Professional Documents
Culture Documents
Tune Report: Count Check Count Unit Category Mass
Tune Report: Count Check Count Unit Category Mass
[No Gas]
Integration Time [sec] 0.1 Acquisition Time [sec] 22.74 Y Axis Linear
Tune Parameters
## Plasma Paramters ##
ParameterName Value Unit ParameterName Value Unit
RF Power 1550 W Nebulizer Pump 0.10 rps
RF Matching 1.70 V S/C Temp 2 ℃
Smpl Depth 8.0 mm Gas Switch Makeup Gas
Carrier Gas 1.05 L/min Makeup/Dilution Gas 0.00 L/min
Option Gas 0.0 %
## Lenses Parameters ##
ParameterName Value Unit ParameterName Value Unit
Extract 1 0.0 V Cell Entrance -40 V
Extract 2 -160.0 V Cell Exit -60 V
Omega Bias -70 V Deflect 17.2 V
Omega Lens 10.8 V Plate Bias -50 V
## Cell Parameters ##
ParameterName Value Unit ParameterName Value Unit
Use Gas false OctP Bias -8.0 V
He Flow 0.0 mL/min OctP RF 170 V
H2 Flow 0.0 mL/min Energy Discrimination 5.0 V
3rd Gas Flow 0 %
[He]
Mass Range Count Response Response Response
(Actual) (Actual) (Required) (Flag)
[cps/ug/l] [cps/ug/l]
59 1000 775 7752.45 5000.00
89 1000 662 6620.38 5000.00
205 5000 3820 38204.12 5000.00
Tune Parameters
## Plasma Paramters ##
ParameterName Value Unit ParameterName Value Unit
RF Power 1550 W Nebulizer Pump 0.10 rps
RF Matching 1.70 V S/C Temp 2 ℃
Smpl Depth 8.0 mm Gas Switch Makeup Gas
Carrier Gas 1.05 L/min Makeup/Dilution Gas 0.00 L/min
Option Gas 0.0 %
## Lenses Parameters ##
ParameterName Value Unit ParameterName Value Unit
Extract 1 0.0 V Cell Entrance -28 V
Extract 2 -170.0 V Cell Exit -70 V
Omega Bias -80 V Deflect 3.4 V
Omega Lens 11.5 V Plate Bias -60 V
## Cell Parameters ##
ParameterName Value Unit ParameterName Value Unit
Use Gas true OctP Bias -18.0 V
He Flow 5.0 mL/min OctP RF 200 V
H2 Flow 0.0 mL/min Energy Discrimination 5.0 V
3rd Gas Flow 0 %