PRODUCT
APPLICATION
Sterilizer of the DGM AND type with
rectangular prism-shaped, horizontally
positioned sterilizing chamber, is designed for
sterilizing with water steam such medical items
as: surgical instruments, operating clothes,
dressing materials, infusion liquids, rubber
products, culture media etc.
MODEL SELECTION
= Standard O = Optional
MODEL & DIMENSIONS
Model Chamber Size Chamber Dimensions
wxhxd Volume wxhxd
mm ltr mm
100-2 400 x400 x650 103
800 x 1615 x 900
100-1 400 x400 x700 109
150-2 400 x400 x950 150
800 x 1615 x 1200
150-1 400 x400 x1000 156
DOOR SELECTION
Single Door (-1)
Double Door (-2)
VERSIONS
CONTROL SYSTEM
The work of the sterilizer and its steam generator is
controlled by Siemens S7-200 controller.
All the activities such as the following: supplying
steam into the chamber, maintaining constant
temperature during the exposure phase, course of
the cycle, opening and closing the door etc.; and for
the version with the built-in steam generator:
automatic filling of the steam generator boiler with
water, inspection of heating elements to ensure a
constant steam pressure, are monitored on an
ongoing basis by the monitoring system. The
controller stores the following sterilization and test
programs: Bowie-Dick test and chamber Leak test. "Graph" screen displays the chamber pressure
and temperature as a graph. The graph allows
Cycle Performance Documentation the operator to monitor the process in real time.
The DGM AND series sterilizer provides two
standard means for cycle documentation:
• Paper printer
• USB flash memory drive
On the insertion side there is a printer that records
the sterilization cycle parameters.
• All the standard parameters of the cycle
are printed in a table.
• The printout includes the following data:
• temperature in the chamber;
• date and time of the cycle start;
• sterilization cycle number;
• operator number;
• serial number;
• set sterilization program parameters
(duration, temperature);
• name and start time of the process stage;
"Table" screen displays the course of operation of "Attention!" screen" displays the current system
the device. This menu displays the program settings messages. This screen appears regardless of
and the degree of its completion. the screen viewed by the user. All messages
displayed in this screen are accompanied by a
sound signal.
CYCLE PROGRESSION
• Venting
During that phase multiple pressure changes are
applied in a pre-programmed fashion. It causes
effective removing of the air from various forms of
load. Initial heating of load is performed.
Venting phase consists of four repeating cycles.
Every cycle starts when required pressure is
"Manual control" screen is used for visual obtained, followed by pulsatory steam introduction
supervision of the control system. It allows and then by full steam introduction.
the operator to monitor the process (completion of
the respective stages) and the operation of • Heating
the device components (valves, pumps, level The purpose of this phase to heat load up to a
sensors). preset exposure temperature. This is accomplished
The software of the sterilizer allows the operator to by filling the chamber with steam to the required
carry out a number of operations in the manual pressure. The pressure depends on a selected
mode. This is particularly convenient in emergency temperature. The duration of this phase depends
situations (i.e. when it is necessary open the door of mainly on the size of load put into a sterilization
the pressurised chamber, etc.), as well as for testing chamber.
and servicing. • Exposure
After equilibration of a temperature selected inside
load, countdown of the exposure time starts.
Condensates forming on load, chamber walls and
doors are removed from the chamber by a draining
installation. Temperature during all the exposure
phase should not be lower than a preset one and
should not be higher more than by 3°C. Duration of
this phase depends on a program selected.
• Decreasing
After the exposure phase is finished the pressure
inside the chamber is lowered to the level required
for a drying phase. Large pressure differential –
between the end of exposure and beginning of
drying – could cause in this phase damage to the
packs (paper-foil packaging) and therefore it is
extended in time.
DGM TOR System (optional)
• Drying
The purpose of a drying phase is to equilibrate the
temperature and humidity of the sterilized load. All System for monitoring and documenting processes
kinds of materials are exposed to high vacuum for (track & tracing) of supplies movement and
some time. After attaining the required pressure processing.
level countdown of the drying time is commenced
automatically – depending on a selected program. • Automation of monitoring and documenting
• Aeration the processes of infection control;
Aeration by means of a sterilizing filter is the • Electronic document management;
subsequent stage which completes the sterilization • Automatic reporting;
cycle. The pressure is then brought from a negative • Flexibility in customizing for a particular
pressure to the atmospheric one. customer;
• Availability of process data transfer from
ENVIRONMENTAL IMPACT the installation to the PC;
• Broad opportunities for data-processing
The sterilizer is composed of 94 % stainless steel, 4 (movement of supplies, as well as the staff
% of non-ferrous metals and 2 % of other materials, and documents).
which – after dismantling the unit by an authorised
person – may be recycled.
A global network of skilled service specialists can Baarerstrasse 8, 6th floor, Postfach 623
provide periodic inspections and adjustments to 6301 Zug, Switzerland
help assure low-cost peak performance. A detailed
service and maintenance plan is included in the Phone + 31 6 48 555 464,
user manual. Fax +41 41 766 77 01
e-mail: trade@dgmmachinery.com
www.dgmmachinery.com
Options
Accessories