MRI Suite Shielding Requirements


General
Magnetic Resonance images. It is very important in image artifacts, which
Imaging (MRI) systems use to prevent stray RF waves could lead to limited
radio frequency (RF) waves from entering the MRI suite. diagnostic value and image
to help construct soft tissue Failure to do so could result retakes.

RF Shielding
Radio Frequency (RF) when addressing the need The shielding effectiveness
shielding works to prevent to connect or penetrate of an RF shield is measured
exterior RFI (Radio certain necessary in dB’s. This unit-less
Frequency Interference) applications: the RF door to measurement uses the
from negatively affecting the gain entry, an RF window to same logarithmic scale used
operation of the MRI and the see inside, an RF on a seismic Richter scale.
digital image being Waveguide Air Vent Hence, an RF shield that
produced. MRI machines (WGAV) penetration for performs at 70 dB, versus
have highly sensitive HVAC requirements, and an one that is working at 80 dB,
sensors that pick up minute electrical filter for electrical or 90 dB, is 10 times or 100
RF signals. These sensors wiring. times better respectively.
can be overridden by
external RF noise generated Any material that conducts All MRI scanners operate
by a variety of electronic electrical current can be between 5 MHz and 128
applications, thus the need used as an RF shielded MHz, which is the frequency
for RF shielding. enclosure. Galvanized steel at which the sensor picks up
and copper have proven the minute RF signal.
An RF shield encompasses over the course of time to Although RF enclosures
all six sides of a room. The work better, are cost shield at a variety of
walls, floor and ceiling are effective, and provide the frequencies, the sensor
all connected and all create design flexibility necessary frequency is used in testing
an RF Shield. The design for today’s architectural to insure that the RF
and construction challenge applications. enclosure is performing for
with RF enclosures occurs that particular MRI.
Magnetic Shielding
As previously indicated, MRI magnetic shielding varies greater influence on the
machines are available in a per magnet and per site, magnet’s function. It is vital
variety of sizes and field depending upon these two to proper imaging that the
strengths. The field strength criteria. As the field strength magnet’s homogeneity does
is the strength of the magnet increases, the size of the not vary as it is tuned to
at its isocenter, or the exact magnetic field grows; hence, PPM, or parts per million.
center of the magnetic field a higher field magnet (1.5T Hence, a large moving steel
measured in Tesla. As the or 3.0T) will have a greater mass with a magnetic field
magnet’s field and site influence on its surrounding can interfere with the
location varies, the area. As the field strength magnet’s ability to image, as
requirement for magnetic decreases (0.2T – 1.0T) the it can distort the magnet’s
shielding varies. The surrounding area has a uniformed magnetic field.