7.00
6.00
5.00
%MC
4.00
3.00
2.00
1.00
0.00
0 60 120 180 240 300 360 420 480 540
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 7.01 %MC 7.09 %MC
Standard deviation (SD) 0.08 0.05
Measurement duration (mean) 8 min 960 min
Description of sample
Ground hazelnuts, brownish powder
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 130 °C 103 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 4g 5g
Sample preparation Stir sample, use spatula to put Dry weighing container with cover in oven
and procedure the sample on the pan then (103°C, 1h), allow to cool in desiccator,
distribute it evenly by gently weigh. Stir sample, add to container, weigh.
shaking the pan Dry in oven (2 h), allow to cool down to
room temperature in desiccator, weigh. Dry
sample two more times in oven (1 h, each
time), cool in desiccator, weigh
6.00
5.00
4.00
%MC
3.00
2.00
1.00
0.00
0 60 120 180 240 300 360
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 5.27 %MC 5.27 %MC
Standard deviation (SD) 0.08 0.07
Measurement duration (mean) 6 min 240 min
Description of sample
Ketchup
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 125 °C 102 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 4g 2g
Sample preparation Stir sample, use spatula to distrib- Dry metal pan with 25 g sand, glass stirrer
and procedure ute evenly on a glass fiber filter and cover in the oven (102 °C, 1 h), allow
to cool in desiccator, weigh. Stir sample,
add to pan, weigh, mix with sand.
Dry in oven (4 h), allow to cool down to
room temperature in desiccator, weigh
80.00
60.00
%MC
40.00
20.00
0.00
0 240 480 720 960 1200 1440 1680
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 69.40 %MC 69.29 %MC
Standard deviation (SD) 0.06 0.06
Measurement duration (mean) 24 min 240 min
Description of sample
Granulate
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 145 °C 102 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 6g 5g
Sample preparation Stir sample, use spatula to put Dry weighing container with cover in
and procedure the sample on the pan then oven (103 °C, 1 h), allow to cool in
distribute it evenly by gently desiccator, weigh. Stir sample, add to
shaking the pan container, weigh.
Dry in oven (2 h), allow to cool down to
room temperature in desiccator, weigh.
Dry sample three more times in oven (1 h,
each time), cool in desiccator, weigh
10.00
8.00
6.00
%MC
4.00
2.00
0.00
0 60 120 180 240 300 360 420 480
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 9.15 %MC 9.01 %MC
Standard deviation (SD) 0.09 0.14
Measurement duration (mean) 7 min 300 min
Description of sample
Powder of milk with 2.7% milk fat
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 110 °C 102 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 3g 2g
Sample preparation Stir sample, use spatula to put Dry weighing container with cover in
and procedure the sample on the pan then oven (102 °C, 1 h), allow to cool in
distribute it evenly by gently desiccator, weigh. Stir sample, add to
shaking the pan container, weigh. Dry in oven (2 h),
allow to cool down to room temperature
in desiccator, weigh. Dry sample two
more times in oven (1 h, each time),
cool in desiccator, weigh
5.00
4.00
3.00
%MC
2.00
1.00
0.00
0 60 120 180 240 300 360 420 480
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 4.61 %MC 4.67 %MC
Standard deviation (SD) 0.10 0.02
Measurement duration (mean) 7 min 240 min
Description of sample
Flakes
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 150 °C 131 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 4g 5g
Sample preparation Grind sample, use spatula to Dry weighing container with cover in oven
and procedure put the sample on the pan then (131 °C, 1 h), allow to cool in desiccator,
distribute it evenly by gently weigh. Grind sample, stir well, add to
shaking the pan container, weigh.
Dry in oven (1.5 h), allow to cool down to
room temperature in desiccator, weigh
12.00
10.00
8.00
%MC
6.00
4.00
2.00
0.00
0 60 120 180 240 300 360 420
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 10.96 %MC 10.91 %MC
Standard deviation (SD) 0.04 0.01
Measurement duration (mean) 6 min 90 min
Description of sample
Made of 100% oranges, from concentrate
Reference method
Oven
Method parameters
Rapid Not applicable
Drying program 140 °C 103 °C
Drying temperature 3 (1 mg/50 s) Not applicable
Switch-off criterion 2g 10 g
Sample weight Stir sample, apply evenly to Dry metal pan (approx. 8 cm) with 10 g
glass fiber filter using a pipette sand, glass stirrer and cover in the oven
(103 °C), allow to cool in desiccator,
weigh. Stir sample, add to pan using a
pipette, weigh, mix with sand, evaporate
in water bath for 30 min.
Dry in oven (3 h), allow to cool down to
room temperature in desiccator, weigh
100.00
80.00
60.00
%MC
40.00
20.00
0.00
0 60 120 180 240 300 360 420
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 89.40 %MC 89.23 %MC
Standard deviation (SD) 0.04 0.06
Measurement duration (mean) 6 min 180 min
Description of sample
Dried pasta made of durum wheat
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program 145 °C 130 °C
Drying temperature 3 (1 mg/50 s) Not applicable
Switch-off criterion 4g 5.5 g
Sample weight Grind sample, use spatula to Dry weighing container with cover in oven
put the sample on the pan then (131 °C, 1 h),
distribute it evenly by gently allow to cool in desiccator, weigh. Grind
shaking the pan sample, stir well, add to container, weigh.
Dry in oven (1.5 h), allow to cool down to
room temperature in desiccator, weigh
12.00
10.00
8.00
%MC
6.00
4.00
2.00
0.00
0 120 240 360 480 600 720 840 960
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 10.90 %MC 10.90 %MC
Standard deviation (SD) 0.10 0.03
Measurement duration (mean) 14 min 90 min
Description of sample
Granulate, fine-grained
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 135 °C 102 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 2g 5g
Sample preparation Stir sample, use spatula to put Dry weighing container with cover in
and procedure the sample on the pan then oven (102 °C, 1 h), allow to cool in
distribute it evenly by gently desiccator, weigh. Stir sample, add to
shaking the pan container, weigh. Dry in oven (2 h),
allow to cool down to room temperature
in desiccator, weigh. Dry sample two
more times in oven (1 h, each time),
cool in desiccator, weigh
12.00
10.00
8.00
%MC
6.00
4.00
2.00
0.00
0 60 120 180 240 300
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 10.82 %MC 10.70 %MC
Standard deviation (SD) 0.12 0.05
Measurement duration (mean) 5 min 240 min
Description of sample
Liquid, homogeneous
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Rapid Not applicable
Drying temperature 130 °C 103 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 3g 4g
Sample preparation Stir sample, apply evenly to Dry metal pan (approx. 8 cm) with 10 g
and procedure glass fiber filter using a pipette sand, glass stirrer and cover in the oven
(103 °C, 30 min), allow to cool in des-
iccator, weigh. Stir sample, add to pan
using a pipette, weigh, mix with sand,
evaporate in water bath for 30 min.
Dry in oven (3 h), allow to cool down to
room temperature in desiccator, weigh
100.00
80.00
60.00
%MC
40.00
20.00
0.00
0 60 120 180 240 300 360 420 480 540 600
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 95.22 %MC 95.32 %MC
Standard deviation (SD) 0.05 0.04
Measurement duration (mean) 8 min 180 min
Description of sample
Powder
Reference method
Oven
Method parameters
Halogen Moisture Analyzer Drying Oven
Drying program Standard Not applicable
Drying temperature 140 °C 130 °C
Switch-off criterion 3 (1 mg/50 s) Not applicable
Sample weight 4g 5g
Sample preparation Stir sample, use spatula to dis- Dry weighing container with cover in
and procedure tribute evenly on the sample pan oven (130 °C, 1 h), allow to cool in
desiccator, weigh. Stir sample well, add
to container, weigh.
Dry in oven (2 h), allow to cool down to
room temperature in desiccator, weigh
12.00
10.00
8.00
%MC
6.00
4.00
2.00
0.00
0 60 120 180 240 300 360 420 480
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 11.33 %MC 11.60 %MC
Standard deviation (SD) 0.06 0.09
Measurement duration (mean) 7 min 120 min
Description of sample
Reference method
Oven
100.00
80.00
60.00
%MC
40.00
20.00
0.00
0 120 240 360 480 600 720 840 960
Time [s]
Measurement results
Halogen Moisture Analyzer Drying Oven
Result (mean) 87.41 %MC 87.50 %MC
Standard deviation (SD) 0.10 0.10
Measurement duration (mean) 14 min 180 min
White Paper:
Moisture Analyzer versus drying oven
The White Paper "Drying Oven vs. Halogen
Moisture Analyzer (HMA)" explains scientifi-
cally how the drying oven method can easily
be replaced by rapid, easy-to-use halogen
moisture analysis.
} www.mt.com/moisture-vs-oven
www.mt.com/moisture-routine-testing
For more information
Mettler-Toledo GmbH
Laboratory and Weighing Technologies
CH-8606 Greifensee, Switzerland
Tel. +41 44 944 22 11
Fax +41 44 944 30 60