(a) Select 'Abdomen' by clicking on the abdomen (<1>) in the figure below and
select the appropriate eXam Plan for the exam by clicking it (<2>) with the
mouse.
<2>
<1>
Figure 5.4-30
following items:
Figure 5.4-31
No. 2B201-208E*K
(c) Detailed eXam Plan information is displayed. Confirm the
scanning conditions.
(a) The start scan key ( key. ) on the keyboard flashes. press the
No. 2B201-208E*K
5-45
Figure 5.4-33
Figure 5.4-34
Select [[Group]]. Cuts from group1 are selected. Adjust the scanning
(cutting) position to group 1.
• To move the cut position, drag the solid line section displayed on the
scanogram.
• To change the scan angle, drag the arrow to the bottom edge
of the solid line section displayed in the scanogram and tilt the
solid line section until you get the desired angle.
No. 2B201-208E*K
5-46
• To change the setting for each cut in the same group, click [[SLICE]].
Figure 5.4-35
Click [[+]]. The next group is selected as the adjustment target. Make
adjustments for the selected group in the same way as in step (c)
above. To return to the previous group, click [[-]].
When the plan adjustment is completed for each group, confirm that
patient safety is assured and then click . THE
following window is displayed. When [[OK]] is clicked in the displayed
window, the table top moves to the first scan position set using the
scan planning program and the
scanning starts.
Figure 5.4-36
No. 2B201-208E*K
5-47
(a) If the table top moves more than 600 mm, the window shown in figure
5.4-36 is displayed again and the table top stops. In this case, confirm
that the patient's safety is assured and then click [[OK]]. The table top
moves to the first cut position.
(b) When the gantry is tilted, the following window is displayed. Hold
down the key on the keyboard while confirming that security
of the patient is assured. The gantry tilts until it reaches the target
angle.
Figure 5.4-37
(c) When table top movement and gantry tilt are completed, the start
scan key on the keyboard lights up. Press the lit key.
If the table top has moved more than 200 mm or if gantry tilt is
performed, a corresponding confirmation window is displayed. In this
case, confirm that the patient's safety is assured before proceeding
with the next procedure. After that, repeat this operation as many
times as necessary.
No. 2B201-208E*K
5-48
(b) Press the 'auto home' key on the operation panel. the gantry
returns to the 0° position, the table top moves to the rear edge of
the patient table and the patient table lowers to its original position.
Figure 5.4-38
No. 2B201-208E*K
5-49
For information on the protocol used to connect a laser (or analog) camera
to , refer to section 7.
Click the print print icon on in the image display area. the window of
film is displayed.
: print key
(Performs printing on the
film.)
: delete key
(delete data
stored.)
: store key
(Data stored.)
NOTE: The area enclosed within the whiteboard is transferred (and printed) onto the
film. Before printing on film, be sure of the area that will be printed on film.
No. 2B201-208E*K
5-50
NOTE: Film printing can be done directly from the image matrix window.
No. 2B201-208E*K
5-51
(1) When the window level and window width are adjusted.
Start
click the [[Get Image]] icon The key can also be used to
or the [[Store]] icon. store image
Do you want
YES
cancel the process of
storage?
Select the frame to be
NOT deleted and click The key also
[[Delete]]. can be used to cancel the
storage process.
YES
Click [[Filming]] The key can be used to
print film.
Click [[Close]].
Figure 5.5-4
No. 2B201 208
5-52
espcearançeãom, evnotlou.me de
(3) When window level and window width have already been adjusted
No. 2B201 208E K
5-53
Start
The end
Figure 5.5-6
No. 2B201 208E K
5-54
Since network data transfer, optical magnetic disk archiving, hard disk data
deletion, etc. are performed in parallel, if the delete image operation is
performed after the network transfer request is formulated, delete image
processing may have priority. Therefore, it is recommended that all datasets
be stored on the day the dataset is acquired and that deletion processing is
done first when the system is used the next day.
No. 2B201 208E K
5-55
<Operating procedures>
(a) Click the icon in the start operation window. in the menu window
displayed, select [[Image data utility]]. The file utility window is displayed.
(d) The ON/OFF selection window is displayed. Click [[OK]] after selecting
[[ON]] or [[OFF]]. When [[ON]] is selected, "P" is displayed in the
protection signal column for the selected file. When [[OFF]] is selected, "P"
displayed disappears from the protection signal column for the file.
NOTE: When data processing is performed in exam mode, all files for the
selected exams are processed at the same time. of this
so the files in the exam are all protected or they are all unprotected. If
protection or protection release is to be performed by file, perform the
necessary operations for each file separately.
No. 2B201-208E K
5-56
After scanning, immediately carry out the necessary processing (printing the
data, storing data to the storage medium such as magnetic-optical disk, etc.)
and then discarding the scanned data. If the scanned data should not be
deleted immediately because image processing etc. will be performed, it is
recommended that the data
are protected.
- Refer
to subsection 5.6.2 "Data protection (file
protection)" on p. 5-61 of this manual.
NOTE: If scanned data is not regularly deleted. If the free area on the system's
hard disk gradually becomes smaller, scanning becomes impossible.
<Operating procedures>
(a) Click the utility icon in the start operation window. In the window
menu displayed, select [[Image data utility]]. The file utility window
is displayed.
(b) Select the study (exam), series, or image files to be deleted by highlighting
them.
(c) Click [[Delete]] at the bottom of the file utility window.
(d) When [[OK]] is clicked on the delete confirmation window shown in the figure
below, the exam data is cleared. Clicking [[Cancel]] ends processing without
deleting the exam displayed in the window.
Study list display of image data on hard disk, selection, and loading are
performed using this window.
To end display image processing, click [[Quit]] at the bottom right of the
window.
<Operation procedure>
(b) Click [[OK]] in the lower right corner of File Selection window. The image of
the selected study is displayed in the image display area.
List of
When an ROI (region of interest) is adjusted on the displayed image, the area, mean
TC number, standard deviation, maximum TC number, and minimum TC number in
the ROI can be calculated.
The ROI types that can be adjusted are Linear ROI (straight line), Rectangular
ROI, Circular ROI, Ellipse ROI, Line ROI, Polygonal ROI, Free ROI, and Angular
ROI.
Rectangular
Circular
Ellipse
Line
Polygonal
Free
Angular
(a) When the graphic editor display icon on the right side of the