You are on page 1of 4

Customer Service Information

Department: Subject: Date:


AFTER SALES TECHNICAL BULLETIN P-079 14.07.2008

Product: Originator: Issued by:


Planmeca Proline XC Timo Müller Mark Niemi

SOFTWARE VERSIONS AND


COMPABILITY INFORMATION OF
PLANMECA PROLINE XC
FILM X-RAY UNIT
CAUTION
The CPU PCB and software chips are sensitive to static electric discharges.

The recommended software combinations are presented in Table 1 and 5.

The spare parts order codes for chips with latest available software versions are:
Generator Processor PCB: 10011479
Keyboard Processor PCB: 10011478

NOTE! PK, PG and GUI software should be updated at the same time
NOTE! See also PLANMECA PROLINE XC X-RAY UNIT GUI SOFTWARE 1.4.3.0.R

TABLE 1: RECOMMENDED SOFTWARE COMBINATIONS

KEYBOARD GENERATOR GUI


Planmeca Proline XC film X-ray unit
PROCESSOR PROCESSOR
Keyboard Processor PCB (code 10009419) and
PK 6.61 PG 6.53 1.4.3
Generator Processor PCB (code 10009814)

PLANMECA Oy Phone: +358-20-7795 500


Asentajankatu 6 Fax: +358-20-7795 555
1
FIN-00880 Helsinki, Finland E-mail: aftersales@planmeca.com
Customer Service Information

TABLE 2: THE LATEST VERSION OF THE KEYBOARD PROCESSOR SOFTWARE (PK)

SOFTWARE DATE OF ADDITION AND MODIFICATION


VERSION RELEASE
PK 6.61 09.07.2008 • Fixed Parameter History
• Fixed DAP values
• Exposure quick settings updated
PK 6.60 31.5.2007 • Ceph ON/OFF selection in service mode corrected so that the
selection can be made easier when error 24 appears.
• Patient positioning lights are activated for 2 minutes (before it was
0,5 min)
• Intermittent lock down problems when going Ceph fixed
PK 6.56 28.12.2005 • Removed 10 minutes timeout on leaving Ceph mode.
• NOTE: Requires PG6.53
PK 6.55 26.9.2005 • Use of child slot of primary collimator in panoramic mode of a
Pan/Ceph X-ray unit.
• Added True Profile TMJ imaging mode (requires a True Profile
TMJ activation key and GUI SW 1.1.0)
• ER 60 internal diagnostics improved
• NOTE: Requires PG6.53
PK 6.54 1.6.2005 • The exposure switch type can no longer be chosen in the Service
Mode i.e. the green Ready indicator is on, when the X-ray unit is
ready for the exposure.
PK 6.53 3.5.2005 • Supports the color GUI.
• Three different jaw sizes and three different shapes can be
selected.
• Segmented imaging can be selected from five different segments.

TABLE 3: THE LATEST VERSION OF THE GENERATOR PROCESSOR SOFTWARE (PG)

SOFTWARE DATE OF ADDITION AND MODIFICATION


VERSION RELEASE
PG 6.53 26.9.2005 • Added error code Er85 (PSU PCF fault, only on PCB 105-10-28
rev. C or later)
• Added error code Er87 (PSU heat limit on, only on PCB 105-10-
28 rev. C or later)
PG 6.52 1.6.2005 • Including a new error code E85 (problems in generator PFC; low
voltage)
• Changes in AEC because of the jaw size selection
PG 6.51 22.11.2004 • Because of new type of Planmeca Proline EC and the EMC
regulations the PCBs have been modified, which requires new
software.
• To be used with PK 6.53.

PLANMECA Oy Phone: +358-20-7795 500


Asentajankatu 6 Fax: +358-20-7795 555
2
FIN-00880 Helsinki, Finland E-mail: aftersales@planmeca.com
Customer Service Information

PLANMECA Oy Phone: +358-20-7795 500


Asentajankatu 6 Fax: +358-20-7795 555
3
FIN-00880 Helsinki, Finland E-mail: aftersales@planmeca.com
Customer Service Information

PLANMECA PROLINE X-RAY UNIT: SOFTWARE AND PCB COMPABILITY INFORMATION AND ORDER CODES

Publication Nr. 10013274 GENERATOR PROCESSOR PCB (Software PG X.XX)


Updated Jul 14, 2008 Order Code 00665069 Order Code 00665180 Order Code 00665181 Order Code 10009814
PCB 105-10-03 PCB 105-10-03 PCB 105-10-03 with PCB 105-10-28
Version –B6 or older Version –C or later Adapter PCB 105-10-14
Version –C or later
Order Code 00665070 PM 2002 CC / PM 2002 Proline EC/CC: No compatible PCBs No compatible PCBs No compatible PCBs
PCB 105-10-06 - PK0 1.67, order code: 00471010
Version B2- or later - PK1 1.67, order code: 00471011
- PG1.69, order code: 00471088
KEYBOARD PROCESSOR PCB (Software PK X.XX)

Order Code 00665070 PM 2002 CC: PM 2002 CC: Proscan (with stepper motor No compatible PCBs
PCB 105-10-06 - PK2.51 order code: 00471087 - PK2.51, order code: 00471087 type ceph lock):
Version –D or later - PG1.69, order code: 00471088 - PG2.07, order code: 00471064 - PK4.54, order code: 00471031
PM 2002 Proline EC/CC: PM 2002 Proline EC/CC Film: - PG4.54, order code: 00471030
- PK3.51, order code: 00471087 - PK3.75, order code: 00471087 Proscan (with solenoid type
- PG1.69. order code: 00471088 - PG2.25, order code: 00471064 ceph lock):
PM 2002 Proline EC/CC Dimax1: - PK4.12, order code: 00471031
- PK3.48, order code: 00471087 - PG4.13, order code: 00471030
- PG3.12, order code: 00471064
Order Code 10002568 No compatible PCBs PM2002 / Proline/ EC/CC/ Film: No compatible PCBs No compatible PCBs
PCB 105-10-06 - PK6.00, order code: 10012715
Version –L or later - PG6.00, order code: 10012716
PM2002/ Proline/ EC/CC/Dimax
2/3:
- PK5.16, order code: 10004422
- PG5.15, order code: 10004421
Order Code 10009419 No compatible PCBs PM2002/ Proline /EC/CC/ Film: No compatible PCBs Proline XC/EC Film:
PCB 105-10-27 - PK6.00, order code: 10012715 - PK6.61, order code: 10011478
- PG6.00, order code: 10012716 - PG6.53, order code: 10011479
PM2002 / Proline /EC/CC/ Dimax Proline XC/EC Dimax 2/3:
2/3: - PK5.62, order code: 10010375
- PK5.16, order code: 10004422 - PG5.60, order code: 10010376
- PG5.15, order code: 10004421
• NOTE! Concerning codes 00471087 and 00471064. When ordering software chips for units equipped with keyboard processor PCB code 105-10-06 Version K or earlier, the required
software version must be given. If the required software is not known then the equipment must be described according to the above table. The spare parts department must be informed
about the X-ray unit model and type (film/digital), sensor type (dimax1/ dimax2 or 3), keyboard processor type and version, generator processor type and version.
• NOTE! Table shows all the possible PCB/ software/machine type combinations. Combinations that are not mentioned in the table are not possible. Planmeca does not guarantee a proper
operation if older software version than the one mentioned in the table is used.

PLANMECA Oy Phone: +358-20-7795 500


Asentajankatu 6 Fax: +358-20-7795 555
4
FIN-00880 Helsinki, Finland E-mail: aftersales@planmeca.com